Monitoring
Mode
O2-Vol.% TI TE
50 60 0,6 1,0
70
40 0,4 0,8 0,8 2 CPAP
25 90 0,2 1,5 15
0,4
21 100 0,1 2x 0,2 30x
IPPV
Insp. Flow V PEEP/CPAP
7,5
40
5 5
10 10
30 50
3 20
The warranty and liability conditions of the general terms and conditions for business transac-
tions of Dräger Medizintechnik GmbH are not extended by this technical documentation.
Safety Regulations:
Reference is hereby made to the observance of the relevant safety provisions, such as the
Medical Equipment Ordinance (Medizingeräteverordnung), the Pressure Container Ordi-
nance (Druckbehälterverordnung), the Technical Rules for Pressurised Gases (Technische
Regeln Druckgase) or the Occupational Health and Safety Provisions (Unfallverhütungsvor-
schriften).
Insofar as reference is made to laws, regulations or standards, these are based on the legal
system of the Federal Republic of Germany.
General
Function Description
1 General 9
2 Block Diagrams 11
2.3 Block Diagram of the Babylog 8000/ Babylog 8000 plus/Babylog 8000 SC 13
3 Airway Monitoring 14
4 Monitoring Functions 17
4.15 Flow Measurement Monitoring (Babylog 8000/ Babylog 8000 plus) .............. 24
10.9 Front PCB (Babylog 8000/Babylog 8000 plus with EL Display) ...................... 48
11 Sensors 54
11.2 Y-Piece with Flow Sensor (Babylog 8000/Babylog 8000 plus) ........................ 54
Test List
1 Test Equipment 58
2 Accompanying Documents 59
7 Functional Check 62
11 Leak Test 69
21 Tested 74
1 Important Information 75
For internal use only. Copyright reserved.
Appendix
Read each step in every procedure thoroughly before beginning any test. Always use the
proper tools and specified test equipment. If you deviate from the instructions and/or
recommendations in this Service Manual, the equipment may operate improperly or unsafely,
or the equipment could be damaged.
This Service Manual does not replace the Instructions for Use. Any use
of the product requires full understanding and strict observation of the
product-specific Instructions for Use.
1 General
Babylog 8000/Babylog 8000 plus has a flow measurement function. The Babylog 8000 SC
can be upgraded to a Babylog 8000 using the "flow measurement conversion kit.
Elektronic unit
pressure AIR
pressure O2
Vol.% O2
Pexsp
Loud speaker
Pinsp
Battery
Fan
O2 ampilfier
O2 sensor
RS232 PCB
Pneumatic Analog
Patiententeil-
Communication PCB
heating
Watch dog PCB Monitoring PCB
PCB
I/O PCB
driver PCB
Pneumatic
Frontadapter PCB
Pneumatic unit
Frontcontroller PCB
Potentiometer field
Monitoring field
Front
Display PCB
For internal use only. Copyright reserved.
LC display
F61733XXT01.fm 24.01.02
Elektronic unit
pressure AIR
pressure O2
Vol.% O2
Pexsp
Loud speaker
Pinsp
Battery
Fan
O2 amplifier
O2 sensor
RS232 PCB
Ccommunication PCB
Patient unit
Analog PCB
Pneumatic
heating
Watch dog PCB Monitoring PCB
Flow PCB
Pneumatic amplifier
PEEP/PIP
I/O PCB
PCB
Frontadapter PCB
Pneumatic unit
Front PCB
Potentiometer field
Monitoring field
ELD converter
Front
For internal use only. Copyright reserved.
EL display
Fig. 2: Block diagram of the Babylog 8000/Babylog 8000 plus with EL display
F61733XXT01.fm 24.01.02
Potentiometer
keys
Front
Communication Battery
E E
V V
Flow sensor
E E E E E E
P P P P O O
+27 V +5 V
PAIR PO2 Pexsp Pinsp O2-Sensor
&
24 Ventile
Valve amplifier
PEEP/PIP valve
Valve monitoring
In the ventilation modes IPPV/IMV and CPAP, the airway pressure (Paw), the flow at the Y-
piece V T (Babylog 8000/Babylog 8000 plus), and the fraction of inspired O2 (FiO2) are
measured.
Depending on the selected mode and the menu settings, the parameters are monitored and
displayed on the screen. Pressure and flow curves as well as their storage can be displayed
graphically on the LC display or EL display. In all ventilation modes, the chronological
sequence of the airway pressure (Paw) is displayed in the LED bargraph.
Two internal pressure sensors measure the pressure at the inspiratory outlet (Pinsp) and the
pressure at the expiratory inlet (Pexp). The airway pressure is calculated as follows:
The following pressures are calculated from the Paw pressure signal:
− Peak pressure (Peak)
− Mean airway pressure (Pmean)
− Positive end-expiratory pressure (PEEP)
The peak pressure is the maximum positive pressure of the most recent respiratory cycle.
After 30 s, at the latest, a new respiratory cycle must be recognized and a new measured
value for the peak pressure must be generated, otherwise the current measured value is no
longer valid and is removed from the display.
The mean airway pressure is the initial value of a software digital filter.
The PEEP is either the pressure value during the expiratory phase at a zero expiratory flow or
the last measured value before the next inspiration. Like the peak pressure, the PEEP is no
longer valid after 30 s and is removed from the display if a new measured value is not
generated within the 30-s period.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
In order to generate a trigger signal, the inspiratory flow must be integrated during
spontaneous breathing and compared to the adjustable trigger threshold.
A O2 sensor in the inspiration line measures the O2 content of the respiratory gas.
Calibration data of the O2 sensor is maintained after switching off the Babylog. If the
Babylog has been out of operation for more than 24 hours, a calibration will automatically be
carried out upon power-on.
The operator may also initiate a calibration manually, e.g. as required after replacing an O2
sensor.
A two-step calibration with 21% and 100% O2 is always carried out in order to achieve a
higher measurement accuracy over the whole concentration range and/or to be able to
recognize a spent sensor cell.
If the O2 measurement fails, the Babylog generates an alarm. This alarm status is shown in
the status field of the screen by the flashing indication "FiO2".
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
A direction-sensitive hot-wire flowmeter integrated into the Y-piece measures the inspiratory
and expiratory flows through the tube. This measurement function must be reactivated each
time after switching on Babylog 8000/Babylog 8000 plus and after each replacement of the
sensor by calibrating the sensor. The flow signal is used to calculate the following values:
− Tidal volume (VT)
− Minute volume (MV)
− Percentage of the MV uptake through spontaneous breathing
− Tube leakage
The tidal volume is the expiratory flow signal applied between two breathing phase cycles.
Unlike the tidal volume, the minute volume is not related to a respiratory cycle. In the same
way the inspiratory MV is calculated for leak rate detection. The expiratory minute volume is
displayed. In addition to the complete minute volume, Babylog also calculates the expiratory
minute volume uptake through spontaneous breathing.
Babylog calculates the percentage by comparing the total minute volume and the minute
volume uptake through spontaneous breathing:
The leakage at the tube can be estimated by comparing the inspiratory and the expiratory
minute volume.
The respiratory rate f is measured through the inspiratory respiratory phase cycles.
All measured values derived from the patient flow become valid only after successful
calibration of the flow sensor.
If the flow sensor or the measuring electronics fails, an alarm is generated. However, Babylog
can still be used without the functions depending on the flow measurement. This alarm status
is shown in the status field of the screen by the flashing indication "flow".
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The system checks whether limit values are kept and all functions are ok. If a function fails or
if the limit values are not met, Babylog will generate an alarm.
4.1.1 O2 Measurement
The measured O2 value is checked against upper and lower limit values. The alarm limits are
automatically set at ±4 vol.% below the set FiO2value. A time delay makes sure no alarm is
activated when the set value (O2) is changed or when the O2 sensor is calibrated.
The absolute sensor voltages are checked. The differential voltage between the two sensor
cells must be lower than (U1 + U2)/8 +1 mV. The output voltage of each individual cell must
be between 9.5 mV and 123.6 mV.
4.1.2 O2 Calibration
The O2 sensor is either automatically calibrated 24 hours after the last calibration or manually
after selection from the mode menu.
If the O2 sensor is replaced during operation the new O2 sensor will be calibrated
automatically. However, Babylog cannot detect a sensor replacement if the unit is switched
off. In this particular case, the operator has to calibrate the new O2 sensor manually.
By switching over a valve the sensor is separated from the respiratory gas flow and purged
with the calibration gas (O2). This leads to a change in concentration at the O2 sensor. This
change in concentration allows the Babylog to recognize the activation of the calibration
valve.
A two-step calibration with 21% O2 and 100% O2 is carried out in order to achieve a higher
measurement accuracy over the whole concentration range and to be able to recognize a
spent sensor cell.
During the calibration procedure the microprocessor system processes (synchronizes) one
of the two O2 sensor channels. The O2 channels are then submitted to a specific sequence
For internal use only. Copyright reserved.
of states controlled by one of the microprocessor systems and monitored for correct
sequence and maximum period by the other. For instance, searching of the calibration values
for 21% or 100% must not last longer than 3.5 minutes each.
