You are on page 1of 8

MLC dosimetric characteristics for small field and IMRT applications

J. E. Bayoutha) and S. M. Morrill


University of Texas Medical Branch, Department of Radiation Oncology, Galveston, Texas 77555
共Received 17 October 2002; accepted for publication 29 June 2003; published 25 August 2003兲
The objective of this work was to measure the performance characteristics of a double-focus
multileaf collimator 共MLC兲 for intensity modulated radiation therapy 共IMRT兲, specifically the
variation in penumbra and leakage for narrow fields as a function of field position over a 20⫻27 cm
space available for segmented MLC IMRT. Measurements were made with 6 MV x rays through a
MLC containing 29 leaf pairs 共27 pairs of 1 cm width兲, and EDR2 film at 10 cm depth in solid
water at 100 cm SAD. Films were digitized with 0.17 mm resolution and converted to dose.
Interleaf and intraleaf transmission were measured along 11 vertical profile locations. Leaf-end
transmission was measured along horizontal profiles for each of 9 different leaf abutments, travel-
ing over a 20 cm range. In-plane penumbra measurements were made through a single leaf re-
tracted, for 7 different leaves. Cross-plane penumbra 共leaf-end兲 measurements were made for all 27
leaf pairs, where the 1 cm field width was placed in 11 different off-axis positions 共20 cm range兲.
Interleaf leakage 共range 1.0%–1.5%兲, intraleaf transmission 共range 0.6%–0.8%兲, and leaf-end
transmission 共range 0.8%–2.7%兲 were consistent for all leaf pairs at a given abutment position. The
penumbra for these 1-cm-wide fields was measured to be 0.36 cm⫾0.03 cm for 99% of the
measurements. In conclusion, the penumbra and leakage of the double-focus MLC were remarkably
consistent for the range of leaf positions studied, producing dosimetric characteristics that are well
suited for IMRT segments where opposing leaf pairs are often separated by 10 mm or less.
© 2003 American Association of Physicists in Medicine. 关DOI: 10.1118/1.1603743兴

Key words: multileaf collimator, double-focused, leakage, transmission, penumbra

I. INTRODUCTION are created using a MLC. Although the general dosimetric


characteristics of MLCs have been described by others8 –14
Three-dimensional conformal radiotherapy 共3DCRT兲 has be- the small field dosimetric characteristics of the MLC are not
come the standard of care for many types of cancers receiv- well known. Furthermore, their behavior for the range of
ing radiation therapy. The Radiation Therapy Oncology off-axis positions often utilized in IMRT is not well known.
Group now relies on this delivery technique in protocols for For example, a given MLC leaf has penumbra off the edge of
the treatment of cancers of the prostate, lung, glioblastoma, the leaf and the end of the leaf. These two sources of pen-
and oropharynx. Optimal dose distributions created using umbra may be different, and each penumbra type potentially
3DCRT produces isodose surfaces that conform closely to changes both in the in-plane and cross-plane direction. Pre-
the planning target volumes, and are often designed to ac- vious studies have not provided data on the magnitude of
complish conformal avoidance, i.e., creating steep dose- these changes, and treatment planning systems typically ig-
gradients between the planning target volumes and nearby nore these effects. The objective of this work was to measure
critical structures. The utility of steep dose-gradients has the MLC performance characteristics for IMRT, specifically
long been exploited in stereotactic radiosurgery; Podgorsak the variation in penumbra and leakage/transmission for nar-
et al.1 studied the dose-gradient as a metric when comparing row fields as a function of field position over the allowable
the dose fall-offs outside the target volume for various radio- position space for IMRT delivery.
surgical techniques. Further efforts in conformal avoidance
have led to the inclusion of ‘‘island blocks’’ within the treat-
ment portal, as described by Eisbrauch et al.,2 Frass et al.,3 II. MATERIALS AND METHODS
Xiao et al.,4 and Verhey.5 This treatment planning goal is a Exposures were made with a Siemens 共Siemens Medical
primary objective for a subset of 3DCRT, intensity modu- Systems, Concord, CA兲 Primart linear accelerator, containing
lated radiation therapy 共IMRT兲.6 29 leaf pairs of which the central 27 leaf pairs are 1 cm wide
One common technique of creating IMRT beams is to centered on the central axis. A complete description of the
utilize a segmentation algorithm that converts the ideal flu- design of this double-focused multileaf collimator was pro-
ence map into deliverable beam segments using a multileaf vided by Das et al.14 Each leaf is capable of traveling 10 cm
collimator 共MLC兲.7 During the segmentation process the leaf beyond the central axis, so when considering this limitation
ends of the MLC, which are used to create the resultant for opposing leaf banks, this design provides 20 cm of al-
segment shapes, may be separated by as little as 5–10 mm lowable position space in the direction of leaf travel and 27
for the delivery of IMRT. One of the many dosimetric chal- cm perpendicular to leaf travel. For this study the 6.5-cm-
lenges for treatment planning systems is to model the dose wide leaf pairs 共1 and 29兲 are not considered. The Primart is
distributions created by these segments accurately when they a single energy linear accelerator that produces 6 MV x rays,

