Professional Documents
Culture Documents
Author Abstract
Youjuan Zhang
Agilent Technologies
This Application Note describes an effective and reliable analytical method for
(Shanghai) Co. Ltd,
the determination of organophosphorus and organochlorine pesticide residues in
Shanghai 200131 P.R. China
fruit and vegetables following the China Standard NY/T 761-2008.1 An unpurged
two-way capillary flow technology (CFT) device was used to split the sample 1:1
onto two columns then to two detectors. Compared with the traditional pretreatment
procedure described in NY/T 761-2008, a simplified Quick, Easy, Cheap, Effective,
Rugged, and Safe (QuEChERS) method was used in this application, and provided
sufficient sample matrix cleanup while preserving low-level analyte detection. Area
repeatability, linearity, and recovery were evaluated for both organophosphorus and
organochlorine pesticides using an Agilent 8890 GC system with four detectors.
Introduction spectrometry methods have obvious instrument to provide greater flexibility
advantages in qualitative analysis, and compared to other lab GCs. For both
The use of pesticides has played an can determine dozens or even hundreds organophosphorus and organochlorine
important role in the prevention and of pesticide residues simultaneously with pesticides analysis, labs do not need
control of pests and the resulting high efficiency. Consequently, the price to change hardware; the only thing
improvement of agricultural yields. of the instrument is relatively expensive. they need to do is to reinstall different
Two common pesticide classes Although gas chromatography has columns onto different detectors.
used in agricultural treatments have less qualitative ability than MS, it is still Sample pretreatment is important
organophosphorus and organochlorine welcomed by many labs because of its for the analysis of multipesticide
chemical structures. selective detectors and low detection residues, which directly affects the work
For the determination of cost. ECD has excellent selectivity for efficiency and sensitivity. The NY/T 761
organophosphorus and organochlorine chlorine, and is a good choice for the method uses traditional, labor-intensive
pesticide residues in fruit and analysis of organochlorine pesticides. extraction and cleanup procedures
vegetables, gas chromatography (GC), FPD has high selectivity for sulfur and for sample pretreatment. The cleanup
gas chromatography/mass phosphorus, and it is a good choice procedure of organochlorine pesticides
spectrometry (GC/MSD), and gas for the analysis of organophosphorus and organophosphorus pesticides is
chromatography/tandem mass pesticides. In the NY/T 761 method, different. That means two different
spectrometry (GC/QQQ) methods are a primary column and confirmation pretreatments are needed for the same
widely used. Correspondingly, China column are used in tandem to achieve sample when organophosphorus and
has issued a series of standards for accurate qualitative analysis and prevent organochlorine pesticides both need
the determination of those pesticides. false-positive results. This dual-column to be tested. The popular sample
NY/T 761-2008 describes a GC method approach also provides quantitative pretreatment method of QuEChERS is
with an electron capture detector results for both organophosphorus the optimal method for high-throughput
(ECD) and flame photometric detector and organochlorine pesticides when sample analysis. Most importantly, the
(FPD). GB/T 19648-20062 describes a used with two FPDs and two ECDs. same sample pretreatment process can
GC/MSD method for 500 pesticides, Normally, labs need two GC systems be used both for organophosphorus and
and GB 23200.113-20183 is a GC/QQQ to follow the NY/T 761 method strictly. organochlorine pesticides, which greatly
method for 208 pesticides. Mass However, the 8890 GC can have two improves the analysis efficiency.
FPDs and two ECDs installed on one
2
Experimental A FPD1 B ECD1
FPD2 ECD2
Instrumentation HP-50+
HP-1
DB-5
DB-17
Inlet Inlet
An 8890 GC with an SSL inlet equipped
with two ECDs and two FPDs was Splitter Splitter
used for this series of experiments. An Retention Retention
gap gap
unpurged two-way CFT device was used
to split the sample 1:1 onto two columns Figure 1. Agilent capillary flow technology two-way splitter without makeup gas (p/n G3181B) and diagram
of instrument setup of simultaneous confirmation from a single injection onto both the primary and
for detection on two detectors. For confirmation columns. (A) FPDs for phosphorus detection, (B) ECDs for chlorine detection.
organophosphorus pesticides analysis,
the primary column and confirmation
column were Agilent HP-50+ and Table 1. Chromatographic conditions.
