You are on page 1of 9

A flattening filter free photon treatment concept evaluation

with Monte Carlo


U. Titt,a兲 O. N. Vassiliev, F. Pönisch, L. Dong, H. Liu, and R. Mohan
M. D. Anderson Cancer Center, The University of Texas, 1515 Holcombe Boulevard, Houston,
Texas 77030
共Received 9 February 2005; revised 28 March 2006; accepted for publication 28 March 2006;
published 10 May 2006兲
In principle, the concept of flat initial radiation-dose distribution across the beam is unnecessary for
intensity modulated radiation therapy. Dynamic leaf positioning during irradiation could appropri-
ately adjust the fluence distribution of an unflattened beam that is peaked in the center and deliver
the desired uniform or nonuniform dose distribution. Removing the flattening filter could lead to
reduced treatment time through higher dose rates and reduced scatter, because there would be
substantially less material in the beam; and possibly other dosimetric and clinical advantages. This
work aims to evaluate the properties of a flattening filter free clinical accelerator and to investigate
its possible advantages in clinical intensity modulated radiation therapy applications by simulating
a Varian 2100-based treatment delivery system with Monte Carlo techniques. Several depth-dose
curves and lateral dose distribution profiles have been created for various field sizes, with and
without the flattening filter. Data computed with this model were used to evaluate the overall quality
of such a system in terms of changes in dose rate, photon and electron fluence, and reduction in
out-of-field stray dose from the scattered components and were compared to the corresponding data
for a standard treatment head with a flattening filter. The results of the simulations of the flattening
filter free system show that a substantial increase in dose rate can be achieved, which would reduce
the beam on time and decrease the out-of-field dose for patients due to reduced head-leakage dose.
Also close to the treatment field edge, a significant improvement in out-of-field dose could be
observed for small fields, which can be attributed to the change in the photon spectra, when the
flattening filter is removed from the beamline. © 2006 American Association of Physicists in
Medicine. 关DOI: 10.1118/1.2198327兴

Key words: Monte Carlo simulations, flattening filter free, treatment head modification, beam line
modification

I. INTRODUCTION dose close to the penumbra; and changes in the lateral dose
profiles, as a function of depth in a water phantom, are in-
Conventional clinical accelerators used in radiotherapy are vestigated in this work.
equipped with a flattening filter 共FF兲. The FF is designed to Several previous investigations of FF free machines have
produce approximately uniform dose distribution across the been undertaken, primarily to investigate the physical prop-
field at a certain depth in a homogeneous phantom. Such a erties of the beam, such as changes in dmax, the location of
field is typically not the best solution for complex treatment maximum dose deposition and energy spectrum changes.3–5
problems. For this reason, additional beam-modifying de- Variation of the beam quality with off-axis distance have
vices are used, such as wedges, compensators, and dynamic been researched by Zefkili et al.6 and Lee et al.7 O’Brien
multileaf collimators 共MLCs兲. MLCs are quite flexible in et al.8 removed the FF from a Therac-6 linear accelerator
terms of shaping fluence distribution across the field. In prin- 共AECL, Atomic Energy of Canada Ltd., Mississauga, On-
ciple, they render the FF unnecessary. tario, Canada兲 in an effort to reduce the treatment time of a
stereotactic radiosurgery procedure. Gillies et al.9 discussed
The FF introduces a very substantial amount of material
engineering modifications to a Therac-6 linear accelerator to
into the beamline. It is one of the most significant sources of
perform dynamic stereotactic radiotherapy, including a head
undesired radiation scatter.1,2 More than 50% of primary ra-
modification, which allowed easy removal of the FF. Sixel
diation can be scattered or absorbed in the FF. Relevant nu- and Faddegon4 calculated photon spectra for small fields on
merical data will be presented later in this study. the Therac-6 linear accelerator as well as depth-dose and
By removing the FF an increase in dose rate can be profiles with and without the FF. Further research was con-
achieved, leading to a shorter irradiation time per treatment ducted to investigate portal imaging properties,10 dose deliv-
fraction. This will also reduce treatment head scatter, out-of- ery enhancement in high-Z material-loaded tumors,11 and an
field exposure of the patient, and leakage dose through the application in helical tomotherapy.12 Fu et al.13 investigated
MLC. The implications of removing the FF with respect to intensity modulated radiation therapy 共IMRT兲 delivery time
increased fluence and dose rates; changes in the out-of-field with and without the FF. They used a BJ-6B linear accelera-

1595 Med. Phys. 33 „6…, June 2006 0094-2405/2006/33„6…/1595/8/$23.00 © 2006 Am. Assoc. Phys. Med. 1595
1596 Titt et al.: FFF Monte Carlo simulation 1596

tor 共in 6-MV photon mode兲 and the Focus 共version 3.2.1,
CMS, Computerized Medical Systems, St. Louis, Missouri,
USA兲 treatment-planning system.
In this paper, we present our first results on the way to a
flattening filter free 共FFF兲 clinical accelerator, based on the
Varian Clinac 2100 共Varian Medical Systems Inc., Palo Alto,
California, USA兲. Numerous FFF irradiations have been
simulated to investigate the properties of the device and the
first results are being presented. The study focuses on the
6-MV photon mode.

