You are on page 1of 9

Color match of machinable lithium

disilicate ceramics: Effects of cement


color and thickness
Eva Niu, DDS, MS,a Marcus Agustin, PhD,b and
R. Duane Douglas, DMD, MSc
Southern Illinois University School of Dental Medicine, Alton, Ill; Shin
Kong Wu Ho-Su Memorial Hospital, Taipei, Taiwan
Statement of problem. The effects of cement color and thickness on the color of machinable lithium disilicate (MLD) ceramic
luted on metal foundation restorations is unknown.

Purpose. The purpose of this study was to evaluate the effects of cement color and thickness on the shade matching of MLD
restorations luted on silver-palladium (Ag-Pd) foundations.

Material and methods. Fifteen 1.5-mm-thick ceramic specimens were made from shade A1 LT lithium disilicate blocks. Five resin
cements with different colors and opacities (Multilink Automix white opaque, Multilink Automix yellow, Nexus3 white opaque,
Nexus3 white, Nexus3 yellow) of 3 thicknesses (300 mm, 100 mm, 50 mm) were sequentially luted to a roughened Ag-Pd alloy
foundation restoration block. Five shade measurements were made with a portable spectrophotometer after optically connecting the
ceramic specimen to each cement foundation block. The color differences (DE) between each specimen assembly and the target
block (a 141412-mm shade A1 LT crystalized e.max block) were recorded with the CIE (Commission internationale de l’éclairage)
L*a*b* system. Clinical significance was determined by comparing color differences to perceptibility and acceptability thresholds
with the 1-sample t test (a¼.05).

Results. Both cement color and cement thickness significantly affected the mean values of the color difference (DE) of lithium
disilicate ceramic assemblies to the target block (P<.001). Among the 5 cements tested, the lowest mean DE was observed for Nexus3
white opaque cement. Overall, the combination of Nexus3 white opaque cement and 100-mm thickness led to the lowest mean DE.
The means of DE were below the clinically perceptible level (DE<2.6) for combinations of Nexus3 white opaque of 100-mm and 300-
mm cement thicknesses (P<.001 and P¼.022, respectively). The means of DE were below the clinically acceptable level (DE<5.5) for
the following combinations: Nexus3 white opaque of 50-mm thickness, Nexus3 white, and Nexus3 yellow for all tested cement
thicknesses and Multilink white opaque or Multilink yellow of either 50-mm or 100-mm cement thickness (P<.001). The means of DE
were above the clinically acceptable level (DE>5.5) for Multilink white opaque and Multilink yellow of 300-mm cement thickness.

Conclusions. The colors of machinable lithium disilicate ceramic restorations placed on Ag-Pd foundation were affected by
both the color and thickness of cements. Among the 5 cements tested in this study, Nexus3 white opaque cement of 100 mm
or 300 mm yielded shade matches below the clinical perceptible threshold (DE<2.6) relative to the target block. (J Prosthet
Dent 2014;111:42-50)

Clinical Implications
White opaque cements are more effective in masking the dark color
of silver-palladium alloy and improve the resultant shade match
of machinable lithium disilicate restorations. Increasing the cement
thickness above 100 mm did not improve the shade match. Nexus3
white opaque at all thicknesses and Multilink white opaque at 50 mm
resulted in the best shade matches.

a
Assistant Professor, Department of Restorative Dentistry, Section of Fixed Prosthodontics, Southern Illinois University, School
of Dental Medicine.
b
Professor, Department of Mathematics and Statistics, Southern Illinois University, Edwardsville.
c
Associate Professor, Chair, Department of Restorative Dentistry; and Section Head, Fixed Prosthodontics, Southern Illinois
University School of Dental Medicine.