All reagents and solvents were HPLC Liner Agilent Ultra Inert, split, low pressure drop, glass wool (p/n 5190-2295)
grade. Acetonitrile (ACN) and hexane Injection 2 µL
were purchased from J&K Scientific Retention Gap 0.5 m × 0.53 mm id deactivated fused silica tubing (p/n 160-2535-5)
LTD. Acetone was purchased from Column1: Agilent DB-5 30 m × 0.25 mm, 0.25 µm (p/n 122-5032)
Column
ANPEL Laboratory Technologies Column2: Agilent DB-17 30 m × 0.25 mm, 0.25 µm (p/n 122-1732)
(Shanghai) Inc. All organophosphorus Carrier Nitrogen, 1 mL/min, constant flow (the column flow is the same for column 1 and 2)
3
Table 2. Data results for organophosphorus pesticides analysis on an Agilent HP-50+ column.
Solutions and standards
Table 2 shows 54 organophosphorus Linearity % RSD (n = 8)
pesticides divided into four groups. Range MDL
No. Name (mg/kg) R2 Low Middle High (mg/kg) % Recovery Group
According to the response values of
1 Dichlorvos 0.05–0.5 0.9982 2.8 1.5 1.5 0.004 108.9 1
each pesticide on the instrument, a
2 Acephate 0.05–0.5 0.9989 4.6 3.1 1.6 0.007 97.7 1
certain volume of single pesticide
3 Dicrotophos 0.05–0.5 0.997 4.3 2.3 1.6 0.007 105.3 1
standard solution of the same group
4 Disulfoton 0.05–0.5 0.9984 3.1 2.7 2.1 0.006 118.2 1
was accurately added and diluted with
5 Dimethoate 0.05–0.5 0.9981 0.6 1.6 1.3 0.002 111.4 1
acetone. The same method was used to
6 Parathion-methyl 0.05–0.5 0.9984 1.6 2.1 1.6 0.003 116 1
prepare four groups of stock solutions
7 Chlorpyrifos 0.05–0.5 0.9982 2.7 1.7 1.3 0.003 115.1 1
of organophosphorus pesticide mixture.
8 Pirimiphos-ethyl 0.05–0.5 0.9985 2.6 1.4 1.4 0.003 111.8 1
Calibration standards were diluted by
matrix blank (see Sample preparation). 9 Fenthion 0.05–0.5 0.999 3 2.4 1.6 0.005 111 1
4
Table 3 shows 41 organochlorine Table 2. Data results for organophosphorus pesticides analysis on an Agilent HP-50+ column (continued).
prepare three groups of stock solutions 47 Trichloronate 0.05–0.5 0.9999 2.3 2.7 1.1 0.004 97.5 4
of organochlorine pesticide mixture. 48 Malathion 0.05–0.5 0.9999 1.9 1.9 1 0.005 98.1 4
Calibration standards were diluted by 49 Isocarbophos 0.05–0.5 0.9999 2.6 1.8 1 0.004 96.9 4
matrix blank (see Sample preparation). 50 Quinalphos 0.05–0.5 0.9999 2.8 1.6 1 0.004 97 4
Table 3. Data results for organochlorine pesticides analysis on an Agilent DB-5 column.
% RSD (n = 8)
Linearity
Range 0.05 0.1 0.5 MDL %
No. Name (mg/kg) R2 mg/kg mg/kg mg/kg (mg/kg) Recovery Group
1 α-BHC 0.05–0.5 0.9996 1.1 1.2 1 0.00003 98.6 1
14-1 Cyfluthrin-1 1
14-4 Cyfluthrin-4 1
15-1 tau-Fluvalinate-1 1
0.05–0.5 0.999 2.8 1.4 0.7 0.0005 105
15-2 tau-Fluvalinate-2 1
21-1 Endosulfan-1 2
0.05–0.5 0.9984 2.1 0.5 0.8 0.00008 94.8
21-2 Endosulfan-2 2
5
Table 3. Data results for organochlorine pesticides analysis on an Agilent DB-5 column (continued).
