Nano Research

DOI 10.1007/s12274-016-1420-3

Toward integrated detection and graphene-based


removal of contaminants in a lab-on-a-chip platform
Andrzej Chałupniak1 and Arben Merkoçi1,2 ()

1
Catalan Institute of Nanoscience and Nanotechnology (ICN2), CSIC and The Barcelona Institute of Science and Technology, Campus
UAB, Bellaterra, 08193 Barcelona, Spain
2
ICREA, Pg. Lluís Companys 23, 08010 Barcelona, Spain

Received: 7 November 2016 ABSTRACT


Revised: 8 December 2016 A novel, miniaturized microfluidic platform was developed for the simultaneous
Accepted: 14 December 2016 detection and removal of polybrominated diphenyl ethers (PBDEs). The
platform consists of a polydimethylsiloxane (PDMS) microfluidic chip for an
© Tsinghua University Press immunoreaction step, a PDMS chip with an integrated screen-printed electrode
and Springer-Verlag Berlin (SPCE) for detection, and a PDMS-reduced graphene oxide (rGO) chip for
Heidelberg 2016
physical adsorption and subsequent removal of PBDE residues. The detection
was based on competitive immunoassay-linked binding between PBDE and
PBDE modified with horseradish peroxidase (HRP-PBDE) followed by the
KEYWORDS monitoring of enzymatic oxidation of o-aminophenol (o-AP) using square wave
electrochemistry, anodic stripping voltammetry (SW-ASV). PBDE was detected with good sensitivity
microfluidics, and a limit of detection similar to that obtained with a commercial colorimetric
graphene oxide, test (0.018 ppb), but with the advantage of using lower reagent volumes and a
flame retardants, reduced analysis time. The use of microfluidic chips also provides improved
linearity and a better reproducibility in comparison to those obtained with
lab on a chip,
batch-based measurements using screen-printed electrodes. In order to design
polydimethylsiloxane
a detection system suitable for toxic compounds such as PBDEs, a reduced
graphene oxide–PDMS composite was developed and optimized to obtain increased
adsorption (based on both the hydrophobicity and π–π stacking between rGO
and PBDE molecules) compared to those of non-modified PDMS. To the best
of our knowledge, this is the first demonstration of electrochemical detection
of flame retardants and a novel application of the rGO-PDMS composite in a
biosensing system. This system can be easily applied to detect any analyte using
the appropriate immunoassay and it supports operation in complex matrices
such as seawater.

1 Introduction developed since the last decade due to their versatility,


ease of fabrication, and a multitude of applications
Lab-on-a-chip (LOC) platforms are being intensively in biosensing and analysis. So far, LOC devices have

Address correspondence to arben.merkoci@icn2.cat


2 Nano Res.

been reported for the detection of protein [1], nucleic diphenyl ether (BDE) and octa-BDE) and 2013 (deca-
acids [2, 3], drugs [4, 5], hormones [6, 7], and many BDE) in USA, European Union, and elsewhere
different organic compounds. Thus, LOC devices are [26, 27].
very promising for health diagnostics, environmental Due to their hydrophobicity, PBDEs can accumulate
screening, and ensuring safety and security. in various tissues and organs, especially in adipose
Various methodologies of LOC fabrication have tissue and their exposure is correlated with liver
recently been developed involving materials such as dysfunction [28]. PBDEs can also be found in breast
paper [8], glass [9, 10], polymers [11, 12], or even pro- milk, and many investigations have demonstrated
teins (e.g., gelatin [13, 14]). The use of a certain material dependence between the persisting contact with
depends on the desired application, physicochemical materials containing PBDE (by dermal absorption,
properties of the analyte to be detected, the need for consumption of contaminated food or inhalation)
biocompatibility, and possibly integration with other and their further accumulation in the body. Besides
devices and materials [11]. the small size and hydrophobicity, the biphenyl
One of the most common approaches in LOC chemical structure of PBDEs is similar to that of the
development is polydimethylsiloxane-based (PDMS) thyroid hormones (3,5-diiodothyronine (T2), 3,3,5-
technology (soft-lithography), where a polymeric triiodothyronine (T3), and 3,3,5,5-tetraiodothyronine
elastomer is used to fabricate chips of different shapes, (T4)), and it has been confirmed that PBDEs bind with
patterns, and properties [15]. alpha- and beta-thyroid hormone receptors. PBDEs
Development of low-volume (in the range of micro- are also linked to neurodevelopment toxicity and
and nano-liters) and portable devices is especially various types of tumors [28, 29].
interesting for environmental screening of potentially Despite the need for monitoring PBDE levels, only
hazardous compounds. The monitoring of various a few methods have been reported so far for their
pollutants present in air, soil, or water requires regular detection. Standard determination of PBDE is based
monitoring, and the use of an LOC device can allow on gas chromatography (GC) separation combined
efficient in-field detection; access to sophisticated with electrode capture detection (GC-ECD) or mass
laboratories is limited, expensive, and time consuming. spectrometry (GC-MS). The use of these methods is
LOC devices have been reported for electrochemical still limited by high equipment cost, the need for a
detection of phenolic compounds [16, 17] and atrazine qualified workforce, and high time consumption [30–32].
[18], optical platforms for pesticides [19], and heavy One of the alternatives is a competitive immunoassay
metal detection [20]. An LOC can be applied not only for PBDE detection that is based on competition
for the analysis of pollutants derived from industry between PBDE molecules and PBDE modified with
and urban areas, but also for toxins originating from horseradish peroxidase (HRP-PBDE) in binding to
the natural environment such as those produced from anti-PBDE antibodies that are coupled to magnetic
marine microalgae [21–23]. beads (MB-Ab). The signal is revealed using a well-
The detection of polybrominated diphenyl ethers known enzymatic reaction that catalyzes 3,3',5,5'-
(PBDEs), a class of halogenated compounds, similar tetramethylbenzidine (TMB) oxidation, resulting in
to polychlorinated diphenyls (PCBs) and commonly a colored product that can be easily measured by
used as flame retardants, needs to be improved. spectrophotometry. This method is relatively simple
PBDEs were introduced into the market in the 1960s and offers an attractive range of detection (0.025–
and used for many years in domestic electronics, toys, 1.0 ppb of PBDE), both in standard solutions and
furniture, polyurethane foams, and building materials environmental samples like soil or seawater.
[24, 25]. Over time, however, the negative effects of Taking advantage of well-developed antibodies and
long-term exposure to PBDEs have been demonstrated, a competitive reaction, we modified the colorimetric
although many examples are not well investigated. kit by performing electrochemical detection of HRP-
As a result, commercial mixtures of PBDEs were PBDE and replacing TMB with o-aminophenol (o-AP)
banned from all applications in 2004 (penta-brominated as a substrate for the electrochemical reaction. We

