You are on page 1of 2

Technical Details G.

Heinemann Medizintechnik GmbH

Power supply 230 V


Frequency: 50 Hz (optional 60 Hz)
Current consumption: max. 10 A
Power consumption: max. 2,3 kVA
Classification acc. 93/42/EEC: IIa
Protection class: I
Electrical measurements according to DIN
every 12 moths
EN 62353:
Impendance of protective earthing system: max. 0,3 Ohm
Leakage enclosure current: max. 0,5 mA
B: ear irrigation handle, suction handle;
BF: compressed air handle, cold light source (Caution: BF
Degree of protection (insulation class)
properties are only given for the cold light source when the cold
(applied parts):
light source is used in conjunction with an insulating cold light
cable)
IP-Code: IP20
F1 = 10 AT (L)
F2 = 10 AT (N)
F3 = 2 AT (suction, Sekretomatik)
F4 = 8 AT (ear irrigation)
F5 = 3 AT (mirror heater, compressor)
F6 = 3 AT (Int. power supply)
Fuses:
F7 = 6 A (24 V) (circuit breaker)
F8 = 10 A (15 V)
F9 = 10 A (12 V)
F10 = 2 A (15 V)
F11 = 3 A (12 V)
F12 = 1 A (6 V)
Noise level: 50 - 56 dB
Compressor, oil free, 12 l/min, 2700 rpm and 138 VA, max. 2.5
Compressed air (Option):
bar; optional 0.2-2.5 bar adjustable
Vacuum pump, oil free, 40 l/min at -0,85 bar / -85 kPa and 150
Suction (Option):
VA (60 l/min optional)
max. 1500 VA (nonstop operation), min. water supply pressure
Water heater (Option): 1.5 bar, max. water supply pressure 4.0 bar, integrated water
filter 140 µm
Water temperature (Option): 37° C +/-3° C, flow rate: max. 500 ml/min, adjustable
Water tank (Option): 5 liter
Light source, halogen (Option): 150 VA, 15 V
Light source, LED (Option): approx. 30 VA, 12 V
Endoscope holder (pre-heated) (Option): 37° C ± 4° C
Instrument tray (pre-heated) (Option): 37° C ± 4° C
Mirror heater: 350 VA
UVC instrument tray light bulb (Option): 12 VDC, 14 VA

HM_VT_Techn-Details-MODULA_17062015_00 Seite 1 von 2


Technical Details G. Heinemann Medizintechnik GmbH

Electrical connection: Power cable: 2 m


Water connection: 3/8", water pressure: 1,5 - 4 bar
Drainage: Ø 40 mm, height above the ground: max. 50 cm
Temperature: -20 to +50° C, Humidity (without condensation):
Ambience conditions (transport, storage):
10 to 95%, air pressure 700 to 1060 hPa
Temperature: +5 to +35° C, Humidity (without condensation):
Ambience conditions (operation):
30 to 65%, air pressure 700 to 1060 hPa
MODULA Duo: 1870 x 970 x 545 mm
MODULA Europa: 1123 x 970 x 545 mm
Dimensions (WxHxD):
MODULA Paris: 915 x 970 x 545 mm
MODULA mini: 410 x 790 x 585 mm
MODULA Duo: max. 250 kg
Total netto weight (depending on equipment MODULA Europa: max. 160 kg
options): MODULA Paris: max. 140 kg
MODULA mini: max. 60 kg
MODULA Duo: 140x120x80 cm (case 1), 80x120x80 cm (case
2)
Dimensions of the transport case (WxHxD): MODULA Europa: 140x140x80 cm
MODULA Paris: 120x120x80 cm
MODULA mini: 80x120x80 cm
MODULA Duo: max. 360 kg
Total brutto weight with the transport case MODULA Europa: max. 270 kg
(depending on equipment options): MODULA Paris: max. 230 kg
MODULA mini: max. 120 kg
CE-sign: 0482
IEC 60601-1:1988+A1:1991+A2:1995 (II. Edition)
IEC 60601-1-2:2001 (EN 60601-1-2)
IEC 60601-1-2: 12/2007
IEC 60601-1:2005 (3rd Edition)
Applied standards: IEC 60601-1-18:1996/A1:2000
EN ISO 10079-2:2009
EN 15223-1
EN ISO 13485:2012
EN ISO 14971:2012
GMDN-Code 44380
HS-Code: 90189084

HM_VT_Techn-Details-MODULA_17062015_00 Seite 2 von 2