The calibration value for 21% must be between 9.2 mV and 26 mV, and for 100% between
43.6 mV and 123.6 mV; this applies to both channels. The zero voltage resulting from both
F61733XXT01.fm 24.01.02
The current measured values of the O2 and AIR gas supplies are considered when the valve
bank is adjusted. The inlet pressure for O2 and AIR is 1.7 bar.
If the pressure falls below the limit value, a visible and audible alarm is generated. The
operator is informed about the cause of the alarm by a plain text message on the display. If
the adjusted O2 concentration is 21% or 100%, only an advisory message (no alarm) is
displayed if the non-added gas type fails.
If the pressure exceeds the limit value, the safety venting mechanism is activated and the
continuous flow is switched off. The Babylog will continue to operate only when the pressure
has decreased to a permissible level.
If the AIR supply fails, the Babylog control switches to O2 supply. If the O2 supply fails,
Babylog switches to AIR supply.
Depending on the selected mode and parameter settings, a certain airway pressure time
profile is set. If the actual profile deviates from the set profile, the Babylog generates an
alarm.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
Babylog recognizes whether or not the hose system has been connected correctly. If the
hose system has been connected incorrectly, the continuous flow is (partly) directed to
atmosphere. No pressure builds up in the hose system. Babylog checks during each
breathing cycle whether a sufficiently high pressure is available during inspiration. Since
inspiratory breaths occur in the mandatory ventilation modes IPPV/IMV, SIPPV and SIMV,
this monitoring does not function in the CPAP mode.
Paw
Pinsp
Tdis
Pdis
PEEP
time
Zeit
Tdis > 25 ms !
During each mandatory breath, the level of the disconnection pressure (Pdis) must be
exceeded without interruption for at least the disconnection time (Tdis), otherwise an alarm
will be generated. Pdis is a function of the settings Pinsp and PEEP. If Pinsp is set
considerably higher than the actually attained pressure, the Pdis alarm level likewise
increases.
Depending on the flow and the inspiration time settings, the level might not be attained any
more. Babylog 8000 generates an alarm even if there is no leakage in the breathing system.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
During the inspiratory and expiratory phases, the airway pressure (Paw) must not exceed the
set pressure limit by more than 10 mbar. If the set pressure limit is exceeded by 10 mbar to
20 mbar, Babylog generates an alarm and, at the same time, reduces the inspiratory breath
time. The remaining time is added to the CPAP phase. If the set pressure limit is exceeded by
more than 20 mbar, an alarm is activated and the breathing system is vented.
In the CPAP phase, the airway pressure must be ±4 mbar of the set PEEP/CPAP. Otherwise
an alarm will be generated. If the limit is exceeded by more than 10 mbar, venting is carried
out as during the inspiratory phase even if the pressure falls below the absolute level of —2
mbar.
These overpressure and low pressure alarms apply to all ventilation modes.
Paw
Stenose 2 (Notentlüftung)
10 mbar
Stenose 1 (Inspiration Exspiration)
10 mbar
Pinsp
Entlüftung
6 mbar
4 mbar
PEEP
4 mbar
-2 mbar time
Zeit
Fig. 5: Alarms
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The dynamic stenosis limit applies to Babylog 8000 units with software version 4.0 and
Babylog 8000 SC units with software version 1.0.
If this limit value is exceeded, the microprocessor system reads in the actual measured value
at 8.3-ms intervals (sampling frequency). The limit value (Pinsp + 5 mbar) is subtracted from
each of the actual measured values and entered in a summer.
As soon as the sum of 40 mbar (stenosis 1) is reached, Babylog switches over from
inspiratory phase to expiratory phase. If a sum of 70 mbar is reached, despite of the safety
measure (stenosis 1), the system will carry out an emergency venting.
Paw
Σ = 40 mbar Stenose 1
Σ = 70 mbar Stenose 2
Pinsp + 5 mbar
Pinsp
PEEP
time
Zeit
8,3 8,3
8,3 ms
The monitoring menu is used to set the upper and lower alarm limits in the range of
0.03 L/min to 15 L/min. The upper and lower alarm limits are continuously compared with
the measured MV value. If the value exceeds the upper alarm limits or falls below the lower
alarm limit, the system will activate an alarm. Monitoring is deactivated during calibration of
the flow sensor (Babylog 8000) to allow the measured MV value to stabilize. If the flow
measurement system fails, the MV monitoring becomes ineffective.
The monitoring menu is used to set the upper and lower alarm limits in the range of
0.00 L/min to 15 L/min. The upper and lower alarm limits are continuously compared with
the measured MV value. If the value exceeds the upper alarm limits or falls below the lower
alarm limit, the system will activate an alarm. Monitoring is deactivated during calibration of
the flow sensor (Babylog 8000) to allow the measured MV value to stabilize. If the flow
measurement system fails, the MV monitoring becomes ineffective.
Babylog has two audible alarm generators (loudspeaker and piezo). During normal operation,
the loudspeaker is the audio interface to the operator in the event of alarms. The piezo is only
used in the event of power or loudspeaker failure.
At the end of the self-test, the loudspeaker is triggered with a test signal. This test signal is
monitored with a microphone. If there is no feedback, Babylog generates error 817.
During operation, Babylog uses comparators to continuously monitors whether the operating
voltages are within specified limits.
When Babylog is switched on, these comparators are checked by a defined changing of the
comparator reference voltage.
Input via rotary potentiometers are processed by two channels. Two microprocessor systems
read the input via independent channels and compare the results.
During operation, the ROM area used is summed up per byte in a long word variable (long
word = 32 bits) which is compared to the check sum stored in the ROM. If the check sums
do not match, an error message will be displayed.
During operation the RAM area used is tested with a simple algorithm. By writing 55 hex or
AAhex in each cell every bit is checked for settability and resettability.
A temperature sensor measures the internal temperature of the Babylog. If the temperature is
not within the range of –20 °C to 75 °C, an alarm will be generated. If the temperature is too
high, a message will inform the operator about a possible failure of the fan.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
An excess-current monitoring function is integrated into the trigger electronics of the valves.
The excess-current monitoring function is tested after switching on Babylog. If a valve is
switched on and operated with the operate voltage for an excessive period of time, the
excess-current monitoring function responds within a specific time.
The relay which enables de-energizing of the whole pneumatics assembly is also tested once
after power-on. For this purpose, the relay is switched on and off once and the voltages at the
relay contact are measured.
During operation, the system checks whether the voltage at the relay contact corresponds to
the switching condition.
The battery is recharged automatically during operation and its charge checked periodically.
During operation, the flow sensor is checked for correct functioning, freedom of damage and
intact contacts.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The priority level is indicated by the different tone sequences of the individual audible alarm.
When an alarm occurs the corresponding message is displayed as plain text message in a
window of the display. The message can be canceled momentarily by pressing the
Reset/Check key (or OK key as of software version 5.n). After a certain time out, the
message appears again if its cause still exists.
If several alarms have occurred simultaneously, the highest-level alarm is shown on the
display. Other alarms will be queued. A message whose cause no longer exist will not be
displayed any more.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The display shows parameters and text messages. The functions of Babylog are set with keys
and potentiometers.
Menü line
In some cases, the graphics and measured-value fields are combined to form one large field.
The Monitoring/Mode key and the six keys below the LC display are used for the menu. The
Monitoring/Mode key toggles between the two main menus, the monitoring menu and the
mode menu. This works from any menu status always reaching the basic status of the other
main menu. The function of the keys depends on the respective menu status and is indicated
in the menu line of the display by an abbreviation or a symbol.
The key "Vent. Mode" activates the menu for the ventilation modes. The key "Vent. Option"
activates the menu for the additional functions of the ventilation modes.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The control elements comprise the potentiometers (rotary knobs) and the keys.
This potentiometer is used to set the fraction of inspired O2 to a value between 21 vol.% and
100 vol.%. The adjusted value becomes effective immediately.
This potentiometer is used to set the inspiratory time to a value between 0.1 s and 2 s. The
adjusted value becomes effective at the end of the current respiratory cycle.
This potentiometer is used to set the expiratory time to a value between 0.2 s and 30 s. The
adjusted value becomes effective at the end of the current respiratory cycle.
This potentiometer is used to set the inspiratory flow to a value between 1 L/min and
30 L/min. The adjusted value becomes effective immediately.
This potentiometer is used to set the inspiratory pressure limit to a value between 10 mbar
and 80 mbar. A value above 40 mbar must be confirmed by pressing the Reset/Check key
(or OK key as of software version 5.n). A message is displayed. This pressure limit is also
effective when a manual breath is applied. The adjusted value becomes effective
immediately.
6.1.6 PEEP/CPAP
For internal use only. Copyright reserved.
This potentiometer is used to set the end-expiratory pressure for controlled ventilation or to
set the continuous positive airway pressure for spontaneous breathing to a value between
0 mbar and 15 mbar (or 25 mbar as of software version 4.n). A value above 8 mbar must be
confirmed by pressing the Reset/Check key (or OK key as of software version 5.n). A
message is displayed. The adjusted value becomes effective immediately.
F61733XXT01.fm 24.01.02
6.2.1 CPAP
Pressing this key will activate the CPAP mode. The current respiratory cycle is stopped
before a new mode becomes active (for example, IPPV/IMV). This key is protected against
unintentional operation by a software-controlled locking. After power-on, Babylog will
automatically enter the most recent mode selected.
