Professional Documents
Culture Documents
Express
PAPER
PAPER
Keywords: electron beam therapy, small circular fields, therapeutic depth, prescription depth, Monte Carlo method, least square fitting,
EGSnrc/BEAMnrc
Abstract
An empirical model for small circular electron fields was developed. This can be of great help in the
treatment planning process for small circular electron fields. A complete dosimetric analysis of the
circular fields defined by electron cutouts diameters (2 cm–9 cm) was done for nominal electron
energies ranging between 6 MeV and 20 MeV using a 3D water phantom and a pin-point ion chamber.
Properties studied included depth dose, in-air inverse-square fall-off, and beam profiles. The Varian
Clianc 2100 C accelerator was modelled, benchmarked and Monte Carlo simulations were performed
using the EGSnrc/BEAMnrc code for the small circular cutouts. A simple exponential model was
found to accurately predict the very important therapeutic depth (90% of Dmax) for the small circular
field size within an accuracy of better than 2 mm in most cases. The model has only two parameters
(d90 and ‘b’). Also, the penumbra widths (90% of the off-axis profiles) of these small circular electron
fields were studied and least square fitted to a simple quadratic model. Full dosimetric profiles of these
small circular electron fields were further studied using the benchmarked Monte Carlo simulations.
This study presents a simple model to predict the very important therapeutic depth (90% of Dmax)
and a recipe to develop such an electron treatment model for any linear accelerator system. Such
predictions can be extremely valuable and time saving prior to treatment planning involving not only
small circular shaped electron fields but also irregularly shaped electron fields.
2
Biomed. Phys. Eng. Express 6 (2020) 065023 I Khan et al
Figure 2. 90% Depth (y) beyond Dmax versus projected circular cutout diameter (x) for SSD of 100 cm. For x<2 cm, the data points
were generated through benchmarked Monte Carlo accelerator model.
Table 1. Values of fitted parameters evaluated using least square regression analysis.
Parameters Energy=6 MeV Energy=9 MeV Energy=12 MeV Energy=16 MeV Energy=20 MeV
points (x<2 cm) are necessary to arrive at the Table 1 shows the values of the least square fitted
functional form of equation (1). parameter ‘b’, d90 for all the nominal electron energies
for the therapeutic depth (90% of Dmax). These values
3. Results can be easily obtained for any particular linear accel-
erator by performing measurements for a few circular
The following simple exponential model (equation (1)) electron cutouts and a few Monte Carlo simulations.
had been found to accurately predict the commonly used The parameter ‘b’ depends on the beam energy and
the therapeutic depth (90% of Dmax) for the circular has been least square fitted to a quadratic model (cubic
field size with projected diameter x (at SSD = 100 cm), term coefficient is close to zero) as shown in figure 3
within an accuracy of better than 2 mm in most cases. with coefficient of determination : R2=0.998. This
model can be developed for any accelerator system and
y = d 90[1 - e-b.x ] (1)
used to predict the ‘b’ parameter for a particular
Here y is the 90% depth for the circular cutout with energy to be used the main modeled equation (1) for
projected diameter: x, d90 is the 90% therapeutic or calculating the therapeutic depth (90% of Dmax).
prescription depth (90% of Dmax) for the open cutout Penumbra widths, which are defined as the 90%
(10×10 cm2 field), which is normally measured field width at the depth of maximum dose were eval-
during the commissioning process. This d90 can also uated and are presented in figure 4. These are plotted as
be estimated approximately by E/3.2 cm, where E is a function of electron cutout size (x) and nominal elec-
the most probable energy in MeV of the electron beam tron energy at 110 cm SSD. The data was least square
at the patient/phantom surface [8]. Figure 2 shows
curve fitted to set of quadratic expressions (shown in
the plot of measured values and also those obtained
figure 4) for each electron energy. The coefficient of
from Monte Carlo modelling for projected diameter:
determination (R2 value), which is a measure of how
x<2 cm. The predicted curves are plotted as well by
curve fitting the functional form using least square good the fitting [10] has been calculated and in almost all
method for electron beam energies ranging from cases are very close to the ideal number of one.