% RSD (n = 8)
Linearity
Range 0.05 0.1 0.5 MDL %
No. Name (mg/kg) R2 mg/kg mg/kg mg/kg (mg/kg) Recovery Group
26-1 Flucythrinate-1 2
0.05–0.5 0.991 1.3 0.4 1 0.0005 92
26-2 Flucythrinate-2 2
37-1 Tetramethrin-1 3
0.05–0.5 0.997 2.2 1.1 2.4 0.0003 85.7
37-2 Tetramethrin-2 3
39-1 Cypermethrin-1 3
39-4 Cypermethrin-4 3
40-1 Fenvalerate-1 3
0.05–0.5 0.998 1.2 0.6 0.6 0.0003 93.9
40-2 Fenvalerate-2 3
6
Sample preparation QuEChERS sample preparation workflow
An apple sample was purchased from Weigh a 10 g sample (±0.01 g) into a 50 mL centrifuge tube.
a local grocery store. Ten grams of
homogenized apple sample were
weighed into a 50 mL centrifuge tube, Add spike solution, and vortex for one minute.*
and two ceramic homogenizers were
added to the sample. QC samples were
Add ACN (10 mL) and two ceramic homogenizers (p/n 5982-9313).
spiked with appropriate amounts of
spiking solution to yield QC samples
with a quantitative concentration of Add a QuEChERS extraction salt packet (p/n 5982-5650), and vortex for one minute
approximately 0.1 mg/kg. Ten milliliters
of acetonitrile were added to the tube. An
Agilent QuEChERS extraction salt packet Centrifuge for five minutes at 5,000 rpm at 10 °C.
(part number 5982‑5650) containing
4 g of MgSO4, 1 g of sodium chloride,
Transfer 6 mL of the upper layer to a dispersive tube (p/n 5982-5056).
1 g of Na-citrate, and 0.5 g of disodium
citrate sesquihydrate was added to
each centrifuge tube for extraction. An Vortex for one minute, and centrifuge for five minutes at 5,000 rpm at 10 °C.
Agilent QuEChERS general fruit and
vegetables dispersive SPE 15 mL tube
(part number 5982-5056) was used for Transfer 1 mL of the upper layer to a tube, then dry the final supernatant under nitrogen.
cleanup. For fruits and vegetables with
high pigments and fats, other types Use 1 mL of appropriate solvents** to reconstitute, vortex for one minute,
of QuEChERS packets are needed for and centrifuge for five minutes at 5,000 rpm at 10 °C.
extraction and cleanup. The details of the
sample preparation procedure are shown
in Figure 2. Transfer the extract to an autosampler vial.
Matrix blanks were prepared in the same * This is for the recovery test. For matrix blanks, skip this step.
** Using acetone for organophosphorus pesticides, while using hexane for organochlorine pesticides.
manner as the samples, except there
was no addition of spike solution. Figure 2. Flowchart of the QuEChERS extraction procedure for apple samples.
7
Results and discussion
Dimethoate Parathion-methyl
A Acetone
Matrix Ditalimfos
Disulfoton Chlorpyrifos
Organophosphorus Acephate Pirimiphos-ethyl
pesticides analysis Fenthion
of pesticide chromatograms in an
apple matrix blank and acetone. Blue
represents a standard prepared in
acetone, while red represents a standard
prepared in the apple matrix blank. It
B
shows that for some compounds, using Methamidophos Acetone
a matrix blank to dilute the working Matrix
120 9
this system. The 54 organophosphorus 100
3
12
10
pesticides were divided into four groups 80
60
for easy and accurate retention time 40
determination. Figures 4 to 7 illustrate 20
the analyses of groups 1, 2, 3, and 4 0
organophosphorus pesticide mixtures on 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
220 Time (min)
the HP-50+ and HP-1 columns. 200
B HP-1
13
180
160 2
140
150 pA
120
5 6
100
1 4 7 8 11 12
80 9
3
60
40 10
20
0
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
Time (min)
8
Matrix-matched calibration standards 220 Group 2
A HP-50+
and spiked QC samples were prepared 200
14 27
by spiking appropriate standard solutions 180
160
into the matrix blank. For 44 compounds 19 26
140
analyzed, the linearity range was in the 15 20 22
150 pA
120 16 28
range of 0.05 to 0.5 mg/kg. For other 100 18 24
21
compounds such as phoxim, phosmet, 80
17
23
25
120 21
an HP‑50+ column. 100 19 26 28
15 24
20 22
Repeatability assessments at three 80 17 16
18
60 25
levels: low, middle, and high were 23
40
obtained for all compounds in apple 20
matrix. For most compounds, the low 0
level was 0.05 mg/kg, the middle level 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
Time (min)
was 0.1 mg/kg, and the high level was
0.5 mg/kg. For low response value Figure 5. Chromatograms of group 2 organophosphorus pesticide standard solution (about 0.1 mg/kg) on
compounds, the low, middle, and high a dual-column system using HP-50+ and HP-1 capillary GC columns.
