Professional Documents
Culture Documents
Digital Inclinometer System: User Manual
Digital Inclinometer System: User Manual
inclinometers
& TILT METERS
INdex Introduction Page 4
Description Page 5
Troubleshooting Page 37
Maintenance Page 39
Symbols
Note The present Manual is issued by SISGEO in English Language and should be available in
other different languages.
In order to avoid discrepancies and disagreements on the interpretation of the meanings,
Sisgeo Srl declares that English Language prevails.
All the information in this document is the property of Sisgeo S.r.l. and should not be used
or reproduced without permission from Sisgeo S.r.l.
We reserve the right to change our product without prior notice.
For any other information regarding inclinometers, please refer to the International Standard
ISO 18674-3.
- dummy probe (C), similar to the inclinometer probe, but without sensor. It allows to check
the accessibility of the inclinometer casing before the survey. Supplied with dummy cable
reel (C.1)
- spiral meter, to measure the twisting of inclinometer casing’s keyways;
- pulley and cable stop (D), used to block the cable and as depth reference.
- KLION data analysis software (see relevant datasheet and manual)
C.1 D
F A C A.1
inclinometer
control cable
cable connector
electrical
connector
B A
gauge length
rubber tip
electrical electrical
connector “2” connector “1”
spring-loaded wheels cable
connector
fixed wheels
(gauge length)
2
1
inclinometer
control cable
• ON/OFF button
• Notification LED
• Charger connector
ON/OFF button
Notification LED
Charger connector
STATUS LED
The LED, through a colour code, shows B.R.A.IN status. Following is the colours code:
• POWER ON: the green flashing LED means that B.R.A.IN is switched on but the
device hasn’t established a Bluetooth connection with B.R.A.IN reel yet;
• BT CONNECTION: the blue flashing LED means that B.R.A.IN successfully connected
with a device;
• ANOMALY: the red flashing LED means that there is a system’s anomaly. On the APP
will be displayed the anomaly’s details;
• LOW BAT: the yellow flashing LED means that B.R.A.IN’s rechargeable battery is
empty and needs to be recharged.
CHARGER connector
The connector on the lower part of the panel allows the recharge of the batteries inside
B.R.A.IN.
Connecting the supplied battery charger, the batteries will be recharged. It is possible to
use B.R.A.IN during the recharge.
ATTENTION: we recommend to use only the original battery charger supplied by SISGEO to
avoid damaging the batteries and B.R.A.IN itself.
NOTE: during the re-charge, ONLY the led on the battery charger will switch on; instead
there will be no led flashing on B.R.A.IN panel.
RESET
It is possible to force B.R.A.IN reset keeping the button on the panel pushed for 10
seconds.
As soon as B.R.A.IN will receive the reset command, the LED will flash repeatedly with the
sequence of two short and one long flash.
When this sequence is shown, release the button: the LED will start to flash green. After
this operation, in some cases will be necessary to pair again the reel with the device.
The smartphone/tablet where the APP is installed must be compatible with BLE (Bluetooth Low
Energy) from version 4.2 onward.
The APP can be downloaded from the APP stores of both O.S. (Google Play Store for Android and
APP Store for Apple). Search for the APP and then click on Install key, the APP will be dowloaded
and then installed on the device. With the Android O.S. it is possible to install the APP also using
directly the APK (Android Application Package) of the app.
The APK file can be supplied by SISGEO in case of connection problems to Google Play Store.
In this situation, it’s necessary to enable the installation of apps of unknown source.
This option is usually in the menu “Advanced Settings - Safety - Unknown source”. Please refer
to your device manual for more information.
NOTE: the association through pairing must be carried out only the first time you connect
B.R.A.IN to a device.
NOTE: the Bluetooth connection between device and B.R.A.IN happens only after the
selection of its B.R.A.IN in the APP; at the end of the test the device will be automatically
disconnected from B.R.A.IN.
SET-UP After a few seconds will be displayed the welcome page with the main menu (also called
home page)
START: it allows the configuration of sites and inclinometer casings, as well as inclinometer
readings execution and the export of the measures taken.
SYSTEM: it allows to obtain information about the probe connected, about integrated reel
electronics, about the APP version and the device. This page summarize all the information
necessary in case of assistance with Sisgeo’s Customer Care.
