You are on page 1of 11

Int. 1. Radiation Oncology Biol. Phys., Vol. 32, No. 4, pp.

1215- 1225, 1995


Copyright 0 1995 Elsevier Science Ltd
Pergamon Printed in the USA. All rights reserved
0360-3016/95 $9.50 + 00

0360-3016(94)00465-X

l Technical Innovations and Notes

AN ITERATIVE FILTERED BACKPRO JECTION INVERSE TREATMENT


PLANNING ALGORITHM FOR TOMOTHERAPY

TIMOTHY W. HOLMES, PH.D., T. ROCK MACRIE, PH.D. AND PAUL RECKWERDT, B.S.
Departmentsof Medical Physicsand HumanOncology, University of WisconsinSchool of Medicine,
1300University Avenue, Madison, WI

Purpose: An inverse treatment planning algorithm for tomotherapy is described.


Methods and Materials: The algorithm iteratively computes a set of nonnegative beam intensity profiles
that minimizes the least-squares residual dose defined in the target and selected normal tissue regions of
interest. At each iteration the residual dose distribution is transformed into a set of residual beam profiles
using an inversion method derived from filtered backprojection image reconstruction theory. These “resid-
ual” profiles are used to correct the current beam profile estimates resulting in new profile estimates.
Adaptive filtering is incorporated into the inversion model so that the gross structure of the dose distribution
is optimized during initial iterations of the algorithm, and the fine structure corresponding to edges is
obtained at later iterations. A three dimensional, kernel based, convolution/superposition dose model is
used to compute dose during each iteration.
Results: Two clinically relevant treatment planning examples are presented illustrating the use of the
algorithm for planning conformal radiotherapy of the breast and the prostate. Solutions are generally
achieved in lo-20 iterations requiring about 20 h of CPU time using a midrange workstation. The majority
of the calculation time is spent on the three-dimensional dose calculation.
Conclusions: The inverse treatment planning algorithm is a useful research tool for exploring the potential
of tomotherapy for conformal radiotherapy. Further work is needed to (a) achieve clinically acceptable
computation times; (b) verify the algorithm using multileaf collimator technology; and (c) extend the method
to biological objectives.

Radiation therapy, Tomotherapy, Treatment planning, Optimization, Image reconstruction, Filtered back-
projection.

INTRODUCTION a configuration of radiation sources that achieves a high


degree of tumor control while simultaneously attempting
The introduction of computers into the field of radiation to minimize the amount of normal tissue damage. This
oncology, about 30 years ago, carried the promise of im- process involves the application of biological response
proved treatment techniques through pretreatment model- data by an oncologist, either implicitly in the form of a
ing and optimization of radiation dose distributions and prescribed dose distribution, or explicitly in the form of
computer controlled treatment delivery. During this time dose responsemodels used to compute bioeffect probabil-
much of the focus has been on the development of more ities for comparing rival treatment plans. In the first in-
accurate and efficient dose calculation algorithms. Until stance, the dose prescription represents a surrogate for
recently, comparatively less emphasishas been placed on biological response. The distinction leads to two optimi-
1) providing optimization algorithms that use these dose zation approaches.Physical optimization methodsassume
calculation models, and 2) treatment delivery technology that the prescribed dose distribution is the biologically
that conveniently allows for clinical realization of opti- optimal dose distribution. The prescription is nearly al-
mized plans. ways physically unrealizable because zero dose is typi-
Radiotherapy optimization is the process of selecting cally assigned to normal tissues. These methods attempt

Reprint requeststo: Timothy W. Holmes, Ph.D., Southern 0. Deasyfor many stimulatingdiscussionson radiotherapyopti-
WisconsinRadiotherapyCenter, 1102John Nolen Drive, Madi- mization. This work wassupported,in part, by NC1grant num-
son, Wisconsin53713. ber CA48902. T. Holmeswas supported,in part, by National
Acknowledgments- We acknowledgeR. H. Huesman,G. T. ResearchService Award CA09206 and a grant from the Wis-
Gullberg, W. L. Greenberg,and T. F. Budinger, the authorsof consinFoundation.
the “Donner Algorithms for ReconstructionTomography,” for Accepted for publication 19 August 1994.
the use of this software (15). We would like to thank Joseph
1215
1216 I. J. Radiation Oncology l Biology 0 Physics Volume 32. Number 4, 1995

