Section 3
-1-
Silver (Ag)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.1923, K2= −0.0032
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.1293, K2= 0.0003
Interference
Chemical interference hardly exists.
-2-
Aluminum (Al)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0114
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0064
Interference
Al is ionized 10 to 15% by N2O-C2H2 flame. 0.1% lanthanum chloride or potassium is
added to the standard or unknown sample as ionization restrainer.
About 10% positive interference is given if Ti and Fe coexists with Al in 10 to 50-fold
concentration.
-3-
Arsenic (As)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9996
K0= 0.0002, K1= 0.0200, K2= −0.0001
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9995
K0= 0.0036, K1= 0.0113, K2= −0.0000
Interference
The greatest interference is diffusion and absorption by Air-C2H2 flame.
The measuring method sending arsine gas directly to Ar-H2 flame or heated quartz shell
is recommended for sensitivity and interference.
-4-
Gold (Au)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0000, K1= 0.0424, K2= −0.0002
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0019, K1= 0.0296, K2= −0.0001
Interference
If Fe, Cu and Ca coexist in large quantities, negative interference is given. However,
chemical interference is generally small.
-5-
Boron (B)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9999
K0= 0.0012, K1= 0.0007, K2= −0.0000
Interference
If Na coexists in large quantities, positive interference is given. Interference can be
decreased by changing the C2H2 quantity of the N2O-C2H2 flame or burner height; but
sensitivity also drops.
-6-
Barium (Ba)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0270
Interference
-7-
Beryllium (Be)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0006, K1= 0.2582, K2= −0.0015
Interference
Some ionization is made by the N2O-C2H2 flame.
Interference by Al can be restrained by the addition of HF.
-8-
Bismuth (Bi)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.0308, K2= −0.0002
[ABS]=K2*[C]2+K1*[C]+K
K0= 0.0000, K1= 0.0225, K2= −0.0001
Interference
Interference is small.
-9-
Calcium (Ca) I
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.1093
Interference
Negative interference is received if Al, Be, P, V, Si, Ti and Zr coexist. Therefore, 0.1 to
0.2% lanthanum chloride or strontium is added to the standard or unknown sample.
- 10 -
Calcium (Ca) II
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.3655
Interference
Negative interference, due to coexist substances in the Air-C2H2 flame, is removed by
using the N2O-C2H2 flame. However, it is ionized and 0.1 to 0.2% potassium chloride is
added to the standard or unknown sample.
- 11 -
Cadmium (Cd)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.6106
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.3207
Interference
Interference is small.
- 12 -
Cobalt (Co)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.1114, K2= −0.0023
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.0840, K2= 0.0014
Interference
If excessive transition elements and heavy metals coexist, interference is given. For
example, if nickel coexists (0.1% or so), 20 to 50% negative interference is obtained.
- 13 -
Chromium (Cr)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.1389
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0799
Interference
Negative interference is given by Fe which coexists in the Air-C2H2 flame. However, it
can be removed by the addition of 0.1 to 0.2% ammonium chloride.
Interference from Cu, Al, Mg, Ba, etc. can be removed by the N2O-C2H2 flame and can
be restrained by changing the acetylene quantity.
- 14 -
Cesium (Cs)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K
K0= 0.0000, K1= 0.0792, K2= −0.0016
Interference
Ionization interference is great in the Air-C2H2 flame.
Therefore 0.1% potassium (potassium chloride) is added to the standard or unknown
sample as restrainer.
Thus, absorption sensitivity is almost doubled.
- 15 -
Copper (Cu)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.1901
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.1261
Interference
Interference is small.
- 16 -
Iron (Fe)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.1147
[ABS]=K1*[C]+K
K0= 0.0000, K1= 0.0632
Interference
Slight interference is received if Ni, Co and mineral acid coexist.
Interference by Si can be removed by the addition of 0.1% calcium chloride.
Interference by citric acid can be restrained by the addition of 0.1 to 0.2% phosphoric acid.
- 17 -
Gallium (Ga)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0025, K1= 0.0045, K2= −0.0000
Interference
- 18 -
Germanium (Ge)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0008, K1= 0.0035, K2= −0.0000
Interference
Interference is small.
- 19 -
Potassium (K)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.4822
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.2126
Interference
Because it is ionized to some degree, 0.1 to 0.2% cesium chloride or rubidium chloride
is added to the standard or unknown sample.
- 20 -
Lithium (Li)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.3385, K2= −0.0187
Interference
Because it is ionized by the Air-C2H2 flame, 0.1% potassium chloride is added to the
standard or unknown sample as a restrainer.
- 21 -
Magnesium (Mg)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 2.1000, K2= −0.7497
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 1.1083, K2= −0.2839
Interference
A negative interference is given if P, Al, Ti and Si are put in the form of an oxyacid, and
coexist in the same or more quantity than Mg. This interference is retrained by the addition
of 0.1% strontium chloride to the standard or unknown sample.
There is no interference of the above elements in the N2O-C2H2 flame. However, it is
ionized and 0.1% potassium chloride or strontium chloride is added to the standard or
unknown sample.
- 22 -
Manganese (Mn)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0020, K1= 0.1883, K2= −0.0049
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0029, K1= 0.1348, K2= −0.0016
Interference
Small interference.
Because Si gives a negative interference, 0.1% calcium chloride is added.
- 23 -
Molybdenum (Mo)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0 r2=0.9981
K0= −0.0107, K1= 0.0151
[ABS]=K1*[C]+K0 r2=0.9992
K0= −0.0007, K1= 0.0097
Interference
The interference is not clear but 0.5% aluminum chloride or 2% ammonium chloride is
added to remove sensitivity differences due to the burner height.
