White Paper

www.vaisala.com

GMP warehouse mapping


step-by-step guidelines for validating
life science storage facilities
Good manufacturing practice (GMP) regulators in the United States,
Canada, European Union, Japan, Australia, and China have sharpened
their focus on warehouse storage and distribution practices. Driving this
trend is a shift in regulatory thinking from quality-by-test to quality-by-
design systems with emphasis on level of risk to product quality and
patient safety. Other drivers include greater demand for storage facilities
due to globalization of manufacturing, increase in temperature-sensitive
biopharmaceuticals, and changes in technology.

Regulators in these countries


require “mapping” the temperature
and relative humidity profiles of
warehouses for environmentally
sensitive life science products. This
step-by-step guide describes how
to map a warehouse to comply
with internationally recognized
GMPs, including many that have
been published or revised recently.
(See the end of this paper for
links to relevant regulations
and guidance documents.)
This guide, intended for use by
any organization involved in
the storage and distribution of
products sensitive to temperature
and humidity in a GMP-compliant
environment, draws on Vaisala’s
extensive customer experience
throughout North America and
Europe. Vaisala solutions are used
in over 150 countries worldwide.
Step by Step –
Good practices for warehouse mapping studies
Vaisala recommends a nine-point process for successful
mapping of a warehouse or other regulated storage space:

These nine steps will help you The validation master plan should:
design and execute a successful
1. Create a mapping plan. They will ensure • State the validation objectives.
Validation Plan that you take into consideration
the most important elements • Identify roles and responsibilities
of quality, metrology, and other
 of validation, especially
understanding where temperature
working groups in the process.

2. Identify areas at risk and humidity pose risks to product • Identify validation activities,
including processes, equipment,
quality. Following these steps will
and space.
 go a long way in demonstrating to
a regulatory inspector that your • Develop documentation
3. Develop Protocol company is GMP compliant. and procedures, including
the company’s response if
Information
a temperature or humidity
Step 1: Create a validation excursion occurs.
 plan
• Determine a validation schedule.
4. Determine sensor The validation plan, or validation
master plan, is the document • Specify the management approval
distribution used to specify the company’s process, especially for adverse
decisions about qualifying every events such as temperature
 aspect of the facility, equipment,
deviations.
and processes to maintain a GMP- • Create change control protocols
5. Select suitable compliant environment. The plan so it’s clear when changes such as
technology should take a risk-based approach, maintenance, new construction,
with a rationale based on verifiable and reconfiguration of racks will

 data. The plan should focus on


where environmentally sensitive
require revalidation.

6. Set up mapping products and materials will be Regulatory Note: GMPs require
equipment stored and whether environmental maintaining temperature
controls can meet specified and humidity within storage
 storage requirements. recommendations printed on
product labels or provided
7. Conduct test and The plan is also a starting point for by raw-material suppliers.
review data regulators to evaluate the rationale These recommendations are
for the company’s goals and derived from known chemical
methods.
 properties and stability testing.

