We are pleased to congratulate you on obtaining the cooling incubator with forced circulation of air
and cooling which enables a timely, spatially precise moderate heating of samples. A unique cooling
system offers an accurate and economical simulation of selected natural processes and reduces
the evaporation of samples. They can be applied in the sphere of biotechnology, botany, zoology,
food processing industry, cosmetics, chemistry etc. where it enables extreme short times of thermal
condition regeneration.
The units meet technical and legislative requirements and they are designed according to respective
EN standards. The units are made of high quality materials by using the latest technology. Each piece
undergoes a careful output control.
Provided you will follow the instructions mentioned here the unit becomes your reliable and powerful
partner.
Now these advantages will be available just for you. This unit will help to solve your everyday
problems and it will become powerful assistant for you. This unit is very easy to use, nevertheless, we
recommend you to read the Instructions for use carefully so that you could use all advantages of this
unit and obtain complete knowledge for its optimal use.
CONTENTS:
20. 3. 2013
1 GENERAL NOTES..................................................................................................................3
2 IMPORTANT INSTRUCTIONS................................................................................................3
2.1 Unpacking and checking.............................................................................................................3
2.2 Before starting the unit..........................................................................................................3
2.3 Operation.....................................................................................................................................4
2.4 Permitted loads........................................................................................................................4
2.5 Wiring and other conditions....................................................................................................4
5 MAINTENANCE.......................................................................................................................27
5.1 Sealing exchange and door setting..........................................................................................27
5.2 A container collecting the condensed water steam................................................................28
5.3 Cleaning the cooling incubator.............................................................................................28
5.4 Melting............................................................................................................................29
5.5 Inspection of electrical components.......................................................................................29
6 TECHNICAL DATA.................................................................................................................30
6.1 C hambers ..........................................................................................................30
6.2 Shelves with exposure lighting...............................................................................................31
10 OPTIONAL EQUIPMENT.......................................................................................................32
10.1 Inner lighting..........................................................................................................................32
10.2 Bushings of diameter 25, 50, 100 mm.........................................................................................32
10.3 Lockable door..........................................................................................................................32
10.4 Left door.................................................................................................................................33
10.5 Independent sensor PT100.......................................................................................................33
10.6 Supportive sw for pc.............................................................................................................33
10.6.1 Recording sw - warmcomm - for pc under windows.......................................................33
10.6.2 Recording printing sw - printer archiv - for pc under windows........................................33
10.7 Potential-free contact for alarm reports.............................................................................33
10.8 The inner switched socket......................................................................................................33
10.9 Exposure lighting a) in the door; b) on the shelves...............................................................34
10.11 Temperature range from - 9.9 °C..............................................................................................34
10.12 Protection of the programme setting by a chip card..............................................................34
10.13 Automatic defrosting...............................................................................................................34
10.14 Signalling of the door opening..............................................................................................34
10.15 Door blocking..........................................................................................................................34
APPENDICES:
EC - DECLARATION OF CONFORMITY
Instructions for use
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 3
Instructions for use
If the unit is not used for a longer time, 2.4 permitted loads
disconnect it from the mains by pulling the
connecting cord out of the socket. size weight [kg]/tray total weight [kg]
22 10 25
The mains cord must not come into contact 55 20 50
with hot parts of the unit. 111 20 50
222 30 70
Protection of the heating cabinet, its 404 30 100
surroundings and the processed material 707 50 130
against inadmissible temperature
exceeding is ensured by a protective
thermostat corresponding to EN 61010-2-
010. 2.5 wiring and other
conditions
The devices must always stand on the legs
(or wheels) on a flat mat (floor). Do not Basic data for connection:
incline! Do not turn over! Mains FC 22 100 - 240V/50(60)Hz
connection: FC 55-707 1x230V/50(60)Hz
Minimum distance of the rear and side wall 1x115V/50(60)Hz
of the unit from other objects and walls is (standard types are marked with bold face)
100 mm. In 22-liter models, free space of Mains voltage variation ±10 %
100 mm should be left above the device in Protection against dangerous contact - I
addition to the above mentioned distance class:
requirements. External circuits isolation:
– double isolation
Shifting the upper sheet of the inner Type of unit plug:
chamber out and in must be carried out – as a standard CEE-7/VII, IEC-83/CH,
carefully, there is a danger of cutting the 16 A/250 V (or another according to the type)
chamber seal through as a consequence of Socket protection:
careless manipulation. – 16 A (acc. tech. parameters in the Instructions fo
use of the unit)
After the device is positioned to the desired Protection IP 20
place, turn and brake the wheels (404 and according to
707 models only). The front wheels shall EN 60529:
be turned forwards while the back ones Overvoltage category according to
backwards, in line with the device side (IEC 664 - EN 61010):
walls, to provide the device with the – II in case of pollution degree 2
maximum possible backward and forward Fuses on the rear Blow-out fuse T16A/1500
stability. wall of the upper
piece
Ambient conditions:
– ambient temperature: +5 °C to +40 °C
– max.relative humidity: 80 % at 31 °C
– maximal altitude: 3000 m
4 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
X (V)1
X (V)1
X (Š)
X (Š)
� � � � �
� � � �
1. Síťový vypínač
2. Ovládací panel
3. Plastový kryt regulátoru
4. Kondenzátor obr. 6: Konektor pro zapojení tiskárny
5. Kompresor (chladivo R134a) FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
9 pin Canon konektor na skříni
5
6. Chladicí trubkový had – 6a, 6b
Kolík Signál
Instructions for use
4
RS 232C must fulfill valid regulations and be
approved by a competent testing laboratory. function and
operating
3.4 operating panel
Function of indicator lights and control keys is
8 9 10 11 12 13
described in paragraph 3 - Description of the
unit.
Program Segment °C
6 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
4.3.1 activating the service mode 4.3.4 service 03 – set the real time
and date
Press simultaneously buttons ▲► for 2 seconds.
LCD shows Service 01. By pressing the buttons Pressing START/STOP you enter the service.
◄ and ► you can change between the single Set the real time and date by pressing buttons
service modes Service 01 – Service 07, Service ◄► and ▲.
41, 58 and 60.
Position 1 2 3 4 5 6 7 8 9 10 11
Content D F m t : _ R R R R -
12 13 14 15 16
M M - D D
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 7
Instructions for use
Cursor in position 6: Select the data separator by Terminate and close the service by START/STOP
the key ▲. There are the following possibilities: pressing.
Dash (-), stroke (/), dot (.), colon (:). Dash is
default. The key ◄ takes you back to the time
setting. 4.3.6 service 05 – select language
Cursor in position 7: Select the data format by Pressing START/STOP you enter the service.
the key ▲. There are the following possibilities To select the language use button ▲. To quit the
for the Czech language: RRRR‑MM‑DD, service press START/STOP again.
RRRR‑DD‑MM, MM‑DD‑RRRR, DD‑MM‑RRRR.