2545 Med. Phys. 30 „9…, September 2003 0094-2405Õ2003Õ30„9…Õ2545Õ8Õ$20.00 © 2003 Am. Assoc. Phys. Med. 2545
2546 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2546

utilizing a straight waveguide with rotating target that yields


a sharper focal spot. The dosimetric characteristics of this
accelerator are similar to the Siemens Primus accelerator.15
Kodak EDR2 ReadyPak film 共Eastman Kodak Company,
Rochester, NY兲 was used for all radiation dose measure-
ments. This film was chosen over gafchromic film for these
measurements because of its size and cost, and over Kodak
XV film because of its wider range and reduced energy
dependence,16 –19 likely due to a reduction in the amount of
silver halide. The films were placed at 10 cm depth in a solid
water phantom 共30⫻30⫻22 cm兲 and 100 cm SAD. All films
were processed together and a unique H&D curve was cre-
ated for each group of measurements. H&D curves were gen-
erated using a single film calibration method similar to that
presented by Childress et al.,19 where eight nonoverlapping
portals of different dose are exposed on a single film. Al-
though the film depth utilized for the H&D curve was
slightly different 共7.5 cm兲 from the measurement depth for
the remainder of the films 共10 cm兲, EDR2 film is insensitive
to the depth difference.19,20 Each film was digitized with 0.17
mm resolution using the RIT113 film dosimetry system 共Ra-
diological Imaging Technology, Colorado Springs, CO兲 and
converted to dose 共optical density range 0.3–1.6兲. The digital
images were manipulated to reduce noise 共3⫻3 medial filter兲
and remove rotation of the film data with respect to the col-
limator coordinate system.
A. Intra- and Interleaf leakageÕtransmission
Three types of leakage/transmission were measured: inter-
leaf, intraleaf, and leaf-end.7 For all measurements, the
leakage/transmission was reported as the percent of the un-
blocked dose. The percentage of the unblocked dose was
computed as the percentage of dose recorded on the film
divided by the dose that would have been delivered to the
film at the same depth and monitor units from a 10⫻10 cm
field. To measure the interleaf and intraleaf transmission, all
27 1 cm leaf pairs were programmed to fully close, produc-
ing a 0 cm gap between opposing leaves. An example of the
field shape is shown in Fig. 1共a兲. The Y jaws travel in the
in-plane direction, while the X jaws travel in the cross-plane
direction. The Y2 jaw is on the side of the central axis closer
to the electron gun, while the X1 leaf bank is on the left-hand
side in the beams-eye-view. In Fig. 1共a兲 leaf pairs 2 through
28 are abutting each other 10 cm off-axis, while leaf pairs 1
and 29 were opened 3 mm, which allowed the Y1 and Y2
jaw set to 13.6 cm and allow full exposure of the leaves. A
single radiographic film was exposed to 6 MV x rays for
6000 monitor units, which produced the image in Fig. 1共b兲.
Eleven vertical dose profiles were measured across this im- FIG. 1. MLC intra- and interleaf leakage/transmission test pattern. 共a兲 MLC
age, 2 cm apart from one another, perpendicular to the leaf field shape in the beams-eye view. Note the Y-jaws 共not shown兲 are set 1
motion direction. The first profile was 5 mm medial to the mm distal to leaf pairs 2 and 28. 共b兲 Resultant film 共dose兲 from 共a兲.
match of the X1 and X2 leaf bank, for the field shape shown
in Fig. 1共a兲, and the eleventh 10.5 cm from the CAX toward
B. Leaf-end leakageÕtransmission
the X2 leaf bank. A single profile shows the difference in
leakage/transmission for each leaf pair, while multiple verti- Nine sequential field shapes were utilized to assess leaf-
cal profiles across the image in Fig. 1共b兲 from left to right end transmission 共i.e., the transmission of photons through
will show the variation of leakage/transmission in the cross- abutted leaf pairs兲. Ten separate films were exposed: one for
plane direction. each of the nine fields, and a tenth was exposed for all nine

