Professional Documents
Culture Documents
discuss the relationship between pressure. 299.35 g/mol) is an antibiotic agent used
LOD, sample weight and the accu- to treat bacterial infections. It belongs to
racy of LOD in thermogravimetric TGA at low pressure the group of β-lactam antibiotics and
measurements. Figure 1 shows schematically how a TGA acts by inhibiting the cell wall synthesis
can be operated at low pressure. The low of various bacteria. Imipenem is avail-
Introduction pressure is created by connecting the fur- able as the monohydrate (CAS 74431-23-5;
In Part 1 of this series [1], we described nace outlet to a suitable vacuum pump the stoichiometric water content of the
two gravimetric techniques (TGA and via a cold trap. The cold trap is only monohydrate is 5.7%).
the halogen moisture analyzer) and the necessary if the decomposition gases are
Karl-Fischer titration for determining the likely to damage the pump. The pressure
water content in materials. In contrast in the TGA is monitored by a pressure
to the Karl-Fischer titration, we saw that gage connected to the purge gas inlet. A
gravimetric methods are not selective be- balance protection gas (usually nitrogen)
cause the mass losses measured during should be used even if the TGA is oper-
drying can be also be caused by the loss ated at low pressure. As usual, the protec-
of other volatile substances or decompo- tive gas inlet is used for this. The gas flow
sition products and not only by the loss should be set using a suitable rotameter – Figure 2. Structural formula of imipenem
(C12H17N3O4S). Possible decomposition
of water. Gravimetric methods are only the METTLER TOLEDO gas boxes cannot products are COx, NHx and H2O. Molar mass:
selective with regard to water if the gas- be used. If a reactive gas is also needed, 299.35 g/mol. Source: https://en.wikipedia.
org/wiki/Imipenem
eous products released during the drying the reactive gas inlet can be used – here
process are continuously identified. This again using a rotameter to set the gas
is possible if the TGA is combined with a flow. The pressure in the TGA should not Figure 3 shows TGA and DTG curves
suitable evolved gas analysis system. be less than 10 mbar. obtained by measuring imipenem in
Figure 1.
Operation of a TGA
at low pressure.
Pressure gauge
Gas supply
Vacuum pump
2 METTLER TOLEDO UserCom 52
nitrogen at a heating rate of 20 K/min Figure 3.
Imipenem at ambi-
at ambient pressure and at 26 mbar. ent and 26 mbar
pressure. TGA
The red curve in the upper diagram in curves and their
first derivatives.
Figure 3 is the TGA curve measured at The red curves
ambient pressure. It shows the typical were measured in
nitrogen at ambient
initial curve shape of substances con-
pressure and
taining moisture (the loss of mass ac- the blue curves
celerates with increasing temperature). at a pressure of
26 mbar.
At about 178 °C, the imipenem rapidly
decomposes. The corresponding DTG
curve in the lower diagram shows that
the release of moisture (including water
of crystallization) and decomposition
overlap. This means that the moisture
content cannot be determined from the
TGA curve. However, if the imipenem
sample is measured at low pressure, in
this case at 26 mbar (blue curves), the
Figure 4.
mass loss due to moisture is separated Repeatability of the
from the mass loss due to decomposition. LOD of four samples
The moisture content can therefore easily of imipenem mea-
sured at a pressure
be calculated from the TGA curve (6.5%, of 26 mbar.
of which 5.7% is water of crystallization).
METTLER TOLEDO UserCom 52 3
the case from 178 °C onward. This means this, a calibration measurement was per- for the water content of the sample.
that above this temperature products are formed using calcium oxalate monohy- The TGA/MS experiment confirms that
released that contribute to the m/z 17 ion drate (the procedure is described in refer- it really is exclusively water.
current but not to the m/z 18 ion current. ence [2]). This gave an MS sensitivity for
Since we expect NHx decomposition prod- water (m/z 18) of 510 nsA per milligram How accurately can the LOD
TA Tip
ucts, we assume that NH3 (m/z 17) ex- water. This allows us to quantify the re- be determined by TGA?
plains why the m/z 17 and m/z 18 curves lease of water up to 178 °C. The result was Pharmaceutical products in particular
differ. The fact that above 178 °C the 4.75%, which is markedly lower than the have LOD values that are usually very low
shape of the m/z 16 (NH2) curve approxi- step height between 50 and 178 °C on the (<1%). The question arises as to how ac-
mately matches the shape of the m/z 17 TGA curve (5.85%). The difference is due curately such low contents can be deter-
curve confirms this hypothesis. We also to the fact that, at the heating rate used mined by TGA and how large the sample
note that CO2 (m/z 44) is released from (20 K/min), imipenem already begins to weight must be in order to measure an
about 150 °C onward. Apparently, the decompose at 150 °C. In fact, there are expected LOD with a particular accuracy.
decomposition of imipenem begins at a two reasons why the value of 4.75% does From the technical point of view, LODs
temperature appreciably below the abrupt not correspond to the actual water content are evaluated as horizontal steps on TGA
mass loss at 178 °C. of the sample. First, water is a possible curves. If a simple horizontal step evalu-
decomposition product of imipenem, and ation is obviously not possible, different
What can we say about the water content second, the release of water initially pres- processes overlap and the evaluation
of the imipenem sample? Up to 178 °C ent (water of crystallization) is not com- of the LOD is at least questionable (ex-
the shape of the m/z 17 and m/z 18 pleted at 178 °C. ample: imipenem at ambient pressure,
curves is the same. The area under the see Figure 3). Let us assume that the LOD
m/z 18 curve is therefore proportional to In this example, the vacuum method can be clearly separated on the TGA curve
the amount of water released. To estimate probably gives the most credible value from decomposition processes (example:
imipenem at 26 mbar, see Figure 3).
Figure 5. In this case, the accuracy with which a
TGA-MS measure- step height can be evaluated is mainly
ments of imipenem given by the resolution and the noise
using open 40-µL
aluminum crucibles. level of the balance. If we furthermore
The heating rate assume a combined standard deviation
was 20 K/min.
of 2 ug, and also assume that the true
LOD should lie within the required ac-
curacy with a probability of 0.95, we get
the minimum sample weights listed in
Table 1. Example: if an LOD of 1% is to
be determined so that the true LOD lies
between 0.99% and 1.01% (required ac-
curacy is 1%) with a probability of 0.95,
a sample weight of at least 40 mg must
be used. Further details can be found in
reference [3].
Table 1.
The table shows Required Accuracy Expected effect in %
the sample weight in % 0.1 0.2 0.5 1 2 5 10
in milligrams nec-
essary to determine 0.1 4000 2000 800 400 200 80 40
a step-shaped LOD
mass loss with a 0.2 2000 1000 400 200 100 40 20
certain step height
and a given 0.5 800 400 160 80 40 16 8
accuracy.