The default values are RRRR‑MM‑DD (Czech),
YYYY‑MM‑DD (English), JJJJ‑MM‑TT (German), 4.3.7 service 06 – setting the
AAAA‑MM‑JJ (French). The key ► takes you switching times of the inner
back to the time setting. socket (optional) and the fan
run
The service shall be terminated and closed by
pressing the button START/STOP. The data A) Switching times setting
format on the display: hour:min_day.month.year. With help of this service you can set the time of
The selected format will be shown both on the the inner socket periodic switching on and off.
LCD display and the printer. Press START/STOP to enter the service.
On the bottom line of LCD there is a read out:
“on: XXPP off: YYQQ“, where
4.3.5 service 04 – set interval on XX is the number of time units during which the
printer socket is on for a part of a period (00 – 99)
YY is the number of time units during which the
To print protocols device can be set up (optional) socket is off for a part of a period (00-99)
with a printer, which is connected by a cable to PP time units for switching on (min or hr)
the printer interface – see chapter Printing the QQ time units for switching off (min or hr)
protocol. Enter the service by START/STOP and Move the cursor by pressing ◄► change the
set the interval: number of the time units or unit by pressing ▲ If
XX = 00, the socket will never be on, if XX ≠ 00
Pressing ▲ select one of the following choices: and YY = 00, the socket is on all the time after
- printer OFF the start.
- switched over to PC (application SW
WarmComm – data are sent to PC) B) Setting the fan run after the programme
- interval 10 s termination
- interval 01 min Leave the “inner socket” setting by means of
- interval 01 h. ◄► and by pressing ▲, set:
a) “Prog” – the fan runs only during the P1-P6
By pressing ◄ move the cursor to the position run;
10/01 seconds, minutes or hours. By pressing b) “TXX°C” – the fan keeps running after the
▲ you can change the printer interval within the programme termination till the temperature
following range: decrease below the preset value XX °C;
c) “Door” - the fan keeps running after the
10 to 50 s (step 10 s) programme termination and switches off after
01 to 59 min (step 1 min) the door opening (applied only for units with a
01 to 12 h (step 1 h). closed door sensor);
d) “alws” – the fan keeps running till the cabinet
To quit the service press START/STOP again. is switched off by the ON/OFF button.
8 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 9
Instructions for use
10 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
b) Shelf lighting (one type of the lighting - The cumulated dose is displayed in kiloLux-
source): hours (eVi if visible light is measured)
See the chapter 6 Parameters, table 6.2. or miliwatt-hours/cm2 (eUV if UV light is
measured).
- The overall time (tVi or tUV) is displayed in
4.3.17 service 60 – racks with minutes, hours, and days.
exposure lighting –
photovalues setting When checking the measured values during the
programme operation (see chapter 4.11), the
There are two possibilities how to set the lighting: light variables are displayed in the following way:
A) No possibility to control the illuminated - Actual value of the visible light (e.g.“L-Vis1” if
shelves: Cumulated light doses and overall the irst channel measures the visible light) in
times at the individual shelves can only be klux,
displayed/zeroed. - Actual value of the ultraviolet light (e.g.
“L- UV_2 if the second channel measures the
Service 60 selection: UV light) in mW/cm2,
- Cumulative visible light dose (e.g. “eVi1”) if
S e r v i c e 6 0 the first channel measures the visible light) in
kluxh kilolux-hours),
P h o t o v a l u e s s e t .
- Cumulative ultraviolet light dose (e.g. “eUV2”)
if the second channel measures the UV light)
Displaying examples after Service 60 activation: in mWh/cm2 (milliwatt-hours per one square
centimetre)
S e r v i c e 6 0 - Overal exposure time (e.g. tVi1, tUV2) in
minutes, hours and days.
1 : e V i 1 6 . 3
B) Illuminated shelves control allowed:
S e r v i c e 6 0 Parameters for the control of the individual
shelves lighting can be displayed/set/zeroed, or
2 : t U V 0 d 0 0 h 0 0 m the last finished light exposure protocol can be
printed.
S e r v i c e 6 0 Position 1: Rack with lighting is selected;
3 : - possibilities: 1, 2, or 3
(the racks are numbered from the top, i.e. the
highest rack has the number 1); position 14
Explanatory notes: then will show the exposure type (UV, Vis – the
- Select the measuring channel (1 to 3) by exposure type is set at the factory or by the
means of ▲ on the second line, first position. service technician – see the Service Instructions,
The given variable type or sign “ -“ (dash) is Service Technician Menu).
displayed after the colon if the channel is not
measuring any light variable. Position 3: The displayed parameter is selected;
- Select the type of variable displayed by means possibilities:
of ▲ on the second line, third position. There a) T-W: Required time of switching on the given
are the following possibilities: “e” (exposure) rack exposure in the format hhh:mm (hours
or “t” (time). The variable displayed is zeroed and minutes);
by X/W pressing. If both the exposure and b) T-X: Actual time of switching on the given rack
time have to be zeroed, exposure (e) must be exposure in the format hhh:mm (hours and
displayed and X/W pressed, and then time (t) minutes),
must be displayed and X/W pressed again. c) e-W: Rack without a photosensor: The input
- The second line, fourth position displays exposure value in a defined reference point,
either “UV” or “Vi” depending on whether measured by external measuring device (kLux
the ultraviolet (UV) or visible (Vi) light for Vis exposure, mW/cm2 for UV exposure).
measurement has been set on this channel. Rack with a photosensor: Overall required
The type of the measured light variable is set exposure dose (kLuxh for Vis lighting, mW/
by the manufacturer or service technician. cm2h for UV lighting),
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 11
Instructions for use
d) e-X : Total exposure dose. 6. After the required temperature is reached, the
Rack without lighting: The dose is calculated lighting is switched on and measurement of
as a product of e-W and T-X (kLuxh for Vis the exposure time (dose) starts.
exposure, mW/cm2h for UV exposure). Rack 7. By means of ◄ or ► buttons, you can
with a photosensor: The dose is an integrated display (on the upper line of the display) all
measured value. parameters during the programme run: T-W,
e) Printout: Last protocol printing (started by X/W T-X, e-W, e-X, and the state of Vis and UV
button). lighting switches (On, Off).
12 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
- If e-W parameter is zero on the rack with Dose specified by T-X parameter
photosensors, but T‑W time is not zero, the ----------------------------------------------------------
dose will be terminated in the moment when
the measured time T-X reaches T-W value. PROTOCOL POLICE Vis-1 RACK PROTOCOL Vis-1
- The last protocol of each rack can be Zacatek mereni: Measur.start:
subsequently printed in the user mode of 22-06-2009 11:17 22-06-2009 11:17
Service 60. In a not terminated dose, the T = 29.9°C T = 29.9°C
paragraph “End of measurement” will not be Konec mereni: Measur. end:
printed. 22-06-2009 11:27 22-06-2009 11:27
- After language switching in Service 5, the T = 29.9°C T = 29.9°C
same protocol can be printed in another Pozadovany čas: Required time:
language. 000:10 000:10
- To print a protocol, the printer must be Celkovy cas: Total time:
switched on in the user service 04 (recording 000:10 000:10
mode). Celkove osvetleni: Total exposure:
- In FC device, RH will also be printed on the 2.71 kLxh 2.71 kLxh
protocol in addition to the temperature at the ---------------------------------------------------------
beginning and end of the dose. Datum: Date:
22-06-2009 11:50 22-06-2009 11:51
Examples of protocol printouts:
JMENO: NAME:
Dose specified by e-X parameter
---------------------------------------------------------- Podpis: Signature:
ATTENTION!
