Professional Documents
Culture Documents
2545 Med. Phys. 30 „9…, September 2003 0094-2405Õ2003Õ30„9…Õ2545Õ8Õ$20.00 © 2003 Am. Assoc. Phys. Med. 2545
2546 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2546
FIG. 2. MLC leaf end leakage/transmission test pattern. This image shows
the resultant dose pattern for all 9 abutment locations for all 27 1 cm leaf
pairs.
C. In-plane penumbra
Two basic types of penumbra 共80% to 20%兲 were studied,
in-plane and cross-plane penumbra. Fig. 3共a兲 shows one of
seven field shapes used to measure in-plane penumbra, FIG. 3. MLC penumbra tests. 共a兲 MLC field shape in beams-eye-view 共1 of
7兲 used to assess in-plane penumbra. 共b兲 MLC field shape in beams-eye
where a single leaf on the X2 leaf bank was fully retracted view 共1 of 11兲 used to assess cross-plane penumbra.
共in this example, leaf 10兲. In this figure, the in-plane penum-
bra is dependent on the Y2 side of leaf 9 共tongue兲 and the Y1
side of leaf 11 共groove兲. The penumbra is dependent on the
D. Cross-plane penumbra
sides of the leaves only when the Y jaws are not abutting the
open port, as is often the case with segmental MLC IMRT.6 Figure 3共b兲 shows the beams-eye-view of one of eleven
A new film was exposed for each of seven leaves retracted field shapes used to measure cross-plane penumbra 共leaf-
共2, 6, 10, 15, 20, 24, and 28兲. A dose profile perpendicular to end兲, where leaves are separated by 1 cm. In this case, the 1
the direction of leaf travel 共vertical兲 passing through the cen- cm opening is centered at 10 cm off the central axis, toward
tral axis was acquired for each film to measure the in-plane the X1 jaw. A new film was exposed for each of eleven field
variation of in-plane penumbra. In addition, the cross-plane shapes, where the center of the exposed port moved from 10
variation of in-plane penumbra was analyzed. Eleven total cm toward the X1 jaw to 10 cm toward the X2 jaw in 2 cm
vertical dose profiles were acquired on each film 共2 cm spac- increments. A dose profile parallel to the direction of leaf
ing, from 9.5 cm toward X1 to 10.5 cm toward X2兲. travel 共horizontal兲 passing through the central axis was ac-
TABLE I. Interleaf leakage variation in the in-plane direction. Each column shows the average and standard deviation of leakage in the in-plane direction for
a specified cross-plane location.
9.5 7.5 5.5 3.5 1.5 ⫺0.5 ⫺2.5 ⫺4.5 ⫺6.5 ⫺8.5 ⫺10.5
Average 共%兲 1.11 1.14 1.14 1.12 1.11 1.10 1.09 1.06 1.03 0.99 0.96
Standard dev 共%兲 0.12 0.13 0.13 0.13 0.14 0.14 0.14 0.14 0.14 0.14 0.13
TABLE II. Interleaf leakage variation in the cross-plane direction. Each column shows the average and standard deviation of leakage in the cross-plane
direction between a specific leaf-pair.
2 3 4 5 6 7 8 9 10
Average 共%兲 0.57 0.96 0.96 1.01 1.02 1.13 1.08 1.05 1.12
Standard dev 共%兲 0.06 0.05 0.05 0.07 0.06 0.07 0.07 0.04 0.07
11 12 13 14 15 16 17 18 19
Average 共%兲 1.09 1.13 1.11 1.24 1.42 1.10 1.10 1.12 1.13
Standard dev 共%兲 0.05 0.06 0.04 0.11 0.06 0.07 0.04 0.07 0.07
20 21 22 23 24 25 26 27 28
Average 共%兲 1.11 1.10 1.05 1.10 1.13 1.04 1.05 1.02 1.12
Standard dev 共%兲 0.05 0.05 0.04 0.06 0.06 0.07 0.08 0.07 0.12
lines around the mean value of 1.9% demonstrates minor This graph shows very little variation of the in-plane penum-
cross-plane variation. There was no difference between in- bra caused by the edge of a leaf, either in the in-plane (p
plane and cross-plane leaf-end transmission (p⫽0.4). The ⫽0.4) or cross-plane direction (p⫽0.3).
leaf-end transmission at 10 cm off the central axis for all leaf Similarly, the cross-plane penumbra caused by the end of
pairs is approximately 0.5% less than the mean, while the a leaf 关as shown in Fig. 3共b兲兴 is provided in Fig. 6共c兲. Each
central leaves are 0.5% greater than the mean for the central line represents a particular leaf, and the data points on the
leaf pairs. Figure 5共b兲 shows the frequency distribution of line indicate the cross-plane position of the penumbra mea-
leaf-end transmission, which is nearly normally distributed surement. These results show less than 0.03 cm from their
about the mean with the 99% confidence interval calculated mean value, regardless of the location of the field edge. Al-
to be 1.9%–2.0%. Seventy percent of the measured values though these figures only show two field edges 共Y2 and X2兲,
are within 0.2% of the mean, 80% are within 0.4%, and 98% the results for all four edges are shown in Fig. 6共d兲. The
are within 0.6%. Since the change in transmission for differ- 80%–20% penumbra for these 1-cm-wide fields was mea-
ent abutment positions is not symmetric about the central sured to have a weighted mean of 0.37 cm⫾0.01 cm.