8. Make modifications


9. Document and
schedule mapping
study
Step 2: Identify areas • Independent energy sources, such Step 3: Develop protocol
at risk as space heaters, air conditioners, information
and fans, which create warm or
To map a warehouse or storage cold spots. Once you’ve identified areas of
space, you first must identify areas risk, develop a protocol for the
• Layout of racks, shelves, and
where product quality may be mapping study that describes the
pallets, which obstruct airflow.
at risk because of unacceptable following, with justifications for
variations in temperature and • Location of HVAC control sensors. each decision:
humidity. Many factors affect For example, a thermostat located
the control or variability of your near a source of heat or cold may
cause the temperature of the • Types of data to be generated –
space. (Because relative humidity for example, temperature, relative
space to fluctuate excessively.
is dependent on temperature, humidity, and measurement
variations in temperature • Locations near sources of heat or intervals. Five-minute intervals
will affect humidity as well.) cold, such as the roof and exterior offer more data to evaluate
Considering each of these factors walls, windows, and loading docks. trends and modify the warehouse
will help you identify risks: setting (see Step 8). Once you
• High-traffic areas where product
are satisfied that temperature
or equipment is moved often.
and humidity are relatively
• Volume of space. A large • Seasonal temperature changes or stable, 15-minute intervals may be
warehouse has different control unusual weather events. adequate for the final mapping.
burdens than a small storage
area, with greater demands • Number of sensors to be used
on the HVAC system and the (see Step 4: Determine Sensor
potential for greater variations Distribution).
in temperature and humidity at Regulatory Note: You can
• Schematic or diagram of sensor
various locations. achieve GMP compliance locations.
through sound justification of
• The capacity of diffusers or fans
your approach to identifying • Duration of study. Your rationale
to adequately circulate air.
risk. The more considerations the and protocol may support a
• Temperature gradients between protocol addresses, the better series of tests, each lasting two
the floor and warmer air near the days during normal operations
your rationale is likely to be.
ceiling. and into a weekend. A different
and equally defensible protocol
might specify a single run over
a two-week period to account
for a variety of activities, such as
opening loading dock doors, in
the warehouse.
• Calibration requirements of the
3
data loggers.
2

5 • Acceptable range of variation


over time and across the space,
which will depend on the product
7 8 stored.
1 4
• Acceptable limits for temperature
10
or relative humidity excursions.
12 13 • Reporting requirements.
6 9
Regulatory Note: Once you
Loading
develop a protocol, follow it
15
Door Door consistently. If the protocol
11 14 changes, document the
reasons.

Figure 1: . The even distribution of 15 sensors is a typical pattern for a


three-dimensional mapping of a small space.
Top
View

Figure 2: Sensors placed in the middle of racks more closely reflect product temperatures. In this example, nine
sensors are located on each double rack in this warehouse measuring 30 meters by 30 meters by 15 meters.

