You are on page 1of 4

ARTICLE BACK TO BASICS

pH meters and their


electrodes: calibration,
maintenance and use
lithium-based glass electrodes are now used
Following consideration of centrifugation and liquid handling in the exclusively for pH-responsive glass electrodes.

February and March issues, respectively, of The Biomedical Scientist, POTENTIOMETRIC MEASUREMENT
Almost all conductors of electricity are
Peter Riddle now directs attention to pH meters, their electrodes and metal or an electrolyte, with the current
being carried by either electrons or ions.
the measurement of hydrogen ion concentration. When current passes from metal to
electrolyte, or from electrolyte to metal, the
type of carrier usually changes suddenly, and
As living processes are dependent on the around 1928 the first commercial pH meter whenever there is an interface between the
unique ionising solvent we know as water, was produced by the Cambridge Instrument metal and ions of that metal in a solution
and as much of classical chemistry was Company. Both electrodes were subject to then an electric potential is produced.
concerned with the study of reactions in interference from the components of redox This potential is called the electric potential
aqueous solution, the ubiquity of the pH systems that reduced the effectiveness of of that metal. An electrode potential is also
meter in present-day biomedical laboratories these electrodes to measure the potential produced when different concentrations of
is scarcely surprising. The concept of response due to hydrogen ions. Membrane an ion are separated by a membrane that is
electrolytic dissociation was introduced over electrodes, including the glass electrode, semipermeable to that ion. Non-metallic
a century ago by Arrhenius, and in 1909 are not subject to these limitations and elements such as hydrogen also have
Sorenson demonstrated the importance of electrode potentials.
hydrogen ion concentration on enzyme In order to measure an electrode potential,
activity and coined the term ‘pH’. Although ‘The hydrogen electrode another voltage source (such as another
that same year saw the first systematic study metal/solution interface) is needed to
of the glass electrode, it was not until the was the first electrode to be measure it against. Each of the electrodes is
development of convenient electronic called a half-cell. The two half-cells arranged
techniques capable of accommodating the used for pH measurement together constitute an electrochemical cell,
high electrical resistance of the glass in which one of the half-cells maintains a
electrode that interest in electrometric pH but it required a supply of constant voltage. The electrode in the half-
determination and its importance in cell with the constant voltage is called the
biological systems became clear. hydrogen gas and thus its reference electrode, whereas the variable
The pH meter is designed to measure the voltage portion is termed the indicator
concentration of hydrogen ions in a solution. application was limited’ electrode. It is possible to measure the
Basically, three parameters are involved in the
measurement: the actual molar concentration
of hydrogen ions, the dissociation constant
of the acid (pKa), and temperature. pH is
defined as the negative of the logarithm to Ag/AgCl/HCl Glass Test or standard KCl/Hg2Cl2/Hg
membrane membrane
the base 10 of the hydrogen ion concentration
(pH = –log10 [H+]).
The hydrogen electrode was the first
electrode to be used for pH measurement but Indicator glass electrode Reference electrode
it required a supply of hydrogen gas and thus
its application was limited. The quinhydrone Fig 1. The composition of the two half-cells that permits the
electrode appeared in the early 1920s and difference in potential to be detected between them.

202 THE BIOMEDICAL SCIENTIST APRIL 2013


ARTICLE

Wire connection point Wire connection point

Glass body
Glass or
plastic body

Seal
Filled with Silver
potassium chloride wire
‘buffer’ solution Silver
wire
Bulb filled with – + + – Very thin glass
potassium – + + – bulb chemically
chloride ‘buffer’ – + + – ‘doped’ with
solution Silver
Silver – + chloride tip +
+ – lithium ions so as
+ to react with
chloride tip – + + – hydrogen ions
– + + ++ + –
Porous junction – – – outside the bulb
– –
Voltage produced across thickness
of the glass membrane

Fig 2. The reference electrode. Fig 3. The indicator (glass) electrode.