II. MATERIALS AND METHODS


The Monte Carlo radiation transport programs used in this
study are BEAMnrc14 and MCNPX 2.5d.15 Both programs
require input files, defining the geometry and various physics FIG. 1. Central-axis depth-dose profile for benchmarking the MCNPX pro-
gram. Relative dose D is shown as a function of depth, d, in the water
parameters. The geometry used in both programs was based phantom. The number of calculated initial electron histories is 108, field size
on the Varian Clinac 2100 treatment head. Although the ge- is 10⫻ 10 cm2, initial mean electron energy is 6.2 MeV, and energy spread
ometry in BEAMnrc was restricted to the major beamline is 3% 共FWHM兲. The statistical uncertainty of the simulation was smaller
than 1% at dmax, and the difference to data measured during the commis-
components, a more detailed model was developed for
sioning of a Varian Clinac 2100C showed a maximum of 1.1%. The depth,
MCNPX, including the structural components and the lead dmax, of the maximum dose differed by 1.2± 0.1 mm between measurement
shielding, in order to investigate realistic out-of-field doses. and simulation.
Both systems use table-based cross-sectional data. The elec-
tron transport was terminated when the kinetic energy of the
electron reached a value below 100 keV 共MCNPX兲 or cal fluctuations in the Monte Carlo results, the calculated
200 keV 共BEAMnrc兲. Photon transport was terminated at depth-dose curves were normalized to the sum of the dose
10 keV in both programs. Initially, the complete radiation value between depths of 5 cm and 20 cm, using the follow-
transport was modeled, starting with electrons stopping in a ing equation:
target to produce bremsstrahlung photons. BEAMnrc and
20 cm
MCNPX both provide a variety of variance reduction tech-
niques, which have been utilized in our simulations. In the D= 兺
5 cm
D共z兲dz, 共1兲
case of BEAMnrc, selective bremsstrahlung splitting was
used, and in the case of MCNPX, a weight window mesh and compared with measurements. Both results, measured
was created to improve the statistics in the receptors and thus and calculated, could then be compared with respect to the
lower the statistical uncertainties. The bremsstrahlung pho- relative value of the maximum dose Dmax and the corre-
tons were transported through the first collimator, a thin sponding depth dmax. Dose distributions from field sizes of
vacuum window, the FF followed by an ionization chamber,
and through a set of jaws into a cubic water phantom 共50
⫻ 50⫻ 50 cm3 in size兲 where all dose profiles were tallied.
The properties of the electron source were determined
based on the work of Cho et al.16 It was assumed that the
primary electron beam is parallel, and the energy spectrum
and lateral spread of the electron fluence are of Gaussian
shape. The energy distribution is centered at 6.2 MeV and
has a 3% full width at half-maximum 共FWHM兲. The lateral
spread is 1 mm FWHM.
For subsequent calculations, phase-space files were used,
to eliminate the CPU-intensive simulation of bremsstrahlung
production. The Varian accelerators at M. D. Anderson Can-
cer Center deliver photons with different energy distribu-
tions, denoted as 6 MV and 18 MV. All data presented in the
following refer to the 6 MV mode. Extensive benchmarking
simulations have been undertaken to compare simulations of FIG. 2. Central-axis depth-dose profile for benchmarking the BEAMnrc pro-
the model including the FF with measurements performed gram, showing relative dose, D, as a function of depth, d, in water. The
during commissioning of a photon treatment machine at M. number of calculated initial electron histories is 2.7⫻ 108, field size is 10
⫻ 10 cm2, initial mean electron energy is 6.2 MeV, and energy spread is 3%
D. Anderson. The measured central axis depth-dose curves
共FWHM兲. The statistical uncertainty of these simulations was smaller than
were normalized to unity at the depth, dmax, of the maximum 0.43% at dmax, and the maximum relative dose difference to measured data
dose deposition, Dmax. To minimize the influence of statisti- was 1.1%. The location of dmax differed by 0.03 mm.

Medical Physics, Vol. 33, No. 6, June 2006


1597 Titt et al.: FFF Monte Carlo simulation 1597

TABLE I. Comparison of calculated 共MCNPX兲 and measured central-axis


depth-dose profiles at various field sizes. In all cases, the initial electron
energy was 6.2 MeV. A denotes the irradiation field size, dmax denotes the
location of the maximum dose, and ⌬Dmax is the relative dose discrepancy
between the measurement and the calculations at dmax.

