The transfer function

Analytic form of transfer functionIn certain cases the transfer function is available as an analytic expression.
One common transfer function used for resistance temperature sensors (to be discussed in Chapter 3) is the
Callendar– Van Duzen equation. It gives the resistance of the sensor at a temperature T as

where the constants A, B, and C are determined by direct measurement of resistance for the specific material
used in the sensor and R0 is the temperature of the sensor at 0 ºC. Typical temperatures used for calibration are
the oxygen point (-182.962 ºC; the equilibrium between liquid oxygen and its vapor), the triple point of water
(0.01 ºC; the point of equilibrium temperature between ice, liquid water, and water vapor), the steam point (100
ºC; the equilibrium point between water and vapor), the zinc point (419.58 ºC; the equilibrium point between solid
and liquid zinc), the silver point (961.93 ºC), and the gold point (1064.43 ºC), as well as others. Consider a
platinum resistance sensor with a nominal resistance of 25 Ω at 0 C. To calibrate the sensor its resistance is
measured at the oxygen point as 6.2 Ω, at the steam point as 35.6 Ω, and at the zinc point as 66.1 Ω. Calculate
the coefficients A, B, and C and plot the transfer function between -200 ºC and 600 ºC.

Solution

1. 6.2 Ω at a temperature of -182.962 ºC (oxygen point).


2. 35.6 Ω at a temperature of 100 ºC (steam point).
3. 66.1 Ω at a temperature of 419.58 ºC (zinc point)

Reordering the Callendar-Van Duzen equation we obtain the following

R (T )
AT + BT 2 + C(T − 100)T 3 = −1,
R0