Pressing this key will activate the controlled ventilation mode. This key is protected against
unintentional operation by a software-controlled locking. If CPAP was set before, the first
mandatory breath is applied after the last expiratory phase has been completed (zero
expiratory flow) or after a zero flow has been measured at the Y-piece for one second.
The CPAP and IPPV/IMV keys deactivate each other. The selected ventilation mode is
indicated by a green LED on the respective key.
Pressing this key will apply a breath at the set inspiratory flow rate and the set pressure limit
(Pinsp). This breath is stopped when the key is released or when a fixed time limit of 5 s is
reached. The next manual or mandatory breath can only be applied after an expiratory phase
(IPPV/IMV) has been completed. If a manual breath is stopped by the time limit, the next
mandatory or manual breath can only be applied after a fixed time-out (5 s).
Pressing this key will silence the audible alarm for 2 minutes. This status is indicated by the
yellow LED on the key. Pressing the key again within the silence time will reactivate the
audible alarm.
Pressing the Reset/Check key (or OK key as of software version 5.n) will confirm or cancel
text messages shown on the display. The message currently shown on the display will be
cancelled and the audible alarm will be silenced.
For internal use only. Copyright reserved.
Pressing the key for a longer period of time (approx. 2 s) will start a display test and an
audible alarm generator test.
F61733XXT01.fm 24.01.02
6.2.8 Keys
The keys below the display have no fixed function. Their function depends on the operating
status of the menu.
This key has not been assigned a function yet. It is reserved for future extensions.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
A distinction is made between an intentional switch-off with the power switch and a power
failure. If a power failure is recognized, the position of the low voltage contacts of the power
switch is evaluated.
When Babylog is switched off, all set parameters and data on the power switch state are
stored. When switching on the Babylog, this information is reloaded and a "cold start" is
carried out. Babylog initializes, carries out its function tests, indicates the results of these
tests, and starts ventilation.
A "warm start" is carried out after a power failure. Babylog does not carry out a self-test, but
starts ventilation immediately. It continues to work with the previously set parameters. A
visible and audible alarm is active during power failure.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
AIR and O2 flow through the filters F 1.1 and F 1.2 and the check valves D 1.1 and D1.2 to
the pressure regulators DR 1.1 and DR 1.2 which regulate the gases to a constant system
pressure.
From the pressure regulators DR 1.1 and DR 1.2 the gases flow to the mixing and flow-
control unit. The gas mixture created there flows through the inspiratory line to the patient.
AIR and O2 are taken downstream of the pressure regulators DR 1.1 and DR 1.2 and
delivered to the solenoid Y 1.1. If one of the gases fails, this solenoid switches over to the
other one.
If the power supply is interrupted or if a stenosis 2 situation has occurred, solenoid Y 1.2
switches over to emergency venting.
The gas required to control the ejector flows through solenoid Y 1.4.
F1.1 DR1.1 1
AIR To mixing and flow-control unit
D1.1
D1.2 DR1.2
2
O2 To mixing and flow-control unit
F1.2
Y1.1
Y1.5
Y1.2
6
Y1.3 3
Y1.4
For internal use only. Copyright reserved.
A continuous flow of gas is delivered to the Y-piece through the inspiratory line. The fraction
of inspired O2 is measured at the O2 sensor S 3.1 through the open pneumatic valve Y 3.1.
Safety valve Y 3.3 prevents overpressure in the ventilation system, for example, if the
expiratory line is blocked.
F3.1 Y3.2
S3.1 Y3.3
D3.1
Insp.
Y3.1
R3.1
3a 10 11a
8.2.1 Inspiration
The PEEP control valve 4 triggers the expiratory valve Y 5.1 causing it to close the expiratory
side. The continuous flow of gas is delivered to the patient’s lung. The airway pressure is
measured by the two relative pressure sensors S 6.3 and S 6.4. The inspiratory pressure limit
is controlled by expiratory valve Y 5.1.
E 6 Pressure measurement
P S6.3
E S6.4
P
R1.1
Y4.1
Y5.1
For internal use only. Copyright reserved.
4 PEEP/PIP valve
5 Expiratory valve
The PEEP control valve Y 4.1 vents the control pressure at the expiratory valve Y 5.1. The
expiratory valve Y 5.1 opens the ventilation system. Expiration takes place through the open
expiratory valve Y 5.1. The ejector, which is driven through the open solenoid Y 1.4, supports
expiration.
E 6 Pressure measurement
P S6.3
E S6.4
P
From
Y1.4
R1.1
Y4.1
Y5.1
4 PEEP/PIP valve
5 Expiratory valve
8.2.3 PEEP
The PEEP/PIP valve is a pressure regulating valve which generates a control pressure of -
19 mbar to +80 mbar. Control pressures are used to trigger the inspiratory and expiratory
valves in order to deliver the peak inspiratory pressure (PIP) during inspiration and the
positive end-expiratory pressure (PEEP) during expiration to the patient-side valves.
An ejector integrated into the valve allows a negative control pressure of up to -19 mbar.
8.2.4 CPAP
The control pressure generated by the PEEP control valve Y 4.1 acts on the control side of
the expiratory valve Y 5.1 and builds up a continuous positive airway pressure in the
ventilation system.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
9.1 O2 Measurement
The O2 sensor continuously measures the fraction of inspired O2 through the open
pneumatic valve Y 3.1.
F3.1 Y3.2
S3.1 Y3.3
D3.1
Insp.
Y3.1
R3.1
3a 10 11a
AIR is supplied to the pneumatic valve Y 3.1 through solenoids Y 1.1, Y 1.2, and Y 1.3. The
pneumatic valve Y 3.1 closes the connection between the O2 sensor S 3.1 and the
inspiratory line. The pneumatic valve Y 3.2 is opened while, at the same time, the O2 sensor
is purged with AIR through the restrictor R 3.2 for approximately 2.5 minutes.
Then, purging with O2 is carried out in the same way. After calibration, the solenoid Y 1.1
switches back to AIR, the solenoid Y 1.3 closes, the pneumatics valve Y 3.2 closes, too, and
the pneumatics valve Y 3.1 reconnects the O2 sensor to the inspiratory line.
Due to the continuous flow, the inspiratory measurement results in a value that is higher than
the value measured at the Y-piece. Since an expiratory measurement results in a value which
is too low, an average value is calculated from the measured values of the relative pressure
For internal use only. Copyright reserved.
A bactericidal metering pipe prevents contamination of the pressure sensor with expiratory
gas.
F61733XXT01.fm 24.01.02
The Pneumatics Control PCB is the output unit of the CPU 68000. It receives serial data
from the I/O PCB and triggers the corresponding valves.
Further functions of Pneumatics Control PCB: voltage supply to the patient system heater,
power supply to the PEEP/PIP valve.
Motherboard
Input logic
Valve block O2
Output
Input block
Pneumatic
The Pneumatics Analog PCB controls and monitors the operate voltage and the withstand
voltage to the valves. The Pneumatics Analog PCB is installed on the Pneumatics
Control PCB.
Current control
Current of
the valve
State of valve
Amplifier control
Stop clock
Switch
regulator
The patient system heater consists of a heating resistor located at the expiratory-valve
connection.
F61733XXT01.fm 24.01.02
The Pressure Sensor Base PCB has four pressure sensors for the measurement of Pinsp,
Pexp, AIR and O2 supply pressures and for coordination of the pressure and the O2
measurement signals.
The O2 Amplifier PCB amplifies the sensor signals directly at the sensor.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The electronic assembly consists of the following printed circuit boards and subassemblies:
− Power supply unit
− Motherboard PCB
− CPU 68000 PCB
− I/O PCB
− Flow PCB (Babylog 8000) (optional)
− Monitoring PCB/Watchdog PCB
− Front Adapter PCB
− Front Controller PCB (Babylog 8000/Babylog 8000 SC with LC display )
− Front PCB (Babylog 8000/Babylog 8000 plus with EL Display)
− Display PCB (Babylog 8000/Babylog 8000 SC with LC display )
− EL display (Babylog 8000/Babylog 8000 plus with EL display)
− Potentiometer field
− Display field
− Communication PCB (optional)
− Interface PCB (optional)
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The power supply unit supplies Babylog with the following voltages:
− +5 V
− +15 V
− –15 V
− +27 V
The power supply unit comprises the mains connection, the ON/OFF switch, and the fuses.
All electrical connections between the individual printed circuit boards are led through the
Motherboard PCB. Cable connections go to the front panel, to the pneumatic assembly, and
to the rear panel, which contains the rechargeable battery, the fan, the loudspeaker, and the
flow-sensor connection.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The CPU 68000 PCB comprises the following components: 68000 CPU, EPROM, RAM,
time-keeper RAM, address decoder, multi function peripheral (four 8-bit timers, RS 232
interface, 8-bit I/O port), DTACK and bus error generator, watchdog, power-on reset,
interrupt controller, bus controller, and bus interface.
The CPU 68000 PCB is clocked with 8 MHz. If the "communication" conversion kit is
installed, the clock frequency is increased to 16 MHz.
