You are on page 1of 72

SERVICE MANUAL

VS 60/90/130
TABLE OF CONTENTS

INTRODUCTION
CONTROL PANEL
PERIODIC MAINTENANCE

START TROUBLE SHOOTING


COMPONENT CHECK
ADVANCED TROUBLE SHOOTING

COMPONENT PLACEMENT
MENU NAVIGATION
WIRING DIAGRAM

INSTALLATION
TECHNICAL DATA
FEEDBACK

SPARE PARTS LIST

Rev D TABLE OF CONTENTS


INTRODUCTION
General
The VS-series consists of vertical sterilizers with pre-
WARNING!
programmed sterilisation programs. The sterilizer
• As the sterilizer works with
works with steam produced in a chamber, fed with wa-
steam under pressure, some
ter from an external connection.
parts of it may become warm.
You should therefore exercise
WARNING TEXTS IN THIS MANUAL
the greatest caution whenever
working on the sterilizer. There are 4 different admonitory levels in the manual:
• Hot steam may flow out of the
• WARNING!
sterilizer when the door is ope-
Factor, circumstance or state that can give rise to
ned.
• When handling materials and the risk of personal injury.
loading the sterilizer, bear in • CAUTION!
mind that the chamber and the Factor, circumstance or state that can give rise to
area close to the chamber en- the risk of damage to machinery or the process.
trance is very warm.
• Liquids must be sterilised in • IMPORTANT!
special programs intended for Clarification that facilitates the understanding of
liquids. The sterilizer's liquid that described in this manual.
program is solely intended for
• TIP!
the sterilisation of liquids in
Factor, circumstance or state that can facilitate
open vessels.
execution or the process.
The following warning symbols is placed on the out-
side of the sterilizer (symbol acc. to ISO 3864)
If national requirements do not accept this symbol, the
symbol must be replaced by a suitable symbol valid in
the country were the sterilizer is used.
The warning symbol indicates that the person, re-
sponsible for the sterilizer must be fully conversant
with all requirements mentioned in the documents for
the sterilizer.

INTRODUCTION
Warranty
A 2 year warranty is provided. However, not on those
IMPORTANT! components considered to be consumables, for ex -
The warranty on the delivered ample, door gaskets. For service work, reference
product is invalidated by faulty
should be made to agreements with the supplier.
installation or an overlooked/
incorrect service interval/mainten- The Product Liability Act only applies when the in-
ance It must be possible to verify structions in this manual have been followed in their
service/maintenance. entirety.

Copyright
Copyright © 2005 Getinge Skärhamn AB. All rights re-
served.
All technical documentation provided by Getinge
Skärhamn AB is copyright protected and the property
of Getinge Skärhamn AB.
This publication may be downloaded from Getinge
Skärhamn's official website:
www.skarhamn.getinge.com.
The technical documentation is supplied as is. The
documentation can contain printing errors, technical
errors or other errors. Getinge Skärhamn AB reserves
the right to make alterations without prior notice.

No part of this publication may be copied without the


expressed permission of Getinge Skärhamn AB.

Manufacturer
Getinge Skärhamn AB Industrivägen 5
SE-47131 Skärhamn Sweden

Disposal
This product consists of (percentage by weight)
76 % Stainless steel
4 % Plastic
9 % Aluminium
5 % Brass
2 % Copper
4 % Electronics
When disposing of the sterilizer, it must be sorted at
source in accordance with the requirements set out in
the WEEE directive (2002(96/EC).
Contact your dealer when disposing of the sterilizer.

INTRODUCTION
CONTROL PANEL
Layout of the control panel
1. PROGRAM SELECTOR button – 6. Green status lamp – program finished
selection of the program
7. Red status lamp – alarm
2. START button – start of the program
8. ENTER button – select items from the menu
3. RESET button – reset a fault
9. Arrow keys – navigate through menus/lists
4. STOP button – emergency stop
10. Pressure gauge – shows the chamber pressure
5. Yellow status lamp – program in progress
11. Display unit

11 5 6 9

1 2 3 4 7 10 8

Display unit
Shows the current status, program progress, number of minutes after program start, number of completed
cycles and error codes.

IN STANDBY MODE SHOWS: SHOWN DURING THE PROCESS:


• Process name (program) • Process name

• Sterile time/temperature • Process time

Menu Home position:


Press the LEFT arrow for at least 4 seconds
Press the DOWN arrow to CYCLE COUNTER (twice)
and then press the ENTER button now press the To specifically read the load sensor temperature
DOWN arrow to read the pressure and load sen- (when the sterilizer features a liquid program),
sor temperature (when the sterilizer features a press the ENTER button at the process name.
liquid program).
Operation, check
INCORRECT OPERATION, CHECKLIST
The table below lists the most common operating
faults that result in an alarm (error code) or that cer-
tain symptoms are experienced.
Check that usage was correct.

ALARM (ERROR CODE) REASON

Close door Door has not been locked before the start button has been pressed. Read
the text below the table, Correct operation, checklist.

Sterile error The sterilizer may have been overloaded. Read the Technical data se c-
tion.

Door open The handle has been activated before the finished signal has been dis-
played. Read the text below the table, Correct operation, checklist.

Process stopped The stop button has been pressed by the operator.

Heating error Can be due to items caught in the door.

No water Perhaps the water supply tap has not been opened

SYMPTOM REASON

Process abnormally long Can be due to:

• no water on the incoming supply.


• sterilizer overloaded. Read the Technical data section.
• leakage.

CORRECT OPERATION, CHECKLIST

1 Placing items in the sterilising chamber. Only use


WARNING! loading equipment especially designed for the
Consider the risk of burns when
VS series. Carefully check that nothing is caught
loading the sterilizer.
between the door and chamber.

Place moisture-sensitive items and low-mass


items at the top of the chamber.
CAUTION!
Place heavy items at the bottom of the sterilis-
Only sterilise items intended for
ing chamber.
sterilisation with steam and that
can withstand at least 121 de- Place dish-shaped items dish side down.
grees.
Place empty containers (bottles, test tubes,
Only use loading equipment espe-
cially designed for the VS series. etc.) with their opening downwards.
Each standard basket may be 2 Close the door and pull the handle clockwise to
loaded with max. 12 kg of goods. lock.
To ensure optimal ergonomics use
a lifting device when processing 3 Press the START button, the cycle starts and will
heavy loads. be completed automatically.

OPERATION, CHECK
4 When the display shows F I N I S H, the green status
lamp comes on and when the pressure gauge
shows zero the process is completed.

5 The display never shows F I N I S H and the green


status lamp for process finished never comes on if
the process is interrupted.

6 As soon as the door is opened F I N I S H is no


longer shown on the display, this provides infor-
mation that if F I N I S H is not shown on the display,
then the goods in the chamber may have been
Recommended accessories loaded by “someone else”, but not sterilised.
Basket, part no. 566687401
Solid, stainless steel container 7 If the red status lamp for an alarm is on after a
48325031 sterilising process, the goods should be consid-
ered as unsterilised and must then be sterilised
again.

When the door is opened F I N I S H is no longer dis-


played. The display now shows:
INSTRUMENT 1 3 4 0 C (last selected program)
00 . 05 . 00 _ 1 3 4 0 C

LIQUID STERILISATION

• A special liquid program and a load sensor in the


WARNING! chamber are required to sterilise liquids. The liq-
Remember the goods are hot after
uid process is solely intended for liquid sterilisa-
sterilising.
tion in open receptacles.
Never attempt to open the door
until the pressure gauge has re- • Bottles should be filled to max. 80 % of their vol-
turned completely to zero. ume.

• Place the load sensor in the largest container.


When there is no space in the liquid for the load
sensor, use a control receptacle where the load
sensor can be placed. The volume of the liquid in
the control receptacle should be as great or grea-
ter than the liquid intended for sterilisation. Boiling
point for the liquid in the control vessel must be
the same as for the liquid to be sterilised.

• Use a receptacle with the smallest possible vol-


ume for the fastest processing.

OPERATION, CHECK
INTERRUPTED PROCESS
IMPORTANT!
If the sterilizer stops the process:
If the process must be cancelled
for one reason or another press 1 Press the RESET button; the audio signal stops.
the STOP button; the red status
2 Wait until the pressure indication shows zero.
lamp will come on and an audio
signal with sound. 3 When the red status lamp has gone out:
Now press the RESET button and
4 Press the RESET button; the alarm text is no
wait for the pressure gauge to
drop to zero. longer displayed
When the red status lamp indicat- If a process in progress has to be interrupted:
ing an alarm has gone out, press
the RESET button. 1 Press the STOP button; the text PROCESS
Open the door. STOPPED is shown on the display. The audio sig-
nal sounds and the red status light comes on.

2 Press the RESET button; the audio signal stops.

3 Wait until the pressure indication shows zero.

4 When the red status lamp has gone out, press the
IMPORTANT! RESET button; the alarm text is no longer dis-
When a process is interrupted, played.
there is a risk of steam streaming
out of the chamber when the door
is opened.

OPERATION, CHECK
PERIODIC MAINTENANCE
General
The manufacturer recommends at full utilization of ca-
WARNING!
pacity that at least the following maintenance opera-
The outer casing may only be
tions should be performed at the specified intervals
removed by technicians with do-
with by service technicians. Monthly and weekly main-
cumented experience.
tenance is carried out by the operator as set out in the
user manual.

WARNING!
Bear in mind when cleaning that
the chamber and the area close to
the chamber entrance is very
warm.

Annual maintenance
TIP! ANNUALLY OR AT LEAST EVERY 400 CYCLES
The following Service kits are UNLESS OTHERWISE STATED.
available: • Replace the filter where appropriate (accessory).
• Disks for solenoid valves Check that the filter is not wet.
• Grease for door gaskets
• Replace the plunger and O-ring on MV-1,MV-2,
MV-6 and MV-8. Refer to the section Component
WARNING! check, 3. Solenoid valves.
The electrical power supply must • Check the door and its flange. Lubricate the mov-
always be disconnected with all ing parts by the door gasket. Use a grease suit-
types of maintenance work. able for use in the food industry.

• Replace the door gasket. At least every 3rd year


or 1200 cycles. See the section Component
check, 20. Door, Door gasket.

• Measure the pressure using a calibrated absolute


test pressure gauge, use the intended flange
marked VT on the sterilizer. See the section
Component check, 12. Absolute pressure se n-
sor.

