Professional Documents
Culture Documents
Purpose. The purpose of this study was to evaluate the effects of cement color and thickness on the shade matching of MLD
restorations luted on silver-palladium (Ag-Pd) foundations.
Material and methods. Fifteen 1.5-mm-thick ceramic specimens were made from shade A1 LT lithium disilicate blocks. Five resin
cements with different colors and opacities (Multilink Automix white opaque, Multilink Automix yellow, Nexus3 white opaque,
Nexus3 white, Nexus3 yellow) of 3 thicknesses (300 mm, 100 mm, 50 mm) were sequentially luted to a roughened Ag-Pd alloy
foundation restoration block. Five shade measurements were made with a portable spectrophotometer after optically connecting the
ceramic specimen to each cement foundation block. The color differences (DE) between each specimen assembly and the target
block (a 141412-mm shade A1 LT crystalized e.max block) were recorded with the CIE (Commission internationale de l’éclairage)
L*a*b* system. Clinical significance was determined by comparing color differences to perceptibility and acceptability thresholds
with the 1-sample t test (a¼.05).
Results. Both cement color and cement thickness significantly affected the mean values of the color difference (DE) of lithium
disilicate ceramic assemblies to the target block (P<.001). Among the 5 cements tested, the lowest mean DE was observed for Nexus3
white opaque cement. Overall, the combination of Nexus3 white opaque cement and 100-mm thickness led to the lowest mean DE.
The means of DE were below the clinically perceptible level (DE<2.6) for combinations of Nexus3 white opaque of 100-mm and 300-
mm cement thicknesses (P<.001 and P¼.022, respectively). The means of DE were below the clinically acceptable level (DE<5.5) for
the following combinations: Nexus3 white opaque of 50-mm thickness, Nexus3 white, and Nexus3 yellow for all tested cement
thicknesses and Multilink white opaque or Multilink yellow of either 50-mm or 100-mm cement thickness (P<.001). The means of DE
were above the clinically acceptable level (DE>5.5) for Multilink white opaque and Multilink yellow of 300-mm cement thickness.
Conclusions. The colors of machinable lithium disilicate ceramic restorations placed on Ag-Pd foundation were affected by
both the color and thickness of cements. Among the 5 cements tested in this study, Nexus3 white opaque cement of 100 mm
or 300 mm yielded shade matches below the clinical perceptible threshold (DE<2.6) relative to the target block. (J Prosthet
Dent 2014;111:42-50)
Clinical Implications
White opaque cements are more effective in masking the dark color
of silver-palladium alloy and improve the resultant shade match
of machinable lithium disilicate restorations. Increasing the cement
thickness above 100 mm did not improve the shade match. Nexus3
white opaque at all thicknesses and Multilink white opaque at 50 mm
resulted in the best shade matches.
a
Assistant Professor, Department of Restorative Dentistry, Section of Fixed Prosthodontics, Southern Illinois University, School
of Dental Medicine.
b
Professor, Department of Mathematics and Statistics, Southern Illinois University, Edwardsville.
c
Associate Professor, Chair, Department of Restorative Dentistry; and Section Head, Fixed Prosthodontics, Southern Illinois
University School of Dental Medicine.
Multilink Ivoclar Vivadent AG White opaque (MWO); 50, 100, 300 mm MWO: N74838;
automix yellow (MLY) MLY: R06513
Nexus3 Kerr Corp White opaque (NWO); 50, 100, 300 mm NWO: 4458298;
white (NXW); yellow (NXY) NXW: 4464989; NXY: 4464987
DL, mm
50 -1.22 (0.13) -3.54 (0.24) -0.87 (0.29) -3.08 (0.26) -3.43 (0.29)
100 -3.14 (0.24) 3.33 (0.32) -3.72 (0.25) -0.03 (0.46) -2.86 (0.22) -3.24 (0.29)
300 6.74 (0.35) -4.88 (0.26) 2.23 (0.23) -2.72 (0.23) -2.73 (0.24)
Da, mm
50 -0.50 (0.04) 0.29 (0.07) -0.04 (0.06) 0.25 (0.07) 0.29 (0.06)
100 0.12 (0.07) -0.60 (0.06) 0.18 (0.07) -0.17 (0.06) 0.40 (0.05) 0.36 (0.06)
300 -0.21 (0.06) -0.30 (0.04) -0.31 (0.07) 0.01 (0.06) 0.04 (0.05)
Db, mm
50 -2.62 (0.26) -3.01 (0.19) -2.41 (0.22) -2.96 (0.17) -3.25 (0.18)
100 -3.63 (0.28) 0.31 (0.27) -2.89 (0.16) -2.15 (0.27) -2.60 (0.16) -3.26 (0.20)
300 3.41 (0.19) -5.41 (0.17) -0.91 (0.34) -3.04 (0.18) -3.03 (0.20)
MWO, Multilink white opaque; MLY, Multilink yellow; NWO, Nexus3 white opaque; NXW, Nexus3 white; NXY, Nexus3 yellow.