Medical Physics, Vol. 30, No. 9, September 2003


2547 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2547

FIG. 2. MLC leaf end leakage/transmission test pattern. This image shows
the resultant dose pattern for all 9 abutment locations for all 27 1 cm leaf
pairs.

fields on a single film. The shape of each field resembled that


shown in Fig. 1共a兲, where the abutment location was trans-
lated by 2.5 cm steps over the course of the sequence. Each
exposure was for 500 MU; Figure 2 shows the single film
exposed for all nine leaf abutment locations. The leaf-end
transmission was not determined from the image in Fig. 2,
but rather from each film receiving a single exposure.
Twenty-seven horizontal dose profiles were measured, one
through the center of each leaf pair, for each of the nine
abutment locations and the maximum transmission through
each abutted leaf pair was recorded. Thus, for an individual
leaf pair we could analyze the leaf-end transmission and its
variation across the IMRT position space.

C. In-plane penumbra
Two basic types of penumbra 共80% to 20%兲 were studied,
in-plane and cross-plane penumbra. Fig. 3共a兲 shows one of
seven field shapes used to measure in-plane penumbra, FIG. 3. MLC penumbra tests. 共a兲 MLC field shape in beams-eye-view 共1 of
7兲 used to assess in-plane penumbra. 共b兲 MLC field shape in beams-eye
where a single leaf on the X2 leaf bank was fully retracted view 共1 of 11兲 used to assess cross-plane penumbra.
共in this example, leaf 10兲. In this figure, the in-plane penum-
bra is dependent on the Y2 side of leaf 9 共tongue兲 and the Y1
side of leaf 11 共groove兲. The penumbra is dependent on the
D. Cross-plane penumbra
sides of the leaves only when the Y jaws are not abutting the
open port, as is often the case with segmental MLC IMRT.6 Figure 3共b兲 shows the beams-eye-view of one of eleven
A new film was exposed for each of seven leaves retracted field shapes used to measure cross-plane penumbra 共leaf-
共2, 6, 10, 15, 20, 24, and 28兲. A dose profile perpendicular to end兲, where leaves are separated by 1 cm. In this case, the 1
the direction of leaf travel 共vertical兲 passing through the cen- cm opening is centered at 10 cm off the central axis, toward
tral axis was acquired for each film to measure the in-plane the X1 jaw. A new film was exposed for each of eleven field
variation of in-plane penumbra. In addition, the cross-plane shapes, where the center of the exposed port moved from 10
variation of in-plane penumbra was analyzed. Eleven total cm toward the X1 jaw to 10 cm toward the X2 jaw in 2 cm
vertical dose profiles were acquired on each film 共2 cm spac- increments. A dose profile parallel to the direction of leaf
ing, from 9.5 cm toward X1 to 10.5 cm toward X2兲. travel 共horizontal兲 passing through the central axis was ac-