NOTE: in order to show B.R.A.IN and the probe’s information, B.R.A.IN must be connected
to the device and the probe to B.R.A.IN. If they are not connected, both fields will result
empty.
MANUALS: it create a connection with SISGEO WEBSITE, on the products page, with its
manuals and the video tutorial.
SITE CREATION
In order to create a new site, proceed as follows:
• select START menu
• select SITE/CASING CONFIGURATION, then SITES and then NEW SITE:
• It will be opened a new page where you can insert the data of the SITE you are about
to create. You can insert a SITE NAME and a SITE DESCRIPTION. Selecting the icon
“POSITION” the device will record automatically the GPS coordinates of the survey
site in SITE characteristics. A check mark will be shown next to POSITION icon.
Selecting INSERT A PHOTO it will be possible to upload and link a photo from the
device’s gallery with the SITE.
MODIFY/DELETE SITE
In order to modify a SITE previously created, proceed as follows:
• select START menu
• select SITE/CASING CONFIGURATION, then SITES and then EDIT next to the SITE
you want to modify:
CASING CREATION
In order to create a new CASING proceed as follows:
• select START menù
• select SITE/CASING CONFIGURATION, select then CASINGS
• Select the SITE where the casing will be created, select CASINGS and then NEW
CASING:
• It will open CASING configuration page. Once you have configured CASING’s
parameters, press ACCEPT to save configuration.
CASING NAME: name to be associated to the CASING. This field will be used for the
identification of the casing in the APP.
CASING DESCRIPTION: descriptive field to be used in order to better identify the CASING
in the APP (optional)
MEASURE TYPE: selection field where can be choosen the kind of measure to be executed
among INCLINOMETER SURVEY, SPIRAL METER SURVEY or T-REX SURVEY
UNIT: it allows to select the unit of measurements used to manage the inclinometer survey
between METERS and FEET. Automatically it will change the unit of measure displayed for
the MEASURING LENGTH and for the STEP LENGTH.
NOTE: this unit of measurement must be selected in agreement with probe’s gauge length.
Infact it’s not correct to have a metric probe with settled “feet” as unit of measurements
and viceversa
MEASURING LENGTH: it allows to insert the length of the casing’s section to be measured
(for details see APPENDIX 5). This value must be a must be a multiple of the value specified
in STEP LENGTH.
• Selecting the icon POSITION the device will record automatically the GPS coordinates
of the survey site in SITE characteristics. A check mark will be shown next to
POSITION icon.
Selecting INSERT A PHOTO it will be possible to upload and link a photo from the
device’s gallery with the SITE.
Selecting ACCEPT you will return to the CASING CONFIGURATION submenu and will
be shown the CASING just created.
MENU shown a page with all the information on the probe, of the reel electronics, of the APP and
of the used device. This page is really useful in case of the intervention of our Customer
Care Department is needed. It can be shared with SISGEO in order to supply the necessary
information on the inclinometer system and receive a proper assistance.
NOTE: in order to obtain all the information it’s necessary that the probe would be connected
to the B.R.A.IN panel. Otherwise no information about the probe will be available.
• Selecting SYSTEM it will be displayed the list of B.R.A.IN system available
• Select the B.R.A.IN you want to connect to, after a few seconds it will be displayed
the page with the information summary.
In the case that the probe is not connected, the SYSTEM INFO will give data of BRAIN reel,
DEVICE and APP revision.
If no BRAIN will be founded from the APP, the SYSTEM INFO will give after 30 seconds
countdown the data of DEVICE and APP only.
To execute a BACKUP:
• Select BACKUP/RESTORE
• Select BACKUP. It will display a window where the system will suggest some
APP that can be used to execute the BACKUP. After the choice of the APP (in the example,
G-drive), select the name of the backup file and the position where to save it. Then press
SAVE. The BACKUP will be saved.
To execute a RESTORE.
• Select BACKUP/RESTORE
• Select RESTORE. A warning popup will appear informing the user that all APP’s
data will be replaced from the ones of the BACKUP: select YES if you wish to proceed. Now
you shall look for the proper BACKUP file in the position where it has been saved. After the
selection, a popup will appear informing that the RESTORE was successful.
NOTE: In order to retrieve a backup saved on a cloud, it’s necessary to download it in advance
on the device memory. Otherwise the RESTORE procedure will show an error message.