• Inspect and rectify any leakage from pipe cou-


plings and instrument connections. See the sec-
tion Advanced trouble shooting, Trouble
shooting leakage.

• Check and clean the water outlet in the chamber.


At least every 800 cycles.

• Check that the monthly and daily maintenance


have been carried out correctly.

PERIODIC MAINTENANCE
• Perform a functional check of the electrical com-
ponents. See the section Component check. At
least every 1200 cycles or every 3rd year.

• Reset the service alarm from the Service mes-


sages menu. (situated under S E-
TUP/SYSTEM/SERVICE in the menu tree)

SAFETY CHECKS

• Check that the door seals against the steam pres-


sure during the sterilising process. See the sec-
tion Component check, 20. Door.

• Check that the sterilizer can not be started when


the door has not been closed. See the section
Component check, 20. Door.

• Check that the door's counter-hold spring works


and prevents crush injuries. See the section
Component check, 20. Door, check the door
spring

• Inform the customer about the significance of the


warning signs.

• Inform the customer of the risks involved with the


program for sterilising liquids, if installed. Particu-
larly the information that it is only permitted to
sterilise liquids inopen vessels.

PERIODIC MAINTENANCE
Monthly maintenance
MONTHLY, OR AS REQUIRED

• Clean the bottom filter. See the section Compo-


nent check, 1. Chamber.

• Clean the chamber. See the section Component


check, 1. Chamber.

• Clean the door gasket and lubricate with the gre-


ase recommended by Getinge.

Daily maintenance
Fig. 1
• Check that the overheating protection sensor is fit-
ted on the heating element, see fig 1. See the
section Component check, 6. Overheating pro-
tection.

• Check that the level guard in the bottom of the


chamber is clean, fig. 2.
See the section Component check, 1. Chamber.

Fig. 2

PERIODIC MAINTENANCE
START TROUBLE SHOOTING
Trouble shooting orientation
START TROUBLE SHOOTING HERE

Choose the error code/codes displayed and the symp-


4 Final 1 Start tom experienced in the table below. Carry out the pre-
verification Choose the scribed actions in the stated order, until the fault has
error code
been identified and corrected.
symptom

CHECK PROCEDURE

1 START
Select the alarm (error code) or symptom in the
tables below.
3 Verify the 2 Action 2 ACTION
action Carry out the actions set out in the table. Carry
out the actions in the stated order, continue until
the fault has been corrected.

a. Check the components. In the event of a


component fault, repair or replace it. Instruc-
IMPORTANT! tions are enclosed with the spare part or are
The actions stated in the tables set out after each check.
are usually specific checks on the
components. However, the actions b. Should the check indicate that the component
can also represent advanced is OK, return to the table in this section and
trouble shooting and indicate the continue with next action as per point 2.
specific component to be checked. 3 VERIFY ACTION
After repair/replacement the action must be veri-
fied. Instructions concerning verification are en-
IMPORTANT! closed with the spare part or are set out after
After the action (for example, each check.
component replacement) the ac-
tion must be verified (e.g. man- 4 FINAL VERIFICATION
ual activation). (ERROR CODE/SYMPTOM VERIFICATION)
In addition, it must be verified that After action verification it must be verified that the
the error code/customer's experi- displayed error code is not displayed again and
enced symptom has been cor- the experienced symptom no longer remains. This
rected. is done by running a program and checking that
Several actions may be needed no error codes are displayed or the symptoms are
with subsequent action verifica- experienced.
tion, remember to always end
with verification against the
error code/symptom.

START TROUBLE SHOOTING


A l ar m ( e r r o r c o d e s v i s i b l e o n t h e d i s p l a y )
EXPLANATIONS
Explanations to references to sections stated in the
tables below:

CC : chapter Component check


Advanced chapter Advanced trouble
trouble shooting: shooting

Other references are given in plain text.

ERROR CODE EXPLANATION ACTIONS IN ORDER OF PRIORITY

GETINGE PACS 300 Process does not start, frozen proc- 1. CC, 19. Software PACS cold start
GETINGE PACS 300 DIS- ess 2. CC, 18. Display
PLAY 3. CC, 17. Circuit board

CLOSE THE DOOR Door unlocked at point of start 1. Operation, check


2. CC, 20. Door, (micro-switch)
3. CC, 17. Circuit board

DOOR OPEN Door open during process 1. Operation, check


2. CC, 20. Door, (micro-switch)
3. CC, 17. Circuit board

POWER FAILURE Power failure during process 1. Check external fuses.


2. CC, 6. Overheating protection
3. CC, 17. Circuit board (Fuse)
4. CC, 17. Circuit board

PT-100 ERROR Error on load sensor during process 1. CC, 15. Load temperature sensor
2. CC, 17. Circuit board AI

SERVICE Service indicator. 1. Contact the service technician

STERILE ERROR Incorrect pressure/steam tempera- 1. Operation, check


ture in the sterile phase 2. CC, 15. Load temperature sensor
3. CC, 12. Absolute pressure sensor,
4. CC, 17. Circuit board

PROCESS STOPPED The process has been stopped with 1. Operation, check
the STOP button 2. CC, 18. Display

PRESSURE SENSOR Pressure sensor incorrectly cali- 1. Advanced trouble shooting, Checks
ERROR brated with error code

UPDATE Battery replacement or circuit board 1. CC, 17. Circuit board


fault

START TROUBLE SHOOTING


ERROR CODE EXPLANATION ACTIONS IN ORDER OF PRIORITY

NO WATER Timeout level guard 1. Operation, check


2. CC, 10. Level guard
3. CC, 3. Solenoid valve MV-3

HEATING ERROR Timeout heater 1. CC, 6. Overheating protection


2. CC, 5. Relay
3. CC, 4. Heater
4. CC, 17. Circuit board

HEATING ERROR Leakage Trouble shoot according to the section


Trouble shooting leakage

MANUAL MODE Component manually activated 1. Operation, check


2. An output on the circuit board has
been set to manual mode. ( menu
tree Diagnostic)
POST TREATMENT ER- Outlet valve clogged CC, 3. Solenoid valve, MV-1
ROR
POST TREATMENT ER- Screen in the chamber clogged CC, 3. Strainer
ROR

START TROUBLE SHOOTING


Symp tom (experienced by user)
EXPLANATIONS
Explanations to references to sections stated in the
tables below:

CC: chapter Component check


Advanced chapter Advanced trouble
trouble shooting: shooting

Other references are given in plain text.

SYMPTOM ACTIONS IN ORDER OF PRIORITY

No ready signal 1. CC, 12. Absolute pressure sensor


2. CC, 3. Solenoid valve MV-2
3. CC, 11. Air filter

Instruments rust in the sterilizer 1. Poor quality instruments


2. Water cleaning filter – type de-ionization, check
(see documentation supplied with the water fil-
ter)

Process time abnormally long 1. See the section installation. Electricity, Water
requirement
2. Advanced trouble shooting, Leakage test, check
3. Operation, check
4. CC, 22. Strainer

Humming noise in standby mode 1. CC, 3. Solenoid valves

Door hard to open/cannot be opened 1. CC, 3. Solenoid valve MV-8


2. CC, 20. Door, (lock)

Pressure gauge shows pressure with the door open 1. CC, 14. Pressure gauge

Steam leaks from the door 1. CC, 20. Door, door gasket
2. CC, 3. Solenoid valve MV-5
3. CC, 21 Pressure relief valve door gasket

Black background on the display 1. CC, 23 Fan.


2. CC, 18 Display

Unstable measurement value and text on display 1. CC, 23 Fan.


2. CC, 18 Display
3. CC 17 Circuit board
4. Check the power supply (Fluctuations?)

START TROUBLE SHOOTING


COMPONENT CHECK
TABLE OF CONTENTS

1. General .........................................................2
2. Chamber.......................................................2
3. Pipe/pipe couplings ........................................2
4. Solenoid valves .............................................3
5. Heater...........................................................5
6. Relay ............................................................6
7. Overheating protection, resetting ....................6
8. Non return valve BV-1 ....................................7
9. Non return valve BV-2 ....................................7
10. Non return valve BV-3 ....................................7
10. Non return valve BV-3 ....................................7
11. Level guard ...................................................8
12. Air filter (accessory) .......................................8
13. Absolute pressure sensor ...............................8
14. Ejector ..........................................................9
15. Pressure gauge .............................................9
16. Load temperature sensor (accessory) ........... 10
17. Pressure relief valve .................................... 10
18. Circuit board................................................ 11
19. Display ........................................................ 12
20. PACS software ............................................ 12
21. Door ........................................................... 13
22. Pressure relief valve, door gasket ................. 16
23. Screen ........................................................ 16
24. Fan ............................................................. 16

COMPONENT CHECK
page 1
1. General
It is extremely important for safe functionality that all
IMPORTANT!
nipples are tightened so that air, which can jeopardize
Only use Teflon tape that is in-
the sterilisation result, can not enter the chamber. If
tended for sealing pipe couplings
pipe couplings must be removed while performing
against steam pressure.
checks, think about the following when assembling:

• Use ample Teflon tape, so that the tape partly


forms a seal and partly seals the threads.

2. Chamber
The chamber walls are cleaned using a mild lime dis-
solving agent. Rinse well. Drain by closing the door,
the bottom valve then opens automatically.

BOTTOM FILTER (ALSO SEE 23. SCREEN)


Lift out the filter. Clean

LEVEL GUARD (ALSO SEE 11. LEVEL GUARD)

Check that the rocker moves up and down easily.

3. Pipe/pipe couplings
Check:
TIP!
Start a 134 °C program. Trouble shoot when the steri-
Check whether steam is let out
lizer is in “Sterilisation” mode.
with an overpressure.
• Use soapy water on cold com- Check whether steam is let out.
ponents
Action:
• Use a mirror on hot compo-
Check according to respective component checks.
nents
Also see 1. General.

Verification:
Check all pipes/pipe couplings again.

COMPONENT CHECK
page 2
4. Solenoid valves
Pay attention to solenoid valves that hum! The noise
may be a forewarning about the coil starting to over-
IMPORTANT!
heat due to it not working in a closed iron circuit. This
All valves open in an energised
can occur when the moving core is influenced by the
state.
magnetic forces from the winding and dirt prevents the
coil from reaching the end position. An air gap is then
formed in the magnetic circuit, which lowers the induc-
tance in the coil resulting in increased current.