The Journal of Prosthetic Dentistry Niu et al


January 2014 43
Matching the shade of ceramic res- popularity for both anterior and posterior closer color match between a specimen
torations to the natural dentition is a crowns because of its superior esthetics, and a target color. The threshold DE
challenge in dentistry. Having no sup- strength, abrasion behavior, and chemical value, which is perceptible to half of the
porting metal substructure, ceramic durability.2,24-31 dentist observers is 2.6 DE units, and the
restorations achieve light scattering and Previous studies have demonstrated acceptability tolerance for half of the
transmission that are similar to that of that increasing ceramic restoration dentist observers is 5.5 DE units.41
natural teeth.1,2 However, the trans- thickness to 2.0 mm or more is effective Compared with subjective visual obser-
lucency of ceramic restorations may be in masking the influence of underlying vations, spectrophotometers offer a 33%
detrimental to the perceived color when discolored tooth structure or metal increase in accuracy and a more objec-
these restorations are luted over dis- foundation material.3,13-17 However, in tive match in 93.3% of patients.39,43-46
colored dentin or metal foundation many clinical situations, achieving 2.0 The Vita EasyShade Compact (Vident)
restorations.3-14 mm of axial reduction is not possible is a small, cordless, portable intraoral
The shade perceived for ceramic without encroaching on the pulp and dental spectrophotometer used as an
restorations is affected by illuminants, compromising the strength of the intraoral and extraoral dental shade
types of ceramic, thickness of the remaining tooth structures. In addition, matching device. Recently, it has been
ceramic materials, shades of the cores or such facial reduction results in a prepa- used for dental shade matching
internal tooth structures, staining, sur- ration that is thin incisally. On such research, and several studies have sup-
face textures, polishing, and color and preparations, overmilling is common ported its reliability and accuracy.43-46
opacities of luting agents.3-20 When a and can result in the compromised sta- The purpose of this study was to
ceramic crown is placed on an bility of restorations.32 When it is not evaluate the effects of cement color and
endodontically treated tooth with a possible to attain sufficient ceramic thickness on the resultant color of
metal post and core, the dark color and thickness to mask underlying metal machinable lithium disilicate (MLD)
opacity of metal foundation restorations cores, the manipulation of the color and specimens on Ag-Pd metal cores. Three
may lead to discoloration of the defini- thickness of cement may be the only null hypotheses were evaluated. First, the
tive restorations.8-12 Among the alloys variable available to clinicians to mask color of MLD ceramic restorations would
for cast posts and cores, gold-colored the effect of the underlying core on the not be affected by the color of the cement
alloys cause a negligible effect on the color of the definitive restorations.32-38 used. Second, the color of MLD ceramic
color of semitranslucent glass ceramic However, there is no agreement in the restorations would not be affected by the
when ceramic thickness is 1.5 mm or literature regarding the predictability of cement thickness. Third, no interactions
more.3,9,11,12 However, because of the cement to modify the color of ceramic would be found between the cement
increased cost of gold, alloys such as restorations. Only 1 study was identified color and thickness and the resultant
silver-palladium (Ag-Pd) and base that evaluated the interrelationship be- color of MLD restorations.
metals are now commonly used for cast tween the effect of cement thickness and
posts and cores. Compared with a gold its ability to mask discolored abut- MATERIAL AND METHODS
alloy, these alloys lack a yellow hue and ments.13 This study did not evaluate the
are somewhat darker and bluish. effects of metal foundations. Fifteen 1.5-mm-thick ceramic speci-
Therefore, they may substantially alter Spectrophotometers are among the mens were made from low-translucent
the resultant color of semitranslucent most accurate instruments for shade A1 lithium disilicate computer-
ceramic restorations.8,11,12,21,22 measuring color in dentistry.39-41 These aided design blocks (IPS e.max CAD LT;
Because colors of a foundation cannot instruments measure spectral reflectance Ivoclar Vivadent AG). The sample size
always be chosen by clinicians, ceramic and can express it in terms of 3 coordi- was determined by the authors’ unpub-
types and thickness and cement color and nate values (L*, a*, b*), which locate lished pilot study with try-in paste. No
thickness become important variables in the object’s color within the CIE (Com- power analysis was performed. The
manipulating the color of ceramic resto- mission internationale de l’éclairage) specimens were made sectioning
rations. The masking effect of a ceramic -L*a*b* color space. The L* coordinate ceramic blocks with a slow speed saw
material depends on its translucency, represents the brightness of an object, (Isomet 1000 Saw; Buehler) with a 0.6-
which is determined by the material itself the a* value represents the red (positive mm-thick diamond blade. Each
and the ceramic thickness.1-5,7,9-17 High- value) or green (negative value) chro- sectioned slice was measured with a
strength ceramic substructure materials macity, and the b* value represents the digital micrometer (IP-65 Digimatic
such as alumina- or zirconia-based ce- yellow (positive value) or blue (negative Micrometer; Mituoyo) and adjusted to a
ramics have high opacity and require value) chromacity.39-42 The color differ- thickness of 1.5 mm 5 mm with 400
translucent veneering porcelain to achieve ence (DE) of 2 objects can then be grit SiC paper. The sectioned slices were
adequate shade matches.1,2,23,24 Among determined by comparing the diferences polished with SiC paper (600 and 800
the semitranslucent glass-ceramic sys- between respective coordinate values of grit). The color-measuring surfaces of
tems, lithium disilicate has gained each object.39 A lower DE value means a the ceramic specimens were finely
Niu et al
44 Volume 111 Issue 1
Table I. Cements used
Cement
Manufacturer Colors Thickness Batch Lot No.

Multilink Ivoclar Vivadent AG White opaque (MWO); 50, 100, 300 mm MWO: N74838;
automix yellow (MLY) MLY: R06513

Nexus3 Kerr Corp White opaque (NWO); 50, 100, 300 mm NWO: 4458298;
white (NXW); yellow (NXY) NXW: 4464989; NXY: 4464987

