Professional Documents
Culture Documents
slit beams
J. P. Balog, T. R. Mackie,a) P. Reckwerdt, M. Glass, and L. Angelos
Department of Medical Physics, 1530 MSC, 1300 University Avenue, Madison, Wisconsin 53706
~Received 21 November 1997; accepted for publication 27 October 1998!
The UW tomotherapy workbench utilizes a convolution/superposition based dose calculation and
optimization program. It specifies the energy fluence that must be delivered from each leaf for each
phantom projection angle. This requires that the spectrum of the radiation emitted from the one-
dimensional MLC ~multileaf collimator! attached to the linear accelerator be determined. The steps
involved in that process are described. The spectrum along the central axis of the slit beam was
determined, as well as the softening with off-axis position. Moreover, the magnitude of the energy-
fluence output had to be quantified on a per MU ~monitor unit! basis. This was done for a single leaf
along the central axis of the beam. Factors, which modify that energy-fluence output, were inves-
tigated. The output increases with off-axis position due to the horns of the beam. The output for a
leaf of interest will also increase if additional leaves are open due to the absence of the tongue-
and-groove effect and penumbra blurring. The energy-fluence increase per leaf increases by 4.9% if
an adjacent leaf is open. No other factors related to the state of additional leaves were found to
significantly increase the energy-fluence output for an individual leaf. © 1999 American Associa-
tion of Physicists in Medicine. @S0094-2405~99!00301-6#
55 Med. Phys. 26 „1…, January 1999 0094-2405/99/26„1…/55/10/$15.00 © 1999 Am. Assoc. Phys. Med. 55
56 Balog et al.: Characterization of the output for slit beams 56
FIG. 3. A top view of six leaves of an MLC are shown. Only leaf ‘‘e’’ is
closed. The scattered photon on the left is able to increase the output under
It was hypothesized that opening more leaves beyond the leaf ‘‘c’’ when the two leaves labeled ‘‘a’’ and ‘‘b’’ are open along with
two adjacent leaves next to a leaf of interest ~LOI! might also leaf c. Had leaf ‘‘e’’ been open also, the energy fluence would have been
increase the energy-fluence output by the LOI. The magni- increased even more. Extra-focal radiation originating from the right-hand
tude of this effect as a function of open leaves had to be side of the collimator needs leaves ‘‘c,’’ ‘‘d,’’ and ‘‘e’’ to be open simul-
taneously for it to contribute under leaf ‘‘c.’’
determined. It was also theorized that if there were a closed
leaf between a LOI and a group of open leaves, that the
closed leaf would act as a septa and block most of the extra
energy fluence that would otherwise result from the group of II. METHODS
open leaves. This concept is illustrated in Fig. 3. A. Energy-fluence spectrum characterization
The initial energy spectrum was determined by fitting a
measured percent depth dose curve with the predictions of
E. Energy-fluence dependence on adjacent slices
the convolution method. A PDD curve for a 10 cm by 10 cm
The effect of energy fluence arriving from open leaf com- field at 100 cm SSD was measured by an Exradin air equiva-
binations from off-axis slices also had to be determined. Spe- lent ion chamber ~Exradin Inc. Lisle, IL! in a Solid Water
cifically, it had to be decided if there was extra-focal radia- phantom ~RMI Corporation, Middleton, WI! with no MLC
tion from adjacent slices, and if so, how was it affected by present. A broad field was initially used because the mea-
the number of adjacent slices treated. The number of adja- surement is much simpler with a larger field. A Monte Carlo
cent slices treated determines the overall longitudinal length composed spectrum from Mohan23 was first assumed. The
of the field. Traditionally, the larger the field length, the Mohan spectrum was projected, attenuated, and convolved in
greater the extra-focal radiation. a large virtual water phantom to yield a percent depth dose
It was hypothesized that there would not be any extra- distribution. The calculated PDD was compared to the mea-
focal radiation as a function of longitudinal field length. This surement and the spectrum was modified until agreement
is because the MIMiC always defines the field width along was achieved.