MV-1

Is fitted in the bottom of the sterilizer and has the task


of evacuating any remaining water in the sterilizer at
the end of the process.

Check:
Check that both the plunger and the area it runs in are
clean. Always change the O-ring that seals between
the coil and housing after removing the coil. Activate
MV-3 and fill the chamber. Activate MV-1 and MV4
(ejector). Check that the water runs out of the cham-
ber again. Measure the voltage across the coil.
Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.
Verification:
Check that the water level in the chamber is main-
tained for a long period (about 15 minutes).

MV-2
Is fitted at the rear, top edge of the sterilizer. Takes
care of pressure equalization after the post-vacuum
phase by releasing air into the chamber. The valve is
open when the sterilizer is in standby mode if the door
is locked. It becomes de-energised as soon as the
door is opened.
Check:
Check that both the plunger and the area it runs in are
clean. Always change the O-ring that seals between
the coil and housing after removing the coil. Activate
MV-4 and MV-2 on the diagnostic menu. The pressure
should not drop appreciably below atmospheric press-
sure. Measure the voltage across the coil.

Check for leakage: Dismantle BV-2. Activate MV-3


and the heater. No steam should come out of MV-2.
Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.
Verification:
Run a sterilisation program. Check for leakage ac -
COMPONENT CHECK
page 3
cording to Check and that pressure equalization takes
place at the end of the program.

MV-3

Is fitted directly on the inlet. Loads the chamber with


water at the start of the process.
Check the opening function:
Activate MV-3 in the diagnostic menu; water should
then flow into the chamber.
Check the closing function:
In standby mode, check that the water level in the
chamber does not rise once the valve to the chamber
has been closed.

Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.

Verification:
Run a sterilisation program. Check that no water en-
ters the chamber after the process has ended and the
door is open.

MV-4
Opens the water flow to the ejector to produce a vac-
uum
Check:
Check that water does not run out of the drain hose in
standby mode
Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.
Verification:
Run a process and check again that water does not
run out of the drain hose in standby mode.

MV-5
Pressurises the door gasket before a program.

Check:
IMPORTANT!
During the process, check that the door seals against
Minor steam leakage can be due steam.
to a leaking gasket and has noth-
ing to do with the MV-5 function Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.

Verification:
Check the door for steam leakage.

MV-6

Vents the chamber while the program is running.


Opens at the given time interval. Also simplifies eva-

COMPONENT CHECK
page 4
evacuation of the chamber pressure after sterilisation
is completed.
Check:
Build up a pressure in the chamber. Open MV-6. Th e
chamber pressure should drop. Check that the cham-
ber pressure is maintained when MV-6 is closed. Pi-
pes on the outlet side should be cooler than on the
pressurised side.
Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.
Verification:
Run a program. Now verify according to Check, see
above.

MV-8

Evacuates the door gasket pressure after the com-


pleted process.
Check:
IMPORTANT!
Close and lock the door. Open MV-5 and open MV-8.
A symptom of poor opening func-
Water should flow out of the drain.
tion in the valve is difficult door
opening. Action:
Clean and replace the plunger and O-ring. Replace
the entire valve if necessary.
Verification:
Run a process. The door should be easy to unlock af-
ter the completed process.

5. Heater
Is fitted in the bottom of the chamber and has the task
TIP!
of heating the water dosed for the process.
Ensure the heater is de-energised
before you measure the resis- Check:
tance. Activate the heater. Measure with a clamp-on amme-
ter, this should show about 8 A/coil. The heater output
is 2000 W x3 coils, which gives a resistance of about
27 O/coil. Check the resistance.

Action:
Replace the heater with incorrect resistance.
If the clamp-on ammeter shows 0 A, and the relay and
circuit board are OK. Replace the heater.
Verification:
Verify by measuring the resistance.

COMPONENT CHECK
page 5
6. Relay
Is fitted in the distribution box. Its task is to connect
the heater to the power supply.
Check:
Activate the heater, measure the coil on the relay us-
ing a voltmeter. The coil is across A1-A2. Measure the
voltage on the outgoing connection 2 - 4 - 6

7 . O v e r h e a t i n g p r o t ec -
tion , r e s e t t i n g
Protects the heater against overheating, the protection
trips at approx. 300 °C. Requires manual resetting.
The sensor is positioned on the heater coil. The reset-
ting button is positioned on the protection. A symptom
that the overheating protection has tripped is the ti-
IMPORTANT! meout alarm, heating error.
Note that if overheating protection
Possible causes:
trips due to a temperature outside
of the working range (0 - 300 °C), • That the sterilizer has been transported in tem-
the sterilizer must cool for a pe- peratures below freezing.
riod so that the resetting button
shall remain pressed in. Wait until • That a fault has occurred in the temperature regu-
the sterilizer is within the working lation of the heater.
range again. • That leakage results in the water being consumed
Note that it may also require a
during the process.
great deal of force to press in the
overheating protection. The unit should always be de-energised when reset-
ting.
Check:
Check that the manual resetting button is pressed in.
Measure the voltage on the incoming and outgoing
pins on the protection. Voltage on the incoming but
not on the outgoing, means that the protection has
tripped or is faulty.
CAUTION!
Action:
Check that the overheating protec-
Replace the overheating protection. Make sure that
tion sensor is fitted correctly on
the sensor is correctly fitted on the heater. Also see
the heater. Otherwise the sensor
daily maintenance
has no function.
Verification:
Check that the sterilizer display is lit.

COMPONENT CHECK
page 6
8. Non return valve
B V- 1
Is fitted between MV-1 and the chamber and has the
purpose of preventing return flow through MV-3
Check:
Fill the chamber with an open door by opening MV-3.
Shut the main tap and let MV-3 remain energised and
open MV-4. Check whether water is running out from
the drain or that the level in the chamber drops over a
long period of time, at least 15 minutes.
Action:
Replace BV-1.

Verification:
As Check, see above.

9. Non return valve


B V- 2
Is fitted between the air filter and MV-2. Its task is to
prevent steam, with a fault on MV-2 being forced thro-
ugh the filter.

Check:
Close and lock the door. Remove the air filter. Build
up a pressure; open MV5 and MV1, close MV1, acti-
vate the heater, open MV2. The pressure should rise
to 1 bar according to the front pressure gauge. Check
that no steam comes out of the air filter holder.

Action:
Replace BV-2. Check that the air filter is not wet.
Verification:
As Check, see above.

10. Non return valve


B V- 3
Is fitted after MV-1 and has the task of preventing wa-
ter from the ejector returning up into the chamber
Check:
Open MV-4 and MV-1. Check that no water enters the
chamber.
Action:
Replace BV-3
Verification:
As Check, see above.

COMPONENT CHECK
page 7
11. Level guard
Is fitted in the lower part of the chamber above the
heater. The task is to check the filling of the chamber.
Check:
Move the rocker arm up and down so the digital input
(Service, Diagnostic, Digital input) should switch be-
tween 1 and 0.

12. Air filter (accessory)


Its task is to supply the sterilizer with sterile filtered air
after the post vacuum phase.
Check:
The air filter must be dry. Check that pressure equali-
zation takes place after the post vacuum phase in the
program.

Action:
Replace the filter.
Verification:
Run a program. Also see section BV-2, check.

1 3 . Ab s o l u t e p r e s s u r e
sensor
Its task is to monitor and control the process.
CAUTION!
Connect the pressure gauge for Check:
the absolute pressure to the outlet Carried out with an independently, calibrated pressure
on the lead-through at the rear of gauge for the absolute pressure, whose values are
the chamber. compared with the values shown on the display. Area:
0 - 4 bar absolute pressure corresponds to 0 - 5 VDC,
output signal with a open door approx. 1.25 V (meas-
ured between the “common” and “out” on the pressure
sensor).
Action:
TIP!
When the deviation is significant, replace the sensor
Start a program and stop it by and calibrate. With minor deviation, calibrate. Calibra-
using the “Pause” switch in the tion is carried out via the menu tree. Step using the
vacuum and sterile phase. Then
UP arrow and the ENTER button through SETUP/
adjust the A-value (sterile phase) SYSTEM/ (password)/ CALIBRATION/ MANUAL CA-
and the B-value (vacuum) LIBRATION. Select the sensor and adjust index A or
B up or down depending on the type of fault indica-
tion. B with vacuum and A with a sterile pressure.

TIP! Verification:
Use the program CS1000 (for PC), Run a sterilisation program. Perform with an inde-
for a basic calibration procedure. pendently, calibrated pressure gauge for the absolute
pressure whose pressure values are compared with
the values shown on the display.
COMPONENT CHECK
page 8
1 4 . E j ector
The ejector is mounted on the pipe from the water in-
let to the drain. The ejector's function is to create an
underpressure in the chamber.

Check:
Activate MV-1 and MV-4 through the diagnostic menu.
The pressure shall, under the condition that BV-3 is
problem free and the water pressure is above 1.5 Bar,
reach a vacuum greater than 800 mBar within 15 min-
utes.

Action:
Clean the ejector. Any dirt or leakage in the ejector
connections cause impaired vacuum capacity.

Verification:
Check the depth of vacuum according to the check
section above.

15. Pressure gauge


CAUTION!
Is placed on the front of the sterilizer. Shows the
The pressure gauge only gives an
chamber pressure even if the power is switched off.
indication of the pressure and
should not be used during calibra- Check:
tion. When the door is open the pressure should show 0
±0.1 bar. Check that the pointer indicates a pressure
higher and lower than zero during a cycle.

Action:
Calibrate the pressure gauge according to the section
Pressure gauge adjustment. If it is not possible, re-
place the pressure gauge.
Verification:
Run a program. Compare the pressure gauge with the
pressure indication on the display.

PRESSURE GAUGE ADJUSTMENT

• Make sure the door is open.

• Open the front.

• Loosen the brass nipple at the rear of the pres-


sure gauge and lift out the pressure gauge.

• Pry loose the pressure gauge glass by inserting a


screwdriver in the notch at the top of the pressure
gauge glass.

• Insert narrow tipped pliers under the pointer and


lift it off.

• Adjust and press back the pointer in the right posi-


tion.

COMPONENT CHECK
page 9
16. Load temperature
sensor (accessory)
TEMPERATURE SENSOR

The sensor is placed in the lead-through to the cham-


ber and is connected to contact piece X2 on the circuit
board. It monitors the sterile process and provides the
display with information about the temperature of the
liquid that sterilises and controls that the liquid is un-
der 95°C before the door can be opened.