Step 4: Determine sensor In mapping a large warehouse, If you don’t have an adequate
distribution set sensors as far as 30 meters number of sensors to map an
apart, with additional sensors in entire warehouse in one study, you
How many sensors will you need to vulnerable areas affected by: may map one section at a time.
map a particular space? Where will Mapping in sections takes longer,
you put them? There are no simple • Heat or cold from external walls, and you may want to extend the
answers. Sensor distribution must solar heating, windows, lighting mapping time for each section to
be adequate to assess temperature compensate for the uncertainty of
• Air circulation or drafts from
uniformity. Good practice means mapping the space in sections. To
entries, traffic, or the HVAC
that you use a sufficient number decide, calculate the equipment
system
of sensors to understand your savings from a sectional mapping
environment, especially areas • Temperature extremes in poorly
approach against the additional
where risk is greatest. insulated areas
time needed to complete the
• Localized effects of space heaters project.
You’ll need to place sensors in and air conditioners
a uniform pattern in all three If high or low relative humidity can
dimensions of the space – top to Anticipate that airflow and adversely affect product quality,
bottom, left to right, and front temperature gradients may vary then you should map for relative
to back. Add additional sensors depending on whether shelves are humidity as well as temperature.
where you suspect cool or warm empty or stocked with product. There are two approaches to
areas exist, as well as near the Taller racks will be subject to wider determining the number and
control sensors and monitoring temperature gradients, requiring location of relative humidity
sensors. Placement of temperature more sensors top to bottom. sensors.
and relative humidity sensors is a
function of the risks identified in You can mount sensors in open
Step 2. areas (for example outside
of racks or aisles) where they
A walk-in chamber or small are convenient to set up. But
warehouse is often mapped in convenience must not take
three dimensions with 15 sensors precedence over effectiveness.
(See Figure 1.) The protocol should Sensors must measure the
include guidelines for the distance conditions that products are
between sensors, for example, no exposed to.
greater than six meters.
Considerations for Regulatory Note:
humidity Understanding the relationships
Compared with temperature between measured parameters
sensors, relative humidity sensors is essential to successful
are far more prone to to lose mapping studies and managing
accuracy, or “drift,” over time. Drift risk in a GMP storage space.
may be caused by poor design,
poor calibration, or contamination Step 5: Select suitable
from water-vapor saturation or technology
chemical vapors. A single errant
reading at recalibration time will Use equipment designed
call attention to your decision for mapping. Software that
to use fewer humidity sensors. accompanies the sensors is used
Starting with fewer humidity to set up the equipment and
sensors creates the risk of download data. Software should
nonconformance, because if one produce tabular and graphical
fails, or is out of specification, that reports that meet all requirements
single sensor will represent a high of 21 CFR Part 11 and comparable
percentage of your total humidity international standards, such as
measurements. Interpolating European Commission Annex 11,
relative humidity from temperature and those contained in European
data will require that a company Union GMP Volume 4.
employee with this specialized
knowledge meet with the auditor When choosing data loggers, look
or inspector. Ideally, your company for the following features:
should minimize the number
of contacts needed during an • Minimum sources of error – that is,
inspection as a way to streamline low measurement uncertainty.
the process and reduce the • High accuracy in the
possibility of a misstatement. measurement range. Vaisala
Determining humidity DL2000 data loggers, for
sensor density If you’re concerned about relative example, are accurate to
The first approach is to use humidity, a more defensible ± 0.1 °C over +20 °C to +30 °C,
mapping strategy is to track with humidity accuracy of ± 1 %RH
comparatively few humidity
temperature and humidity at all in 10 to 80 %RH.
sensors distributed throughout the
warehouse (as few as one for every locations with data loggers that • Sensitivity to small temperature
six temperature sensors). In this record both measurements. It’s changes (high resolution). The
case you will rely on temperature important to use high-quality more rapid the response, the
uniformity to make the case that data loggers that are stable and more closely the data point can
regularly calibrated. be associated with the time of the
humidity is also within bounds. This
measurement.
approach should be based on a
history of temperature mapping in Mapping with integrated • Long-term stability, particularly
different seasons with consistent temperature and relative humidity for relative humidity sensors.
results. With this history, a specialist sensors offers several advantages Low-quality equipment needs to
with an understanding of humidity over deducing humidity from be calibrated before and after
measurement can effectively make temperature. Mapping both every study.
the case to an auditor or inspector temperature and humidity at all
• Traceable calibration performed
that humidity measurements are sensor locations provides a more within the measurement range
not needed at all data points. If you quantitative map of the entire and with equipment using an
decide to follow this strategy and storage space for inspectors and unbroken chain of comparisons
cut back on the number of humidity auditors to easily comprehend to an internationally recognized
sensors, it’s crucial to place the without detailed explanation. And standard such as that of the
relative humidity excursions will National Institute of Standards
few humidity sensors you do use
be easier to identify with more and Technology (NIST).
in areas with poor air circulation,
between HVAC fans or diffusers, humidity data points. • Clear, comprehensive, and
and where temperature is most accessible calibration records.
variable.
Regulatory Note: GMPs Step 6: Set up mapping • Ensure mapping software
equipment access has been secured and
require written procedures
authenticated. Access privileges
for calibrating, inspecting, After you’ve identified risk restrict who is allowed to use the
and checking automated, areas and determined sensor application.
mechanical, and electronic distribution, it’s time to set up
equipment (21 CFR 211.68). • Ensure the software reads and
mapping equipment and conduct records hardware and firmware
International standards a test of the storage space. model, version, and serial number.
such as ISO/IEC 17025:2017 The purpose of this initial test
“General Requirements for is to determine where variable • Ensure the warehouse area
the Competence of Testing and data logger locations are
conditions exist, and where
and Calibration Laboratories” accurately described. A schematic
temperature and humidity are
are recognized best-practice or diagram helps ensure
uniform and suitable for product consistent sensor placement in
references for calibration. storage. Work through the subsequent mapping studies.
following checklist and document
each step: • Regular sample intervals have
been determined. Intervals
typically run between five and 15
• Equipment has been calibrated. minutes.
Document by whom, when, and
the next calibration date. This • Study duration has been
confirms that the data logger determined. All data loggers are
performs within the calibrated set to begin and end at the same
measurement range. time.