to automate conventional treatment planning normally A. Tomotherapy


performed by a dosimetrist. Biological optimization meth-
ods determine the dose distribution that has the best bio-
logical “figure-of-merit” without any a priori bias regard-
ing what the optimal dose distribution should look like
(18). The underlying biological mechanisms of radiation
damage and repair must be properly modeled. At this
time there is still great uncertainty in the actual form of
the biological response model, owing to an insufficient
amount of reliable clinical response data (7). Conversely,
the physical processes of particle interactions in matter
are well understood and can be accurately modeled, yet
the prescribed dose distribution may not correspond to
the optimal biological effect.
Conventional treatment planning attempts to arrive at
the best approximation of the desired dose distribution
(or the best biological figure-of-merit) by determining the
set of beam profiles and orientations through trial and
error. Historically, this process has been carried out manu-
ally by a dosimetrist. The limitations of equipment, time,
and personnel required to fabricate custom blocks and
compensators and execute treatment delivery has re-
stricted the maximum number of beam portals to be clini-
cally practical (< 4-6). Until recently, the technology of
automatically creating and delivering a large number of B. Tomotherapy Unit
modulated beams was not available to most clinics. This
probably explains why automated optimization methods
have never found popular application even though a num-
ber of approaches were described from the mid-1960s to
the mid-1980s (1, 13, 14, 20, 26, 27, 29, 32, 33).
A renewed interest in automated optimization methods
has occurred since the late 1980s due in part, no doubt, to
the commercial availability of computer-controlled linear
accelerators and multileaf collimators. This new technol-
ogy may make it possible to conveniently treat a configu-
ration of a large number of beams that realize the promise
of conformal radiotherapy techniques initially developed
over 20-30 years ago by Green and colleagues in England
(6, 16), and Takahashi in Japan (34). Several efforts are
underway to achieve conformational dose distributions
using computer-controlled collimation systems. Several
investigators have proposed methods based on modulat-
ing the profile of beam intensity by subdividing the treat-
ment volume into subfields that are differentially modu-
lated, using either asymmetric collimators or two-dimen-
sional (2D) multileaf collimators (5, 19, 25, 37). These Fig. 1. (A) Tomotherapy is carried out by rotating the x-ray
beamwhile the patientis translatedthoughthe planeof irradia-
methods should allow the physical realization of modu- tion. (B) Schematicof the tomotherapyunit proposedby Ma&e
lated beam profiles computed by any of a number of et al. (23).
inverse radiation treatment planning algorithms (2, 4, 12,
18, 21, 30, 35, 36) including the algorithm described in
this work. through the gantry bore (Fig. la). A schematic drawing
We have proposed an integrated approach to treatment showing the main components of a hypothetical tomother-
planning, delivery, and verification of conformal radio- apy unit is given in Fig. lb. The on-board CT scanner is
therapy, called tomotherupy (23). Tomotherapy is analo- used to simultaneously image the anatomy during treat-
gous to spiral computed tomography (CT) scanning where ment. Image detectors are used to detect transit profiles
a megavoltage x-ray fan beam is continuously rotated on of the megavoltage beam and the data used along with the
a ring gantry while the patient is simultaneously translated CT imanes of anatomv, to reconstruct dose distributions as
Tomotherapy l T. W. HOLMES et al. I225