Check:
Loosen the connection on the circuit board and meas-
ure the resistance on the sensor with an ohmmeter.
The sensor's resistance should at 20 ° be about 120
ohm. The higher the temperature, the higher the resis-
tance.

Action:
With a resistance above 150 ohm or 0 ohm. Replace
the sensor.

Check for leakage:


Check that there is no steam leakage from the sen-
sor's lead-through – silicone seal.

Action:
With leakage – replace the silicone seal.

1 7 . P r e s s u r e r e l i e f v a l ve
The function of the pressure relief valve is to relieve
the pressure in the chamber, should this rise uncon-
trolledly.
Check:
Start a process. When “sterilisation” is reached, close
MV-1 and MV-6 as well as start the heater. Check us-
ing an independently calibrated pressure gauge for
absolute pressure that the opening pressure is 3.7
±0.14 bar (absolute pressure). Check that the pres-
sure relief valve does not leak water or steam at nor-
mal sterilisation pressure – max 3.15 bar.
Action:
Clean the valve. Replace the valve if necessary.

Verification:
Run up the chamber pressure, activate the heater and
WARNING! close MV1 and MV6. Read when the pressure relief
Adjustments or actions that alter valve trips. Check that the valve does not leak during
the opening pressure on the a program
pressure relief valve may only be
performed by persons with docu-
mented experience.

COMPONENT CHECK
page 10
18. Circuit board
Is located in the casing to the right. Its task is to bridge
CAUTION!
all input and output signals with the software PACS.
The display unit and the circuit
The circuit board houses, among others, the flash
board unit must always be kept in
memory that holds the PACS software and the bat-
shielded bags until the compo-
tery.
nents are to be installed in the
sterilizer. Make sure not to be Check:
electrostatically charged when
• Incoming voltage on X23.
these parts are handled, by e.g.
touching an earthed heater or by • Jumpers are located correctly: (X28=Flash,
using an ESD wrist strap. X29=Normal, X30=ON).

• Fuse 500 mA.

• Voltage on the battery, 3.6 V.

• Connect the cables and connectors.

Computer communication:
On connection COM 1 ( X24) the board is equipped
with a serial communication port (RS232) for connec-
tion e.g. to a computer.
Action:
WARNING! Replace the battery.
Lithium battery. Explosion risk.
Replace the battery with the same Make a backup of the software according to the sec-
type as recommended by Getinge. tion Software PACS, if possible.

Cut off the battery solder tags as close to the battery


as possible.

Do not solder on the circuit board!


TIP!
Battery Solder in the new battery on the old solder tags.
When battery replacement is nec- Check that the new battery's voltage is 3.6 V
essary and no battery is available,
the sterilizer can temporarily be Perform a cold start according to Cold start.
run with jumper X29 (see section Action:
Cold start ) in cold start mode. If the above does not work, correct or replace the
The calendar and process counter card.
do not work in cold start mode.
Verification:
Run a prepacked program.

COMPONENT CHECK
page 11
19. Display
Is placed on the front of the sterilizer. Its task is to
show values and act as the interface to the PACS
software. Connected X26 on the circuit board. The
keypad in the decal is connected to contacts J1 and
J3 on the display card.
Check:
Check the cables including connection X26. Check
jumper JP1.
Should be fitted between pins 1 and 2.

Action:
Replace the display card.
Verification:
Run a program and make sure the display communi-
cates without interference.

20. PACS software


Is installed in the flash memory on the circuit board.
Its task is to monitor and govern the entire sterilizer
and the processes run on it. PACS also logs events
and alarms.

Check:
Run a process.
Action:
MAKE A BACKUP/LOAD A BACKUP.
Follow the following when the processes are to be sa-
ved on a PC:

Connect your PC to contact X24 on the sterilizer use


the communications cable.

Start the CS1000 program. State the password for le-


vel 3.

Write down all calibration values.

Go to the menu TOOLS-PACS to RAM/ UPLOAD TO


FILE.

Save the backup file on the hard drive in the PC.

COMPONENT CHECK
page 12
TO LOAD THE BACKUP FILE FROM THE PC TO
THE STERILIZER:

Connect your PC to contact X24 on the sterilizer use


IMPORTANT!
the communications cable.
All calibration values are lost when
a new backup is loaded. Therefore Start the CS1000 program. State the password for le-
note all values on a paper or the vel 3.
like.
Go to the menu TOOLS-PACS to RAM/ DOWNLOAD
TO FILE.

Enter all previously noted calibration values in the new


backup.

Go to the same menu and select SAVE RAM TO


FLASH.

Verification:
Read the software version on the display.

COLD START

Switch off the power and place the jumper X29 in po-
sition Cold.

Switch on the power again, wait a few seconds.

Switch off the power, place the jumper in the Normal


position and switch on the power again.

21. Door
Door is locked mechanically by a bayonet mount and
10 flanges. The door gasket seals through the gasket
groove being pressurised at the start of the process
which then presses against the door.

Open and close the sterilizer door several times in a


row and check that there is no play on the door and
hinge. Also check that the door spring has not be-
come too weak or has broken.
Check:
Check that a process can not be started if the door is
not closed correctly. Check the hinge joints, micro-
switch and door lock.

COMPONENT CHECK
page 13
Action:

• Clean and lubricate the door's moving parts.

• Check the wear to the hinge joint, change if nec-


essary.

• Check the micro-switch by activating it manually


IMPORTANT!
and read the display (switches between the pro-
The door must be fully open (110°)
to carry out the action. gram parameter and “S T A R T O K ” on the display)

• Check the door lock moves easily and does not


jam. Check especially that it reaches all the way in
the deactivated position.

Verification:
Open and close the door. The door's movement
should not be more resistant than it can be closed us-
ing normal force. In general as in Action.
Check the door spring:
Check that the door spring has not become too weak
or has broken by opening the door. The door spring
shall prevent the door from shutting.

Action:

• Remove the door's plastic cover

• Open the door as far as possible

• Loosen the nut (6)

• Loosen the index screw (7). Remove the door

• Remove both clasps holding the spring axle. Se-


cured with six nuts under the axle
7 6 8 • Remove the circlip (8) located at the end of the
spring axle.

• Pull off the axle

• Change the spring. Press the spring leg in the


door holder.

• Refit in the reverse order

• Centre the door with the help of the screw (7)

Verification:
Open the door and check that the door spring pre-
vents the door from shutting.

DOOR GASKET
Door is locked mechanically by a bayonet mount and
10 flanges. The door gasket seals through the gasket
groove being pressurised at the start of the process
which then presses against the door.
Check:
Check that the gasket is in one piece and that the

COMPONENT CHECK
page 14
contact surfaces are clean and lubricated with a food
industry grease.

Action:
The door gasket should be replaced after max 1200
cycles/once every 3rd year. Close the water connec-
tion before replacing the door gasket.

Open the door

Remove the gasket using a blunt tool

Check that the gasket groove is clean

Clean the groove if necessary

Grease the new gasket

Press the new gasket in the groove fully

Verification:
IMPORTANT! Close the door and run a program 134 °C check for
Check that steam does not blow leakage. Also see section Component check, 4. So-
out around the door when the lenoid valves, MV-5 and MV-8.
pressure is built up.

COMPONENT CHECK
page 15
22. Pressure relief va l -
ve, door gasket
Used to release the water pressure in the door gasket
groove if the pressure exceeds 4.8 bar
Check:
Check that the pipe after the pressure relief valve is
not hot during “sterilisation”.
Action:
Replace the pressure relief valve

Verification:
Run a new process and check whether the pipe is hot

23. Screen
Is positioned in the bottom of the chamber to prevent
a stoppage in the outlet.
Check:
Remove the screen, by lifting it straight up from the
chamber
Action:
Clean the screen

Verification:
Make a visual inspection that the screen is clean.

2 4 . Fan
The fan is positioned behind the casing and is con-
nected to contact piece X16 on the circuit board. Its
task is to lower the temperature under the casing
when the process is run.
Check:
Loosen the casing and check that the fan rotates
when a process is run. Check that the fan is voltage
fed.

Action:
Replace the fan

COMPONENT CHECK
page 16
ADVANCED TROUBLE SHOOTING
This section concerns advanced trouble shooting. The
different sub-sections below follow on from the previ-
ous sectionSTART TROUBLE SHOOTING. This me-
ans that all trouble shooting starts there.

Flow chart

ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION


Process description
Steam is generated in the sterilizer through an exter-
nal water feed that may be preceded by a softening fil-
ter.
5. Condensation is led off to the drain.

1. The door gasket is pressurized by MV-5 opening


6. 7. and releasing water into the gasket groove.
8.
2. The sterilizer is filled with water until the level gu-
ard cuts in.

3. The heater heats the water and the pressure ri-


3. ses.

4. The steam pressure is monitored by the pressure


1. 2. 9.
sensor which, via a PID-regulator, controls the
4. heater.

5. The temperature of the steam is monitored by the


10. pressure sensor (through a theoretical calcula-
tion).

6. The solenoid valve MV-1 is used to empty the wa-


ter from the chamber.

7. The steam pressure is released via MV1 and MV6

8. The water pressure powers an ejector, which cre-


ates an underpressure in order to lower the
chamber pressure below atmospheric pressure.
MV4 releases the water pressure on the ejector.

9. MV-2 is used to release the air in the chamber.

10. MV-8 releases the door gasket's water pressure.


The door can now be opened.

ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION


LIQUID LOAD

11. The door gasket is pressurised.


15.
12. The sterilizer is filled with water until the level gu-
ard cuts in.
16.
17. 13. The heater heats the water and the pressure ri-
ses.

14. The steam pressure is monitored by the pressure


sensor which, via a PID-regulator, controls the
13.
heater.
11. 12. 18. 15. The load sensor monitors the temperature of the
liquid to be sterilised. The load sensor tempera-
14. ture determines when the sterilizer switches to
sterilisation.

16. The solenoid valve MV-1 is used to empty the wa-


ter from the chamber.

17. The steam pressure is released via MV1 and MV6


through ramping to prevent decoction.

18. When the liquid reaches a temperature below


95°C (controlled by the load sensor) MV-8 re-
leases the gasket pressure and the door can be
opened. The process is complete.