Do not reduce the fan speed if it is not
necessarily needed. Reduced fan speed may
cause reduced accuracy!
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 13
Instructions for use
4.4.2 lcd read out shown The unit operates on the set temperature until
defined run down time (max. 9999hr 59min) is
Program Segment °c
over.
P1 / – 00 . 0 00 : 00
4.5.1 set p2
100
– With help of buttons ◄► put the cursor to the
RH 5 6 position Program and switch to the program P2.
Set the temperature at the position °C. Move the
cursor to the position and set the run down
time between 00hr:00min to 9999hr.
4.4.3 start p1
Set the ventilator rate at the position tfrom 10,
Press START/STOP, a sound will appear and the 20... up to 100 % (the value of 100 % only is pre-
“windmill“ in the first digit of LCD starts running set in PM models).
clockwise.
ATTENTION!
See the following LCD read out: Do not reduce the ventilator rate if not necessary.
Program Segment °c The reduced ventilator rate could influence the
regulation of temperature and RH negatively!
*1 / – °C °C . °C h h : mm
100 4.5.2 lcd read out shown
Program Segment °c
P2 / – / 00 . 0 00 : 00
RH 5 6
4.5.3 start p2
4.4.4 information on the set
parameters during the Press START/STOP, a sound will appear and the
device’s running “windmill“ in the first digit of LCD starts running.
Program Segment °c
Press X/W and all information on the set
parameters are shown at once. All possible
parameters in all possible programs can be set
*2 / – / °C °C . °C h h : mm
including the one which is presently active, as
long as the change is effective. 100
RH 5 6
4.5 program p2
14 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
Press START/STOP.
4.6.3 start p3
4.6 program p3
Press START/STOP, a sound will appear and the
The delayed start begins after the defined run “windmill“ in the first digit of LCD starts running
down time has been reached. clockwise. See the following LCD read out:
[The function starts for example at the weekend.
Thus your goods are ready on the first day of
return to work.] 4.6.3.1 P3 and segment 1 in the graph
are active
P3= Program Segment °c
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 15
Instructions for use
4.6.3.3 p3 and the third segment in the Digit 11 on LCD, placed between temperature
graph are active and dwell time values, shows the following
information:
Program Segment °c
r = ramp value, can be set on digits 12 – 16
* 3 – / – °C °C . °C h h : mm and corresponds with temperature change
speed in °C/min
max. value is +3.20 °C/min, when T0 < T1,
100 max. value is –1 °C/min, when T0 > T1
If r = 00.0, no ramp active!
RH 5 6 (Variations of maximum ramp rates by
different temperatures may appear due to
= third
třetí segment
segment physical limits of heating power.)
t = time available in position (00hr:00min
– 9999 hr), at T2 the dwell time can be
The preset temperature has been reached. If unlimited
the procedure duration is unlimited, the time c = cycles shown in position , digits 12 – 13 on
passed is displayed. If the procedure duration is LCD (available only at T2, if the dwell time is
limited, the time remaining till its end is displayed. unlimited).
After the programme ends, the temperature in Possible number of cycles is 01 to 99,
the chamber and the text “End” are displayed. eventually ∞ which is the next position after 99
Both the programme end indication and the (the cycles are possible on condition, that after
temperature in the chamber disappear after any the end of the interval on T2 the temperature
panel key pressing. drops below the value T0).
16 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
4.7.2 set the segment for T1 4.7.2.6 set the fan speed at the
temperature T1
4.7.2.1 set temperature T1
Put the cursor to the position and change
Put cursor to the position Program and switch the fan speed between 10, 20...100 % (In case of
to P4. By pressing ◄►place the cursor on the volume 22 only 100 %).
graph and make sure that T1 ( /¯\_ ) is activated.
If not, activate it with button ▲. ATTENTION!
Change to the position °C - setting the Do not reduce the fan speed if it is not
temperature - and set the temperature between. necessarily needed. Reduced fan speed may
cause reduced accuracy!
Program Segment °c
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 17
Instructions for use
4.7.3.5 r
ead out of T2 and the ramp See the following LCD read out:
The graph shows the actual state of the program.
Program Segment °c
T1 T2
P4 / – / – 00 . 0 r +/- 0 . 00 S1 S2 S3 S4
100 P4 =
RH 5 6
Program Segment °c
100 –
* 4 / / – °C °C . °C h h : mm
RH 5 6
100
4.7.3.8 set the fan speed at T2 RH 5 6
4.7.4 start p4 * 4 / – / – °C °C . °C h h : mm
ATTENTION!
100
If cycling is required, „c“ must be on LCD RH 5 6
when starting the program.
= S3
Press START/STOP, a sound will appear and the
“windmill“ in the first digit of LCD starts running °C°C°C = actual temperature
clockwise.
18 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
hh:mm = time running since the end of T1 till it Each of the segments 1 to 40 can be set
reaches T2. as the last and the infinite, that is after the
end temperature Ti has been reached, and
simultaneously after the time ti is over either the
4.7.4.4 S4 is active, T2 is reached, program is finished or followed by another cycle.
dwell time at T2 is running Each of the segments 1 to 40 can be set as
the last and infinite, that is the program at this
Program Segment °c segment must be finished by START/STOP
* 4 / – / – °C °C . °C h h : mm button.
t0 t1 t2 t3 t4 t5 t6 t7 tn=max 40
tn=max 15
T7
T4 S5 Sn
T3 S4 S6 S7
T5 T6
T1 T2
S0 S1
S2 S3
F G H I
t A: P5 name of the program
The program consists of 41 segments. Segment B: XX number of the set segment
0 is for the delayed start, it is defined only by C: X type of the set segment:
the time t0. If t0 is 0, there is no delayed start. x time of the zero segment is set to zero,
segment will be skipped over
Segments 1 to 40 are of the same qualities, each > not the last, the end length - see E
of them is defined only by the end temperature, >l the last, the end length - see E
time or speed (ramp), fan rotation speed and ∞ the last, infinite length
depending on the type of the unit also by relative D: XX . X required temperature of the set segment
humidity and chamber light. (! the end temp.)
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 19
Instructions for use
E: TXX : XX The required length of the set segment, 2) The programme runs, for example, in
or
segment 5. You want to change the setting
R ± X . XX The ramp = the required speed of the set
segment; or
in the running segment 5. Press X/W and
tXX .XX The calculated length of the set segment;
change the setting of segment 5 according
ramp is default; or to your needs. Wait for several seconds. The
cXX the number of cycles of the whole programme returns from the setting mode to
program the beginning of the working segment 5 and
F: XXX the fan speed of the set segment starts working.
G: XX disengaged 3) The programme is running. Power supply is
H: XXX the light of the set segment interrupted – either by the main device switch,
I: XX . XX the month and day at the moment or in the supply mains. After the power supply
(impossible to be set in P5) is restored, the device continues automatically
from the same point as before interruption.