6 MeV to 20 MeV. The only real unknown in this Figure 5 shows the measured and Monte Carlo
model here is the parameter ‘b’, to be determined generated data for 12 MeV central axis, percent depth
using least curve fitting. The coefficient of determina- doses for all the small circular cutouts of diameters
tion for the curve fitting (R2 value) was close to 0.99 for ranging from 2 cm to 10 cm. Figure 6 shows a compar-
all the electron energies [9]. ison of the measured and the Monte Carlo generated
3
Biomed. Phys. Eng. Express 6 (2020) 065023 I Khan et al
Figure 3. The parameter ‘b’ of the equation (1) as a function of electron energy along with a least square fitted quadratic model for
predicting it for any console energy: E.
Figure 4. Measured and fitted quadratic equations for 90% penumbra widths at 110 cm SSD.
data for off-axis profiles of 9 MeV electron beams circular sizes of 2 cm and 3 cm is not as great as the
again for all the circular cutouts of diameter ranging agreement for field sizes higher than 3 cm. This slight
from 2 cm to 10 cm. The simulations were done using disagreement may be due the uncertainties involved in
100 million histories to give error bars of sizes less than measurement for very small field sizes. The bench
±1%. The simulation time taken for each cutout was marked Monte Carlo data for percent depth dose may
around 2 to 3 hours on an Intel i5-4570@3.2 GHz pro- be trusted more in these cases.
cessor with 4GB RAM.
The Monte Carlo method needs to be able to predict 4. Discussion and conclusions
a dose distribution in agreement with the measurement
within clinically acceptable tolerance level [11, 12]. A The formulation developed in this work accurately
commonly accepted accuracy level is ±2% or ±2 mm as describes the clinical dosimetric properties of small
stated in ICRU Report 42. [13]. Great agreements (dif- circular electron fields. It can be used to select the
ferences less than 2%) between the measurements and circular cutout diameter needed to provide the desired
the Monte Carlo data have been found in both the cases coverage (using 90% field widths) at the very impor-
studied in figures 5 and 6. The agreement for very small tant therapeutic depth (90% of Dmax), beyond which
4
Biomed. Phys. Eng. Express 6 (2020) 065023 I Khan et al
Figure 5. Measured and Monte Carlo data for 12 MeV, central axis percent depth dose profiles.
Figure 6. Measured and Monte Carlo data for 9 MeV off-axis profiles for circular cutouts with diameters ranging from 2 cm to 5 cm.
the dose falls off very rapidly. This can be done without and energies have been narrowed down, the fitted
relying on usually used approximate rules i.e. estimat- equations of the measured data for penumbra widths
ing it as E/3.2 cm, where E is the most probable energy can be used to close the search for the right cutout and
in MeV of the electron beam at the patient/phantom the right energy. Modelled equation (1) has only two
surface [8] or trying various combinations of energy parameters that of d90 for the open 10×10 field
and cutout size in a treatment planning system. This (which is usually measured during the commissioning
formulation can greatly help in providing adequate process) and the independent parameter ‘b’. The pos-
dose at the right target depth under the skin and have sible values of parameter ‘b’ (shown in table 1) were
intended target coverage. determined using the empirical data and nonlinear
Equation (1) can be effectively used for predicting regression method [9]. It may be noted that the para-
the depth (90% of Dmax), represented as: y for small meter ‘b’ does not vary much for higher energies ran-
circular cutouts with projected diameter: x and greatly ging from 12 MeV to 20 MeV. Parameter ‘b’ may be
help in the decision of choosing the right circular cut- related to the scattering power of the electron beam,
out size and energy. Once a set of relevant cutout sizes which is higher for lower energies. This parameter can
5
Biomed. Phys. Eng. Express 6 (2020) 065023 I Khan et al