| www.editorialmanager.com/nare/default.asp
Nano Res. 3

believe that the electrochemical reaction will provide a composite of PDMS and rGO (rGO-PDMS) was
an opportunity to assay PBDEs with better sensitivity developed and used to fabricate a microfluidic chip,
and may lead to improved automation by integration where post-measurement samples are injected and
into an LOC device. The LOC system and the detection organic traces, including excess PBDE, are immobilized
approach are shown in Fig. 1. and adsorbed onto the channel walls to improve
The role of nanomaterials in LOC and other devices the safety of the whole system. Such non-covalent
has grown significantly in recent years for both sensing interactions between PBDEs and carbon materials
and environmental removal of contaminants. In this (based on π–π stacking) were previously used for novel
context, we take advantage of the adsorption and extraction methods [33, 34] and micromotor-based
removal capabilities of reduced graphene oxide (rGO) removal of PBDEs [35]. Various composites of PDMS
toward certain contaminants. As a proof of concept, with carbon materials have also been developed, but

Figure 1 LOC system for detection and removal of PBDEs. (a) Components of the LOC device with visualization of the microfluidic
structures (10× magnification objective); (b) side and top view of the assembled LOC device on a polycarbonate substrate with a
screen-printed carbon electrode (SPCE); (c) assay procedure: (1) on-chip mixing and incubation of magnetic beads coupled to anti-PBDE
antibodies (MB-Ab) with samples containing PBDE, (2) on-chip mixing and incubation of MB-Ab with HRP-PBDE, (3) immobilization
of the immunocomplex in chip using an external magnet and washing of unbound molecules, (4) incubation with substrate solution
(o-AP + H2O2), and (5) electrochemical detection of the product of enzymatic reaction (3-aminophenoxazone (3-APZ)).

www.theNanoResearch.com∣www.Springer.com/journal/12274 | Nano Research


4 Nano Res.

most of them are based on sophisticated protocols from Eco Chemie B.V., Utrecht, the Netherlands).
that have been optimized to obtain conductive rather Microfluidic chips with an integrated SPCE electrode
than highly adsorptive materials [36, 37]. were connected to the Autolab PGSTAT-12 with a
To the best of our knowledge, this is the first specially adapted electrical edge connector. A syringe
approach showing an LOC device with an electro- pump (Pump 11 Elite; Harvard Apparatus, USA) was
chemical approach for PBDE detection. Herein, a novel used as a driving unit for microfluidic system. The
rGO-PDMS composite is used for the first time in an syringe pump was equipped with 0.5 mL glass syringes
LOC device for biosensing and removal purposes. (Gastight®, Hamilton, USA). Centrifuge SIGMA 2-16PK
was used for rGO purification and a magnetic stirrer
(IKA lab disc white) was used during the reduction
2 Experimental
process. Plasma cleaner PDC-002 was used as a source
2.1 Reagents of oxygen plasma for chip fabrication. Microfluidic chip
operation was optimized using an optical microscope
PBDE colorimetric immunoassay kit was purchased (Olympus IX71, Germany) coupled with a CCD camera
from Abraxis (Pennsylvania, USA). PBDE stock solution (Olympus DP72, Germany) with the bright field mode,
was purchased from AccuStandard Inc. The screen- 10× magnification.
printed inks (carbon sensor paste C2030519P4 and
silver–silver chloride paste C2130809D5) were purchased 2.3 Experimental methodology
from Gwent Group, UK. Polycarbonate sheets were
purchased from Vink Plastics S.L.U., Spain. PDMS Unless otherwise stated, all optimization experiments
(Sylgard® 184 Elastomer kit) was purchased from regarding microfluidic chips were carried out on five
Ellsworth Adhesives Iberica, Spain. Graphene oxide independent chips. Optimization of electrochemical
stock solution (5 mg/mL) was purchased from Angstron measurements was carried out using five independent
Materials (Dayton, USA). A negative photoresist screen-printed electrodes. Final calibration curves
(SU8-50) and an appropriate developer were purchased were obtained based on three repetitions. When the
from Microchem, USA. Flexible photolithography colorimetric method was used, each sample was run
masks were provided by Microlitho, UK. Four-inch in duplicate according to the manufacturer´s procedure.
silicon wafers were purchased from NOVA Electronic All errors of the results obtained were expressed as a
Materials. Hydrogen peroxide, o-aminophenol, HPLC relative standard deviation (RSD).
grade methanol, phosphate buffered saline (PBS),
2.4 Fabrication and operation of LOC devices
3-aminopropyltriethoxylane (APTES), L-ascorbic acid
(L-AA), and microscopic glass slides were purchased Fabrication of screen-printed electrodes was based on
from Sigma. Britton–Robinson (BR) buffer (0.04 M, already reported design and procedures, using poly-
pH 5.7) was prepared as described elsewhere [38]. All carbonate sheets as the substrate [39]. The microfluidic
commercial reagents were of analytical grade and mold was prepared using a previously reported
handled according to the material safety data sheets methodology of photolithography process [15, 39],
suggested by the suppliers. where a 4-inch silicon wafer was spin coated with
a negative photoresist SU8-50 and patterned by
2.2 Instruments
photolithography using a flexible mask.
Screen-printed carbon electrodes (SPCE) were fabricated PDMS was mixed with its curing agent in the mass
by screen-printing technology using a screen-printer ratio 10:1. After intensive stirring for 5 min, an excess
(DEK 248, UK). Electrochemical measurements were of air bubbles was removed using vacuum. Afterward,
carried out using computer-controlled Autolab PDMS was poured onto the silicon wafer with a given
PGSTAT-12 (302 N-High performance, potentiostat/ design of the microfluidic chip and kept on the hotplate
galvanostat) with a general-purpose electrochemical at 70 °C for 4–5 h until completely polymerized. After
software operating system (GPES version 4.9.007, polymerization, proper shapes of PDMS were cut out