NOTE: the APP and his content automatic backup can be executed automatically through the
operating system base functions. In this case it will be necessary to tell the O.S. to execute
the backup of B.R.A.IN APP and its content. For further information please refer to the O.S.
manufacturer user manual.
NOTE: with iOS, before restoring data you have to activate the APP “File” on the device.
NOTE: a BACKUP file produced with Android device can be used for a RESTORE in an iOS
device and viceversa.
MEASUREMENTS Inclinometer casing survey consists in a number of readings at regular steps of the tilt of
the inclinometer probe.
Probe shall be moved incrementally along the casing, whereby for ISO 18674-3 each
increment shall not exceed the gauge length of the probe (500mm, 1000 mm or 2 ft).
reference
1 The inclinometer casing should be installed with one keyway in the reference direction.
In vertical boreholes, the reference direction should be the direction of the anticipated
keyway
principal movement OR the keyway closer to geographic north.
The keyway in the reference direction shall be the reference keyway and shall be durably
marked, for example, by cutting a notch at the top of the guide tube into that keyway or by
4 2 identifying it by a water-resistant and wear-resistant mark.
Inclinometer’s casing keyways, once installed, are identified with numbers 1-2-3-4,
clockwise; the keyway in the reference direction is the key way “1”.
3 The below scheme shows SISGEO probe sign convention: the lower wheel and letter A are
the probe’s reference side. If you place the probe vertically on an horizontal flat surface and
tilt it in the arrows direction, channel A and B readings (the 2 orthogonal axis) take different
signs.
Probe signs convention is reverse if compared with ISO18674-3. Using KLION data
elaboration software, the sign is converted and the elaboration is in compliance to the ISO
standard.
reference wheel
B
(lower)
A-
B+
B-
A+
reference wheel
where:
A1B1 is the run with probe’s reference wheel in keyway 1
A3B3 is the run with probe’s reference wheel in keyway 3 (reverse or conjugated run)
A2B2 is the run with probe’s reference wheel in keyway 2
A4B4 is the run with probe’s reference wheel in keyway 4 (reverse or conjugated run)
The first complete survey is called “reference reading” and is the survey compared to
which will be calculated all the future displacements of the inclinometer’s casing.
For establishing a reliable reference set, there should be, in the reference reading, at least
two repetitions of the initial measuring survey whereby each survey consists of a normal
(0°) and a reverse (180°) run along the guide tube (usually A1B1 - A3B3), under repeatable
conditions.
We suggest to take reference readings on four keyways (A1B1 - A3B3 - A2B2 - A4B4),
because sometimes casing could deform, preventing probe’s readings, while the
perpendicular keyways can still be used.
Before inserting the inclinometer probe in the casing, we advice to check casing’s accessibility
using the dummy probe, which has to slide along the casing with the reference wheels put
in all the 4 keyways.
The stability of the readings is critical for a correct measurement. Therefore, at the beginning
of a survey, the measuring probe shall be brought to the point furthest from the collar of the
guide tube (lower part of the casing) and left there to adjust to the ambient temperature until
the readings are considered to be sufficiently stable. We suggest a minimum stabilization
time of 10 minutes.
Advices:
• Use always the same inclinometer system (probe-B.R.A.IN reel) to survey the casing.
• Use the pulley and cable stop.
• When you move from a level to the next one, record the data only when the values are stable.
• Avoid pulls and excessive speed in the probes’ recovery.
The inclinometer reading consists in a survey, at regular steps, of the inclinometer casing
tilt compared to the horizontal.
The step is usually the same of the wheels assembly distance (gauge length).
For establishing a reliable reference set, there should be, in the reference measurement,
at least two repetitions of the initial measuring survey whereby each survey consists of a
normal and a reverse run along the inclinometer tube under repeatable conditions.
The inclinometer casing should be installed with one keyway in the reference direction.
In horizontal and inclined boreholes, the reference direction should be the “top” of the
borehole. The keyway which is in the reference direction shall be the reference keyway and
shall be durably marked, for example, by cutting a notch at the top of the guide tube into
that keyway or by identifying it by a water-resistant and wear-resistant mark. Inclinometer’s
casing keyways, once installed, are identified with numbers 1-2-3-4, clockwise; the keyway
in the reference direction is the keyway “1”.
connector 1
ght
e len
gaug
connector 2
A<0
reference keyway
spring-loaded
wheel
4 2
fixed wheel
Advices:
• Use always the same inclinometer system (probe-B.R.A.IN reel) to survey the casing.