potential difference between these two concentration through the Nernst equation, is desired, a 0.1 mol/L or 1 mol/L potassium
electrodes and calculate the concentration which, in its simple form, is E=KT log chloride solution electrode is preferred
of ions in the solution of the indicator (C1/C2) where K = Constant 2.3 (R/zF) because it reaches its equilibrium potential
(measuring) electrode. and if either C1 or C2 is known then the more quickly and its potential depends less
For example, if a silver wire is immersed in concentration can be calculated from the on temperature than does the saturated type.
a solution of silver chloride, ionisation of the measurement of the potential developed. The silver–silver chloride electrode is a very
silver metal occurs, with the formation of In making an electrical connection reproducible electrode. The solution
silver ions (Ag+) and electrons. An electric between the reference electrode and the surrounding the electrode should be
potential now exists between the wire and the sample solution via a salt bridge, a negligible saturated with potassium chloride and silver
solution. If two half-cells are used, each with a but reproducible potential is produced. This chloride.
silver wire or foil immersed in a different silver potential develops at the interface between
solution, and the two solutions are connected two non-identical solutions and is called the INDICATOR ELECTRODES
through a meter, a difference in potential liquid–liquid junction potential. A saturated The glass electrode was the first and still most
can be detected between them (Fig 1). potassium chloride solution is normally used commonly used electrode for measuring
As the potential of each solution depends on as a salt bridge in the reference electrode hydrogen ion activity. A glass electrode consists
the concentration of silver ions in it, the because many ions diffuse from the salt of a small bulb of special glass that contains a
concentration of ions in one solution can be bridge against which the sample ions must solution of known hydrogen ion concentration
predicted if the value for the other solution diffuse. Thus, diffusion of ions from the (eg 0.1N HCL or KCL ‘buffer solution’) and an
and the difference in potential between sample to the junction is negligible and the internal reference electrode (usually calomel
them are known. A temperature difference junction potential is ≤1 mV and is or silver–silver chloride; Fig 3). The sensor is
between the two half-cells would affect the reproducible. normally a platinum wire.
reproducibility of measurement. Other minor The pH electrode is manufactured with
technical factors, such as coating (eg protein), REFERENCE ELECTRODES different properties, depending on the
will affect the measurements. The standard hydrogen electrode is the application. Those used in biomedical
Potentiometric methods are based on international standard but is seldom used for laboratories should be linear at about pH 7.0.
the quantitative relationship between the routine work because a supply of hydrogen Wide-range pH electrodes (1–14) tend to be
potential of a cell as given by the following gas is required and more convenient types non-linear towards the ends of the range.
distribution of potential: (together with reliable calibration buffers) are Some are designed to be linear for low pH
Ecell = Ereference + Eindicator + Ejunction available. The saturated calomel electrode values, while others are linear for high pH
is a widely used reference electrode that values. These properties are controlled by the
Because the reference and junction potentials contains an inert element (eg platinum) in glass formula.
are constant, the indicator potential can be contact with mercury, mercurous chloride
determined. The potential of the indicator (calomel) and a solution of potassium COMBINATION ELECTRODE
electrode can then be related to chloride of known concentration. However, It is inconvenient in many applications to
concentration. Electrode systems used in in view of environmental considerations, the have to contend with two measuring
biomedical laboratories have precalibrated mercury electrode is rarely used these days electrodes (reference and indicator).
read-out devices that give results in and such electrodes are not suitable for Combination electrodes are available into
concentration units. varying temperatures or temperatures above which both the reference (usually Ag–AgCl)
The cell potential is related to 60˚C (Fig 2). When a high degree of accuracy and indicator electrodes are incorporated.