4) When switching off by START/STOP or
display - running of the program ON/OFF and re-switching on, the programme
will start running from the very beginning.
A B C D E
Fan run after the programme end:
The fan run can be set in Service 06. Immediate
fan switch-off after the programme end has been
preset in all models equipped with a fan.
A:∗5 the rotating symbol and the program Requirement: Operation for 5 working days under
number temperature of 20 °C, RH = 50 %, light 100 %,
B: X the number of the relevant segment 2 weekend days under temperature of 4°C,
C: X the type of the set segment: RH = 60 %, light off. This mode of operation is
> not the last,
>| the last, the end length - see E
to be repeated cyclically with maximum possible
∞ the last, infinite length final repetitions.
D: XX . X the temperature in the chamber at the
moment Setting in program P5
E: XX : XX the segment time at the moment -
P 5 0 0 x - - . - 0 0 : 0 0
hours, minutes all segments - counting
down from the set value towards zero, 1 0 0 O F F
in case of an infinite length of the last
segment counting up from zero in case
the time is higher than 99 hr. 59 min >>: P 5 0 1 > 2 0 . 0 T 0 0 : 0 0
>> is indicated 1 0 0 O N
F: XXX the fan speed at the moment
G: XX disengaged
P 5 0 3 > 2 0 . 0 T 9 9 : 0 0
H: XXX the light at the moment
1 0 0 O N
P5 programme continuation after the preset >
P 5 0 4 0 4 . 0 T 0 0 : 0 0
parameters change or after the programme
interruption: 1 0 0 O F F
1) The programme runs, for example, in segment
5. You want to change the setting in the P 5 0 5 >| 0 4 . 0 T 4 8 : 0 0
segment that has already passed c 9 9
(e.g. segment 4) or in a segment that will run 1 0 0 O F F
(e.g. segment 6). Press X/W, change the
setting according to your needs. Meanwhile, Caution!
the programme continues running in segment In time and temperature setting in programme
5. Wait for several seconds. The programme 5, it is recommended to observe the following
returns from the setting mode to the working principles:
segment 5 and continues working. If you need to change the temperature in an
already set chain of segments, pay attention to
20 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
whether the nature (rising or decreasing) and/or 4. The whole P6 programme (all segments at
the subsequent or preceding ramps (marked “R“ once) can be shifted by means of Service 64
or “t“) are in accord with the expected change. by any time interval both to the future and past
If not, cancel the ramp and set it again only after and there is no need to rewrite all segments
the temperature setting. before a new start of the programme.
The segment temperature shall be set at first
and only then the ramp. If the ramp (“R“ or “t“) is
decisive for the segment, then we should switch 4.9.2 programme setting
to time (“T“ or “r“)at first if we want to change the
temperature. Segment 0 is compulsory, i.e. it cannot be
When setting a chain of segments, we should switched off. This condition is necessary for
proceed from the lower segment to the higher one. defining both the programme beginning and
overall duration, which is necessary for cycling.
4.9 program p6 When setting the delayed start, year is set in the
bottom line. The programme start thus can be set
4.9.1 description of P6 and its even to the past. Year is always displayed at the
control same place (7th position on the bottom line); the
sequence month-day (DD-MM or MM-DD) is set
T in Service 3
P 6 0 0 > 2 2 . 5 0 9 : 0 0
t0 t1 t2 t3 t4 t5 t6 t7 tn=max
tn=max 1540
1 0 0 2 0 0 9 1 8 - 0 2
T4
T7
S5 Sn or
T3 S4 S6 S7
T5 T6 S3 P 6 0 0 > 2 2 . 5 0 9 : 0 0
S2
T1 T2
S0 S1
1 0 0 2 0 0 9 0 2 - 1 8
t
In other segments, RH and/or lighting is set in
valid for products since 22.04.2009 with 41BB the bottom line. Years will be calculated at the
and newer memory). programme start, e.g.
P 6 0 1 > 2 2 . 5 0 9 : 0 0
P6 programme works with segments the duration
of which is determined by real time and calendar 1 0 0 O F F O N 1 8 - 0 2
date.
1. P6 programme consists of 40 segments and or
follows strictly the real time, i.e. follows the set P 6 0 1 > 2 2 . 5 0 9 : 0 0
programme independently of the time of start, 1 0 0 O F F O N 0 2 - 1 8
power supply failures, restarts after an error,
etc.
2. The segment 0 (so-called “delayed start“) is Each of the segments 1 to 40 can be set as
compulsory and time and year is set in it; it the last one, i.e. after the end temperature Ti is
can even be set to the past and means the reached and, at the same time, time ti lapses,
“programme beginning“. The segments 1 either the programme ends or the subsequent
to 39 have equal properties; each of them cycle follows.
is determined by the final temperature, time Each of the segments 1 to 40 can be set as the
or speed (ramp), ventilator rotation speed, last and indefinite one, i.e. the programme must
relative humidity (RH) and, according to the be terminated at this segment by the START/
device model, by the exposure lighting of the STOP button.
chamber as well. If there is an active selection of cycles on the
3. If cycling is switched on, the segment end display in the moment of the programme start,
times are shifted after the end of each cycle (see the Figure Display – programme selection
(forward one duration of the programme). – E: cXX), then – during the programme run –
The control of parameters during the device the device will shift to the beginning again after
run (after X/W pressing), therefore, shows termination of the last segment and the whole
updated parameters. programme is repeated. The overall number of
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 21
Instructions for use
the whole cycle courses equals to the numerical displej - state of the program selection
value after the letter “c“.
The cycles number value is related to the whole A B C D E
programmed sequence. It can be displayed in the
presence of any segment but it does not relate
to the displayed sequence only, but to the whole
sequence!
The minimum number of cycles is 1. If only one
cycle is selected, the programme will run as if it
was started without cycles.
The programme can be started by the “Start” F G H I
button. It will always be started from segment A: P6 name of the program
1 (if the zero segment time is zero) regardless B: XX number of the set segment
of the segment displayed. However, it is of key C: X type of the set segment:
importance for successful inclusion of cycling x segment will be skipped over
in the activated process that the display must > not the last, the end length - see E, I
>l the last, the end length - see E, I
show “c XX” at the moment of the programmed ∞ the last, infinite length
sequence start, where XX is the number of D: XX . X the required temperature of the set
intended courses of the whole sequence. For segment (! the end temp.)
example, the number 2 means that the sequence E: TXX : XX Hour and minute of the end of the set
will be carried out and then repeated once more. segment; or
If there is not “c XX” displayed at the moment R ± X .XX The ramp = the required speed of the set
of switching on by means of the START/STOP segment; or
button, but time or ramp of the actually displayed r± X .XX The calculated hour and minute of the
segment are displayed, the whole sequence will end of the set segment; ramp is default;
run only once. or
The time is not determined as a segment cXX the number of cycles of the whole
duration but it is determined by the hour, minute, F: XXX the fan speed of the set segment
day, and month of its end, i.e. by the real time. G: XX disengaged
If the set time and date of the segment is lesser H: XXX the light of the set segment
than the time and date of the previous segment, I: XX . XX the month and day at the moment
(impossible to be set in P5)
the segment will end no sooner than in the
subsequent year. The actual time and date
(including the year) must be set to the system display - running of the program
(regulator) in the service mode. Rising to the set
relative humidity starts after the set temperature A B C D E
is reached.