Vivadent AG) according to the manu-


Photopolymerizing
facturer’s recommendations.
light Thimble of A foundation block was made by
digital investing a 12142-mm acrylic resin
micrometer pattern (Duralay; Reliance Dental) with
Ceramic phosphate-bonded investment (Ceram-
Cement
Ag-Pd alloy block igold; Whip Mix Corp) and was cast in
Ag-Pd alloy (Albacas; Jelenko) with an
Thimble of
digital induction casting machine (Fornax T;
micrometer BEGO Dental), according to the alloy
manufacturer’s recommendations. One
Under magnification flat surface of the Ag-Pd foundation
Immersion
block was roughened with 200 grit SiC
oil
paper, rinsed with ethanol, and air-
Cement
dried. Five different color cements from
2 manufacturers were used in the
Spectrophotometer study (Table I). The cement layer was
sequentially bonded to the roughened
Ceramic surface of the Ag-Pd foundation block.
Ag-Pd alloy block Cement thicknesses of 300, 100, and 50
Custom-made specimen aligning index mm were controlled by pressing the sur-
1 Diagram of specimen fabrication and specimen face of a cover glass with the flat end of
assemblies. A, Cement thickness was controlled after photo- a digital micrometer (Digital micrometer
polymerizing cement through cover glass. B, Then, specimen IP-65; Mitutoyo) after loading the mixed
assemblies were placed on custom-made specimen aligning cements onto the foundation block. The
index for color measurements. cement was completely polymerized
with a photopolymerizing machine
Table II. DE of specimen assemblies to target block with different colors and (Optilux 501; Kerr Corp) over the cover
cement thickness glass for 90 seconds. The cover glass
Reference was then carefully separated from the
cement, which left the cement intact on
(without
the foundation block (Fig. 1).
cement) MWO MLY NWO NXW NXY The ceramic specimen was then
optically connected to the cement layer
50 mm 2.94 (0.24) 4.68 (0.18) 2.67 (0.18) 4.29 (0.18) 4.74 (0.17)
with a drop of immersion oil (immer-
100 mm 4.82 (0.18) 3.41(0.33) 4.71 (0.20) 2.21 (0.30) 3.89 (0.16) 4.63 (0.19)
sion oil type A, refractive index¼1.515;
300 mm 7.56 (0.29) 7.30 (0.21) 2.44 (0.28) 4.08 (0.15) 4.09 (0.16) Nikon) without actual bonding to the
MWO, Multilink white opaque; MLY, Multilink yellow; NOW, Nexus3 white opaque; NXW, Nexus3 ceramic surface. Shade measurements
white; NXY, Nexus3 yellow. were performed by positioning the flat
Values are given as mean (SD).
surface of a portable spectrophotom-
polished with 1200 grit SiC paper. One mandrel and served as the target block eter (Vita EasyShade Compact; Vident)
A1 LT e.max CAD block (C14, for the study. Ceramic specimens and against the center of the flat ceramic
121414 mm) was polished with the target block were crystallized in a surface of each specimen assembly.
1200 grit SiC paper after removing the ceramic furnace (Programat CS; Ivoclar After 5 shade measurements with the

The Journal of Prosthetic Dentistry Niu et al


January 2014 45
Table III. DL, Da, Db of specimen assemblies to target block with different colors and cement thickness
Reference
(without cement) MWO MLY NWO NXW NXY

DL, mm
50 -1.22 (0.13) -3.54 (0.24) -0.87 (0.29) -3.08 (0.26) -3.43 (0.29)
100 -3.14 (0.24) 3.33 (0.32) -3.72 (0.25) -0.03 (0.46) -2.86 (0.22) -3.24 (0.29)
300 6.74 (0.35) -4.88 (0.26) 2.23 (0.23) -2.72 (0.23) -2.73 (0.24)

Da, mm
50 -0.50 (0.04) 0.29 (0.07) -0.04 (0.06) 0.25 (0.07) 0.29 (0.06)
100 0.12 (0.07) -0.60 (0.06) 0.18 (0.07) -0.17 (0.06) 0.40 (0.05) 0.36 (0.06)
300 -0.21 (0.06) -0.30 (0.04) -0.31 (0.07) 0.01 (0.06) 0.04 (0.05)

Db, mm
50 -2.62 (0.26) -3.01 (0.19) -2.41 (0.22) -2.96 (0.17) -3.25 (0.18)
100 -3.63 (0.28) 0.31 (0.27) -2.89 (0.16) -2.15 (0.27) -2.60 (0.16) -3.26 (0.20)
300 3.41 (0.19) -5.41 (0.17) -0.91 (0.34) -3.04 (0.18) -3.03 (0.20)

MWO, Multilink white opaque; MLY, Multilink yellow; NWO, Nexus3 white opaque; NXW, Nexus3 white; NXY, Nexus3 yellow.
Values are given as mean (SD).