pub-214, University of California, Berkeley. CA; October 26. Mantel, J.; Perry, H. Automatic variation of field size and
1977. dose rate in rotation therapy. Int. J. Radiat. Oncol. Biol.
16. Jennings, W. A. The tracking cobalt project: From moving- Phys. 2:307-317; 1977.
beam therapy to three-dimensional programmed irradiation. 27. McDonald, S. C.; Rubin, P. Optimization of external beam
In: Orton, C. G., ed., Progress in medical radiation physics, radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2:307-
chap. 1-I. New York: Plenum Press; 1985:144. 317; 1977.
17. Kak, A. C.; Slaney, M. Principles of computerized tomo- 28. Reckwerdt, P. Superposition/convolution speed improve-
graphic imaging. New York: IEEE Press; 1988. ments using run-length raytracing. (Abstr.) Med. Phys.
18. Kallman, P., Lind, B. K.; Brahme, A. An algorithm for 19:784; 1992.
maximizing the probability of complication-free tumor con- 29. Redpath, A. T.; Vickery, B. L.; Wright, D. H. A new tech-
trol in radiation therapy. Phys. Med. Biol. 37:871-890; nique for radiotherapy planning using quadratic program-
1992. ming. Phys. Med. Biol. 21:781-791; 1976.
19. Lane, R. G.; Loyd, M. D.; Chow, C. H.; Ekwelundu, E.; 30. Rosen, I. I.; Lane, R. G.; Morrill, S. M.; Belli, J. A. Treat-
Rosen, I. I. Custom beam profiles in computer-controlled ment plan optimization using linear programming. Med.
radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 22: 167- Phys. 18:141-152; 1991.
174; 1991. 31. Rosenberger, F. U.; Mathews, J. W.; Johns, G. C.; Drzy-
mala, R. E.; Purdy, J. A. Use of transputers for real time
20. Langer M.; Leong, J. Optimization of beam weights under
dose calculation and presentation for three-dimensional ra-
dose-volume restrictions. Int. J. Radiat. Oncol. Biol. Phys.,
13:1255-1260; 1987. diation treatment planning. Int. J. Radiat. Oncol. Biol. Phys.
25:709-719; 1993.
21. Lind, B. Properties of an algorithm for solving the inverse
32. Sonderman, D.; Abrahamson, P. G. Radiotherapy design
problem in radiation therapy. Inverse Problems 6:415 -426; using mathematical programming models. Op. Res.
1990. 33:705-725; 1985.
22. Luenberger, D. G. Linear and nonlinear programming. 2nd 33. Starkshall, G. A constrained least-squares optimization
ed. Reading, MA: Addison-Wesley; 1989. method for external beam radiation therapy treatment plan-
23. Mackie, T. R.; Holmes, T. W.; Swerdloff, S.; Reckwerdt, ning. Med. Phys. 11:659-665; 1984.
P.; Deasy, J. 0.; Yang, J.; Paliwal, B.; Kinsella, T. Tomo- 34. Takahashi, S. Conformation radiotherapy: Rotation tech-
therapy: A new concept for the delivery of conformal radio- niques as applied to radiography and radiotherapy of can-
therapy using dynamic compensation. Med. Phys. 20: 1709- cer. Acta Radiol. Suppl. 242:1-142; 1965.
1719; 1993. 35. Webb, S. Optimization of conformal radiotherapy dose dis-
24. Mackie, T. R.; Reckwerdt, P. J.; Gehring, M. A.; Holmes, tributions by simulated annealing. Phys. Med. Biol.
T. W.; Kubsad, S. S.; Thomadsen, B. R.; Sanders, C. A.; 34:1349-1370; 1989.
Paliwal, B. R.; Kinsella, T. J. Clinical implementation of 36. Webb, S. Optimization by simulated annealing of three-
the convolution/superposition method. In: Proceedings of dimensional, conformal treatment planning for radiation
the 10th International Conference on Computers in Radio- fields defined by a multileaf collimator: II. Inclusion of
therapy, Lucknow, India. Lucknow, India: Alpana Arts; two-dimensional modulation of the x-ray intensity. Phys.
1990:322-325. Med. Biol. 37: 1689- 1704; 1992.
25. Mageras, G. S.; Podmaniczky, K. C.; Mohan. R. A model 37. Zacarias, A. S.; Lane, R. G.; Rosen, I. I. Assessment of a
for computer-controlled delivery of 3-D conformal treat- linear accelerator for segmented conformal radiation ther-
ments. Med. Phys. 19:945-953; 1992. apy. Med. Phys. 20:193-198; 1993.
1218 I. J. Radiation Oncology 0 Biology l Physics Volume 32. Number 4, 1995

WA(f)
- lIA(9 ---
R(f) ....-

---I

-0.5 -0.4 -0.3 -0.2 -0.1 0 0.1 0.2 0.3 0.4 0.5
SpatialFrequency

-0.5 -0.4 0.3 -0.2 -0.1 0 0.1 0.2 0.3 0.4 0.5 -0.5 -0.4 -0.3 -0.2 -0.1 0 0.1 0.2 0.3 0.4 0.5
SpetlalFrequenq SpatialFrequency

Fig. 2. Frequency spectra of the quantity 1f 1/A(f) for: (A) boCo; (B) 6 MV; (C) 10 MV; and (D) 24 MV x-rays,
Also shown are the spectra for the frequency ramp 1f 1 and the inverse of the filter A(f).

