Professional Documents
Culture Documents
What is a Biosensor?
Biosensor is a powerful and innovative device that is used to convert biological reactions into
measurable signals to detect target analytes. This technological device consists of three
components, the biomolecule, transducer and output system, as depicted in Figure 1.1.
Biomolecules (sensitive bioreceptors) such as enzymes, or bioaffinity groups like antibodies,
tissues and DNA probes are immobilized to support materials. These molecules recognize the
target analytes like enzyme substrates, complementary DNA, antigens, whole cells including
microbial, plant, animal cells, subcellular organelles and lectins. The transducers are used to
convert the signals resulting from the interaction between the biomolecule and target analyte into
observable signals. The signal is then displayed through the output system. The main features of
biosensors are stability, cost, sensitivity, and reproducibility.
Classification of Biosensors:
The different types of biosensors are classified based on the sensor device as well as the biological
material that is discussed below.
1. Electrochemical Biosensor: Generally, the electrochemical biosensor is based on the
reaction of enzymatic catalysis that consumes or generates electrons. Such types of
enzymes are named as Redox Enzymes. The substrate of this biosensor generally includes
three electrodes such as a counter, reference, and working type.
2. Optical Biosensors: Optical Fibers play an important role in Optical Biosensors. The
optical fibers allow detection of the sensing elements based on the different properties of
light like absorption, scattering and fluorescence.
One of the main advantages of using optical biosensors is their non-electrical nature. This
allows them to analyze multiple elements on a single layer just by varying the wavelength
of the light.
3. Piezoelectric Biosensors: are also known as Acoustic Biosensors as they are based on the
principle of sound vibrations i.e. acoustics. When a mechanical force is applied on a
piezoelectric biosensor, they produce an electrical signal.
Biomolecules
Need for immobilization: The basic requirement of a biosensor is that the biologic material bring
the physicochemical changes in close proximity to a transducer. In this direction, immobilized
enzyme technology has played a major role. Immobilization not only helps in forming the required
close proximity of the biomaterial with the transducer but also in stabilizing it for reuse.
The biologic material has been immobilized directly on the transducer or, in most cases, in
membranes, which can subsequently be mounted on the transducer. Biomaterials can be
immobilized either through adsorption, entrapment, covalent binding, crosslinking, or a
combination of all these techniques. Selection of a technique and support depends on the nature of
enzyme, nature of substrate, and configuration of the transducer used. The choice of support and
technique for the preparation of membranes often has been dictated by the low diffusional
resistance of the membrane. The choice of technique also depends on the biomaterial of interest.
Thus, enzymes and antibodies have normally been immobilized either by covalent binding or
through adsorption followed by crosslinking.
Various methods are available for immobilization of biomolecules (Figure 1.3). The most
commonly used biomaterial immobilization techniques for designing and development of specific
sensors are physical adsorption, entrapment, inter molecular cross-linking and covalent binding:
Support materials
The term support material usually refers to the solid insoluble matrix where biomolecules can be
attached to. Support materials are widely implemented in an attempt to conserve the tertiary
structure of enzymes used in immobilization, which contributes to the stability and reactivity of
the enzymes. Support materials must display good chemical and mechanical stability, as they can
alter the properties of the immobilized biomolecules. Generally, an ideal support needs to be a
mesoporous material to provide wide surface area and high porosity to increase the amount of the
immobilized biomolecule and to protect it from outer obstructions.
As biomolecules bind to the functional groups, the support materials should also contain groups
that can be easily modified. A variety of organic and inorganic materials that can be used as
supports are shown in Table I and Table II.
In addition to organic and inorganic supports, there are also hybrid support materials produced by
joining inorganic organic, organic-organic and inorganic-inorganic materials to combine their
individual properties.
Table I: The properties of inorganic support materials
As new biosensing techniques are being developed, it is crucial to measure the efficiency of a
device. Different criteria such as selectivity, precision, sensitivity, working range, limit of
detection (LOD), reusability and time factors can be used to quantify the efficiency.
• Selectivity, ability of a biomolecule to detect a certain analyte without reacting with others,
is the most important feature of a biosensor.
• In addition to being selective to a specific analyte, an ideal biosensor should possess good
precision, which is the ability of providing the same results for each measurement, and
high sensitivity (slope of the calibration curve), which is the minimum quantity of analyte
that can be detected.
• The sensor should also have a wide working range (linear range), the range of analyte
concentration that the sensor can detect and a low LOD, the minimum quantity of an
analyte that is able to induce an identifiable output signal. LOD depends on the affinity
between the recognition element and the analyte, which is expressed via the dissociation
constant “Kd”. A lower Kd value means a higher affinity between the recognition element
and the analyte.
• Reusability represents how many times a sensor can operate before the degradation of the
biomolecule, and this is another desired feature for a biosensor. The reusability of a sensor
also requires high precision in each usage.
Response time, regeneration time and shelf life are key time factors.
• In the context of biosensors, response time is defined as the time needed for the system to
come to equilibrium to make a reliable measurement, and should be short. After an
interaction with the sample, the sensor needs some time to return to its working state. This
is called the regeneration time and should also be quick.