Bus-Interface 16 MHz
Adress- 8 MHz
decoder
Reset
Power-on-
Reset
Interrupt EPROM CPU 68000
Control-Bus Interrupt
Digital-Bus 68000
Controller Co-
RAM 12 MHz
prozessor
Adreß-Bus
Bus-
Controller Time DTACK-,
Keeper Bus-Error-
RAM Generator
Daten-Bus
Watchdog
Adreß-Bus
Seriell Multi
Function
Powerfail Peripheral
I/O
Digital I/O
Optionsschalter
Status CPU
The I/O PCB measures the analog signals O2 concentration, inspiratory pressure, expiratory
pressure, AIR supply pressure, and O2 supply pressure, and triggers the loudspeaker and
the PEEP/PIP valve.
Parallel/seriell
DTACK DTACK Converter
generator
Control bus
Sound
Adress
Digital bus 6800
generator
decoder
Adress bus
Control-Bus
Adreß-Bus D U
Data bus Daten-Bus A I
PIO
S A
Muxer &
H D
The Flow PCB in Babylog 8000/Babylog 8000 plus measures the flow. It provides the
following function blocks: sensor bridges with sensor amplifiers, analog-to-digital converter,
microprocessor Z80 minimum system, Z80-68000 bus interface.
Adress
decoder 6 MHz
DTACK DTACK
generator EPROM Z80
64 k / 32 k CPU
Control bus
Adress
RAM
Digital bus 68000
decoder Timer
8k/2k
Control bus
Adress bus
Adress bus Dual-Port-
RAM Watchdog
Data bus
Data bus
System reset
Analog 1
S A
Muxer & PIO
Analog 4 H
D
Flow sensor
Status flow
Control bus
Adress bus
Data bus
The Monitoring PCB measures and monitors all measurement and status signals. The
Monitoring PCB compares these signals with the signals from the CPU 68000. If an error
For internal use only. Copyright reserved.
occurs, the Monitoring PCB switches off the valves, activates the audible alarm generator,
and resets the CPU PCB.
A Z80 microprocessor (minimum system) controls the Monitoring PCB. The Monitoring PCB
communicates with the CPU 68000 via a bus interface.
F61733XXT01.fm 24.01.02
− O2 concentration
− Potentiometers (O2, Vinsp, TI, TE, Pinsp, PEEP/CPAP)
− Pressure Pinsp
− System temperature
− Supply voltages (potentiometer reference, rechargeable battery, GNDA, GND
reference, 27 V)
− Valve voltage
8k Timer
Control bus
Adress bus
Dual-
Adress bus Port- Watchdog
RAM
Data bus
Analog Input
S&H Watch dog
PIO
PCB
(Filter)
A
I/O connections
D
Digital Input
Status
PIO display
+
voltages
-
URef
sound
decoder
rechargeable battery
The Front Adapter PCB connects the CPU 68000 to the front panel. The front panels
provides the control and display elements.
The Front Adapter PCB reduces the 16-bit data bus to 8 bits, generates control signals for
the front panel, and controls the LC display (VIDEO RAM and LCD controller) or the EL
display.
The power supply and the analog signals are led from the Motherboard PCB to the front
panel through the Front Adapter PCB.
DTACKQ DTACK
generator
decoder Controller
Adress bus
Dual-
Data bus Port-
RAM
Daten-Bus (16 bit)
Frontcontroller
Potentiometer monitoring
For internal use only. Copyright reserved.
The Front Controller PCB reads in the keys of the potentiometer and display field, controls
the LEDs of the potentiometer and display field, powers the LC display with operating voltage
(contrast), controls the LC display backlight, and generates the control signal for the
bargraph.
key port
Control bus
and
Adress bus LEDs of the front panel
Alarm LEDs
Test A/D converter
Amplifier
Bargraph
Bargraph of the front panel
UB
URef
A
Potentiometer of the front panel
D
6x
Adress
decoder
Fig. 22: Block diagram of the Front Controller PCB (Babylog 8000/8000 SC with LC
For internal use only. Copyright reserved.
display)
F61733XXT01.fm 24.01.02
The Front PCB reads in the keys of the potentiometer and display field, controls the LEDs of
the potentiometer and display field, powers the EL display with operating voltage (contrast),
and generates the control signal for the bargraph.
EL display, complete
DC +5 V, +12 V
DC
EL display converter
Cable of EL display
EL display
key port
Alarm LEDs
Test A/D-Wandler
LED driver
Amplifier
Bargraph
Bargraph of the monitoring field
UB
URef
A
Potentiometer of the potentiometer field
D
6x
Adress
decoder
The Display PCB contains the control electronics for the LED array of the bargraph.
Data bus
Adress bus
Modul 06
The EL display (Electro Luminescence Display) consists of the EL display, the EL display
converter, and the EL display connecting cable. The EL display displays the signals and other
information. The EL display converter generates the (200-240 V a.c.) voltage for the
luminescence voltage of the EL display. The EL display and the converter are ”matched”.
EL display
EL display converter
pn 45363667
For internal use only. Copyright reserved.
The Communication PCB controls an external printer, a computer, and an analog recorder.
The Communication PCB is equipped with an RS232 interface, two 12-bit D/A outputs, and
one digital trigger output. All outputs are electrically isolated from the electronics. Software
version 3.00 or higher, a CPU PCB with 16 MHz (the CPU PCB with 8 MHz was standard
until end of 1992) and the Interface PCB with outputs on the rear panel are minimum
requirements for operation.
System connector
Analog
RS 232
Output
Trigger
Nurse call
A
D
Register
2,5 kV electrical isolation
DC
DC
Parallel/seriell
converter
MFP*
decoder
Adress
DTACK
Control bus
For internal use only. Copyright reserved.
Adress bus
Data bus
DTACK
The Interface PCB is a passive printed circuit board with the following connections:
− Flow-sensor cable
− 2 analog outputs
− Trigger output
− RS 232
+15 V
V1
R1
X 1 / 16 X4 Analog 1
V2
+15 V
V3
R1
X1/8 X5 Analog 2
+15 V V4
V5
R1
X 1 / 12 X6 Trigger
V6
X1/5 X3/2
X1/4 X3/3
X3/5
X 1 / 14 X3/7 RS 232
X3/8
X3/4
X 1 / 15
X 1 / 16 X3/6
X3/1
X8/1 X7/1
X8/2 X7/9
X8/3 X7/2
X8/4 X 7 / 10
For internal use only. Copyright reserved.
X8/5 X7/3
X8/6 X 7 / 11 Flow-Sensor
X8/7 X7/4
X8/8 X 7 / 12
X 8 /15 X7/8
X 8 / 16 X 7 / 15
F61733XXT01.fm 24.01.02
The Y-piece with integrated flow sensor is used in pediatric respiratory gas systems. This Y-
piece connects the respiratory gas hoses of Babylog 8000/Babylog 8000 plus to the patient
catheter connection.
Inspiration Exspiration
The flow sensor works according to the hot-wire flowmeter principle. It has an additional
measuring wire for direction recognition.
For internal use only. Copyright reserved.
F61733XXT01.fm 24.01.02
The flow sensor incorporates two hot-wire flowmeters. A separate wire is used for direction
recognition. Both wires are evaluated on the Flow PCB by two separate measuring channels.
UB
V7
Flow test
Comparator
12 Bit
D 24
R 54 RDR Hot wire
PI regulator
Hot wire
glowing N7
+ Umeß
R 72 R 70
The measuring bridge comprises the resistors R 54, R 72, R 70, and the measuring wire of
the sensor.
The measuring wire is kept at a constant resistance. This means that the amount of heat
dissipated by the gas flow has to be readjusted. As system deviation the voltage passes
through the node R 54/R72 and the measuring wire R 70. If the voltage is adjusted to "0",
the current flowing through the measuring wire and the R70 is a measure for the heat
dissipated. The controllable power source is made with V7. N7 is a controller with PI
behavior. To adjust the bridge a current can be injected via the D/A converter D 24 into the
node R54/R72. The current through the bridge is changed such that the wire begins to glow.
Burning out of the wire is prevented by a current limiter. The current limiter compares the
voltages and switches the current source off, if required. Current limitation is effected
For internal use only. Copyright reserved.
through the comparator which is connected to the input of V7. The current through the flow
measuring bridge is changed for test purposes. R 70 is a current measuring resistor (four-
wire measurement) which measures the heating current (proportional voltage).
F61733XXT01.fm 24.01.02
The O2 sensor is a double cell which functions according to the fuel-cell principle, that is, it is
an electrochemical cell which generates a voltage by means of an ion current. The cell
consists of a capsule which incorporates electrolytes, lead anodes and two gold cathodes
with Teflon diaphragm.
cathode 1
Plastic housing
O2 Alkali-Elektrolyt
anode
cathode 2
- J
The oxygen to be measured diffuses through the Teflon diaphragm, undergoes a chemical
reaction at the gold cathode and produces lead oxide and water at the lead anode. During
this chemical process an electric voltage is generated which is proportional to the oxygen
partial pressure. The gold cathodes are negative, the lead anode is positive. The internal
resistance is determined by the surface of the electrodes, the oxygen diffusion velocity, and
the distances. Under normal condition, the internal resistance is 700 ohms.