1 400 200 80 40 20 8 4
2 200 100 40 20 10 4 2
5 80 40 16 8 4 1.6 0.8
10 40 20 8 4 2 0.8 0.4
4 METTLER TOLEDO UserCom 52
The sample weights given in the table
Tips and tricks for the determination of LOD by TGA
only take into account the uncertainties
that have to do with the weighing pro- • Samples can absorb or release moisture during the wait time on the sam-
cess. If the moisture content of a sample ple turntable. In an experiment, it is therefore important to enter the mass
is low, sorption or desorption processes of the crucible as well as the sample mass. Changes in mass that occur
during the wait time on the sample turntable or during the insertion of the
that occur during sample preparation
sample into the TGA furnace can then be quantified.
and insertion of the sample into the TGA
also contribute to the uncertainty of the • If the TGA is equipped with a sample robot, aluminum crucibles can be
measured moisture content. The con- hermetically sealed with a special lid. In this case, the crucible is opened
tribution depends on the sample under by a needle in the sample robot (the lid piercing kit) immediately before
investigation and is difficult to estimate. the measurement. Sorption or desorption processes during the wait time
can thereby be prevented.
Possibilities as to how these contribu-
tions can be quantitatively estimated • If ceramic crucibles are used, they can be covered with an aluminum lid.
and reduced are summarized in the The sample robot removes the lid after inserting the crucible into the fur-
green text area. nace. Sorption or desorption processes during the wait time on the sample
turntable can then be practically excluded.
Conclusions
LOD determinations can be carried out the experiment. This is the only way to [2] Ni Jing, Determination of the water
content of an ionic liquid, UserCom 40,
by TGA at ambient pressure and at low quantify mass changes that occur during
24–26.
pressure (minimum 10 mbar). Any prod- the wait time on the sample turntable or [3] K. Fritsche, M. Schubnell, Reliable
ucts released during drying can be iden- during the insertion of the sample into determination of sample mass – the
concept of minimum sample weight,
tified by coupling the TGA to a suitable the TGA furnace. UserCom 41, 14–17.
gas analysis system. Sufficiently large
sample masses must be used for mate- References
rials with low moisture contents. If the [1] M. Schubnell, C. De Caro, Moisture
materials exhibit sorption or desorption content, water content, loss on drying,
Part 1: What exactly is meant and
processes at room temperature, the mass how are these quantities determined? For more information:
of the empty crucible must be entered in UserCom 51, 1–8. www.mt.com/ta-moisture
METTLER TOLEDO UserCom 52 5
Polymorphism in metallic alloys, Part 2:
Measurement of the heat capacity of metastable
phases by TOPEM™ and Flash DSC
Applications
The article describes how the specific heat capacity of metastable phases can be determined by temperature-
modulated DSC using a metal alloy as an example. It also shows how the temperature range in which the
specific heat capacity of theses phase can be measured is significantly extended by Flash DSC.
Introduction we have to determine the Gibbs free ener- ing at several hundred degrees per second.
In Part 1 of this series of articles on the gy as a function of temperature [2]. This Flash DSC measurements with this alloy
monotropic polymorphism of metallic al- means we have to determine the specific enabled us to demonstrate the existence
loys, we showed how the transformation heat capacity, cp, of all the modifications of monotropic polymorphism in metal
kinetics and melting points of different involved. The problem that arises in the alloys for the very first time [3, 4, 5].
metastable modifications can be investi- determination of the specific heat capac-
gated by Flash DSC [1]. By polymorphism ity by conventional DSC is that the heat- DSC
of a metal alloy, we mean the occurrence ing rate used is relatively low and that The transformation enthalpies and
of different crystalline forms of the solid modifications can relatively easily trans- temperatures were measured at a heat-
alloy with the same macroscopic chemi- form. In this article, we show how the ing rate of 10 K/min using a METTLER
cal composition. These forms are modi- specific heat capacity of unstable modi- TOLEDO DSC 3+. The sample weight was
fications that differ in their atomic ar- fications can be determined by Flash DSC 32.93 mg. All the DSC and TOPEM mea-
rangement. We also refer to structures as and stochastic temperature-modulated surements were performed in 40-μL alu-
modifications if they are composed of dif- DSC (TOPEM). minum standard crucibles.
ferent types of crystals in sub-microscopic
dimensions (e.g. eutectic systems). The Experimental details TOPEM
decisive factor is that the entire modifica- The specific heat capacity was measured
tion participates in the transformation. Sample with high accuracy using stochastically
The experiments were performed with a temperature-modulated DSC ( TOPEM)
To describe the stability of the modifica- gold-based metal alloy (Au70Cu5.5Ag7.5Si17). [6, 7]. The temperature program can be
tion and understand the phase behavior, This forms a metallic glass on rapid cool- described by
6 METTLER TOLEDO UserCom 52
gram used in the experiments described Figure 2.
Specific heat ca-
here had the following additional param- pacity curves of the
eters: δT0 = 1 K and βu = ±2 K/min. A gold-based alloy,
calculation window width of 300 s was determined using
TOPEM.
used for the evaluation. All other method
and evaluation parameters were the de-
fault settings.
Flash DSC
The Flash DSC measurements were per-
formed with a Flash DSC 1 equipped with Figure 3.
a UFS 1 sensor. To achieve heating and Flash DSC heating
curves measured
cooling rates of several thousand kelvins at 1000 K/s.
per second, the heat capacity of the DSC The “blank” curve
furnace and the sample must be very low. was measured
at 0.5 K/s.
In addition, the thermal contact between
the sample and the sensor must be very
good. The Flash DSC technique is de-
scribed in references [9, 10, 11]. The mass
of the samples used was about 1.5 μg.
METTLER TOLEDO UserCom 52 7
During the relatively low heating and Measurement of the specific heat 1000 K/s (red curve in Figure 3) to mea-
cooling rates, metastable modifications capacity by Flash DSC sure the glass and the supercooled melt
tend to transform to more stable modi- above the glass transition.
fications. These transformations cause Creation and measurement
changes in the heat capacity during of different modifications The x1 modification was created by heat-
Applications
the measurement and are therefore a Flash DSC is an experimental technique ing the glass that had previously been
basic problem in the determination of that could be used to extend the tempera- cooled at 2000 K/s to 440 K and then im-
the c p of metastable modifications by ture range of metastable systems because mediately cooling it at 1000 K/s. Three
conventional DSC. For this reason, it the transformation temperatures of meta- temperature cycles were then performed
is not possible to determine reliable c p stable modifications are shifted due to at ±1000 K/s between 300 K and 500 K
data over a wide temperature range (see the high heating rates used [5]. In order for stabilization. The measurement
Figure 2). Equations (2) and (3) can be to measure the supercooled melt over curves of the last two cycles were identical
used for extrapolation and interpola- the greatest possible temperature range, (blue curve in Figure 3). The x2 modifi-
tion over a wide temperature range but the sample was first cooled from the cation was created by cooling from 750 K
the data is associated with a relatively equilibrated melt at 2000 K/s to 300 K. at 0.5 K/s. The heating measurement was
large uncertainty. A heating segment was then performed at then performed at 1000 K/s (black curve
in Figure 3).