Medical Physics, Vol. 30, No. 9, September 2003


2548 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2548

result from a leaf that is too narrow or not angled properly


along its ‘‘double-focus’’ fan line. Table II shows interleaf
leakage between each leaf pair, obtained by averaging the
data in Fig. 4 for all cross-plane measurements 共i.e., for all
lines兲 for each peak. Each column shows the average and
standard deviation of leakage in the cross-plane direction be-
tween a specific leaf-pair. With respect to Fig. 4, these results
show the variation of inter-leaf leakage for a given peak for
all curves. Here the largest variation is seen between leaves
14 –15 and 28 –29, where their standard deviation is nearly
doubled, as seen qualitatively by visual inspection of Fig.
1共b兲. The leakage through the first and last leaf-pair show
less transmission, possibly due to the reduced transmission
FIG. 4. Inter- and intraleaf leakage across all 27 1 cm leaves, along eleven
different cross-plane locations.
allowed by the proximity of the Y jaw. The variation in the
cross-plane direction is less than that seen in the in-plane
direction (p⫽0.004, two-sample t test兲, demonstrating the
quired for each film to measure the cross-plane variation of double-focus MLC design yields leakage between a leaf and
cross-plane penumbra. In addition, the in-plane variation of its neighbor that is slightly more consistent in the direction
cross-plane penumbra was analyzed. A total of 27 horizontal parallel to the direction of leaf travel than perpendicular.
dose profiles were acquired, one through each leaf pair 共2 The valley of each dose profile is the sum of penumbra
through 28兲, for each of the eleven leaf positions. and intraleaf transmission, which was between 0.6% and
0.9% 共5–95 percentile兲 with a mean of 0.8%⫾0.1%. Table
III. RESULTS III provides intraleaf leakage variation in the in-plane direc-
tion. The standard deviation of leakage in the in-plane direc-
A. Intra- and Interleaf leakageÕtransmission
tion for each cross-plane location varies by less than 0.1%.
Interleaf leakage is shown in Fig. 4, where the peak of Table IV shows the results for intraleaf leakage through each
each dose profile represents the interleaf leakage. Each line leaf at several locations in the cross-plane direction. These
represents a dose profile perpendicular to the direction of leaf results demonstrate the standard deviation of leakage in the
travel, for one of several off-axis 共cross plane兲 positions. cross-plane direction for leaf varies by approximately 0.05%.
Although the general trends of the data are shown in Fig. 4,
detailed analysis of the results requires specific data provided
B. Leaf-end leakageÕtransmission
in Tables I–IV. The mean value and the standard deviation of
interleaf leakage along each profile are listed in Table I. The The effects of leaf-end transmission can be seen qualita-
average interleaf leakage for all 27 peaks was 1.1%⫾0.1%, tively in Fig. 2, where the vertical lines are caused by leaf-
ranging from 0.9% to 1.3% 共5–95 percentile兲. These results end transmission and the horizontal lines are the result of
show that the in-plane variation of interleaf leakage is small; interleaf transmission. Although the two sets of transmission
with a standard deviation of approximately 10% of the mean lines appear similar in exposure, the interleaf transmission
value. These results are also observed graphically, as the occurred for all 9 exposures, where the horizontal lines are
variation from one peak to the next for a given line is small. the sum of leaf-end transmission occurred for one of the 9
A spike in the results is seen at an in-plane distance of 0.5 exposures and intraleaf transmission for the remaining 8 ex-
and 1.5 cm from the central axis. This increase in leakage posures. Quantitative results of the leaf-end transmission
corresponds to the space between leaf pairs 13–14 and 14 – measurements are shown in Figs. 5共a兲 and 5共b兲, as taken
15. Inspection of Fig. 1共b兲 reveals an enhanced interleaf from individual films of leaf abutment. The average leaf-end
leakage dose at this location that increases in magnitude fur- transmission for all leaf pairs at all measurement points was
ther away from the central axis. This may be caused by a 1.9%⫾0.3%, with a range of 1.4%–2.5% 共5–95 percentile兲.
slight variation in the mechanical fabrication of leaf X2.14 Each line in Fig. 5共a兲 represents the transmission values be-
and/or spatially dependent variation in the separation be- tween leaf pairs 2–28 at a single cross-plane location. Minor
tween the leaves allowing additional leakage. Not to be con- variation is seen along a given line, indicating little in-plane
fused with the leaf position calibration, this variation could variation of leaf-end transmission. The grouping of these

TABLE I. Interleaf leakage variation in the in-plane direction. Each column shows the average and standard deviation of leakage in the in-plane direction for
a specified cross-plane location.

