CSI number 2006-47
Place, Date Männedorf, 2006-08-09
Author Christian Brunner
Number of pages 2, plus appended pages from Freedom EVO service manual
Affected
instrument(s) Freedom EVO series (EVO, EVO clinical) with PosID-3
Subject PosID-3 changes on X drive
Situation
The PosID-3 X-drive has been changed to have the cable holder chain running completely
underneath the worktable. This improvement reduces the noise level of the PosID-3 without
having an impact on the reading performance of the PosID-3.
The slotted holes (A) in the PosID-3 base had to be repositioned, which required a change
on new EVO-2 chassis.
A B
OLD
NEW
Cable chain
holder OLD NEW
This change has been implemented on new EVO-2 instruments as of serial numbers:
Should you have questions about this CSI do not hesitate to contact the Expertline.
Install the PosID-3 onto the base See section 8.13.4, 8-165
Scope The upgrade of the PosID-3 base described in this section is applicable on
Freedom EVO-2 with serial no. lower or equal to the numbers in the table below:
PosID-3 Base The figure below shows the PosID-3 base versions before and after the upgrade.
Differences Die “old” version (A) has the cable holder chain in the back behind the X-shaft.
The “new” version (B) has the cable holder chain in front of the X-shaft.
B
C
E D
C
F
Mount the self-adhesive plastic tape (PTFE) on the backside of the worktable
as shown in Fig. 5-17, 5-20 and make sure that you have the tape in a
distance (Y) of 5 mm ( 0.2 in.) from each edge of the worktable.
Cut the tape length (X) according to the following table:
Note: The “new” PosID-3 base will only be fixed with two screws in the front.
The two PosID-3 base fixing bolts in the back are removed (see step 4, 5-19)
8.13.7 X-Axis
Note: For Freedom EVO-2 instruments with appropriate serial numbers the X-
drive assembly cannot be just replaced. The PosID-3 first has to be upgraded with
the new base.
Refer to cross references above.
Note: The X-drive assembly includes the X-motor, the second encoder and the
home position sensor (fork-type light barrier).
A B
Fig. 8-205 X-belt lock/tensioner
3 Loosen the two screws (B) of the belt tensioner (A) on the left side of the
PosID-3 to slacken the tension of the X-belt (C).
D
C
Installing To install the X-drive assembly, proceed in reverse order as described for
removal.
Pay attention to the following:
A X-belt
B Second encoder
C X-motor
D D Idler pulley
E Fixing holes for cable holder
chain (upper end)
C
Tests and To ensure operating readiness, perform the following setups and tests:
Settings Refer to the “Instrument Software Manual”.
Set defaults in EEPROM
Autodetection of home position
Autorange
Determine connect positions (scaling and offset)
Autodetection of no tube sensor position
Set up reading positions
Reading positions test
No tube sensor test
Barcode reading test