where If 1 is a frequency ramp, A(f) is the frequency The Butterworth filter is used in the optimization
space representation of the 1D dose kernel, and algorithm to control both the maximum spatial resolu-
(,/-)-I is the Butterworth lowpass filter where tion of the beam profiles and indirectly the maximum
$ is the filler cutoff frequency and the exponent 2M con- spatial resolution of the optimized dose distribution
trols the filter shape (8). The quantity 1f (/A(f) is plotted during a given iteration (Fig. 3). The cutoff frequency
in Fig. 2 for @‘Cogamma rays and 6 MV, 10 MV, and f0 is adaptively increased from a small starting value
24 MV x-rays. The frequency spectra of (f I and A(f)-’ of 0.0625 up to a maximum value less than or equal to
are included for comparison. Note that to first order the the Nyquist frequency of 0.5. In general, it is desirable
shape of the fast inversion filter resembles a frequency to limit the amount of oscillations in beam profiles
ramp with slope between 2-4 for the range of energies because smooth profiles are easier to realize clinically.
used clinically. The factor (p/p)-‘exp( + PX) provides ap- The philosophy adopted here is to optimize the gross
proximate energy conservation. To simplify matters it is structure of the dose distribution during early iterations
assumed that the patient is water equivalent; therefore and the penumbra at later iterations by slowly increas-
p/p and p are defined for water and f is the average depth ing the frequency cutoff when no progress is being
to the center of the tumor for all beams intersecting the made as measured by the relative change in the objec-
tumor. Typically, the scale factor is in the range 20-40 tive function:
for pwlp= 0.05 cm*/g and an average attenuation factor
of 1.o-0.5. (Eq. 9)
Tomotherapy 0 T. W. HOLMES et al. 1219

A.FastlnwtionFilter B.BulterworthFilter
-I-
1.8 &
1.8 I
1.4 I
!
1.2
I
1
0.8 !
I
0.8
I
0.4
0.2
0
-0.5 Xl.4 0.3 4.2 -0.1 0 0.1 0.2 0.3 0.4 0.5 -0.4 -0.2 0 0.2 0.4
SpatialFrequency Saptbl
Frequency

C.BandlirnitedFastlnrersionFilter D.AdaptiwBandlimiting
! I""'""""'1

h 0.45
e
0.8 ; 0.4
i 0.35
0.8 k9 0.3
< 0.25
t 0.2
5 0.15
z 0.1
0.05
0
-0.4 -0.2 0 0.2 0.4 0 1 2 3 4 5 8 7 8 Q 10 11 12 13 14 15
Saptial
Frequency No.hllons

Fig. 3. Illustration of adaptive bandlimiting of the fast inversion filter. (A) The fast inversion filters. (B) The
Butterworth filters corresponding to M = 4 and f,, given by Eq. 10. (C) The product of the 6 MV fast inversion
filter and the Buttetworth filters for each value of f,,. (D) Plots of f0 and f(d) as a function of iteration number
for a typical treatment planing problem.

The cutoff frequency is updated when the relative amplified by the inversion process because the filtering
change in the objective function falls below an empiri- operation is a method of deconvolution. These errors are
cally determined threshold of 20% unless, of course, the largely due to inaccurate representation of continuous
maximum cutoff frequency has been reached, or if the edges of regions-of-interest by discrete pixels and show
solution is diverging as measured by A < 0. The initial up as high frequency oscillations in the beam profile (Fig.
value of 0.0625 corresponds to averaging over eight pen- 4a). One solution is to use antialiased edge representation
cil beams. Subsequent cutoff values are obtained by re- at the existing grid resolution. The improvement possible
ducing the number of pencil beams averaged over (n) by is shown in Fig. 4b. Better results are also possible using
one: a higher calculation grid resolution. The major drawback
of this solution is that the computation time is signifi-
n= (8,7,6,5,4,3,2,1). (Eq.lO) cantly lengthened if applied to all steps in the algorithm.
f0=$
We compromise by computing the dose distribution at a
0.5 cm resolution, but perform the inversion at a higher
Small errors in the dose projections will be significantly resolution of 0.125 cm by resampling the dose distribution
I. J. Radiation Oncology 0 Biology l Physics Volume 32, Number 4, 1995