Values are given as mean (SD).
Table IV. Results of 2-way ANOVA for mean DE values of combinations of performed to assess the effects of
different cement color and cement thicknessa cement color and thickness on the
resultant DE. Post hoc comparisons
Sum of Mean were performed to group the effect of
Source Squares df Square F P the cements used and their thicknesses
on the resultant DE. Clinical significance
Cement used 235 4 58.8 1197 <.001
was determined by comparing color
Cement thickness 82.2 2 41.1 836 <.001 differences to the perceptibility and
Interaction 186 8 23.3 473 <.001 acceptability thresholds as determined
Error 10.3 210 0.05 in a study by Douglas et al41 (DE>2.6
was considered clinically perceptible,
Total 514 224
whereas DE>5.5 was considered clini-
a
S¼0.22; R ¼ 97.99%; R (adjusted)¼97.86%.
2 2
cally unacceptable).
device, the next ceramic specimen was foundation block with immersion oil
placed on the cement surface after only. Throughout the study, the shade RESULTS
adding a drop of immersion oil. After was measured by the same operator with
the color of the 15 ceramic-cement- the same portable spectrophotometer. The mean values of DE and the dif-
foundation assemblies was measured, The CIE-L*a*b* color space from each ference of each CIE coordinate (DL, Da,
the edge of the cement layer was loos- color measurement was calculated and Db) between each ceramic specimen
ened with a chisel and cleanly peeled recorded in terms of the 3 CIE coordi- assembly and the target block are listed
off from the foundation block with nate values (L*, a*, b*). Color differ- in Tables II and III. Results from the
tweezers. The foundation block was ences between each ceramic specimen- 2-way ANOVA demonstrated that both
evaluated under magnification (4.5) cement-foundation assembly and the the color and thickness of cement
to ensure a clean cement-bonding sur- target block (DE) were calculated as the significantly affected the color differences
face. The next designated thickness of average of 75 determinations (5 mea- (DE) of MLD ceramic-Ag-Pd assemblies
cement was bonded to the foundation surements for each ceramic specimen- (P<.001). A significant interaction was
block as previously described. cement-foundation assembly and 15 present between the color and thickness
The same procedures were repeated ceramic specimens with a ceramic of the cement (Table IV). The general
for 5 cements at 3 cement thicknesses. thickness of 1.5 mm). linear model was used to further inves-
As a reference, the color of the ceramic Statistical analyses were performed tigate the data (Table V). The estimated
Ag-Pd foundation assembly was me- with statistical software (SPSS, v12.0.1; value of the color-cement thickness
asured without cement by optically SPSS Inc). Two-way ANOVA (a¼.05) of each cement was calculated by
connecting the ceramic specimens to the and the general model procedures were subtracting the value of the intercept
Niu et al
46 Volume 111 Issue 1
Table V. General linear analysis for mean DE values of combinations of different color and thickness varied (Fig. 2). Also,
cement colors and thicknessesa the results from the grouping of the
means of DE from different cement
Sum of Mean
color-thickness combinations with the
Source df Squares Square F P Tukey method are presented in Table VII.
Except for the MWO of 300 mm, ceramic
Model 14 2525 180 2998 <.001
assemblies with Nexus3 white opaque
Error 1110 67 0.06
(NWO) cement (MWO-NWO) resulted
Corrected total 1124 2592 in the closest color match to the target
R ¼0.97; coefficient variance¼5.78; root mean standard of error¼0.25; standard error of mean¼4.24.
a 2 block. Among the 5 cements tested,
NWO yielded the lowest mean DE.