the longitudinal direction at any instance and can never vary Next, a measure of the spectrum and magnitude variation
for a fixed slice width. If there were extra slices treated with off-axis position was obtained for the slit beam. A 1 cm
above and below the slice of interest, the energy fluence deep (d max for the 4 MV beam! dose profile obtained when
from them would be the same nature as the energy fluence the MIMiC was attached and irradiated through open leaves
defined from the slice of interest. The net effect, should just was measured with a Scanditronix diode ~model FP1104,
be a superposition of off-slice energy-fluence characteristics. Scanditronix Corp., Sweden! in a water tank. This profile
The convolution/superposition-based treatment-planning al- was used to predict the initial estimate of the ‘‘horns’’ of the
gorithm would inherently account for this effect of phantom beam. The increase in energy fluence with off-axis position
scatter on dose output. at this shallow depth was assumed to equal the increase in
measured dose with off-axis position. An initial off-axis D. Energy-fluence dependence on additional open
beam softening was assumed as well from previous models. leaves
The assumed energy fluence was projected, and a dose dis- The extra energy-fluence increase for an individual leaf
tribution was calculated. The magnitude of the horns and the due to extra-focal radiation passing through additional open
degree of beam softening were modified in an iterative fash- adjacent leaves was measured with the Scanditronix diode
ion until the calculated dose matched both the shallow dose that was in a solid water mini-phantom. The diode was at a
profile and a 30 cm deep ~same measurement conditions as 1 depth of 1 cm and the phantom was 1 cm wide along the
cm! dose profile acceptably well. fan-beam direction. The diode and phantom were precisely
Finally, the full-width-half-maximum distance of a pre- centered in the beam by measuring the collected charge vs
sumed Gaussian focal spot was varied so that the measured phantom actuator lateral position. ~The phantom actuator
dose penumbra matched the calculated dose penumbra. was able to position all phantom/detector experiments to
within 0.1 mm.! The SSD was 93 cm ~the workbench iso-
B. Energy-fluence output magnitude per single open center!. The increase in detector response as a function of
leaf open adjacent leaves was recorded.
The extra-focal radiation was measured by an additional
The magnitude of the energy-fluence output had to be method. The Exradin chamber was placed in a solid water
quantified. It was done so on a per leaf basis. The dose at a phantom ~20 cm by 20 wide and 5 cm deep! at a depth of 1
depth of 1 cm in a 10 cm by 10 cm field was determined by cm. The long axis of the chamber was set vertical ~Fig. 1!.
a TG21 calibration24 ~performed at a depth of 5 cm, con- The chamber was placed outside the width of the MIMiC
verted to 1 cm by PDD data!. The dose per monitor unit leaves. The accelerator was turned on. Once the accelerator
~MU! for an individually opened central leaf was determined reached its peak output rate ~187 MU/min! a central leaf on
relative to the 10 cm by 10 cm field by film and diode do- the MIMiC was opened. The phantom actuator then moved
simetry. The convolution code predicted the dose per energy the chamber past the open leaf and past the other side of the
fluence to the same point under identical geometrical condi- MIMiC leaves. ~The direction of travel was the fan-beam
tions ~one open leaf along the central axis at d max). The en- direction.! The accelerator was then turned off. The transla-
ergy fluence emitted per MU for this leaf is, therefore, equal tion time was 80 s. The charge collected by the ion chamber
to the measured dose per MU divided by the calculated dose was recorded. The collected charge represents the integral of
per energy fluence. the dose profile. The profile integral charge ~PIC! produced
Due to the horns, the energy-fluence output is greater for by the single open leaf is defined as
off-axis positions. The actual energy-fluence per MU ratio
for an arbitrary, single open leaf is the product of the above-
mentioned calculation multiplied by the relative increase in
PIC5 E ~ DQ/DX ! * dx, ~3!
energy fluence as a function of the leaf’s off-axis position:
where DQ/DX is the charge collected per measurement
~ Energy Fluence/MU! leaf i 5 ~ Energy Fluence/MU! point and dx is the ion chamber translation along the fan-
beam direction.
* Off Axis Factorleaf i . ~1! This measurement was relatively insensitive to slight mis-
alignments between the long axis of the MIMiC and the
C. The tongue-and-groove effect direction of the ion chamber translation. The long axis of the
The absence of the tongue-and-groove effect was mea- ion chamber is 2.1 cm long. The collected charge should not
sured by irradiating XV film through two simultaneously be a strong function of slight vertical positioning errors since
opened adjacent leaves. Twenty MU were applied. The film the ion chamber is wider than the slit beam.
was at an SSD of 100 cm and a depth of 1 cm. Since these This was repeated for various combinations of open and
two leaves were opened together, there was not a common closed leaves. An illustration of a PIC measurement is de-
region of overlap between them that blocked the radiation. picted in Fig. 4. This measurement strategy produced an in-
Forty MU were then delivered to another film when the dependent measurement of the T&G effect as previously
same two leaves were opened sequentially, twenty MU for measured by film. It should also be more accurate whereas
each open leaf. These leaves would exhibit a T&G effect, the film dose produced an informative image.
which would manifest itself in the dose distribution. A PIC was obtained with all leaves closed so that the
The resultant profiles were scanned with a laser densito- charge resulting from leakage through closed leaves could be
meter with a 50 micron spot size, Model SI Personal Dosim- subtracted. Therefore, the net PIC represents a measure of
eter ~Molecular Dynamics, Sunnyvale, CA!. The scanned op- the total energy fluence exiting from open leaves of the
tical density values were converted to dose by use of optical MIMiC.
density/dose values obtained from irradiated calibration
films. Integration of the dose profiles across the leaf fan- E. Energy-fluence dependence on adjacent slices
beam direction ~Fig. 1! yields the profile integral dose ~PID!.