PRE-TREATMENT

The purpose of pre-treatment is to remove air from the


chamber and from the goods. Air prevents the requi-
site contact between the steam and micro-organisms
to be killed.
Pre-treatment consists of steam injection and evacua-
tion pulses in different processes depending on the
program selected. In addition, the humidification es -
sential to killing mainly takes place during the pre-
treatment stage.

Pre-treatment process:

• Blow through (1)


MV-1 opens, the chamber is filled with water and
the heater then heats the water. MV-6 opens dur-
ing the preset time.

• Pressure increases (2)


MV-6 closes and the heater heats. The pressure
rises to a preset level. The sterilisation pressure
should be reached during the final increase in
pressure. Sterilisation begins when the sterilisa-
tion pressure is reached.

ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION


STERILISATION
Sterilisation takes place by maintaining the saturated
steam's relation (boiling point at a specific pressure).

The sterilisation phase is in progress during a pre-


programmed number of minutes at the preselected
temperature/pressure.

Sterilisation process:

• MV-6 opens during the preset time intervals.

• The pressure is controlled by means of the heater,


which heats as required.

POST TREATMENT
The purpose of post treatment is to reduce the mois-
ture content in the goods. During post treatment the
pressure drops to atmospheric pressure or lower –
depending on the selected program. Post treatment is
completed by air being drawn in until the chamber is
at atmospheric pressure.
Post treatment process:

• MV-1 opens and releases the remaining water.

• MV-4 opens and the ejector starts.

• MV-6 opens after an additional few of minutes.

• When the right vacuum is reached (800 mBar)


time metering starts.

• When the time elapses MV-1, MV-4 and MV-6


close. MV-2 opens and air is drawn in and the
pressure rises towards atmospheric pressure.
MV-8 opens and the door gasket pressure is re-
leased. The process is finished.

ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION


PROCESS DESCRIPTION - LIQUID LOAD

PRE-TREATMENT
The purpose of pre-treatment is to remove air from the
chamber and from the goods. Air prevents the requi-
site contact between the steam and micro-organisms
to be killed.

Pre-treatment consists of steam injection and evacua-


tion pulses in different processes depending on the
program selected. In addition, the humidification es -
sential to killing mainly takes place during the pre-
treatment stage.
Pre-treatment process:

• Blow through (1)


MV-1 opens, the chamber is filled with water and
the heater then heats the water. MV-6 opens dur-
ing the preset time.

• Pressure increases
MV-6 closes and the heater heats. The pressure
rises to a preset level just above the sterilisation
pressure to accelerate the heating of the liquid.
When the load sensor temperature approaches
the sterilisation temperature, the steam pressure
drops to the sterilisation pressure. Sterilisation
starts.

ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION


STERILISATION
Sterilisation takes place by maintaining the saturated
steam's relation (boiling point at a specific pressure).

The sterilisation phase is in progress during a pre-


programmed number of minutes at the preselected
temperature/pressure. The load sensor monitors the
temperature of the liquid.
Sterilisation process:

• MV-6 opens during the preset time intervals.

• The pressure is controlled by means of the heater,


which heats as required.

POST TREATMENT

The purpose of post treatment is to lower the tem-


perature of the liquid below boiling point at atmos-
pheric pressure.

Post treatment process:

• MV-4 opens and the ejector starts.

• MV-1 opens at the preset time intervals to ensure


a slow pressure decrease and prevent decoction
of the liquid. The remaining process water is re-
leased in connection with ramping.

• When atmospheric pressure is reached and the


temperature of the liquid (controlled by the load
sensor) is below 95°C, MV-1 and MV-4 close.
MV-2 opens and the atmosphere in the chamber
and the atmosphere is the room are kept at the
same level. MV-8 opens and the door gasket
pressure is released. The process is finished.

ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION


T r o u b l e s h o o t i n g l e a k-
age
1 Check for door leakage as set out in the section
IMPORTANT!
Check component, 20. Door.
Trouble shooting takes place:
• At overpressure 2 Check for leakage from the pressure relief valve
Follow the method prescribed in as set out in the section Check component, 16.
Check component, 2. Pipe/pipe Pressure relief valve.
couplings.
3 Check for leakage from the solenoid valves as set
out in the section Check component, 3. Sole-
noid valves.

TIP! 4 Check for leakage from the pipe couplings as set


Check whether steam is let out out in the section Check component 2.
with an overpressure. Pipe/pipe couplings.
• On cold components: 5 Check for leakage on the temperature sensor's
use soapy water
lead-throughs as set out in the section Check
• On hot components
component, 15. Load temperature sensor.
use a mirror
6 Check for leakage from the ejector as set out in
the section Check component, 13. Ejector.

Calibration instruction
1 Connect a calibrated test pressure gauge to the
connection labelled PT on the sterilizer.

2 Start a 134°C program.

3 Let the program run to post treatment

4 When the vacuum has stabilized, calibrate the B


parameter for the pressure sensor.

5 Check that the external pressure gauge shows


the same pressure as on the sterilizer's display.

6 Allow the sterilizer to finish.

7 Start a new 134°C program. Let the program run


to “sterilisation”. Calibrate the A parameter for the
load sensor. Check that the external pressure
gauge shows the same pressure as on the steri-
lizer's display.

8 The sterilizer will calculate the theoretical sterile


temperature, based on the calibrated pressure va-
lues.

9 Allow the program to finish. Check the vacuum


setting during the post vacuum phase.

ADVANCED TROUBLE SHOOTING, TROUBLE SHOOTING LEAKAGE


MENU NAVIGATION
Orientation
Navigation through the menus is done using the four
arrow keys UP arrow, DOWN arrow, LEFT arrow,
RIGHT arrow and the ENTER key.

Navigate to your choice using the UP arrow and


DOWN arrow. The selection is indicated on the dis-
play by a cursor below the value to be changed.

To return to the previous level, press the left arrow


when the menu position is not selectable.

Press the ENTER key to activate your menu selec-


tion.
When a menu selection has been activated, one of
two actions can be performed.

• When the selection is a new menu position, a fur-


ther selection can be made.

• When the value can be edited, use the arrow keys


TIP! as set out above.
You can return at anytime to the
The selection can either be shown on one or two
main menu by pressing the left
rows. When one row is shown, this means the row
key and holding it down for 4 sec-
can be selected and activated using the ENTER key.
onds.
When two rows are shown, the first row shows the se-
lected name and the second row the value linked to
the selection.

MENU NAVIGATION
Edit values
Press the ENTER key to set the display in edit mode,
i.e. make the selected value editable. The cursor is
positioned to the far left below the first character when
the value is selectable.

EDIT PRESET VALUES


Use the UP arrow and DOWN arrow to change a va-
lue. Different values are shown each time the arrow
key is pressed.
Save the value by pressing the ENTER key.

EDITING NUMERICAL VALUES


The value can consist of several characters. The cha-
TIP! racter with the cursor below can be edited.
When adjusting the sterile or post
vacuum times, the value can not The highlighted character can be increased or de-
be set below the preset value. creased by using the UP arrow or DOWN arrow.
Use the RIGHT arrow or LEFT arrow to select the
next editable character to the right respective left.
When you have finished, save the values by pressing
the ENTER key.

MENU NAVIGATION
THE MENU TREE
Main menus
Main selection: Program name, Program version,
Phase, Cycle counter and Setup menu.

Navigate to your choice using the UP arrow and


DOWN arrow. The selection is indicated on the dis-
play by a cursor below the value to be changed.

To return to the previous level, press the LEFT arrow


when the menu position is not selectable.
Press the ENTER key to activate your menu selec-
tion.

1. PROGRAM NAME
Program name: Selected program

Params 1 Params 2: Sterilisation time & temp (or


Heating up)
Select program: Sub-menu to selected program.
Important, the sterilizer has a PROGRAM SELEC-
TOR button for this on the front.
Change parameter: Possibility to change the sterile
time and post vacuum time. Password 558387

2. PROGRAM VERSION
E.g. 050121

3. PHASE
E.g. PRE- TREATMENT

4. CYCLE COUNTER

Possibility to read the sterilizer's counter, pressure


and temperatures. E.g. 0210

Setup menu
5. SETUP MENU

Print Last Cycle: Not used.


System: To access the System menu.
Password: State

System menu: See the section System menu.


About: Software version

MENU TREE, MAIN MENUS


System menu
Navigate to your choice using the UP arrow and
DOWN arrow. The selection is indicated on the
display by a cursor below the value to be changed.

To return to the previous level, press the LEFT arrow


when the menu position is not selectable.

Press the ENTER key to activate your menu


selection.

The sterilizer's menu for installation specific


settings and unique machine settings.
Main selection: Calendar, Calibration,
configuration, Dipswitch, Service and Flash
memory.

1. CALENDAR MENU
Time and date settings

2. CALIBRATION MENU

Only possible with manual calibration


See separate section Calibration menu

3. CONFIGURATION MENU

See separate section Configuration menu

4. DIPSWITCH MENU
Not used.

5. SERVICE MENU
See separate section Service menu

6. FLASH MEMORY MENU

Save adjustments and changes to the flash memory.


Should always be done once adjustments and
calibrations have been made.

MENU TREE, SYSTEM MENU, ETC


Calibration menu
Main selection: Manual calibration, Automatic ca-
libration and Compensation table.

Navigate to your choice using the UP arrow and


DOWN arrow. The selection is indicated on the dis-
play by a cursor below the value to be changed.

To return to the previous level, press the LEFT arrow


when the menu position is not selectable.

Press the ENTER key to activate your menu selec-


tion.

1. MANUAL CALIBRATION
Always use MANUAL CALIBRATION. To calibrate,
see the section Advanced trouble shooting, Cali-
bration
00 Not used
01 Not used

02 Not used
03 Pressure: Step the cursor to each value to be
changed. Verify with Enter

04 Not used
05 Not used
06 Not used

07 Load sensor: Step the cursor to each value to be


changed. Verify with Enter

2. AUTOMATIC CALIBRATION

Not used

MENU TREE, CALIBRATION MENU


Configuration menu
Main selection: Calendar, Calibration, configura-
tion, Dipswitch, Service and Flash memory.

Navigate to your choice using the UP arrow and


DOWN arrow. The selection is indicated on the dis-
play by a cursor below the value to be changed.

To return to the previous level, press the LEFT arrow


when the menu position is not selectable.

