Professional Documents
Culture Documents
Key words: Monte Carlo simulations, flattening filter free, treatment head modification, beam line
modification
I. INTRODUCTION dose close to the penumbra; and changes in the lateral dose
profiles, as a function of depth in a water phantom, are in-
Conventional clinical accelerators used in radiotherapy are vestigated in this work.
equipped with a flattening filter 共FF兲. The FF is designed to Several previous investigations of FF free machines have
produce approximately uniform dose distribution across the been undertaken, primarily to investigate the physical prop-
field at a certain depth in a homogeneous phantom. Such a erties of the beam, such as changes in dmax, the location of
field is typically not the best solution for complex treatment maximum dose deposition and energy spectrum changes.3–5
problems. For this reason, additional beam-modifying de- Variation of the beam quality with off-axis distance have
vices are used, such as wedges, compensators, and dynamic been researched by Zefkili et al.6 and Lee et al.7 O’Brien
multileaf collimators 共MLCs兲. MLCs are quite flexible in et al.8 removed the FF from a Therac-6 linear accelerator
terms of shaping fluence distribution across the field. In prin- 共AECL, Atomic Energy of Canada Ltd., Mississauga, On-
ciple, they render the FF unnecessary. tario, Canada兲 in an effort to reduce the treatment time of a
stereotactic radiosurgery procedure. Gillies et al.9 discussed
The FF introduces a very substantial amount of material
engineering modifications to a Therac-6 linear accelerator to
into the beamline. It is one of the most significant sources of
perform dynamic stereotactic radiotherapy, including a head
undesired radiation scatter.1,2 More than 50% of primary ra-
modification, which allowed easy removal of the FF. Sixel
diation can be scattered or absorbed in the FF. Relevant nu- and Faddegon4 calculated photon spectra for small fields on
merical data will be presented later in this study. the Therac-6 linear accelerator as well as depth-dose and
By removing the FF an increase in dose rate can be profiles with and without the FF. Further research was con-
achieved, leading to a shorter irradiation time per treatment ducted to investigate portal imaging properties,10 dose deliv-
fraction. This will also reduce treatment head scatter, out-of- ery enhancement in high-Z material-loaded tumors,11 and an
field exposure of the patient, and leakage dose through the application in helical tomotherapy.12 Fu et al.13 investigated
MLC. The implications of removing the FF with respect to intensity modulated radiation therapy 共IMRT兲 delivery time
increased fluence and dose rates; changes in the out-of-field with and without the FF. They used a BJ-6B linear accelera-
1595 Med. Phys. 33 „6…, June 2006 0094-2405/2006/33„6…/1595/8/$23.00 © 2006 Am. Assoc. Phys. Med. 1595
1596 Titt et al.: FFF Monte Carlo simulation 1596
tor 共in 6-MV photon mode兲 and the Focus 共version 3.2.1,
CMS, Computerized Medical Systems, St. Louis, Missouri,
USA兲 treatment-planning system.
In this paper, we present our first results on the way to a
flattening filter free 共FFF兲 clinical accelerator, based on the
Varian Clinac 2100 共Varian Medical Systems Inc., Palo Alto,
California, USA兲. Numerous FFF irradiations have been
simulated to investigate the properties of the device and the
first results are being presented. The study focuses on the
6-MV photon mode.