• When you move from a level to the next one, record the data only when the values are
stable.
• Avoid pulls and excessive speed in the probes’ recovery.
After the configuration of at least a SITE and a CASING it will be possible to execute the
inclinometer measurement.
Proceed as follows in order to execute the inclinometer measurement:
• Select START
• Select SURVEY, then the SITE and lastly the CASING where the inclinometer survey
will be executed:
If there are already measurements for the CASING, it will be possible to select a measure in
order to use it as reference reading for the next measure it will be executed.
Otherwise, select NO REFERENCE READING in case you don’t want to refer the inclinometer
survey to any previous reading.
Clicking on the top-right arrow, it will be displayed the page showing the last options to be
configured before the reading:
• SURVEY NUMBER: by default it will be suggested the number “0” if in the current
casing there are no other surveys. if there would be other surveys, the proposed
number automatically will be the next one referring to the last survey
• DELAY: by default it is suggested “Manual”, so, the operator shall tap during the survey
in order to record the measure. Otherwise it will be suggested time steps in order to
execute the survey automatically.
• DIRECTION: it will be possible to select whether to execute the survey starting from
the bottom (UPWARD) or from the top (DOWNWARD)
Pressing on CONFIRM the survey will start.
In case you enabled, in the APP settings, the measure stability notification (you find more
info in the chapter “SETTING MENU”), it will be displayed a warning telling the user to wait
at least 10 minutes before executing the measure. This will allow the probe to sabiliize its
temperature at the bottom of the hole.
Pressing on OK, will start the probe’s reading page.
It will display a timer that runs a 10 minutes countdown. The light blue button warns the user
about the time to be waited until the inclinometer measuerement can start, as suggested
by SISGEO.
It’s anyway possible to start immediately the inclinometer survey, without waiting the 10
minutes, pressing on the light blue button, that will become green confirming the start of the
inclinometer survey. For the detailed description of reading page, please refer to APPENDIX 6.
In order to record the measure, it’s necessary to press on READ button. Once pressed, the
data will be recorded and the device will vibrate and ring, confirming the measure taken.
Positioning the probe at the chosen depths and pressing READ on each occasion, the
measures will be recorded.
In case a wrong measure is taken, and you need to do it again, you just need to press on the
BACK button, and you can take another one. The previous wrong measure will be overwritten.
It’s also possible to interrupt the inclinometer survey, pressing the button ABORT.
In this case a pop up will ask if you want to keep the already recorded data, or delete them.
NOTE: also saving data, it won’t be possible to restart the interrupted inclinometer survey.
Once all the measure steps have been executed, the READ button will become grey and can’t
be selected anymore. The button SAVE, grey and not-selectable until now, it will become
green and it will be possible to save the survey just ended.
Before the ending of the survey and the data saving, it will be possible to insert some text
notes about to the survey that will be recorded in the XML file.
Once you select CONFIRM, it will be confirmed the survey saving through a page where it
will be possible to view immediately the graph of the survey just ended (selecting DATA).
At the end select FINISH.
It will show the list of of the readings executed with reference of date and time.
From the list select the SURVEY you want to check the readings: it will open a table with all
the readings at their measure steps.
Swiping left on the page you will pass from the table, to the graphs of the measure.
ANDROID iOS
Below you can find the basic information for data processing.
The values (digits) displayed and stored by B.R.A.IN APP can be converted in degrees
according to the following formula:
digits = (sin α) x K
and so:
α = ARCSIN ( digits / K )
In order to calculate the probe displacement from vertical at each step, proceed with the
following formula:
where:
DA, i = probe local displacements in A-axis direction at the generic step “i”
DB, i = probe local displacements in B-axis direction at the generic step “i”
Now, to calculate the resultant displacement and its direction at the generic step “i”, use
the following formula:
where:
where:
The same formula can be used also to calculate the total cumulative displacement in A-axis
and B-axis, by replacing “Di” with “DA, i” or “DB, i”
The total cumulative displacement calculation must be done considering that one end
of the casing is not subject to movements (usually deepest point as reference point), or
considering the absolute displacements of the casing taken, for example, by means of
topographic measurements.