8 METTLER TOLEDO UserCom 52
In Figure 5, the c p values of the melt capacity data should be known over the [5] J.E.K. Schawe, S. Pogatscher,
measured with TOPEM were compared widest possible temperature range. This J.F. Löffler, Thermodynamics of poly-
morphism in a bulk metallic glass:
with those of the heat capacity (Cp) from range can be extended using Flash DSC Heat capacity measurements by fast
the Flash DSC measurements (Figure 4) measurements because the high scan- differential scanning calorimetry,
Thermochimica Acta 685 (2020)
in a temperature interval around 700 K ning rates shift the transformation tem-
178518.
and a shift factor k calculated as shown peratures. The problem that arises with [6] TOPEM™-The new multi-frequency
in eq (5): the Flash DSC is that blank curves can- temperature-modulated technique,
UserCom 22, 6–8.
not be directly measured and that the
C [7] J. Schawe, DSC measurements
Eq (5) k = p∕cp sample mass is not accurately known. of metals by DSC and TGA/DSC,
The article describes how the thermal UserCom 46, 8–11.
[8] J.E.K. Schawe, T. Hütter, C. Heitz,
All the Flash DSC curves were multiplied losses of the Flash DSC measurements I. Alig, D. Lellinger, Stochastic tempera-
by this factor. The curves are in excellent can be corrected and how specific heat ture modulation: A new technique in
agreement with the data determined from capacities can be obtained. This is done temperature-modulated DSC, Thermo-
chim. Acta 446 (2006) 147–155.
the TOPEM measurements (Figure 5). by comparing the Flash DSC data with [9] J. Schawe, The revolutionary Flash
The temperature range with measured cp c p values measured using temperature DSC 1: maximum performance for
data is thereby extended, especially for modulation in a conventional DSC. metastable materials, UserCom 32,
12–16.
the x1 modification and the supercooled A detailed discussion of the method pre- [10] J. Schawe, Fast Scanning DSC:
melt. The measurement data can now be sented here is given in reference [5]. Selected application examples,
UserCom 49, 16–20.
described more accurately using the ap-
[11] J. Schawe, Fast Scanning DSC:
proximation functions (eqs (2) and (3)). References Practical tips and information,
This means that further studies on the [1] J. Schawe, Monotropic polymorphism UserCom 50, 14–19.
in metallic alloys, Part 1: Flash DSC [12] C. Quick, J.E.K. Schawe,
thermodynamics of phase behavior can investigation of the melting of meta- P.J. Uggowitzer, S. Pogatscher,
carried out with greater certainty. stable phases, UserCom 51, 19–21. Measurement of specific heat capacity
[2] D. Giron, Thermal analysis and calo- via fast scanning calorimetry-accura-
rimetric methods in the characterisa- cy and loss corrections, Thermochim.
Summary and conclusions tion of polymorphs and solvates, Acta 677 (2019) 12–20.
The specific heat capacity of metals Thermochim. Acta 248 (1995) 1–59.
[3] S. Pogatscher, D. Leutenegger,
can be measured by TOPEM or ADSC
J. E. K. Schawe, P. J. Uggowitzer,
with an accuracy of about 2% [7]. With J. F. Löffler, Solid-solid phase
metastable modifications, this is how- transitions via melting in metals,
Nat. Commun. 7 (2016) 11113.
ever only possible in a relatively small [4] S. Pogatscher, D. Leutenegger,
temperature range because the modi- J.E.K Schawe, M. Pavithran, R.
fications can undergo transformation Schaeublin, P. Uggowitzer, J. Löffler,
Monotropic polymorphism in a glass-
during the measurement. For ther- forming metallic alloy, J. Phys.: For more information:
modynamic reasons, the specific heat Condens. Mat. 30 (2018) 234002. www.mt.com/ta-fdsc
METTLER TOLEDO UserCom 52 9
Innovation and standardization:
The development of a standard for the Flash DSC
Applications
The application of a new measurement procedure in routine analysis is frequently tied to the existence of a
corresponding standard. The article describes the development of international standards, which can save
time-consuming, costly analytical method validation.
Importance and advantages as articles of trade and on the other hand, Standardization as
of standardization metal bars served as an early currency or pacemaker for innovation
A standard describes a uniform, widely means of payment (see Figure 1) [2]. Standardization has to the present day
recognized and applied procedure to still lost none of its importance for the
produce or perform something that has Even today, metal bars are still manu- easier exchange of products and services.
proven itself in comparison with other factured in different shapes and sizes In the course of technical progress in
procedures [1]. Standardization is the for the exchange of goods. Besides this, the age of the industrialization, it has,
process of developing and implementing bars of precious metals, especially gold, however, become increasingly important
standards. The availability of a standard serve as a safe investment. Quite early to describe new inventions and technical
can replace method validation in analy- on, however, coins made of metals of developments by uniform standards in
sis, which is required for quality assur- uniform size became commonly used order to make them applicable for a wid-
ance, accreditation or approval. This is as currency. Their individual value was er user group. The development of video
particularly important for new measure- stamped on them and they were easier technology is a good example of this.
ment techniques such as the Flash DSC. to handle due to their small size and low Worldwide, three systems known as PAL,
In this sense, the manufacture of met- weight. These early forerunners of stan- SECAM and NTSC were used to display an-
als in the shape of uniform bars in early dardization however had less to do with alog color television images; their global
Christian times can be looked on as an the introduction of new techniques but distribution is shown in Figure 2 [3].
early example of standardization. On the were rather a contract to facilitate the These systems enabled color television
one hand, it facilitated the use of metals exchange of goods and services. technology to spread throughout the
world. In the meantime, of course, these
Figure 1. analog standards have been replaced by
Celtic iron bars. digital techniques (DVB).
Source: © LAD,
Y. Mühleis,
https://books. In the field of instrumental analysis,
ub.uni-heidelberg. standardization also plays an important
de/heibooks
role by making innovative investigation
techniques available for less experienced
users. These new methods can only assert
themselves in industrial routine analy-
sis when standardized procedures for the
application, execution and evaluation of
new measurement techniques are avail-
able. Both instrument manufacturers
and potential users then profit from such
new developments.
10 METTLER TOLEDO UserCom 52
developed measurement techniques Figure 2.
Distribution of
will be discussed using the Flash DSC as analog video sys-
an example. tems by country.
Standardization
and interest groups
Standardization in the technical sense
describes the agreement, publication and
application of requirements, specifica-
tions, and guidelines by recognized orga-
nizations. Standards should be based on
secure knowledge derived from science,
technology and experience and ensure NTSC
that products, processes and services PAL or transition from SECAM to PAL
only SECAM
are suitable and accepted world-wide for not specified
the given purpose. Standardization can
help maximize the compatibility, safety,
Figure 3.
repeatability and quality of products and Overview of the ISO
services. member countries.
