Professional Documents
Culture Documents
net/publication/236278017
CITATIONS READS
9 450
7 authors, including:
31 PUBLICATIONS 308 CITATIONS
Celltrion
151 PUBLICATIONS 2,150 CITATIONS
SEE PROFILE
SEE PROFILE
All content following this page was uploaded by Suk Ju Cho on 15 October 2015.
| Original Article |
Jae Sung Ryu, MD, Seung Woo Baek, MD, Cheol Hee Jung, MD, Suk Ju Cho, MD*,
Eu Gene Jung, MD, Hae Kyoung Kim, MD, and Jae Hun Kim, MD
Background:
Medical doctors who perform C-arm fluoroscopy-guided procedures are exposed to X-ray radiation. Therefore,
radiation-protective shields are recommended to protect these doctors from radiation. For the past several years,
these protective shields have sometimes been used without regular inspection. The aim of this study was to
investigate the degree of damage to radiation-protective shields in the operating room.
Methods:
This study investigated 98 radiation-protective shields in the operation rooms of Konkuk University Medical
Center and Jeju National University Hospital. We examined whether these shields were damaged or not with
the unaided eye and by fluoroscopy.
Results:
There were seventy-one aprons and twenty-seven thyroid protectors in the two university hospitals. Fourteen
aprons (19.7%) were damaged, whereas no thyroid protectors (0%) were. Of the twenty-six aprons, which have
been used since 2005, eleven (42.3%) were damaged. Of the ten aprons, which have been used since 2008,
none (0%) was damaged. Of the twenty-three aprons that have been used since 2009, two (8.7%) of them
were damaged. Of the eight aprons used since 2010, one (12.3%) was damaged. Of the four aprons used since
2011, none (0%) of them were damaged. The most common site of damage to the radiation-protective shields
was at the waist of the aprons (51%).
Conclusions:
As a result, aprons that have been used for a long period of time can have a higher risk of damage.
Radiation-protective shields should be inspected regularly and exchanged for new products for the safety of
medical workers. (Korean J Pain 2013; 26: 142-147)
Key Words:
fluoroscopy, radiation exposure, radiation-protective shields.
Received November 19, 2012. Revised January 8, 2013. Accepted January 9, 2013.
Correspondence to: Jae Hun Kim, MD
Department of Anesthesiology and Pain Medicine, Konkuk University Medical Center, 4-12 Hwayang-dong, Gwangjin-gu, Seoul 143-729,
Korea
Tel: +82-2-2030-5470, Fax: +82-2-2030-5449, E-mail: painfree@kuh.ac.kr
This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://
creativecommons.org/licenses/by-nc/3.0/), which permits unrestricted non-commercial use, distribution, and reproduction in any medium,
provided the original work is properly cited.
Copyright ⓒ The Korean Pain Society, 2013
Ryu, et al / The Survey about the Degree of Damage of Radiation-Protective Shields in Operation Room 143
www.epain.org
144 Korean J Pain Vol. 26, No. 2, 2013
www.epain.org
Ryu, et al / The Survey about the Degree of Damage of Radiation-Protective Shields in Operation Room 145
Fig. 3. Degree of utilization of radiation-protective shields. (A) Apron. (B) Thyroid protector. KUMC: Konkuk University
Medical Center, JNUH: Jeju National University Hospital. *P < 0.05 Apron versus Thyroid protector. †P < 0.05 Pain clinic
versus Surgery.
lized during surgery than during pain management. In which were introduced in 2009 found to be damaged. An
professors/fellows, the degree of utilization of aprons used increase in the number of aprons might result in a de-
by surgery personnel was significantly lower than that of crease in the frequency of use for a single apron. At JNUH,
pain physicians. The degree of utilization of thyroid pro- the overall rate of damage to the aprons was 90%. The
tectors by surgery personnel was significantly lower than total number of aprons at JNUH was thirteen, among them
that of the pain clinicians. Almost all pain clinicians wore ten aprons were introduced in 2005. As a result, a longer
thyroid shields during procedures, but fewer than half of period and more frequent use of the same protector was
the surgery personnel wore them (Fig. 3). thought to be a possibility for the increase in damage. The
most common sites of damage to the aprons were at the
DISCUSSION waist. This is probably due to waistbands causing tension
on the aprons’ waist sections, which are frequently folded
At KUMC, the rate of damage to the aprons that had while walking and sitting (Fig. 4). There were four two-
been introduced in 2005 was the highest because physi- piece garments at KUMC. There were two damaged kilts
cians had used them for three years without receiving any whereas there was no damaged vest (Fig. 2B). Male resi-
new supplies. There were no damages found in aprons dents (surgeons) tend to wear kilts without vests at KUMC
which were introduced after 2008, and only two aprons therefore kilts were used more frequently than vests.