ues are nonphysical. The profiles can be optionally modi-


!
INITIAL- fied to account for blocking of normal tissue and critical
/ . A. I
! ESTIMATE+-- structures. The beam profiles are subsequently used to
/ compute dose using a 3D superposition calculation (24,
! 28). The current dose estimate is then compared against
!
upper and lower dose constraints defined for the target
region and any regions-of-interest specified in the normal
tissue. The iteration is terminated if a feasible solution is
.-.
reached, wherein the current dose estimate falls between
F the upper and lower dose limits for all regions-of-interest.
If a feasible solution is not achieved, then the algorithm
proceeds to compute the residual dosein the target region-
of-interest and any unblocked normal tissue regions-of-
interest in preparation for updating the beam profiles. The
residual dose is used to compute the objective function,
-20 -15 -10 -5 0 5 10 15 20 which is used to terminate the planning process if it in-
Distance(cm)
creases in successive iterations. If it decreasesthen the
residual dose in the regions of interest is transformed into
residual beam profiles by the inversion method on a slice-
by-slice basis. New beam profile estimates are computed
/ INITIAL-
25 8. ESTIMATE+ by subtracting the residual beam profiles from the current
profile estimatesand a new iteration is begun. Upon termi-
nation the dose calculation can be optionally verified us-
ing an alternative dose calculation model. This is useful

INlllALlZE ENERQY FLUENCE

-20 -15 -10 -5 0 5 10 15 20


Oklance(cm)
Fig. 4. Inversion results for different edge representations of
the external contour. The test consisted of simulating the irradia-
tion of a 30 cm diameter circular uniform phantom by 72 identi-
cal uniform 24 MV beam profiles equally spaced about 360
degrees. (A) Binary edge. (B) Antialiased edge. The calculation BLOCKING CONSTRAINT
grid was 64 x 64 pixels at 0.5 cm resolution.
I

to the higher resolution. The dose residual is computed FORWARD DOSE CALCULAYION c
and extracted from the target and normal tissue using high
resolution region-of-interest masks. The high resolution
projections of the dose residual are obtained and then
resampledto the lower resolution of the multileaf collima-
tor and filtered by I. This modification to the algorithm
is discussedfurther in Holmes and Mackie (11).
The IFB algorithm is implemented as the iterative se-
quence of steps listed in the flowchart shown in Fig. 5.
At the beginning of the iteration the beam profiles are INVERSE DOSE CALCULATION
initialized by inverting the dose prescription. Typically
this is represented as nonzero dose in the target and zero
dose to normal tissues. During each iteration and prior to
computing dose, the beam profiles are verified to seethat
they consist of nonnegative values, becausenegative val- Fig. 5. Flowchart of the iterative optimization algorithm.
Tomotherapy 0 T. W. HOLMES et al. 1221

if approximations are incorporated in the dose calculation Blocking Strategies


used in the iteration to speed up the inverse planning No Blocks

process.

RESULTS
The definition of beam blocking is one of the initial
steps in the treatment planning process and can have a
significant impact on the results obtained using the IFB
algorithm. Figure 6 illustrates the possible choices of
blocking for a C-shaped target and nearby organ at risk.
I 7
I rl
Note that if the organ at risk is not blocked, it is included Critical Structure Blocked
with the target in the determination of the dose residual; Fig. 6. Diagramshowingthe different blocking strategiesfor a
otherwise if the organ at risk is blocked, the residual is C-shapedtarget with nearby squaresensitivestructure.
computed only over the target region. Dose profiles are
compared in Fig. 7 corresponding to the different types
of blocking for two beam energies. The profiles corre-
spond to a horizontal line through the center of the target

Co!imabdbn,Crilk9SbW%mW0d@d
1.2 1.2 1

I
1

0.6

ij O4

0.4

0.2
/
Od be
0
-15 -10 -5 0 5 IO 15
rmlca (all)

Cdimati,CIkdStWumNotBkdied Wmated,NoBloding
1.2 , I 1.2 >
I I I '1

24MV- - 1 24W- .

0.6
5-t
9

8 O4

0.4

0.2

0
-15 -10 -5 5 10 15
WaA (an)

Fig. 7. Dose profiles for different blocking strategies for the phantom shown in Fig. 6. In all cases the target is
included in the determination of the residual. In panels A-C the normal tissue is blocked by collimating the beam
to the target boundaries. In these examples the critical structure is (A) ignored, (B) blocked, and (C) unblocked,
but includedin the determinationof the residual.(D) Whenboth structureswereunblockedthe algorithmperformed
poorly.
1222 I. J. Radiation Oncology l Biology 0 Physics Volume 32. Number 4, 1995