Paw
Trigger -10 0 20 40 60 mbar
Alarm
g
man. Reset
Insp. Check
Monitoring
Mode
O2-Vol.% TI TE
50 60 0,6 1,0
70
40 0,4 0,8 0,8 2 CPAP
25 90 0,2 1,5 15
0,4
Copyright reserved. Reproduction allowed for non-commercial purposes only.
7,5
40
5 5
10 10
30 50
3 20
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Check List
Check List
Rear Panel
Fan cover
Power cable
Copyright reserved. Reproduction allowed for non-commercial purposes only.
Side Panels
GBK61733XXTL1.fm 24.01.02
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Potentiometer field
Display field
Inspiratory block
O2 sensor mount
Expiratory Valve
Crater
Expiratory diaphragm
Ejector
Copyright reserved. Reproduction allowed for non-commercial purposes only.
Silencer
GBK61733XXTL1.fm 24.01.02
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Y-piece with flow sensor 8410185 with flow sensor set 8410179 (only in the
Babylog 8000/Babylog 8000 plus)
Flow sensor (ISO) 84 11 015 with flow sensor set 8410179 (only in the
Babylog 8000/Babylog 8000 plus)
Flow sensor (ISO 15, low dead space) 84 11 130 with flow sensor set 84
10 179 (only in the Babylog 8000/Babylog 8000 plus)
Trolley 0.5 B
Humidifier
Hinged arm
Water trap Ex
Drug nebulizer
GBK61733XXTL1.fm 24.01.02
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
7 Functional Check
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Test set-up
Bellows
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Test set-up
T-piece
Bellows
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Variant A
(not in the USA) 0 mbar /
cmH2O
Y 3.3
Inspiratory block
Variant B
0 mbar /
cmH2O
Y 3.3
Inspiratory block
Fig. 3: Test 1
Copyright reserved. Reproduction allowed for non-commercial purposes only.
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Remove the bottom plate from the Babylog 8000/Babylog 8000 plus/
Babylog 8000 SC and disconnect the hose from connection 4 on the gas
connection block. Apply a pressure of 1.5 bar ±0.1 bar into this hose (see
Figure below).
4
Valve block
4x1.5 Si nf hose
PEEP/PIP-
Valve
Fig. 4: Test 2
GBK61733XXTL1.fm 24.01.02
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Variant A
(not in the USA)
25 mbar /
cmH2O
1.5 bar
/21.3 psi
Y 3.3
Inspiratory block
Variant B
100 mbar /
cmH2O
1.5 bar
/21.3 psi Y 3.3
Inspiratory block
Important: If during Test 1 the variant A and during Test 2 the variant B
(pressure values possibly too high) is recognized, then it is a B variant with a
non-permissible spring mounted in the valve Y 3.3. This spring must be
removed and the test be repeated. Mount the bottom cover.
• Slowly compress the bellows and generate the following pressure:
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Communication conversion kit y/n (is displayed under the operating hour
indication after power-on).
Important: 8306065 or 8411108 (see nameplate option field) without
central-alarm function. The sockets for the communication interface may be
present even if this function is not available.
• O2 calibration.
• Flow sensor calibration (only in the Babylog 8000).
LED and horn test (press and hold the “Reset Check” : after 2 s, each LED
and the stand-by horn are triggered.).
Important:
Alarm LEDs = 1 yellow and 2 red LEDs
Trigger LEDs = 3 yellow LEDs
Display backlight.
• Testing the keys (check respective reaction of the equipment).
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
60 1 2
1 80 0
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
CPAP
30 15
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
30 2 2
10 var var
• Select the functions “setting” and “measured value” from the “list”
menu:
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
CPAP
60
var 30 5
Fig. 9: Mixer
• Select the following inspiratory flows from the “list” menu: 1.2 / 1.3 /
Copyright reserved. Reproduction allowed for non-commercial purposes only.
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
• Put your hand on the upper right corner of the rear panel of the
Babylog 8000/Babylog 8000 plus/Babylog 8000 SC.
O2 Failure Alarm
Copyright reserved. Reproduction allowed for non-commercial purposes only.
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
Power cable
Mains fuse-unit
(the specifications on the fuse-unit must match those on the rating plate)
21 Tested
Date: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK61733XXTL1.fm 24.01.02
This Test List does not replace inspections and servicing by Dräger Medizintechnik GmbH
1 Important Information
Follow the instructions in this manual and carefully read and observe all statements following
a warning or info symbol. If you ignore the warning statements, this may lead to personal
injury and/or device damage.
Before removing the rear panel, unplug the power plug from the mains supply and
disconnect the device from the central gas supply.
After performing any repair or inspection on medical devices, carry out a final check before
returning the devices to normal use.
For internal use only. Copyright reserved.
GBK61733XXV01.fm 24.01.02
• Pull out the cooling air filter 1 from its mount in the rear panel.
• Insert a new cooling air filter or wash the old one in warm water with a few drops of dish-
washing liquid. Dry well.
• Insert the cooling air filter 1 into the mount (make sure it is free of folds).
• Remove the slotted-head screws 1 in the cover on the right-hand side and carefully pull
the cover out.
• Pull the spent O2 sensor capsule out of its mount.
• Insert the new O2 sensor capsule into the mount (the ring-shaped conductors must
show towards the cover).
• Push the cover in place and secure it with the screws 1.
• Calibrate the O2 sensor manually.
IPPV
IMV
Special funktions
IPPV
IMV
For internal use only. Copyright reserved.
sensor calibration
Flow FiO2
GBK61733XXV01.fm 24.01.02
After approx. 5 min, the display “O2 Cal” disappears from the status field and calibration is
completed.
• Use, for example, a spherical-head 2.5 mm screwdriver to remove the old lip seals 2.
Insert the new lip seals and make sure they fit properly.
2 2 2
• Re-mount the expiratory valve. To do so, lift up the toggle and place the expiratory valve
For internal use only. Copyright reserved.
4 4
2
For internal use only. Copyright reserved.
3
GBK61733XXV01.fm 24.01.02
• Re-assemble the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC. The battery for
power failure alarm is recharged automatically during operation.
After replacing the battery, allow the device to operate for at least half an hour in order to
sufficiently charge the battery.
No acoustic alarm will sound in the event of power failure if the battery
is not sufficiently charged. Risk of severe injury or death. Allow the
device to operate for at least half an before clinical application.
For internal use only. Copyright reserved.
GBK61733XXV01.fm 24.01.02
• Disconnect the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC from the mains
supply.
• Disconnect the gas supply hoses from the central gas supply/gas cylinder.
• Disengage the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC from the trolley.
To do so, loosen the knurled nuts 1 on the rear panel.
1 1
2 2 2 2
• Remove the bottom screws 3 from the fixing straps on the left and right panel.
3 3
Fig. 13: Rear Panel of the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC
• Gently lift the top part and place it next to the bottom part.
• Remove the screws 5.
• Remove the gas supply connection plate 6.
• Remove the two filters and the gasket. Loosen the screws 7 on the supply block.
7 7
5 5
6
For internal use only. Copyright reserved.
7 7
Fig. 15: Pneumatic Assembly of the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC
Tilting the supply block excessively will cause damage to the hoses
and/or printed circuit boards. Do not tilt the supply block more than
necessary.
For internal use only. Copyright reserved.
GBK61733XXV01.fm 24.01.02
9 9 9 9
• Replace the pressure reducers 10 for AIR and O2 including the O-rings 11 (part
number D19080) and 12 (part number M23154).
10
12
11
Air
B
O2
1
• Connect to the air supply only and switch on the Babylog 8000/
Babylog 8000 plus/Babylog 8000 SC.
• Adjust the following parameters:
Mode = CPAP; PEEP/CPAP adjuster = 0 mbar; insp. flow = 1 L/min
For internal use only. Copyright reserved.
GBK61733XXV01.fm 24.01.02
• Carry out the measurement with a manometer and a secretion sightglass at the
pneumatics outlet.
• Slowly turn the adjusting ring of the air pressure reducer clockwise. The message “no
compressed air” on the LCD disappears.
• Set the pressure reducer to 1.7 -0.1 bar.
• Set the insp. flow adjuster to 30 L/min. The manometer should indicate at least 1.3 bar.
For internal use only. Copyright reserved.
GBK61733XXV01.fm 24.01.02
• Carry out the measurement with a manometer and a secretion sightglass at the
pneumatics outlet.
• Slowly turn the adjusting ring of the O2 pressure reducer clockwise. The message “no
oxygen” on the LCD disappears.
• Set the pressure reducer to 1.7 -0.1 bar.
• Set the insp. flow adjuster to 30 L/min. The manometer should indicate at least 1.3 bar.
For internal use only. Copyright reserved.
GBK61733XXV01.fm 24.01.02
C C
R3.2
R2.1 R2.2 R2.10 O2
D3.1
F1.1 DR1.1
1a Y3.3
Air
1 7 Y3.1
D1.1 9a
8 Insp.
F1.2 DR1.2 9
O2
2 2a
D1.2 R2.11 R2.12 R2.20 R3.1
nc Y2.11 Y2.12 Y2.20 3a 10 11a 11
12/00
Y1.1
no 6 Pressure
Y1.5 c E
P measurement
S6.3
no n 6 Valve block
Y1.2 E
6 P S6.4
c R3.3
3
Y1.3 12a
4 4a R1.1 12
Y1.4 D4.1
10b
10a Exp.