Table IV. Results of 2-way ANOVA for mean DE values of combinations of performed to assess the effects of
different cement color and cement thicknessa cement color and thickness on the
resultant DE. Post hoc comparisons
Sum of Mean were performed to group the effect of
Source Squares df Square F P the cements used and their thicknesses
on the resultant DE. Clinical significance
Cement used 235 4 58.8 1197 <.001
was determined by comparing color
Cement thickness 82.2 2 41.1 836 <.001 differences to the perceptibility and
Interaction 186 8 23.3 473 <.001 acceptability thresholds as determined
Error 10.3 210 0.05 in a study by Douglas et al41 (DE>2.6
was considered clinically perceptible,
Total 514 224
whereas DE>5.5 was considered clini-
a
S¼0.22; R ¼ 97.99%; R (adjusted)¼97.86%.
2 2
cally unacceptable).
device, the next ceramic specimen was foundation block with immersion oil
placed on the cement surface after only. Throughout the study, the shade RESULTS
adding a drop of immersion oil. After was measured by the same operator with
the color of the 15 ceramic-cement- the same portable spectrophotometer. The mean values of DE and the dif-
foundation assemblies was measured, The CIE-L*a*b* color space from each ference of each CIE coordinate (DL, Da,
the edge of the cement layer was loos- color measurement was calculated and Db) between each ceramic specimen
ened with a chisel and cleanly peeled recorded in terms of the 3 CIE coordi- assembly and the target block are listed
off from the foundation block with nate values (L*, a*, b*). Color differ- in Tables II and III. Results from the
tweezers. The foundation block was ences between each ceramic specimen- 2-way ANOVA demonstrated that both
evaluated under magnification (4.5) cement-foundation assembly and the the color and thickness of cement
to ensure a clean cement-bonding sur- target block (DE) were calculated as the significantly affected the color differences
face. The next designated thickness of average of 75 determinations (5 mea- (DE) of MLD ceramic-Ag-Pd assemblies
cement was bonded to the foundation surements for each ceramic specimen- (P<.001). A significant interaction was
block as previously described. cement-foundation assembly and 15 present between the color and thickness
The same procedures were repeated ceramic specimens with a ceramic of the cement (Table IV). The general
for 5 cements at 3 cement thicknesses. thickness of 1.5 mm). linear model was used to further inves-
As a reference, the color of the ceramic Statistical analyses were performed tigate the data (Table V). The estimated
Ag-Pd foundation assembly was me- with statistical software (SPSS, v12.0.1; value of the color-cement thickness
asured without cement by optically SPSS Inc). Two-way ANOVA (a¼.05) of each cement was calculated by
connecting the ceramic specimens to the and the general model procedures were subtracting the value of the intercept

Niu et al
46 Volume 111 Issue 1
Table V. General linear analysis for mean DE values of combinations of different color and thickness varied (Fig. 2). Also,
cement colors and thicknessesa the results from the grouping of the
means of DE from different cement
Sum of Mean
color-thickness combinations with the
Source df Squares Square F P Tukey method are presented in Table VII.
Except for the MWO of 300 mm, ceramic
Model 14 2525 180 2998 <.001
assemblies with Nexus3 white opaque
Error 1110 67 0.06
(NWO) cement (MWO-NWO) resulted
Corrected total 1124 2592 in the closest color match to the target
R ¼0.97; coefficient variance¼5.78; root mean standard of error¼0.25; standard error of mean¼4.24.
a 2 block. Among the 5 cements tested,
NWO yielded the lowest mean DE.
Table VI. Overall, the combination of Nexus3
Standard t white opaque cement and 100 mm
thickness resulted in the lowest mean DE
Parameter Estimate Error Value P
(Tables II, VIII; Fig. 2).
Intercept 2.94 0.03 104.05 <.001 For ceramic assemblies with MWO,
the lowest mean value of DE was
MLY 50 1.74 0.04 43.32 <.001
observed at 50-mm cement film thick-
NWO 50 -0.27 0.04 6.81 <.001
ness, followed by 100 mm, then 300 mm
NXW 50 1.34 0.04 33.51 <.001 (Table II, Fig. 2). Based on the results of
NXY 50 1.80 0.04 44.91 <.001 the 1-sample t tests, it was statistically
MWO 100 0.47 0.04 11.65 <.001 significant that the mean DE was below
MLY 100 1.76 0.04 44.00 <.001 the clinically perceptible level (DE<2.6)
NWO 100 -0.77 0.04 19.18 <.001
for the combinations of NWO with 100-
mm and 300-mm cement film thickness
NXW 100 0.95 0.04 23.79 <.001
(P<.001, P¼.022) (Table VIII). The
NXY 100 1.68 0.04 42.05 <.001 mean DE for all the other cement
MWO 300 4.62 0.04 115.30 <.001 and thickness combinations, except
MLY 300 4.35 0.04 108.64 <.001 for MWO and MLY with 300-mm
NWO 300 -0.50 0.04 12.50 <.001 cement thickness, was within clinically
NXW 300 1.14 0.04 28.45 <.001 acceptable levels (DE<5.5). The mean
DE was above the clinically acceptable
NXY 300 1.15 0.04 28.72 <.001
level (DE>5.5) for MWO and MLY of
MWO, Multilink white opaque; MLY, Multilink yellow; NWO, Nexus3 white opaque; NXW, Nexus3 300-mm cement thickness (P<.001).
white; NXY, Nexus3 yellow.
** Intercept: MWO 50 ¼ 2.94.
DISCUSSION
thickness combination minus ΔE for

Reference: ΔE for MWO X 50 μm


ΔE for each cement color/cement

5 This study evaluated the masking


effects of 2 commercial resin cements
4
of different colors and thicknesses when
MWO X 50 μm