• The shelf life expresses the time range that a biosensor can be used to achieve reliable
results and should be as long as possible. Some of these characteristics are illustrated in
Figure 2.1.
Figure 2.1: Some characteristics of Biosensor
Notes: (a) LOD, linear range and sensitivity; (b) selectivity (selective to analyte 2); (c)
response time; (d) reusability (number of use) and shelf life (days).
In addition to the characteristics explained above, a successful biosensor must also exhibit the
following features.
• The reaction should not be affected by changes in pH value, temperature and stirring.
This feature shall allow the detection without pretreatment or with minimal treatment.
• The biosensor should be inexpensive, tiny, portable and reusable in addition to being easy
to use.
• If the biosensor is going to be used invasively, the probe should be small to reduce tissue
damage, and should also be biocompatible, possessing no antigenic or toxic effects.
• The sensor should also be sterilizable. The response should be correct, reproducible,
precise and linear over the working range. The response should also be free from noise,
which is often a problem for tools measuring voltage.
b. Amperometric
In contrast to potentiometric sensors, where the potential generated across a membrane is used to
convey information on the analyte concentration, amperometric biosensors operate at a fixed
potential with respect to a reference electrode and the current generated by the oxidation or
reduction of species at the surface of the working electrode is measured. Amperometric biosensors
are based on redox enzymes; thus, their appeal is caused by the availability of a large number of
oxido-reductase enzymes that can act on fatty acids, sugars, amino acids, aldehydes, and phenols.
These enzymes use molecular oxygen as an electron acceptor and produce hydrogen peroxide m
the reaction with then substrates.
2. Thermal
Thermal enzyme sensors are based on the principle that the heat evolved in an enzymatic reaction
can be utilized to calorimetrically determine the amount of substrate reacted. Thus, thermometric
indicators or transducers require only a single reaction step producing sufficient or measurable
heat. In thermal enzyme sensors, the enzyme is attached directly to the temperature transducer, a
thermistor, either by crosslinking or by entrapping the enzyme in a membrane enclosing the
thermistor. Alternatively, the enzyme is placed
in a temperature-controlled column and the heat of reaction is measured by
recording the increase in temperature between the inlet and outlet as the sample
flows through the column.
3. Optical
In biosensors based on optical methods, the change in optical properties, such as UV/vis
absorption, bio- and chemiluminescence, reflectance and fluorescence brought by the interaction
of the biocatalyst with the target analyte, is monitored optically.
Potential Applications
The main application areas of enzyme biosensors are m human and animal health care, food and
fermentation industries, environmental monitoring, agriculture and defense.
The second area in which enzyme-biosensors are expected to play a major role is in environmental
monitoring. There is an increasing demand for measuring several classes of substances, including
toxic chemicals, such as polychlorinated biphenyls, polyaromatic hydrocarbons, phenols, dioxins,
organic peroxides, and so forth, pesticides, and heavy metals in water, soil, and air. Enzyme
biosensors will provide rapid, easy to use, and cost-effective on-site testing.
Figure 2.2: Schematic representation of the diffusion of the substrate S and the products P and H+
in the enzymatic layer on a pH electrode.
As shown from Fig. 2.4, an ISFET consists of an ion-sensitive membrane and an FET-structure in
which the metal electrode (gate) is removed and its function is taken over by the solution under
investigation. Transistors for ISFETs are manufactured according to standard planar silicon
technology, and frequently Si3N4/Si02 /Si structures are used.
The operation principle of an ISFET is the formation of a potential difference at the
membrane/solution interface influencing the width of the source-drain channel of the FET. Since
the value of the interfacial potential difference depends on the concentration of corresponding ions
in solution, the change of the FET characteristics (e.g., drain-source-current ID) will be a measure
of the analyte concentration.
2.2.7. Enzyme Biosensors Based on Electron Transfer Between Electrode and Immobilized
Peroxidases
Peroxidases are widely spread in nature and are classified as oxidoreductases E.C. l. 11.1.X, where
X is determined by a biologic reducer. Heme-containing peroxidases are divided into two
superfamilies, viz., plant and mammalian (see Table 1). The latter includes myeloperoxidase,
lactoperoxidase, thyroid peroxidase, and prostaglandin H synthetase. The superfamily of plant
peroxidases consists of yeast cytochrome c peroxidase (CCP), plant ascorbate peroxidases, fungal
peroxidases, and classic plant peroxidases. Plant enzymes in general are more stable than others,
and among them horseradish peroxidase (HRP) is the most commonly used in practical analytical
applications.
Peroxidases oxidize a wide range of electron donors, such as amines and phenols. One-electron
and proton transfer results in formation of substrate radicals, which could interact with each other
forming polymeric products or attack the enzyme itself causing its inactivation. Rapid inactivation
of peroxidases in the reaction course is a well-known phenomenon and this is one of the main
limitations of sensitivity m peroxidase-based assays.
When reaction la proceeds on an electrode surface Compound I is reduced into ferriperoxidase by
a heterogeneous electron transfer (ET) directly from the electrode material as well as by means of
electron mediators. Both approaches result in a reduction current related to the concentration of
hydrogen peroxide in the contacting solution.