41
140 35
120 34
100 31 39 36
32 38
80 33 37
60 40
40
20
0
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
Time (min)
Figure 6. Chromatograms of group 3 organophosphorus pesticide standard solution (about 0.1 mg/kg) on
a dual-column system using HP-50+ and HP-1 capillary GC columns.
9
As described in the Sample preparation 170 A HP-50+ Group 4
section, QC samples were spiked with 160
150
appropriate amounts of spiking solution 140
130 49 54
to yield QC samples with quantitative 120 46-2
50
110
concentration of 0.1 mg/kg (for low 100 44 47
48
150 pA
response value compounds such as 90
80 51
43 45
phoxim, acquired at the 0.4 mg/kg 70
60
52 53
46-1
level). Recoveries were determined on 50
40
an HP-50+ column, and the results for 30
20
all organophosphorus pesticides were 10
between 70.4 and 118.2%. Table 2 0
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
lists the recoveries for the individual Time (min)
pesticide. Most of the compounds, 200
B HP-1
47
QuEChERS, as shown in Figure 8.
100
49 50
80 44
46-1 46-2 48
60 53
45 51
43
40
20
0
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
Time (min)
80
60
40
20
0
Profenofos
Propetamphos
Paraoxon
Parathion
Isofenphos
Methidathion
Phosfolan-methyl
Dichlorvos
Acephate
Disulfoton
Dimethoate
Parathion-methyl
Chlorpyrifos
Pirimiphos-ethyl
Fenthion
Ditalimfos
Phosmet
Trichlorfon
Ethoprophos
Phorate
Omethoate
Diazinon
Fonofos
Chlorpyrifos-methyl
Fenitrothion
Bromophos
Bromophos-ethyl
Ethion
Coumaphos
Methamidophos
Famphur
Phosalone
Azinphos-ethyl
Naled
Sulfotep
Terbufos
Monocrotophos
Dichlofenthion
Fenchlorphos
Pirimiphos-methyl
Mevinphos
Phosphamidon
Azinphos-methyl
Trichloronate
Malathion
Quinalphos
Isocarbophos
Tetrachlorvinphos
Phosfolan
EPN
Phoxim
Triazophos
Pyrazophos
Dicrotophos
10
Organochlorine pesticides analysis Group 1
Similar to the analysis of 4,500 A DB-5 4 5
1
organophosphorus pesticides, 4,000
Hz
10
system for organochlorine pesticides 2,000
9
2,000
10
1,500
1,000 13
3 12 15
500 14
2 11
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34
Time (min)
Group 2
5,500 17
A DB-5 18
17
5,000 18
4,500
4,000
16
3,500
3,000
Hz
2,500 22
2,000 21-2
16 20 24
1,500
1,000 19 21-1 26
500 23 25
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34
Time (min)
5,000 B DB-17
4,500 17
18
4,000
3,500
3,000
Hz
2,500
2,000 22
16
1,500 20 21 23,24
1,000 19 26
500 25
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34
Time (min)
Figure 10. Chromatograms of group 2 organochlorine pesticide standard solution (0.1 mg/kg) on a
dual‑column system using DB-5 and DB-17 capillary GC columns.