3 MWO luting MLD ceramic to Ag-Pd founda-


2 MLY tion alloy. Based on the results of this
NWO study, all 3 null hypotheses were rejec-
1 NXW ted. The color differences between the
0 NXY target block and the ceramic assemblies
–1 50 μm 300 μm were affected by both the color and
100 μm
thickness of the resin cements, and an
–2
interaction existed between the cement
color and thickness on the resultant
2 Means of DE in combinations of different cements and color of MLD restorations (Tables IV, V,
film thicknesses subtracted from mean values of DEMWO50. VI; Fig. 2).
(Multilink white opaque [MWO] 50 mm) 4.68-2.94¼1.74 (Table VI). A graph was Several studies reported that Ag-Pd
from the original mean value. For drawn based on the results of the general alloy and base metals significantly re-
example, the estimated value of Multilink linear analysis to demonstrate the duce the L* value of ceramic restora-
yellow (MLY) 50 mm was calculated as changes of color differences as cement tions,8,11,18 although their effect on
The Journal of Prosthetic Dentistry Niu et al
January 2014 47
Table VII. Grouping information with Tukey method yellowness (b*) varied and the a* values
tended to be relatively unchanged. In this
Mean Mean Groupinga study, the control (the MLD ceramic-Ag-
Pd assemblies with no cement) was
MWO300 7.56 A
darker and bluer than the target block
MLY300 7.30 A
(Table III). White opaque cements
NXY50 4.74 B demonstrated better masking ability
MLY100 4.71 B than cements of other colors. Both NWO
MLY50 4.68 B and MWO cements effectively increased
NXY100 4.63 B the L* and b* values of the MLD
NXW50 4.29 C
ceramic-Ag-Pd assemblies and created
better shade matches in the tested
NXY300 4.09 C D
cement thicknesses (Tables II-VII, Fig. 2).
NXW300 4.08 C D
Compared with MLD ceramic-Ag-Pd as-
NXW100 3.90 D semblies with no cement (reference),
MWO100 3.41 E white or yellow cement with low opacity
MWO50 2.94 F (Nexus3 white-Nexus3 yellow [NXW-
NWO50 2.67 F G NXY]) caused minimal change in the L*
and b* values of the specimens and did
NWO300 2.44 G H
not improve the resultant shade match,
NWO 100 2.21 H
regardless of the cement thickness. Yel-
MWO, Multilink white opaque; MLY, Multilink yellow; NWO, Nexus3 white opaque; NXW, Nexus3 low cement with higher opacity (MLY)
white; NXY, Nexus3 yellow. significantly reduced the values of L* and
a
Means that do not share same uppercase letter are significantly different.
b* compared with the control and
further compromised the resultant color
Table VIII. One-sample t test for comparing mean DE values of different match of ceramic assemblies as cement
combinations to clinically perceptible level (2.6 DE units) and clinically acceptable thickness increased. The results of this
level (5.5 DE units) (n¼15)
study concur with those from a color
95% study of lithium disilicate ceramic with
SE Upper try-in paste, which recommended using
white opaque cement for cobalt-
Variable Mean SD Mean Bond T P
chromium (Co-Cr) foundation mate-
Test of m¼2.6 vs <2.6 rial.21 Among cements tested in this
NWO 100 2.21 0.30 0.08 2.34 -5.03 <.001 study, Nexus3 white opaque cement
(NWO) at all thicknesses resulted in the
NWO 300 2.44 0.28 0.07 2.57 -2.22 .022
best color match overall, followed by the
Test of m¼5.5 vs <5.5
MWO with cement thickness of 50 mm
NWO 50 2.67 0.18 0.05 2.75 -61.64 <.001 and 100 mm (Table VII, Fig. 2).
MWO 50 2.94 0.24 0.06 3.05 -41.44 <.001 Although the use of opaque cement
MWO 100 3.41 0.33 0.09 3.56 -24.57 <.001 to mask the dark shade from metal al-
NXW 100 3.90 0.16 0.04 3.97 -38.88 <.001 loys has been suggested,8,21,33 little in-
formation is available regarding the
NXW 300 4.08 0.15 0.04 4.15 -36.72 <.001
appropriate cement thickness to
NXY 300 4.09 0.16 0.04 4.17 -34.37 <.001
improve color match.13 A study by Vichi
NXW 50 4.29 0.18 0.05 4.37 -26.78 <.001 et al13 evaluated the effects of cement
NXY100 4.63 0.19 0.05 4.71 -18.00 <.001 thickness on the color of glass ceramic.
MLY 50 4.68 0.18 0.05 4.76 -17.24 <.001 The researchers reported that differ-
MLY 100 4.71 0.20 0.05 4.80 -15.55 <.001 ences in cement thickness (0.1 or 0.2
NXY 50 4.74 0.17 0.04 4.82 -16.98 <.001 mm) had little effect on the resultant
colors of leucite-reinforced ceramic
Test of m¼5.5 vs >5.5
restorations. However, only translucent
MLY 300 7.30 0.21 0.05 7.20 32.88 <.001
cements of different colors were used
MWO 300 7.56 0.29 0.08 7.43 27.38 <.001 in the study by Vichi et al,13 and dis-
DE, color differences; SD, standard deviation; SE, standard error; MWO, Multilink white opaque; colored substrates did not include
MLY, Multilink yellow; NWO, Nexus3 white opaque; NXW, Nexus3 white; NXY, Nexus3 yellow. metal foundations.
Niu et al
48 Volume 111 Issue 1
Based on the results of the current color-thickness combination with the of the luting agent to the ceramic
study, white opaque cement should be intercept group (MWO 50), the interac- specimens but only to the foundation
used to mask the dark color of Ag-Pd tion between cement color and thickness block. To simulate a ceramic-cement-
alloy to achieve a clinically acceptable was clearly demonstrated in Table IV. In foundation lamination, the bonded
shade match (5.5DE units). Cement this study, for Nexus3 cements (NWO, cement-foundation complex was opti-
thickness for the best color match de- NXW, NXY), increasing cement thickness cally connected to the ceramic with
pends on the color, brightness, and from 100 to 300 mm had minimal influ- immersion oil. However, pilot testing of
opacity of the cements. Increasing the ence on the resultant colors of MLD this methodology demonstrated that
cement thicknesses of Multilink cements ceramic-Ag-Pd assemblies. However, for the use of oil did not affect the color of
substantially affected the resultant color Multilink cements (MWO, MLY), the ceramic assemblies. Third, a
of ceramic restorations, whereas, for increasing cement thickness from 100 mm portable spectrophotometer was used
Nexus3, the influence of thickness was to 300 mm led to an increase in that color to measure colors. Although this spec-