Schematics and Diagrams
Y4.1 Y5.1
1 Gas supply block
5 4 PEEP/PIP valve
5a
5 Expiratory valve
Version 0 valid up to serial no. ARDK (Sep. 91)
incl. spring (25 mbar in Y 3.3)
incl. pressure limiting valve (85 mbar) in PEEP/PIP
Page 92
valve,
Y 3.3 is triggered by PEEP/PIP pressure
GBK61733XXD01.fm 24.01.02
235 bl
Detail X
Detail Z
170 ye
1 glued with silicone
2-2
DR1.2 DR1.1
10 ±1
(glue Elastosil
Gas supply block E41/Wacker)
Hose 2x1 Si gn.
10 lg
2 Y1.2
5 Hose 4x1.5 Si
3 Y1.5
Y1.1 tr. 20 lg
Y1.3 Y1.4 4 M 22 286 Valve block View U View V
100 30 O2 Hose 2x1.5 Si ye. Hose 2x1.5 Si
Air
40 230 ye 3 lg bl. 3 lg
20
120 68 00 187
85 S6.1
60 View X View Y
7a Hose 2x1 Si
12/00
Y2.20 Y2.10
5a 100 9
Hose 2.7 x 0.65
12 8 7 - PA 11 w M25778
3a 170 bl.
X
50 gn PA tube 6 x 1 bl ye
10b 11
84 01 083 10
2 x 1 Si rd. gn. or
80 gn 9a 2 x 1.5 Si tr. ye. bl
4 x 1.5 Si tr.
Page 93
5 x 2 Si tr.
GBK61733XXD01.fm 24.01.02
6 Pressure measurement
E E 3 Inspiratory block
P P F3.1
8a E Y3.3
7a R2.1 O2
R2.2 R2.10
F1.1 DR1.1
1a
D3.1
Air
7 Y3.1
D1.1 9a
8 Insp.
F1.2 DR1.2 9
O2
2 2a
D1.2 R2.11 R2.12 R3.1
R2.20
12/00
4
nc R1.1
10a 10b
Y1.4 Exp.
Y4.1
Y5.1
1 Gas supply block
5 4 PEEP/PIP valve
5a
5 Expiratory valve
Version 1 up to ARDL (Oct. 91)
Page 94
235 bl.
2-2
10 ±1
Gas supply block
2
5 Y1.2 Hose 2x1 Si gn.
3 10 lg Hose 4x1.5 Si tr.
Y1.5
100
Y1.1 20 lg
Y1.3 Y1.4 4 Valve block
M 22 286
40 O2 View U View V
Air
230 ye Hose Hose
20 2x1.5 Si 2x1.5 Si
68 00 187 ye. 3 lg bl. 3 lg
85
60 S6.1
68 00 187 7a
12/00
150 2a View Y
U 1a View X
D4.1 20
Hose 2x1 Si Hose 2x1 S
PEEP/PIP S6.2
8a rd. 3 lg rd. 3 lg
valve 230 84 06 810 100 rd
V Y2.11 Y2.1
4a Z
Y4.1 X S6.3
84 01083
Y2.20 Y2.10
100
5a
9 Hose 2.7 x 0.65
12 3a 170 bl. 8 7 - PA 11 w M 25778
50 gn
PA tube 6 x 1 bl. ye
10b 11 10
2 x 1 Si rd. gn. or
9a
2 x 1.5 Si tr. ye. bl.
Page 95
4 x 1.5 Si tr.
5 x 2 Si tr.
GBK61733XXD01.fm 24.01.02
6 Pressure measurement
E E 3 Inspiratory block
P P F3.1
D3.1
R2.1 R2.2 R2.10 O2
F1.1 DR1.1
1a R3.2
Air
7
D1.1 9a
8 Insp.
F1.2 DR1.2 9
O2 Y3.1
D1.2 2a R2.11 R2.12
R2.20 R3.1
12/00
R1.1
nc 10a
Exp.
Y1.4 Y4.1 10b
Y5.1
1 Gas supply block
5 4 PEEP/PIP valve 5a
5 Expiratory valve
Version 2 (as of 1992)
without spring (25 mbar in valve Y 3.3)
Page 96
235 bl
2-2
Gas supply block (glue Elastosil
10 ±1
E41/Wacker)
2
Y1.2 Hose 2x1 Si
3 5
gn. 10 lg
Y1.1 Y1.5 Hose 4x1.5 Si tr.
Y1.3 Y1.4 M 22 286 Valve block 20 lg
4
110 40 100 O2 Air View V
230 ye View U Hose
30 20 Hose 2x1.5 Si
2x1.5 Si bl. 3 lg
120 68 00 187 ye. 3 lg
85 S6.1
60
7a
130 68 00 187
12/00
U 2a 1a
40
View X View Y
D4.1 S6.2
84 06 810 Hose 2x1 Si
8a rd. 3 lg Hose 2x1 Si
PEEP/PI 230 100 rd Y2.11 Y2.1 or. 3 lg
Pvalvel V
4a Z S6.3
Y1.4
11a Y2.12 Y2.2
10a 100 or
X
Pressure Z S6.4
sensors
12a
20
Y
70
Schematics and Diagrams
170 bl
Y2.20 Y2.10
5a
100
9 Hose 2.7 x 0.65
12 8 7
3a - PA 11 w M 25778
10b
11 PA tube 6 x 1 bl. ye
X 10
210 gn
2 x 1 Si rd. gn. or
2 x 1.5 Si tr. ye. bl.
9a
4 x 1.5 Si tr
Page 97
5 x 2 Si tr.
GBK61733XXD01.fm 24.01.02
Fig.
7: Function Diagram of Babylog 8000/Babylog 8000 SC „Version 3“
6 Pressure measurement
D3.1
R2.1 R2.2 R2.10 O2
F1.1 DR1.1
1a R3.2
Air
7 9a
D1.1
8 Insp.
F1.2 DR1.2 9
O2 Y3.1
2a
D1.2 R3.1
R2.11
12/00
R2.12 R2.20
nc Y1.1
10 11
Y2.11 Y2.12 Y2.20 11a
Y1.5 no 4,8m
c 6 Pressure
nc Y1.2 E
6 Valve block P measurement
R3.3 S6.3
6 no
c E S6.4
nc 3
P
D 4.1
Y1.3
4 4a 12
12a
Schematics and Diagrams
R1.1
10a
nc Exp.
Y1.4 Y4.1
10b
Y5.1
1 Gas supply block
5 4 PEEP/PIP valve
5a
5 Expiratory valve
Page 98
235 bl.
2-2
(glue Elastosil
10 ±1
E41/Wacker)
2
Y1.2 Hose 2x1 Si
3 5
gn. 10 lg
Y1.1 Y1.5 Hose 4x1.5 Si tr.
Y1.3 Y1.4 4 M 22 286 Valve block 20 lg
7a
130 68 00 187 2a 1a View X View Y
40
U
D4.1
84 06 810 S6.2 Hose 2x1 Si Hose 2x1 Si
PEEP/PIP 230 8a rd. 3 lg or. 3 lg
100 rd Y2.11 Y2.1
valve V
4a Z
Y1.4 S6.3
11a Y2.12 Y2.2
10a 100 or
X
Pressure Z
sensors S6.4
20
12a
Schematics and Diagrams
70 170 bl. Y
9a
5 x 2 Si tr.
GBK61733XXD01.fm 24.01.02
6 Pressure measurement
D3.1
R2.1 R2.2 R2.10 O2
F1.1 DR1.1
1a R3.2
Air
7 9a
D1.1
8 Insp.
F1.2 DR1.2 9
O2 Y3.1
2a
12/00
D1.2 R3.1
R2.11 R2.12 R2.20
nc Y1.1
10 11
Y2.11 Y2.12 Y2.20 11a
Y1.5 no c
4,8
6 Pressure measurement
nc Y1.2 E
6 Valve block R3.3 P
S6.3
no
6
c E S6.4
nc 3
P
Y1.3 D 4.1
4 4a 12a 12
R1.1
Schematics and Diagrams
nc 10a
Y1.4 Exp.
Y4.1 10b
1 Gas supply block Y5.1
5 4 PEEP/PIP valve
5a
5 Expiratory valve
Page 100
The restrictor for the purge flow is integrated in the silicone hose (84 11 459)
Fig. 10: Tubing Diagram of Babylog 8000/ Babylog 8000 SC (Version 3 with conversion kit Gas supply block
1 170 ye Detail X Detail Z
8409111
60gn DR1.2 DR1. glued with silicone
„purge flow“)
2-2
(glue Elastosil
10 ±1
E41/Wacker)
2
Y1.2 Hose 2x1 Si
3 5
gn. 10 lg
Y1.1 Y1.5 Hose 4x1.5 Si tr.
Y1.3 Y1.4 4 M 22 286 Valve block 20 lg
110 40 100 O2 Air View U View V
230 ye Hose
30 20 Hose 2x1.5 Si
2x1.5 Si bl. 3 lg
120
68 00 187 ge. 3 lg
85
60 S6.1
7a
130 68 00 187 2a
U 1a View X View Y
40
D4.1
84 06 810 S6.2 Hose 2x1 Si Hose 2x1 Si
12/00
4a Z
Y1.4 S6.3
11a Y2.12 Y2.2
10a 100 or
X
Pressure Z
20 sensors S6.4
12a
70 Y
170 bl.