Cross-plane distance from central axis 共cm兲

9.5 7.5 5.5 3.5 1.5 ⫺0.5 ⫺2.5 ⫺4.5 ⫺6.5 ⫺8.5 ⫺10.5

Average 共%兲 1.11 1.14 1.14 1.12 1.11 1.10 1.09 1.06 1.03 0.99 0.96
Standard dev 共%兲 0.12 0.13 0.13 0.13 0.14 0.14 0.14 0.14 0.14 0.14 0.13

Medical Physics, Vol. 30, No. 9, September 2003


2549 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2549

TABLE II. Interleaf leakage variation in the cross-plane direction. Each column shows the average and standard deviation of leakage in the cross-plane
direction between a specific leaf-pair.

Gun side of leaf

2 3 4 5 6 7 8 9 10

Average 共%兲 0.57 0.96 0.96 1.01 1.02 1.13 1.08 1.05 1.12
Standard dev 共%兲 0.06 0.05 0.05 0.07 0.06 0.07 0.07 0.04 0.07

Gun side of leaf

11 12 13 14 15 16 17 18 19

Average 共%兲 1.09 1.13 1.11 1.24 1.42 1.10 1.10 1.12 1.13
Standard dev 共%兲 0.05 0.06 0.04 0.11 0.06 0.07 0.04 0.07 0.07

Gun side of leaf

20 21 22 23 24 25 26 27 28

Average 共%兲 1.11 1.10 1.05 1.10 1.13 1.04 1.05 1.02 1.12
Standard dev 共%兲 0.05 0.05 0.04 0.06 0.06 0.07 0.08 0.07 0.12

lines around the mean value of 1.9% demonstrates minor This graph shows very little variation of the in-plane penum-
cross-plane variation. There was no difference between in- bra caused by the edge of a leaf, either in the in-plane (p
plane and cross-plane leaf-end transmission (p⫽0.4). The ⫽0.4) or cross-plane direction (p⫽0.3).
leaf-end transmission at 10 cm off the central axis for all leaf Similarly, the cross-plane penumbra caused by the end of
pairs is approximately 0.5% less than the mean, while the a leaf 关as shown in Fig. 3共b兲兴 is provided in Fig. 6共c兲. Each
central leaves are 0.5% greater than the mean for the central line represents a particular leaf, and the data points on the
leaf pairs. Figure 5共b兲 shows the frequency distribution of line indicate the cross-plane position of the penumbra mea-
leaf-end transmission, which is nearly normally distributed surement. These results show less than 0.03 cm from their
about the mean with the 99% confidence interval calculated mean value, regardless of the location of the field edge. Al-
to be 1.9%–2.0%. Seventy percent of the measured values though these figures only show two field edges 共Y2 and X2兲,
are within 0.2% of the mean, 80% are within 0.4%, and 98% the results for all four edges are shown in Fig. 6共d兲. The
are within 0.6%. Since the change in transmission for differ- 80%–20% penumbra for these 1-cm-wide fields was mea-
ent abutment positions is not symmetric about the central sured to have a weighted mean of 0.37 cm⫾0.01 cm.
axis, the variation seen is likely most dependent on MLC leaf The range in penumbra measured was between 0.31 and 0.43
position calibration accuracy. mm for all cases. The mean cross-plane penumbra was
statistically different (p⬍0.001) for the X1 bank
(0.358⫾0.0007 cm) compared to the X2 bank (0.369
C. Penumbra ⫾0.0007 cm).
Figure 6共a兲 shows a vertical dose profile 共in-plane兲
through each of the retracted leaves 关as shown in Fig. 3共a兲兴, IV. DISCUSSION
where the tongue and the groove of each opening generate an The greatest significance of this work may be found in
independent penumbra. This graph demonstrates the con- establishing the dosimetric characteristics of the MLC in or-
stancy of projected leaf width and penumbra for each leaf, der to allow the physicist to review the treatment planning
regardless of its off-axis distance in the in-plane direction. system’s accuracy in describing these effects. Nelson et al.21
That result is further shown in Fig. 6共b兲, where the in-plane of the RPC showed that all IMRT treatment planning sys-
penumbra was tested for cross-plane dependence. Each line tems they analyzed struggled to accurately model the dose in
represents a single leaf, and the data points on the line indi- organs at risk adjacent to planning target volumes, regions
cate the cross-plane position of the penumbra measurement. presumably shielded by the MLC leaves during the majority

TABLE III. Intraleaf leakage variation in the in-plane direction. Each column shows the average and standard deviation of leakage in the in-plane direction for
a specified cross-plane location.