A / cm2 dsim
max / cm dmeas
max / cm ⌬Dmax
FIG. 4. Comparison of simulated central-axis depth-dose 共per initial par- dent particle兲 by a factor of 2 at least. The ratio of the flat-
ticle兲 profiles in a 40⫻ 40 cm2 field, with and without the FF, normalized to tening filter-free simulation to the results with the FF 共field
100 at the shoulder of the FFF dose falloff. Besides the overall dose in-
crease, the profile from the filter-free simulation clearly shows a sharp spike size 10⫻ 10 cm2兲 showed two different ranges; in both, the
in the build-up region, which is attributed to excessive electron fluence. ratio is almost linearly dependent on the depth z. First, a
steep decline in the build up region could be observed 共from
a ratio of 3.2 at z = 0.05 cm to 2.4 at z = 1.2 cm兲 followed by
calculated with MCNPX using a field size of 10⫻ 10 cm2, a shallow curve 共from 2.4 at z = 1.5 cm to 2.2 at z = 20 cm
and Fig. 2 shows the same setup calculated with BEAMnrc. and to 2.1 at z = 40 cm兲. The depth of maximum dose dmax
Tables I and II list the results of the comparison of the simu- was found at 1.36 cm compared to 1.44 cm when calculated
lations with measured data. with a FF. Both these results indicate that irradiations with-
Lateral beam profiles were also simulated for various field out the flattening filter have an increased low-energy compo-
sizes at numerous depths. Figure 3 shows the comparison of nent. The ratios of central-axis depth-dose with and without
Monte Carlo calculations to measured data for a field size of the FF at the field size of 40⫻ 40 cm2 showed the same
10⫻ 10 cm2. The differences between the measurement and shape. The first steep decline showed a ratio of 3.7 at z
the simulations in lateral field size at the 50% dose level, = 0.05 cm and 2.4 at z = 1 cm; the second, a shallow curve
⌬x50, were determined to be 0.5 mm or less for MCNPX starting at 1 cm showed the ratios of 2.1 at z = 20 cm and 2.0
calculations, whereas ⌬x50 was about 0.1 mm in the BEAM- at z = 40 cm. From the depth-dose profile simulations of
nrc calculations. The penumbra widths, P90−10 and P80−20, larger field sizes 共40⫻ 40 cm2兲, an additional contribution to
calculated from the 90% level to the 10% level and from the surface dose could be clearly identified showing a spike
80% to 20%, respectively, showed a systematic bias: most of in the build-up region 共Fig. 4兲. To identify the source of the
the simulations resulted in smaller penumbras compared to spike and to eliminate it, the photon fluences and the electron
the measured ones, whereas the penumbra widths from fluences were further investigated.
MCNPX and BEAMnrc calculations only differed by a frac-
tion of a millimeter. Table III summarizes the results of the
comparisons.
TABLE IV. Photon fluences ␥ per initial electron at the downstream surface
of the ionization chamber 共IC兲 and at the upstream surface of the water
phantom 共WP兲 for two different field sizes A as well as with and without the
FF in the beam. All values at the WP are averaged fluence over the total
surface of the water phantom 共50⫻ 50 cm2兲, and the IC values represent the
averaged fluence through the entire surface of the ionization chamber win-
dow. The average statistical uncertainty of the calculation was smaller than
3%.
FIG. 5. Electron fluences, e, per initial particle, tallied at the upstream
A ␥ at IC 共FF兲 ␥ at IC 共FFF兲 ␥ at WP 共FF兲 ␥ at WP 共FFF兲 surface of the water phantom, as a function of electron energy Ee. The field
共cm2兲 共cm−2 n−1兲 共cm−2 n−1兲 共cm−2 n−1兲 共cm−2 n−1兲 size is 40⫻ 40 cm2. The solid line shows the fluence calculated with a
flattening filter free beamline. Introduction of a 2-mm copper foil 共dotted
10⫻ 10 6.02⫻ 10−5 7.82⫻ 10−5 8.75⫻ 10−8 1.93⫻ 10−7 line兲 reduced the total electron fluence to 22% above the total fluence with
40⫻ 40 6.02⫻ 10−5 7.82⫻ 10−5 1.33⫻ 10−6 1.99⫻ 10−6 the FF in the beamline 共dashed-dotted line兲. The nylon foil reduced e to
46% above this curve.
The electron fluence, e, and the photon fluence, ␥, per FFF lateral dose profiles were calculated for several field
initial particle, n, in units of cm−2 n−1 were tallied and aver- sizes using the copper foil as an electron filter. In standard
aged over the whole surface of the monitor chamber window, photon treatment machines, the shape of the lateral profiles
and the total area of the upstream surface of the water phan- shows a depth dependency. It is particularly strong for fields
tom 共50⫻ 50 cm2兲 for field sizes 10⫻ 10 cm2 and 40 larger than 10⫻ 10 cm2. At shallow depths, i.e., at the depth
⫻ 40 cm2. Tables IV and V list all electron and photon flu- of maximum dose deposition, dmax, an increased shoulder
ences. A notable increase in the electron fluence on the ion- can be found, whereas at larger depths, beyond a depth of
ization chamber window is observed when the FF is re- 10 cm, the shoulder tends to decrease below the dose level at
moved. Many of these electrons reach the phantom surface the central axis. This behavior is mainly due to significant
and contribute to the above-mentioned dose spike in the variation of beam quality across the field. The flattening filter
build-up region. We describe this phenomenon in the Discus- thickness decreases with increasing off-axis distance. For
sion section. this reason the photon spectrum is hardened further on the
To investigate possible effects on the ionization chamber central axis relative to the shoulder area. Figure 7 shows an
performance, e was tallied according to the particle direc- example of this change in shape by overlaying 20⫻ 20 cm2
tion, so the component of backscattered electrons entering field-size profiles. The dose is normalized to the central axis
the ionization chamber could be identified. The amount of and the field width x⬘ is normalized to x50. The shapes of
backscattered electron fluence in the monitor chamber was lateral dose profiles calculated without the FF show much
calculated to be 3.4% 共±0.2% 兲 with the FF and 3.1% less dependency on the depth, as can be seen in Fig. 8.