A / cm2 dsim
max / cm dmeas
max / cm ⌬Dmax

4⫻4 1.55 1.64 0.70%


10⫻ 10 1.44 1.56 1.13%
20⫻ 20 1.39 1.36 2.42%
40⫻ 40 1.26 1.26 3.60%

4 ⫻ 4 cm2, 10⫻ 10 cm2, 20⫻ 20 cm2, and 40⫻ 40 cm2 have


been investigated.
The measured lateral dose profiles were normalized to FIG. 3. Benchmark comparison of lateral dose profiles, calculated with MC-
NPX and BEAMnrc at a depth of 6.0 cm in a water phantom using a field
unity on the central axis, and the calculations were normal- size of 10⫻ 10 cm2 with a measured dose profile. The maximum relative
ized to the sum of the dose values between the two 50% difference, ⌬max, in the central high dose region of the profile, excluding the
locations in the penumbra, −x50 and +x50, penumbra and beyond, shows values between 1.2% and 1.8%. The penum-
bra widths between the 80% level and the 20% level are denoted by P80–20.
x50
D= 兺 D共x兲dx,
−x50
共2兲
ence, electron filters of various materials were simulated up-
and compared to the measured values. The size of the pen- stream of the ionization chamber, and their influence on
umbra in the 90% to 10% interval and in the 80% to 20% photon and electron fluences was evaluated.
interval, the differences of x50 from measurements and simu- The statistical uncertainty of the simulations 共one ␴ inter-
lations, as well as the maximum differences at the center of val兲 was below 1% for the depth-dose profiles of all field
the beam and at the field edges were evaluated for three field sizes calculated with BEAMnrc and for field sizes of up to
sizes, 10⫻ 10 cm2, 20⫻ 20 cm2, and 40⫻ 40 cm2. 20⫻ 20 cm2 calculated with MCNPX. The 40⫻ 40 cm2 field
Removing the FF from the model enabled us to investi- size resulted in a statistical uncertainty of up to 3% when
gate the properties of an FFF machine with respect to MCNPX was used. The simulations were performed until the
changes in the central axis depth-dose profiles, the lateral statistical uncertainties reached values lower than 1% at and
dose profiles at various depths in the water phantom, and around the central beam axis.
increases in the electron and photon fluences and the respec-
tive energy fluences at the downstream surface of the moni- III. RESULTS
tor chamber and at the surface of the water phantom.
A. Benchmark calculations
The electron and photon fluences at the monitor chamber
and at the phantom were of particular interest. Increased Central axis depth-dose profiles of various field sizes with
electron fluence in the ionization chamber could result in the FF in the beamline were computed with MCNPX and
saturation of the detector and, hence, lead to nonlinearities in with BEAMnrc to benchmark the Monte Carlo system. Fig-
the ratio of the delivered therapeutic dose to monitor units ure 1 shows an example of a central-axis depth-dose curve
共MU兲 measured. The electron fluence at the surface of the
phantom also had to be minimized, in order to avoid exces-
sive dose in the build-up region. To reduce the electron flu- TABLE III. Comparison of measured and calculated lateral dose profiles. The
field size is denoted as A, ⌬x50 is the lateral difference measured at the 50%
dose point in the penumbra, and P90–10 as well as P80–20 describe the width
of the penumbra measured from the 90% point to the 10% point and from
TABLE II. Comparison of calculated 共BEAMnrc兲 and measured central-axis the 80% to 20%, respectively. The maximum relative differences are given
depth-dose profiles at various field sizes. Initial electron energy was again by max共⌬D兲.
6.2 MeV. A denotes the irradiation field size, dmax denotes the location of
the maximum dose, and ⌬Dmax is the relative dose discrepancy between the Method A / cm2 ⌬x50 / mm P90–10 / mm P80–20 / mm max共⌬D兲
measurement and the calculations at dmax.
Measurement 10⫻ 10 11 5.5
A / cm2 dsim dmeas ⌬Dmax MCNPX 10⫻ 10 0.5 11 3.5 1.2%
max / cm max / cm
BEAMnrc 10⫻ 10 0.1 8.5 3.4 1.1%
Measurement 20⫻ 20 12 6
4⫻4 1.50 1.64 0.30% MCNPX 20⫻ 20 0.1 7 3.5 2.0%
10⫻ 10 1.50 1.56 1.10% BEAMnrc 20⫻ 20 0.1 10 3.3 1.0%
20⫻ 20 1.37 1.36 0.04% Measurement 40⫻ 40 11
40⫻ 40 1.50 1.29 1.60% BEAMnrc 40⫻ 40 0.2 7 6.0%

Medical Physics, Vol. 33, No. 6, June 2006


1598 Titt et al.: FFF Monte Carlo simulation 1598

TABLE V. Electron fluences ␾e per initial electron at the downstream surface


of the ionization chamber 共IC兲 and at the upstream surface of the water
phantom 共WP兲 for two different field sizes A as well as with and without the
FF in the beam. All values at the WP are averaged fluence over the total
surface of the water phantom 共50⫻ 50 cm2兲, and the IC values represent the
fluences averaged over the entire surface of the downstream ionization
chamber window. The average statistical uncertainty of the calculation was
smaller than 3%.