APRIL 2013 THE BIOMEDICAL SCIENTIST 203


ARTICLE

The pH membrane is usually recessed in, ‘Both standard moving-pointer INSTRUMENT CALIBRATION
and protected by, a plastic housing. Such It is essential that pH electrodes are
electrodes will withstand more physical abuse meters and digital display calibrated regularly using the meter and
than will conventional pH electrodes. reference electrode with which they are to be
The pH of a solution is a function of its units are common, although used in the laboratory. The procedure is as
temperature. Voltage output from the follows:
electrode changes linearly in relation to most new instruments 1 Turn instrument on and allow adequate
changes in pH, and the temperature of the time to warm up (15–30 min)
solution determines the slope of the graph. incorporate digital displays’ 2 Position electrodes in holder and plug into
pH applications therefore require some form instrument
of temperature compensation to ensure 3 Set function switch to pH
standardised pH values. Some manufacturers potential difference between the two half-cells 4 Prepare pH buffer solutions for integer pH
use a temperature-compensating electrode. is measured by a very sensitive voltmeter. values. These are available in dry form in
This is a resistance thermometer that is When the instrument is calibrated against foil packets and are prepared using
inserted into the solution under test, together standards and adjusted for the temperature deionised water. Preprepared liquid buffers
with the glass and reference electrodes. effect, the hydrogen ion activity can be read can also be used
The electronic mechanism of the amplifier very accurately. Owing to the high resistance 5 Set temperature compensation control for
is designed so that when the pH meter is of the glass electrode, electronic amplification temperature of the buffers (not room
used in conjunction with the resistance is necessary for measurement of the potential temperature). Most modern meters and
thermometer the pH is independent of and a very sophisticated, stable and sensitive electrode systems have automatic
temperature, so long as the temperature measuring device is required. This is what we temperature-compensation correction
range is within ±5˚C of the temperature of now know as a pH meter. 6 Insert electrodes in pH 7.0 buffer
the buffers used to calibrate the instrument. The voltage produced by the electrodes 7 Adjust calibration control until meter
Other instruments use electronic initially is amplified and compared against a indicates pH 7.0 – modern pH meter
compensation for temperature changes. standard reference voltage. The output of the models working in calibration mode often
In these instruments the temperature of the voltage comparator goes to the display recognise the buffer and take necessary
calibration buffers or sample is set with a electronics, which convert the comparator action automatically
control on the instrument. output to a suitable form for the display unit. 8 Do not readjust calibration control for
Both standard moving-pointer meters and steps 9 to 12
MEASURING DEVICE digital display units are common, although 9 Remove electrodes, flush with deionised
When the indicator electrode and the most new instruments incorporate digital water and blot gently to remove excess
reference electrode are immersed in a displays. The critical portion of most pH water
solution containing hydrogen ions, the small PathPracNov7_PathPrac_Nov7
meters is a potentiometer circuit. 08/11/2012
10 Start with 09:25
the highest Page and
pH buffer 1 place

Medical
EUROIMMUN UK Ltd. Laboratory
Diagnostics

Antibody diagnostics in
autoimmune nephropathies

Primary MGN* Lupus nephritis

Anti-dsDNA
Anti-PLA2R Anti-nucleo-
somes

Goodpasture‘s
Renal vasculitis
syndrome

Anti-PR3
Anti-GBM
Anti-MPO

ELISA
PLA2R GBM
dsDNA-NcX MPO
Nucleosomes PR3-hn-hr

EUROLINE
ANA
MPO/PR3/GBM

IIFT
Transfected cells (PLA2R)
Purified antigens** (PR3, MPO, GBM)
Cultured cells (DNA, nucleosomes, ANCA)
Tissue sections (GBM)