Fig.8. Prostate example. Isodose lines shown are 110, 100, 75, 50, and 25%. (A) Central slice. (B) Off-axis slices
1 cm and 3 cm superior and inferior to the central slice.

and critical structure. These results suggest that it is best prostate, its location deep inside the body, and the rela-
to at least collimate the beam to the outline of the target tively moderate size of adjacent organs at risk (bladder
region to reduce the effect of the beam energy on the and rectum) allow a variety of treatment techniques to be
resulting dose distribution. Furthermore, the algorithm used, from multiple static fields to arcs. In contrast, the
reached an acceptable level of dose uniformity in the superficial location of the breast and the large size of
target when the organ at risk was also blocked (Fig. 7b), adjacent organs at risk (lung, heart, and the contralateral
although with increased dose to the normal tissue as a breast) typically constrain one to a pair of tangential paral-
consequence. lel-opposed fields. For the prostate, the goal of conformal
The IFB inverse treatment planning algorithm was used radiotherapy is generally the improvement of local control
to compute the conformal isodosedistributions for a pros- via dose escalation; whereas for the breast, especially
tate and a breast shown in Figs. 8 and 9, respectively. early stage disease,it is reduced normal tissue complica-
These examples represent two different classesof prob- tions, becauseonly marginal improvement in local control
lems for conformal radiotherapy. The small size of the is felt to be possible using dose escalation.
Tomotherapy 0 T. W. HOLMES et al 1223

Fig. 9. Breast example. Isodose lines shown are 110, 100, 75, 50, and 25%. (A) Central slice. (B) Off-axis slices
2 cm and 4 cm superior and inferior to the central slice.

To simplify the problem, and test feasibility of the pute on a workstation.’ A larger number of iterations was
algorithm, the critical structures were ignored and the obtained in the case of the prostate because it is a smaller
target volume was the only region-of-interest used to structure and requires fewer pencil beams to be optimized
compute the residual dose. The planning objective in both than the breast volume. Due to the long calculation times,
cases was to minimize the least-squared residual dose in the iterations were terminated prior to convergence to a
the target volume to achieve a dose uniformity of 2 5% minimum of the objective function, and consequently
in the target volume. The prostate dose distribution was only a 10% dose uniformity was achieved in the target
achieved in 9 iterations using 64 beam directions equally for both examples. The dose distributions for the central
spaced about 360”. The breast dose distribution was axis and four off-axis planes are provided for each exam-
achieved in 5 iterations using 32 beam directions equally ple. Although + 5% dose uniformity was not achieved,
spaced about 360”. Both examples took over 20 h to com- the dose distributions are nevertheless highly conformal.

’ DEC Station 5000/200, Digital Equipment Corp, Maynard,


MA.
1224 I. J. Radiation Oncology 0 Biology 0 Physics Volume 32. Number 4, 1995

Furthermore, the dosevolume histograms for these exam- ations to achieve a solution for 2D problems, whereas
ples indicate that a sensitive structure receives no more Webb (35) has reported using several million iterations
than about 30-50% of the target dose to one-third of its using simulated annealing to solve similar problems. The
volume. This might be reduced further by including the major limitation of gradient algorithms is that they are
critical structures in the optimization. Further work is lessgeneral than simulated annealing-they are only ap-
needed to compare these results against other treatment propriate for problems where the objective function is
delivery methods using coplanar and noncoplanar beam unimodal; that is, it has only a single minimum value
directions, with and without intensity modulation of the that is the global minimum. We have found that the IFB
beam profiles. algorithm will likely converge to a feasible solution when
the number of pencil beams (e.g., the “degrees-of-free-
dom”) are comparable to the number of dose constraints
DISCUSSION
in the target, in analogy with the general principle used
Long computation times are the major problem facing in image reconstruction that the total number of projection
optimization of large numbers of beams.The overall time ray samples should be roughly equal to the number of
of the IFB algorithm is dominated by the 3D superposition pixels in the circle of reconstruction (17). This is consis-
dose calculation, which is repeated during each iteration tent with requiring that the matrix D be square and its
using the new beam profile estimates. This could be re- determinant nonzero so that its inverse D-’ will exist (22).
duced somewhat by incorporating useful approximations Consequently, if this criteria is met for a given tomother-
that do not compromise accuracy of the dose calculation. apy inverse treatment planning case, then simulated an-
For example, if heterogeneities are not important then nealing may not be required, although further research
approximate dose calculations that are similar to filtered comparing the two algorithms is needed.
backprojection image reconstruction (3, 10) could be
used. Another solution is to use parallel processing com-
CONCLUSION
puter technology to accelerate the dose computation (31)
by assigning the computation of a subset of one or more We have described an inverse treatment planning algo-
beams to a dedicated central processing unit. rithm for tomotherapy based on filtered backprojection
A secondary, but important, issue is the number of image reconstruction. The algorithm is demonstrated us-
iterations needed to achieve a solution. An advantage of ing two clinically relevant treatment planning examples
gradient search algorithms, such as the algorithm de- of the prostate and breast. Future work is necessary to
scribed here and those described in Bortfeld et al. (2) and reduce the computation time, to extend the method to
Brahme (4) over simulated annealing, is the small number biological objectives, and to perform experimental verifi-
of iterations needed to obtain an acceptable solution. For cation, including comparison with other inverse treatment
example, the IFB algorithm typically requires lo-20 iter- planning algorithms.