A ␾e at IC 共FF兲 ␾e at IC 共FFF兲 ␾e at WP 共FF兲 ␾e at WP 共FFF兲


共cm2兲 共cm−2 n−1兲 共cm−2 n−1兲 共cm−2 n−1兲 共cm−2 n−1兲

10⫻ 10 4.63⫻ 10−7 2.87⫻ 10−6 8.88⫻ 10−10 3.13⫻ 10−9


40⫻ 40 4.59⫻ 10−7 2.94⫻ 10−6 4.81⫻ 10−9 1.93⫻ 10−8

FIG. 4. Comparison of simulated central-axis depth-dose 共per initial par- dent particle兲 by a factor of 2 at least. The ratio of the flat-
ticle兲 profiles in a 40⫻ 40 cm2 field, with and without the FF, normalized to tening filter-free simulation to the results with the FF 共field
100 at the shoulder of the FFF dose falloff. Besides the overall dose in-
crease, the profile from the filter-free simulation clearly shows a sharp spike size 10⫻ 10 cm2兲 showed two different ranges; in both, the
in the build-up region, which is attributed to excessive electron fluence. ratio is almost linearly dependent on the depth z. First, a
steep decline in the build up region could be observed 共from
a ratio of 3.2 at z = 0.05 cm to 2.4 at z = 1.2 cm兲 followed by
calculated with MCNPX using a field size of 10⫻ 10 cm2, a shallow curve 共from 2.4 at z = 1.5 cm to 2.2 at z = 20 cm
and Fig. 2 shows the same setup calculated with BEAMnrc. and to 2.1 at z = 40 cm兲. The depth of maximum dose dmax
Tables I and II list the results of the comparison of the simu- was found at 1.36 cm compared to 1.44 cm when calculated
lations with measured data. with a FF. Both these results indicate that irradiations with-
Lateral beam profiles were also simulated for various field out the flattening filter have an increased low-energy compo-
sizes at numerous depths. Figure 3 shows the comparison of nent. The ratios of central-axis depth-dose with and without
Monte Carlo calculations to measured data for a field size of the FF at the field size of 40⫻ 40 cm2 showed the same
10⫻ 10 cm2. The differences between the measurement and shape. The first steep decline showed a ratio of 3.7 at z
the simulations in lateral field size at the 50% dose level, = 0.05 cm and 2.4 at z = 1 cm; the second, a shallow curve
⌬x50, were determined to be 0.5 mm or less for MCNPX starting at 1 cm showed the ratios of 2.1 at z = 20 cm and 2.0
calculations, whereas ⌬x50 was about 0.1 mm in the BEAM- at z = 40 cm. From the depth-dose profile simulations of
nrc calculations. The penumbra widths, P90−10 and P80−20, larger field sizes 共40⫻ 40 cm2兲, an additional contribution to
calculated from the 90% level to the 10% level and from the surface dose could be clearly identified showing a spike
80% to 20%, respectively, showed a systematic bias: most of in the build-up region 共Fig. 4兲. To identify the source of the
the simulations resulted in smaller penumbras compared to spike and to eliminate it, the photon fluences and the electron
the measured ones, whereas the penumbra widths from fluences were further investigated.
MCNPX and BEAMnrc calculations only differed by a frac-
tion of a millimeter. Table III summarizes the results of the
comparisons.

B. Simulations without the flattening filter, fluence


evaluation
Initial simulations of irradiations without the FF showed
an overall increase of the dose on the central axis 共per inci-

TABLE IV. Photon fluences ␾␥ per initial electron at the downstream surface
of the ionization chamber 共IC兲 and at the upstream surface of the water
phantom 共WP兲 for two different field sizes A as well as with and without the
FF in the beam. All values at the WP are averaged fluence over the total
surface of the water phantom 共50⫻ 50 cm2兲, and the IC values represent the
averaged fluence through the entire surface of the ionization chamber win-
dow. The average statistical uncertainty of the calculation was smaller than
3%.
FIG. 5. Electron fluences, ␾e, per initial particle, tallied at the upstream
A ␾␥ at IC 共FF兲 ␾␥ at IC 共FFF兲 ␾␥ at WP 共FF兲 ␾␥ at WP 共FFF兲 surface of the water phantom, as a function of electron energy Ee. The field
共cm2兲 共cm−2 n−1兲 共cm−2 n−1兲 共cm−2 n−1兲 共cm−2 n−1兲 size is 40⫻ 40 cm2. The solid line shows the fluence calculated with a
flattening filter free beamline. Introduction of a 2-mm copper foil 共dotted
10⫻ 10 6.02⫻ 10−5 7.82⫻ 10−5 8.75⫻ 10−8 1.93⫻ 10−7 line兲 reduced the total electron fluence to 22% above the total fluence with
40⫻ 40 6.02⫻ 10−5 7.82⫻ 10−5 1.33⫻ 10−6 1.99⫻ 10−6 the FF in the beamline 共dashed-dotted line兲. The nylon foil reduced ␾e to
46% above this curve.

Medical Physics, Vol. 33, No. 6, June 2006


1599 Titt et al.: FFF Monte Carlo simulation 1599

Further contributions were 2.6% from the primary collima-


tor, 1.3% from the vacuum window, and 16.5% from all
remaining beamline components. To prevent these excess
electrons from reaching the phantom surface and saturating
the ionization chamber, two different foils were simulated in
the position of the removed FF. An 11-mm nylon slab and a
2-mm copper foil were tested at a field size of 40⫻ 40 cm2
concerning electron stopping and photon transparency. Fig-
ure 5 shows ␾e as a function of electron energy, Ee, at the
upstream surface of the water phantom, and Fig. 6 shows the
photon fluence, ␾␥, as a function of photon energy E␥ at the
same location. Table VI lists all fluence values from the ny-
lon and the copper foil calculations at both locations, the
FIG. 6. Photon fluences ␾␥, per initial electron, at the upstream surface of ionization chamber, and the water phantom. The copper foil
the water phantom as a function of photon energy E␥. The field size for these removes more electrons and low energy photons, including
simulations was 40⫻ 40 cm2. The solid line shows the fluence calculated characteristic x rays, from the beam.
with a flattening filter free beamline. Introduction of a 2 mm copper foil
共dotted line兲 reduced the total photon fluence by 9%, and results with the
nylon foil showed a reduction of 7%.
C. Lateral dose profiles FF vs FFF

