You are on page 1of 11

D. Gowrisankar et al / J Biomed Sci and Res.

, Vol 2 (2), 2010,89-99

Validation and Calibration of Analytical Instruments


a
D.Gowrisankar, bK.Abbulu, cO.Bala Souri, K.Sujana*
a
Department of Pharmaceutical Analysis, Andhra University, Visakhapatnam.
b
Department of Pharmaceutics, Mallareddy Institute of pharmaceutical sciences, Hyderabad,
c
Department of Quality control, Neuland laboratories, Hyderabad.
Abstract:
Analytical instruments are used for a specific analysis. So regular performance verifications are made to ensure that
the instrument to be used is suitable for its intended application. All equipments used in the production of products
shall be properly Validated and Calibrated “to demonstrate that it is suitable for its intended purpose”.
Key words: Calibration, GC, HPLC, UV-Visible spectrophotometer, Validation.
Definition of Validation: USP: Installation Qualification (I.Q):
“Validation of an analytical procedure is the The purpose of I.Q is to check the
process by which it is established, by installation site/environment, confirms
laboratory studies, that the performance equipment specifications and verifies the
characteristics of the procedure meet the condition of installed equipment.
requirements for the intended analytical I.Q protocol shall include the following:
applications.”  Confirmation of the specifications of
Validation Protocol: the analytical equipment.
A written plan stating how validation will be  Confirmation and maintenance of
conducted and defining acceptance criteria. documents (Instruction manuals,
For example, the protocol for a qualification protocol and certificates).
manufacturing process identifies processing  Confirmation of installation site and
equipment, critical process conditions
parameters/operating ranges, product  Confirmation of delivered equipment.
characteristics, sampling, test data to be  Confirmations of Software and
collected, number of validation runs, and Firmware i.e., verify that the equipment
acceptable test results. is consistent with
Equipment validation: Demonstrate that actual versions displayed when power is
equipment used in validation studies is turned on.
suitable for use and is comparable to In I.Q, connect each unit (Electrical system,
equipment used for routine analysis Flow line system) and confirm that the
Calibrated (as applicable) connections are correct.
Qualifications should have been performed Any problems identified in I.Q must be
Installation Qualification investigated and appropriate actions must be
Operation Qualification taken. All such actions must be documented
Performance Qualification and approved by higher authority.
Routine maintenance performed- proper Operational Qualification (O.Q):
working order O.Q includes procedures and documentation
Qualification: Action of proving and of O.Q of analytical instrument. When all
documenting that equipment or ancillary procedures are executed and all items pass
systems are properly installed, work the inspection, it is verified that the system
correctly, and actually lead to the expected operates to satisfy the intended purpose.
results. Qualification is part of validation, O.Q protocol shall include the following:
but the individual qualification steps alone  Operation check on each unit
do not constitute process validation.

89
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

o Due to modular nature of the 3. Determination of the test intervals, e.g.,


system, the operation of each (a) - Everyday.
unit is checked properly. (b) - Every time the system is used.
 Operation check on overall system (c) - Before, between and after a series of
o Confirm that the system runs.
controller and work station 4. Define corrective actions on what to do if
control each unit during the system does not meet the established
analysis and that the analysis criteria.
results meet the prescribed Definition of Calibration: ICH
criteria. The demonstration that a particular
 Software and Firmware check instrument or device produces results within
Here Firmware checking is specified limits by comparison with those
conducted based on version display and produced by a reference or traceable
Software certificate of Compliance. The standard over an appropriate range of
Software and Firmware must be properly measurements.
managed and change procedures must be Calibration of HPLC: Various Calibration
properly clarified. parameters are:
Any problems identified in O.Q must be  Flow rate accuracy
investigated and appropriate actions must be  Injector accuracy
taken. All such actions must be documented  System Precision
and approved by higher authority.  Wavelength accuracy
Prior to implementing O.Q, check the  Detector linearity
system configuration, determine the items to  Injector linearity
be evaluated and record them in O.Q record  Gradient Performance Check
and have them approved.  Column oven temperature accuracy
Performance Qualification (PQ): Flow Rate Accuracy:
The objective is to ensure that the 1. Prime all the solvent lines with Milli Q
instrument is performing within specified water.
limits. Hence documented verification that 2. Set the flow rate to 0.500 ml/m.
the equipment and ancillary systems, as 3. Wait for about 15 m to stabilize the
connected together, can perform effectively system and ensure that the pressure is stable.
and reproducibly based on the approved 4. Insert the outlet tubing into a 10 ml
process method and specifications. volumetric flask and start the stop watch
The PQ represents the final qualification of simultaneously.
equipment or system. This incorporates a 5. Stop the stopwatch when the lower
range of testing to simulate your production meniscus reaches the 10 ml mark on the
process options and provide assurance that flask.
systems and operating documentation are 6. Record the elapsed time.
capable of subsequent process validation 7. Similarly check the flow for 1.0 ml/m
activities. It is used to establish and or and 2.0 ml/m.
confirm; Acceptance criteria: The time taken to
1. Definition of performance criteria and collect the water should be with in ± 2.0% of
test procedures. the actual value.
2. Selection of critical parameters, with
predefined specifications.