Early work on metal-dispersed carbon biosensors focused on the utility of platinized carbon for
enhancing the sensitivity of oxidase- or dehydrogenase-based devices. Palladium-dispersed or
sputtered particles displayed similar improvements. The dramatically improved sensitivity towards
the liberated peroxide or NADH species is attributed to the tine dispersion of the metal
microparticles m the carbon matrix. Such dispersion results in highly catalytic surfaces, as long as
the particle size is comparable to that of the electrical double layer. Unfortunately, the dispersed
platinum or palladium particles are also highly electrocatalytic toward interfering substances.
Investigation has led to the identification of dispersed rhodium, ruthenium and iridium particles as
selective promoters of the peroxide (and to a lesser extent NADH) reactions.
Enzyme sensors based on conductimetric measurement exploit the fact that the changes in
substrate and product concentrations resulting from the catalytic action of some enzymes may be
accompanied by a net change in solution electrical conductivity.
There are a number of mechanisms by which an enzymatic reaction can lead to a conductivity
change. The magnitude of the conductivity change, and to some extent the suitability of a particular
enzyme for the construction of a conductimetric biosensor, depends on the reaction mechanism,
the background conductivity, and the nature of the buffer used.
Electrical conductivity measurement is one of the oldest methods of the analytical chemist.
Conductimetric biosensors are based on the measurement of enzymatically catalyzed changes in
conductivity, but the devices differ in important ways from the analytical chemist’s conductivity
cell.
The thermistor-based calorimeters popularly known as “enzyme thermistors” (ET) were designed
to cover a broader range of applications (6), which included determination of metabolites,
monitoring of enzyme reactions, bioprocess monitoring, characterization of immobilized
biocatalysts, bio-separations, reactions in nonaqueous medium, environmental control, and more
recently investigation of ionic interactions.
The metabolites determined include alcohols, glucose, and lactate, using alcohol oxidase, glucose
oxidase, and lactate oxidase, respectively. The detection limit for these compounds is in the
submicromolar range m pure solutions. The normal procedure is to coimmobilize the oxidases
with catalase to increase total heat production, reduce oxygen consumption, and eliminate
hydrogen peroxide.
The immune system is an attractive subject in much scientific research because of its ability to
process information. The primary function of an immune system is to recognize and identify all
cells and molecules in the system and sort these biological entities as either harmful or not harmful,
as part of the system's defense mechanism. In the presence of foreign substances (i.e. antigens),
specialized immune system cells produce immunoglobulins (i.e. antibodies) that specifically bind
these antigens. This phenomenon has a number of functions including the development of sensors.
A sensor that is based on the concept of immunology (also known as an immunosensor) uses an
antibody (as a bioreceptor) for specific molecular recognition of antigens and subsequently forms
a stable immunocomplex. The immunocomplex is determined and measured by coupling this
reaction to the surface of a transducer (e.g. thermistor). The transducer detects and converts the
reaction to an electrical signal where it can be processed, recorded and viewed. Transducers can
be classified into different categories based on the signal output (e.g. electrical) or alteration in
properties (e.g. mass change) upon the formation of antigen–antibody complexes. Ideally, an
immunosensor should be designed with the following specifications: (i) ability to identify target
antigens quickly; (ii) ability to generate immunocomplexes without the need to add supplementary
reagents; (iii) ability to give results with high reproducibility; and (iv) ability to easily detect the
target in real samples. In immunosensors, detection of the target analyte may be direct by observing
the formation of immunocomplexes, or indirect by utilizing a label, for example, enzymes or gold
nanoparticles to observe a binding event.
Optical immunosensors exploit other methods for the detection of analytes that are based on
chemiluminescence, light absorbance, fluorescence, phosphorescence, light polarization and
rotation, with surface plasmon resonance being the most widely employed. Optical-based
immunosensors observe a variation in phase, polarization speed or frequency of input light that
corresponds to the formation of an antigen–antibody complex. The concept involved in optical
sensors for the identification of analytes is based on the higher dielectric permittivity acquired by
all proteins, cell and DNA, than air and water causing these biomolecules to reduce the propagation
speed of the electromagnetic fields flowing through them. Because all molecules contain atomic
nuclei and electrons in varying orbital states, these molecules are able to interact with the
electromagnetic fields that pass through them. By placing these molecules in oscillating
electromagnetic fields analogous to the propagation of light, electrons within the molecules will
vibrate due to the force subjected to them. Free electrons will then polarize in the presence of
light's magnetic field generating polarization current, resulting from the movement of electrons
where it moves much slower though a biomolecule than in free space. Typically, optical
immunosensors employ light either coming from a laser, diode or white-hot light bulb and enable
observations of any alterations in the attributes of the light reflected from or passed through the
sensor. Measurements can be concurrently or successively carried out by casting light upon the
sensor at various angles on a single sensor plane. An advantage of this type of sensor is the fact
that changes in the light's characteristic is measured and all the light is sourced externally, thus
making it energy efficient; only especially low illumination power is required to generate signals.
Several kinds of transducing components can be used to generate optical change: a grating
couplers resonant mirror, surface plasmon resonance (SPR) interferometry, reflectrometric
interference spectroscopy, ellipsometry, and total internal reflection fluorescence (TIRF).