Press the ENTER key to activate your menu selec-


tion.

1. LANGUAGE DATE UNIT

Settings related to countries

Language setting: Display language

Date format

Pressure format: NOTE! Must not be changed! The


sterilizer is calibrated based on bar.

Temperature format: NOTE! Must not be changed!


The sterilizer is calibrated base don degrees Celsius.

2. PRINTER

Printing Mode: Not used.

Logging interval low: Logging interval for the total


process excluding the sterilisation phase.

Logging interval high: Logging interval for the sterili-


sation phase only.

3. ALARM CLOCK

Not used

4. PACS ADDRESS

Possibility to give the sterilizer an Id. When there is


more than on sterilizer with the printer in the same lo-
cation. 1-99

5. COMMUNICATION

Settings for the Com-ports (3) RS232 or RS485

Set for communication with the service software.

Must not be changed!

MENU TREE, CONFIGURATION MENU


Service menu
Main selection: Error log, Service messages and
Diagnostics.

Navigate to your choice using the UP arrow and


DOWN arrow. The selection is indicated on the dis-
play by a cursor below the value to be changed.

To return to the previous level, press the LEFT arrow


when the menu position is not selectable.

Press the ENTER key to activate your menu selec-


tion.

1. ERROR LOG

20 most recent error codes

The type of error code, in which phase of the program


and when.

2. SERVICE MESSAGES

Resetting of the service alarm (alarm every 400 proc-


ess)

3. DIAGNOSTICS

Manual test of components. Set the value under MAN


and control with S:1 or S:0

Analog input: Shows the water, pressure and tem-


perature values

Analog output: Not used.

Digital input: Shows the door switch and an external


level switch

Digital output: Check of all components

Test User Flag: Not used

Test System Flag: Not used

Test Printer: Not used

Test LED/buzzer: Functional check of the sterilizer's


audio signal and LEDs.

MENU TREE, SERVICE MENU


Component placement,
internal
1 Circuit board

3 Ejector

4 Solenoid valve MV-3

5 Non return valve BV-1

6 Solenoid valve MV-4

7 Solenoid valve MV-5

8 Solenoid valve MV-6

9 Solenoid valve MV-8

10 Heater

11 Relay

12 Overheating protection, ÖV300

13 Chamber

14 Level guard

15 Solenoid valve MV-1

16 Absolute pressure sensor

17 Pressure relief valve (Chamber)

18 VT/PT-outlet

19 Pressure relief valve (Door gasket)

20 Solenoid valve MV-2

21 BV-2

22 BV-3

COMPONENT PLACEMENT, INTERNAL


23

24

COMPONENT PLACEMENT, INTERNAL


COMPONENT PLACEMENT
Component placement,
outside
1. Water supply

2. Water connection

3. Water outlet

4. Handle

5. Control panel

6. Reset button, overheating protection

7. Printer port RS232 (accessory)

8. Mains supply

9. Door lock

COMPONENT PLACEMENT, OUTSIDE


CIRCUIT BOARD DIAGRAM

X1 Absolute pressure sensor


X2 Temperature sensor
X3
X9
X10:A1 Switch (Door)
X10:A3 Level guard water
X16:2 L1 (Phase)
X16:3 Fan
X16:4 MV-1
X16:5 MV-2
X17:3 MV-5

X17:4 MV-4
X17:5 MV-3
X18:3 MV-6
X18:4 Relay (Heater)
X18:5 MV-8
X19
X20
X21
X23
X23:1 L1 (Phase)
X23:2 Neutral
X24 PC communication
X25
X26 Display
X27
X32

CIRCUIT BOARD DIAGRAM


WIRING DIAGRAM

CIRCUIT BOARD DIAGRAM


INSTALLATION
INTERNAL TRANSPORT
CAUTION!
Installation must be carried by The easiest method to move the sterilizer is using a
technicians with documented pallet truck or a barrow with a minimum lifting limit of
experience. 120 kg.

UNPACKING
Check that the packaging is intact.
Check that the supplied goods correspond with the
order.
Check that the sterilizer is free from defects.
Transport damage must be reported immediately to the
carrier.

Setup
WARNING! PREPARATIONS
Make sure that the area where the Setup ought to be carried out by technicians with
sterilizer is to be located can documented experience.
withstand a load of at least 250 kg.
The sterilizer must be located indoors with a maximum
ambient temperature of 35 °C and a relative, no
condensing, humidity of maximum 95 %.
CAUTION!
PLACEMENT
Ambient temperature, max 35 °C.
Ambient humidity, max 95 %. The sterilizer is a floor model and must stand free, it
must not be enclosed. It should be positioned so that
loading and unloading can be performed without risk. It
must be possible to open the door fully. See the
recommended distances in the picture
Follow local safety regulations.
Once the placement of the sterilizer has been
established, anchor the sterilizer to the floor by each of
the three feet. This prevents the sterilizer from tipping. A
13 mm hole has been drilled in each foot. Use a screw
and plug intended for the surface in question. Make sure
there is a slight backward incline, i.e. the side of the door
should be slightly lower than the front, about 2 - 3 mm.
This is accomplished by placing a washer under each of
the front feet.

INSTALLATION
VS 90/60 1550mm
VS130 1750mm

130mm

INSTALLATION
Electrical connection
IMPORTANT! Prior to carrying out any connection work, ensure that
The electrical connection of the the characteristics stated on the nameplate correspond
unit must be carried out by trained to the mains supply to which it is to be connected. In
personnel case of uncertainty as to the unit supply voltage,

VOLTAGE FLUCTUATION:
The electrical components tolerate a supply voltage
fluctuation of between -10 and +10 %. For larger
fluctuations, it is necessary to use a voltage stabilisation
system placed between the circuit-breaker and the
processor unit transformer. In this case, contact the
sales office.

POWER-SURGE PROTECTION:

A power-surge protection device (over current


protection) must be placed on the sterilizer’s AC power
supply. This protection is the responsibility of the
customer.

CIRCUIT-BREAKER:
A device enabling the sterilizer to be electrically isolated
must be inserted on all the live conductors (including
neutral). This switch or circuit-breaker must be part of
the building’s electrical equipment, positioned near the
unit, and must be easily accessible and marked as the
power switch for the unit, while stating its function.
If the unit is connected in polyphase. Its characteristics
are as follows:
• Breaking power 63 A breaking the 3 phases as well as
neutral
Earth terminal making it possible to connect the isolating
device to the earth conductor without braking it.
The on and off positions must be marked: I on and O off

• Box IP 54

CAUTION! TYPE- Power (kW) Voltage (V) Current (A)


The sterilizer must be declared to VS 60 - 130 6.1KW 230V Single phase 30A
your electricity supplier in case of
a 230 single phase connection. VS 60 - 130 6.1KW 230X3V 20A
VS 60 - 130 6.1KW 400X3V + N 16A
Connect the sterilizer to the earth and according to the
specification of the manual. The installation must be
protected in accordance with the regulations in force.
Failure to comply with this instruction may result in
electrocution of the user and damage to the unit.

CONNECTION POWER SUPPLY


Open the front box.

INSTALLATION
400V 3 phases connection
5-wire cable (3 phase + Neutral + earth)
No jumper

230V 3 phases connection


4-wire for 230V (3 phase (3 phase + earth)
Jumper between L2-N

230V single phase connection


3 wire (1phase + Neutral + earth).
Jumper between L1-L2-L3

With appropriate section to 60227 and 60245.


Pass the power cable through the rear housing provided
for this purpose.
Connect the power cable earth connector to the terminal
WARNING! marked with the earth symbol
Never work with energized
Connect the other end of the cable to the circuit-
equipment.
breaker/cut-off device located on the outside.
The sterilizer must be earthed.
Change of voltage:

This adaptation requires certain parts to be changed;


use the “Jumper kit” furnished with the machine to
perform this change. For further information see the
electrical diagram, or get in touch with the selling
company.
Terminal connection:
Also see “Wiring diagram X1”
400 V 3 Phases Jumper between E-G-H
230 V 3 Phases Jumper between A-G, B-E, C-H

230 V Mono Phase Jumper between N-A-B-C

INSTALLATION
Water connection
The incoming water must have a pressure between 1.5
CAUTION! bar and 6 bar with a temperature of about 15 °C.
The water supply must be fitted Water quality <4 dH
with a shutoff tap, as the sterilizer
must not be left without The sterilizer is connected to a ½" water connection with
supervision with the water a shutoff tap. The sterilizer must not be connected with
pressurized. hoses less than 12 mm. The hoses supplied with the
sterilizer are fitted with Gaz ½" G connections. The
hoses (two) are labelled “Inlet” and “Drain”.
WARNING! The drain should be connected to a 6" well, with a
Risk of hot water from the drain capacity of 10 litres/min that can withstand 100° C.
hose, exercise the greatest Condensation can occur.
possible caution.
The drain hose must end at least 3 cm above the water
surface in the well.
The drain hose must not go higher than to the lower
edge of the chamber.

Diverse
Should the sterilizer be transported in temperatures
below -5 °C, the overheating protection on the sterilizer
may need to be reset. The error code Heating error
when a process is started indicates a symptom for this.

Keep all instructions and description of operation close


to the sterilizer.
Make sure the operator has read the User manual.

All other documentation supplied with the sterilizer must


be looked after by the operator. The documentation shall
be kept with the sterilizer until its service life comes to an
end.

INSTALLATION
TECHNICAL DATA
Dimensions/weight
Width/Depth 570/700 mm

Height VS60/VS90/VS130 1030/1030/1230 mm


Load height VS60/VS90/VS130 940/940/1140 mm
Weight VS60/VS90/VS130 100/106/115 kg

Operating weight VS60/VS90/VS130 107/113/122 kg

Capacity
Chamber, diameter 444 mm
Chamber, depth VS60/VS90/VS130 384/590/858 mm

Chamber, volume VS60/VS90/VS130 60/90/130 L


Instrument load VS60/VS90/VS130 12/24/36 kg

Electrical specifications
Electrical connection according to rating plate

Control voltage according to rating plate


Total output 6 kW (3 x 2 kW)
Power consumption 5 A (3 x 400 V), 9 A (3 x 230 V)

Heat emission while operational, closed door about 615 W


Heat emission in standby, open door 0W

Other data
Chamber, calculation pressure 3.0 bar
Chamber, provi sions AFS 1999:4 (ÅKN 71), Isbest PED 97/23 EEC)

Chamber/door, material EN 1.4404, AISI 316L


Water quality Max 4dH
Water pressure 1.5 - 6 bar

Water and drainage connections ½”


Water consumption during steam phase 7 litres
Water consumption per process 75 - 100 litres

Sound level less than 53 dB


Not intended for the USA market

TECHNICAL DATA
FEEDBACK
Getinge Skärhamn works continuously to improve its
Type designations, select
products.
• ? VS60
Would you like to help us by submitting your points of
• ? VS90 view, tips or criticism? Copy this page and send or fax
• ? VS130 it to us.
All suggestions that result in the improvement of our
Serial number:
product will be rewarded. Getinge Skärhamn reserves
_______________________________ the right to assess whether the suggestion is an im-
provement
Points of view, tips or criticism:

_______________________________ Points of view were submitted by:

_______________________________ Name .................................................


Telephone: .................................................
_______________________________
Address: .................................................
_______________________________
.................................................
_______________________________ .................................................