METTLER TOLEDO UserCom 52 11
Structure of the ISO bodies The development of a standard is possible 1. Proposal stage. Submission of
The Flash DSC is handled in TC 61 “Plas- despite existing patent protection if the a proposal for a new standard-
tics” under the more than 250 technical owner of the patent guarantees to grant ization project
committees. TC 61 is made up of a total licenses at standard market conditions. In this first step, the proposal for a new
of 31 full members and 38 observers and standardization project is presented
Applications
is divided into 11 sub-committees (SC). Before a standardization project is accept- and discussed in the national and ISO
The Flash DSC appears again in SC 5 ed, adequate resources must be available. working groups, and a decision taken
“Physical-Chemical Properties”. SC 5 is First, appropriate standards bodies have as to whether an official standardiza-
divided into 4 working groups (WG), of to be defined for the intended project and tion project should be requested.
which WG 8 “Thermal Analysis” deals at least 2⁄3 of the full members must be in
with the Flash DSC. favor. Furthermore, a project leader and 2. Preparatory stage. Inclusion of
at least 5 national experts must be named a new standardization project in
The national standards institutes, such for the development of the standard and the ISO work program
as DIN in Germany are similarly orga- a project plan must be prepared that in- The proposal for a new standardiza-
nized and often differ only in the names cludes the individual project milestones. tion project is made to ISO by a full
of the bodies and a slightly modified In addition, acceptance of the project re- member. Before the ballot about
classification. quires the creation of a draft work pro- acceptance, a project leader must be
gram that includes all the important ele- named and the field of application
Criteria for the acceptance of new ments of the standard. Normally written must be defined. Ideally however, a
standardization projects [3] and spoken communication is in English. complete first working draft should
Important criteria for justifying the in- be presented. The ballot lasts three
clusion of new standardization projects Stages in the development months and the standardization proj-
in the ISO work program are sufficient of a standardization project [3] ect must be supported by at least 2⁄3 of
need, market relevance, and a binding The development of an ISO standard the full members. In addition, at least
project plan. It should also be noted that can be divided into 6 stages whereby, in 5 national experts must be named for
all copyrights must be transferred to ISO some circumstances, the optional stages participation. Furthermore, the time-
and the use of trade names should as far (marked with a red star *) can be skipped frame including milestones for the
as possible be avoided. (see Figure 4): development of the standard must be
Figure 4.
Overview of the
individual stages
Comparing development processes for an IS
in the development
of an ISO standard. 6 stages Action Balloting time Default path Shortest path
Stages marked
with a red star (*)
• 3-month ballot by default
1
are optional.
Proposal Proposal to start • 2-month ballot possible
NP NP
a new project • TC/SC resolution for Straight to DIS
NP revision & amendments
2 Preparatory
WD *
Expert consensus
within working group WD
3 Committee
CD *
Committee consensus
• 3 or 4 month vote
possible
• Can be skipped
CD
4 Enquiry
DIS
National consensus
• 2-month translation
• 3-month ballot
24 months to reach
DIS
12 months to reach
DIS
5
• Skipped by default
Approval YES or NO vote • Can be introduced FDIS
• 2-month ballot
FDIS *
12 METTLER TOLEDO UserCom 52
defined. If the project is rejected, no Figure 5.
Review cycle of
further work occurs. It can however be Maintenance Periodic a standard.
revised and resubmitted at any time. of the standard review
3. Committee stage
Taking into account the comments
received in the ballot about the new
standardization project, a new work-
Confirmation Vote on the
ing draft is prepared on which the
of the revision review
full members vote within 2 months.
In the ballot, a consensus is sought,
but a majority of at least 2⁄3 is needed
for acceptance. All full members are
obliged to participate in the ballot
Vote on the Revision of the
and any rejections must be justified.
revision standard
If the project is rejected, the respon-
sible body decides whether a second
ballot can be considered after a suit-
able revision or whether the project The stages in the development process dard, information on its use, or special
should be withdrawn. of an ISO standard are summarized in advantages can be put in front of the
Figure 4. standard. Mandatory for each standard is
4. Enquiry stage a section at the beginning that describes
After assessing the comments received In many cases, the development of inter- the application field and that presents a
in the draft stage, the project leader nationally valid standards by ISO runs list of normative references as well as a
prepares a revised draft which must be parallel with the European Standards collection of terms and definitions that
voted on within 3 months. Here again a Organization CEN as EN ISO standards. are used in the particular standard.
consensus is sought, but at least 2⁄3 ap- A special feature of these standards is
proval and not more than ¼ rejection is that their inclusion in the national stan- What follows can be different depend-
necessary. Rejection without explana- dardization bodies of the CEN member ing on the purpose of the standard. For
tion and abstentions are not taken into countries is mandatory. the standardization of test methods,
account. If no technical comments are however, sections usually follow that de-
raised in the ballot, then the standard Review cycle of standards [3] scribe the measurement principle, define
is published after editorial review, The current status of ISO standards is pe- equipment requirements, sampling and
otherwise another formal final ballot riodically reviewed every 5 years. In ad- sample preparation, and how to perform
is necessary after taking the techni- dition, every full member can request an and evaluate measurements. Finally, the
cal objections into account. unscheduled revision. A revision requires test report contains information on the
a simple majority of the member coun- expected accuracy of the results and a
5. Approval stage tries. As a result of the review, the standard list of necessary details.
In this ballot, only technical com- is either confirmed unchanged, revised or
ments are allowed and the accep- withdrawn. If a revision is requested, the Optionally, normative or informative ap-
tance criteria (at least 2⁄3 approval responsible ISO body decides in which of pendices to describe special aspects of a
and maximum ¼ rejection) are the the above-mentioned stages the revision standard and literature references can be
same as in the previous stage. If the should begin. The typical review cycle of appended.
ballots in the committee, enquiry, a standard is shown in Figure 5.
or final draft stages are not in favor, Example of the newly developed
the responsible body decides whether Development of a standard standard for the Flash DSC [5]
to repeat the process after suitable To facilitate the readability and applica- The standard is entitled “Plastics – Fast
revision or to withdraw the project. tion of standards, formatting guidelines Differential Scanning Calorimetry – Chip
have been agreed that standardize the Calorimetry”. In the application field,
6. Publication stage structural organization of standards as it is stated that the standard is valid for
Once the standard has successfully well as the formatting of figures and non-adiabatic fast DSC instruments that
passed the enquiry or approval stage, tables. operate according to the principle of
the final text is released by the power compensation. In addition, the
project leader and published by ISO If desired, an introduction that describes application is limited to an open sample
within 6 weeks. the scientific background of the stan- geometry in which samples are positioned
METTLER TOLEDO UserCom 52 13
directly on the measuring sensors with- Future prospects [6] https://www.iso.org/directives-and-
out being sealed in a crucible so that ex- The publication of a standard for the ap- policies.html, ISO/IEC Directives, Part
2 (2018), Principles and rules for the
tremely high heating and cooling rates of plication of the Flash DSC for polymers structure and drafting of ISO and IEC
up to several thousand kelvins per second makes the necessary know-how available documents
can be achieved. Furthermore, a series for using the method in routine analysis. [7] https://www.iso.org/standard/77496.
Applications
The determination of the filler content in polymers is frequently performed by thermogravimetric analysis
(TGA). In this article, we discuss different methods for doing this using carbon-fiber reinforced PEEK as
an example.
14 METTLER TOLEDO UserCom 52
Δm PEEK – Δmcomposite
ξCF [%] = 100 · Δm PEEK (1) Figure 1.
TGA, DSC and DTG
curves of PEEK
where ΔmPEEK and Δmcomposite are the per- (blue), carbon
centage mass loss steps. The step heights fibers (green, only
TGA) and the
given in Figure 1 yield carbon fiber and composite (red),
PEEK contents of 15.6% and 84.4%. The measured in
residual carbon from the pyrolysis of nitrogen.
METTLER TOLEDO UserCom 52 15
ence temperature). The strong exotherm fore be automatically used to determine upper temperature limits correspond to
produced in the combustion of the PEEK the carbon fiber content from a measure- about the temperatures at which carbon
carbon heats the sample to temperatures ment in oxygen. (both residual carbon and the carbon
well above the reference temperature. fibers in the composite) starts to burn.