Recently, techniques exploiting the use of piezoelectric crystal instruments have been designed
and developed for immunosensor utilization. The piezoelectric immunosensor makes use of the
mass sensitivity of piezoelectric quartz crystal in response to the specific antibody–antigen
interaction. This type of immunosensor is becoming popular because of the simplicity, high
sensitivity, specificity, speed, label-free, stability and safety of this technique and is being used in
a number of sectors such as in clinical diagnosis and environmental pollution supervision. Another
advantage of the piezoelectric immunosensor is its ability to detect analytes in real time, which
obeys the pseudo-first order kinetics. The basic principle involved in the mass sensitive
immunosensor is based on the change in mass during the attachment of the analyte to the binding
sites of the antibody. This mass change can be examined by quartz crystal, a main constituent of
the piezoelectric sensor. The piezoelectric crystal oscillates at a specific frequency in conjunction
with the use of an electrical signal at a certain frequency. By applying electrical voltage to the
quartz crystal via two electrodes, the orientation of the crystal is altered and this causes a distortion
in the crystal lattice that causes a mechanical oscillation at a characteristic vibrational
frequency i.e. the crystal's natural resonant frequency. In the event of the formation of
immunocomplexes, the surface of the crystal is loaded with an extra mass, which changes the
frequency of oscillation of the crystal and the mass change can be determined electrically.
Biological reactions are often accompanied by liberation or absorption of heat. This concept is
exploited in the thermal immunosensor to measure temperature change from evolved or absorbed
heat as a consequence of a specific analyte–antibody/antigen reaction as the detection mode.
Variations in temperature can straightforwardly be converted to an electrical signal, which makes
the electrical method the most efficient procedure to determine temperature when constructing a
thermometric immunosensor. The entire heat evolved or absorbed can be measured by evaluating
the molar enthalpy and sum of products produced in the biochemical reaction. The transducer
commonly chosen for the thermometric sensor is a thermistor acquiring an extremely high negative
temperature coefficient of resistance.
The temperature-based detection mode is often coupled with an enzyme thermistor since virtually
all reactions involving enzymes are accompanied with enthalpy changes. It is a common practice
to combine thermal detection of enzymatic reactions with the flow-injection assay (FIA) method
that leads to the development of the thermometric immunosensor. An advantage of this principle
has paved the way for the use of theoretically any type of enzyme conjugated with detection
antibodies.
2.3.6 Fiberoptic lmmunosensors with Continuous Analyte Response
Fiberoptic fluorescence signal transmission has several advantages for immunosensor design:
physical flexibility for remote sensing, no risk of electrical interference, high signal-to-noise ratio
with little attenuation over distance, and the capacity to both measure several analytes with
fluorescence from a single fiber and bundle fibers without significant crosstalk. For many
immunosensors these possibilities have offered little advantage because they have been designed
for single use, or they have required regeneration that usually cannot be accomplished in situ.
Basis of the lmmunosensor
The indicator system is membrane-encapsulated at the end of an optical fiber and depends on
changes in fluorescence energy transfer between the donor, B-phycoerythrin (BPE), and the
acceptor, Texas Red (TR). The intensity of fluorescence from BPE labeled with phenytoin (BPE-
PHT) or other analyte is decreased when the antibody, labeled with TR, binds to phenytoin. Figure
1A shows the condition when antibody-hapten binding places the BPE and TR in close proximate
geometry for efficient energy transfer and thus quench of BPE fluorescence. In the absence of
analyte phenytoin, antibody molecules are fully packed around BPE and fluorescence quenching
is maximal. When free phenytoin is present it competes with the BPE-PHT for TR- antibody. A
portion of the TR-antibody binds to analyte phenytoin, making it unavailable to quench BPE,
increasing the fluorescence signal (Fig. 1B). At equllibrium, the concentration of free phenytoin
in the test solution produces a characteristic fraction of TR-antibody bound to BPE-PHT, with a
unique and reproducible fluorescence intensity. In this design the main determinants of response
time are the effective kdls, the reagent solution viscosity, and membrane interactions. Simulations
indicate that the association rate constant (kass) has little impact on the time to reach equilibrium
after a change in analyte concentration, whereas a large kdrs is essential for a rapid response.
Changing kass by three orders of magnitude has little effect on the simulated response time.
However, if the kass is held constant at 4 x 106 L/mol/s, increasing the kdls, one order of magnitude,
for example from 4 x 10-4 to 4 x 110-3 s-1, results in a 10-fold reduction in response time (162 to
16 min). Because kdls rates may vary by as many as eight orders of magnitude, antibodies differ in
suitability for continuous measurement systems and should be chosen Judiciously to achieve an
appropriate sensor response time. The phenytoin sensor described here uses antibodies with a kdls
of 4 x 10-3 s-l, and intact sensors have response times of 15 m in aqueous solution. Reagent viscosity
can impact response time because the free analyte must diffuse into and throughout the reagent
chamber to affect a change m fluorescence. For example, when dextran 70 kDa is included in the
sensor for plasma or blood to balance oncotic pressure in the specimen, the response time is 1 h.
Because of viscosity effects, and other unpredictable interactions between reagents and the
encapsulation membrane, sensor performance may differ substantially from theoretical. The best
approach for sensor design may be to select antibodies with a relatively large kdls and satisfactory
specificity, then test intact sensors to determine if the response time is appropriate.