axis, the variation seen is likely most dependent on MLC leaf The range in penumbra measured was between 0.31 and 0.43
position calibration accuracy. mm for all cases. The mean cross-plane penumbra was
statistically different (p⬍0.001) for the X1 bank
(0.358⫾0.0007 cm) compared to the X2 bank (0.369
C. Penumbra ⫾0.0007 cm).
Figure 6共a兲 shows a vertical dose profile 共in-plane兲
through each of the retracted leaves 关as shown in Fig. 3共a兲兴, IV. DISCUSSION
where the tongue and the groove of each opening generate an The greatest significance of this work may be found in
independent penumbra. This graph demonstrates the con- establishing the dosimetric characteristics of the MLC in or-
stancy of projected leaf width and penumbra for each leaf, der to allow the physicist to review the treatment planning
regardless of its off-axis distance in the in-plane direction. system’s accuracy in describing these effects. Nelson et al.21
That result is further shown in Fig. 6共b兲, where the in-plane of the RPC showed that all IMRT treatment planning sys-
penumbra was tested for cross-plane dependence. Each line tems they analyzed struggled to accurately model the dose in
represents a single leaf, and the data points on the line indi- organs at risk adjacent to planning target volumes, regions
cate the cross-plane position of the penumbra measurement. presumably shielded by the MLC leaves during the majority
TABLE III. Intraleaf leakage variation in the in-plane direction. Each column shows the average and standard deviation of leakage in the in-plane direction for
a specified cross-plane location.
9.5 7.5 5.5 3.5 1.5 ⫺0.5 ⫺2.5 ⫺4.5 ⫺6.5 ⫺8.5 ⫺10.5
Average 共%兲 0.84 0.85 0.85 0.84 0.83 0.82 0.80 0.78 0.75 0.72 0.70
Standard dev 共%兲 0.07 0.07 0.08 0.08 0.08 0.08 0.08 0.08 0.09 0.09 0.08
TABLE IV. Intraleaf leakage variation in the cross plane direction. Each column shows the average and standard deviation of leakage in the cross-plane
direction through a specific leaf.
X2 bank leaf
2 3 4 5 6 7 8 9 10
Average 共%兲 0.53 0.69 0.72 0.76 0.79 0.81 0.80 0.82 0.85
Standard dev 共%兲 0.07 0.04 0.05 0.06 0.06 0.05 0.05 0.05 0.05
X2 bank leaf
11 12 13 14 15 16 17 18 19
Average 共%兲 0.85 0.85 0.86 0.90 0.84 0.86 0.85 0.86 0.86
Standard dev 共%兲 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05
X2 bank leaf
20 21 22 23 24 25 26 27 28
Average 共%兲 0.85 0.82 0.82 0.82 0.80 0.78 0.74 0.74 0.65
Standard dev 共%兲 0.05 0.05 0.05 0.05 0.06 0.06 0.06 0.06 0.05
of IMRT treatments. Furthermore, these regions are fre- Interleaf leakage, intraleaf leakage, and leaf-end penum-
quently within a high dose-gradient region where accurate bra have been previously described for fields of 10⫻10 cm
modeling of the dose profile from 5–10 mm from the field and larger for the MLC collimators found on linear accelera-
edge may be critical. tors from Elekta 共Huq et al.9兲, Siemens 共Das et al.14兲, and
Varian 共Galvin et al.22兲. These measurements tend to be
through the central axis of the beam and for larger field sizes.
Das et al.14 showed the average intraleaf leakage to be 0.92
⫾0.12% and the average interleaf leakage to be 1.17
⫾0.13% for the same MLC design studied here. Our results
are consistent with these, in both the in-plane and cross-
plane direction, regardless of the measurement position with
respect to the central axis. Also for comparison, the transmis-
sion through leaves reported for MLCs on General Electric,
Philips, and Varian linear accelerators were in the order of
1.2%, 5%, and 4%, respectively for a 6 MV beam.9,14,22
The leaf-end transmission that we have measured has not
been reported previously in the literature for the double-
focused MLC design provided by Siemens. Klein et al.23 re-
ported this value for a single-focused MLC with rounded leaf
edges as large as 28% for 18 MV photons on the central axis,
reducing to 12% for the furthest points measured off-axis.
The values presented here for the double-focused MLC were
substantially more consistent, as would be expected. The
magnitude of the leaf-end transmission will be greatly de-
pendent on the leaf position calibration accuracy of the
MLC. At our institution, we utilize a leaf position calibration
procedure24 that produces 95.5% of the leaf-positions to be
within 1 mm and a mean within 0.4 mm.