共±0.2% 兲 without the FF, for the 10⫻ 10 cm2 field. For the Whereas the penumbras of fields from flattening filter free
40⫻ 40 cm2 field, the values were lower, 1.7% 共±0.1% 兲 with calculations do not show significant differences when com-
flattening filter and 1.2% 共±0.1% 兲 without. pared to simulations with flattening filter, the out-of-field
For the FFF case, the relative contribution of electrons, dose contribution close to the treatment-field edge shows sig-
originating from various beamline components, was evalu- nificant differences in small fields. Figure 9 shows the lateral
ated just below the jaws. The majority of the electrons profiles of a 4 ⫻ 4 cm2 treatment field. It can be seen that the
共74.2%兲 were primary electrons from the linear accelerator, out-of-field dose from the FFF accelerator is substantially
whereas the materials of the target button contributed 5.4%. lower, and it falls off faster with distance. This means that a
TABLE VI. Fluence comparison of FFF operation using electron stopping foils, consisting of 11 mm of nylon
and 2 mm of copper. All electron and photon fluences are averaged over the complete surface of the ionization
chamber 共IC兲 window and over the total upstream surface of the water phantom 共WP兲. The field size is 40
⫻ 40 cm2. The average statistical uncertainty of the calculations was smaller than 3%.
11 mm Nylon 2 mm Cu No foil
FIG. 7. Lateral dose profiles calculated for a 20⫻ 20 cm2 field size using the
FIG. 9. Comparison of out-of-field dose calculated with 共solid兲 and without
FF in the beamline. The dose, D, was normalized to the central-axis dose
共dashed兲 the FF for a 4 ⫻ 4 cm2 treatment field. The symbols show the
level, and the lateral width, x⬘, was normalized to x50, the 50% location in
Monte Carlo results, while the lines depict fitted curves to these results.
the penumbra. Note, that with increasing depth in the water phantom the
shape of the profile changes significantly.
of 4 mm, hence leading to an enlargement of the measured higher compared to the FF mode. With the nylon foil, the
penumbra. Also, the transport cutoff energy for electrons values were somewhat less favorable; the electron fluence
contributed to the discrepancy by preventing transport of would be decreased by a factor of 2.7, but would still be
electrons below certain energies. about 48% above the flattened beam, whereas the photon
The simulations of the FFF machine showed a maximum fluence would be 40% higher.
possible dose rate increase on the central axis by more than a At the downstream surface of the ionization chamber, e
factor of 2 compared to irradiation with the FF. Comparison can be reduced by a factor of 5.3, when a 2-mm copper foil
of the photon fluences, ␥, averaged over the phantom sur- is used; this corresponds to 22% higher than the FF value.
face, without and with the FF, showed a ratio FFF/FF of 3.52 The nylon foil decreased the electron fluence by a factor of
for a field size of 10⫻ 10 cm2, whereas this ratio was 1.49 4.4, which corresponds to 46% above the FF value.
for a field size of 40⫻ 40 cm2. These results show that the best choice of electron stop-
These results are promising in terms of the possibility of ping foils would be the copper foil with a thickness of 2 mm
reducing the beam-on time by about a factor of 2 and thus which would result in normal performance of the ionization
cutting the exposure of the patient to leakage photons from chamber, i.e., there would be no danger of saturation because
the treatment head into half. In small treatment fields deliv- the dynamic range of this device can handle a 22% signal
ered by a conventional accelerator, about 50% of the photons increase.18
would be absorbed or scattered in the FF, and add to the The lateral profiles calculated with and without the FF in
out-of-field dose. With the FF removed, most of these pho- the beamline show significant differences 共Figs. 7 and 8兲.