| www.editorialmanager.com/nare/default.asp
Nano Res. 5

from the template, washed in miliQ water, followed be automated by simply passing the sample through
by isopropanol and dried on the hotplate at 70 °C for the channel, where mixing supports both (1) PBDE
10 min. In order to modify the surface, oxygen plasma binding to MB-Ab and (2) HRP-PBDE binding. The
treatment was performed. Meanwhile, polycarbonate design of IMMU-CHIP was previously used and its
sheets (either with or without screen-printed electrodes) effectiveness (not reported here) in biochemical
were washed with miliQ water, followed by isopropanol, reactions was confirmed.
ethanol, plasma treatment, and incubated in 2% Using an optical microscope, the process of mixing
APTES solution in order to silanize the surface (1 h). was observed, so as to match the best parameters and
Afterward, both parts (PDMS and polycarbonate) were make sure that all chemicals used in the assay are
simply attached to each other, forming a permanent compatible with the materials used for the LOC device.
bonding that was strengthened by incubation on the As all the reagents seemed to be transparent for
hotplate at 70 °C for 1 h. Prior to the assay, the chips microscopy, Trypan blue (10%) was used to facilitate
were connected with each other using polymeric tubes. the distinction between two different solutions during
A general scheme of the system as well as dimensions mixing. The washing step was also assessed in order
of microfluidic channels are shown in Fig. 1. The system to identify an optimum flow rate and buffer volume
consists of three microfluidic chips: immunoassay (Fig. S1 in the Electronic Supplementary Material
chip (IMMU-CHIP), detection chip (DET-CHIP), and (ESM)).
removal chip (GO-CHIP).
IMMU-CHIP consisted of channels with repeatable 2.6 Optimization of the on-chip magneto-
curvatures and three inlets (MB-Ab, PBDE, and HRP- immunoassay
PBDE). The role of curvatures was to support the IMMU-CHIP was fabricated as described in section
diffusion between mixed reagents. In this chip, an 2.4, and it was washed by priming with 1 mL of washing
immunoreaction between PBDE/HRP-PBDE and MB- buffer (water + 0.05% Tween-20) prior to use. A
Ab occurred. The PDMS layer was attached either to dedicated inlet was connected with a syringe containing
glass or polycarbonate. reagents of magneto-immunoassay for PBDE detection.
DET-CHIP consisted of a single channel and the Two types of optimization experiments were per-
chamber (10 mm length, 3 mm width, and approximate formed. The first experiment checked the occurrence
volume of 15 μL) where the immunocomplex (MB- of non-specific adsorption of PBDE to PDMS and
Ab-PBDE/HRP-PBDE) was immobilized by an external changes in non-specific adsorption in relation to the
magnet placed underneath the chip. This chip was flow rate. A solution of 1.0 ppb PBDE was flushed
bonded to polycarbonate with SPCE. The DET-CHIP through the whole chip and collected in a microcen-
could be employed as a part of the LOC system as well trifuge (Eppendorf tube) at the outlet. A colorimetric
as a separate chip just for various electrochemical test was then performed following the standard pro-
measurements. cedure (S1 in the ESM). Results were expressed as a%
GO-CHIP was fabricated using the rGO-PDMS of signal relative to the control sample of 1.0 ppb of
composite. The design used to fabricate this chip was PBDE.
based on a broad channel with pillars along its surface The second experiment checked the possible
so as to increase surface-area-to-volume ratio and adsorption of magnetic beads and HRP-PBDE onto
improve the adsorption performance through the the microchannel. The first inlet was connected to a
microfluidic channel. The fabrication process was based PBDE blank solution (0.0 ppb), the second to MB-Ab,
on the patterning of the SU-8 photoresist. and the third to HRP-PBDE. The solutions were injected
2.5 Optical evaluation of the on-chip immunoreaction and mixed in a chip with each other at different flow
performance rates, collected in an Eppendorf tube, and proceeded
as in standard procedure for colorimetric detection of
The motivation for using a microfluidic chip for mixing PBDE. Results were expressed as a% of signal obtained
(IMMU-CHIP) is the possibility that the process can by using a control sample containing MB-Ab, HRP-

www.theNanoResearch.com∣www.Springer.com/journal/12274 | Nano Research


6 Nano Res.

PBDE, and 0.0 ppb of PBDE. is the absorbance of control sample with the initial
concentration of 1.0 ppb; and B0(0,0ppb) is the absorbance
2.7 Preparation of rGO-PDMS composite of the blank control sample containing 0.0 ppb solution.
An already reported approach for GO reduction 2.8 Electrochemical detection
using L-AA was employed [40]. L-AA was added
to 200 mL GO solution to reach the concentration of Electrochemical detection of HRP is inspired by
50 mg/mL. Samples were adjusted to desired pH using already reported works [38, 41]. The competitive
NaOH solution and incubated for 48 h with continuous immunoassay results show that the level of HRP
magnetic stirring (600 rpm). After reduction, solution is inversely proportional to the level of PBDE. HRP
was transferred to 15 mL centrifugal tubes and catalyzes the redox reaction where hydrogen peroxide
centrifuged for 1 h, 12,000 rpm, and 4 °C. The super- is converted to water molecules, while a given substrate
natant was discarded, while pellets were washed is oxidized. In this approach (Fig. 1(c)), o-AP was
with miliQ water. This step was repeated twice using chosen as the substrate for the enzymatic reaction
miliQ and twice using 96% (v/v) ethanol. Purified due to its compatibility with other assay reagents and
solution of rGO was immediately used for further suitable electrochemical properties. Oxidation of o-AP
experiments. leads to the formation of 3-aminophenoxazone (3-APZ)
Two components of the PDMS kit (elastomer and that can be reduced on the SPCE surface and detected
curing agent) were mixed in a volume ratio of 10:2. in several electrochemical techniques such as cyclic
After stirring, a desired amount of rGO was added voltammetry (CV), differential pulse voltammetry
and vigorous stirring was continued for about 10 min. (DPV), chronoamperometry (AMP), and square wave
Afterward, rGO-PDMS was poured onto a silicon anodic stripping voltammetry (SW-ASV). As the PBDE
wafer with the design of a given chip. The wafer with detection kit based on the HRP-PBDE molecule is
rGO-PDMS was kept for 1 h at room temperature (RT) optimized for a relatively narrow PBDE concentration
in order to remove air bubbles entrapped within the range, SW-ASV was chosen because this technique is
polymer before polymerization. Then, the wafer was reportedly much more sensitive than others. Never-
kept on a hotplate and incubated at 70 °C for 4–5 h theless, some examples of PBDE detection using
when the rGO-PDMS was completely polymerized. CV and DPV are shown in Fig. S2 in the ESM. The
Microfluidic chips were fabricated using the same following parameters were applied: deposition potential
procedure. (–0.6 V); deposition time 20 s; equilibration time 15 s;
Analysis of the rGO-PDMS composite (the effect potential range from 0.0 to –0.6 V; step potential 4 mV;
of the pH, flow rate, and rGO concentration) was amplitude 25 mV; and frequency 25 Hz.
performed to determine the ability to adsorb a Detection using the microfluidic chip was preceded
standard solution of PBDE (1.0 ppb). All parameters by measurements based on bare SPCE so as to
were optimized to maximize the adsorption yield. determine the optimal detection protocol. Briefly, the
Experiments were based on passing the PBDE sample immunoassay was performed as usual; after washing
through the rGO-PDMS chip (with a single channel an immunocomplex, samples were reconstituted in
of 5 cm length, 500 μm width, and 100 μm depth) and 10 μL of PBS buffer and 100 μL of substrate solution
then determining the concentration of PBDE using (o-AP + H2O2 in BR buffer) and incubated for 30 min.
After incubation, 50 μL of the sample was pipetted onto
the colorimetric kit. As this kit is a competitive assay,
the SPCE surface and electrochemical measurements
the relative adsorption was determined using the
were performed. Following this procedure, incubation
following formula (Eq. (1))
time, o-AP concentration, H2O2 concentration, and
ADS = A1(1.0ppb) – A0(1.0ppb)/B0(0,0ppb) (1) deposition time in SW-ASV were optimized.
Then, electrochemical detection was performed using
where A1(1.0ppb) is the absorbance of sample used in DET-CHIP. Parameters for detection were chosen
chip, with the initial concentration of 1.0 ppb; A0(1.0ppb) based on optimization in static measurements. In this