Materials Used
Instrumentation
1. The optical components of the immunosensor are shown in Fig. 2. 1. Excitation source: a
100-mW air-cooled argon Ion laser (model 5500A from Ion Laser Technology, Salt Lake
City, UT) tuned to 5 14.5 nm and attenuated to 0.1% with a neutral-density filter.
2. Reference channel beam splitter (conventional glass microscope slide at 45” to the incident
beam) and photometer
3. Off axis parabolic mirror (#02POAOl5, Melles Gnat, Irvine, CA) of 66-mm focal length
with a 2-mm hole drilled on the optical axis.
4. Focusing lens, fused silica with a 25-mm focal length (#OlLQB028, Melles Griot).
5. Fiber positioner (#FP-2, Newport, Fountain Valley, CA).
6. Emission monochrometer set at 577 nm.
7. Optical fiber 100/l 10 l.un core/cladding fiber (Superguide G #B2-0007-20, Fiberguide
Industries, Stirling, NJ)
8. The remaining optical components (breadboard, photomultiplier tubes, photometers, lens
mounts, shutter, and filters) are standard.
Affinity Biosensing Based on Surface Plasmon Resonance Detection
The idea that receptors could be incorporated with potentiometric electrodes to produce biosensors
was proposed as early as 1975. Neurotransmitter receptor proteins are excellent candidates as
sensing elements in biosensors because of their high affinity and selectivity for specific ligands.
They can recognize families of chemicals of physiological, pharmacological and toxicological
significance, ranging from amino acids and peptides to therapeutics, drugs of abuse and toxicants.
However, the highly complex structures of neuroreceptors, their generally labile nature at room
temperature, and the difficulty in obtaining sufficient quantities of receptor protein for biosensor
studies have restricted attempted biosensor applications of these proteins until recently.
UNIT III
The DNA-based sensors should have a rapid response time and should be quantitative, sensitive,
suitable for automation, cost effective, disposable and solve analytical problems in a wide range
of industrial contexts in order to be commercially viable.
Several kinds of optical transducers can form the basis of a DNA biosensor. Such transducers
include fluorescence, surface plasmon resonance and Raman spectroscope.
Fiber Optics
The fluorescent DNA stain ethidium bromide (EB) is a commonly used dye for the detection of
DNA. The fluorescent ethidium cation strongly associates with dsDNA by intercalation into the
base stacking region and, in some cases, the major groove of the double helical structure. EB has
an absorption maximum of 510 nm; detection of dsDNA is achieved by exposing it to an EB
solution, washing off the unbound EB, and measuring the fluorescence intensity by UV visible
spectrometry. The fluorescence intensity is directly proportional to the amount of EB-intercalated
DNA. Biosensors can be designed to use this principle.
Resonant Mirror
Optical methods, such as those using surface plasmon resonance (SPR) and monomode dielectric
wave guides, have been extensively developed in recent years. The resonant mirror is an
evanescent wave sensor which has been designed to combine the simple construction of SPR
devices with the enhanced sensitivity of wave guiding devices. Detection of enzyme substrate,
antibody-antigen, and protein-cell interactions by the resonant mirror device has been reported.
This technology has recently been applied to direct and rapid detection of DNA-DNA
hybridization. Biotinylated oligonucleotide probes were immobilized on the sensor’s surface, via
streptavidin, and hybridization of a complementary target oligonucleotide (40-mer) was monitored
in real time. The interaction at the sensory surface was shown to be sequence specific under
conditions of low stringency.
Raman Spectroscopy
Raman spectroscopy is a useful tool for chemical analysis because of its excellent capability for
chemical group identification. One limitation of conventional Raman spectroscopy is its low
sensitivity; hence, effective measurements often require powerful and costly laser sources for
excitation. However, this method has attracted renewed attention because of the recent observation
that Raman scattering efficiency can be enhanced by factors of up to 10& when a compound is
adsorbed on or near special metal surfaces. This modified technique is known as surface-enhanced
Raman scattering (SERS) spectroscopy. In 1994, Vo-Dinh et al. reported the development of a
new type of DNA probe based on surface-enhanced Raman scattering detection. These surface-
enhanced Raman gene (SERG) probes do not require the use of radioactive labels, while having a
great potential for sensitivity and selectivity.
Apparently electrochemical transduction-based DNA sensors are easy to construct, needing only
two electrodes and a voltammeter; however, the sensitivity and specificity of such constructs is
poor.
1. The fluorescence polarization measurements are made using an SLM 8000C scanning
spectrofluorometer, manufactured by SLM Aminco Instruments, Inc * (Urbana, IL). Other
spectrofluorometers with similar specifications and components can also be used This
instrument uses a 450-W xenon arc lamp as the excitation source and a double-grating
monochromator for the selection of the excitation wavelength. A single-grating
monochromator monitored the fluorescence. The excitation and emission paths contained
adjustable Glan-Thompson polarizers and the normal 90’ fluorescence geometry was used.
Photomultipliers monitored two channels; a reference channel monitoring the xenon lamp,
consisting of a concentrated solution of rhodamine B, which served as a quantum counter,
and the fluorescence signal channel. The fluorescence signal was measured and normalized
by the reference signal to minimize the effects of lamp fluctuations.