• Equipment has been validated. • Data loggers link to an audit


Installation qualification and trail file for traceability. This is an
operation qualification (IQ/OQ) is essential requirement to show
typically provided by the mapping that the data is trustworthy.
system supplier. • Data loggers are functional and
positioned in defined locations.

Regulatory Note: GMPs require


the use of calibrated equipment
and calibration records. If you
gathered data in electronic
form, these records must
meet regulations for electronic
records as defined in 21 CFR
Part 11, in EC Annex 11, and in
European Union GMP Volume 4.
Step 7: Conduct a test • Raw data with times and dates. A graphical overview can help
and review data • Calculated values such as
identify high-risk locations,
temperature minimum, maximum, especially where problems may
You’ll need to establish the occur sporadically. For example, a
and average.
reporting information you’ll use temperature spike may be linked
to evaluate the test. When the • A graph of all sensors over the to a time when loading doors
test is complete, the software test period. were open.
will read the secure files from the • Instrument settings.
data loggers, show recorded data, Such a variation might indicate
perform calculations, and graph • Calibration information.
a risk from routine workplace
the results selected for a mapping • Date and time of the test. activity or suggest the need for a
study report. The test document buffer zone.
will typically show the information • Space for review and approval
signatures on printed reports.
in Figure 3:
Regulatory Note: It’s better
to present a summary graph
Trend data from each sensor can
with a clear conclusion than a
be compiled in a single graph to
very detailed report that may
provide an overview. Preset lines,
generate additional questions.
such as acceptable minimum
and maximum limits, can aid the
analysis.

Figure 3. The mapping report can show high and low limits to quickly visualize thresholds.
Step 8: Make Step 9: Document and HVAC changes, and similar
modifications schedule the mapping modifications to the environment
study require additional mapping.
Use the results from the initial Seasonal changes and extreme
test to identify locations where After you adjust for environmental weather may justify mapping the
the product may be exposed to variability in the warehouse, it’s warehouse with greater frequency
unacceptable conditions. Then time to conduct and document a or rescheduling a test for a more
make adjustments—for example, mapping study for approval. “seasonable” temperature. For
to the storage racks or HVAC example, the validation plan may
system—to correct this variation. call for a mapping study in July,
How long should mapping last?
Or simply decide where products when temperatures are typically
will not be stored. For example, As with your initial mapping test, hottest. But if July is unseasonably
many warehouses have a mezzanine there’s no fast rule. Your rationale cool, it may make sense to delay
level designated off-limits for raw and protocol may support a single mapping until a warm spell in
materials or finished goods because long study, or a series of shorter August. The validation plan should
HVAC controls are ineffective there. studies. Either way, it’s important provide enough flexibility to
Name and describe these locations to measure the environment capture weather extremes. For
and modify the validation plan. Also, during a range of different work example, depending on the climate
modify your validation protocol in activities in the warehouse, such in your area, your plan might
light of the results from your initial as loading, moving product, and call for mapping when summer
mapping test. periods such as weekends when temperatures exceed 30°C and
little activity might occur. winter temperatures fall below 0°C.
Regulatory Note: Modifications
to a newly commissioned How often should you map Regulatory Note: Maintaining
warehouse don’t need to a space? useful records is integral to
appear in the inspection Some protocols call for mapping meeting GMPs. Records must
record. But once your company every three months while others be stored securely and easily
approves a validation master can justify mapping yearly or even retrieved for review. They
plan, then the plan must less frequently. The validation must be gap-free. They must
document all subsequent master plan should anticipate provide an audit trail. Records
changes. variables that can change storage may be paper, electronic, or
conditions after completion a combination. If they are
of a warehouse qualification. electronic records, they must
Warehouse construction, major meet the requirements of 21
CFR Part 11 or EC Annex 11.
even these guidance documents Pharmaceutical Inspection