| www.editorialmanager.com/nare/default.asp
Nano Res. 7

case, when the substrate solution is injected into the already reported as effective adsorbents of a range of
microfluidic channel, it passes through the chamber molecules. In this work, we tried to combine the
where magnetic beads with the immunocomplex are ability of rGO to adsorb PBDE with the convenient
immobilized; therefore, the enzymatic reaction was properties of PDMS chips by using an LOC device
performed in flow and the product passed further with increased adsorption performance (Fig. 2). Similar
along the channel up to the electrode surface where approaches are already reported, where graphene
it was electrochemically deposited and detected. In materials and PDMS were used as the composite, but
addition, the optimum flow rate was analyzed in they are used for its electrical properties and none was
order to determine the best sensitivity and limit of applied for the adsorption and removal applications
detection. in LOC devices [36, 42]. As presented, the rGO-PDMS
composite does not need to be conductive; therefore,
a much simpler fabrication protocol can be used that
3 Results and discussion
utilizes the adsorption property of rGO, which is
attractive for lab-on-a-chip applications.
3.1 rGO-PDMS composite
Furthermore, as the adsorption mechanism is based
Graphene oxide and other carbon derivatives are on π–π stacking, the rGO-PDMS composite can be

Figure 2 rGO-PDMS composite (a) fabrication process; (b) example of rGO-PDMS-glass chip filled with color inks; (c) example of
flexible solid piece of rGO-PDMS; (d) the effect of rGO concentration (w/w) on PBDE adsorption; (e) the effect of pH during reduction
process on PBDE adsorption; (f) the effect of flow rate on PBDE adsorption using rGO-PDMS and bare PDMS.

www.theNanoResearch.com∣www.Springer.com/journal/12274 | Nano Research


8 Nano Res.

employed for various biomolecules such as DNA or flow rate will decrease the adsorption yield. However,
proteins as both structures contain abundant aromatic comparison of PDMS to rGO-PDMS shows that in
rings. As a proof of concept, we performed the a higher flow rate (90 μL/min), the effect of rGO is
analogous test as for PBDEs but measured the distinct (with a significant increase of adsorption)
adsorption of a model protein (HRP) and the effect compared to that achieved at lower flow rates
of rGO adsorption can be clearly observed (Fig. S3 in (30 μL/min).
the ESM). Final measurements in the LOC system were com-
We identified pH as an important factor facilitating pleted with the rGO-PDMS chip fabricated using
rGO reduction and thus the efficiency of PDMS the design shown in Fig. 1 (GO-CHIP) and under
adsorption. Experiments based on PBDE adsorption the optimal conditions mentioned above. Due to a
onto the rGO-PDMS channel (Fig. 2(e)) showed that significantly increased contact surface area between
an increase of pH leads to an increase in adsorption; the liquid and the rGO-PDMS channel, its adsorption
this effect peaks at pH 7. Above pH 7, there is no capacity is higher than that of a single-channel chip.
significant improvement in absorption. We tested PBDE in the concentration range 1–
It is not surprising that an increase in the rGO 1,000 ppb and identified that the total adsorption
concentration in the PDMS will make adsorption capacity is between 100 and 1,000 ppb, both in the
stronger as more rGO will be exposed to the channel standard buffer solution and seawater (Fig. S4 in the
surface. However, any compound added to PDMS ESM). We believe that this is sufficient for real sample
may affect its ability to polymerize. We found (Fig. 2(d)) treatment as the usual level of PBDE in seawater
that when the rGO concentration is lower than 1% is significantly lower than 10 ppb [43–45]. Moreover,
(w/w), there is no improvement in adsorption and a decrease in the flow rate may further improve
probably a slight decrease compared to that obtained the adsorption capability. In this work, as all of the
with bare PDMS. A concentration of 4.5% (w/w) is operations are performed simultaneously, the flow rate
optimal as it results in the highest adsorption of PBDE value is determined by the other steps performed in
(around 70%). When the concentration is higher than the IMMU-CHIP and DET-CHIP.
4.5% but lower than 10%, the adsorption capacity Apart from the adsorption capability, the possible
significantly decreases and the composite keeps its reuse of the platform was tested. We found that solvents
physical properties such as durability and hardness. introduced to the GO-CHIP channel can extract the
If the concentration of rGO is higher than 10% (v/v), previously adsorbed PBDE. Acetone was the most
polymerization proceeds and the gas bubbles that are suitable solvent, as this solvent enabled desorption
always slowly released during the polymerization of of PBDE in four subsequent cycles and recovery of
PDMS prevent the formation of a flat, homogenous nearly 100% PBDE (Fig. S5 in the ESM). The ease of
layer of rGO-PDMS. Such a high concentration of fabrication indicates that this platform is more feasible
rGO-PDMS also changes some physical properties as when considered as disposable, rather than reusable
the polymer is too flexible and spongy. This effect platform.
may be related to the presence of rGO, but our opinion
3.2 Electrochemical detection of PBDE (bare SPE)
is that it is due to the presence of excessive solvent
(water from the rGO suspension) in PDMS, which Development of a lab-on-a-chip platform for electro-
evaporates during polymerization and affects the chemical detection was preceded by experiments
structure of the polymer. Taking into account these involving just screen-printed electrodes. Figure 3 shows
observations, we decided to perform the last steps of parameters that were optimized in order to find the best
rGO purification in ethanol solution, as it is compatible conditions for PBDE determination. The most crucial
with PDMS and can evaporate very quickly without parameter is the concentration of o-AP (Fig. 3(a)). If it is
affecting the polymerization. higher than 0.12 mM, it contributes to non-enzymatic
Flow rate is a major factor that influences microfluidic oxidation and reduces the signal-to-noise ratio, similarly
processes. We proved (Fig. 2(f)) that an increase of as reported elsewhere [38]. For this reason, 0.6 mM

| www.editorialmanager.com/nare/default.asp
Nano Res. 9

Figure 3 Optimization of electrochemical detection in static conditions using bare SPCE: (a) optimization of substrate (o-AP)
concentration, (b) optimization of hydrogen peroxide concentration, (c) optimization of incubation time (enzymatic reaction), and (d)
optimization of deposition time in square wave anodic stripping voltammetry. All results expressed as signal-to-noise ratio calculated as
a ratio between the signal represented by control sample (0.0 ppb of PBDE) and substrate solution.