Y2.20 Y2.10 Hose 2.7 x 0.65
Schematics and Diagrams
5a - PA 11 w M 25778
100
PA tube 6 x 1 bl. ye
12 3a 9
8 7 2 x 1 Si rd. gn. or
6 Pressure measurement
„Version 3.10“
D3.1
R2.1 R2.2 R2.10 O2
F1.1 DR1.1
1a R3.2
Air
7 9a
D1.1
8 Insp.
F1.2 DR1.2 9
O2 Y3.1
2a
D1.2
12/00
R1.1
10a
nc Exp.
Y1.4
Y4.1 10b
1 Gas supply block Y5.1
5 4 PEEP/PIP valve
5a
5 Expiratory valve
Page 102
Version 3.10
Version 3.10 with leakage compensation (avoids a subsequetn pressure rise at the pressure
reducers.
GBK61733XXD01.fm 24.01.02
Fig. 12: Tubing Diagram of Babylog 8000/Babylog 8000 plus/Babylog 8000 SC „Version
235 bl.
Gas supply
3.10“
1 170 ye
8409111 Detail X
60gn DR1. DR1. Detail Z
2-2
glued with silicone
(glue Elastosil
10 ±1
2 E41/Wacker)
5 Y1.2 Hose 2x1
3 Si gn. 10 lg
Y1.
Y1.1 M 22 286 Hose 4x1.5 Si
Y1.3 Y1.4 Valve block
4 tr. 20 lg
110 40 100 84 11 411 O2 Air
View U View V
Hose
30 20 160 ye Hose 2x1.5
120 2x1.5 Si
68 00 187 Si bl. 3 lg
40 bl.
85 S6.1 ge. 3 lg
60 7a
130 68 00 187 2a
12/00
20 1a View Y
40
U
20 View X Hose 2x1
D4.1 84 06 810 S6.2 Si rd. 3 lg Hose 2x1 Si
PEEP/PI 230 8a or. 3 lg
100 rd Y2.11 Y2.1
Pvalve V
4a Z
Y1.4 S6.3
11a Y2.12 Y2.2
10a 100 or 20
X
Z 84 11 411
Pressure S6.4
20
sensors 12a
70
110 bl.
Y
Schematics and Diagrams
2 Type of Changes
This service manual reflects the state of technology which applied in August 1996.
Bitte hier für Inhaltsverz. klicken/ Please, press here for table of contens
Sach-Nr. Benennung
Part No. Description
8409200 BABYLOG 8000 D/E 220/240V
BABYLOG 8000 G/E 220/240V
7 PLATTE 8409191
PLATE
8 GASANSCHLUSS 8409955
GAS SUPPLY
9 SCHALTVENTIL 8409960
PILOT VALVE
10 PEEP/PIP-VENTIL,KOMPL. 8409174
PEEP/PIP VALVE, CPL.
11 INSPIRATIONSBLOCK 8409100
INSPIRATION BLOCK
12 EXSPIRATIONSVENTIL 8408950
EXSP.-VALVE BO 8000
15 ELEKTRONIK,VOLLST. 8305417
ELECTRONIC, CPL.
16 ELEKTRONIK,VOLLST. 8305417
ELECTRONIC, CPL.
18 VENTILBANK 8409961
DIGITAL VALVE ARRAY
Datei/File: 61733.020
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 7 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
4 KOMPAKTSCHIENE,KOMPL. 2M85337
COMPACT RAIL, CPL.
ADAPTER-SCHLAUCHHEIZUNG 8411097
ADAPTER-DUAL SERVO HOSE HEATER
Datei/File: 61733.001
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 9 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
Datei/File: 61733.002
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 10 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
3 SCHRAUBE 2M17288
SCREW
4 SEITENWAND 6H 2M17345
SIDE PANEL 6H
6 BODENBLECH 8410037 10
8 SCHALLDAEMPFER 8409990
SOUND ABSORBER
9 FILTERMATTE 8305367 10
FILTER MAT
Datei/File: 61733.003
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 13 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
12 TUELLE 6800187
T-SOCKET
13 T-STUECK 8401083
T-PIECE
16 Y-STECKER G60423
Y-CONNECTOR
17 REDUZIERSTUECK M22286
REDUCER
21 STECKKUPPLUNG 8404948
PLUG-IN COUPLING
Datei/File: 61733.004
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 16 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
3 GASANSCHLUSS 8409955
GAS SUPPLY
4 FUSS 2M17302
BASE
5 PUFFER 6800478
BUFFER
6 AUFKLEBER 8302721
ADHESIVE LABEL MULTIWARN
8 PEEP/PIP-VENTIL,KOMPL. 8409174
PEEP/PIP VALVE, CPL.
11 VENTILBANK 8409961
DIGITAL VALVE ARRAY
14 GEHAEUSE 8410029
HOUSING
15 SETZMUTTER CH17646
NUT
16 NIETMUTTER 8305346
RIVET NUT
31 BODENBLECH 8410037 10
BASE PLATE
32 UEBERDRUCKVENTIL 8410568
PRESSURE CONTROL VALVE
33 KREUZHALTERUNG 8404154
CROSS SUPPORT
Datei/File: 61733.005
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 19 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
2 SETZMUTTER M8 M19399
SET NUT M8
4 LASCHE 2M17311
bis Geräte-Nr. ARKL-xxxx
LINK
up to device no. ARKL-xxxx
5 BLECH 2M17305
ab Geräte-Nr. ARKM-xxxx
SHEET METAL
from device no. ARKM-xxxx
7 KLINKE 2M17303
CATCH
8 LASCHE 2M17310
LINK
9 SETZMUTTER CH18043
NUT
14 PLATTE 8409191
PLATE (WITH INSPIRATION BLOCK)
PLATTE Bild/Picture 7
PLATE
Datei/File: 61733.006
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 21 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 PLATTE 8410024
PLATE
2 TUELLE 8400964
SOCKET
3 VERSCHLUSSPLATTE 8409951
LOCKING PLATE
4 SCHEIBE 8300401
DISC
6 DICHTRING D04873
COPPER RING
7 EINSCHRAUB-STECKANSCHLUSS M25210
SCREW-IN PLUG-TYPE CONNECTION
8 PATIENTENTEILHEIZUNG 8305480
PATIENT PART HEATING
11 TUELLE 8402393
SOCKET
12 SCHEIBE 8402392
WASHER
15 O-RING 2M10540
O-RING SEAL
16 LIPPENDICHTUNG 8408442
LIP SEAL
17 GEWINDESTIFT 8410026
THREADED PIN
GASANSCHLUSS Bild/Picture 8
GAS SUPPLY
Datei/File: 61733.007
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 24 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 DRUCKREGLER 8409199
PRESSURE REGULATOR
3 O-RING M23154
O-RING SEAL
4 O-RING D19080
O-RING SEAL
8 WINKEL 8409506
ELBOW
10 SCHALTVENTIL 8409960
PILOT VALVE
11 RUNDSCHNURRING M09361
TOROIDAL SEALING RING
12 O-RING M16958
O-RING SEAL
14 STECKANSCHLUSS M30960
PLUG-TYPE CONNECTION
15 EINSCHRAUB-STECKANSCHLUSS M25210
SCREW-IN PLUG-TYPE CONNECTION
16 MIKRO-MAGNETVENTIL 8406545
MICRO-ELECTROVALVE
18 TUELLE 8402462
SOCKET
20 L-SCHWENKANSCHLUSS G12065
L-SWIVEL CONNECTION
21 PRUEFDIODE 8407732
TEST DIODE
23 FLACHDICHTUNG 8408204
FLAT PACKING
24 FILTER 8408208
FILTER (FOR 8411848)
25 AUFNAHME 8410474
RECEIVER
28 VENTILKEGEL D21138
VALVE CONE
29 STELLSCHRAUBE 8400775
ADJUSTING SCREW
KABELBINDER 8712007
ohne Abbildung
CABLE CLIP (2,4X92)
without illustration
SCHALTVENTIL Bild/Picture 9
PILOT VALVE
Datei/File: 61733.008
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 27 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
2 RUNDSCHNURRING M09361
TOROIDAL SEALING RING
3 O-RING M16958
O-RING SEAL
DOSIERVENTIL 55 8409965
METERING VALVE 55
DOSIERVENTIL 75 8409966
METERING VALVE 75
PEEP/PIP-VENTIL,KOMPL. Bild/Picture 10
PEEP/PIP VALVE, CPL.
Datei/File: 61733.009
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 29 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 PEEP/PIP-VENTIL 8409150
PEEP/PIP VALVE
1 RAT-PEEP/PIP-VENTIL,(8409150) 8410255
REP.EXCH.PEEP/PIP-VALV(8409150
3 PLATTE 8407746
PLATE
4 HALTER,KOMPL. 8409207 10
SUPPORT, CPL.