Range of parameters:
time t1 to t40 each of segments up to a
year
ramp r1 to r40 0 to 3.2 °C/min for heating
–1.0 to 0 °C/min for cooling F G H I
temperature T1 to T40 0.0 to 99.9 °C
fan speed V1 to V40 10 to 100 % with 10 % steps A: ∗ 6 the rotating symbol and the program
number
relative RH1 to RH40 10 % to 90 % with 10 % steps
B: X the number of the relevant segment
humidity
C: X the type of the set segment:
light L1 to40 0 % (OFF) to 100 % (ON)
> not the last,
with 1 % steps
>| the last, the end length - see E
number of C 2 to 99 (eventually ∞ - which ∞ the last, infinite length
cycles is the next position after 99
D: XX . X the temperature in the chamber at the
moment (°C)
E: XX : XX the segment time at the moment -
hours, minutes all segments - counting
down from the maximum value towards
zero, in case of an infinite length of the
last segment counting up from zero in
case the time is higher than 99 hr. 59
min >>:>> is indicated
22 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
F: XXX the fan speed at the moment (%) segments that should already be terminated
G: XX disengaged (%) will be skipped, duration of the actual segment
H: XXX the light at the moment (%) will be adapted (shortened).
2. Programme without cycles; the actual time
is higher than the last segment time. The
CAUTION! programme will be terminated (all segments
When setting the time and temperature will be skipped because the programme
parameters in the programs 6 it is suitable to “ended“ during the power supply failure); LCD
follow the following principles: displays “End of P6“.
If there is a need to change the temperature in an 3. Programme with cycles; the actual time is
already set chain of segments, pay attention to lower than the last segment time. See point
whether the nature (rising or decreasing) and/or 1 above. The programme will continue with a
the subsequent or preceding ramps (marked “R“ shortened “actual“ segment; subsequent cycle
or “t“) are in accord with the expected change. If runs normally.
not, cancel the ramp and set it again only after 4. Programme with cycles; the actual time
the temperature setting. is higher than the last segment time. The
The segment temperature shall be set at first programme definition is shifted by the
and only then the ramp. If the ramp (“R“ or “t“) is appropriate number of P6 programme
decisive for the segment, then we should switch durations so that the programme could
to time (“T“ or “r“) at first if we want to change the continue.
temperature. The operation then is like in point 3 above.
When setting a chain of segments, we should
proceed from the lower segment to the higher one.
4.9.5 automatic shifting of p6
programme beginning in
4.9.3 programme start cycling
After the programme starting (by the START/ After each cycle is terminated, the programme
STOP button or order from WC3), years of other definition is recalculated (all segments are shifted
segment will be calculated and the test of the last forward by one programme duration). When
segment time will be then performed. checking the parameters during the programme
run by X/W pressing, the actual P6 setting is
Three situations may occur: displayed. After the last cycle or programme
1. The actual time is lower than the zero without cycling ends, the parameters will not
segment time – waiting phase is started; the be shifted, i.e. the last parameters according
whole programme will run. to which the programme was running are
2. The actual time is higher than the zero always displayed after the programme end or
segment time but lower than the last segment interruption (with or without cycles).
time – all segments that should be terminated
at the given time will be skipped and that
segment which corresponds to the actual one 4.9.6 manual shifting of p6
will be started. Duration of such segment will programme beginning - user
be adapted (shortened) so as to terminate at service 64
the set time ; LCD displays the time remaining
to the segment end. Durations of all segments remain unchanged.
3. The actual time is higher than the zero
S e r v i c e 6 4
segment time – warning audible signal
(Operator Error) sounds and the programme S h i f t i n g p r o g P 6
does not start.
After the programme is started (START/STOP –
see), time and whole date of the zero segment
4.9.4 power supply failure are displayed.
S e r v i c e 6 4
After a power supply failure, the following
situations may occur: 0 9 : 0 0 2 0 0 9 - 0 2 - 1 8
1. Programme without cycles; the actual time
is lower than the last segment time. Those
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 23
Instructions for use
or P 6 0 3 >| 2 2 . 0 T 0 9 : 0 0
c 0 0
S e r v i c e 6 4
1 0 0 O F F 0 2 . 1 9
0 9 : 0 0 2 0 0 9 - 1 8 - 0 2
Example 2: Day and night phase with different
or temperatures.
Requirement: from 9:00 AM till 5.00 PM
S e r v i c e 6 4
Temperature 25 °C, light ON 100 %
0 9 : 0 0 0 2 - 1 8 - 2 0 0 9 from 9:00 AM till 5:00 PMof the next day;
temperature 22 °C, humidity 70 % RH, light OFF
or indefinite number of cycles.
S e r v i c e 6 4
Programme start on 18.02.2009 at 9:00 AM. At
0 9 : 0 0 1 8 - 0 2 - 2 0 0 9 the programme start, the display must show the
setting „c oo“(cycling activation).
Any parameter can be changed (a year can
P 6 0 0 > - - . - 0 9 : 0 0
be changed by one year forward or backward
as compared to the actual year) and by X/W 1 0 0 2 0 0 9 0 2 : 1 8
pressing, the programme will be shifted by the
P 6 0 1 > 2 5 . 0 0 9 : 0 0
difference between the new and original segment
time (“Done” is displayed). 1 0 0 O N 0 2 : 1 8
S e r v i c e 6 4 * P 6 0 3 > 2 2 . 0 1 7 : 0 0
D o n e 1 0 0 O F F 0 2 : 1 8
24 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
Setting in P6 programme future and past. The setting is described in the
Start on Monday 30.10 at 08:00 AM P6 programm.
P 6 0 0 > - - . - 0 8 : 0 0
1 0 0 2 0 0 9 1 0 : 3 0 4.10 measured variables
displaying during the
Monday morning at 08:15 AM programme run
P 6 0 1 > 2 0 . 0 T 0 8 : 1 5
In devices with a greater number of measured
1 0 0 O N 1 0 . 3 0 variables, all variables cannot be displayed at
once. Their values may be displayed during the
Saturday morning at 08:00 AM programme run consequently by pressing the ◄
or ► button. The given variable is displayed for
P 6 0 2 > 2 0 . 0 T 0 8 : 0 0
several seconds. Its displaying can be extended
1 0 0 O N 1 1 . 0 4 by repeated ▲ pressing.
In devices with a greater number of measured
Saturday morning at 08:15 AM variables, all variables cannot be displayed at
P 6 0 3 > 0 4 . 0 T 0 8 : 1 5 once. Their values may be displayed during the
1 0 0 O F F 1 1 . 0 4
programme run consequently by pressing the ◄
or ► button. The given variable is displayed for
several seconds. Its displaying can be extended
Monday morning at 08:00 AM by repeated ▲ pressing.