The double-focused MLC also demonstrated consistency
in penumbra. Das et al.14 studied the change in penumbra for
a 10⫻10 cm field at several off-axis positions of the field
center. Specifically, they considered the penumbra along the
center leaf, traveling from 5 cm over-travel to 10 cm open,
and reported ⭐0.5% variation in the penumbra. Our results
show the penumbra is consistent both for the leaf-end as well
as across the side-edge of the leaves. The magnitude of the
FIG. 5. Leaf-end transmission 共a兲 through abutted leaf pairs, for all 27 1 cm
leaf pairs. Each line represents 1 of 9 different abutment locations. 共b兲 fre- penumbra measured for fields 1 cm wide 共0.37 cm penum-
quency distribution of leaf-end transmission. bra兲 was significantly less than the values reported by Das
FIG. 6. In-plane and cross-plane penumbra created by the side and end of the leaf, respectively. 共a兲 An in-plane dose profile through each of 7 different
retracted leaves. 共b兲 In-plane penumbra measured at 11 different cross-plane locations. Each line represents 1 of 7 retracted leaves. 共c兲 Cross-plane penumbra
measured at 11 different cross-plane locations. Each line represents 1 of 27 leaves on the X2 leaf bank, each having a different in-plane position. 共d兲 Combined
penumbra results for all four field edges measured for all locations studied.
acteristics of a commercial multileaf collimator,’’ Med. Phys. 22, 241– modulated radiation therapy delivered by a static multileaf collimator,’’
247 共1995兲. Med. Phys. 29, 1687–1692 共2002兲.
10 17
J. R. Palta, D. K. Yeung, and V. Frouhar, ‘‘Dosimetric considerations for I. Chetty and P. Charland, ‘‘Dosimetric characterization of Kodak ex-
a multileaf collimator system,’’ Med. Phys. 23, 1219–1224 共1996兲. tended dose range film in megavoltage photon beams,’’ Med. Phys. 29,
11
S. L. Meeks, F. J. Bova, S. Kim, W. A. Tome, J. M. Buatti, and W. A. 1207 共2002兲.
18
Friedman, ‘‘Dosimetric characteristics of a double-focused miniature M. Bieda, A. Sarkar, and I. Das, ‘‘Accuracy of Kodak’s EDR2 film in
multileaf collimator,’’ Med. Phys. 26, 729–733 共1999兲. measuring clinical dosimetry data,’’ Med. Phys. 29, 1264 共2002兲.
19
12
M. Stasi, B. Baiotto, F. Palamara, P. Gabriele, and G. Scielzo, ‘‘Dosim- N. Childress, L. Dong, and I. Rosen, ‘‘Rapid radiographic film calibration
etric characterization of a multileaf collimator,’’ Radiol. Med. 共Torino兲 for IMRT verification using automated MLC Fields,’’ Med. Phys. 29,
1366 共2002兲.
97, 382–388 共1999兲. 20
13 J. Bayouth, S. Zhou, and S. Morrill, ‘‘Analysis of variation in H&D curve
M. Pasquino, V. Casanova Borca, and S. Tofani, ‘‘Physical-dosimetric
creation for EDR2 film dosimetry,’’ Med. Phys. 30, 1450 共2003兲.
characterization of a multileaf collimator system for clinical implemen- 21
A. Nelson, P. Balter, W. Hanson, and G. Ibbott, ‘‘Experience with an
tation in conformational radiotherapy,’’ Radiol. Med. 共Torino兲 101, 187– IMRT head and neck QA phantom,’’ Med. Phys. 29, 1366 共2002兲.
192 共2001兲. 22
J. M. Galvin, A. R. Smith, and B. Lally, ‘‘Characterization of a multi-leaf
14
I. J. Das, G. E. Desobry, S. W. McNeeley, E. C. Cheng, and T. E. Schulth- collimator system,’’ Int. J. Radiat. Oncol., Biol., Phys. 25, 181–192
eiss, ‘‘Beam characteristics of a retrofitted double-focused multileaf col- 共1993兲.
limator,’’ Med. Phys. 25, 1676 –1684 共1998兲. 23
E. E. Klein, W. B. Harms, D. A. Low, V. Willcut, and J. A. Purdy, ‘‘Clini-
15
R. Mitra, J. Bayouth, and I. Das, ‘‘Startup characteristics and dosimetry cal implementation of a commercial multileaf collimator: Dosimetry, net-
of small monitor unit segments in step and shoot IMRT delivery for two working, simulation, and quality assurance,’’ Int. J. Radiat. Oncol., Biol.,
types of digital linear accelerators,’’ Med. Phys. 28, 1202 共2001兲. Phys. 33, 1195–1208 共1995兲.
16 24
X. R. Zhu, P. A. Jursinic, D. F. Grimm, F. Lopez, J. J. Rownd, and M. T. J. E. Bayouth, D. Wendt, and S. M. Morrill, ‘‘MLC quality assurance
Gillin, ‘‘Evaluation of Kodak EDR2 film for dose verification of intensity techniques for IMRT applications,’’ Med. Phys. 30, 743–750 共2003兲.