www.epain.org
146 Korean J Pain Vol. 26, No. 2, 2013
Aprons have more folds, so can break easily. As a result, scattered x-rays that strike it [7]. According to another
aprons should be stored on a hanger in order to minimize study, in C-arm fluoroscopy-guided procedures, exposure
folds which can lead to damage. In addition, aprons should in terms of mREM should be 2,690 mREM outside the
be inspected fluoroscopically on an annual basis to check apron with 0 mREM inside the apron, whereas it should
for holes and a reduction in shielding integrity [7]. be 4.03 mREM outside the apron for the entire body (head
In this study, the degree of utilization of radia- and trunk) with an effective dose equivalent exposure of
tion-protective shields for surgeons was significantly lower 0 mREM per year inside the apron [12].
than that of pain physicians. This is thought to be due to The International Commission on Radiological Protec-
the low awareness or knowledge of radiation safety. In ad- tion (ICRP) has set an annual maximum permissible radia-
dition, this result seems to be related to the surgeon wear- tion dose to reduce damages caused by radiation exposure,
ing an aseptic gown over the radiation-protective shields. with the ICRP strongly suggesting that radiation exposure
In these cases, the doctor may feel excessive heat and be within these set limits [13]. The annual allowance of ra-
discomfort. Although the number of thyroid shields is less diation exposure is 50 mSv. In Europe, however, exposure
than that of aprons, there has been no lack of thyroid should not exceed 20 mSv/year on average over 5 years;
shields for clinicians to use. Therefore, the lack of damage in the U.S., regulations limit exposure to only 10 mSv/year
to the thyroid protectors may not be related to the lower over a person’s entire life [14].
number of thyroid shields, but instead may be related to The cardinal principles of radiation protection are: (1)
the lower utilization of thyroid shields than that of aprons. maximize the distance from the radiation source; (2) use
The amount of radiation exposure is proportional to shielding materials; and (3) minimize exposure time [10].
the duration of exposure, while being inversely proportional Precautions should be taken to reduce exposures as much
to the square of the distance from the radiation source [9]. as possible. Potential decreases in radiation exposure can
This is also proportional to the damaged areas of the be accomplished by decreased exposure time; increased
aprons. If shielding materials collapses inside the apron, distance; increased shielding with aprons, thyroid gland
the apron cannot protect the wearer from radiation covers, gloves, and glasses; beam collimation; using the
exposure. Radiation exposure results from the primary low-dose option available on some C-arm units; inverting
beam and the scatter radiation reflected off the floor and the C-arm; and surgeon control of the C-arm [15].
the patient [10]. The practitioner receives the highest dos- Available patient dose reduction technologies include
age overall in comparison to the assistants and scrub low fluoroscopy dose rate settings, low frame rate pulsed
nurses [11]. Standing 1 m away before obtaining an image fluoroscopy, spectral beam filtration, and the use of in-
can significantly reduce the levels of radiation exposure to creased X-ray beam energy [14]. Application of collimation
the staff members [1]. In most areas, radiation-protective can exclude areas of greatly varying radiodensity to im-
shields require a thickness of at least 0.5 mm of lead- prove the image quality by reducing the range of densities
equivalents to be used, which attenuates over 90% of the included in the field [16]. Using the pulsed mode and
www.epain.org
Ryu, et al / The Survey about the Degree of Damage of Radiation-Protective Shields in Operation Room 147
low-dose modes together showed a significant decrease in 6. Shope TB. Radiation-induced skin injuries from fluoroscopy.
the dose of radiation absorbed than in any other mode [17]. Radiographics 1996; 16: 1195-9.