o-AP was chosen for further experiments, providing 3.3 Lab-on-a-chip optimization
satisfactory sensitivity. The effect of the other com-
ponent, the substrate solution (hydrogen peroxide), The next step in the LOC platform development was
upon reaction sensitivity (Fig. 3(b)) was also studied; evaluation of the IMMU-CHIP performance. Figure 4(a)
thus, the concentration of 2 × 10–4 M was used, as shows the process of mixing between the solution
reported elsewhere [38]. of PBDE and MB-Ab suspension. It can be seen that
Incubation time is the time when redox reaction is when solutions meet each other in the microfluidic
catalyzed by HRP. The time providing the highest channel, they do not mix and flow separately. However,
SNR (Fig. 3(c)), 30 min, is quite similar to the time of due to the curvature of the channel, the solution
reaction in a commercial colorimetric method (20 min). is further conveyed from laminar into turbulent
Further time extension does not lead to any sensitivity flow; as a result, the diffusion between them occurs
improvement. quite efficiently and facilitates the immunoreaction.
An increase in the deposition time (Fig. 3(d)) leads Figure 4(a)(3) shows the ending section of the mixer
to a proportional increase of SNR. However, if the chip, where the color of the solution is already uniform;
deposition time exceeded 60 s, an enormous decrease hence, a reaction between PBDE and MB-Ab is likely.
of SNR is observed due to electrochemical oxidation Figure 4(b) presents the results of flow rate
of the substrate (o-AP), which prevails over the effect optimization, where a colorimetric kit was used as a
of the enzymatic reaction (HRP catalysis). Thus, 20 s reference technique. Regarding PBDE adsorption, an
was chosen as the optimum deposition time, which increase of flow rate is linked with an increase of PBDE
provides good sensitivity and reduces the risk of recovery, which means that no analyte is lost during
false-positive results. the assay. For the flow rate equal or higher than

www.theNanoResearch.com∣www.Springer.com/journal/12274 | Nano Research


10 Nano Res.

Figure 4 Optimization of LOC operation. (a) Visualization of mixing in IMMU-CHIP between reagents in the (1) intersection where
MB-Ab meet PBDE, (2) middle part of the chip where mixing process partially occurred, and (3) terminal section of the channel where
samples are fully mixed. Trypan-blue (10%) was used to visualize PBDE solution; pictures were taken using an optical microscope with
20× magnification; (b) optimization of the flow rate in the IMMU-CHIP was based on a colorimetric test regarding the PBDE recovery
and immunocomplex (HRP-PBDE-Ab-MB) signal when passing the microfluidic channel; and (c) optimization of the flow rate during
enzymatic reaction in electrochemical detection of HRP-PBDE (DET-CHIP). All results are expressed as the signal-to-noise ratio
calculated as a ratio between the signal represented by substrate solution without PBDE as a control sample.

100 μL/min, PBDE adsorption is no longer observed. ments), in chip detection (DET-CHIP) requires a high
Regarding the on-chip immunoreaction, the efficiency level of attention to the flow rate, as the main factor
of the reaction is not very high when the low flow affecting sensitivity and limit of detection. The flow
rate is applied (60% of control signal for the flow rate of the sample particularly supports the enzymatic
of 5 μL/min), which is probably related to protein reaction, where substrate solution is mixed with the
adsorption or the physical loss of magnetic beads. On immunocomplex and 3-APZ is obtained, as well as
the other hand, when very high flow rates are applied, the deposition process, where 3-APZ is accumulated
the efficiency of the reaction is also not satisfactory, onto the electrode surface. In electrochemical stripping
probably due to the reduced time that limits the (the detection step), the flow is stopped and this step
occurrence of the immunoreaction. Thus, analysis is performed in static conditions.
of both curves shows that the flow rate of about In order to find the optimal conditions, some
100 μL/min offers a compromise, as PBDE adsorption measurements of different flow rates (defined by taking
is not observed and the reaction between HRP-PBDE into account the constant volume of the injected
and MB-Ab works with the maximum efficiency. o-AP/H2O2/PBS solution, 30 μL) as well as the desired
Compared to electrochemical measurements per- time of the enzymatic reaction were carried out.
formed just by drop casting (batch/static measure- Figure 4(c) shows that an increase in the flow rate

| www.editorialmanager.com/nare/default.asp
Nano Res. 11

correlates with a decrease in the signal-to-noise ratio. SW-ASV is represented by curves in which negative
This is obviously caused by the reduced time that is a current is inversely proportional to the PBDE concen-
key factor in enzymatic reactions. Flow rates between tration. Each peak is characterized by a specific
0.5 and 3 μL/min allowed to perform reaction with potential of about –0.35 V that shifts slightly toward
a relatively high signal-to-noise ratio, similar to the –0.3 V when the concentration of HRP-PBDE increases.
optimal condition (30 min) of static measurements. We have also observed that the presence of salts (e.g.,
The final experiments were performed using the flow PBS) influences the potential value by shifting it
rate of 2 μL/min, where the reaction time is only 15 min. toward –0.4 V, without affecting the signal intensity
Similar to static measurements, the time of 20 s was (data not shown). All measurements were performed
found to be optimal for the deposition step (data not in the presence of PBS (o-AP + H2O2 and PBS in a ratio
shown). The process of deposition was performed 10:1) so as to keep a reproducible value of potential.
with a flow rate of 90 μL/min in order to transfer all In order to validate the developed technique, we
of the solution from the chip, where an enzymatic simultaneously performed the commercial colorimetric
reaction occurred. We have observed that such a test (Fig. 5(b)). The limit of detection (LOD) of this
relatively high flow rate supports the sensitivity of technique is about 0.019 ppb PBDE and is in accordance
the assay as the background current is much lower with the results obtained by electrochemical mea-
than that in static measurements. This results from surement, where the LOD is about 0.018 ppb, both
the amount of molecules adsorbed onto the electrode for the whole assay performed in a chip and only for
during the deposition step that is lower at the higher the detection step performed in a chip. This is not
flow rate. surprising as the LOD of this method is limited
by the immunoassay performance, rather than HRP
3.4 Complete detection of PBDE using the lab-on-
detection as the concentration of the labeling protein
a-chip platform
is sufficient, in both samples with the highest and
Figure 5 presents the final results of all methods com- lowest concentrations of PBDE. Such a limit of detection
pared. As it can be seen in Fig. 5(a), the signal of seems sufficient for all possible applications, taking

