Hart Mode Users Guide: Documenting Process Calibrator
Hart Mode Users Guide: Documenting Process Calibrator
744
Documenting Process Calibrator
HART Mode Users Guide
PN 691292
December 1998
© 1998 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Table of Contents
Title Page
Introduction ..................................................................................................................... 1
How to Contact Fluke ..................................................................................................... 2
Connecting to a HART Transmitter ................................................................................ 3
Supported vs. Generic Transmitters ............................................................................... 6
Communication Operations ............................................................................................ 8
Viewing Process Variables ........................................................................................ 8
Setup Operations ....................................................................................................... 8
Basic ..................................................................................................................... 8
Sensor................................................................................................................... 10
Device Identification .............................................................................................. 10
HART Output ........................................................................................................ 11
HART Information ................................................................................................. 11
Service Operations .................................................................................................... 12
Abort Softkey ............................................................................................................. 12
Interaction between Analog Mode and HART Mode ...................................................... 12
Understanding HART Calibration ................................................................................... 13
744 HART Mode Menus for Adjustment......................................................................... 13
i
744
HART Mode Users Guide
ii
List of Tables
iii
744
HART Mode Users Guide
iv
List of Figures
v
744
HART Mode Users Guide
vi
Documenting Process Calibrator HART Mode
1
744
HART Mode Users Guide
2
Documenting Process Calibrator HART Mode
Connecting to a HART Transmitter
3
744
HART Mode Users Guide
4
Documenting Process Calibrator HART Mode
Connecting to a HART Transmitter
-PS S
+PS
MEAS
mA SETUP
M
SOURCE
V V TC -IN Z
Hz RTD
7 8 9 CLEAR
( ZERO)
+IN
4 5 6
1 2 3
ENTER
0 .
V mA mA V
RTD
RTD
CAT
30V SOURCE MEAS 300V
MAX 30V 30V MAX
TC
MAX MAX
pe01c.eps
5
744
HART Mode Users Guide
Note
Sensor trim is provided for supported
transmitters, with a few exceptions, as identified
in the list of supported transmitters that you can
pe07s.bmp
view on the display.
4. Press u or d to highlight the desired manufacturer,
To display a list of supported transmitters and software and press e. A list of model numbers appears.
versions:
5. Press u or d to highlight the model number, and
1. If the calibrator is connected to a HART transmitter press e. A list of software versions appears.
with the Active Device screen showing, press the
Abort softkey, followed by More Choices.
6
Documenting Process Calibrator HART Mode
Supported vs. Generic Transmitters
7
744
HART Mode Users Guide
pe08s.bmp
pe09s.bmp
8
Documenting Process Calibrator HART Mode
Communication Operations
(Example) • Tag
PV • PV units
PROCESS % Range • LRV, URV
Basic • Sensor serial number
AO • Damping
SV, TV, QV • Sensor lower and
• Transfer function
upper limits
• Sensor minimum
Sensor span
(temperature devices only)
• Sensor type
• Sensor connections
• Software version
• Final assembly number
SETUP Device • Tag
Identification • Message
• Date
• Description
• Write protect
HART • Alarm state
Loop test Output • HART poll address
• HART burst mode
Pressure • Manufacturer • HART burst
zero trim • Model command
SERVICE
HART • Device HART ID
Output trim Information • Software revision
• Hardware revision
Sensor trim • Number of preambles
ABORT
pe03f.eps
Figure 2. 744 HART Mode Menu Tree
9
744
HART Mode Users Guide
pe11s.bmp
pe10s.bmp
10
Documenting Process Calibrator HART Mode
Communication Operations
pe12s.bmp
pe13s.bmp
11
744
HART Mode Users Guide
12
Documenting Process Calibrator HART Mode
Understanding HART Calibration
13
744
HART Mode Users Guide
744 Calibrator
4-20mA
Physical Input Computation Output
loop
variable
HART transmitter
pe02f.eps
Figure 3. Block Diagram of a HART Transmitter
14
Documenting Process Calibrator HART Mode
Calibrating a Supported HART Transmitter
If you are starting a calibration procedure from a 3. You are presented with a set of choices for analog
loaded Task, connect to the HART transmitter mode. Use the u and d keys to select one of the
first and establish communication before you measure/source choices. This is where you select
press Task in analog mode. measure mA (analog output), or measure PV, which
does not involve the Output stage of the transmitter.
The following procedure assumes that you are familiar with See Figure 3.
using a 740 Series Documenting Process Calibrator to
calibrate analog transmitters, and are not running a loaded 4. Press e.
Task.
5. Press the As Found softkey.
1. Make the appropriate measuring, sourcing, and HART
interface connections between the calibrator and the 6. Use the u and d keys followed by e to select
transmitter, and press r to establish an instrument calibration procedure. You’ll see that
communication. the calibration template is preloaded with appropriate
data. You can make changes to any item if you wish.
2. Press r to switch to analog mode.
15
744
HART Mode Users Guide
10. Press the As Left softkey and repeat the calibration 2. Press a softkey to command the transmitter to set its
procedure to verify that the transmitter passes. analog output to the corresponding value, or manually
enter a value with the numeric keys.
Loop Test
In the middle of the screen the transmitter acknowledges
The Loop Test feature sends a command to the transmitter the selected PVAO (digital representation of the Output
to set its output stage to a specified value. Use this to stage). The calibrator shows the measured value at the top
check the calibration of the output stage, or to check for of the screen, so you can use Loop Test as a quick way to
proper reading on an external loop reading device. check the calibration of the transmitter’s Output stage.
16
Documenting Process Calibrator HART Mode
Calibrating a Supported HART Transmitter
pe16s.bmp
17
744
HART Mode Users Guide
2. Use the u and d keys to select the Sensor Trim 5. Enter the trim value you want the calibrator to
operation, and press e. produce, and press e, or in the case of a
pressure transmitter (as above), press Fetch to load
3. Follow the instructions on the display. For pressure, the value being measured by the Fluke pressure
there is a message to connect a pressure module to module into the dialog box.
the calibrator, and another to press the c key to
zero the pressure module. 6. Press Trim.
Note
Select trim points near the range limits unless the
transmitter’s documentation specifies otherwise.
If you get an error message that for the selected
trim point there is an excess correction,
sometimes you can solve the problem by doing a
series of Sensor Trims in small steps from the
pe18s.bmp previous trim point toward the desired trim point.
18
Documenting Process Calibrator HART Mode
Cloning a Transmitter
Cloning a Transmitter 4. Press the Store Page softkey to load the settings into
the calibrator’s memory (not the transmitter).
Cloning copies the Basic Setup information from one
transmitter to another. You can clone generic and 5. Disconnect the calibrator from the transmitter and
supported transmitters. connect it to the transmitter you want to configure.
Press the Abort softkey to establish communication.
To use the calibrator to clone a transmitter:
6. Press the Setup softkey, then select Basic to return to
1. Connect the calibrator to the transmitter you want to the Basic Setup screen.
clone. You only need a communication connection.
7. Press the Recall Page softkey to refresh the calibrator
2. From the Active Device screen, press the Setup display with the cloned parameter settings. The
softkey. settings are not transmitted to the transmitter at this
time.
3. Use the u or d keys to select Basic from the list on
the screen and press e. 8. Use the u or d and ekeys to select
parameters one at a time, change them, or leave them
as cloned, and press the Send softkey to transmit
them to the transmitter.
pe09s.bmp
19
744
HART Mode Users Guide
20
Index
21
744
HART Mode Users Guide
22