The electron fluence, ␾e, and the photon fluence, ␾␥, per FFF lateral dose profiles were calculated for several field
initial particle, n, in units of cm−2 n−1 were tallied and aver- sizes using the copper foil as an electron filter. In standard
aged over the whole surface of the monitor chamber window, photon treatment machines, the shape of the lateral profiles
and the total area of the upstream surface of the water phan- shows a depth dependency. It is particularly strong for fields
tom 共50⫻ 50 cm2兲 for field sizes 10⫻ 10 cm2 and 40 larger than 10⫻ 10 cm2. At shallow depths, i.e., at the depth
⫻ 40 cm2. Tables IV and V list all electron and photon flu- of maximum dose deposition, dmax, an increased shoulder
ences. A notable increase in the electron fluence on the ion- can be found, whereas at larger depths, beyond a depth of
ization chamber window is observed when the FF is re- 10 cm, the shoulder tends to decrease below the dose level at
moved. Many of these electrons reach the phantom surface the central axis. This behavior is mainly due to significant
and contribute to the above-mentioned dose spike in the variation of beam quality across the field. The flattening filter
build-up region. We describe this phenomenon in the Discus- thickness decreases with increasing off-axis distance. For
sion section. this reason the photon spectrum is hardened further on the
To investigate possible effects on the ionization chamber central axis relative to the shoulder area. Figure 7 shows an
performance, ␾e was tallied according to the particle direc- example of this change in shape by overlaying 20⫻ 20 cm2
tion, so the component of backscattered electrons entering field-size profiles. The dose is normalized to the central axis
the ionization chamber could be identified. The amount of and the field width x⬘ is normalized to x50. The shapes of
backscattered electron fluence in the monitor chamber was lateral dose profiles calculated without the FF show much
calculated to be 3.4% 共±0.2% 兲 with the FF and 3.1% less dependency on the depth, as can be seen in Fig. 8.
共±0.2% 兲 without the FF, for the 10⫻ 10 cm2 field. For the Whereas the penumbras of fields from flattening filter free
40⫻ 40 cm2 field, the values were lower, 1.7% 共±0.1% 兲 with calculations do not show significant differences when com-
flattening filter and 1.2% 共±0.1% 兲 without. pared to simulations with flattening filter, the out-of-field
For the FFF case, the relative contribution of electrons, dose contribution close to the treatment-field edge shows sig-
originating from various beamline components, was evalu- nificant differences in small fields. Figure 9 shows the lateral
ated just below the jaws. The majority of the electrons profiles of a 4 ⫻ 4 cm2 treatment field. It can be seen that the
共74.2%兲 were primary electrons from the linear accelerator, out-of-field dose from the FFF accelerator is substantially
whereas the materials of the target button contributed 5.4%. lower, and it falls off faster with distance. This means that a

TABLE VI. Fluence comparison of FFF operation using electron stopping foils, consisting of 11 mm of nylon
and 2 mm of copper. All electron and photon fluences are averaged over the complete surface of the ionization
chamber 共IC兲 window and over the total upstream surface of the water phantom 共WP兲. The field size is 40
⫻ 40 cm2. The average statistical uncertainty of the calculations was smaller than 3%.

11 mm Nylon 2 mm Cu No foil

␾e at IC 共cm−2 n−1兲 6.70⫻ 10−7 5.60⫻ 10−7 2.94⫻ 10−6


␾e at WP 共cm−2 n−1兲 7.15⫻ 10−9 6.16⫻ 10−9 1.93⫻ 10−8
␾␥ at IC 共cm−2 n−1兲 7.47⫻ 10−5 7.37⫻ 10−5 7.82⫻ 10−5
␾␥ at WP 共cm−2 n−1兲 1.87⫻ 10−6 1.83⫻ 10−6 1.99⫻ 10−6

Medical Physics, Vol. 33, No. 6, June 2006


1600 Titt et al.: FFF Monte Carlo simulation 1600

FIG. 7. Lateral dose profiles calculated for a 20⫻ 20 cm2 field size using the
FIG. 9. Comparison of out-of-field dose calculated with 共solid兲 and without
FF in the beamline. The dose, D, was normalized to the central-axis dose
共dashed兲 the FF for a 4 ⫻ 4 cm2 treatment field. The symbols show the
level, and the lateral width, x⬘, was normalized to x50, the 50% location in
Monte Carlo results, while the lines depict fitted curves to these results.
the penumbra. Note, that with increasing depth in the water phantom the
shape of the profile changes significantly.

