Anaesthetic Gas Scavenging System HTM02-01/HTM2022/ISO7396-2
Anaesthetic Gas Scavenging System HTM02-01/HTM2022/ISO7396-2
AGS SYSTEMS
24/08/16
Issue 9
Page 1
Anaesthetic Gas Scavenging System power is available to the 24V control and indication
The Anaesthetic Gas Scavenging (AGS) System shall circuit. A ‘HAND/OFF/AUTO’ switch shall be provided
comply with HTM 02-01, HTM2022 and either EN ISO to control operation of the pump, running the pump
7396-2 or BS 6834. The AGS system shall be a dedicat- continuously when selected to ‘HAND’. When selected
ed, specifically designed active extraction and dispos- to ‘AUTO’, control of the pump shall be passed to the
al system for waste anaesthetic gas. It shall provide a remote start switch panels. Operation of any of the re-
maximum flow rate of 80 l/min (EN ISO 7396-2) or 130 l/ mote start switches shall activate the pump. The pump
min (BS 6834) with a 1 kPa resistance to flow, and a min- shall continue to run until all remote switches are se-
imum of 50 l/min (EN ISO 7396-2) or 80 l/min (BS 6834) lected ‘OFF’.
with a 2 kPa (EN ISO 7396-2) or 4 kPa (BS 6834) resist-
ance to flow at each terminal unit, irrespective of the The starter/isolator panel shall incorporate a thermal
number of terminal units in use. The AGS system shall protection overload device. The thermal protection
use dedicated radial blowers in a simplex or duplex overload device shall also monitor the electrical pow-
configuration. The AGS pump assemblies shall be skid er supply and phase input. In the event of a fault, the
mounted and included on the skid shall be the simplex overload device shall break the circuit to the pump, pre-
or duplex pump(s), motor control unit(s) with starter/ venting operation until the system is manually re-set.
isolator, moisture drain flask and flexible connector(s) Operation of the overload device shall also break the
to connect the plant to the pipeline. Each pump shall circuit to the remote start switch panels, extinguishing
include an electric motor and directly coupled impel- the green running indicator.
ler assembly. Impeller bearings in the pump(s) shall not
require lubrication. The pump(s) shall be air cooled and
rated for continuous operation. Simplex starter/isolator panels c/w alarm pressure
switch and duplex units incorporate line pressure
Vacuum/Flow Regulating Valve switch. This line pressure switch monitors vacuum lev-
A vacuum/flow regulating valve shall be provided and els and provides an additional control of the remote
positioned at the pump, comprised of a spring-load- start switch and starter/isolator panel green ‘RUNNING’
ed plate valve and inlet silencer. The valve should be indicators. The pressure switch shall also include a digi-
changeable with the pipeline inlet in order to provide tal display providing an accurate readout of the vacuum
flexibility on site. The plate shall control air ingress into level in the pipeline in order to assist with installation/
the pipeline system, thereby controlling the vacuum commissioning and annual
level within. re-commissioning.
An optional air inlet filter shall be available should the Simplex installations shall use remote start switches
air quality be poor/dusty offering further protection that include a red ‘PLANT EMERGENCY’ indicator. This
against dirt ingress into the pump. indicator shall illuminate on all remote start switch pan-
els if the vacuum level falls below the pressure switch
Additional in line vacuum/flow regulating valves may set point level when the pump has been called, or if the
be installed if required and shall be determined by the overload trips. The on/off rocker switch shall include a
pipeline designer. The vacuum/flow regulating valve green illuminated surround to indicate ‘mains on’.
shall ensure a maximum vacuum of 200mb below at-
mospheric pressure is not exceeded and shall be fac- Duplex installations shall use remote start switches
tory preset at 150mb. that include an amber ‘PLANT FAULT’ indicator. This
shall illuminate, if either pump is set to ‘HAND’, or if one
of the overloads trip. A red ‘PLANT EMERGENCY’ indi-
Control System cator shall also be provided and shall illuminate on all
Each motor control panel shall incorporate an emer-
remote start switch panels if the vacuum level falls be-
gency panel isolation switch facility, which controls all
low the pressure switch set point level when the pump
electrical power to the exhauster unit, remote start
has been called. The on/off rocker switch shall include a
switch panels and system indication lights. All control
green illuminated surround to indicate ‘mains on’.
and status indication circuitry shall be limited to 24V
a.c. A green ‘POWER ON’ indicator shall be fitted to the
starter/isolator panel, and shall illuminate whenever
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005208
24/08/16
Issue 9
Page 2
Terminal Units
Terminal unit shall be provided with an adjustable ori-
fice to allow balancing of the terminal unit flows during
commissioning. Venturi style terminal units are not ac-
ceptable. Terminal units shall not be connected to the
medical vacuum system.