REFERENCES
1. Bahr, G. K.; Kereiakes, J. G.; Horwitz, H.; Finney, R.; 8. Hamming, R. W. Digital filters. 3rd ed. Englewood Cliffs,
Galvin, J.; Goode, K. The method of linear programming NJ: Prentice Hall; 1989.
applied to radiation treatment planning. Radiology 91:686- 9. Holmes, T. W.; Mackie, T. R. A comparison of three in-
693; 1968. verse treatment planning algorithms. Phys. Med. Biol.
2. Bortfeld, T.; Btirkelbach, J.; Boesecke, R.; Schlegel, W. 39:91-106; 1994.
Methods of image reconstruction from projections applied 10. Holmes, T. W.; Mackie, T. R. A filtered backprojection
to conformation therapy. Phys. Med. Biol. 35:1423-1434; dose calculation method useful for inverse treatment plan-
1990. ning. Med. Phys. 21:303-313; 1994.
3. Bortfeld, T.; Schlegel, W.; Rein, B. Decomposition of pen- 11. Holmes, T. W.; Ma&e, T. R. Suppression of artifacts in
cil beam kernels for fast dose calculations in three-dimen- optimized beam profiles determined by the iterative filtered
sional treatment planning. Med. Phys. 20:31 l-318; 1993. backprojection inverse treatment planning algorithm. (sub-
4. Brahme, A. Optimization of stationary and moving beam mitted to Phys. Med. Biol.); 1993.
radiation therapy techniques. Radiother. Oncol. 12: 129- 12. Holmes, T. W.; Ma&e, T. R.; Simpkin, D.; Reckwerdt, P.
140; 1988. A unified approach to the optimization of brachytherapy
5. Convery, D. J.; Rosenbloom, M. E. The generation of inten- and external beam dosimetry. Int. J. Radiat. Oncol. Biol.
sity-modulated fields for conformal radiotherapy by dy- Phys. 20:859-873; 1991.
namic collimation. Phys. Med. Biol. 37:1359- 1374; 1992. 13. Hope, C. S.; Laurie, J.; Orr, J. S.; Halnan, K. E. Computer
6. Davy, T. J. Physical aspects of conformation therapy using optimization of 4 MeV treatment planning. Phys. Med.
computer-controlled tracking units. In: Orton, C. G., ed. Biol. 10:365-373; 1965.
Progress in medical radiation physics, chap. l-11. New 14. Hope, C. S.; Laurie, J.; Orr, J. S.; Halnan, K. E. Optimiza-
York: Plenum Press: 1985:45-94. tion of x-ray treatment planning by computer judgment.
7. Emami, B.; Lyman, J.; Brown, A.; Coia, L.; Goitein, M.; Phys. Med. Biol. 12:531-542; 1967.
Munzenrider, J. E.; Shank, B.; Solin, L. J.; Wesson, M. 15. Huesman, R. H.; Gullberg, G. T.; Greenberg, W. L.; Bud-
Tolerance of normal tissue to therapeutic irradiation. Int. J. inger, T. F. Users manual: Donner algorithms for recon-
Radiat. Oncol. Biol. Phys. 21:109-122; 1991. struction tomography. Lawrence Berkeley Laboratory, LBL
Tomotherapy l T. W. HOLMES et al. I225