2. A UV-VIS spectrophotometer, used for measurement of the DNA absorption at 260 nm for
concentration determinations, is available from Perkin-Elmer Corp. (Norwalk, CT), Milton
Roy (Rochester, NY), and many other instrument companies.
3. Standard buffer: 8 mM Tris-HCl, 50 mM NaCl, and 1 mM EDTA in nanopure distilled
water; adjust pH to 7.0 with 3 M HCl. Autoclave the buffer for 20 min at 20 psi.
4. Calf thymus DNA (sodium salt): Take one vial (2 mg) and dissolve m 50 mL of buffer by
placing the DNA and buffer m a screw-capped vial and agitate for at least 12 h. Determine
the concentration of DNA in solution by measuring the absorption at 260 nm in a 1 -cm
path length quartz cuvet (1.00 absorption units of duplex DNA is assumed equal to 50
𝜇g/mL of DNA in solution). The typical DNA concentration range is between 32 and 33
𝜇g/mL.
5. Dilute suspension of glycogen.
6. Acridine orange: 2 x 10-5 M in buffer
Cell-Based Biosensor
Cell-based biosensor can detect biochemical effects directly via living cells and convert these
effects into digital electrical signals by sensors or transducers. Hence, it serves as the bridge
between biology and electronics. Cell-based biosensors combine living cells and sensors or
transducers for cellular physiological parameter detection, pharmaceutical effect analysis,
environmental toxicity test, etc. In contrast to molecule-based approaches, cell-based biosensors
have a broad spectrum of detection capabilities. Moreover, in addition to analyte sensing and
detecting, cell-based biosensors can provide the advantages of rapid and sensitive analysis for in
situ monitoring with cells. Cells naturally encapsulate molecular sensor arrays. Enzymes,
receptors, and ion channels, all with a stable status, could respond to their corresponding analytes
via a native cellular mechanism. Compared with molecular biosensors, cell-based biosensors are
expected to respond optimally to bioactive analytes. Therefore, cell-based biosensors provide a
useful tool to study the physiological effects of analytes.
Because of the obvious advantages of cell-based biosensors, e.g., long-term recording in a
noninvasive way, fast response time, and label-free experimentation, they have been widely
utilized in many fields such as cellular physiological analysis, pharmaceutical evaluation,
environmental monitoring, and medical diagnosis. In these applications, cell lines and primary
cultured cells are mainly selected as the cell sources. Cell lines divide actively in vitro, which
offers the convenience of preparation and culture, if the desired cell type is available.
The conventional cell-based biosensors usually use cell population recording to study the behavior
of cell populations on the basis of sensor platforms. The recorded data are often the collective
responses of many cells detected by the sensor electrodes, which can be applied for cell−cell
interaction study, such as cell invasion and barrier function assessment.
Cell-based biosensors enable analysis and monitoring of drug-ligand interactions, effect of
bioactive agents, environmental toxicity studies, etc. in a more physiologically relevant way. With
the use of suitable substrates, it is possible now to use cell-based biosensors for in situ (ex vivo in
some cases) monitoring with living cells in their native environment. Different cell types such as
bacteria, yeast, fungi, algae, and other higher eukaryotes including fish, rat and human cells have
been utilized for biosensor fabrication (Figure 3.1). Microbial cells, including bacteria, fungi,
yeast and algae, have largely been utilized in biosensors for water quality monitoring and toxicity
assessment.
Figure 3.1: Schematic showing the different cell types (left) and functional strategies (right)
utilized in cell-based biosensors.
Types of immobilization
Passive immobilization
Several types of immobilization strategies have been explored for attachment of living cells onto
transducer/detecting platforms (Figure 3.2). In general, the factors that help in determining a
particular immobilization strategy for a biosensor include: 1) the cell types used as bioreceptors,
2) the type of material/substrate onto which cells are to be immobilized, 3) toxicity of the substrate,
4) cell viability and growth requirements, and 5) the application of the biosensor. For example,
applications such as monitoring of biochemical oxygen demand (BOD) and cell migration rely on
the viability and metabolism of the cells used, hence it is imperative that immobilization should
not hinder the cell growth or viability.
The functional strategy of a biosensor is also an important determinant of immobilization method.
For example, bio-luminescence mediated detection of target analytes (e.g. heavy metals)
necessitates the microbial cells to be metabolically active and expressing the bioluminescent gene
in a stable manner. Thus, such biosensors often involve immobilization strategies that do not affect
cell viability, such as physical adsorption, hydrogel/sol gel entrapment/encapsulation and biofilm
formation. Since such techniques rely on the cells’ natural ability to adhere to surfaces, they are
collectively referred to as passive immobilization techniques.