estimation in the entrance region was expected. Comparative


significant reduction in the out-of-field dose and enhanced calculations of bremsstrahlung production with identical
sparing of critical tissues and organs close to small treatment physics options and geometries showed that MCNPX pro-
fields can be achieved. duced up to 20% more photons with energies below 200 keV
compared to BEAMnrc, which showed far better agreement
D. Total scatter factor with experimental data. At a depth of 5 cm and more, both
the MCNPX and the BEAMnrc depth-dose calculations
The total scatter factors, Sc,p, defined as the dose rate at a
showed excellent agreement with measured data. The data
reference depth for a given field size, divided by the dose
obtained from BEAMnrc show considerably better agree-
rate at the same depth for the reference field size 共10
ment at smaller depths, only differences of 0.04% through
⫻ 10 cm2兲17 were calculated for several square fields. Both,
1.6% in this region could be observed. The lateral dose pro-
an FFF treatment head and a standard setup with a FF have
files show very good agreement in the high-dose region close
been simulated. Table VII lists all values, including the sta-
to the central axis, ±1% to ±2% at field sizes below 20
tistical uncertainties, describing the 1␴ confidence interval.
⫻ 20 cm2, whereas for larger fields 共40⫻ 40 cm2兲, the uncer-
tainties are somewhat larger. At the treatment-field edges, the
IV. DISCUSSION
difference could be up to 6%. The penumbras 共80% to 20%兲
The depth-dose calculations with MCNPX, which were show reasonable agreement with the measurements, showing
used for benchmarking this Monte Carlo program, show that differences from 1.0 mm to 2.7 mm. The measurements were
the program overestimates the maximum dose deposition performed with an ionization chamber, which had a diameter
from 0.7% at a field size of 4 ⫻ 4 cm2 to up to 3.6% at a field
size of 40⫻ 40 cm2 compared to measurements. This over-
TABLE VII. Total scatter factors, Sc,p, for square fields, normalized to dmax of
a 10⫻ 10 cm2 field for the FFF system and a standard Varian Clinac 2100
setup with the FF. The statistical uncertainties describe the 1␴ - confidence
interval.

A / cm2 Sc,p 共FFF兲 Sc,p 共FF兲

2⫻2 0.927± 0.48% 0.899± 0.45%


3⫻3 0.950± 0.51% 0.939± 0.53%
4⫻4 0.972± 0.53% 0.951± 0.52%
6⫻6 0.979± 0.59% 0.972± 0.53%
8⫻8 0.982± 0.63% 0.987± 0.61%
10⫻ 10 1 1
12⫻ 12 1.007± 0.63% 1.011± 0.59%
15⫻ 15 1.011± 0.67% 1.023± 0.58%
20⫻ 20 1.032± 0.62% 1.043± 0.67%
FIG. 8. Lateral dose profiles of a 20⫻ 20 cm2 field calculated in the FFF 25⫻ 25 1.037± 0.66% 1.057± 0.66%
machine. The dose, D, was normalized to the central axis dose level, and the 30⫻ 30 1.029± 0.66% 1.061± 0.75%
lateral width x⬘ was normalized to x50. Note that in this case, the depth- 35⫻ 35 1.044± 0.66% 1.072± 0.74%
dependency of the dose inside the field seen in the case with the FF has 40⫻ 40 1.047± 0.70% 1.053± 0.76%
virtually disappeared.