P 6 0 4 >| 0 4 . 0 T 0 8 : 0 0
c 9 9 Example: FRIOCELL with one flexible
1 0 0 O F F 1 1 . 0 6 temperature sensor:
1. Cycle: xx/yy where xx means the number of
Note: The 15-minute delay between 8:00 AM the just running cycle; yy is the overall
and 8:15 AM was chosen on condition that it is number of the preset cycles
sufficiently long enough for the transfer from one 2. Temp_1 Regulatory sensor temperature [oC]
temperature to another If it is not the case, an 3. Humid RH [%]
experimentally ascertained delay must be set. 4. Temp_2 Flexible sensor temperature [oC]
5. T_Sfty_Th Protective sensor temperature
[oC] (if the control of the difference between
4.9.8 summer/winter time the measured temperature and the regulatory
sensor has been preset).
The regulator does not take account of the
time changes. When the time is changed to the
summer/winter time, the user must change the 4.11 error messages
time manually in Service 3, or the time can be set
in a remote way. The programme of the recording Error messages are placed on LCD and have the
WarmComm 3-F SW synchronizes the regulator following meaning:
time with the time of a PC, if connected, with an
installed recording SW (the set regulator time Error 00 PT 1 (regulatory sensor) – not
equals the PC time). Any time change during the connected, broken, or short-circuited.
Call the service.
programme run, however, does not influence its
course because the durations of the segments Error 01 PT 2 (flexible sensor a) – not
connected, broken, or short-circuited.
are calculated only at the programme start (and Call the service.
in the examples above, the day and night phase
Error 02 temperature is below the preset lower
will be changed at 6:00 PM and 10:00 AM). limit. See service 41 in the instructions
To match the times fully, the programme must for use.
be restarted either by stopping and immediate Error 03 temperature is above the preset lower
starting, or by power supply failure simulation (by limit. See service 41 in the instructions
the power supply switch). for use.
Error 04 protective thermostat is activated;
Service 64 – P6 programme start shifting heating is disconnected. See service
The whole P6 programme (all segments at once) 07 in the instructions for use.
can be shifted by any time interval both to the
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 25
Instructions for use
Error 05 protective thermostat setting. See to interrupt the running process by a hard
service 07 in the instructions for use. error.
The error can be caused by starting a
programme whose temperature is out If a “soft” error occurs, you can proceed in
of the limits specified by the protective following ways:
thermostat setting, or by a HW defect.
1) By depressing X/W only the acoustic signal
Error 10 controlling programme. is suppressed so that it does not disturb. The
Call the service.
humidity generating elements are blocked,
Error 11 communication
the unit does not regulate the humidity. Other
Call the service.
functions are kept and controlled.
Error 12 PT 3 (flexible sensor b) – not
2) By depressing START/STOP
connected, broken, or short-circuited
Call the service - both the acoustic signal is suppressed
- and the humidity generating elements are
Error 19 – PT 4 (flexible sensor c) – not
connected, broken, or short-circuited unblocked. During the “soft” error the button
Call the service. START/STOP has not the function of stopping
Error 20*) door cannot be blocked the operation. After depressing START/
See service 58 and chapter 10.15 in STOP, if the soft error cause is eliminated,
the Instructions for use. the unit can continue the operation inclusive
Error 21*) door cannot be unblocked the humidity. If another error occurs in
See service 58 and chapter 10.15 in future, it will be signalized only optically, not
the Instructions for use. acoustically.
Error 27 PT 5 (flexible sensor d) – not 3) By disconnecting the unit from mains and
connected, broken, or short-circuited connecting it again (e.g. with the main switch)
Call the service. during the operation the elements of the
Error 28 great temperature difference between humidity automatics are initialized and the
the regulatory sensor and protective unit starts running automatically in the same
thermostat way as in case of the fall-out loss of electrical
Call the service. energy and the operation continues from
the point of interruption. In this case another
*).......... This is a soft error eventual “soft” error will be signalized both
**)........ Error type may be set by the user. acoustically and optically.
* It is a soft error. The difference between the If “soft” errors occur repeatedly, please inform
soft error and the hard one is, that after the your service organization.
soft error occurs and is reported the whole
running process is not stopped, only the
system is eliminated, whose element has 4.12 printing the protocol
been detected as defective. Other systems
continue their controlled function – heating, DPT 6333 printer is recommended for records
cooling, ventilator, lighting. printing. Other types can be used as well, such
as CITIZEN, model iDP 3110-24 RF-A. Printing
If such an error occurs – the regulator into a PC application (Printer Archiv) is allowed
(without interrupting the operation) informs – as well.
acoustically and optically by means of display By setting the printer interval you give the time
– the operator. The acoustic signal is the intervals of printing actual values of temperature
same as in case of the hard error, the optical in the chamber. Length of this interval is
signal differs (from reporting the hard error) displayed in the head of the protocol.
by flickering of the appropriate message on Printer interval can be set - see paragraph Basic
the display. In case of hard error it is shown Setting, point Service 04.
permanently. 1. Connect the printer to the device by means
of a cable with a serial connector (RS-232C);
The soft error is used, because its cause use the printer socket (on the rear side of the
can be removed by a simple correction and extension piece/device). Then connect the
it is not necessary to interrupt the running other end of the cable to the connector in the
process. E.g. the hose of the drain from the rear part of the printer. Connect the printer
steam generator is rolled up improperly, which to the electric power mains by means of an
prevents the water to go out when turbidity is adapter. Some types require a switch on the
removed, on the other hand it is not necessary right side of the printer to be switched on.
26 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
The POWER and SEL indicators light up. The 4.13 checking the function of
printer is in the ON-LINE mode. Paper is fed the safety thermostats
by pressing the LF button (in the OFF-LINE
mode only, i.e. after the SEL button pressing – According to the procedure described in Service
the SEL indicator turns off. To renew the ON- 07 – Setting the safety thermostat, set the limit
LINE mode, the SEL button must be pressed temperature 85 °C and confirm with the chip card
– the SEL indicator lights up). The printer can SO. In P1 set the temperature 80 °C and start
print in the ON-LINE mode only! the program. After reaching the set temperature
2. Follow Service 04 – Printing Interval Setting wait ca. 10 minutes, stop the program and
and set the required interval either up or enter the Service 07; in Service 07 set the limit
down. temperature to 70 °C, confirm the setting with
There is a head printed by the printer, the the chip card SO. The unit operates according to
head comprises: type of unit, set temperature, the description in Service 07 -Setting the safety
eventually set ramp, number of cycles, thermostat: on the display “Safety thermost” etc.
ventilator rate, RH, light and the selected
printing interval. Below the head following
values are printed: 4.14 additional functions
time since the start of the program, real
temperature in the chamber (below eventually 4.14.1 door opening control
temperature on the flexible sensor), RH.
The door opening is signalled as follows:
A new head is printed if: If the alarm is switched off and the door is
- the cooling incubator is started opened during the programme run, the display
- some parameter has been changed will show a flashing message “Door open”. If the
3. You can stop the printing by setting the alarm is switched on, it sounds in addition to the
printing interval to Printer off. displayed message.