The hypothesis that additional slices would not contribute
PID5 E D ~ x ! dx. ~2!
extra-focal radiation was investigated. The resultant dose
profile from a 5 cm vertical translation of a phantom across
FIG. 8. The measured vs the predicted dose profile for sets of two leaves
open at a depth of 1 cm. The measured profile was obtained with an ion
FIG. 6. The agreement between the measured and predicted PDD curves for
chamber in a water tank. The profile distance is in the fan-beam direction.
a 10 cm by 10 cm field at an SSD of 100 cm. The depth is measured in the
beam central-axis direction.
FIG. 10. The measured 5 cm vertical translation profile vs the predicted 5 cm FIG. 12. A comparison of the predicted profiles resulting from a 5 cm ver-
vertical translation profile for all leaves open. tical translation and a 3 cm vertical translation for the same translation speed
as the measured 5 cm translation profile, 16 s/cm.
IV. DISCUSSION
The energy-fluence modeling from a one-dimensional
multileaf collimator is an involved, but an entirely manage-
able problem. The methodology is straightforward and logi-
FIG. 11. The charge collected across all open leaves for 249 MU. In one
cal. The energy spectrum is characterized along the entire
case, the ion chamber was translated 5 cm vertically during irradiation, and slit, and accounts for off-axis softening. Additional dose pro-
in the other it was stationary ~and moved vertically between irradiations!. files were calculated and measured for various leaf patterns
2
made as small as possible. Leakage has been incorporated T. Mackie, T. Holmes, S. Swerdloff, P. Reckwerdt, J. Deasy, J. Yang, B.
into the treatment-planning algorithm. Paliwal, and T. Kinsella, ‘‘Tomotherapy: A new concept for the delivery
of conformal radiotherapy,’’ Med. Phys. 20, 1709–1719 ~1993!.
The difference between the two-leaves-open and all- 3
T. R. Mackie, T. W. Holmes, P. J. Reckwerdt, and J. Yang, ‘‘Tomo-
leaves-open profile ~Fig. 13! is due almost exclusively to therapy: optimized planning and delivery of radiation therapy,’’ Int. J.
phantom scatter ~as discussed there is very little extra-focal Imaging Syst. Technol. 6, 43–55 ~1995!.
radiation!, which cannot be modified except by beam energy. 4
F. Khan, The Physics of Radiation Therapy ~Williams and Wilkins, Bal-
Presently, the specification for leakage through the pri- timore, MD, 1994!, pp. 201–202.
5
mary collimator and both pairs of jaws for existing linear- T. R. Mackie, H. H. Liu, and E. C. McCullough, ‘‘Calculating output
accelerator collimators is 0.1% of the useful beam. The to- factors for photon beam radiotherapy using a convolution/superposition
method based on a dual source photon beam model,’’ Med. Phys. 24,
motherapy jaws under development will try to reduce this
1975–85 ~1997!.
amount by a factor of 10 using a thicker primary collimator. 6
J. Balog, Ph.D. thesis, University of Wisconsin at Madison ~1998!.
One last point about phantom scatter and leakage integra- 7
J. Yang and T. R. Mackie, ‘‘An investigation of tomotherapy beam de-
tion should be made. The relative dose outside the primary livery,’’ Med. Phys. 24, 425–436 ~1997!.
energy fluence, compared to the dose inside, is more than a 8
A. Brahme, ‘‘Optimization of stationary and moving beam radiation
simple scalar multiple of the number of slices treated times therapy techniques,’’ Radiother. Oncol. 12, 129–140 ~1998!.
9
the static dose outside the primary energy fluence. This is N. H. Barth, ‘‘An inverse problem in radiation therapy,’’ Int. J. Radiat.
because the peak dose for a translation profile results prima- Oncol., Biol., Phys. 18, 425–431 ~1990!.
10
M. Goitein, ‘‘The inverse problem,’’ Int. J. Radiat. Oncol., Biol., Phys.
rily from the average dose inside the primary energy fluence.
18, 489–491 ~1990!.
This is less than the peak dose produced from a static open 11
S. Webb, ‘‘Optimization of conformal radiotherapy dose distributions by
leaf. The smaller the slit beam is, the less is the average dose simulated annealing: 2. Inclusion of scatter in the 2D technique,’’ Phys.
inside the area irradiated by the slit beam, thereby reducing Med. Biol. 36„9…, 1227–1237 ~1991!.
efficiency for moving slit-beam delivery. 12
S. M. Morrill, R. G. Lane, G. Jacobson, and I. I. Rosen, ‘‘Treatment
planning optimization using constrained simulated annealing,’’ Phys.