Figure 5 Comparison of optical and electrochemical detection of PBDE. (a) Example of SW-ASV measurements of different
concentrations of PBDE (0.0–1.0 ppb) using the LOC system; (b) calibration curves showing the detection of PBDE using a colorimetric
test as well as different modes of electrochemical detection; and (c) electrochemical detection in seawater samples spiked with PBDE
standard solutions.

www.theNanoResearch.com∣www.Springer.com/journal/12274 | Nano Research


12 Nano Res.

into account the amount of PBDEs found globally; for are removed so there is minimal influence on
instance, seawater in Jiaozhou Bay in China (6.59 ppb) electrochemical detection. As already mentioned, the
[43], coastal water of the Aegean Sea (0.03–0.05 ppb) presence of salts (ionic strength) does not affect the
[44], or various indoor environments in Barcelona, signal intensity. As the system is based on magneto-
Spain (7.3–13.9 ppb) [45]. immunoassay, the use of magnetic beads in a chip
Compared to the optical detection method, the also supports an efficient method of capturing analytes
electrochemical approach has higher method sensitivity from a complex matrix such as seawater.
and is clearly seen by the slope of the calibration
3.6 Comparison with other methods and future
curves (Fig. 5(b)). Furthermore, a major difference
perspectives
regarding linearity and reproducibility can be observed
by comparing measurements performed on a bare Comparison between the colorimetric and electro-
electrode with those in a microfluidic chip (either just chemical methods regarding the use of reagents is
detection or the whole assay). Detection based just shown in Table S1 in the ESM. Owing to the use of
on bare electrodes was linked with a much worse microfluidics, less reagents are consumed and a
coefficient of determination (R2 = 0.974 compared to significant amount of time is saved. By using a syringe
R2 = 0.994) and higher errors amongst all repetitions pump as a driving unit, automation of the system
of the experiment in the microfluidic chip. This is is obtained. Moreover, instead of a syringe pump,
probably related to the nature of square wave anodic programmable microfluidic flow controllers might be
stripping voltammetry, as good performance of this applied for this platform, providing full automation
technique depends on the diffusion and contact between of the process and eliminating the need of a highly
the electrode surface and the sample. Apparently, this qualified workforce. This is crucial, as PBDE is a small
works better in microfluidics where the deposition molecule, which can be detected mainly in the com-
process is performed under flow and the presence of petitive immunoassay format. This kind of assay is
the microfluidic channel keeps a constant volume of generally more sophisticated than that for label-free
liquid in contact with the working electrode and methods; thus, more time and attention are usually
prevents the evaporation of the sample by limiting required to perform the analysis. Although simple
any influences from the external environment. This and sensitive label-free methods for PBDE detection
confirms the already reported statements on improved based on pattern recognition are reported, the risk of
performance of the SW-ASV in microfluidics compared cross-reactivity and the need of data separation limit
to that of the bare screen-printed electrodes. It is the practical application of these methods [46]. Another
especially meaningful taking into account the use of example of label-free detection of PBDE is electro-
home-made electrodes fabricated using more eco- chemical impedance spectroscopy, which although
friendly materials (carbon, silver/silver-chloride) than specific, provides much higher LOD (1.3 ppb) compared
electrodes containing toxic mercury [39]. to that reported in our work [47].
Another possible future improvement of the
3.5 Validation of the lab-on-a-chip device in
presented platform is its high-throughput applications.
seawater samples
By miniaturizing all of the chips and multiplying
In order to justify the practical application of the them, it would be possible to run several assays in
presented LOCs device, we evaluated seawater samples parallel without an extension of the total size of the
spiked with the PBDE standard solution, as described sensing system.
in S2 in the ESM. Results are shown in Fig. 5(c), and
as there is clear evidence of good recovery yield
4 Conclusions
(94.1%–103.0%), it can be assumed that this method is
fully compatible with seawater samples. This is possible This work describes the synthesis of a novel, integrated
due to washing steps where all the trace chemicals microfluidic platform for simultaneous capture,

| www.editorialmanager.com/nare/default.asp
Nano Res. 13

detection, and removal of polybrominated diphenyl PDMS composite, comparison of electrochemical and
ethers. The presented system consists of three micro- optical methods) is available in the online version of
fluidic chips: the IMMU-CHIP, where the analyte is this article at http://dx.doi.org/10.1007/s12274-016-1420-3.
captured and immunoreaction occurred; the DET-CHIP,
where the level of HRP-PBDE is quantified using
References
SW-ASV; and the GO-CHIP, which is responsible for
the removal of potentially toxic PBDEs. [1] Medina-Sánchez, M.; Miserere, S.; Morales-Narváez, E.;
The LOD achieved was 0.018 ppb, which is in Merkoçi, A. On-chip magneto-immunoassay for Alzheimer’s
accordance with that of the currently available methods biomarker electrochemical detection by using quantum dots
for PBDE detection. Electrochemical detection per- as labels. Biosens. Bioelectron. 2014, 54, 279–284.
formed in the microfluidic chip showed better sensitivity [2] Marasso, S. L.; Giuri, E.; Canavese, G.; Castagna, R.;
compared to that of the commercial colorimetric Quaglio, M.; Ferrante, I.; Perrone, D.; Cocuzza, M. A
method and better reproducibility and linearity than multilevel lab on chip platform for DNA analysis. Biomed.
those of electrochemical measurements performed Microdevices 2011, 13, 19–27.
simply using a bare screen-printed electrode. [3] Lee, H. H.; Yager, P. Microfluidic lab-on-a-chip for microbial
identification on a DNA microarray. Biotechnol. Bioprocess
The practical impact of this work is twofold: Firstly,
Eng. 2007, 12, 634–639.
the proof that electrochemical detection of PBDEs can
[4] Kurbanoglu, S.; Mayorga-Martinez, C. C.; Medina-Sánchez,
be performed, and secondly, that the rGO-PDMS
M.; Rivas, L.; Ozkan, S. A.; Merkoçi, A. Antithyroid
composite in microfluidic devices can be employed
drug detection using an enzyme cascade blocking in a
as an efficient adsorbent of various compounds. To the
nanoparticle-based lab-on-a-chip system. Biosens. Bioelectron.
best of our knowledge, both of the issues mentioned 2015, 67, 670–676.
above are not yet reported. [5] Kimura, H.; Ikeda, T.; Nakayama, H.; Sakai, Y.; Fujii, T.
Furthermore, the small size, durability, and suitability An on-chip small intestine-liver model for pharmacokinetic
for operation with seawater samples suggest that studies. J. Lab. Autom. 2015, 20, 265–273.
the presented platform can be employed in marine [6] Ozhikandathil, J.; Badilescu, S.; Packirisamy, M. Detection
environmental analysis; for example, as an integrated of bovine growth hormone using conventional and lab-on-
component of buoys. Combining this with commercially a-chip technologies: A review. Int. J. Adv. Eng. Sci. Appl.
available miniaturized potentiostats (electrochemical Math. 2015, 7, 177–190.
detection) and microfluidic driving units will result [7] Ozhikandathil, J.; Packirisamy, M. Nano-islands integrated
in a small and fully automated detection platform. evanescence-based lab-on-a-chip on silica-on-silicon and
polydimethylsiloxane hybrid platform for detection of
recombinant growth hormone. Biomicrofluidics 2012, 6,
Acknowledgements 46501.
[8] Medina-Sánchez, M.; Cadevall, M.; Ros, J.; Merkoçi, A.
We acknowledge FP7 EU Project “SMS” (No. 613844).
Eco-friendly electrochemical lab-on-paper for heavy metal
ICN2 acknowledges support from the Severo Ochoa
detection. Anal. Bioanal. Chem. 2015, 407, 8445–8449.
Program (MINECO, No. SEV-2013-0295) and Secretaria
[9] da Costa, E. T.; Santos, M. F. S.; Jiao, H.; do Lago, C. L.;
d’Universitats i Recerca del Departament d′Economia
Gutz, I. G. R.; Garcia, C. D. Fast production of microfluidic
i Coneixement de la Generalitat de Catalunya (2014 devices by CO2 laser engraving of wax-coated glass slides.
SGR 260). The authors would also like to thank Dr. Electrophoresis 2016, 37, 1691–1695.
Mariana Medina Sánchez for microfluidic mold fabri- [10] Ambrosi, A.; Guix, M.; Merkoçi, A. Magnetic and
cation that was employed in GO-CHIP development. electrokinetic manipulations on a microchip device for
bead-based immunosensing applications. Electrophoresis
Electronic Supplementary Material: Supplementary 2011, 32, 861–869.
material (protocol of colorimetric detection of PBDE, [11] Medina-Sánchez, M.; Miserere, S.; Merkoçi, A. Nanomaterials
data treatment, measurements in seawater, alternative and lab-on-a-chip technologies. Lab Chip 2012, 12, 1932–
electrochemical methods, protein adsorption to rGO- 1943.