To calculate the probe displacement from horizontal at each step, proceed with the
following formula:
D i = L x sin α i
where:
In order to calculate the total cumulative vertical displacement, proceed with the following
formula:
where:
The total vertical cumulative displacement calculation must be done considering that one
end of the casing is not subject to movements, or considering the absolute displacements
of the casing taken, for example, by means of topographic measurements.
Check-sums should be used as the basis for the acceptance test of the readings and for
identifying erroneous measurements.
If readings are incorrect or in doubt, the inclinometer survey shall be repeated.
The check-sum is the sum of the normal and reverse readings taken at the same depth “i”.
Ideally, the sum is zero since the readings have opposite signs.
Basically, however, variations in the positioning of the probe and zero offsets of the probe
contribute to non-zero check-sums.
The check-sums can be affected by sediments in the keyways, by inclinometer casing
deformation and from the wheel positioning near the couplings.
Common practice is plotting the check-sums of all measuring points over the inclinometer
guide tube.
In normal circumstances, each combination of probe and guide tube has a unique check-
sum signature, which is repeatable in each follow-up measurement.
Small variations do not usually indicate a problem, however, spikes in the plot are indicative
of inconsistencies in the readings.
SISGEO suggests a check of the inclinometer system in our laboratory in case of check-
sum values higher than 50digit.
KLION software automatically calculates the check-sums and generates plots.
For any further information regarding the check-sums interpretation, please refer to “Mikkelsen P.E.
2003. Advances in inclinometer data analysis. Proceed. 6th Int. Sympos. Field Meas. in Geomechanics,
15–18 September 2003, Oslo, pp. 555–567, Lisse (Balkema)”
Wrong
Wrong choice of the reference Use the correct reference
measures
put in charge B.R.A.IN reel using
Low battery voltage
the supplied charger.
https://support.sisgeo.com/
Send back the instrument/equipment with the complete accessories, using suitable
packaging, or, even better, the original ones.
The shipping costs shall be covered by the sender.
Please return to the following address, with correct delivery documentation, reporting the
RMA code received:
SISGEO S.r.l.
Via F.Serpero, 4/F1
20060 MASATE (MI)
How to correctly connect probe and cable in order to avoid any connector damage:
1 2
3 4
C N.C. C Grey
D N.C. D Yellow
Shield
first cable 5 1
mark 10 1
15 1 1
20 1
25 1 1
30 1 1
35 1 1 1
40 2
45 1 2
50 1
55 1 1
1250 mm (with gauge length equal to 1000 mm)
1000 mm (with gauge length equal to 500 mm)
60 1 1
65 1 1 1
70 1 1
75 1 1 1
80 1 1 1
85 1 1 1 1
90 2 1
95 1 2 1
100 2
105 1 2
110 1 2
115 1 1 2
120 1 2
125 1 1 2
130 1 1 2
135 1 1 1 2
140 2 2
145 1 2 2
150 3
155 1 3
160 1 3
gauge length
165 1 1 3
center line
170 1 3
between
175 1 1 3
wheels
180 1 1 3
assembly
185 1 1 1 3
190 2 3
195 1 2 3
200 4
92 mm
first cable 10 1
mark 20 1
30 1 1
40 2
50 1
60 1 1
70 1 1
80 1 1 1
90 2 1
100 2
110 1 2
4 ft (48 in) with gauge length equal to 2 ft (24 in)
120 1 2
130 1 1 2
140 2 2
150 3
160 1 3
170 1 3
180 1 1 3
190 2 3
200 4
210 1 4
220 1 4
230 1 1 4
240 2 4
250 5
260 1 5
270 1 5
280 1 1 5
290 2 5
300 6
gauge length
center line
between
wheels
assembly
3.6 in
1 2
- Remove the cap from the - Align the pushing rod connecting
inclinometer cable connector tool with the cable connector
3 4
- Stick the tool on the connector - Tighten the two screws with allen
wrench N.5
5 6
- Unscrew the protecting cap from - Insert carefully the connector and
the inclinometer probe connector tighten the nut
7 8
- Allign the female connector of the - Tighten the nut of the joint.
first pushing rod to the male joint of For the next rods, proceed as for
the connecting tool step 7 and 8
cable mark
measure length
CASING name
SITE name
READ button
SAVE button
(active when survey is finised)
BACK button
(it erase the last reading)
DATA button
(it show all collected readings
and check-sums values)
NOTE