90
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

PQ Relationships

Table 1: I.Q And O.Q And P.Q Report

System Name:
System ID No.:
Installation site:
Performer_______________Signature _______________Date:______

Reviewer_______________Signature________________Date:______

Manager_______________Signature________________Date:______

Table 2: Flow Rate Accuracy

Set Flow Actual time required to collect Acceptance criteria (in m)


up to the mark in m
0.5 ml/m 20.0 19.6 – 20.4
1.0 ml/m 10.0 9.8 – 10.2
4.9 – 5.1
2.0 ml/m 5.0

91
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

Table 3: Chromatographic Conditions for System Precision

Column C18,150mm×4.6 mm,5µm


Flow rate
Detection 1.0 ml/m
injection volume
Run time UV at 272 nm
Column oven temperature
20 µl
Mobile phase
15 m

25°C ± 2°C

Water: Methanol (70:30v/v)


Injector Accuracy: 10ml of this solution into a 100ml
1. Connect the pump and detector inlet volumetric flask and dilute to the volume
with union. with mobile phase.
2. Prepare mobile phase consisting of a Procedure: Inject blank, followed by
mixture of water and Methanol (70:30 v/v) standard preparation in 6 replicates. Note
3. Set a flow rate of 0.5 ml/m and a run down the areas and retention times.
time of 1 m. Now calculate the %RSD of retention time
4. Set the column temperature at 25± and peak areas for 6 replicates injections.
2°C. Acceptance criteria: The %RSD of
5. Fill a standard HPLC vial to 2/3rd with retention time & peak area should be <1.0%.
Milli-Q water. Seal the vial properly with a Wavelength Accuracy:
cap. Procedure: Create and instrument method
6. Weigh the vial and record the weight with a wavelength in nm and inject blank,
as W1 grams. followed by Standard preparation and note
7. Place the vial in the chromatographic down the height or absorbance.
system and perform 6 injections of 50µl Acceptance criteria: The maximum
volume from this vial. absorbance should be ±2nm.
8. Weigh the vial again and note the PDA Detector Accuracy:
weigh after the injections as W2 grams. Select 3D mode and set the wavelength
Calculate the mean volume injected per range as 200-400nm.Inject 20 µl of standard
injection as follows: preparation once into the chromatographic
Mean injected volume (µl) = (W1 – W2) system. Extract and record the
×100/6 chromatograms at wavelengths of 202 to
Acceptance criteria: The mean injected 208nm with an interval of 1nm and at 269 to
volume should be 50.0±1.0 µl. 275 nm with an interval of 1nm.Note down
System Precision: the height or absorbance.
Standard Preparation: Accurately weigh Acceptance criteria: The maximum
and transfer about 60mg of Caffeine into a absorbance should be at 205±2nm and
100ml volumetric flask. Dissolve and dilute 272±2nm.
to the volume with mobile phase. Transfer