Medical Physics, Vol. 33, No. 6, June 2006


1601 Titt et al.: FFF Monte Carlo simulation 1601

of 4 mm, hence leading to an enlargement of the measured higher compared to the FF mode. With the nylon foil, the
penumbra. Also, the transport cutoff energy for electrons values were somewhat less favorable; the electron fluence
contributed to the discrepancy by preventing transport of would be decreased by a factor of 2.7, but would still be
electrons below certain energies. about 48% above the flattened beam, whereas the photon
The simulations of the FFF machine showed a maximum fluence would be 40% higher.
possible dose rate increase on the central axis by more than a At the downstream surface of the ionization chamber, ␾e
factor of 2 compared to irradiation with the FF. Comparison can be reduced by a factor of 5.3, when a 2-mm copper foil
of the photon fluences, ␾␥, averaged over the phantom sur- is used; this corresponds to 22% higher than the FF value.
face, without and with the FF, showed a ratio FFF/FF of 3.52 The nylon foil decreased the electron fluence by a factor of
for a field size of 10⫻ 10 cm2, whereas this ratio was 1.49 4.4, which corresponds to 46% above the FF value.
for a field size of 40⫻ 40 cm2. These results show that the best choice of electron stop-
These results are promising in terms of the possibility of ping foils would be the copper foil with a thickness of 2 mm
reducing the beam-on time by about a factor of 2 and thus which would result in normal performance of the ionization
cutting the exposure of the patient to leakage photons from chamber, i.e., there would be no danger of saturation because
the treatment head into half. In small treatment fields deliv- the dynamic range of this device can handle a 22% signal
ered by a conventional accelerator, about 50% of the photons increase.18
would be absorbed or scattered in the FF, and add to the The lateral profiles calculated with and without the FF in
out-of-field dose. With the FF removed, most of these pho- the beamline show significant differences 共Figs. 7 and 8兲.
tons could contribute to the target dose. The FFF profiles, in contrast to the flattened beam, exhibit a
The thickness of the bremsstrahlung target 共6-MV mode兲 strong variation across the field exceeding a factor of 2 in
that is given in the available technical documentation is not large fields. The shape of these profiles exhibits a relatively
sufficient to completely stop the incident primary electrons. small variation with depth. The shape of the flattened beam
changes substantially: at shallow depths, it is concave near
This can be confirmed by a simple continuous slowing-down
the center and becomes convex as the depth increases. The
calculation. In a conventional accelerator, these electrons are
out-of-field dose calculated without the flattening filter is
blocked by the FF. If the FF is removed, the electron fluence
smaller outside the field edge for small field sizes. Note that
on the monitor chamber surface increases by a factor of 6.2
in the region up to about 30 cm near the treatment field,
based on Monte Carlo simulations 共Table V兲. The increase
scatter from the treatment beam dominates over treatment-
poses the risk of placing the chamber outside the range of its
head leakage.19 Hence, a patient would be exposed to signifi-
reliable operation. Electron fluence also increases on the
cantly less radiation to normal tissue.
phantom surface 共Table V兲 indicating a potential risk of de-
The calculated total scatter factors, Sc,p, defined as the
livering an elevated skin dose as can be seen in Fig. 4. The
dose rate at a reference depth for a given field size 共x
number of electrons reaching the phantom strongly depends
⫻ x cm2兲 divided by the dose rate at the same point and
on the field size: the calculations show that electron fluence
depth for the reference field size 共10⫻ 10 cm2兲, showed less
in the 40⫻ 40 cm2 field is about 6 times higher than that in a systematic variation in the case of an unflattened beam 共FFF兲
10⫻ 10 cm2 field 共Table V兲. This is because primary elec- compared with the values for the system with FF. Sc,p for the
trons leaving the target have a broad angular distribution and, unflattened beam ranges from 0.927 at 2 ⫻ 2 cm2 to 1.047 at
after further collisions downstream, can be considered as 40⫻ 40 cm2 compared to a range of 0.899 for the 2
originating from a volume source located upstream of the ⫻ 2 cm2 field size to 1.053 at 40⫻ 40 cm2 for the FF setup.
field defining jaws. The energy distribution of electrons Fits to the calculated Sc,p data yielded
reaching the phantom peaks around 1.6 MeV 共Fig. 5兲. The
shape of the peak is characteristic of the spectrum broaden- 0.8894 + 0.175 01x
f共x兲 = 共3兲
ing of a monoenergetic beam due to ionization collisions 1 + 0.163 88x
共energy straggling兲. Its location is consistent with the above-
for the FFF beamline shown as the solid curve in Fig. 10 and
mentioned continuous slowing-down calculation.
On the one hand, these excess electrons should be pre- 0.8552 + 0.1479x
f共x兲 = 共4兲
vented from reaching the ionization chamber and the phan- 1 + 0.132 74x
tom; on the other hand, any mechanism used to accomplish
this should avoid attenuation and scattering of the brems- for the FF setup shown as the dashed line in Fig. 10.
The smaller systematic variations in Sc,p for the FFF ma-
strahlung photons. Knowing the energy spectrum of elec-
chine is mainly due to reduction of scattered photon or head
trons 共Fig. 5兲, and therefore their ranges, we chose to inves-
scatter component of the total scatter factor.
tigate two different electron-stopping foils 共2-mm copper and
11-mm nylon兲.
By introducing the copper foil into the beam, the electron V. CONCLUSIONS
fluence, averaged over the phantom surface 共50⫻ 50 cm2兲 A Monte Carlo model to simulate the 6-MV beam of a
could be decreased by a factor of 3.1 at the phantom surface, Varian Clinac 2100 with and without a FF has been built,
whereas it was still 28% above the value with the FF in tested, and validated using experimental commissioning data
place. The photon fluence at this location would be 37% measured at M. D. Anderson. Depth-dose profiles, lateral