- 24 -
Sodium (Na)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 1.0522
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.5462
Interference
Ionization by the Air-C2H2 flame can be decreased by adding 0.1% potassium chloride
or cesium chloride to the standard or unknown sample.
- 25 -
Nickel (Ni)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.1067
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0831
Interference
A negative interference is received if high concentrations of Fe and Cr coexist.
The 232.0 nm calibration curve bends at high concentrations owing to Ni232.14 nm.
The 341.5 and 352.4 nm wavelengths are suitable for the high concentration calibration
curve. (Refer to Section 1 - 2 - 2.1.a.)
- 26 -
Lead (Pb) I
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0545
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0401
Interference
A higher interference is given if 1% or more of the ion element coexists in a 5 ppm of Pb.
Because a negative interference is given if a 10-fold carbonate, phosphate, iodine
compound, or fluorine compound coexist, 0.1M EDTA is added.
- 27 -
Lead (Pb) II
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0235
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.0149
Interference
A higher interference is given if 1% or more of the ion element coexists in a 5 ppm of Pb.
Because a negative interference is given if a 10-fold carbonate, phosphate, iodine
compound, or fluorine compound coexist, 0.1M EDTA is added.
- 28 -
Palladium (Pd)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0001, K1= 0.0625, K2= −0.0015
Interference
If Al, Co or Ni coexists, 0.1 to 0.5% EDTA is added.
- 29 -
Platinum (Pt)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
• Measuring conditions
Current ; 14 mA/0 mA Burner height ; 7 mm
Wave length ; 266.0 nm Burner angle ; 0 deg
Slit width ; 0.5 nm Fuel gas flow ; 1.8 l/min
Lighting mode ; BGC-D2 Type of oxidant ; air
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.0062, K2= −0.0000
Interference
Sensitivity is affected if other precious metals coexists, and the effect can be reduced by
the addition of 0.1 to 0.2% lanthanum chloride.
- 30 -
Rubidium (Rb)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= 0.0005, K1= 0.1571, K2= −0.0051
Interference
- 31 -
Rhodium (Rh)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9996
K0= 0.0008, K1= 0.0213, K2= 0.0001
Interference
- 32 -
Antimony (Sb)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0 r2=0.9991
K0=-0.0058, K1= 0.0124
[ABS]=K1*[C]+K0 r2=0.9993
K0=-0.0061, K1= 0.0088
Interference
Spectroscopic interference is received if Cd or Cu coexists in high concentrations.
- 33 -
Selenium (Se)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9999
K0= 0.0024, K1= 0.0209, K2= −0.0001
Interference
- 34 -
Silicon (Si)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9999
K0= 0.0033, K1= 0.0044, K2= −0.0000
Interference
Ionization can be removed by the addition of 0.1% potassium chloride or sodium
chloride. Silicon absorption can be restrained under 1% coexistence of HF, boric acid and
potassium.
- 35 -
Tin (Sn) I
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0 r2=0.9968
K0= –0.0092, K1= 0.0026
[ABS]=K1*[C]+K0 r2=1.0000
K0=-0.0002, K1= 0.0014
Interference
- 36 -
Tin (Sn) II
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0 r2=0.9986
K0= 0.0063, K1= 0.0017
Interference
- 37 -
Tin (Sn) III
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0 r2=0.9989
K0= –0.0057, K1= 0.0052
[ABS]=K1*[C]+K0 r2=0.9997
K0= –0.0046, K1= 0.0030
Interference
Interference from other elements is hardly received.
- 38 -
Tin (Sn) IV
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0 r2=0.9995
K0= 0.0068, K1= 0.0031
Interference
Interference from other elements is hardly received.
- 39 -
Strontium (Sr)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9998
K0= 0.0004, K1= 0.0684, K2= –0.0008
Interference
If Si, Al, Ti, Zr, P, etc. coexist in the Air-C2H2 flame, a negative interference is
received. 0.5% lanthanum chloride or calcium chloride is added to the standard or
unknown sample.
No interference is received by the N2O-C2H2 flame, but ionization occurs. Therefore,
0.5% potassium chloride is added to the standard or unknown sample.
- 40 -
Tellurium (Te)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= –0.0004, K1= 0.0177, K2= –0.0000
Interference
- 41 -
Titanium (Ti)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0
K0= 0.0000, K1= 0.0077, K2= –0.0000
[ABS]=K2*[C]2+K1*[C]+K0
K0= –0.0000, K1= 0.0024, K2= –0.0000
Interference
If excessive hydrofluoric acid, F2 or Al coexists, sensitivity rises. Therefore, 0.1M
ammonium fluoride is added to the standard or unknown sample.
0.1% potassium chloride is added to restrain ionization.
- 42 -
Thallium (Tl)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=1.0000
K0= –0.0026, K1= 0.0163, K2= –0.0000
Interference
Interference hardly exists.
- 43 -
Vanadium (V)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K2*[C]2+K1*[C]+K0 r2=0.9999
K0= –0.0020, K1= 0.0126, K2= –0.0000
Interference
Sensitivity rises if a lot of Al, Ti, excessive Fe, or phosphoric acid coexists. If these
elements are the main components, the matrix of the standard and unknown samples is
matched.
0.1% potassium chloride is added to restrain ionization.
- 44 -
Zinc (Zn)
Standard substance and standard solution preparation method
Refer to Section 2 - 3 for the Standard solution preparation method
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.6430
[ABS]=K1*[C]+K0
K0= 0.0000, K1= 0.5098
Interference
Interference hardly exists.
- 45 -
- 46 -