This is illustrated in Figure 3. It shows Between 500 and about 600 °C, the TGA The carbon fiber content determined in
Applications
the TGA and TGA/DSC curves from Fig- and DTG curve shapes for PEEK and this way thus agrees with the value cal-
ure 2 plotted as dotted-line curves (here composite are practically the same and culated from the measurements in ni-
the temperature corresponds to the refer- are only due to the decomposition of trogen within the limits of uncertainty
ence temperature). The solid-line curves PEEK (Figure 2). If the first maxima on given by the choice of the temperature
are the curves obtained when the corre- the DTG curve (indicated by the verti- limits for the step evaluation
sponding curves are plotted as a function cal dotted lines in Figure 2) are chosen
of the sample temperature. Comparison as upper temperature limits for the step In principle, the PEEK content (and
of the curves shows that during the com- evaluation of the first decomposition hence the carbon fiber content) in the
bustion of the PEEK carbon the sample step, a carbon fiber content of 15.9% is composite can also be determined by
temperature increases to temperatures at obtained from the two step heights using DSC. Both the step heights of the glass
which combustion of the carbon fibers eq 1. Comparison with the correspond- transition as well as the enthalpies of
can take place. Equation 1 cannot there- ing TGA/DSC curves shows that these fusion of PEEK in the composite and in
the sample of pure PEEK can be used for
this. Measurement curves of PEEK and
Figure 4.
DSC curves of the the composite are shown in Figure 4.
composite (red) If the enthalpies of fusion are evaluated,
and of PEEK (blue), the PEEK content of the composite cor-
measured at
10 K/min in responds to the ratio of the peak areas
nitrogen. (normalized to sample mass) of the PEEK
melting peak of the composite and the
pure PEEK sample. This yields a PEEK
content of 85.5%. The evaluation of the
areas of the melting peaks depends on
the choice of the baseline (Line, Spline,
etc.) as well as on the integration limits.
We estimate the associated uncertainty of
the PEEK content to be about ±4%.
16 METTLER TOLEDO UserCom 52
Table 1.
TGA DSC Filler content of
Filler content
Step evaluation Enthalpy of fusion Step height Tg PEEK and carbon
fibers in a compos-
ite. Blue-shaded
In nitrogen 15.6% ± 0.2% boxes: Results
14.5% ± 4% 13.9% ± 10% based on mea-
surements of the
In oxygen 15.9% ± 1.0% individual compo-
(independent of (independent of
nents (PEEK and
the atmosphere) the atmosphere)
N2 to 800 °C carbon fibers) and
15.8% ± 0.3% the composite.
O2 from 300 °C
Green-shaded box:
Determination from
a TGA measurement
The measurement of the composite in oxygen exhibits two distinct but not tainties involved are significantly larger of the composite.
The uncertainties
sample in oxygen showed that the com- completely separated exothermic mass than the uncertainties arising from the are based on
bustion of the residual carbon from the loss steps. They correspond to the com- TGA measurements. estimates.
PEEK and the carbon fibers cannot be bustion of the residual carbon of PEEK
separated. If only the composite is avail- (on the DSC curve at about 510 °C) and If only the composite is available, its
able, an experiment will have to be de- the carbon fibers (on the DSC curve at composition can only be determined
signed in which the combustion of the about 710 °C). This yields a value of from a TGA measurement. The method
two types of carbon (residual carbon 15.9% (step height of the second mass for doing this consists of three segments:
from PEEK and carbon fibers) can be loss step) for the carbon fiber content. In the first segment, the sample is pyro-
separated. This is possible if the two types Due to the overlap of the two combustion lized in nitrogen. In the second segment,
of carbon burn in different temperature processes we estimate the uncertainty of the residual sample is cooled under py-
ranges. The results for the measurements the step evaluation to be about ±0.3%. rolysis conditions to about 300 °C. In
in oxygen prove that this is the case for It should be possible to improve the sepa- the third segment, the sample is heated
this particular composite. A method ration of the two combustion processes in oxygen. This enables the combustion
frequently used in such cases involves by varying the heating rate during the processes of the residual polymer carbon
heating the sample in nitrogen to a tem- second heating segment. and the carbon fibers to be separated.
perature at which pyrolysis is complete
(in our example about 900 °C), cooling Summary and conclusions The results are summarized in Table 1.
the residual sample in nitrogen to 300 °C We have shown how the composition The values in the blue-shaded boxes are
(in this cooling segment the sample mass of a composite consisting of PEEK and based on a comparison of measurements
remains unchanged), and then heating it carbon fibers can be determined by TGA. of the composite and its individual com-
in oxygen to 800 °C (combustion of car- If the individual constituents of the com- ponents (PEEK, carbon fibers). The value
bon). Several TGA and TGA/DSC curves posite are also available in addition to in the green-shaded box results from the
obtained for the PEEK composite in such the composite, the carbon fiber content measurement of the composite using the
an experiment are shown in Figure 5. The is most easily and accurately determined (three segment) method described above.
first TGA heating run is the same as the by performing TGA measurements in
TGA curve shown in Figure 1. On cool- an inert atmosphere. Alternatively, the
ing, the sample mass remains constant carbon fiber or the PEEK content can be For more information:
as expected. The second heating run estimated by DSC. However, the uncer- www.mt.com/ta-techniques
METTLER TOLEDO UserCom 52 17
TGA-IST-GC/MS-FTIR analysis
of polypropylene
Applications
Nicolas Fedelich
18 METTLER TOLEDO UserCom 52
were identified using the NIST/EPA/NIH Figure 3.
Total ion chromato-
mass spectral database [2]. gram (TIC) of the
gas sample taken
The Bruker FTIR spectrometer was at 450 °C for PP-FR
with the names of
equipped with a gas cell, and was con- some of the identi-
nected to the outlet of the IST16 interface fied decomposition
products.
by an additional transfer line. This en-
abled FTIR spectra of the decomposition
gases to be continuously recorded during
the TGA experiment (online). The trans-
fer line and the gas cell were heated at
250 °C. The IR spectra were scanned from
4000 to 400 cm-1 at a resolution of 4 cm-1;
averaging 16 spectra each time.
Results
METTLER TOLEDO UserCom 52 19
calcium carbonate in the PP-FR sample a second experiment, marked with blue PP-FR are practically identical (in par-
(calcination of CaCO 3 does not begin crosses on the black curve in Figure 2). ticular alkenes, e.g. 2,4-dimethyl-1-hep-
until about 600 °C and is not of inter- tane, see also Figure 4).
est here). GC/MS measurements
In a second experiment, gas samples The total ion chromatogram (TIC) of the
Applications
The simultaneously measured DSC were collected for both materials at the gas sample taken at 380 °C for PP-FR
curves show the melting of the polypro- temperatures determined before and shows a prominent peak at 18.747 min
pylene at about 160 °C. The different spe- then analyzed by GC/MS. The results are (besides numerous smaller peaks).
cific enthalpies of fusion are the result of shown in Figures 3 to 6. They show the The MS analysis identified the peak as
the different compositions of the two ma- total ion chromatograms (TICs) of gas 1,3-benzenedicarbonitrile (see Figure 6).
terials (PP-ref: 100% polypropylene and samples taken at 450 °C (Figure 3) and This substance was only detected in the
PP-FR: 70% polypropylene, 20% flame 380 °C (Figure 5). Each peak in the two temperature range 300 to 380 °C and
retardant, 10% CaCO3). chromatograms corresponds to a dif- only for PP-FR. It is therefore a typical
ferent decomposition product; the most decomposition product of the flame re-
The two TGA curves were used to de- important decomposition products are tardant used.
fine the TGA temperatures at which gas labeled next to the peaks. At 450 °C, the
samples were collected in the IST16 (in decomposition products of PP-ref and FTIR measurements
Figure 7 shows the TGA curves of the two
samples together with their correspond-
Figure 6.