Medical Physics, Vol. 33, No. 6, June 2006


1602 Titt et al.: FFF Monte Carlo simulation 1602

following members of Varian Medical Systems for their sup-


port during the course of this project: Jyrki Alakuijla, Jeff
Amacker, Yves Archambault, Cal Huntzinger, Scott Johnson,
Armin Langenegger, Stan Mansfield, Richard Morse, and Ri-
chard Stark.
a兲
Author to whom correspondence should be addressed. Electronic mail:
utitt@mdanderson.org
1
J. Deng, S. B. Jiang, A. Kapur, J. Li, T. Pawlicki, and C.-M. Ma, “Photon
beam characterization and modelling for Monte Carlo treatment plan-
ning,” Phys. Med. Biol. 45, 411–427 共2000兲.
2
M. K. Fix, M. Stampanoni, P. Manser, E. J. Born, R. Mini, and P. Rüeg-
segger, “A multiple source model for 6 MV photon beam dose calcula-
tions using Monte Carlo,” Phys. Med. Biol. 46, 1407–1427 共2001兲.
3
K. E. Sixel and E. B. Podgorsak, “Buildup region and depth of dose
maximum of megavoltage x-ray beams,” Med. Phys. 21, 411–416
FIG. 10. Total scatter factor Sc,p as a function of field size. The open squares 共1994兲.
4
show data calculated for a standard Varian Clinac 2100 with FF, and the K. E. Sixel and B. A. Faddegon, “Calculation of x-ray spectra for radio-
solid squares represent the results for the FFF machine. Uncertainties of the surgical beams,” Med. Phys. 22, 1657–1661 共1995兲.
5
MC simulations show the 1␴ confidence interval. The curves are fits to the C. Guillerminet, E. Duverger, R. Gschwind, L. Makovicka, A. Monnier,
simulation data. F. Hamon, and C. David, “Theoretical and experimental study of spectral
distortions at the output of an accelerator for medical use,” Radiat. Phys.
Chem. 61, 611–613 共2001兲.
beam profiles, and photon and electron fluences as well as 6
S. Zefkili, C. Kappas, and J. C. Rosenwald, “On-axis and off-axis, pri-
total scatter factors have been calculated for the FFF ma- mary dose component in high energy photon beams,” Med. Phys. 21,
chine and compared to data calculated for a standard treat- 799–808 共1994兲.
7
P. C. Lee, “Monte Carlo simulation of the differential beam hardening
ment head with the FF.
effect of a flattening filter on a therapeutic x-ray beam,” Med. Phys. 24,
The results of the Monte Carlo simulation suggest that the 1485–1489 共1997兲.
removal of the FF could result in higher central-axis dose 8
P. F. O’Brien, B. A. Gillies, M. Schwartz, C. Young, and P. Davey, “Ra-
rates and, hence, shorter IMRT treatments. The out-of-field diosurgery with unflattened 6-MV photon beams,” Med. Phys. 18, 519–
dose 共i.e., head scatter, MLC leakage, etc.兲 to the patient can 521 共1991兲.
9
B. A. Gillies, P. F. O’Brien, R. McVittie, C. McParland, and H. Easton,
be decreased, because the beam-on time of the treatment unit
“Engineering modifications for dynamic stereotactically assisted radio-
can be decreased by up to a factor of 2. Close to the treat- therapy,” Med. Phys. 20, 1491–1495 共1993兲.
ment field, a significant decrease in out-of-field dose from 10
A. Tsechanski, A. F. Bielajew, S. Faermann, and Y. Krutman, “A thin
in-field-scatter photons could be observed for small treat- target approach for portal imaging in medical accelerators,”Phys. Med.
ment fields. Biol. 43, 2221–2236 共1998兲.
11
J. L. Robar, S. A. Riccio, and M. A. Martin, “Tumor dose enhancement
Excess electrons, which would normally be scattered and
using modified megavoltage beams and contrast media,” Phys. Med. Biol.
stopped in the FF would reach the patient. However, they can 47, 2433–2449 共2002兲.
be filtered out with a thin copper foil without significantly 12
R. Jeraj, T. R. Mackie, J. Balog, G. Olivera, D. Pearson, J. Kapatoes, K.
compromising the enhanced photon dose rate. Ruchala, and P. Reckwendt, “Radiation characteristics of helical tomo-
With the data from the simulations done so far, a therapy,” Med. Phys. 31, 396–404 共2004兲.
13
W. Fu, J. Dai, Y. Hu, D. Han, and Y. Song, “Delivery time comparison for
treatment-planning system 共TPS兲 will be commissioned for
intensity-modulated radiation therapy with/without flattening filter: a
an FFF accelerator. The accuracy of the TPS standard dose planning study,” Phys. Med. Biol. 49, 1535–1547 共2004兲.
calculation algorithms will be investigated and, if necessary, 14
D. W. O. Rogers, C. M. Ma, B. Walters, G. X. Ding, D. Sheikh-Bagheri,
modifications will be made to ensure reliable results. To a and G. Zhang, “BEAMnrc user manual,” NRCC Report No. PIRS-
large extend, the investigation will be based on the Monte 0509共A兲 revG.
15
H. G. Hughes, H. W. Egdorf, F. Gallmeier, R. C. Little, R. E. Prael, E.
Carlo model of the accelerator. After the TPS is commis-
Snow, L. S. Waters, and M. White, “MCNPX User’s Manual,” TPO-E83-
sioned and fully tested, treatment plans will be developed for G-UG-X-00001, Rev. 0 共1999兲.
both the conventional and the FFF accelerators. A compara- 16
S. Cho, O. Vassiliev, S. Lee, H. Liu, G. Ibbott, and R. Mohan, “Reference
tive analysis of the plans will be made in terms of the quality photon dosimetry data and reference phase space data for the 6 MV pho-
of deliverable treatment, including an analysis of dose distri- ton beam from Varian Clinac 2100 series linear accelerators,” Med. Phys.
32, 137-148 共2005兲.
butions, estimation of risk for late effects and comparisons of 17
F. M. Khan, The Physics of Radiation Therapy, 3rd ed. 共Lippingcott Wil-
treatment delivery times. Further plans also include a similar liams and Wilkins, New York, 2003兲, p. 179.
18
study for the 18-MV beam. S. Johnson, Varian Medical Systems Inc., Palo Alto, California, USA
共private communication兲.
19
ACKNOWLEDGMENTS M. Stovall, C. R. Blackwell, J. Cundiff, D. H. Novak, J. R. Palta, L. K.
Wagner, E. W. Webster, and R. J. Shalek, “Fetal dose from radiotherapy
This work was supported in part with a research grant with photon beams: Report of AAPM Radiation Therapy Committee Task
from Varian Medical Systems. The authors wish to thank the Group No. 36,” Med. Phys. 22, 63–82 共1995兲.

Medical Physics, Vol. 33, No. 6, June 2006

You might also like