5 BUCHSE 8408312
SOCKET
INSPIRATIONSBLOCK Bild/Picture 11
INSPIRATION BLOCK
Datei/File: 61733.010
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 31 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 INSPIRATIONSGEHAEUSE 8409160
INSPIRATION HOUSING
2 SCHRAUBE 8409158
SCREW
3 O-RING 8403962
O-RING SEAL
4 O-RING E20566
O-RING SEAL
5 DROSSELSCHRAUBE E08481
CHOKE SCREW
8 DICHTRING 8409187
PACKING RING
9 O-RING D20119
O-RING SEAL
10 O-RING E22651
O-RING SEAL
11 KRATER 8409105
CRATER
12 SCHRAUBE 8409157
SCREW
13 DICHTSCHEIBE 8409823
SEALING WASHER
14 TELLER 8409167
PLATE
15 MEMBRAN 2M08762 2
DIAPHRAGM
16 DICHTPFROPFEN E09765
nicht in allen Geraeten vorhanden
SEALING STOPPER
existing not in all devices
17 FEDER R20262 10
nicht in allen Geraeten vorhanden
SPRING
18 DECKEL 8409101
COVER
22 FEDER T03886
SPRING
23 DISTANZSTUECK 8409109
SEPARATOR
24 MEMBRANE 8403352 4
DIAPHRAGM
25 ROSETTE 8406059
ROSETTE
27 DECKEL 8409102
COVER
29 MUTTER 8409988
NUT
31 DECKEL 8409188
COVER
36 SCHRAUBE 8409989
SCREW
37 MIKROSCHALTER 8407265
MICROSWITCH
38 BLECH 8409177
SHEET METAL
40 RUECKSCHLAGVENTIL,UGR. M27711
bis Geräte-Nr. ARHJ-xxxx
NONRETURN VALVE
up to device no. ARHJ-xxxx
41 FILTEREINSATZ 8411342
bis Geräte-Nr. ARKD-xxxx
FILTER CARTRIDGE
up to device no. ARKD-xxxx
42 O-RING R31296
O-RING SEAL
43 TUELLENAUFNAHME 8408942
NOZZLE RECEIVER
46 ZYLINDERSCHRB.DIN913-M2X4-A4 1316575
CHEESE HEAD SCR.DIN913-M2X4-A4
50 RING 8410476
RING
EXSPIRATIONSVENTIL Bild/Picture 12
EXSP.-VALVE BO 8000
Datei/File: 61733.011
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 36 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 EXSPIRATIONSGEHAEUSE 8408940
EXSPIRATION HOUSING
2 SCHEIBE 8410820
WASHER
3 STAUROHR 8409204
PRESSURE TUBE
4 SCHRAUBE 8409157
SCREW
5 O-RING R31296
O-RING SEAL
6 TUELLENAUFNAHME 8408942
NOZZLE RECEIVER
9 ZYLINDERSCHRB.DIN913-M2X4-A4 1316575
CHEESE HEAD SCR.DIN913-M2X4-A4
10 HEBEL 8409949
LEVER
13 TYPENSCHILD D17428
TYPE PLATE
15 O-RING E20274
O-RING SEAL
16 SAUGROHR 8408948
SUCTION PIPE
17 MISCHROHR 8408947
MIXING TUBE
18 O-RING R22383
O-RING SEAL
19 BUCHSE 8410016
SOCKET
20 GEHAEUSE 8409987
HOUSING
21 ZYLINDERSCHRB.DIN913-M2X4-A4 1316575
CHEESE HEAD SCR.DIN913-M2X4-A4
22 DAEMPFER 8409986
DAMPER
24 KUGELKAEFIG 8409934
BALL RETAINER
25 MUTTER M19189
NUT
28 FEDER M25184
SPRING
29 UEBERGANGSSTUECK 8409939
ADAPTER
30 ROHR 8409171
TUBE
32 BLECH 8409170
SHEET METAL
37 DECKEL 8409064
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 38 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
38 MEMBRAN 8409184 2
DIAPHRAGM
39 DICHTRING R10247
PACKING RING
40 KRATER 8408945
CRATER
41 O-RING R31296
O-RING SEAL
42 O-RING 8403962
O-RING SEAL
Datei/File: 61733.012
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 40 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
Datei/File: 61733.013
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 41 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
BABYLOG-DISPLAY,F. 8305870
BABYLOG DISPLAY, F.
BABYLOG-DISPLAY,AM. 8306027
BABYLOG DISPLAY, AM
1 ANZEIGENFELD,VOLLST. 8305380
DISPLAY, CPL.
1A ANZEIGEFELD,VOLLST.F. 8305867
DISPLAY, CPL., F
2 DISPLAY 8305340
DISPLAY
11 HALTER 8305568
SUPPORT
13 SCHRAUBE 8408347
SCREW
14 POTENTIOMETERFELD,VOLLST. 8305390
POT FIELD, CPL.
15 LED,GRUEN 8305024
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 43 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
23 SCHARNIER M29632
HINGE
ELEKTRONIK,VOLLST. Bild/Picture 15
ELECTRONIC, CPL.
Datei/File: 61733.014
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 45 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
ELEKTRONIK,VOLLST. Bild/Picture 16
ELECTRONIC, CPL.
Datei/File: 61733.015
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 46 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
4 KABEL 1817167
CABLE
6 FILTERMATTE 8305367 10
FILTER MAT
7 LAUTSPRECHER 8305490
LOUDSPEAKER (1827790)
8 BLECH 8305457
SHEET METAL
10 BATTERIEANSCHLUSSKABEL 8305489
BATTERY CONNECTING CABLE
11 LUEFTER 8305488
FAN
INTERRUPTENCODER 8306390
INTERRUPT ENCODER
Datei/File: 61733.016
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 50 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 PATIENTENTEILHEIZUNG 8305480
PATIENT PART HEATING
6 ATEMDRUCKSENSOR 8305623
SENSOR
7 VORDRUCKSENSOR 8305624
ADMISSION PRESSURE SENSOR
VENTILBANK Bild/Picture 18
DIGITAL VALVE ARRAY
Datei/File: 61733.019
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 52 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1 VENTILBLOCK 8409962
VALVE BLOCK
9 DOSIERVENTIL 75 8409966
METERING VALVE 75
10 DOSIERVENTIL 55 8409965
METERING VALVE 55
11 DOSIERVENTIL 40 8409964
METERING VALVE 40
12 TUELLE 8408197
SOCKET
15 WINKELSTECKANSCHLUSS M30961
ANGLE CONNECTION
18 DICHTRING R15422
PACKING RING
20 GIMETALL-LAGER 2M06157
RUBBER METAL CONNECTION
Datei/File: 61733.017
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 55 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
Datei/File: 61733.018
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 56 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
1a WENDEBLATT HF 8411170
REVERSAL SHEET HF
9 SCHLAUCHANSCHLUSS,UGR. 8409685
HOSE CONNECTION
10 FEDER E20373
SPRING
11 TOPF 8403976
POT
13 STOESSEL 7 8409615
TAPPET 7
14 FEDER M09934
SPRING
15 KAPPE 8410309
CAP
17 FEDER 6850050
SPRING
18 KAPPE 8410308
CAP
25 SENSORGEHAEUSE 8410186
SENSOR HOUSING
26 DICHTRING 8409699
PACKING RING
32 WASSERBEHAELTER 8409660
WATER TANK AQUAMOD
35 BEHAELTER,KOMPL.,UGR. 8409858
RECEPTACLE, CPL.
36 DECKEL,KOMPL.,UGR. 8409731
COVER, CPL.
38 BEHAELTER,KOMPL.UGR. 8409725
RECEPTACLE, CPL.
39 GELENKARM 8409825
HINGED BRACKET
42 KLEMME,KOMPL. 8410075
BINDER, CPL.
45 RAENDELSCHRAUBE 8410077
KNURLED SCREW
46 GEWINDESTUECK 8410067
SCREW THREAD
49 SCHLAUCHKLEMME 2 8404761
TUBE CLIP 2
50 ANSCHLUSS 8410066
CONNECTION
51 RUNDSCHNURRING 2M08777
O-RING SEAL
52 WINKEL 8410237
ELBOW
54 PLATTE 8410238
PLATE
55 KONUS 8411018
CONE
56 SCHRAUBE 2M15332
SCREW
58 RASTE 8411077
CATCH
59 KLAMMER 8411076
CLAMP
60 ZYLINDERSCHRB.BZ2,9X13DIN7971 1334948
CHEESE HEAD SCR.DIN7971 2,9X13
Datei/File: 61733.021
Druckdatum: 20.02.2001 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 6173.3 Ausgabe/Edition
Spare parts list 20.02.01
BABYLOG 8000/ 8000 SC/ 8000 PLUS Seite/Page 62 von 62
BABYLOG 8000/ 8000 SC/ 8000 PLUS
2 PLATTE 8410123 10
PLATE
3 SCHIENE 2M17541
BAR
4 VERSCHLUSS-STOPFEN G60455
VENT PLUG
7 ROHR 2M13388
TUBE
9 SCHRAUBE 2M17338
SCREW
10 SCHILD 8408289
PLATE
17 DOPPELLENKROLLE 2M85379
DOUBLE GUIDE ROLL
19 KOMPAKTSCHIENE,KOMPL. 2M85337
COMPACT RAIL, CPL.
Moislinger Allee 53 – 55
D-23542 Lübeck
Federal Republic of Germany
6173.3 e
Subject to modification.
Will not be replaced in the event of modifications.