minimal (Tables II, V, VI, VII; Fig. 2). The difference from 3.40 to 7.56 DE units trophotometer has been used in
mean DE values for Nexus3 white opa- and 4.71 to 7.30 DE units (Table III). numerous color studies, it is not
que cement were consistently lower than These changes associated with an considered as accurate as a spectror-
for MWO cement, regardless of the increased cement thickness created a adiometer.41 Color measurements of
different cement thicknesses used. A resultant shade that exceeded the clini- translucent materials made with a small
color difference below the clinical cally acceptable threshold of 5.5 DE window spectrophotometer or color-
perceptible level (DE 2.6) was achieved units. Therefore, with Multilink cements, imeter can be subject to deviations
for Nexus3 white opaque at 100-mm and precise control of cement thickness below caused by the edge-loss effect.39,41 To
300-mm thickness (NWO 100 and NWO 100 mm is imperative for esthetic results. minimize the impact of edge loss, the
300, respectively). For MWO cement, In patients in whom a uniform cement ceramic specimens for the study were
the best color match was achieved at 50- thickness under 100 mm cannot be made at a large dimension (1214
mm thickness, in which a mean color attained (facial undercuts from previous mm) with flat ground surfaces. If edge
difference for ceramic-Ag-Pd assemblies restorations), the use of a cement such as loss affected the results, these errors
to the target block was slightly above the NWO may be indicated because its would be equal for all specimens
clinically perceptible level (DE¼2.94) resultant color is minimally affected by because the same ceramic specimens
(Tables VII, VIII; Fig. 2). Furthermore, varying thickness. These observations were used throughout the study and,
increasing cement thickness from 100 concurred with the findings of a previous therefore, would have no effect on
mm to 300 mm presented little benefit to study, which reported that color and data derived from differential color
the resultant color match for MLD opacity varied among different compos- computations.
ceramic assemblies (Tables II, VII; ite luting cements of like nominal Only shade A1 LT ceramic was used
Fig. 2). In particular, for both Multilink shades.36 If Multilink cements are used to in this study, and the results may not
cements, increasing the cement thick- prepare facial undercuts, then blocking be applicable to other shades and
ness from 100 mm to 300 mm yielded the out the undercut on the tooth or foun- translucencies of MLD ceramics.
poorest color matches. For MWO, this dation with composite resin before digi- Different shades of the same ceramic
was due to its brightness and possibly tal or conventional impression taking is with the same thickness can present
the high opacity of the cement. Although imperative for uniform color results. Us- differences in light transmission,
white opaque resin cement can mask the ing try-in paste before final cementation which results in differences in optical
underlying dark color of the Ag-Pd core, may be worthwhile, although 1 study performance and color parameters.19
based on the results of this study, has questioned the true agreement in How the resultant shades of other
increasing cement thickness to 300 mm color between try-in pastes and the translucencies and shades of MLD ce-
did not improve the resultant shades. In corresponding cements.47 ramics are affected by the variables
addition, thicker cement layers have This study has several limitations. examined in this study requires further
been reported to be related to decreased First, only a ceramic thickness of 1.5 investigation.
bond strength and the fracture strength mm was tested. Previous studies
of ceramic systems.35,38 demonstrated that the color of under-
CONCLUSION
Substantial shade differences were lying substrate structures and cement
observed when comparing resultant have minimal influence on the resultant
Within the limits of this study, the
shades of ceramic-Ag-Pd assemblies color if the ceramic thickness is 2 mm
following conclusions were drawn:
with cements of the same nominal or more, whereas the same factors
colors from 2 commercial manufacturers substantially alter the color of ceramic 1. The color of MLD ceramic speci-
(such as NWO vs MWO, NXY vs restorations when the ceramic thickness mens mock luted on Ag-Pd foundations
MLY) (Tables II, III; Fig. 2). After the is 1 mm.3,7,9,13 Second, the study was affected by both the color and the
pairwise comparison of each cement design did not include actual bonding thickness of resin cements.
The Journal of Prosthetic Dentistry Niu et al
January 2014 49
2. With the exception of 300-mm 10. Li Q, Yu H, Wang YN. Spectrophotometric 27. Shokry TE, Shen C, Elhosary MM,
evaluation of the optical influence of core build- Elkhodary AM. Effect of core and veneer
MWO cement, white opaque cements
up composites on all-ceramic materials. Dent thicknesses on the color parameters of two
yielded better shade matches than the Mater 2009;25:158-65. all-ceramic systems. J Prosthet Dent 2006;
other color cements tested. The optimal 11. Shimada K, Nakazawa M, Kakehashi Y, 95:124-9.
cement thickness for color match varied Matsumura H. Influence of abutment mate- 28. Brodbelt RH, O’Brien WJ, Fan PL. Trans-
rials on the resultant color of heat-pressed lucency of dental porcelains. J Dent Res
among cements of different opacities lithium disilicate ceramic. Dent Mater J 1980;59:70-5.
and colors. 2006;25:20-5. 29. Holand W, Schweiger MS, Frank M,
3. Among the 5 cements tested in 12. Nakamura T, Saito O, Fuyikawa J, Rheinberger V. A comparison of the
this study, the lowest mean DE was
Ishigaki S. Influence of abutment substrate microstructure and properties of IPS
and ceramic thickness on the color of heat- Empress 2 and the IPS Empress glass-
observed for NWO cement. The com- presses ceramic crowns. J Oral Rehabil ceramic. J Biomed Mater Res 2000;53:
bination of NWO cement at 100-mm 2002;29:805-9. 297-303.
13. Vichi A, Ferrari M, Davidson CL. Influence of 30. Otto T, De Nisco S. Computer-aided