Summary can lag behind technological


advances. In a race to keep
Convention and Pharmaceutical
Inspection Cooperation Scheme:
up, regulatory agencies and • PIC/S GMP Guide Part I: Guide
The keys to a successful
industry stakeholders worldwide to GMP for Medicinal Products
warehouse mapping study
revise their interpretations of Section 3.19
include creating a validation
GMPs, developing new guidance • PIC/S GMP Guide Part II: Guide
plan and protocol, with
documents. So it’s imperative to GMP for Medicinal Products
justification for each step.
to keep abreast of the changing Sections 7.42 and 10.1
standards.
Document changes to the plan Health Canada
and protocol. • GUI 0069: Guidelines for
Links to resources Temperature Control of Drug
Identify areas of risk in your Products During Storage and
International Conference on
warehouse to determine the Transportation
Harmonisation:
distribution of sensors and U.S. FDA:
• ICH Q7 - GMP Guidance for Active
duration of the mapping.
Pharmaceutical Ingredients • 21 CFR Part 210 cGMP in
Manufacturing, Processing,
Select reliable technology • ICH Q9 – Quality Risk
Packing, or Holding of Drugs
suitable to the task. Management
• 21 CFR Part 211 cGMPs for
• ICH Q10 Pharmaceutical Quality
Finished Pharmaceuticals
Modify your storage space System
to ensure you are mapping a • 21 CFR Part 820 Quality System
United States Pharmacopeia:
controlled environment. Regulation
• USP Chapter 1079 Good Storage
• 21 CFR Part 600 Biological
and Distribution Practices for
Document and schedule Products
Drug Products
mapping studies to account • 21 CFR Part 111 cGMPs in
• USP Chapter 1118 Monitoring
for changes in the warehouse Manufacturing, Packaging,
Devices – Time, Temperature, and
environment. Labeling, or Holding Operations
Humidity
for Dietary Supplements
Keep records in a manner that International Society of
• 21 CFR Part 11 Electronic Records;
they are secure and accessible. Pharmaceutical Engineering:
Electronic Signatures
• ISPE Good Practice Guide –
• Pharmaceutical CGMPs for the
Document that your protocol Controlled Temperature Chamber
21st Century – A Risk-Based
was followed consistently, and Mapping and Monitoring
Approach
re-evaluate your procedures Parenteral Drug Association:
ASTM (formerly American Society
periodically. • PDA Technical Report No. 52 – for Testing and Materials):
Guidance for Good Distribution
• ASTM E2500 Standard Guide
Practices for the Pharmaceutical
for Specification, Design, and
Regulations and guidance Supply Chain
Verification of Pharmaceutical and
Warehouse mapping regulations European Commission: Biopharmaceutical Manufacturing
require documented evidence • EC Guidelines on Good Systems and Equipment
that an environment is in a state Distribution Practice of Medicinal
of control and suitable for the Products for Human Use
products stored there. Regulatory • Eudralex Volume 4 Good
agencies and independent Manufacturing Practices –
organizations also issue non- Medicinal Products for Human
binding guidance documents and Veterinary Use, Annex 11:
that can provide greater detail Computerized Systems
than regulations in applying
current regulations. However,

Please contact us at Ref. B211170EN-B ©Vaisala 2020


This material is subject to copyright protection, with all copyrights
www.vaisala.com/contactus retained by Vaisala and its individual partners. All rights reserved.
Any logos and/or product names are trademarks of Vaisala or its
individual partners. The reproduction, transfer, distribution or
storage of information contained in this brochure in any form
without the prior written consent of Vaisala is strictly prohibited.