Entrapment in sol gel/hydrogel is also a prominent cell immobilization technique. Some of the
most commonly used gel matrices are calcium alginate (or, alginate-starch), agarose and chitosan
Nanomaterials have emerged as promising candidates for cell immobilization. Magnetic Fe3O4
nanoparticles have been employed for labeling Bacillus subtilis cells electrostatically and
subsequently immobilizing them on ultramicroelectrode array (UMEA) using external magnetic
field. This modified UMEA was subsequently utilized as a BOD microsensor and shown to be
regenerable and reproducible with better sensitivity than dissolved oxygen (DO)-based and
mediated BOD sensors. Another study has been reported on the immobilization of E. coli cells on
nanoporous gold for selective detection of sulfide. Silicon-based nanostructures have been utilized
for both immobilization of breast cancer cells and improving sensitivity of the biosensor in terms
of monitoring drug effects at very low concentrations (2 nM). Thus, nanomaterials have shown
potential as robust substrates that serve the dual purpose of cell immobilization and signal
enhancement.
Active immobilization
Active immobilization techniques are an efficient alternative to passive immobilization since these
techniques involve the utilization of chemical linkers or other agents for immobilization of living
cells onto a suitable substrate, e.g. glutaraldehyde crosslinking. These methods allow tighter
binding of the cells to transducer surface leading to enhanced sensitivity. A recent study highlights
the significance of covalent (or active) immobilization of cells for improving the stability and
sensitivity of an optical biosensor against methyl parathion. Biomolecules that interact directly
with receptors and other moieties present on the cell surface have been used to immobilize bacterial
cells on various substrates. Leonard et al. developed an antimicrobial susceptibility testing (AST)
platform by immobilizing E. coli cells on Si micropillar arrays via wheat germ agglutinin (WGA).
Bacteria-specific antibodies have also been utilized for efficient immobilization on plastic and
glass surfaces.
Some immobilization strategies are unique for higher eukaryotic cells, such as utilization of the
extracellular matrix (ECM, a semi-solid network of various proteins and fluids to which cells are
bound in a multicellular organism) and its components for immobilization. It has been reported
that surface modification by ECM components enables uniform immobilization of cells while
maintaining their viability, which is a crucial factor in many biosensing applications. Collagen,
fibronectin and laminin together with peptides such as poly-L-lysine have been predominantly
used for the immobilization of various cell types including mouse germ cells, taste receptor cells,
myocytes and MCF-7 breast cancer cells.
Microbial biosensors
Many different types of microorganisms have been utilized for biosensor fabrication over the
years, including bacteria, photosynthetic algae and yeast cells with prominent applications in
environmental monitoring and antimicrobial susceptibility testing. Some of these biosensors
exploit the natural ability of a particular microorganism to sense a specific analyte while several
others are dependent on the introduction of specifically designed genetic constructs in a host
microorganism for detection of bioavailable analytes. Different types of transducing/detecting
systems have been utilized in microbial biosensors, electrochemical and optical being the most
prominent. Electrochemical readouts exploit the activation/production of redox active species by
the microbes for biosensing. In some cases, redox-active mediators may be externally
supplemented to improve electron transfer across the electrode. Thus, genetic modification is not
always required in electrochemical sensing. However, electrochemical systems are often
vulnerable to high noise and background due to interfering species. Optical readouts, on the other
hand, rely heavily on genetic modification for the microbes to produce a visual/fluorescent
response against a specific stimulus. Most genetic modifications involve the introduction of an
analyte-specific promoter upstream to a ‘bioreporter’ gene in the host microbial cells. Upon
exposure to the target, the promoter gets activated and the bioreporter gene expresses fluorescent
proteins. Thus, optical biosensors are generally more specific than electrochemical ones, however
they require a number of additional steps involving gene modification and hence, increase the cost
of the device considerably.
Electrochemical Biosensor
The electrochemical biosensor is one of the typical sensing devices based on transducing the
biochemical events to electrical signals. In this type of sensor, an electrode is a key component
that is employed as a solid support for immobilization of biomolecules and electron movement.
Thanks to numerous nanomaterials that possess the large surface area, synergic effects are enabled
by improving loading capacity and the mass transport of reactants for achieving high performance
in terms of analytical sensitivity.
The electrochemical biosensor is the analytical devices that transduce biochemical events such as
enzyme-substrate reaction and antigen-antibody interaction to electrical signals (e.g., current,
voltage, impedance, etc.). Since Clark developed the 1st version of electrochemical biosensor for
blood glucose, various types of biosensor have consecutively been introduced and commercialized
for diverse applications. In this electrochemical biosensor, an electrode is a key component, which
is employed as a solid support for immobilization of biomolecules (enzyme, antibody and nucleic
acid) and electron movement. Various chemical modification methods are applied for this purpose
via amine- and carboxyl (1-ethyl-3-(3-dimethylaminopropyl)carbodiimide: EDC), aldehyde-
(hydrazide) and thiol (maleimide), depending on the chemical groups on the electrode in the
presence of or absence of supporting materials. Since inappropriate immobilization may cause loss
of activity, less specificity, and low biocompatibility, it is crucial not only to maintain orientation
and biological activity of the biomolecules upon immobilization. In addition, employing proper
functional material for the electrode is a key process for the high performance of biosensors.
Optical Biosensors
An optical biosensor is a compact analytical device containing a biorecognition sensing element
integrated with an optical transducer system (Figure 3.2). The basic objective of an optical
biosensor is to produce a signal which is proportionate to the concentration of a measured
substance (analyte). The optical biosensor can use various biological materials, including enzymes,
antibodies, antigens, receptors, nucleic acids, whole cells and tissues as biorecognition elements.