Cross-plane distance from central axis 共cm兲

9.5 7.5 5.5 3.5 1.5 ⫺0.5 ⫺2.5 ⫺4.5 ⫺6.5 ⫺8.5 ⫺10.5

Average 共%兲 0.84 0.85 0.85 0.84 0.83 0.82 0.80 0.78 0.75 0.72 0.70
Standard dev 共%兲 0.07 0.07 0.08 0.08 0.08 0.08 0.08 0.08 0.09 0.09 0.08

Medical Physics, Vol. 30, No. 9, September 2003


2550 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2550

TABLE IV. Intraleaf leakage variation in the cross plane direction. Each column shows the average and standard deviation of leakage in the cross-plane
direction through a specific leaf.

X2 bank leaf

2 3 4 5 6 7 8 9 10

Average 共%兲 0.53 0.69 0.72 0.76 0.79 0.81 0.80 0.82 0.85
Standard dev 共%兲 0.07 0.04 0.05 0.06 0.06 0.05 0.05 0.05 0.05

X2 bank leaf

11 12 13 14 15 16 17 18 19

Average 共%兲 0.85 0.85 0.86 0.90 0.84 0.86 0.85 0.86 0.86
Standard dev 共%兲 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05

X2 bank leaf

20 21 22 23 24 25 26 27 28

Average 共%兲 0.85 0.82 0.82 0.82 0.80 0.78 0.74 0.74 0.65
Standard dev 共%兲 0.05 0.05 0.05 0.05 0.06 0.06 0.06 0.06 0.05

of IMRT treatments. Furthermore, these regions are fre- Interleaf leakage, intraleaf leakage, and leaf-end penum-
quently within a high dose-gradient region where accurate bra have been previously described for fields of 10⫻10 cm
modeling of the dose profile from 5–10 mm from the field and larger for the MLC collimators found on linear accelera-
edge may be critical. tors from Elekta 共Huq et al.9兲, Siemens 共Das et al.14兲, and
Varian 共Galvin et al.22兲. These measurements tend to be
through the central axis of the beam and for larger field sizes.
Das et al.14 showed the average intraleaf leakage to be 0.92
⫾0.12% and the average interleaf leakage to be 1.17
⫾0.13% for the same MLC design studied here. Our results
are consistent with these, in both the in-plane and cross-
plane direction, regardless of the measurement position with
respect to the central axis. Also for comparison, the transmis-
sion through leaves reported for MLCs on General Electric,
Philips, and Varian linear accelerators were in the order of
1.2%, 5%, and 4%, respectively for a 6 MV beam.9,14,22
The leaf-end transmission that we have measured has not
been reported previously in the literature for the double-
focused MLC design provided by Siemens. Klein et al.23 re-
ported this value for a single-focused MLC with rounded leaf
edges as large as 28% for 18 MV photons on the central axis,
reducing to 12% for the furthest points measured off-axis.
The values presented here for the double-focused MLC were
substantially more consistent, as would be expected. The
magnitude of the leaf-end transmission will be greatly de-
pendent on the leaf position calibration accuracy of the
MLC. At our institution, we utilize a leaf position calibration
procedure24 that produces 95.5% of the leaf-positions to be
within 1 mm and a mean within 0.4 mm.
The double-focused MLC also demonstrated consistency
in penumbra. Das et al.14 studied the change in penumbra for
a 10⫻10 cm field at several off-axis positions of the field
center. Specifically, they considered the penumbra along the
center leaf, traveling from 5 cm over-travel to 10 cm open,
and reported ⭐0.5% variation in the penumbra. Our results
show the penumbra is consistent both for the leaf-end as well
as across the side-edge of the leaves. The magnitude of the
FIG. 5. Leaf-end transmission 共a兲 through abutted leaf pairs, for all 27 1 cm
leaf pairs. Each line represents 1 of 9 different abutment locations. 共b兲 fre- penumbra measured for fields 1 cm wide 共0.37 cm penum-
quency distribution of leaf-end transmission. bra兲 was significantly less than the values reported by Das