www.theNanoResearch.com∣www.Springer.com/journal/12274 | Nano Research


14 Nano Res.

[12] Li, S. G.; Xu, Z. G.; Mazzeo, A.; Burns, D. J.; Fu, G.; [23] Zheng, G. X.; Wang, Y. H.; Wang, Z. M.; Zhong, W. L.;
Dirckx, M.; Shilpiekandula, V.; Chen, X.; Nayak, N. C.; Wang, H.; Li, Y. J. An integrated microfluidic device in
Wong, E. et al. Review of production of microfluidic devices: marine microalgae culture for toxicity screening application.
Material, manufacturing and metrology. In Proceedings of Mar. Pollut. Bull. 2013, 72, 231–243.
the SPIE 6993, MEMS, MOEMS and Micromachining III, [24] Hooper, K.; McDonald, T. A. The PBDEs: An emerging
Strasbourg, France, 2008. environmental challenge and another reason for breast-milk
[13] Bhise, N. S.; Manoharan, V.; Massa, S.; Tamayol, A.; monitoring programs. Environ. Health Perspect. 2000, 108,
Ghaderi, M.; Miscuglio, M.; Lang, Q.; Zhang, Y. S.; Shin, 387–392.
S. R.; Calzone, G. et al. A liver-on-a-chip platform with [25] Andrade, N. A.; McConnell, L. L.; Torrents, A.; Ramirez,
bioprinted hepatic spheroids. Biofabrication 2016, 8, 014101. M. Persistence of polybrominated diphenyl ethers in
[14] He, J. K.; Chen, R. M.; Lu, Y. J.; Zhan, L.; Liu, Y. X.; Li, agricultural soils after biosolids applications. J. Agric. Food
D. C.; Jin, Z. M. Fabrication of circular microfluidic Chem. 2010, 58, 3077–3084.
network in enzymatically-crosslinked gelatin hydrogel. Mater. [26] Guo, W. H.; Holden, A.; Smith, S. C.; Gephart, R.; Petreas, M.;
Sci. Eng. C-Mater. Biol. Appl. 2016, 59, 53–60. Park, J. S. PBDE levels in breast milk are decreasing in
[15] Xia, Y. N.; Whitesides, G. M. Soft lithography. Ann. Rev. California. Chemosphere 2016, 150, 505–513.
Mater. Sci. 1998, 28, 153–184. [27] Ward, J.; Mohapatra, S. P.; Mitchell, A. An overview of
[16] Mayorga-Martinez, C. C.; Hlavata, L.; Miserere, S.; López- policies for managing polybrominated diphenyl ethers
Marzo, A.; Labuda, J.; Pons, J.; Merkoçi, A. An integrated
(PBDEs) in the Great Lakes Basin. Environ. Int. 2008, 34,
phenol “sensoremoval” microfluidic nanostructured platform.
1148–1156.
Biosens. Bioelectron. 2014, 55, 355–359.
[28] Darnerud, P. O.; Eriksen, G. S.; Jóhannesson, T.; Larsen, P. B.;
[17] Mayorga-Martinez, C. C.; Hlavata, L.; Miserere, S.;
Viluksela, M. Polybrominated diphenyl ethers: Occurrence,
López-Marzo, A.; Labuda, J.; Pons, J.; Merkoçi, A. Nanos-
dietary exposure, and toxicology. Environ. Health Perspect.
tructured CaCO3-poly(ethyleneimine) microparticles for
2001, 109, 49–68.
phenol sensing in fluidic microsystem. Electrophoresis 2013,
[29] Porterfield, S. P. Vulnerability of the developing brain to
34, 2011–2016.
thyroid abnormalities: Environmental insults to the thyroid
[18] Medina-Sánchez, M.; Mayorga-Martinez, C.; Watanabe, T.;
system. Environ. Health Perspect. 1994, 102, 125–130.
Ivandini, T.; Honda, Y.; Pino, F.; Nakata, K.; Fujishima, A.;
[30] Ahn, K. C.; Gee, S. J.; Tsai, H. J.; Bennett, D.; Nishioka, M.
Einaga, Y.; Merkoçi, A. Microfluidic platform for environ-
G.; Blum, A.; Fishman, E.; Hammock, B. D. Immunoassay
mental contaminants sensing and degradation based on
for monitoring environmental and human exposure to
boron-doped diamond electrodes. Biosens. Bioelectron. 2016,
the polybrominated diphenyl ether BDE-47. Environ. Sci.
75, 365–374.
Technol. 2009, 43, 7784–7790.
[19] Tan, H. Y.; Loke, W. K.; Nguyen, N. T.; Tan, S. N.; Tay,
[31] Butryn, D. M.; Gross, M. S.; Chi, L. H.; Schecter, A.; Olson,
N. B.; Wang, W.; Ng, S. H. Lab-on-a-chip for rapid
J. R.; Aga, D. S. "One-shot" analysis of polybrominated
electrochemical detection of nerve agent Sarin. Biomed.
diphenyl ethers and their hydroxylated and methoxylated
Microdevices 2014, 16, 269–275.
[20] Ibarlucea, B.; Díez-Gil, C.; Ratera, I.; Veciana, J.; Caballero, analogs in human breast milk and serum using gas
A.; Zapata, F.; Tárraga, A.; Molina, P.; Demming, S.; chromatography-tandem mass spectrometry. Anal. Chim.
Büttgenbach, S. et al. PDMS based photonic lab-on-a-chip Acta 2015, 892, 140–147.
for the selective optical detection of heavy metal ions. [32] Li, Z.; Li, C.; Lin, D.; Kang, W. J.; Pan, D. Y.; Wu, M. H.
Analyst 2013, 138, 839–844. GC/MS analysis of polybrominated diphenyl ethers in
[21] Feng, C. Y.; Wei, J. F.; Li, Y. J.; Yang, Y. S.; Wang, Y. H.; vegetables collected from Shanghai, China. In Proceedings
Lu, L.; Zheng, G. X. An on-chip pollutant toxicity deter- of the 2013 International Conference on Material Science
mination based on marine microalgal swimming inhibition. and Environmental Engineering (MSEE 2013), 2013, pp
Analyst 2016, 141, 1761–1771. 292–296.
[22] Zheng, G. X.; Li, Y. J.; Qi, L. L.; Liu, X. M.; Wang, H.; [33] Liu, Q.; Shi, J. B.; Sun, J. T.; Wang, T.; Zeng, L. X.; Zhu,
Yu, S. P.; Wang, Y. H. Marine phytoplankton motility sensor N. L.; Jiang, G. B. Graphene-assisted matrix solid-phase
integrated into a microfluidic chip for high-throughput dispersion for extraction of polybrominated diphenyl ethers
pollutant toxicity assessment. Mar. Pollut. Bull. 2014, 84, and their methoxylated and hydroxylated analogs from
147–154. environmental samples. Anal. Chim. Acta 2011, 708, 61–68.