Decomposition ing Gram–Schmidt (GS) curves. A GS
product of PP-FR at curve displays the area under the FTIR
a retention time of spectra as a function of temperature and
18.747 min. Above:
the measured mass corresponds qualitatively to the DTG
spectrum. Below: curve. The GS curve is used to set the
the mass spectrum
temperatures at which the FTIR spectra
from the database
(best fit with the of the decomposition gases are exam-
measured mass ined in more detail (normally these are
spectrum) and the
result of the data-
the peak temperatures of the GS curve).
base search – The dashed vertical lines in Figure 7
the decomposition mark the temperatures at which we ana-
product is most
probably 1,3-ben-
lyzed the FTIR spectra more closely (see
zenedicarbonitrile. also Figures 9 and 10).
20 METTLER TOLEDO UserCom 52
Figure 8.
Overview of the
FTIR spectra mea-
sured during the
TGA measurement.
The time axis cor-
responds to the
temperature.
Figure 9.
The spectra at 460 °C for PP-ref and PP-FR The diagram shows
differ only insignificantly from one an- the FTIR spectra for
PP-ref and PP-FR
other and show that mainly decomposi-
measured at 460 °C
tion of polypropylene occurs around this and the reference
temperature. spectrum of 2,4-
dimethyl-1-heptene
(best fit) from the
The situation is quite different at 380 °C: NIST/EPA database.
the FTIR spectra for PP-ref and PP-FR
differ considerably at this temperature
(see Figure 10). It can be seen that the
decomposition of PP-FR produces ammo-
nia and CO2 in addition to 2,4-dimethyl-
1-heptene. Ammonia is also a typical
decomposition product of the flame re-
tardant. The FTIR measurements also
confirm the result of the GC/MS analysis
namely that the initial decomposition of
PP-FR at lower temperatures is due to the
decomposition of the flame retardant. Figure 10.
The diagram shows
GC/MS and FTIR the FTIR spectra
for PP-ref and
The peak areas of the TIC peaks for the PP-FR at 380 °C
different decomposition products in the and reference
spectra from the
15 storage loops allow emission profiles
NIST/EPA database
to be created for each substance detected. for 2,4-dimethyl-
The FTIR spectra can be used to calculate 1-heptene, ammonia
and CO2.
temperature-dependent IR intensities in
selected frequency ranges and produce
so-called chemigrams. Figure 11 dis-
plays emission profiles for 2 substances
(1,3-benzenedicarbonitrile and 2,4-di-
methyl-1-heptene) and chemigrams cal-
culated for 3 different infrared frequency
ranges, 979–910 cm-1 (ammonia), 2440–
2120 cm-1 (CO2) and 3260–2640 cm -1
(C–H stretching vibrations). Also included
METTLER TOLEDO UserCom 52 21
the decomposition of polypropylene (e.g.
ammonia, 1,3-benzenedicarbonitrile),
but are produced by the decomposition of
the flame retardant.
Applications
22 METTLER TOLEDO UserCom 52
Determination of the constituents
of cement clinker by TGA
Dr. Angela Hammer
TGA is a simple technique that can be used to determine the content of individual constituents in a mixture
of different materials that do not influence one another. This is illustrated in the following example using a
sample from the cement industry.
Introduction Figure 1.
TGA and DTG
The content of minerals [1], for example curves of the
calcium carbonate in inorganic mix- decomposition
tures, can easily be determined by TGA of CaCO3.
[2] if their decomposition profiles dif-
fer. The following example shows how
the content of calcium carbonate, cal-
cium hydroxide, magnesium carbonate
and magnesium hydroxide in a sample
from the cement industry (cement clin-
ker: kiln-fired constituent of cement)
can be determined. Each pure substance
decomposes in a different temperature
range so their individual contents in a
mixture can be determined if the ma-
terials do not influence one another
during decomposition. The method was
checked by investigating a mixture with
known contents similar to those of the Table 1.
cement clinker. CaCO3 Ca(OH)2 MgCO3 ·xH20 Mg(OH)2 Overview of the
Moisture 30–200 °C 30–200 °C 30–150 °C 30–200 °C different decompo-
sition processes
Experimental details Dehydration 250–500 °C 150–300 °C and temperature
Samples of pure minerals (CaCO 3 ranges.
>99%, Ca(OH)2 >96%, MgCO 3 hydrate Dehydroxylation 500–750 °C 280–800 °C
(pharmaceutical secondary standard), Decarboxylation 600–800 °C 350–500 °C
Mg(OH)2 >99%) and an unknown mix-
ture were investigated by TGA/DSC be- Table 2.
Molar mass (g/mol) Factor Molar masses of
tween 30 and 1000 °C at a heating rate of the substances
CaCO3 100.0869 2.273
10 K/min in a nitrogen atmosphere. The investigated and
measurements were performed using Mg(OH)2 53.3197 2.96 the factors used
to determine the
sample weights of 10 to 50 mg in 70-µL Ca(OH)2 74.093 4.11 substances investi-
aluminum oxide crucibles. To check the gated in mixtures.
MgCO3 84.3139 1.916 The percentage
method, a mixture containing low con-
weight loss of the
tents of CaCO3 , Ca(OH)2 , Mg(OH)2 and CaO 56.0774
relevant decompo-
Al 2 O 3 as an inert matrix was blended MgO 40.3044 sition step is
multiplied by the
and measured using the same tempera- CO2 44.01 relevant factor.
ture program.
Ca 40.078
Measurements and results O 16
The pure substances were first measured H 1
to determine the temperature range in
Mg 24.305
which they decompose (Figures 1–4).
METTLER TOLEDO UserCom 52 23
• CaCO3 shows a decomposition step
(decarboxylation) from CaCO3 to CaO
between 600 and 800 °C in addition
to the initial loss of moisture between
30 and 200 °C.
Applications
24 METTLER TOLEDO UserCom 52
Figures 5 and 6 show the TGA curves of Figure 5.
TGA and DTG curves
the cement clinker (Figure 5) and a mix- of the decomposi-
ture of known composition (Figure 6). tion of an unknown
The calculated content of cement clinker mixture (cement
clinker).
together with the contents of the known
mixture (by weight, and measured by
TGA) is summarized in Table 3. The re-
sults showed that the contents determined
for the known mixture agree well with the
amounts weighed in.