cement-film thickness resulted in the
ceramic and cement thickness on the direct ceramic restorations: 10-year pro-
best color match of ceramic assemblies masking of various types of opaque posts. spective clinical study of CEREC CAD/
with the A1 target block. J Prosthet Dent 2000;83:412-7. CAM inlays and onlays. Int J Prosthodont
4. Increasing the cement thickness of 14. Douglas RD, Przybylska M. Predicting por- 2002;15:122-8.
celain thickness required for dental shade 31. Herrguth M, Wichmann M, Reich S. The
white opaque cement from 100 mm to matches. J Prosthet Dent 1999;82:143-9. esthetics of all-ceramic veneered and mono-
300 mm did not improve the resultant 15. Azer SS, Rosenstiel SF, Seghi RR, lithic CAD/CAM crowns. J Oral Rehabil
shade match. Johnston WM. Effect of substrate shades on 2005;32:747-52.
the color of ceramic laminate veneers. 32. Bindl A, Windisch S, Mormann WH.
J Prosthet Dent 2011;106:179-83. Full-ceramic CAD/CAM anterior crowns
16. Alghazzawi TF, Lemons J, Liu PR, Essig ME, and copings. Int Comp Dent 1999;2:
REFERENCES Janowski GM. Evaluation of the optical 97-111.
properties of CAD-CAM generated yttria- 33. Dozic A, Tsagkari M, Khashayar G,
1. Hefferman MJ, Aquilino SA, Diaz-Arnold AM, stabilized zirconia and glass-ceramic Aboushelib M. Color management of porcelain
Haselton DR, Stamford CM, Vargas MA. laminate veneers. J Prosthet Dent 2012;107: veneers: influence of dentin and resin cement
Relative translucency of six all-ceramic 300-8. colors. Quintessence Int 2010;41:567-73.
systems. Part I: core materials. J Prosthet 17. Dozic A, Kleverlaan CJ, Meegdes M, 34. Alqahtani MQ, Aljurais RM, Alshaafi MM.
Dent 2002;88:4-9. ven der Zel J, Feilzer AJ. The influence of The effects of different shades of resin luting
2. Kelly JR, Nishimura I, Campbell SD. Ceramics porcelain layer thickness on the final shade cement on the color of ceramic veneers. Dent
in dentistry: historical roots and current of ceramic restorations. J Prosthet Dent Mater J 2012;31:354-61.
perspectives. J Prosthet Dent 1996;75:18-32. 2003;90:563-70. 35. Manso AP, Silva NR, Bonfante EA, Pegoraro TA,
3. Chaiyabutr Y, Kois JC, LeBeau D, 18. Barath VS, Faber FJ, Westland S, Dias RA, Carvalho RM. Cements and adhesives
Nunokawa G. Effects of abutment color, Niedermeier W. Spectrophotometric analysis for all-ceramic restorations. Dent Clin North Am
cement color, and ceramic thickness on of all-ceramic materials and their interaction 2011;55:311-32.
the resulting optical color of a CAD/ with luting agents and different backgrounds. 36. Chang J, Da Silva J, Sakai M, Kristiansen J,
CAM glass-ceramic lithium disilicate- Adv Dent Res 2003;17:55-60. Nagai SI. The optical effect of composite
reinforced crown. J Prosthet Dent 19. Peixoto RT, Paulinelli VM, Sander HH, luting cement on all ceramic crowns. J Dent
2011;105:83-90. Lanza MD, Cury LA, Poletto LT. Light 2009;37:937-43.
4. Azer SS, Ayash GM, Johnston WM, transmission through porcelain. Dent Mater 37. De Azevedo Cubas GB, Camacho GB,
Khalil MF, Rosenstiel SF. Effect of esthetic 2007;23:1363-8. Demarco FF, Pereira-Cenci T. The effect of
core shades on the final color of IPS 20. Wang H, Xiong F, Zhenhua L. Influence luting agents and ceramic thickness on the
Empress all-ceramic crowns. J Prosthet Dent of varied surface texture of dentin color variation of different ceramic against a
2006;96:397-401. porcelain on optical properties of porcelain chromatic background. Eur J Dent 2011:
5. Carossa S, Lombardo S, Pera P, Corsalini M, specimens. J Prosthet Dent 2011;105:242-8. 245-52.
Rastello ML, Preti G. Influence of posts and 21. Ge J, Wang XZ, Feng HL. Influence of 38. Cekic-Negas I, Canay S, Sahin E. Bonding
cores on light transmission through different different post core materials on the color of of resin core materials to lithium disilicate
all-ceramic crowns: spectrophotometric and Empress 2 full ceramic crowns. Chin Med J ceramics: the effect of resin cement film
clinical evaluation. Int J Prosthodont (Engl) 2006;119:1715-20. thickness. Int J Prosthodont 2010;23:
2001;14:9-14. 22. Stevenson B, Ibbetson R. The effect of the 469-71.
6. Goracci C, Ferrari M. Current perspectives on substructure on the color of samples/resto- 39. Chu SJ, Trushkowsky RD, Paravina RD.
post systems: a literature review. Aust Dent J rations veneered with ceramic: a literature Dental color matching instruments and sys-
2011;56(suppl 1):77-83. review. J Dent 2010;38:361-8. tems. Review of clinical and search aspects.
7. Denissen H, Dozic A, van Waas M, Feilzer A. 23. Vichi A, Louca C, Corciolani G, Ferrari M. J Dent 2010;38s:e2-16.
Effects of 5 manipulative variables on the Color related to ceramic and zirconia 40. Öngül D, Şermet B, Balkaya MC. Visual and
color of ceramics used for computer- restorations: a review. Dent Mater 2011; instrumental evaluation of color match ability
generated restorations. Quintessence Int 27:97-108. of 2 shade guides on a ceramic system.
2007;38:401-8. 24. Spear F, Holloway J. Which all-ceramic J Prosthet Dent 2012;108:9-14.
8. Trushkowsky RD. Esthetic and functional system is optimal for anterior esthetics? J Am 41. Douglas RD, Steinhauer TJ, Wee AG.
consideration in restoring endodontically Dent Assoc 2008;139(suppl):19-24S. Intraoral determination of the tolerance of
treated teeth. Dent Clin North Am 2011;55: 25. Shiraishi T, Wood DJ, Shinozaki N, dentists for perceptibility and acceptability
403-10. van Noort R. Optical properties of base of shade mismatch. J Prosthet Dent
9. Volpato CA, Monteiro S, de Andrada MC, dentin ceramics for all-ceramic restorations. 2007;97:200-8.
Fedel MC, Petter CO. Optical influence of the Dent Mater 2011;27:165-72. 42. CIE (Commission Internationale de I’Eclair-
type of illuminant, substrate and thickness of 26. Ahn JS, Lee YK. Difference in the translucency age). Colorimetry-technical report. CIE pub.
ceramic materials. Dent Mater 2009;25: of all-ceramics by the illuminant. Dent Mater no. 15: 2004. 3rd ed. Vienna: Bureau Central
87-93. 2008;24:1539-44. de la CIE; 2004. p. 18.