11
For the analysis of organochlorine 3,000 28
Group 3
38
A DB-5 33
pesticides, some compounds such 2,800 34
2,600 39
as cyfluthrin and cypermethrin have 2,400
isomers. The retention times of those 2,200
2,000
37
isomers were close, and it was difficult 1,800 27
29
35 37
1,600
to achieve baseline separation, as shown
Hz
1,400 30 36
in Figure 12. For these compounds, the 1,200
1,000
setting of integration parameters was 800
31 38 40
41
39
particularly important. Because the 600
400
32
37
standards purchased from the vendor 200
were also isomer mixtures, and no 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44
other compounds eluted between the 3,000
Time (min)
B DB-17 33
isomers, those isomers were integrated 2,800 34
2,600 28
as one peak for quantitative analysis. 2,400
36
Figure 13 shows that, in OpenLab CDS 2,200 3334
2,000
2.3 software, the Area Sum function can 1,800 27
35
help integrate the isomers. 1,600
Hz
29
1,400 36
1,200 30
1,000
800 31
600 38 39 40
32 35 41
400 37
200
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44
Time (min)
Figure 11. Chromatograms of group 3 organochlorine pesticide standard solution (0.1 mg/kg) on a
dual‑column system using Agilent DB-5 and DB-17 capillary GC columns.
2,000 Cyfluthrin
1,900
1,800
1,700
1,600
1,500
1,400
1,300
1,200
1,100
Hz
1,000
900
800
700
600
500
400
300
200
24.2 24.3 24.4 24.5 24.6 24.7 24.8 24.9 25.0 25.1 25.2 25.3
Time (min)
Figure 12. Chromatogram of cyfluthrin isomers using the Area Sum function for integration.
12
Matrix-matched calibration standards
and spiked QC samples were prepared
by spiking appropriate standard solutions
into the matrix blank. The spiking
concentration for calibration standards
were between 0.05 and 0.5 mg/kg in
apple matrix. The data were processed
with OpenLab CDS 2.3 software. Table 3
shows the results on a DB-5 column, the
R2 values >0.991 for all organochlorine
pesticides. The area RSD values for
eight replicates at three levels were
below 4%, with the typical RSD below
2%. Compared to the NY/T 761 method,
the optimized extraction and cleanup
procedure was validated by running
spiked samples at 0.1 mg/kg level.
Acceptable recoveries were achieved Figure 13. Integration table for cyfluthrin isomers.
for most of the analytes. Recoveries
were between 77.3 and 118.6%. Table 3
also shows the MDL results for the
41 compounds. S/N was used for MDL
calculation. The results were better than
the NY/T 761 method reference results.
80
60
40
20
0
p,p'-DDT
o,p'-DDT
Propanil
Vinclozolin
Bifenthrin
cis-Permethrin
Cyfluthrin
α-BHC
Simazine
δ-BHC
Heptachlor
Aldrin
o,p'-DDE
p,p'-DDE
o,p'-DDD
Iprodione
tau-Fluvalinate
β-BHC
γ-BHC
Pentachloronitrobenzene
Endosulfan
p,p'-DDD
Dicofol
Lambda-cyhalothrin
Permethrin
Flucythrinate
Dicloran
Hexachlorobenzene
Atrazine
Triadimefon
Procymidone
Butachlor
Dieldrin
Endrin
Tetramethrin
Fenpropathrin
Cypermethrin
Fenvalerate
Deltamethrin
Chlorothalonil
Chlorobenzilate
13
Conclusion References
An Agilent 8890 GC configured with four 1. China National Standard NY/T
detectors (two FPDs and two ECDs) 761-2008, Determination of
was used to screen organophosphorus Organophosphorus, Organochlorine,
and organochlorine pesticides in fruit Pyrethroid and Carbamate Residues
and vegetables. Splitting the samples in Vegetables and Fruits.
into two different columns then to 2. China National Standard GB/T
two detectors facilitated selectivity, 19648-2006, Determination of
identification, and confirmation of 500 Pesticides and Metabolites
organophosphorus pesticides and Residues in Vegetables and Fruits,
organochlorine pesticides from single Gas Chromatography/Mass
injections of each extract, increasing Spectrometry Method.
laboratory productivity.
3. China National Food Safety Standard
This Application Note demonstrates GB 23200.113-2018, Determination
excellent sensitivity, area repeatability, of 208 Pesticides and Metabolites
peak shape, and resolution for both Residues in Foods of Plant Origin,
organophosphorus and organochlorine Gas Chromatography-Tandem Mass
pesticides, which shows that this Spectrometry Method.
four‑detector system is an ideal platform
for the NY/T 761-2008 method.
www.agilent.com/chem