Conclusions
TGA is a simple technique that can be
used to determine the quantitative con-
tent of individual constituents in a mix-
ture of different materials that do not
influence one another. In the example de-
scribed, moisture, Mg(OH)2, Ca(OH)2 and
CaCO3 were identified and quantified in a
Figure 6.
sample of unknown composition. The ce- TGA and DTG curves
ment clinker was found to contain 0.3% of the decomposi-
moisture, 0.3% Mg(OH)2, 0.8% Ca(OH)2 tion of a mixture of
known composition.
and 0.2% CaCO 3 in addition to other
inorganic constituents (about 98.4%).
The method was checked by preparing
and analyzing a mixture containing
(known) small amounts of Mg(OH)2 ,
Ca(OH)2 and CaCO3.
References
[1] Lehrbuch der anorganischen Chemie, Table 3.
Unknown mixture
Hollemann Wiberg, 80. Auflage, Substance Mixture with known contents Content of inorganic
Walther de Gruyter & CO., 1971,
(cement clinker) minerals in an
page 469. Weighed in Measured unknown mixture
[2] Barbara Lothenbach, Paxel Durdzinski, (cement clinker)
Klaartje De Weerdt, A Practical Moisture 0.3% 0.2% and in a mixture
Guide to Microstructural Analysis of containing known
Cementitious Materials, Chapter 5, Mg(OH)2 0.3% 1.2% 1.3% contents of
CRC Press. substances.
Ca(OH)2 0.8% 0.6% 0.4%
MgCO3 –
CaCO3 0.2% 1.3% 1.4%
For more information: Other inorganic
98.4% 96.9% 96.7%
www.mt.com/ta-tga constituents
METTLER TOLEDO UserCom 52 25
Exhibitions, Conferences and Seminars
GTAP webinar February 4–5, 2021 Sweden
RQ2020 Joint RQ 17 and ISMANAM 27 conference August 22–27, 2021 Warsaw (Poland)
www.mt.com/ta-webinars
Webinars For information on our current program of webinars, please visit
www.mt.com/ta-ondemand
Handbooks Our specialist application handbooks provide valuable background information www.mt.com/ta-handbooks
www.mt.com/ta-videos
• or check our training videos
www.mt.com/ta-techniques
TA News You can find the latest product news and an overview of our thermal analysis services at www.mt.com/ta-news
METTLER TOLEDO offers comprehensive support to ensure you get reproducible results www.mt.com/ta-services
TA Services
in day-to-day analysis www.mt.com/ta-gtap
Tip Sign up for the quarterly thermal analysis knowledge newsletter www.mt.com/ta-knowledge
Kurs TAK11 TMA (Deutsch) 15. März 2021 Course TAK12 TMA (English) March 22, 2021
Kurs TAK11 DMA Basic (Deutsch) 15. März 2021 Course TAK12 DMA Basic (English) March 22, 2021
Kurs TAK11 Administration (Deutsch) 16. März 2021 Course TAK12 Administration (English) March 23, 2021
Kurs TAK11 DMA Advanced (Deutsch) 16. März 2021 Course TAK12 DMA Advanced (English) March 23, 2021
Kurs TAK11 TGA Basic (Deutsch) 16. März 2021 Course TAK12 TGA Basic (English) March 23, 2021
Kurs TAK11 TGA Advanced (Deutsch) 17. März 2021 Course TAK12 TGA Advanced (English) March 24, 2021
Kurs TAK11 DSC Basic (Deutsch) 17. März 2021 Course TAK12 DSC Basic (English) March 24, 2021
Kurs TAK11 DSC Advanced (Deutsch) 18. März 2021 Course TAK12 DSC Advanced (English) March 25, 2021
Kurs TAK11 TGA-FTIR (Deutsch) 18. März 2021 Course TAK12 TGA-FTIR (English) March 25, 2021
Kurs TAK11 TGA-GC/MS (Deutsch) 18. März 2021 Course TAK12 TGA-GC/MS (English) March 25, 2021
Kurs TAK11 SW Advanced (Deutsch) 19. März 2021 Course TAK12 SW Advanced (English) March 26, 2021
Kurs TAK11 TGA-MS (Deutsch) 19. März 2021 Course TAK12 TGA-MS (English) March 26, 2021
Kurs TAK11 TGA-Micro GC/MS (Deutsch) 19. März 2021 Course TAK12 TGA-Micro GC/MS (English) March 26, 2021
Kurs TAK11 Kinetics / TMDSC (Deutsch) 19. März 2021 Course TAK12 Kinetics / TMDSC (English) March 26, 2021
Information and Course Registration / Auskunft und Registration bei: Mrs Perla Irmler (or Mrs Veronica Feroce-Ring), Mettler-Toledo GmbH, Analytical, Nänikon,
Tel: ++41 44 806 73 57, Fax: ++41 44 806 72 60, e-mail: perla.irmler@mt.com (or veronica.feroce-ring@mt.com)
26 METTLER TOLEDO UserCom 52
Local TA Customer Courses
Schweiz/Deutschland Nähere Informationen unter www.mt.com/labtalk oder durch: Frau Janine Schindler, Mettler-Toledo GmbH, Giessen, Tel: ++49 641 507 405, e-mail: labtalk@mt.com
TA-Kundenkurse
und Seminare DSC-Workshop 09. + 10.03.2021 Giessen
TGA-Workshop 23. + 24.03.2021 Giessen
DSC-Workshop 05. + 06.10.2021 Giessen
TGA-Workshop 26. + 27.10.2021 Giessen
France Renseignements et inscriptions par: Paul Maillard, Mettler-Toledo S.A., 18-20 Av. de la pépinière, 78222 Viroflay Cedex, Tél: ++33 1 30 97 16 54, e-mail: FormationClients@mt.com
Cours et séminaires
d’Analyse Thermiquee Analyse Mécanique Dynamique – DMA 15 mars 2021 Viroflay (78), France
Analyse thermo gravimétrique – TGA 16 mars 2021 Viroflay (78), France
Logiciel STARe 17 mars 2021 Viroflay (78), France
Calorimétrie différentielle à balayage – DSC Essentiel 18 mars 2021 Viroflay (78), France
Calorimétrie différentielle à balayage avancée – DSC Perfectionnement 19 mars 2021 Viroflay (78), France
Italia Per ulteriori informazioni Vi preghiamo di contattare: Ottavio Lugaresi, Mettler-Toledo S.p.A., Novate Milanese, Tel: ++39 02 33 332 204, Fax: ++39 02 356 2973
Corsi e Seminari e-mail: ottavio.lugaresi@mt.com
di Analisi Termica Corsi per Clienti: DSC base A richiesta, date da definire Milano
Corsi per Clienti: DSC avanzato A richiesta, date da definire Milano
Corsi per Clienti: TGA A richiesta, date da definire Milano
Corsi per Clienti: TMA A richiesta, date da definire Milano
Corsi per Clienti: DMA A richiesta, date da definire Milano
Eventi: Plast 4–7 maggio 2021 Milano
España Para detalles de los cursos y seminarios, por favor contacte con: Francesc Català, Mettler-Toledo S.A.E., Tel: ++34 93 223 76 00, e-mail: francesc.catala@mt.com
Cursos y Semianros