_______________________________ E-mail: .................................................

_______________________________
Send to:
_______________________________
Getinge Skärhamn AB
_______________________________
Industrivägen 5
_______________________________ SE-47131 Skärhamn, Sweden

_______________________________ Fax:+46 (0)304 600 229


E-mail: info@seska.getinge.com
_______________________________

_______________________________

_______________________________
Thanks in advance.
_______________________________

_______________________________

_______________________________

_______________________________

_______________________________

_______________________________

_______________________________

_______________________________

_______________________________

FEEDBACK
Change Sign. Approved by - Date
12 New revision om pos 9 and 11 skaal

11

10

14
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person

5
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

9 7 6
14 1 48330307 B TRYCKGIVARE / PRESSURE SENSOR
13 13 1 48330318 TILLBEHÖR / ACCESSORIES VS
12 1 48330316 ÖVERKÅPA / TOP COVER VS
11 1 48330314 A KÅPA / COVER VS
10 1 48330308 DÖRRDETALJER / DOOR COMPONENTS
9 1 48330309 B ISOLERKÅPA / COVER
8 1 48330301 A KAMMARE / CHAMBER
7 1 48330312 B ELLÅDA / DISTRIBUTION BOX
6 1 48330311 A KRETSKORT / PRINT BOARD
1 st angel projection
Vyplaceringsmetod

5 1 48330313 A DÖRRVENTIL / DOORVALVE


4 1 48330306 C LUFTPAKET / AIRVALVE
3 1 48330305 C EJEKTOR / EJECTOR
2 1 48330303 C SÄKERHETSVENTIL/SAFETY VALVE
1 1 48330304 A VATTENPAKET / WATER PACKAGE
Pos Qty Drawing no. Rev Description
Designed by Checked by Date Replaces Replaced by Status
AL 2005-10-03

Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A2 VS D 1/


1
Change Sign. Approved by - Date
Pos 30 and 37 new AL
4
19 13 12 33 32 23 22 11 9 A ( 0,60 : 1 )

26 24

5
38

36
29
24
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person

18
21

31
20
A
17
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

38 8 483435803 BRICKA / WASHER AZ 4,1X8,2X0,5


37 6 483430328 SKRUV / SCREW M5x8 FZB
36 1 566692301 SLANG / HOSE
35 2 4836111 STIFT/CABLE CLIP
34 1 48325023 ISOLERING / INSULATION
33 1 483436601 BRICKA / WASHER 8X25X1,5 FZB
32 1 48320174 Bricka / Washer
16 31 1 48325006 KONSOL DÖRRBRYTARE
30 6 483435804 BRICKA / WASHER AZ 5,1X9,4X0,6
10 29 1 483430305 SKRUV / SCREW MRX 3X12 FZB
28 1 483433505 MUTTER / NUT M6M 5.6 M6 FZB
28 27 1 483435805 BRICKA / WASHER AZ 6,4X11,2X0,7
26 1 483433502 MUTTER / NUT M6M 5,6 M3 FZB
25 2 483433503 MUTTER / NUT M6M 5,6 M4 FZB
24 4 483269502 MUTTER / NUT M6M 5.6 M6 FZB
23 1 483435311 BRICKA / WASHER 8,4X16X1,5 FZB
2 30 37 22 1 725537540 INSEXSKRUV M6X30 / SCREW M6X30
21 1 483430306 SKRUV / SCREW MRX 3X16 FZB
20 1 4835503 MIKROBRYTARE / MICROSWITCH
27 1 19 1 4836101 NIVÅVAKT / LEVEL SWITCH
14 18 1 566782504 BRICKA TEFLON / WASHER
17 1 4834056 NIPPEL UTV / NIPPLE 1/2" x 3/8"
16 1 4835642 VINKEL / BOW 1/2"
25 15 15 1 4834762 PLUGG / PLUG 1/2"
14 1 4836097 VINKEL / BOW C5 8x1/8"
13 1 4836116 KULA M6X25 / BALL M6X25
12 1 566713801 WIRE
11 1 566716601 DELNINGSBULT / SCREW
10 1 566791801 SIL / STRAINER
9 1 566687001 DÖRR PACKNING / DOOR GASKET
3 8 1 568508472 KAMMARE / CHAMBER 130L
35
7 1 568508471 KAMMARE / CHAMBER 90L
6 7 8
6 1 568508470 KAMMARE / CHAMBER 60L
1 st angel projection
Vyplaceringsmetod

5 1 566718301 PACKNING / GASKET


4 1 566720001 ÖVERHETTNINGSSKYDD/OVERH PROTECT
3 1 566790901 KÅPA / COVER
2 1 566687201 MUTTER / NUT M45
1 1 48325017 ELEMENT
Pos Qty Drawing no. Rev Description
Designed by Checked by Date Replaces Replaced by Status
AL 2005-09-26

KAMMARE / CHAMBER
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A2 48330301 A 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person Pos 2,3,4 and 5 is new skaal

1
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

H10
3

4 2
5
1 st angel projection
Vyplaceringsmetod

5 1 4835714 PLUGG / PLUG 12mm


4 1 4835732 T-KOPPLING / T-PIECE
3 1 4836151 KOPPLING / COUPLING 8X1/2"
2 1 4836150 T-KOPPLING / T-PIECE
1 1 4836088 SÄKERHETSVENTIL/SAFETY VALVE
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
skaal 2004-12-06

SÄKERHETSVENTIL/SAFETY VALVE
Frame Drawing no. Revision Sheet

©COPYRIGHT GETINGE AB 2001 A3 48330303 C 1/


1
Change Sign. Approved by - date

Sida med vit H12


This drawing remains our property and may not be copied or made the base of production.shown or given to a third person

plastkula placeras
mot pos 6.
H11 5 12
11
2

3
6

A P
9
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

8 H2
4 MV-4 H5
MV-5 2

7 9

H6
MV-3 1

2
H13

12 2 725537640 SKRUV / SCREW 6X35 A2-70


11 1 4836123 T-KOPPLING / T-PIECE 8/6
9 2 4835820 VINKEL / BOW 1/2" -15
8 1 4836096 VINKEL / BOW 8xINV1-4
7 1 566715501 SÄKERHETSVENTIL/SAFETY VALVE
6 1 4834189 T-KOPPLING / T-PIECE
1 st angel projection
Vyplaceringsmetod

5 2 4834037 NIPPEL/NIPPLE 1/4 X 1/4


4 1 4834100 MAGNETVENTIL / SOLENOID VALVE
3 2 4834492 MAGNETVENTIL / SOLENOID VALVE
2 3 4835581 NIPPEL UTV / NIPPLE 1/2" x 1/4"
1 1 48325002 FÖRDELARSTYCKE / DISTRIBUTION
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
skaal 2004-12-06

VATTENPAKET / WATER PACKAGE


Frame Drawing no. Revision Sheet

©COPYRIGHT GETINGE AB 2001 A3 48330304 A 1/


1
1 2 3 4 5 6

10
H6
13 H8 17
A A

MOUNTED UP ON VS130
4
H5
6

1 MV-1
5
B 4 B
H4 H3
P A

H7 H1 H1 H2

8 14 9
3 7
12
2 18 16 18 2 483430524 SKRUV / SCREW MC6S 6X 14 A2
17 1 4836143 VINKEL / BOW 15-15
11 15 16 2 483435803 BRICKA / WASHER AZ 4,1X8,2X0,5
15 2 483430309 SKRUV / SCREW MRX 4X 10 FZB
C 14 1 4836123 T-KOPPLING / T-PIECE 8/6 C
13 1 4835820 VINKEL / BOW 1/2" -15
12 1 4834039 NIPPEL/NIPPLE 3/8
11 1 48320172 KONSOL MV / BRACKET
10 1 48320171 EJEKTOR / EJECTOR
9 1 4836099 T-KOPPLING / T-PIECE 8/6
8 1 4834037 NIPPEL/NIPPLE 1/4 X 1/4
7 1 4834189 T-KOPPLING / T-PIECE
6 2 4835814 BACKVENTIL/NON RETURN VALVE
5 1 4834056 NIPPEL UTV / NIPPLE 1/2" x 3/8"
4 2 4835581 NIPPEL UTV / NIPPLE 1/2" x 1/4"
3 1 4834207 T-KOPPLING / T-PIECE
2 1 4834038 NIPPEL/NIPPLE 3/8 x 1/4
1 1 4834100 MAGNETVENTIL / SOLENOID VALVE
Pos Qty Drawing no. Rev Description Note
D Designed by Checked by Date Replaces Replaced by Status D
skaal 2004-11-30

EJEKTOR / EJECTOR
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330305 C 1/


1
1 2 3 4 5 6
1 2 3 4 5 6

Sealing = SGS676
A A

MV-2

11 7
5

A 4

2 7
P
B MV-6 12 B

H10

P H9
1 6

2 9
A
C 8 C

H3 12 2 483430524 SKRUV / SCREW MC6S 6X 14 A2


11 1 4839254 BRICKA / WASHER
3 10 1 4835804 VINKEL / BOW C5 8x1/4"
10 9 2 483435803 BRICKA / WASHER AZ 4,1X8,2X0,5
8 2 483430309 SKRUV / SCREW MRX 4X 10 FZB
7 2 4834038 NIPPEL/NIPPLE 3/8 x 1/4
6 1 4836123 T-KOPPLING / T-PIECE 8/6
5 1 48320172 KONSOL MV / BRACKET
4 1 4834687 BACKVENTIL / NONE RETURN VALVE
3 2 4834100 MAGNETVENTIL / SOLENOID VALVE
2 2 4834037 NIPPEL/NIPPLE 1/4 X 1/4
1 1 4834189 T-KOPPLING / T-PIECE
Pos Qty Drawing no. Rev Description
D Designed by Checked by Date Replaces Replaced by Status D
skaal 2004-12-06