Medical Physics, Vol. 30, No. 9, September 2003


2551 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2551

FIG. 6. In-plane and cross-plane penumbra created by the side and end of the leaf, respectively. 共a兲 An in-plane dose profile through each of 7 different
retracted leaves. 共b兲 In-plane penumbra measured at 11 different cross-plane locations. Each line represents 1 of 7 retracted leaves. 共c兲 Cross-plane penumbra
measured at 11 different cross-plane locations. Each line represents 1 of 27 leaves on the X2 leaf bank, each having a different in-plane position. 共d兲 Combined
penumbra results for all four field edges measured for all locations studied.

et al.14 for square field sizes of 5 to 35 cm per side 共0.75–1.0


a兲
Electronic mail: john.bayouth@utmb.edu
1
E. Podgorsak, G. Pike, A. Olivier, M. Pla, and L. Souhami, ‘‘Radiosur-
cm penumbra兲, as seen by the Gaussian shape of the dose
gery with high energy photon beams: A comparison among techniques,’’
profile for these narrow fields. This indicates the change in Int. J. Radiat. Oncol., Biol., Phys. 16, 857– 865 共1989兲.
penumbra is driven by additional scatter for the larger field 2
A. Eisbrauch, L. H. Marsh, M. K. Martel, J. A. Ship, R. K. Ten Haken, A.
sizes and not due to the position of the leaves or jaw. T. Pu, B. A. Frass, and A. S. Lichter, ‘‘Comprehensive irradiation of head
and neck cancer using conformal multisegment fields: Assessment of
target coverage and noninvolved tissue sparing,’’ Int. J. Radiat. Oncol.,
Biol., Phys. 41, 559–568 共1998兲.
3
V. CONCLUSIONS B. A. Fraass, M. L. Kessler, D. L. McShan, L. H. Marsh, B. A. Watson,
W. J. Dusseau, A. Eisbruch, H. M. Sandler, and A. S. Lichter, ‘‘Optimi-
Dosimetric characteristics of small and IMRT fields have zation and clinical use of multisegment intensity-modulated radiation
been presented for a double-focus MLC. The data can be therapy for high-dose conformal therapy,’’ Semin. Radiat. Oncol. 9,
used to validate treatment planning algorithms for these 60–77 共1999兲.
4
Y. Xiao, J. Galvin, M. Hossain, and R. Valicenti, ‘‘An optimized forward
fields. AAPM Task Group 507 reports that MLC designs us-
planning technique for intensity modulated radiation therapy,’’ Med.
ing nonfocused leaf ends produce a variation of transmission Phys. 27, 2093–2099 共2000兲.
through abutted leaf pairs 共as a function of leaf position兲 as 5
L. Verhey, ‘‘Comparison of three-dimensional conformal radiation
great as a factor of 2, and changes in the penumbra up to 3 therapy and intensity-modulated radiation therapy systems,’’ Semin Ra-
mm as a function of leaf position. This work shows the pen- diat. Oncol. 9, 78 –99 共1999兲.
6
Intensity Modulated Radiation Therapy Collaborative Working Group,
umbra and leakage of the double-focus MLC to be remark-
‘‘Intensity-modulated radiotherapy: Current status and issues of interest,’’
ably consistent for the range of leaf positions studied. Al- Int. J. Radiat. Oncol., Biol., Phys. 51, 880–914 共2001兲.
though some statistical differences of penumbra and leakage 7
A. Boyer, P. Biggs, J. Galvin, E. Klein, T. LoSasso, D. Low, K. Mah, and
were identified, these differences are not dosimetrically sig- C. Yu, ‘‘Basic applications of multileaf collimators. Report of the AAPM
nificant for our current applications. These dosimetric char- Radiation Therapy Committee Task Group No. 50,’’ 共Medical Physics,
acteristics are well designed for small field IMRT segments 2001兲.
8
M. R. Sontag and T. H. Steinberg, ‘‘Performance and beam characteris-
like those often seen for segmental MLC6 IMRT field seg- tics of the Siemens Primus linear accelerator,’’ Med. Phys. 26, 734 –736
ment shapes, where opposing leaf pairs are often separated 共1999兲.
9
by 10 mm or less. M. S. Huq, Y. Yu, Z. P. Chen, and N. Suntharalingam, ‘‘Dosimetric char-