4. In case of a mains fall-out and restoring the The alarm can be cancelled temporarily by
supply again there is → Mains fall-out printed. pressing any ◄ ▲ ► button. By the door closing,
The printing intervals start to be counted since the alarm is renewed; i.e., the operator will be
the moment of restoring the current supply. informed acoustically about any subsequent door
5. In case of fall-out of the printer or switching opening again.
the printer off there is no report printed after The fact of the door opening during the
the switching on or restoring the supply. programme run is registered in the connected
printer record or in the SW WarmComm record in
DPT-6333-V24 does not require any microswitch the connected PC.
setting by the user.
Setting the DIP microswitches of the printer See also Service 62 – Alarm Setting.
CITIZEN iDP 3110: the second DIP microswitch
from the left is in the upper position, the other
three ones are in the lower position OFF. 4.14.2 keyboard blocking
Setting the DIP microswitches of the printer
CITIZEN CBM 910: OFF / ON / OFF / ON / OFF / The keyboard can be blocked (to prevent
OFF / ON / OFF (from the left to the right). an incidental activation of some function) by
Setting the DIP microswitches of the printer concurrent pressing the X/W and ◄► buttons.
CITIZEN CBM910 II: OFF/ON/OFF/OFF/OFF/ Unblocking is performed in the same way.
OFF/ON/OFF (from the left to the right).
After any change of setting, the printer must be
switched off and on either by the switch, or by
disconnecting from the power supply. 5 maintenance
A different type of the printer must be adjusted
with the help of the printer´s manual and the data 5.1 sealing exchange and door
on the interface RS232C shown on the case. setting
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 27
Instructions for use
28 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
Inspection activities:
Check of the electrical equipment, especially
the power supply, the terminal clamps and
VOLUME 22: the protective clamp. Undamaged state of the
Unscrew the screws in the rear chamber part. conductors insulation (e.g. due to rubbing,
By hitching up towards the door, remove the burnout etc.) and firm connection of the
bottom and ceiling from the instrument. Then, it conductors in the clamps. The resistance of the
is possible to put out the back. In case of leakage protective connection shall be < 0.1 Ohm; the
of some contaminated stuff into the chamber the resistance of the power supply shall not be taken
user is responsible for a proper decontamination into account.
of all contaminated surfaces with a suitable and
approved disinfecting agent. If the cabinet is placed in another environment,
Before using another cleaning or the inspection must be performed in accordance
decontamination method, with the exception with the local standards.
of the by us recommended methods, the user
should consult the producer, whether the A service overhaul depends on an agreement
considered method could not cause a damage to between the customer and the service
the device. organization.
5.4 melting
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 29
Instructions for use
6 technical data
6.1 chambers
FRIOCELL (FC) M
Technical data volume cca l 22 55 111 222 404 707
Inner dimensions
width cca mm 224 400 540 540 540 940
Chamber, stainless steel
depth cca mm 307 370 370 520 520 520
height cca mm 296 350 530 760 1410 1410
Trays, stainless steel *) racks max. No 4 4 7 10 19 19
standard equipment pcs. 2 2 2 2 2 2
Min. distance included 60 70 70 70 70 70
between trays cm
Maximal load screens kg/screen 10 20 20 30 30 50
steel sheet racks kg/rack 10 20 20 30 30 20
in total kg/case 25 50 50 70 100 130
Number of doors outer and inner door No 1/1 1/1 1/1 1/1 1/1 2/2
External dimensions width cca mm 406 620 760 760 1010 1460
(including door and depth cca mm 592 640 640 790 790 790
handle) height cca mm 605N 820N 1100K 1330K 1920K 1920K
Mass net cca kg 33 80 101 135 233 273
brut cca kg 38 99 131 156 261 305
Electricity max. power **) W 130 1130 1130 1130 2250 2500
mains 50/60 Hz V 100- 230 230 230 230 230
Protective system 240 IP 20 IP 20 IP 20 IP 20 IP 20
IP 20
Technical data
Working temperature***) from 0.0 °C to °C °C 5-70 99 99,9 99,9 99,9 99,9
Temperature accuracy at 10 °C accuracy in cca (±) °C <0,3 <0,5 <0,5 <0,5 <1 <1
space <0,3 <0,5 <0,5 <0,5 <1 <1
at 37 °C cca (±) °C <0,1 <0,2 <0,2 <0,3 <0,3 <0,4
accuracy in time
Heating/up time to to 37 °C from the
ambient temperature min <10 23 24 25 26 27
Cooling/down time from 22 °C to 10 °C min <31 <14 <21 <21 <21 <21
Recovery time after at 37 °C and at 50 min 4 4 4 4 4 4
1 min. door open °C
Heat emission at 37 °C cca W 50 62 70 97 123 148
Noise level of the
dB 53 46 46 50 56 58
complete FC
Note:
All technical data are related to 22 °C of ambient temperature and ± 10 % voltage swing.The volume 22 cools ca.
20 °C under the ambient temperature (it uses the Peltier effect.) Peltiér´s module serves for cooling and heating.
* Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
**) Compressor + condenser + electromagnetic valves + ventilator (s).
***) When using probes for measuring of photo values, the upper operation temperature is limited to 75°C. The
maximal adjustable temperature for protective thermostat is 80 °C.
30 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
All data apply to an empty chamber (without samples on the screens) and the ambient temperature of
20 - 22 °C; 100 % of the fan revolutions; supply voltage 230 V ± 10 %.
The actual quantities of the photometric quantities during the experiment must be measured by an
independent measuring device.
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 31
Instructions for use
service
The guarantee period is marked on the certificate 9 how to eliminate
of warranty.
wrappings and
The warranty refers to manufacturing defects or
material defects on condition that: a unit out off
• the product has been installed and used in
accordance with the Instructions for use, service
• the cause of the defect was not a insufficient
maintenance, unqualified intervention in the Pallet – in a refuse incinerating plant
device or damage through external impacts. Cardboard –waste for recycling
Unit out off service:
The warranty does not relate to natural wear of
material and to consumer material, such as e.g. For the European Union Member Countries:
door sealing, materials for recording devices, If the user stops to use the labelled product, and
accumulators etc. If a defect occurs, draw claims the product will become useless for him
to a warranty repair directly at the nearest MMM
service center. Please mention the device name
and type, its serial number and how the defect then the product must not be disposed in
appears (error message, printer record).Provided municipal waste and is subjected to the regime in
the warranty conditions are fulfilled there follows, compliance with the national rules on liquidation of
according to the consideration of the service electrical and electronic equipment, corresponding
center, the cost-free repair or replacement of the to WEEE (Waste Electric and Electronic
defective part. Equipment) Directive in version force (in the Czech
Republic it is Waste Law No. 185/2001 Code of
If the prescribed maintenance is complied with, Law in version in force). Please, ask for detailed
the device life exceeds 10 years. At the same information about correct liquidation of electrical
time, this is the period for which the manufacturer and electronic equipment by your seller or supplier.
bears objective responsibility within the meaning
of EU Directive no. 85/374/EEC for any potential For countries outside of European Union:
damage caused by the device operation. If you The symbol given above is valid only in European
decide to continue using the device after the Union countries. Please, ask for detailed
said period expiration, ask the manufacturer or information about correct liquidation of electrical
an authorized service for an expertise whether and electronic equipment by authorities in your
or not the device can still be used regarding its country or by your seller of electrical equipment.
wear and tear, technical condition and service
availability.