92
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

Detector Linearity: areas. Plot a curve for the volume injected


Standard Preparation: Accurately weigh Vs peak area.
and transfer about 60mg of Caffeine into a Acceptance criteria: The plot should be
100ml volumetric flask. Dissolve and dilute linear and regression coefficient (R2) should
to the volume with mobile phase. not be less than 0.99.
 Detector linearity solution 1(0.06 Gradient performance check:
mg/ml): Transfer 10ml of Standard Add 5ml of acetone to 1000ml of methanol
Preparation into a 100ml volumetric flask filter and degas. Connect the pump and
and dilute to the volume with mobile phase detector inlet with union. Set the detector
 Detector linearity solution 2(0.048 wave length at 254 nm.
mg/ml): Transfer 8ml of Standard Place Channels A and C in methanol and
Preparation into a 100ml volumetric flask channel B and D in 0.5% acetone in
and dilute to the volume with mobile phase. methanol. Set binary gradient with a total
 Detector linearity solution 3(0.03 flow rate of 2.0ml/m.Set gradient program
mg/ml): Transfer 5ml of Standard as shown below for channels A, B and C, D
Preparation into a 100ml volumetric flask individually.
and dilute to the volume with mobile phase.
 Detector linearity solution 4(0.24 Table 4: Gradient Performance Check:
mg/ml): Transfer 4ml of Standard
Preparation into a 100ml volumetric flask Time A or C B or D (0.5%
and dilute to the volume with mobile phase. (m) (Methanol) Acetone in
 Detector linearity solution 5(0.012 Methanol)
mg/ml): Transfer 2ml of Standard Initial 100.0 0.0
Preparation into a 100ml volumetric flask
and dilute to the volume with mobile phase. 4.00 100.0 0.0
Procedure: Inject blank, followed by
Detector linearity solutions and record the 4.01 90.0 10.0
peak responses of Caffeine standard plot
between the concentration Vs the peak 8.00 90.0 10.0
responses.
Acceptance criteria: The plot should be 8.01 100.0 0.0
linear and regression coefficient (R2) should
not be less than 0.99. 12.00 100.0 0.0
Injector Linearity:
Standard Preparation: Accurately weigh 12.01 75.0 25.0
and transfer about 60mg of Caffeine into a
100ml volumetric flask. Dissolve and dilute 16.00 75.0 25.0
to the volume with mobile phase.
Transfer 10ml of Standard Preparation into a 16.01 100.0 0.0
100ml volumetric flask and dilute to the
volume with mobile phase. 20.00 100.0 0.0
Procedure: Inject 5 µl of the mobile phase
as blank injection. Purge all the channels at a flow rate of
Inject 5 µl, 10 µl, 20 µl, 50 µl and 80 µl of 2ml/m for about 5 m. Set the flow rate at
the Standard Preparation and record the peak 2.0ml/m and wait until the base line is
stable. Set the gradient profile for A and B

93
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

and run the gradient profile by injecting Flow rate accuracy:


“0.0”volume of methanol. 1. Connect the digital flow meter to the
Record the height of the peaks. Consider the detector outlet port.
height of the peak resulting from B at 100% 2. Set the carrier gas (Helium) flow and
concentration as 100 and calculate the wait till it reaches the set flow.
percentage height of other peaks. Perform 3. Note the observed flow in replicate.
the gradient performance check similarly for 4. Repeat the procedure for other carrier
channels C and D. gases such as Hydrogen and Air.
Consider the height of the peak resulting 5. Record the result in GC calibration
from D at 100% concentration as 100 and protocol.
calculate the percentage height of other Acceptance criteria: The flow rate of
peaks. carrier gas should be ±10% of set flow.
Calculations:
Height (%) of B/D =Height of B/D peak Table 5: Flow Rate Accuracy:
×100/Height of full scale peak S.No. Carrier Acceptance
Acceptance criteria: The calculated gas criteria in
percentage composition (Height (%) should ml/m
be with in±1.0% of the set composition. 1. Helium 125
Column Oven Temperature Accuracy: 2. Hydrogen 40
It is evaluated with a calibrated digital 3. Air 400
thermometer at 30°Cand 60°C .Place the
thermometer probe in the column oven and Column Oven Temperature Accuracy:
set the column oven temperature at 1. Connect the column to the detector port.
30°C.Wait till the temperature stabilizes. 2. Place the thermometer probe in the
Record the temperature displayed on the column oven and set the column oven
thermometer. Similarly performs the column temperature at 40°C.Wait till the
oven temperature accuracy test at 60°C. temperature stabilizes.
Acceptance criteria: The resulting oven 3. Note the observed temperature as read
temperature from the thermometer display by the probe in triplicate over a period of 10
should be within ±2°C of the set m.
temperature. 4. Repeat the procedure for 100°C, 150°C
NOTE: a) For oven Temperature Accuracy, and 190°C.
Chromatographic conditions and mobile Acceptance criteria: The resulting oven
phase refer to system precision test. temperature from the thermometer display
should be within ±2°C of the set temperature
Calibration of Gas Chromatogrphy: System Precision:
Various Calibration parameters are: Preparation of Standard solution:
 Flow rate accuracy Transfer 20 ml of Methanol, Ethanol and
 Column oven temperature accuracy Acetone into 100ml volumetric flask and
 System precision make up with Ethyl acetate
 System precision for head space auto Procedure: Inject blank, followed by
sampler Standard preparation in 6 replicates. Note
 Detector linearity down the areas and Retention times.
 Detector noise and drift test