tons could contribute to the target dose. The FFF profiles, in contrast to the flattened beam, exhibit a
The thickness of the bremsstrahlung target 共6-MV mode兲 strong variation across the field exceeding a factor of 2 in
that is given in the available technical documentation is not large fields. The shape of these profiles exhibits a relatively
sufficient to completely stop the incident primary electrons. small variation with depth. The shape of the flattened beam
changes substantially: at shallow depths, it is concave near
This can be confirmed by a simple continuous slowing-down
the center and becomes convex as the depth increases. The
calculation. In a conventional accelerator, these electrons are
out-of-field dose calculated without the flattening filter is
blocked by the FF. If the FF is removed, the electron fluence
smaller outside the field edge for small field sizes. Note that
on the monitor chamber surface increases by a factor of 6.2
in the region up to about 30 cm near the treatment field,
based on Monte Carlo simulations 共Table V兲. The increase
scatter from the treatment beam dominates over treatment-
poses the risk of placing the chamber outside the range of its
head leakage.19 Hence, a patient would be exposed to signifi-
reliable operation. Electron fluence also increases on the
cantly less radiation to normal tissue.
phantom surface 共Table V兲 indicating a potential risk of de-
The calculated total scatter factors, Sc,p, defined as the
livering an elevated skin dose as can be seen in Fig. 4. The
dose rate at a reference depth for a given field size 共x
number of electrons reaching the phantom strongly depends
⫻ x cm2兲 divided by the dose rate at the same point and
on the field size: the calculations show that electron fluence
depth for the reference field size 共10⫻ 10 cm2兲, showed less
in the 40⫻ 40 cm2 field is about 6 times higher than that in a systematic variation in the case of an unflattened beam 共FFF兲
10⫻ 10 cm2 field 共Table V兲. This is because primary elec- compared with the values for the system with FF. Sc,p for the
trons leaving the target have a broad angular distribution and, unflattened beam ranges from 0.927 at 2 ⫻ 2 cm2 to 1.047 at
after further collisions downstream, can be considered as 40⫻ 40 cm2 compared to a range of 0.899 for the 2
originating from a volume source located upstream of the ⫻ 2 cm2 field size to 1.053 at 40⫻ 40 cm2 for the FF setup.
field defining jaws. The energy distribution of electrons Fits to the calculated Sc,p data yielded
reaching the phantom peaks around 1.6 MeV 共Fig. 5兲. The
shape of the peak is characteristic of the spectrum broaden- 0.8894 + 0.175 01x
f共x兲 = 共3兲
ing of a monoenergetic beam due to ionization collisions 1 + 0.163 88x
共energy straggling兲. Its location is consistent with the above-
for the FFF beamline shown as the solid curve in Fig. 10 and
mentioned continuous slowing-down calculation.
On the one hand, these excess electrons should be pre- 0.8552 + 0.1479x
f共x兲 = 共4兲
vented from reaching the ionization chamber and the phan- 1 + 0.132 74x
tom; on the other hand, any mechanism used to accomplish
this should avoid attenuation and scattering of the brems- for the FF setup shown as the dashed line in Fig. 10.
The smaller systematic variations in Sc,p for the FFF ma-
strahlung photons. Knowing the energy spectrum of elec-
chine is mainly due to reduction of scattered photon or head
trons 共Fig. 5兲, and therefore their ranges, we chose to inves-
scatter component of the total scatter factor.
tigate two different electron-stopping foils 共2-mm copper and
11-mm nylon兲.
By introducing the copper foil into the beam, the electron V. CONCLUSIONS
fluence, averaged over the phantom surface 共50⫻ 50 cm2兲 A Monte Carlo model to simulate the 6-MV beam of a
could be decreased by a factor of 3.1 at the phantom surface, Varian Clinac 2100 with and without a FF has been built,
whereas it was still 28% above the value with the FF in tested, and validated using experimental commissioning data
place. The photon fluence at this location would be 37% measured at M. D. Anderson. Depth-dose profiles, lateral