Niu et al
50 Volume 111 Issue 1
43. Lehmann KM, Devigus A, Igiel C, 46. Lasserre JF, Pop-Ciutrila IS, Colosi HA. Corresponding author:
Wentaschek S, Azar MS, Scheller H. Repeat- A comparison between a new visual Dr Yi-Hua (Eva) Niu
ability of color-measuring devices. Eur J method of color matching by intraoral Shin Kong Wu Ho-Su Memorial Hospital
Esthet Dent 2011;6:428-35. camera and conventional visual and Department of Dentistry
44. Llena C, Lozano E, Amengual J, Forner L. spectrometric methods. J Dent 95, Wen Chang Road, Shih Lin district, Taipei
Reliability of two color selection 2011;39(suppl 3):e29-36. TAIWAN
devices in matching and measuring tooth co- 47. Akgungor G, Akkayan B, Gaucher H. Influ- E-mail: niueva@gmail.com
lor. J Contemp Dent Pract 2011;12:19-23. ence of ceramic thickness and polymerization
45. Khashayar G, Dozic A, Kleverlaan CJ, mode a resin luting agent on early bond Copyright ª 2014 by the Editorial Council for
Feilzer AJ. Data comparison between two strength and durability with a lithium dis- The Journal of Prosthetic Dentistry.
dental spectrophotometers. Oper Dent ilicate ceramic system. J Prosthet Dent
2012;37:12-20. 2005;94:234-41.

Access to The Journal of Prosthetic Dentistry Online is reserved for print subscribers!

Full-text access to The Journal of Prosthetic Dentistry Online is available for all print subscribers. To activate your individu-
al online subscription, please visit The Journal of Prosthetic Dentistry Online. Point your browser to http://www.journals.
elsevierhealth.com/periodicals/ympr/home, follow the prompts to activate online access here, and follow the instruc-
tions. To activate your account, you will need your subscriber account number, which you can find on your mailing
label ( note: the number of digits in your subscriber account number varies from 6 to 10). See the example below in
which the subscriber account number has been circled.
Sample mailing label

This is your subscription


account number

Personal subscriptions to The Journal of Prosthetic Dentistry Online are for individual use only and may not be trans-
ferred. Use of The Journal of Prosthetic Dentistry Online is subject to agreement to the terms and conditions as indicated
online.

The Journal of Prosthetic Dentistry Niu et al

You might also like