94
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

Table 6: Chromatographic Conditions for System Precision

Column 30m × 0.32mm,1.8µ,DB-624


Detector Flame ionization detector
Injector temperature 180°C
Detector temperature 250°C
Flow mode Pressure
Carrier Gas flow rate Helium 25 kpa
Oven program 50°C(hold 5 m) raise to 10°0C
Split ratio 1:10
Injection volume 0.2 µl
Hydrogen flow 40 ml/m
Air flow 400 ml/m

Table 7: Chromatographic Conditions For Head Space Auto Sampler

Column 30m × 0.32mm,1.8µ,DB-624


Detector Flame ionization detector
Injector temperature 220°C
Detector temperature 260°C
Flow mode Pressure
Carrier Gas flow rate Helium 25 kpa
Oven program 40°C(hold 5 m) raise to 200°C(hold 5 m)
Split ratio 1:10
Injection volume 0.2 µl
Hydrogen flow 40 ml/m
Air flow 400 ml/m

Table 8: Head Space Conditions

Vial equilibrium 22 m
Vial pressure 0.5 m
Loop fill 0.5 m
Loop equilibrium 0.05 m
Inject 1.00 m
GC cycle time 38 m
Oven temperature 80°C
Loop temperature 100°C
Vial pressure 10.8 psi

Acceptance criteria: The %RSD of System precision for head space auto
retention time should be not more than sampler:
1.0%& peak area should be not more than Preparation of standard solution: Prepare
5.0%. a standard mixture solution by taking
Methylene dichloride (0.6g), Chloroform

95
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

(0.06g),Trichloroethane (0.08g),1,4-Dioxane Detector Noise and Drift Test:


(0.38g) in 50ml volumetric flask containing After GC is ready run the system up to 15 m
about 40ml of Dimethyl formamide.Finally through single run. After completion of run
makeup to volume with DMF(Solution-A). calculate noise and drift through software.
Procedure: Take 0.5 ml of standard Acceptance criteria:
solution-A in 6 different vials and seal with Noise NMT: 100 µV
septum, then magnetic caps and crimp. Place Drift NMT: 2500 µV/hr
these vials on head space sampler; prepare a Calibration of UV-Visible
blank vial also. Load the vials in head space spectrophotometer:
sampler tray. Blank vials followed by the Spectral calibration: visible spectral region
standard vials.  Ensure-the socket of the power cord of
Acceptance criteria: The %RSD of the instrument is inserted properly
retention time should be NMT 1.0%& peak -cuvettes are clean
area should be NMT 15.0%.  Switch ON the instrument. Allow 15m
Detector linearity: to warm up.
Preparation of standard solutions:  Keep dummy cuvette in position of
 Detector linearity solution A: Transfer sample holder.
10ml of each Methanol, Ethanol and  Set the λ to 485 nm and press %T
Acetone into a 100ml volumetric flask and button.
dilute to the volume with Ethyl acetate.  Press 0%T in appropriate direction to
 Detector linearity solution B: Transfer adjust 0.00 reading on read out.
15ml of each Methanol, Ethanol and  Now remove dummy cuvette from
Acetone into a 100ml volumetric flask and sample holder. Close the lid.
dilute to the volume with Ethyl acetate.  By adjusting coarse and fine control set
 Detector linearity solution C: Transfer a reading of around 80.0 on read out
20ml of each Methanol, Ethanol and  Now set the value of wavelengths in
Acetone into a 100ml volumetric flask and increments of 0.1 nm up to λ of 487 nm and
dilute to the volume with Ethyl acetate. read the value of %T at each increment of λ
 Detector linearity solution D: Transfer  Draw a curve %T Vs λ.
25ml of each Methanol, Ethanol and  If the peak value of %T is occurring at a
Acetone into a 100ml volumetric flask and λ 486.1 ± 0.5 nm, the spectral calibration of
dilute to the volume with Ethyl acetate. the instrument in the visible spectral region
 Detector linearity solution E: Transfer is proper.
30ml of each Methanol, Ethanol and  This can be confirmed by repeating the
Acetone into a 100ml volumetric flask and above steps with a maximum value of %T of
dilute to the volume with Ethyl acetate. around 30.0 on the read out and λ setting
Procedure: Inject blank, followed by from 655 to 657 nm.
Detector linearity solutions and record the  If the maximum %T is obtained at a λ
peak responses .Draw a standard plot 656.2 ± 0.5 nm, the spectral calibration of
between the concentrations Vs the peak
the instrument in the visible spectral region
responses. is confirmed to be proper.
Acceptance criteria: The plot should be Spectral calibration: U.Vspectral region:
linear and regression coefficient (R2) should
 Keep blank (distilled water) filled
not be less than 0.99.
cuvette and sample (benzene vapor) filled
cuvette.