Medical Physics, Vol. 30, No. 9, September 2003


2552 J. E. Bayouth and S. M. Morrill: MLC dosimetric characteristics 2552

acteristics of a commercial multileaf collimator,’’ Med. Phys. 22, 241– modulated radiation therapy delivered by a static multileaf collimator,’’
247 共1995兲. Med. Phys. 29, 1687–1692 共2002兲.
10 17
J. R. Palta, D. K. Yeung, and V. Frouhar, ‘‘Dosimetric considerations for I. Chetty and P. Charland, ‘‘Dosimetric characterization of Kodak ex-
a multileaf collimator system,’’ Med. Phys. 23, 1219–1224 共1996兲. tended dose range film in megavoltage photon beams,’’ Med. Phys. 29,
11
S. L. Meeks, F. J. Bova, S. Kim, W. A. Tome, J. M. Buatti, and W. A. 1207 共2002兲.
18
Friedman, ‘‘Dosimetric characteristics of a double-focused miniature M. Bieda, A. Sarkar, and I. Das, ‘‘Accuracy of Kodak’s EDR2 film in
multileaf collimator,’’ Med. Phys. 26, 729–733 共1999兲. measuring clinical dosimetry data,’’ Med. Phys. 29, 1264 共2002兲.
19
12
M. Stasi, B. Baiotto, F. Palamara, P. Gabriele, and G. Scielzo, ‘‘Dosim- N. Childress, L. Dong, and I. Rosen, ‘‘Rapid radiographic film calibration
etric characterization of a multileaf collimator,’’ Radiol. Med. 共Torino兲 for IMRT verification using automated MLC Fields,’’ Med. Phys. 29,
1366 共2002兲.
97, 382–388 共1999兲. 20
13 J. Bayouth, S. Zhou, and S. Morrill, ‘‘Analysis of variation in H&D curve
M. Pasquino, V. Casanova Borca, and S. Tofani, ‘‘Physical-dosimetric
creation for EDR2 film dosimetry,’’ Med. Phys. 30, 1450 共2003兲.
characterization of a multileaf collimator system for clinical implemen- 21
A. Nelson, P. Balter, W. Hanson, and G. Ibbott, ‘‘Experience with an
tation in conformational radiotherapy,’’ Radiol. Med. 共Torino兲 101, 187– IMRT head and neck QA phantom,’’ Med. Phys. 29, 1366 共2002兲.
192 共2001兲. 22
J. M. Galvin, A. R. Smith, and B. Lally, ‘‘Characterization of a multi-leaf
14
I. J. Das, G. E. Desobry, S. W. McNeeley, E. C. Cheng, and T. E. Schulth- collimator system,’’ Int. J. Radiat. Oncol., Biol., Phys. 25, 181–192
eiss, ‘‘Beam characteristics of a retrofitted double-focused multileaf col- 共1993兲.
limator,’’ Med. Phys. 25, 1676 –1684 共1998兲. 23
E. E. Klein, W. B. Harms, D. A. Low, V. Willcut, and J. A. Purdy, ‘‘Clini-
15
R. Mitra, J. Bayouth, and I. Das, ‘‘Startup characteristics and dosimetry cal implementation of a commercial multileaf collimator: Dosimetry, net-
of small monitor unit segments in step and shoot IMRT delivery for two working, simulation, and quality assurance,’’ Int. J. Radiat. Oncol., Biol.,
types of digital linear accelerators,’’ Med. Phys. 28, 1202 共2001兲. Phys. 33, 1195–1208 共1995兲.
16 24
X. R. Zhu, P. A. Jursinic, D. F. Grimm, F. Lopez, J. J. Rownd, and M. T. J. E. Bayouth, D. Wendt, and S. M. Morrill, ‘‘MLC quality assurance
Gillin, ‘‘Evaluation of Kodak EDR2 film for dose verification of intensity techniques for IMRT applications,’’ Med. Phys. 30, 743–750 共2003兲.

Medical Physics, Vol. 30, No. 9, September 2003

You might also like