Seminarios de TA: Aplicaciones DSC en polímeros 23 de marzo de 2021 On-line
Seminarios de TA: Seminarios de TA Aplicaciones DSC en Farma 15 de junio de 2021 On-line
Seminarios de TA: Aplicaciones de la técnica TGA 14 de septiembre de 2021 On-line
Seminarios de TA: Aplicaciones de las técnicas TMA y DMA 23 de noviembre de 2021 On-line
Congresos: GEP-SLAP 2020-21 (www.gep-slap2020.eu) 9–13 de mayo de 2021 San Sebastián
Brasil Para maiores informações sobre os seminários, por gentileza entre em contato com: Mr. Pierre Delmorio, Tel: +55 11 98709-4688, e-mail: pierre.delmorio@mt.com
Seminários e
workshops de TA Análise Térmica Aplicada a Indústria Química Março, 2021 São Paulo
Introdução da Análise Térmica e suas principais aplicações Maio, 2021 São Paulo
Análise Térmica aplicada à Indústria Alimentícia Agosto, 2021 São Paulo
India For details of training courses and seminars, please contact: Pradeep Palshikar, Mettler-Toledo India Private Limited, Amar Hill, Saki Vihar Road, Powai, Mumbai 400 072
TA Customer Courses Tel: +91-22-28031 111 / 42910380 / 28031 380, Fax: +91-22-2857 5071, e-mail: pradeep.palshikar@mt.com
and Seminars Introduction to MT Thermal Analyzers / Focused Segments: All Each Month in 2021 Not specified
Thermal Analysis Tips and Hints / Focused Segments: All February 2021 Not specified
Innovative Applications of Thermal Analyzers to Pharmaceutical Industry / Focused Segments: Pharma May 2021 Not specified
Choosing the right crucibles for DSC and TGA analysis / Focused Segments: All August 2021 Not specified
Applications of Thermal Analyzers to Polymers / Focused Segments: MPE, Chemical (Polymer and Plastics) November 2021 Not specified
Malaysia For details of training courses and seminars, please contact: Ms. Lee Ching Huey, Mettler-Toledo (M) Sdn. Bhd, Unit 1-01, Lot 8, Jalan Astaka U8/84, Seksyen U8,
TA Customer Courses Bukit Jelutong, Shah Alam 40150 Selangor, Malaysia, Tel: (+603)-78445888, Fax: (+603)-78458773, e-mail: mtmymarketing@mt.com
and Seminars Good Thermal Analysis Practice – DSC (Central) March 16, 2021 MTMY Shah Alam
Good Thermal Analysis Practice – TGA (Central) June 15, 2021 MTMY Shah Alam
Good Thermal Analysis Practice – DSC and TGA (Northern) April 6, 2021 MTMY Penang
Good Thermal Analysis Practice – DSC and TGA (Southern) August 23, 2021 MTMY Johor
Singapore For details of training courses and seminars, please contact: Lim An Loon, Mettler-Toledo (S) Pte Ltd., 28 Ayer Rajah Crescent, #05 – 01, Singapore 139959
TA Customer Courses Tel: +65 68900011, Fax: +65-68900012, e-mail: gina.chan@mt.com
and Seminars TA School July 2021 MTSG
Thailand For details of training courses and seminars, please contact: Saowanij Kulthiwalert, Mettler-Toledo (Thailand) Limited, Tel: +66-2-7230338, Fax: +66-2-7196479
TA Customer Courses e-mail: saowanij.kulthiwalert@mt.com
and Seminars Advanced polymer applications October 4–8, 2021 Rayong and Center
Performance QC/Library software/LabX October 4–8, 2021 Rayong and Center
TGA – Evolved gas analysis October 4–8, 2021 Rayong and Center
Increased productivity by automation (Robot) October 4–8, 2021 Rayong and Center
Promote data integrity via TA products October 4–8, 2021 Rayong and Center
Flash DSC for Electronics segment October 4–8, 2021 Rayong and Center
Korea For details of training courses, workshop and seminars, please contact: Inyoung Kim at Mettler-Toledo Korea, Tel: ++82 2 3498 3535, e-mail: inyoung.kim@mt.com, www.mt.com
TA Customer Courses
and Seminars 열분석 Application Webinar 2021년 3월 온라인
열분석 School 2021년 5월 온라인 또는 오프라인
열분석 Application Webinar 2021년 6월 온라인
열분석 Application Webinar 2021년 9월 온라인
열분석 School 2021년 11월 온라인 또는 오프라인
열분석 Application Webinar 2021년 12월 온라인
METTLER TOLEDO UserCom 52 27
China For details of training courses and seminars, please contact: Nicole Dong at Mettler-Toledo Instruments (Shanghai) Co. Ltd.
TA Customer Tel: ++86 21 6485 0435, Fax: ++86 21 6485 3351, e-mail: huicong.dong@mt.com
Courses and TA Course (热分析用户培训) 2021/01/15–2021/01/16 Live broadcast
Seminars
TA Course (热分析用户培训) 2021/03/17–2021/03/20 Beijing + Live broadcast
TA Course (热分析用户培训) 2021/04/14–2021/04/17 Guangzhou+ Live broadcast
TA Course (热分析用户培训) 2021/05/14–2021/05/15 Chengdu + Live broadcast
TA Course (热分析用户培训) 2021/06/16–2021/06/19 Shanghai + Live broadcast
TA Course (热分析用户培训) 2021/09/15–2021/09/18 Beijing+ Live broadcast
TA Course (热分析用户培训) 2021/10/13–2021/10/16 Guangzhou+ Live broadcast
TA Course (热分析用户培训) 2021/11/17–2021/11/20 Shanghai+ Live broadcast
TA Course (热分析高级用户培训) 2021/12/17–2021/12/18 Shanghai
TA Webinar (在线研讨会) 2021/01/20 Live broadcast
TA Webinar (在线研讨会) 2021/03/10 Live broadcast
TA Webinar (在线研讨会) 2021/04/07 Live broadcast
TA Webinar (在线研讨会) 2021/05/21 Live broadcast
TA Webinar (在线研讨会) 2021/06/22 Live broadcast
TA Webinar (在线研讨会) 2021/07/16 Live broadcast
TA Webinar (在线研讨会) 2021/07/23 Live broadcast
TA Webinar (在线研讨会) 2021/09/22 Live broadcast
TA Webinar (在线研讨会) 2021/10/20 Live broadcast
TA Webinar (在线研讨会) 2021/11/24 Live broadcast
For further information regarding meetings, products or applications, please contact your local
METTLER TOLEDO representative and visit our Website www.mt.com
Editorial Team
Dr. A. Hammer Dr. D. P. May Dr. R. Riesen Dr. J. Schawe Dr. V. Chaudhary Dr. S. Domann S. Dandekar Dr. T. Oberholzer Dr. G. Raihani U. Jörimann
Chemist Chemist Chem. Engineer Physicist Polymer Engineer Journ. and Chem. Physical Chemist Biochemist Biologist Electr. Engineer
N. Fedelich Dr. E. Hempel Dr. M. Wagner Dr. M. Schubnell Dr. A. Bach Dr. S. Kakade N. Jing Dr. T. Dennenwaldt Dr. M. Reber Dr. J. Chee
Chem. Engineer Physicist Chemist Physicist Chemist Chemist Chemist Chemist Chemist Chemist