Paramount Paramount Calibration Centre

Calibration Laboratory : L/12, GIDC, Opp. Fire Station, Odhav, Ahmedabad – 382415, Gujarat.
Ph : (079) 2287 1922, Mobile : 98985 69851 E-mail : paramountcalibration@yahoo.co.in
Calibration facilities for all types of Process, Quality and Laboratory related Instruments

Calibration Certificate
Name of Customer & Address : Alfa Pumps (P) Limited
58-59, Shreenathji Ind. Estate, Kujad-Bakrol Road, Daskroi, Ahmedabad

Certificate No. : Paramount/54380 Date of Issue : 12-11-2020


Date of Calibration : 12-11-2020 Suggested Due Date : 11-11-2021
Name of Instrument : Digital Vernier Caliper
Instrument No. : 0113312 Range : 0 – 300 mm
Least Count : 0.01 mm Make : Mitutoyo
Visual Inspection : Ok Initial Error : Nil
Gape of External Jaw : Within 4 micron Gape of Internal Jaw : Within 4 micron
Environment Condition : 20  2C / 40 to 60% Rh Clients Reference : Nil
Calibration Method No. : PCC/006 Reference IS No. : IS 3651
 Slip Gauge Box (87 pcs), Make : Aditya, Serial No. 150, Identification No. :
Standard Reference Used
2/1511/1, Grade : “0”.
Slip Gauge Box, calibrated by Tanson Instrument, (NABL Accredited Lab),
Traceability Certificate No. TI/C/SGS/033/19, dated 11th March 2019, valid up to 10th
March 2021.
The uncertainty of measurement is for a confidence probability of not less than 95%.
Results of Calibration
Standard Reading Observed by Digital Vernier Caliper in mm Absolute Error in mm
Dimension in mm External Internal External Internal
0.00 0.00 0.00 0.00 0.00
25.00 25.00 25.00 0.00 0.00
50.00 50.00 50.00 0.00 0.00
75.00 75.00 75.00 0.00 0.00
100.00 100.00 100.00 0.00 0.00
150.00 150.00 150.01 0.00 0.01
200.00 199.99 200.01 0.01 0.01
300.00 299.99 300.01 0.01 0.01

Our objective is Quality Calibration not only Calibration Certificate.


Note : This report refers only to particular item (s) submitted for calibration. Paramount Calibration Centre is not
responsible for any change in calibration data and performance specification on account of mal functioning of instruments
covered by this report. The calibration results reported are valid at the time of under stated conditions of the
measurements. This reports may not be reproduced in full / part without prior permission of Paramount CalibrationCentre
Paramount Paramount Calibration Centre
Calibration Laboratory : L/12, GIDC, Opp. Fire Station, Odhav, Ahmedabad – 382415, Gujarat.
Ph : (079) 2287 1922, Mobile : 98985 69851 E-mail : paramountcalibration@yahoo.co.in
Calibration facilities for all types of Process, Quality and Laboratory related Instruments

Calibration Certificate
Name of Customer & Address : Alfa Pumps (P) Limited
58-59, Shreenathji Ind. Estate, Kujad-Bakrol Road, Daskroi, Ahmedabad

Certificate No. : Paramount/54381 Date of Issue : 12-11-2020


Date of Calibration : 12-11-2020 Suggested Due Date : 11-11-2021
Name of Instrument : Measuring Tape
Instrument No. : MT – 01 Range : 0 – 5 Meter
Least Count : 1 mm Make : Freemans
Spring Function : Ok Clients Reference : Nil
Environment Condition : 20  2C / 40 to 60% Rh Calibration Method No. : PCC/088
Standard  Slip Gauge Box (87 pcs), Make : Aditya, Serial No. 150, Identification No. : 2/1511/1,
Reference Grade : “0”.
Used  Master Setting Rod (200 mm, 300 mm, 500 mm), Make : Mitutoyo, Accuracy :  0.01 mm.
 Slip Gauge Box, calibrated by Tanson Instrument, (NABL Accredited Lab), Certificate No.
TI/C/SGS/033/19, dated 11th March 2019, valid up to 10th March 2021.
Traceability
 Master Setting Rod, calibrated by Asian Technology (NABL Accredited Lab), Certificate
th th
No. CC223920600008853P, dated 6 July 2020, valid up to 5 July 2021.
The uncertainty of measurement is for a confidence probability of not less than 95%.
Results of Calibration
Measured Length in mm Absolute Error in microns
0 – 500 00
500 – 1000 10
1000 – 2000 30
2000 – 3000 40
3000 – 4000 40
4000 – 5000 60

Our objective is Quality Calibration not only Calibration Certificate.


Note : This report refers only to particular item (s) submitted for calibration. Paramount Calibration Centre is not
responsible for any change in calibration data and performance specification on account of mal functioning of instruments
covered by this report. The calibration results reported are valid at the time of under stated conditions of the