Table VI. Overall, the combination of Nexus3
Standard t white opaque cement and 100 mm
thickness resulted in the lowest mean DE
Parameter Estimate Error Value P
(Tables II, VIII; Fig. 2).
Intercept 2.94 0.03 104.05 <.001 For ceramic assemblies with MWO,
the lowest mean value of DE was
MLY 50 1.74 0.04 43.32 <.001
observed at 50-mm cement film thick-
NWO 50 -0.27 0.04 6.81 <.001
ness, followed by 100 mm, then 300 mm
NXW 50 1.34 0.04 33.51 <.001 (Table II, Fig. 2). Based on the results of
NXY 50 1.80 0.04 44.91 <.001 the 1-sample t tests, it was statistically
MWO 100 0.47 0.04 11.65 <.001 significant that the mean DE was below
MLY 100 1.76 0.04 44.00 <.001 the clinically perceptible level (DE<2.6)
NWO 100 -0.77 0.04 19.18 <.001
for the combinations of NWO with 100-
mm and 300-mm cement film thickness
NXW 100 0.95 0.04 23.79 <.001
(P<.001, P¼.022) (Table VIII). The
NXY 100 1.68 0.04 42.05 <.001 mean DE for all the other cement
MWO 300 4.62 0.04 115.30 <.001 and thickness combinations, except
MLY 300 4.35 0.04 108.64 <.001 for MWO and MLY with 300-mm
NWO 300 -0.50 0.04 12.50 <.001 cement thickness, was within clinically
NXW 300 1.14 0.04 28.45 <.001 acceptable levels (DE<5.5). The mean
DE was above the clinically acceptable
NXY 300 1.15 0.04 28.72 <.001
level (DE>5.5) for MWO and MLY of
MWO, Multilink white opaque; MLY, Multilink yellow; NWO, Nexus3 white opaque; NXW, Nexus3 300-mm cement thickness (P<.001).
white; NXY, Nexus3 yellow.
** Intercept: MWO 50 ¼ 2.94.
DISCUSSION
thickness combination minus ΔE for
Niu et al
50 Volume 111 Issue 1
43. Lehmann KM, Devigus A, Igiel C, 46. Lasserre JF, Pop-Ciutrila IS, Colosi HA. Corresponding author:
Wentaschek S, Azar MS, Scheller H. Repeat- A comparison between a new visual Dr Yi-Hua (Eva) Niu
ability of color-measuring devices. Eur J method of color matching by intraoral Shin Kong Wu Ho-Su Memorial Hospital
Esthet Dent 2011;6:428-35. camera and conventional visual and Department of Dentistry
44. Llena C, Lozano E, Amengual J, Forner L. spectrometric methods. J Dent 95, Wen Chang Road, Shih Lin district, Taipei
Reliability of two color selection 2011;39(suppl 3):e29-36. TAIWAN
devices in matching and measuring tooth co- 47. Akgungor G, Akkayan B, Gaucher H. Influ- E-mail: niueva@gmail.com
lor. J Contemp Dent Pract 2011;12:19-23. ence of ceramic thickness and polymerization
45. Khashayar G, Dozic A, Kleverlaan CJ, mode a resin luting agent on early bond Copyright ª 2014 by the Editorial Council for
Feilzer AJ. Data comparison between two strength and durability with a lithium dis- The Journal of Prosthetic Dentistry.
dental spectrophotometers. Oper Dent ilicate ceramic system. J Prosthet Dent
2012;37:12-20. 2005;94:234-41.
Access to The Journal of Prosthetic Dentistry Online is reserved for print subscribers!
Full-text access to The Journal of Prosthetic Dentistry Online is available for all print subscribers. To activate your individu-
al online subscription, please visit The Journal of Prosthetic Dentistry Online. Point your browser to http://www.journals.
elsevierhealth.com/periodicals/ympr/home, follow the prompts to activate online access here, and follow the instruc-
tions. To activate your account, you will need your subscriber account number, which you can find on your mailing
label ( note: the number of digits in your subscriber account number varies from 6 to 10). See the example below in
which the subscriber account number has been circled.
Sample mailing label
Personal subscriptions to The Journal of Prosthetic Dentistry Online are for individual use only and may not be trans-
ferred. Use of The Journal of Prosthetic Dentistry Online is subject to agreement to the terms and conditions as indicated
online.