*
Membranous glomerulonephritis
**
EUROPLUS

EUROIMMUN UK Ltd. · 131 - 139 The Broadway · London · SW19 1QJ


Tel 020 8540 7058 · E-mail mail@euroimmun.co.uk · Internet www.euroimmun.co.uk

204 THE BIOMEDICAL SCIENTIST APRIL 2013


ARTICLE

electrodes in the solution. Read and record Table 1. Some recommended cleaning solutions for glass electrodes.
pH value. The highest pH buffer
represents the lowest hydrogen ion Deposit Cleaning solution
concentration. Calibration in this manner General deposits Mild detergent
minimises test solution carry-over between
Inorganic coatings Commercial glass cleaning solution (not strongly acidic)
measurements
11 Repeat steps 9 and 10 with each Metal compounds Acid solution (not stronger than 1 mol/L)
successively lower-value buffer and record Oil or grease Complexing agent (EDTA) or suitable solvent
all results
Resins, lignins Acetone, alcohol or detergent (not strongly alkaline)
12 Plot calibration curve.
Proteins, blood etc Enzyme solutions (eg pepsin in 0.1 mol/L HCl)
Most modern instruments have a built-in Stubborn deposits Hydrogen peroxide, sodium hypochlorite
automatic buffer recognition facility and will
CAUTION: Care must be exercised when using solvents to clean electrodes that have a plastic body
automatically identify and set to the
or a plastic protective skirt.
appropriate temperature-corrected calibration
values. This can be over-ridden in some
instruments to allow free entry of the actual wet). This does not apply to combination r Handled carefully – the normal lifetime of
buffer pH. or gel electrodes, as these must be stored glass electrodes is approximately two years.
in a concentrated solution of KCl only.
ROUTINE USE Never store your electrode in water (see Occasionally, functional failure occurs before
A considerable variety of instruments from below). Always rinse thoroughly with mechanical failure. This is recognised by a
several manufacturers is available and the deionised water after use. If the response gradually increasing electrode response time,
user is advised to follow the operating of a glass electrode has become sluggish, with increasingly erratic readings. This is a
instructions supplied with the instrument. the recommended treatment (which different effect from electrode shock, which
The glass electrode can be used with should only be performed when other also produces increased response time.
strong acids; however, it is attacked by strong measures have failed) consists of Electrode shock is produced by dipping the
alkaline solutions. Therefore, glass electrodes 1 minute in 20% ammonium bifluoride electrode into a high-concentration solution
should never be left in alkaline solutions for solution, followed by 15 seconds in and then immediately afterwards into a low-
longer than is necessary to measure the pH. 6 mol/L hydrochloric acid. Care should be concentration solution, or vice versa. Thus, if
The glass electrode responds rapidly to large exercised when carrying out this one tries to measure pH 2 and then pH 11, an
pH changes in buffered solutions, but the treatment as the risk of the formation of increased response time should be expected.
response is slower in poorly buffered, or hydrofluoric acid is present. The electrode
unbuffered, solutions. Equilibrium is reached should then be rinsed thoroughly and ELECTRODE CLEANING
slowly and may require several seconds. soaked for 24 hours in water or in an The solution used to clean pH electrodes
Poorly buffered solutions should be stirred acidic buffer solution. depends on the presence of possible
vigorously during measurement to prevent r Electrodes that have been allowed to contaminants. Mechanically intact electrodes
stagnation at the electrode. dry out (often indicated by a hard, dry may show slow response due to clogging or
Measurements can be made in partly deposit of KCl crystals) should be soaked coating. Table 1 shows some recommended
aqueous solutions but the degree of hydration overnight in warm deionised water. Liquid cleaning solutions for glass electrodes.
of the outer surface of the membrane will junctions with fibres or ceramic pins More detailed information about electrode
alter the potential across the membrane. occasionally can become blocked due to care and maintenance can be obtained from
Hence, values obtained in non-aqueous, or crystallisation (eg KCl). If soaking in KCl the manufacturer.
highly ionic, solutions will be incorrect. solution does not solve the problem,
raising the temperature to the maximum ELECTRODE STORAGE
ELECTRODE CARE AND allowable for the reference system will As a general rule, store the pH electrode in
MAINTENANCE often help. Other types of blockage can the same solution as the reference electrolyte
Many types of pH electrode are available also occur (eg in the form of a precipitate of the electrode. In most cases this is a
but the standard glass or epoxy-bodied [black] of silver chloride or mercury 3 mol/L KCl solution. Most manufacturers
combination electrode is ideal for the majority sulphide in the porous pin). Gentle use of supply a plastic protection cap which is filled
of tests carried out on aqueous solutions abrasive paper can sometimes remove the with this solution. Close off the filler opening
with a reasonable ionic strength at ambient precipitate. In other cases, chemical if there is one. Never store your electrode in
temperatures and with limited use in strongly procedures such as soaking the electrode water as this will cause ions to leach out of
acidic or alkaline solutions. for a few hours in an acidic solution of the glass membrane, leading to a sluggish
The following general guidelines indicate thiourea (1 mol/L thiourea in 0.1 mol/L response.
the care and maintenance required for pH HCl) can be used.
electrodes: r Ensure that the electrode is used and RECONDITIONING ELECTRODES
r To dry the electrode, use clean soft tissues stored within its specified temperature Older electrodes, or electrodes that have been
and blot the liquid from the electrode. range. Extreme changes in temperature stored dry, may need to be ‘reconditioned’.
r Immerse in pH 4 buffer for short-term between samples will affect response time, Recondition an electrode by soaking in pH
storage. For longer-term storage use the and electrodes used above their 4.01 buffer or electrode storage solution for
same solution as the reference electrolyte temperature range will age rapidly. at least 30 minutes. r
of the electrode. In most cases this is a r Ensure that air bubbles are not trapped
3 mol/L KCl solution. Most manufacturers at the bottom of the electrode. If present, Other articles by Peter Riddle
supply a plastic protection cap which is bubbles should be removed by holding the (prriddle@hotmail.co.uk) in this series of
filled with this solution. Close off the filler electrode vertically and gently tapping the updates are on centrifugation (February,
opening if there is one. If the electrode will electrode body. If the air bubbles are page 76) and liquid-handling devices (March,
be not used for a long period of time, you trapped by KCl crystals, heating the page 138). Next month the focus turns to
may store it dry to prevent ageing (ageing electrode gently to 60˚C (maximum) in laboratory balances.
takes place only when the electrode is a water bath may also prove beneficial.

APRIL 2013 THE BIOMEDICAL SCIENTIST 205

You might also like