CE Marking
number
Options
Part
The standard range of BeaconMedæs Anaesthetic Gas
Scavenging System are ‘CE’ marked under the Medical
Devices Directive 93/42/EEC with approval from noti-
fied body no. 0088 (Lloyd’s Register Quality Assurance). Inlet Filter & Housing - Factory 0000 0038 19
Under this directive, the specified products are classi- Inlet Filter & Housing - Retrofit 8092 1303 87
fied as Class IIb Medical Devices.
pump @ 1m dB(A)
Supply frequency
Connection (mm)
supply/pump (A)
Length (mm)
Part number
Height (mm)
Width (mm)
Weight (kg)
Voltage (V)
Phase ~
rent (A)
Format
(mm2)
Model
(Hz)
AGS-520-D/3 4153 1254 00 400 3 1.8 9 2.5 16 900 525 1160 96 0.75 60 54
Duplex
AGS-1560-D/3 4153 1255 00 400 3 3.3 23 2.5 16 900 525 1160 121 1.5 64 54
AGS-2860-D/3 4153 1256 00 400 3 7.2 48 4 16 900 525 1160 139 3 69 54
50 Hz
AGS-520-S/3 4153 1251 00 400 3 1.8 9 2.5 16 525 525 1160 56 0.75 60 54
AGS-1560-S/3 4153 1252 00 400 3 3.3 23 2.5 16 525 525 1160 68 1.5 64 54
AGS-2860-S/3 4153 1253 00 400 3 7.2 48 2.5 16 525 525 1160 77 3 69 54
AGS-650-D/3 4153 1261 00 380 3 1.8 10 2.5 16 900 525 1160 96 0.9 64 54
Duplex
AGS-2210-D/3 4153 1262 00 380 3 3.3 23 2.5 16 900 525 1160 121 1.8 69 54
AGS-3770-D/3 4153 1263 00 380 3 7.2 48 4 16 900 525 1160 139 3.6 70 54
60 Hz
AGS-650-S/1 4153 1257 00 220 1 5.5 18 2.5 16 525 525 1160 56 0.9 64 54
Simplex
AGS-650-S/3 4153 1258 00 380 3 1.8 10 2.5 16 525 525 1160 56 0.9 64 54
AGS-2210-S/3 4153 1259 00 380 3 3.3 23 2.5 16 525 525 1160 68 1.8 69 54
AGS-3770-S/3 4153 1260 00 380 3 7.2 48 2.5 16 525 525 1160 77 3.6 70 54
Notes…
1. Noise levels are averages, measured in accordance with ISO 517:1996 Free Air Aspiration and ISO 2151 Mean Sound Level
2. Service connections are copper to BS EN 13348 and relate to the external pipe diameter
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005208
24/08/16
Issue 9
Page 3
AGS Pump Assembly Details - Models Including Additional BMS Alarm Contacts
Connection (mm)
supply/pump (A)
Length (mm)
Part number
Height (mm)
Width (mm)
Weight (kg)
Voltage (V)
1m dB(A)
Phase ~
Format
Model
BMS 00/BMS
AGS-2210-S/3 4153 1259
380 3 3.3 23 2.5 16 525 525 1160 68 1.8 69 54
- BMS 00/BMS
AGS-3770-S/3 4153 1260
380 3 7.2 48 2.5 16 525 525 1160 77 3.6 70 54
- BMS 00/BMS
number
Options
Part
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005208
24/08/16
Issue 9
Page 4
Outlet (2x)
7 3 4 A
1 1
B
1150
2 x Cable entrance
965
for power supply
5 5
6
C
900 400
Installation instruction 2.
The AGS air inlet can be re positioned by swapping the copper stub (item 7) and
Inlet regulating valve (item 2) as indicated by Alt.1 or Alt.2.
Alt.1 Use Permabond MH052 on joints.
Inlet
Alt.2
7 E
requirements of the List of Prohibited Substances
Notes.
All materials supplied are in compliance with the
Ref Description
1 Flexible hose
2 Flow regulating valve
copied, used for manufacturing or communicated to any other person or company.
This document is our property and shall not without our permission be altered,
3 Pressure switch
4 Seal
Tolerances, if not indicated, according to:
5 Non return valve
6 Drain flask ATLAS COPCO STANDARD CLASS
7 Copper stub pipes 54mm dia. Name BM AGS Installation diagram Confidentiality Class
acc. to 1102 K
Material
1820 Confidential
Treatment INV G
CONFIDENTIAL:
6070002446
0
Ed
Posi-
Modified from Date Intr./Appd. Parent 3D model Ed . Version 3D - Des checked. Prod checked. Approved. Date
tion
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005208
24/08/16
Issue 9
Page 5
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005208
the right is reserved to change the specification of the 24/08/16
items described herein at any time. Please contact us Issue 9
for further information and up to date specifications.
Page 6
Anaesthetic Gas Scavenging System Simplified Schematic
Multiple Suite System with Simplex Blower
Typical AGS System Schematic - Simplex
Exhaust
Pressure
Switch
C
posal S posal S
Side Valve
Dis Dis
ys
ys
S
S
AGS
AGS
te
te
m
m
Channel
dæs
dæs
88
88
Me
Me
00
00
on
on
ac ac
Be Be
Blower
To Control
Terminal Units
Panel
Accessories
Terminal
number
number
Note: Not all system components required are shown
Units
Part
Part
AGS Gem Revolve 1st fix
8102340218
AGSS Exhaust Silencer 1.5” 1824664
- Mod/Wall
AGSS Exhaust Silencer 2” 1824665
Gem Revolve Curved Bedhead 1st fix
assembly
8102340238 AGSS Exhaust Silencer 2.5” 1824666
Gem Revolve Mod/Wall 1st fix assembly AGSS Exhaust Silencer 3” 1824667
- Extended Rear Entry
8102340988
AGSS Exhaust Silencer 4” 1824668
Gem 2nd fix assembly - BS 8102340208
Anti-microbial Trim Plate
c/w Bezel and Screws
2006149 AGS Terminal Unit Adjustment Tool 2004569
BeaconMedæs
Telford Crescent,
Staveley, Derbyshire
S43 3PF, England
Tel: +44 (0) 1246 474 242
Fax: +44 (0) 1246 472 982
www.beaconmedaes.com
E-mail: gbn.info@beaconmedaes.com 0088