10 optional
8 transport and equipment
storage 10.1 inner lighting
Device will be prepared for transport by Inner space is illuminated with thermo-resistant
a competent person (who also disconnects bulbs - their switch is placed on the door. It is not
the device from the mains). Device must be recommended for the volume 22.
transported and stored in original wrapping. If you
send the device back (for reparation or change in
case of reclamation), use the original wrapping. 10.2 bushings of diameter 25, 50,
Otherwise you overtake the responsibility for 100 mm
event. damaging during the transport and the
producer will reclaim compensation for event. The bushings are normally placed cca in the
middle of the side (right or left) wall of the
32 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Instructions for use
chamber. The bushes are metallic, closed with The program enables to follow the regulation
a special plastic plug from the outer side, which on-line, to store the regulation course to a file
enables passing through of wires etc. from the on a disc and to view the already stored files, to
outer space to the chamber. send e-mails in case of not receiving data from
(For the volume 22 only the bushings 25 and the cabinet and to send reports of excessive
50 mm.) values being reached within the specified time
interval.
Recommendation: the used unit should be The instructions for the programme use are
provided with a bushing with corresponding supplied with the installation programme.
dimensions, if user wishes to measure Hardware requirements: Common PC with an
temperature or RH (with CLC) inside the chamber operational software Windows 2000, XP and
by means of sensors, that are connected with higher. One free port RS 232 (COM) is required
an independent measuring device by means for each connected device. The maximum length
of wires; the user pulls the wires through the of the connecting cable is 15 m.
bushing.
10.6.2 recording printing sw -
printer archive - for pc
10.3 lockable door
under windows
The lock is placed on the upper part of door’s
surface near the closing mechanism. The Printer Archive programme serves for text
recording from the device print outlet. It serves
as a direct substitution of a physical printer.
10.4 left door The data are recorded to a file on a PC disc
and the programme offers more possibilities of
This is the symmetric mirror version of the right the recorded data handling including archiving
door. The cabinet of 404 liter is not delivered in or printing on a table PC printer. Hardware
this version. requirements: Common PC with an operational
software Windows 95 and higher. One free port
RS 232 (COM) is required for each connected
10.5 independent sensor PT100 device. The maximum length of the connecting
cable is 15 m. For more detailed information
Independent flexible sensor is used for about HW, ask the seller.
measuring of material temperature directly in
the chamber. The date of temperature on this
sensor (PT2) is displayed on LCD in the position 10.7 potential - free contact
of time date, while these two data are altering. for alarm reports
The corresponding printed protocol has two It’s outlet is placed in the connector in the rear
lines, first line - beginning with digit 1 - shows foot, a voltage up to 24 V/ 1 A can be connected
the temperature on the sensor of regulator to it.
(PT1), second line beginning with digit 2 shows
the temperature on the sensor PT2. If there is
a contact failure between PT2 and the regulator, 10.8 the inner switched socket
the display reports error 01.
It is not recommended for the volume 22.
The socket is placed inside the chamber on the
10.6 supportive sw for pc side wall. Details of the switching control see par.
Basic setting of the units FRIOCELL – Service
10.6.1 recording sw - warmcomm - 06, connection and other details are given in the
for pc under windows following diagram and text.
FRIOCELL_np_en_1307_mmm_V2.05_B2V_M 33
Instructions for use
Connection wiring beetween plug and built-in 10.11 temperature range from
equipment - 9.9 °C
TO THE This option is not available in PM models; it
1 EQUIPMENT
K PŘÍSTROJI is neither available in other model sizes in
PLUG
VIDLICE L1 (230V/1A)
combination with the exposure lighting in the
N
3 2 door in both ON and OFF state, and in the
shelves in the ON state.
Warning:
The plug must be connected only by a person 10.12 PROTECTION OF THE
with a sufficient qualification. PROGRAMME SETTING BY A CHIP
When the plug is plugged in and out of the CARD
socket, or when manipulated with the inner
equipment, FRIOCELL must be switched off (the The parameters of the selected programme may
mains switch in the position -„0“). only be changed after an SO card inserting. The
The socket-plug contact must be tightened in programmes may be selected arbitrarily but their
order to prevent disfunctioning of the protective parameters may not be changed (temperature,
system. time, ...). If the card is not inserted, the
After disconnecting the socket-plug contact fit programmes are marked with lower-case letters
carefully the protection covers on the socket and on the display (p1 to p6); after the chip card is
the plug so that impurities and humidity would not inserted, they are displayed in capital letters (P1
get inside. to P6).
34 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
EMERGENCY DO
Instructions for use (use it only in case of fai
ATTENTION!
Blocked door may not be opened forcibly. The
blocking system can be damaged mechanically.
If there is a need of an emergency door opening
(e.g. in case of power failure), follow the
Emergency Door Opening.
Y DOOR OPENING
Emergency door opening
2. To push in the arrow direction and thus to
(use it only in case of failure or power failure) push away the blocking rod (10 -15 mm is
e of failure or power failure) enough). In case of standard model – of the
1. To insert a thin object (screwdriver, rod etc.) right door in the arrow direction, in case of
through 2. theTonon-standard
push in themodel
arrow– direction and –thus
it to pu
od etc.) through the the
air air
hole holeofofthe
the door
door (in the bottom
enough). In case(always
of the left door
of standard model – of the r
door surface) under the angle of ca 45 ° in the would be reversely in the direction
e angle of cadistance
45 ° in of
theca distance
90 mm from of the
ca door
90 mm from
edge. 3. fromTo open
the handle the
non-standarddoor by means
model
to the door – of thethe
hinges). of leftdoor
doorhand
– it
If it would
from betonot
handle thepossible to move away the
door hinges).
would be necessary to unscrew the whole bl
screws) and to push it away as a whole. Afte
necessary (backward blocking mechanism a
36 FRIOCELL_np_en_1307_mmm_V2.05_B2V_M
Date
2012-10-01
European Directive 2006/95/EC, 2004/108/EC
Manufacturer
MMM Medcenter Einrichtungen GmbH, Semmelweisstrasse 6, D-82152 Planegg / München
electrically heated incubator intended for heating and cooling of laboratory materials in the temperature range 0 °C
cooling incubator: up to 99,9 °C (type FCP-B2V: +5 °C up to 70 °C)
trade name: FRIOCELL
type: FC-B2V FC-B2V-M FCP-B2V
model: FC 55, 111, 222, 404, 707 FC 55, 111, 222, 404, 707 FC 22
meets all the provisions of the Directives no. 2006/95/EC, 2004/108/EC, which apply to him.
This declaration is valid for all of the above products, that are marketed after the date of issue and is valid until changes in
the facts contained therein.
Standards applied:
Safety requirements for electrical equipment for measurement, control and laboratory use - Part
EN 61010-1:2001
1: General requirements
Safety requirements for electrical equipment for measurement, control and laboratory use - Part
EN 61010-2-010:2003
2-010: Particular requirements for laboratory equipment for the heating of material
Electrical equipment for measurement, control and laboratory use - EMC requirements Part 1:
EN 61326-1:2006
General requirements