NPC Metrology Laboratory
CERTIFICATE OF CALIBRATION
CERTIFICATE NO.: 2019-026-002
for
Micro-Ohmmeter
Schuetz Messtechnik Model MR5-200C
Serial No.: 1237-5884
Submitted by:
EURO-ASIA ENERGY SOURCE INCORPORATED
2nd Flr. No. 265 Unit N Cortes Bldg. Cor.
Set. Fernadez & Tomas Morato Sts.,
Brgy. Laging Handa, Quezon City
This certifies that the instrument was calibrated by comparison against the standards
maintained by the NPC-Central Metrology Laboratory using procedure PES-ESD-PM-012-
WI-029.
‘The standards used aretraceable to the National Metrology Laboratory of the Department of
Science and Technology and to the National Institute of Standards and Technology through
Ark One Solutions, Ine.
The test data in the attached Calibration Report were those obtained during the time of
calibration.
Itis the responsibility of the user to determine the suitability of the instrament for its intended
use,
‘Testing and Metrology Services Division
Date: 22 April 2019
Page 1 of 2
Quezon Avenue corner BIR Road, Diliman, 1100 Quezon City, Philippines
Tel. Nos. (632) 921-3541 1079 ® FaxNo. (632) 921-2468 P.O. Box ci
Website: www.napocor.gov.phNPC Metrology Laboratory
CALIBRATION REPORT
2019-026-002
Page _2 of 2
Reference: _WP-ESD-TMD-019-026 Date: April 3, 2019
Description: _Micro-Ohmmeter Make: —_Schuetz Messtechnik,
MR5-200C
Serial No: 1237-5884 TagNo.: N/A
Resistance Measurement Test Results
RANGE stD, MEAS. NOMINAL AS FOUND (UUT) AS LEFT (UUT)
INPUT UNCERT. TOLERANCE READING DEV. ~—sREADING _ODE.
(wo) wa) 2) (ua) ua) ua) (uo) (ua)
200 ng t oan 5.396 113.2 -0.2 113.2 -0.2
(ma) @) (ua) (mo) (mo) (mo) (mo) (ma)
2 0.001 + 0.040 + 0.050 1.000 0.000 1.000 0.000
20 0.001 0.040 + 0.230 1.00 0.000 1.00 0.000
0.009960 + 12948 + 0.499 9.95 0.0103 9.95 0.0103
@ @ (mia) (ma) (ma) fare) (mo) (ma)
0.2 on £ 00173 + 5.0 99.3 o7 99.3 07
@) @) (mo) (a) @ @ @ @
2 on * 00173 + 0.023 0.099 0.001 0.099 0.001
1.0 + 0015 + 0.030 0.999 001 0.999 0.001
‘Test Standards
Description Tag No. Calibration Due Date __ Calibration Traceability
Resistor , EMORO/3i6 3L-SR-or 7/26/2019 ‘Ark One/NIST, ITDI/NML
Shunt Resistor, EIS/SR300 2L-SR-07 8/14/2020 Ark One/NIST, ITDI/NML
Current Shunt, Fluke/Y5020 2L-SH-o1 2/11/2020 Ark One/NIST, ITDI/NML
Standard Resistor, L&N/4015B 2L-SR-01 2/8/2020 ITDI/NML.
Standard Resistor, L&N/4020B 2L-SR-02 2/8/2020 TTDI/NML
Test Conditions:
Temperature: 23 42 °C Relative Humidity: < 70% RH
Test Method:
Measurement of the preset resistance of Standard Resistors and Current Shunts.
Remarks:
Readings are within tolerance limits,
Declaration and Notes:
1 This report applies only to the instrument identified above (Make, Model, Tag No, and Serial No),
and the data contained herein are those that were obtained at the time of calibration,
2 This report may be reproduced only in full and withthe permission from this Laboratory.
3. The quoted uncertainty ofthe standard value is obtained at 95% confidence level
Test Performed by: Reviewed by, = _—
Gemelo L. Ballon Castro
Technician Prificipa) Engineer C