pub-214, University of California, Berkeley. CA; October 26. Mantel, J.; Perry, H. Automatic variation of field size and
1977. dose rate in rotation therapy. Int. J. Radiat. Oncol. Biol.
16. Jennings, W. A. The tracking cobalt project: From moving- Phys. 2:307-317; 1977.
beam therapy to three-dimensional programmed irradiation. 27. McDonald, S. C.; Rubin, P. Optimization of external beam
In: Orton, C. G., ed., Progress in medical radiation physics, radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2:307-
chap. 1-I. New York: Plenum Press; 1985:144. 317; 1977.
17. Kak, A. C.; Slaney, M. Principles of computerized tomo- 28. Reckwerdt, P. Superposition/convolution speed improve-
graphic imaging. New York: IEEE Press; 1988. ments using run-length raytracing. (Abstr.) Med. Phys.
18. Kallman, P., Lind, B. K.; Brahme, A. An algorithm for 19:784; 1992.
maximizing the probability of complication-free tumor con- 29. Redpath, A. T.; Vickery, B. L.; Wright, D. H. A new tech-
trol in radiation therapy. Phys. Med. Biol. 37:871-890; nique for radiotherapy planning using quadratic program-
1992. ming. Phys. Med. Biol. 21:781-791; 1976.
19. Lane, R. G.; Loyd, M. D.; Chow, C. H.; Ekwelundu, E.; 30. Rosen, I. I.; Lane, R. G.; Morrill, S. M.; Belli, J. A. Treat-
Rosen, I. I. Custom beam profiles in computer-controlled ment plan optimization using linear programming. Med.
radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 22: 167- Phys. 18:141-152; 1991.
174; 1991. 31. Rosenberger, F. U.; Mathews, J. W.; Johns, G. C.; Drzy-
mala, R. E.; Purdy, J. A. Use of transputers for real time
20. Langer M.; Leong, J. Optimization of beam weights under
dose calculation and presentation for three-dimensional ra-
dose-volume restrictions. Int. J. Radiat. Oncol. Biol. Phys.,
13:1255-1260; 1987. diation treatment planning. Int. J. Radiat. Oncol. Biol. Phys.
25:709-719; 1993.
21. Lind, B. Properties of an algorithm for solving the inverse
32. Sonderman, D.; Abrahamson, P. G. Radiotherapy design
problem in radiation therapy. Inverse Problems 6:415 -426; using mathematical programming models. Op. Res.
1990. 33:705-725; 1985.
22. Luenberger, D. G. Linear and nonlinear programming. 2nd 33. Starkshall, G. A constrained least-squares optimization
ed. Reading, MA: Addison-Wesley; 1989. method for external beam radiation therapy treatment plan-
23. Mackie, T. R.; Holmes, T. W.; Swerdloff, S.; Reckwerdt, ning. Med. Phys. 11:659-665; 1984.
P.; Deasy, J. 0.; Yang, J.; Paliwal, B.; Kinsella, T. Tomo- 34. Takahashi, S. Conformation radiotherapy: Rotation tech-
therapy: A new concept for the delivery of conformal rdio- niques as applied to radiography and radiotherapy of can-
therapy using dynamic compensation. Med. Phys. 20: 1709- cer. Acta Radiol. Suppl. 242:1-142; 1965.
1719; 1993. 35. Webb, S. Optimization of conformal radiotherapy dose dis-
24. Mackie, T. R.; Reckwerdt, P. J.; Gehring, M. A.; Holmes, tributions by simulated annealing. Phys. Med. Biol.
T. W.; Kubsad, S. S.; Thomadsen, B. R.; Sanders, C. A.; 34:1349-1370; 1989.
Paliwal, B. R.; Kinsella, T. J. Clinical implementation of 36. Webb, S. Optimization by simulated annealing of three-
the convolution/superposition method. In: Proceedings of dimensional, conformal treatment planning for radiation
the 10th International Conference on Computers in Radio- fields defined by a multileaf collimator: II. Inclusion of
therapy, Lucknow, India. Lucknow, India: Alpana Arts; two-dimensional modulation of the x-ray intensity. Phys.
1990:322-325. Med. Biol. 37: 1689- 1704; 1992.
25. Mageras, G. S.; Podmaniczky, K. C.; Mohan. R. A model 37. Zacarias, A. S.; Lane, R. G.; Rosen, I. I. Assessment of a
for computer-controlled delivery of 3-D conformal treat- linear accelerator for segmented conformal radiation ther-
ments. Med. Phys. 19:945-953; 1992. apy. Med. Phys. 20:193-198; 1993.

You might also like