Med. Biol. 36„10…, 1341–1361 ~1991!.
V. CONCLUSION 13
M. P. Carol, ‘‘Peacock™: A system for planning and rotational delivery
The energy-fluence output from a 1D MLC attached to a of intensity-modulated fields,’’ Int. J. Imaging Syst. Technol. 6, 56–61
linear accelerator can be modeled. The spectrum was deter- ~1995!.
14
mined along the central axis. The decrease in average J. D. Bourland and E. L. Chaney, ‘‘A finite-size pencil beam model for
photon dose calculations in three dimensions,’’ Med. Phys. 19, 1401–
energy-fluence energy with off-axis position was also deter-
1412 ~1992!.
mined. The energy-fluence output was quantified per single 15
A. R. Bleier, M. P. Carol, B. H. Curran, A. A. Kania, J. D. Scherch, and
open leaf per MU for leaves aside the central axis. Due to the E. S. Strenick, ‘‘Dose calculation in Peacock plan,’’ NOMOS Corp. pub-
horns of the beam, the energy-fluence output per MU in- lication, NOMOS Corp., Sewickly, PA 15143 ~Sept. 1995!.
16
creases for off-axis locations. The energy-fluence output per S. Webb and M. Oldham, ‘‘A method to study the characteristics of 3D
MU for an LOI will also increase if adjacent leaves are open dose distributions created by superposition of many intensity-modulated
due to the absence of the T&G effect. It was determined that beams delivered via a slit aperture with multiple absorbing vanes,’’ Phys.
Med. Biol. 41, 2135–2153 ~1996!.
there is not significant extra energy fluence if additional 17
M. Oldham and S. Webb, ‘‘Intensity-modulated radiotherapy by means of
leaves are open. As demonstrated from the insensitivity of static tomotherapy: a planning and verification study,’’ Med. Phys. 24,
the calculated penumbra on source size, and the lack of 827–36 ~1997!.
extra-focal radiation, the NOMOS MIMiC provides excellent 18
T. Holmes, Ph.D. thesis, University of Wisconsin at Madison ~1993!.
beam collimation. There is negligible extra-focal radiation 19
T. Holmes and T. R. Mackie, ‘‘A filtered backprojection dose calculation
from off-axis slices. This simplifies the planning process. method for inverse treatment planning,’’ Med. Phys. 21, 303–313 ~1994!.
20
The multiplicative nature of the effects of leakage and phan- T. W. Holmes, T. R. Mackie, and P. Reckwerdt, ‘‘An iterative filtered
tom scatter when many slices are treated was also demon- backprojection inverse treatment planning algorithm form tomotherapy,’’
Int. J. Radiat. Oncol., Biol., Phys. 32, 1215–1225 ~1995!.
strated. Leakage is the limiting factor in the ability of con- 21
T. R. Mackie, P. Reckwerdt, and N. Papanikolaou, ‘‘3-D photon beam
ventional linacs with primary collimators to deliver low dose dose algorithm,’’ in Proceedings from the 1996 AAPM Summer School
values when many slices are treated. Leakage from adjacent ~Medical Physics Publishing, Madison, WI, 1996!, pp. 201–222.
slices should be incorporated into any intensity-modulated 22
P. J. Reckwerdt, T. R. Mackie, J. Balog, and T. R. Mcnutt, ‘‘Three di-
planning system. mensional inverse treatment optimization for tomotherapy,’’ in Proceed-
ings from the XII International Conference on the Use of Computers in
Radiation Therapy ~Medical Physics Publishing, Madison, WI, 1997!, pp.
ACKNOWLEDGMENTS 420–422.
23
This work was funded by a contract with General Electric R. Mohan, C. Chui, and L. Lidofsky, ‘‘Energy and angular distributions
of photons from medical linear accelerators,’’ Med. Phys. 12, 592–597
Medical Systems and NIH Grant No. CA489.
~1985!.
24
a! Task Group 21, Radiation Therapy Committee, American Association of
Electronic mail: mackie@macc.wisc.edu
1 Physicists in Medicine, A Protocol for the determination of absorbed dose
M. Carol, W. H. Grant, III, D. Pavord, P. Eddy, H. S. Targovnik, B.
Butler, S. Woo, J. Figura, V. Onufrey, R. Grossman, and R. Selkar, ‘‘Ini- from high energy photon and electron beams,’’ Med. Phys. 10, 741–769
tial clinical experience with the Peacock intensity modulation of a 3-D ~1983!.
25
conformal radiation therapy system,’’ Stereotactic & Functional Neuro- J. P. Balog, T. R. Mackie, and D. Wenman, ‘‘Multi-leaf collimator inter
surgery 66„1–3…, 30–4 ~1996!. leaf transmission,’’ Med. Phys. ~submitted!.