In this investigation, the shielding material in pro- 7. Schueler BA. Operator shielding: how and why. Tech Vasc
Interv Radiol 2010; 13: 167-71.
tective garments can develop cracks and holes over time,
8. Yaffe MJ, Mawdsley GE, Lilley M, Servant R, Reh G.
although they may not be visible externally. To help pre- Composite materials for x-ray protection. Health Phys 1991;
vent this, aprons should be stored on hangers with mini- 60: 661-4.
mum folding. Also, garments should be monitored annually 9. Poletti JL, McLean D. The effect of source to image-receptor
with fluoroscopy to check for holes and reduced shielding distance on effective dose for some common X-ray
integrity [7]. To make fluoroscopy use safer, education projections. Br J Radiol 2005; 78: 810-5.
10. Botwin KP, Freeman ED, Gruber RD, Torres-Rames FM,
about, attention to, and practices of radiation safety
Bouchtas CG, Sanelli JT, et al. Radiation exposure to the
should be required for all pain physicians [18].
physician performing fluoroscopically guided caudal epidural
This study had some limitations. We did not investigate steroid injections. Pain Physician 2001; 4: 343-8.
the frequency of use of each radiation protector, but only 11. Hellawell GO, Mutch SJ, Thevendran G, Wells E, Morgan RJ.
the periods of usage of the radiation protectors. It was im- Radiation exposure and the urologist: what are the risks? J
possible to assess how many times each protector had Urol 2005; 174: 948-52.
been used. Therefore, we supposed that the period was re- 12. Manchikanti L, Cash KA, Moss TL, Pampati V. Radiation
exposure to the physician in interventional pain management.
lated to the frequency of use. In this study, only two hos-
Pain Physician 2002; 5: 385-93.
pitals were involved. If more institutes were included, the 13. The 2007 Recommendations of the International Commission
data might be better than our current findings. on Radiological Protection. ICRP publication 103. Ann ICRP
In conclusion, radiation-protective shields, which have 2007; 37: 1-332.
been used for a long time, can have a higher risk of 14. Miller DL, Vañó E, Bartal G, Balter S, Dixon R, Padovani R,
damage. Radiation-protective shields should be inspected et al. Occupational radiation protection in interventional
radiology: a joint guideline of the Cardiovascular and Inter-
regularly and exchanged for new products for the radiation
ventional Radiology Society of Europe and the Society of
safety of medical workers.
Interventional Radiology. J Vasc Interv Radiol 2010; 21:
607-15.
REFERENCES 15. Singer G. Occupational radiation exposure to the surgeon.
J Am Acad Orthop Surg 2005; 13: 69-76.
1. Kim TW, Jung JH, Jeon HJ, Yoon KB, Yoon DM. Radiation 16. Powys R, Robinson J, Kench PL, Ryan J, Brennan PC. Strict
exposure to physicians during interventional pain procedures. X-ray beam collimation for facial bones examination can
Korean J Pain 2010; 23: 24-7. increase lens exposure. Br J Radiol 2012; 85: e497-505.
2. Le Heron J, Padovani R, Smith I, Czarwinski R. Radiation 17. Cho JH, Kim JY, Kang JE, Park PE, Kim JH, Lim JA, et al.
protection of medical staff. Eur J Radiol 2010; 76: 20-3. A study to compare the radiation absorbed dose of the
3. Tonnessen BH, Pounds L. Radiation physics. J Vasc Surg C-arm fluoroscopic modes. Korean J Pain 2011; 24:
2011; 53: 6S-8S. 199-204.
4. Mroz TE, Yamashita T, Davros WJ, Lieberman IH. Radiation 18. Park PE, Park JM, Kang JE, Cho JH, Cho SJ, Kim JH, et
exposure to the surgeon and the patient during kyphoplasty. al. Radiation safety and education in the applicants of the final
J Spinal Disord Tech 2008; 21: 96-100. test for the expert of pain medicine. Korean J Pain 2012;
5. Cousins C, Sharp C. Medical interventional procedures- 25: 16-21.
reducing the radiation risks. Clin Radiol 2004; 59: 468-73.
www.epain.org