Service Information
1
2000
Optima 2000 Optics
The echelle grating has 79 lines/mm with a blaze angle of 63.8°.
The grating is used in the higher orders, where the high
efficiency and high dispersion allow a relatively short focal
length resulting in a compact optical system.
3
2000
2000 Optical Path with DWS
4
2000
Scanning CCD Detector Overview
2 mm CCD arr ay
Analyte Measurement
Section Output Register for the
analytical measurement
3 mm
l
5
2000
2000
Reference and Analyte Displays
6
2000
Optima2000 Spectrometer Overview
P5RBSPE1-Board Valves,Flow sensor
Prism & Grating 5phase Motor-
5phase Motors Drivers P5RACOBD P5RACONN- Transfer X, Y, Shutter,
- Board Board & DualView Motors,
Shutter sensor
Power- DC
Supply / DC
2phase Motor- Slit Motor
TPU
Drivers
P5RDCCDS-Board
Customer QSPI 8chan. Temperture,
Interface 12bit Pressure, Peltier-
FPGA ser. A/D Control
...
Micro- ACT-
Processor 1280
Sys tem
pSOS-Debug (68332) CCD-MOS- Ham amatsu-
Interface Driver CCD - Detector
RAM
Timing-
1MB
RAM
Diagnostic- Flash
Interface 1MB
16bit A/D Signal- MUX
BRAM FIFO AD976 Conditioner
8kB
P5RANEON-Board
Com munication
PC Interface LED‘s
Neon Lamp
7
2000
Optima2000 Spectrometer Controller Board
8
2000
Optima 2000 Installation
9
2000
10
2000
11
2000
Cabling Connections
12
2000
13
2000
14
2000
15
2000
Removing the Spray Chamber
16
2000
Installing the Injector
17
2000
Injector Positioning
Injector Positioning
18
2000
19
2000
20
2000
21
2000
22
2000
23
2000
Connecting the Sample Tubing
24
2000
Connecting the Sample Tubing
25
2000
Connecting the Drain Tubing
26
2000
Connecting the Drain Tubing
27
2000
Adjusting the pressure on the Sample tubing
28
2000
Adjusting the Pressure on the Drain tubing
29
2000
Setting the Shear Gas Nozzle Position
30
2000
Aligning the Torch Position
Set Torch
Alignment
position for
-3 when using
a Scott Spray
Chamber
31
2000
Plasma Parameters
32
2000
Plasma Parameters for Organic Settings
33
2000
View X,Y
Align. Do
both Axial
& Radial.
34
2000
Optima 2000 Power on Routine
35
2000
36
2000
Default Power on Routine
37
2000
Default Power on Routine
38
2000
Default Power on Routine
Step 4. Continued
4.
39
2000
Default Power on Routine Continued
40
2000
Default Power on Routine Continued
41
2000
Optima2000 Spectrometer Controller Board LED‘s
Led18 Led17 Led16 Led15 Led14 Led13 Led12 Led11 Led10 Led9 Led8 Led7 Led6 Led5 Led4 Led3 Led2 Led1 Led0
DMMY Ta sk is runni ng
CNTA Ta sk is runni ng
Not used /powerup
Flow s ensor/Presure sensor off
Valve 1 off
Valve2 off
Power on
Halt => microproces sor stopped
PLE => m easurement is running , can be modified by command „serv:led:sig x“ x = [ lintest,tclk,fwrt,treserv,starttim,tramsel,measrun,read]
42
2000
43
2000
PCB Interconnect
44
2000
Power Block Diagram
45
2000
Power Distribution
46
2000
47
2000
48
2000
CCDS Sensor Design
49
2000
Scanning CCD Detector Overview
2 mm CCD arr ay
Analyte Measurement
Section Output Register for the
analytical measurement
3 mm
l
50
2000
2000
Photon Detection
51
2000
CCD Peltier Control
52
2000
53
2000
Prism Temperature Sensor
54
2000