LUFTPAKET / AIRVALVE
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330306 C 1/


1
1 2 3 4 5 6
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person or a competetive firm without

H14 7

5
our permission.Getinge AB S-310 44 GETINGE, SWEDEN

H14
3
2

7 1 483431006 SKRUV / SCREW RXS B10X9,5


6 3 4836111 STIFT/CABLE CLIP
5 1 4835751 FÄSTPLATTA / CLAMP PLATE
4 1 4835803 KOPPLING / COUPLING 4XM5
1 st angel projection
Vyplaceringsmetod

3 1 4835734 KOPPLING / COUPLING 4X1/8"


2 1 4834206 BUSSNING / BUSHING
1 1 4835997 TRYCKGIVARE / TRANSMITTER
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
skaal 2004-12-06

TRYCKGIVARE / PRESSURE SENSOR


Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A4 48330307 B 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person
21
10
28
2 27 26 20

22

23
3
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

25
1

10 5

4 14
28 2 483430802 PINNSKRUV / STUD SCREW M6X57
27 2 483430804 PINNSKRUV / STUD SCREW M6X80
26 4 483269502 MUTTER / NUT M6M 5.6 M6 FZB
25 4 483433505 MUTTER / NUT M6M 5.6 M6 FZB
24 1 4836115 KULA / BALL M10 X 40
8
23 1 566717901 FJÄDER / SPRING
22 1 566783301 RÖR / PIPE
6 21 1 483529601 MUTTER LÅS / CHECK NUT M8 A2
18 17 16 20 2 566782501 BRICKA TEFLON / WASHER
19 1 566717801 FJÄDER / SPRING
18 1 566782601 RÖR / PIPE
17 2 566782505 BRICKA TEFLON / WASHER
15 16 2 735311809 SPÅRRING / LOCKING WASHER
15 1 566798001 AXEL / AXIS
14 1 566715470 LUCKFÄSTE / BRACKET
13 2 566714201 GÄNGAD BYGEL / CLAMP
12 4 4832695 MUTTER FLÄNS / NUT M4 A2
12
11 1 48325021 LUCKARM / HOOD HANDLE
10 4 483436303 BRICKA / WASHER 8,4X16
9 2 731431240 MUTTER / NUT M6M M8 A2
19
8 4 483434708 MUTTER / NUT M6M 5.6 M8 A2
13
7 1 566714801 FÄSTE / BRACKET
6 2 566714202 GÄNGAD BYGEL / CLAMP
1 st angel projection
Vyplaceringsmetod

5 1 566720101 VEVSTAKSHUVUD / ROD END


4 2 483430527 SKRUV / SCREW M6S 8X20
8 3 3 483436302 BRICKA / WASHER SS 3576 6,4X12 RFR
7 10
9 2 3 725537540 INSEXSKRUV M6X30 / SCREW M6X30
11 1 1 566717701 LUCKFÄSTE / HOLDER DOOR
Pos Qty Drawing no. Rev Description
Designed by Checked by Date Replaces Replaced by Status
AL 2005-09-29

24 DÖRRDETALJER / DOOR COMPONENTS


Frame Drawing no. Revision Sheet

©COPYRIGHT GETINGE AB 2001 A3 48330308 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person Pos 17 is new AL 2005-11-22

4
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

5 6

1 2

9
7 8
8

17 1 4836162 FLÄKTSKYDD / FAN GUARD


10 16 2 483431006 SKRUV / SCREW RXS B10X9,5
15 1 4835614 FLÄKT / FAN
14 1 48325040 KONSOL FLÄKT / PLATE FAN
13 13 2 4835900 SLANGKLÄMMA/HOSE CLAMP 15X24
12 1 4835449 SLANGNIPPEL / HOSE NIPPLE
11 1 4836122 SLANG AVLOPP / HOSE OUTLET
11 10 1 4834210 SLANGNIPPEL / HOSE NIPPLE
12 9 3 483435805 BRICKA / WASHER AZ 6,4X11,2X0,7
8 7 483430524 SKRUV / SCREW MC6S 6X 14 A2
15 14 16 8 7 13
7 2 560785904 GUMMIKLAMMER / CLAMP
6 1 48325019 KONSOL WIRE / BRACKET
1 st angel projection
Vyplaceringsmetod

5 19 483438601 POPNIT / RIVET 4,0X10 A2


17 11 4 19 483590301 NIT MUTTER / RIVET M6
8 3 1 48325016 FOT FÄSTE / BASE
2 1 48325015 ISOLERKÅPA / COVER 130L
1 1 48325020 ISOLERKÅPA / COVER 60/90L
Pos Qty Drawing no. Rev Description
3 Designed by Checked by Date Replaces Replaced by Status
AL 2005-09-29

ISOLERKÅPA / COVER
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330309 B 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person

3
2
4
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

1 9 11 483430309 SKRUV / SCREW MRX 4X 10 FZB


8 3 483431011 SKRUV / SCREW RXS B8X9 5 FZB
7 2 483433302 SKRUV / SCREW RXB B8X13 FZB
6 1 483556102 K-KANAL/CABLE PIT 18X18
1 st angel projection
Vyplaceringsmetod

5 1 483435803 BRICKA / WASHER AZ 4,1X8,2X0,5


4 1 4836113 HYLSHUS / HOUSING 4 POL
3 1 4836110 HYLSHUS / HOUSING 20 POL
2 1 501931700 KRETSKORT / PRINT BOARD PACS
1 1 48325014 ELLÅDA / DISTRIBUTION BOX
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
AL 2005-02-18

KRETSKORT/PRINT BOARD
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330311 A 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person Infört byglar skaal
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

14

9 6
13

2 12
11
10

14 1 570110027 BYGEL / JUMPER 230V / 1PHASE


1 13 3 570110022 BYGEL / JUMPER 230V / 3PHASE
12 1 483578605 BYGEL / JUMPER 400V / 3PHASE
11 1 4836109 MUTTER / NUT M16
10 2 483431003 SKRUV / SCREW RXS B8X19 FZB
8 9 1 483579102 KABELFÖRSKRUVN/ CABLE FITTING
8 2 483579205 MUTTER / NUT 28
7 2 483579105 KABELFÖRSKRUVN/ CABLE FITTING
7 5 6 1 4836103 KONTAKTOR / CONTACTOR
1 st angel projection
Vyplaceringsmetod

5 2 483578604 KOPPLINGSPLINT/TERMINAL
3 4 1 483578603 KOPPLINGSPLINT/TERMINAL
3 6 483578602 KOPPLINGSPLINT/TERMINAL
4
2 1 48325007 DIN SKENA / RAIL
1 1 4836102 LÅDAPLAST/PLASTIC BOX
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
skaal 2004-12-13

ELLÅDA / DISTRIBUTION BOX


Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330312 B 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person

1
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

H11

2
P 6
A

H4

5 3

6 2 483430524 SKRUV / SCREW MC6S 6X 14 A2


1 st angel projection
Vyplaceringsmetod

5 2 483430309 SKRUV / SCREW MRX 4X 10 FZB


4 2 483435803 BRICKA / WASHER AZ 4,1X8,2X0,5
3 1 48320172 KONSOL MV / BRACKET
2 2 4835617 KOPPLING / COUPLING 8X1/4"
1 1 4834100 MAGNETVENTIL / SOLENOID VALVE
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
AL 2005-04-06

DÖRRVENTIL / DOORVALVE
Frame Drawing no. Revision Sheet

©COPYRIGHT GETINGE AB 2001 A3 48330313 A 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person Pos 11 and 12 is removed. Pos 14 is new. SH

3 2 7 1 6 5
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

10 9 4 8

14 5 483269502 MUTTER / NUT M6M 5.6 M6 FZB


13 4 483430524 SKRUV / SCREW MC6S 6X 14 A2
10 3 483433503 MUTTER / NUT M6M 5,6 M4 FZB
9 3 483910601 BRICKA / WASHER 4,3X9,0X0,8
8 3 483910504 DISTANS / DISTANCE 50-140 14mm
7 1 4836137 KOPPLING / COUPLING 8x1/4"
6 1 4835746 MANOMETER
1 st angel projection
Vyplaceringsmetod

5 1 48325001 DEKAL / DECAL VS 60,90,130


4 1 4839690 DISPLAY
13 3 1 48325009 KÅPA / COVER VS 130
2 1 48325026 KÅPA / COVER VS 60/90
1 1 48325011 KÅPA ÖVERDEL / TOP COVER
Pos Qty Drawing no. Rev Description
Designed by Checked by Date Replaces Replaced by Status
SH 2005-09-15

KÅPA / COVER VS
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330314 A 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person

2 1
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN
1 st angel projection
Vyplaceringsmetod

3
3 1 48325028 DEKAL / STICKER OPEN - CLOSED VS
2 4 483562102 MUTTER / NUT M6
1 1 48325018 LOCK / COVER VS 60,90,130
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
SH 2005-09-19

ÖVERKÅPA / TOP COVER VS


Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330316 1/


1
Change Sign. Approved by - date
This drawing remains our property and may not be copied or made the base of production.shown or given to a third person
or a competetive firm without our permission.Getinge AB S-310 44 GETINGE, SWEDEN

1 2 3

5
1 st angel projection
Vyplaceringsmetod

5 1 567169101 FÄSTE / HOLDER


4 1 566714401 BOTTENPLÅT / BOTTOM PLATE VS
3 3 566687401 KORG / BASKET VS 60,90,130 VS 130
2 2 566687401 KORG / BASKET VS 60,90,130 VS 90
1 1 566687401 KORG / BASKET VS 60,90,130 VS 60
Pos Qty Drawing no. Rev Description Note
Designed by Checked by Date Replaces Replaced by Status
SH 2005-09-26

TILLBEHÖR / ACCESSORIES VS
Frame Drawing no. Revision Sheet

COPYRIGHT GETINGE AB 2001 A3 48330318 1/


1

You might also like