96
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

Table 9: Chromatographic Conditions For Detector Linearity

Column 30m ×0.32mm,1.8µ,DB-624


Detector Flame ionization detector
Injector temperature 180°C
Detector temperature 250°C
Flow mode Pressure
Carrier Gas flow rate Helium 25 kpa
Oven program 50°C(hold 5 m) raise to 100°C
Split ratio 1:10
Injection volume 0.2 µl
Hydrogen flow 40 ml/m
Air flow 400 ml/m
 Set the λ to 253 nm and press  Place blank0.1N H2SO4 cuvette and
absorbance button. 60ppm K2Cr2O7 as sample
 Adjust blank to 0.000 on the read out by  Set λ exactly to 257 nm, if the value of
using coarse and fine adjustment Absorbance of sample at the set λ is
 Now place the sample into optical path, 0.864±0.005, the instrument is measuring
value of Absorbance of sample at the λ set Absorbance properly.
appears on the read out. %Transmittance:
 Again set the values of wavelengths As the value of %T is delivered from
increments of 0.1nm up to a λ of Absorbance itself, if the instrument is
255nm.Measure the A at each increment. If measuring Absorbance properly it is deemed
maximum A is obtained at λ 253.9 ± 0.5nm that it measures %T properly.
the “spectral calibration” of the instrument Concentration:
in U.V region is confirmed to be proper.  Place blank 0.1N H2SO4 cuvette and 60
Photometric Calibration: ppm K2Cr2O7 (0.06006g/l of 0.1NH2SO4) as
Absorbance: Visible region standard and 20 ppm (0.02002 g/l of 0.1N
 Place dummy cuvette in sample holder H2SO4) as sample.
and set %T to “zero”. Now remove dummy  Press “Concentration” push button and
cuvette, by using fine & coarse control set a adjust Concentration control to 600 for
reading exactly 40.0 on the read out. standard on read out.
 Press Absorbance push button. If the  Now place sample holder into optical
maximum absorbance obtained at λ of path, if the value o f Concentration
485nm is 0.398 ± 0.002, the photometric appearing on the read out for sample is
calibration of instrument is confirmed to be 200±5, the instrument is measuring
proper. “concentration” properly.
 To confirm, repeat above steps, and set Calibration of Wavelength:
10.00 on read out a) Holmium filter: For routine calibrations,
 Press Absorbance botton.If the λ at 485 holmium filter is satisfactory. Record the
nm is 1.000±0.002 then it is confirmed the absorption spectrum from 500 to 230 nm
photometric performance in the visible using slowest scan speed and narrowest slit
region is proper. setting. Identify 3 fused absorption bands
Absorbance: U.V region