Surface plasmon resonance (SPR), evanescent wave fluorescence and optical waveguide
interferometry utilize the evanescent field in close proximity to the biosensor surface to detect the
interaction of the biorecognition element with the analyte. Optical detection in these sensors is
performed by exploiting the interaction of the optical field with a biorecognition element.
Figure 3.3: Schematic illustration of LSPR transmission (left) and reflection (right) modes.
Figure 3.4: Schematic diagram of an evanescent wave planar optical waveguide chip
In these biosensors, the biological recognition and the consequent binding event occur within the
confines of an evanescent wave. The evanescent wave arises from the manner in which light
behaves when confined in an optical waveguide or fibres. Guided light is totally internally reflected
when it meets the interface of the waveguide/fibre and a surrounding medium with a lower index
of refraction, as a result an electromagnetic field called an evanescent wave extends out from the
interface into the lower index medium. The evanescent wave decays exponentially with distance
from the surface, generally over the distance of 100 nm to approximately a wavelength. Since the
evanescent wave is such a near-surface phenomena, detection employing evanescent wave
excitation to generate the fluorescent signal is surface-sensitive, meaning that only fluorescent
molecules near the surface are excited (Figure 3.4). This geometric limitation can help to minimize
unwanted background signal from a bulk sample while only enhancing the signal from
fluorophores captured on the surface. A profuse variety of biosensors was developed based on this
principle with a wide array of applications ranging from clinical diagnostics to biodefence to food
testing.
Biosensors generally provide a rapid and convenient alternative to conventional methods for
monitoring chemical substances in fields as diverse as medicine, environmental monitoring, and
food processing. A biosensor is a device that combines a biologic-sensing element with an
electronic signal-transducing element. As biologic-sensing elements, enzymes, antibodies,
receptors, organelles, and microorganisms, as well as animal and plant cells, can be used. Among
these biologic-sensing elements, microorganisms have several advantages, as listed below (I): 1.
Wide range of suitable pH and temperature. 2 Long lifetime. 3. Low cost (because no isolation of
the active enzyme is needed) Purified biomolecules, such as enzymes, are used as the sensing
element in most biosensors. Although enzymes show high specificity for their substrates, they are
generally expensive and have poor stability. Therefore, microorganisms have been employed as
the sensing element of biosensors to solve these problems. They are suitable for on-line
environmental monitoring that requires long-term stability. Microbial biosensors have been
developed for assaying biochemical oxygen demand (BOD), a value related to the total content of
organic material in waste water. Microbial sensors, of course, have some disadvantages, as listed
below: 1. Long response time. 2. Long time to return to baseline. 3. Lower selectivity (because a
microorganism contains many enzymes)
Therefore, it is necessary to optimize measurement and storage conditions to shorten the
measurement time of microbial sensors. The low selectivity of microbial sensors can be
complemented by use of membranes, such as a gas-permeable membrane. More recently
developed microbial sensors are reviewed by Karube et al. The principle of microbial sensors is
shown in Fig. 1. A typical microbial sensor consists of an oxygen electrode to which a membrane
containing immobilized aerobic microorganisms is attached. The respiratory and metabolic
functions of the microorganisms are used to detect particular substances. Immobilization of the
microorganisms is a key step in the fabrication of microbial sensors. Immobilization should not
harm the microorganism, and rt should, ideally, improve their stability for use in continuous
monitoring. When analytes are degraded by immobilized microorganisms, respiratory activity
increases and dissolved oxygen is consumed. The oxygen concentration is monitored using the
oxygen electrode. A large decrease m current coming from the oxygen electrode indicates high
metabolic activity of the immobilized microorganisms, which in turn indicates a high
concentration of analyte. Some sensors use microorganisms whose respiration is inhibited by the
target analyte, as in the case of sensors for cyanide or pesticides.
Immobilization of Microorganisms
Immobilization methods must not harm the immobilized microorganisms and yet provide a sensor
with a good stability under the prevailing conditions of temperature, ionic strength, pH, redox
potential, and chemical composition. Major methods of immobilization are listed as follows: 1.
Physical adsorption method. 2. Entrapment method. 3 Crosslinking method. 4 Covalent binding
method.
Methods of microorganism immobilization are usually modified versions of methods used for
immobilizing enzymes. Immobilization of microorganisms is also carried out in reactors used in
the food industry (e.g., beer and wine production) and in waste water treatment reactors. The most
widely used methods of immobilization are physical adsorption or entrapment methods. Since
chemical methods often damage the cell membrane and decrease biological activity, those methods
have seldom been successfully applied.
Oxygen Electrodes
One of the most suitable transducers for microbial sensors is an electrode, especially the Clark-
type oxygen electrode. The oxygen electrode measures the change in dissolved oxygen
concentration resulting from the respiration of the immobilized microbes. Clark-type electrodes
are either polarographic or galvanic in nature. Polarographic electrodes consist of a platinum
cathode and a silver anode, both of which are immersed in an electrolyte solution of saturated
potassium chloride. A suitable polarization voltage between the anode and the cathode is -0.6 or -
0.7 V, and oxygen is selectively reduced at the cathode at this voltage.