| www.editorialmanager.com/nare/default.asp
Nano Res. 15

[34] Zhang, H.; Lee, H. K. Plunger-in-needle solid-phase detection of hepatitis B surface antigen. Anal. Biochem.
microextraction with graphene-based sol–gel coating as 2013, 441, 1–7.
sorbent for determination of polybrominated diphenyl ethers. [42] Xu, R. Q.; Lu, Y. Q.; Jiang, C. H.; Chen, J.; Mao, P.; Gao,
J. Chromatogr. A 2011, 1218, 4509–4516. G. H.; Zhang, L. B.; Wu, S. Facile fabrication of three-
[35] Orozco, J.; Mercante, L. A.; Pol, R.; Merkoçi, A. Graphene- dimensional graphene foam/poly(dimethylsiloxane) composites
based Janus micromotors for the dynamic removal of and their potential application as strain sensor. ACS Appl.
pollutants. J. Mater. Chem. A 2016, 4, 3371–3378. Mater. Interfaces 2014, 6, 13455–13460.
[36] Chen, M. T.; Tao, T.; Zhang, L.; Gao, W.; Li, C. Z. Highly [43] Ju, T.; Ge, W.; Jiang, T.; Chai, C. Polybrominated diphenyl
conductive and stretchable polymer composites based on ethers in dissolved and suspended phases of seawater and
graphene/MWCNT network. Chem. Commun. 2013, 49, in surface sediment from Jiaozhou Bay, North China. Sci.
1612–1614. Total Environ. 2016, 557–558, 571–578.
[37] Shahzad, M. I.; Giorcelli, M.; Shahzad, N.; Guastella, S.; [44] Lammel, G.; Audy, O.; Besis, A.; Efstathiou, C.; Eleftheriadis,
Castellino, M.; Jagdale, P.; Tagliaferro, A. Study of carbon K.; Kohoutek, J.; Kukučka, P.; Mulder, M. D.; Přibylová, P.;
nanotubes based polydimethylsiloxane composite films. In Prokeš, R. et al. Air and seawater pollution and air-sea gas
Proceedings of the 6th Vacuum and Surface Sciences exchange of persistent toxic substances in the Aegean Sea:
Conference of Asia and Australia (VASSCAA-6), Islamabad, Spatial trends of PAHs, PCBs, OCPs and PBDEs. Environ.
Pakistan, 2013. Sci. Pollut. Res. 2015, 22, 11301–11313.
[38] Lee, A. C.; Liu, G. D.; Heng, C. K.; Tan, S. N.; Lim, T. M.; [45] Cristale, J.; Hurtado, A.; Gómez-Canela, C.; Lacorte, S.
Lin, Y. H. Sensitive electrochemical detection of horseradish Occurrence and sources of brominated and organophosphorus
peroxidase at disposable screen-printed carbon electrode. flame retardants in dust from different indoor environments
Electroanalysis 2008, 20, 2040–2046. in Barcelona, Spain. Environ. Res. 2016, 149, 66–76.
[39] Medina-Sánchez, M.; Miserere, S.; Marín, S.; Aragay, G.; [46] Xu, D.; Zhu, W.; Wang, C.; Tian, T.; Li, J.; Lan, Y.; Zhang,
Merkoçi, A. On-chip electrochemical detection of CdS quantum G. X.; Zhang, D. Q.; Li, G. T. Label-free detection and
dots using normal and multiple recycling flow through discrimination of poly-brominated diphenylethers using
modes. Lab Chip 2012, 12, 2000–2005. molecularly imprinted photonic cross-reactive sensor arrays.
[40] Zhang, J. L.; Yang, H. J.; Shen, G. X.; Cheng, P.; Zhang, J. Y.; Chem. Commun. 2014, 50, 14133–14136.
Guo, S. W. Reduction of graphene oxide via L-ascorbic [47] Radhakrishnan, R.; Suni, I. I.; Bever, C. S.; Hammock, B. D.
acid. Chem. Commun. 2010, 46, 1112–1114. Impedance biosensors: Applications to sustainability and
[41] Nourani, S.; Ghourchian, H.; Boutorabi, S. M. Magnetic remaining technical challenges. ACS Sustainable Chem. Eng.
nanoparticle-based immunosensor for electrochemical 2014, 2, 1649–1655.