97
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

Table 10: Chromatographic Conditions

Column 15m ×0.53mm,3.0µ,DB-1


Detector Flame ionization detector
Injector temperature 150°C
Detector temperature 200°C
Flow mode Pressure
Septum purge flow 3 ml/m
Oven program 60°C
Split ratio 1:10
Hydrogen flow 40 ml/m
Air flow 400 ml/m
Run time 15 m
centered on 452.2nm and single band at maximum wave length is observed at 370
around 360.9nm.Instruments with accurately nm.
calibrated λ scales will show λ max at 453.2, Limit of stray light:
418.4, 360.9, 287.5, 279.4 and 241.5nm. Weigh accurately 1.2g of dried Potassium
b) Holmium per chlorate solution: Prepare chloride in 100 ml volumetric flask and
a solution of Holmium (III) per chlorate by makeup to mark with Double distilled water.
dissolving 0.5g of holmium oxide in 2.4 ml Measure the absorbance at 200 nm.
perchloric acid (72% AR grade) by warming Acceptance criteria: Tolerance limit NLT
gently and diluting to 10ml with water. 2.0
Record the absorption spectrum from 500- Resolution:
230nm.The wavelengths of principal bands Prepare 0.02%v/v solution of Toluene and
(Absorbance- 0.4) should be 485.8, 450.8, make up with Hexane. Scan the wavelength
416.3, 361.5, 287.1, 278.7, 241.1 nm. from 250 to 280nm.Maximum absorbance is
c) Discharge lamps: A low pressure 269 nm and Minimum absorbance is 266nm
discharge lamp is suitable. Record the Acceptance criteria: Ratio limit NLT 1.5
transmission spectrum from 600 to 240 nm Photometric linearity:
of Mercury lamp place near the entrance to  Weigh accurately 100mg of Potassium
monochromator, using minimum slit setting chromate in 100ml volumetric flask and
and slowest scan speed. The principal dissolve in 0.05N Potassium hydroxide
emission lines of Mercury are at 579.0, solution. Make up with the same solvent.
576.9, 546.1, 435.8, 404.5, 364.9 and 253.7  From the above solution take 20ml and
nm. make up to 500ml with 0.05N Potassium
d) Prepare standard solution by dissolve 100 hydroxide solution.
mg of Potassium dichromate in 0.05N  Now prepare dilution of 4,8,16,24,32
Potassium hydroxide solution in 100ml µg/ml
volumetric flask. Make up to volume with  Measure the absorbance at 370nm using
the same. From the standard solution take blank.
20ml and make up to 500ml with 0.05N Acceptance criteria: The plot should be
Potassium hydroxide solution. Now scan the linear and regression coefficient (R2) should
wave length from 340 to 400nm using blank NLT 0.999.
0.05N Potassium hydroxide solution. The

98
D. Gowrisankar et al / J Biomed Sci and Res., Vol 2 (2), 2010,89-99

Table 11: Schedule for Calibration/Inspection of Some Major Instruments


INSTRUMENT INTERVAL (MONTHS)
HPLC 3±7 days
Gas Chromatography 3±7days
UV-Visible spectrophotometer Monthly once ±3days
IR spectrophotometer Monthly once ±3days
NMR spectrophotometer 6 ±15days
Flourimeter 3±7 days
Polarimeter Monthly once ±3days
PH meter daily
Analytical balance daily
Abbreviations: to ensure that equipment is calibrated, clean
HPLC : High Performance Liquid and suitable for its intended use.
Chromatography Procedures for the use of facilities should
GC : Gas Chromatograph ensure that materials are handled in a
ICH : International Conference on manner that minimizes the risk of
Harmonization contamination and cross-contamination. So
C18 : Octadecyl silane column validation and calibration is very important
DMF : Dimethyl formamide for analytical instruments.
DB-624: 6% Cyano Propyl Phenyl- References:
94%Dimethyl Polysiloxane [1] Lloyd R.Snyder,Joseph J.Kirkland, Joseph
NMR : Nuclear Magnetic Resonance L.Glanjch.Practical, HPLC method
development, A Wiley – interscience
IR : Infra red publication,1997.
The calibration of the entire analytical [2] USP 30 NF 25 “The official compendia of
instrument or its components (which ever is standards”, Asian edition, 2007, vol: 1.
appropriate) should be performed after any [3] A.H.Beckett, J.B.Stenlake, Practical
major maintenance. Pharmaceutical chemistry, CBS Publishers
and distributors, IV-Edition, part two, 2004,
Conclusion: During all phases of clinical p.no.326.
development, including the use of small- [4] Y.Anjaneyulu, R.Marayya, Quality Assurance
scale facilities or laboratories to and Quality Management in Pharmaceutical
manufacture batches of APIs for use in industry, Pharmabook Syndicate, 2005.
clinical trials, procedures should be in place [5] ICH Guidelines Q2BValidation of Analytical
Procedures: Methodology (1996).

99

You might also like