Professional Documents
Culture Documents
Preparative Pump
Reference Manual
s1
Copyright Agilent Warranty WARNING
Technologies 2000
The information For details of safety,
All rights reserved. contained in this see Safety Information
Reproduction, adaption, document is subject to on page 251.
or translation without change without notice.
prior written permission Warning Symbols Used
is prohibited, except as Agilent Technologies In This Book
allowed under the makes no warranty of
copyright laws. any kind with regard to
this material,
Part No. G1361-90000 including, but not
limited to, the implied
!
warranties or The apparatus is marked
Edition 11/00
merchantability and with this symbol when
fitness for a particular the user should refer to
Printed in Germany
purpose. the instruction manual
in order to prevent risk
Agilent Technologies of harm to the operator
shall not be liable for and to protect the
errors contained herein apparatus against
or for incidental or damage.
consequential damages
in connection with the
furnishing, performance,
or use of this material. Indicates dangerous
voltages and risk of
electrical shock.
Indicates a protective
counductor terminal.
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn
Germany
Agilent 1100 Series Preparative Pump
Reference Manual
In This Book
This manual contains technical reference information about the Agilent 1100 Series
preparative pump. The manual describes the following:
• Installation,
• Optimizing performance,
• Troubleshooting,
• Repairing,
• Specifications and
4
Contents
2 Optimizing Performance
How to optimize the preparative pump to achieve best
performance 41
5
Contents
functions 49
Status Indicators 52
Power Supply Indicator 53
Instrument Status Indicator 53
Error Messages 55
Time-out 56
Shutdown 57
Remote Time-out 58
Synchronization Lost 59
Leak 60
Leak Sensor Open 61
Leak Sensor Short 62
Compensation Sensor Open 63
Compensation Sensor Short 64
Fan Failed 65
Open Cover 66
Pressure Above Upper Limit 67
Pressure Below Lower Limit 68
Pressure Sensor not Connected 69
Wrong Data from Pressure Sensor 70
Motor Drive Temperature too High 71
Motor Drive Fault 72
Motor Maximum Power Fault 73
6
Contents
Motor Direction 74
Pump Off While Analysis 75
Index Missing 76
Position Error 77
Spurious Index 78
Encoder not Connected 79
Purge Valve (EMPV) Fuse Blown 80
Electro-Magnetic-Proportional-Valve (EMPV) Missing 81
No Pump Head 82
Wait for Volume Timed Out 83
Wait for Pressure Timed Out 84
Composition Cluster Partner Lost 85
Solvent Counter Zero 86
No Motor Motion 87
Seal Wash Current Low 88
Seal Wash 89
Wait for Encoder Position Timed Out 90
24V Supply for External Devices Low 91
Pressure Test 92
Description 92
Positioning the Blank Nut 94
Running the Pressure Test 95
Running the Test from the ChemStation 95
Running the Test from the Control Module 95
Evaluating the Results 97
Potential Causes of Pressure Test Failure 98
Leak Test 99
Description 99
Running the Leak Test 102
Running the Test from the ChemStation 102
7
Contents
8
Contents
9
Contents
10
Contents
11
Contents
8 Specifications
Performance specifications of the preparative pump 245
12
1
Power Consideration
The preparative pump power supply has wide-ranging capability (see Table
1). Consequently there is no voltage selector in the rear of the preparative
pump. There are also no externally accessible fuses.
WAR NI N G To disconnect the preparative pump from line power, unplug the power
cord. The power supply still uses some power, even if the power switch
on the front panel is turned off.
Power Cords
Different power cords are offered as options with the preparative pump. The
female end of each power cord is identical. It is plugged into the power input
socket at the rear of the preparative pump. The male ends of the power cords
are available with different designs. The power cord, which is delivered with
your preparative pump is designed to match the wall socket of your
particular country or region.
WAR NI N G Never operate your instrumentation from a power outlet that has no
ground connection. Never use a power cord other than the power cord
designed for your region.
14
Installing the Pump
Installing the Pump
WAR NI N G Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
Bench Space
The preparative pump dimensions and weight (see Table 1) allow the
instrument to be placed on almost any laboratory bench. The instrument
requires an additional 2.5 cm (1.0 inch) of space on either side, and
approximately 8 cm (3.1 inches) at the rear for the circulation of air, and
room for electrical connections. Ensure the preparative pump is installed in a
horizontal position.
If a complete Agilent 1100 Series system is to be installed on the bench, make
sure that the bench is designed to carry the weight of all the modules. For a
complete system including multiple pumps, Agilent 220 micro plate sampler
or an LC-MSD it is recommended to position the modules in two or more
stacks.
Environment
Your preparative pump will work within specifications at ambient
temperatures and relative humidity as described in Table 1.
CA UT IO N Do not store, ship or use your preparative pump under conditions where
temperature fluctuations may cause condensation within the preparative
pump. Condensation will damage the system electronics. If your preparative
pump was shipped in cold weather, leave it in its box, and allow it to warm up
slowly to room temperature to avoid condensation
15
Installing the Pump
Installing the Pump
Weight 15.0 kg
Dimensions 200 × 345 × 440 mm
(height × width × depth) (8 × 13.5 × 18 inches)
Line voltage 100 – 240 VAC, ± 10 % Wide-ranging capability
16
Installing the Pump
Installing the Pump
CA UT IO N In case you need to ship the preparative pump at a later date, it is useful to
keep the packing materials for a later shipment.
Damaged Packaging
Upon receipt of your preparative pump, inspect the shipping containers for
any signs of damage. If the containers or cushioning material are damaged,
save them until the contents have been checked for completeness and the
preparative pump has been mechanically and electrically checked. If the
shipping container or cushioning material is damaged, notify the carrier and
save the shipping material for the carrier’s inspection.
CA UT IO N If there are signs of damage to the preparative pump, please do not attempt to
install the preparative pump.
Delivery Checklist
Ensure all parts and materials have been delivered with the preparative
pump. The delivery checklist is shown in Table 2 - 5. To aid in parts
identification, please see chapter 5 “Parts and Materials”. There are two
different kits that can possibly be delivered with the pump:
1 A Basis kit (G1361-68708) is delivered with every isocratic configuration
or if the pump is the first pump in a system (isocratic configuration, order
No. G1361A).
2 A Gradient kit (G1361-68707) is delivered with every additional pump
that is added to a system’s first pump (gradient configuration, order No.
G1391A).
Please report missing or damaged parts to your local Agilent Technologies
sales and service office.
17
Installing the Pump
Installing the Pump
Description Quantity
Preparative Pump 1
Corrugated waste tubing 1
Power cable (local) 1
Manual (local) 1
Solvent cabinet 1
Basis kit (see Table 4) 1 with every FIRST (G1361A) pump of a system, only!
Description Quantity
Preparative Pump 1
Corrugated waste tubing 1
Power cable (local) 1
Gradient kit (see Table 5) 1 with very ADD-ON (G1391A) pump of a system!
Table 4 Basis Kit Contents G1361-68708 (shipped with the first pump of a system)
18
Installing the Pump
Installing the Pump
Table 4 Basis Kit Contents G1361-68708 (shipped with the first pump of a system)
19
Installing the Pump
Installing the Pump
Table 5 Gradient Kit Contents G1361-68707 (shipped with every additional pump in a
gradient system)
20
Installing the Pump
Optimizing the Stack Configuration
If your preparative pump is part of a complete 1100 series system, you can
ensure optimum performance by limiting the configuration of the system
stack to one of the following configurations. The configurations shown on
the following pages are typical configurations. They optimize the systems
flow paths, ensuring minimum delay volumes.
NOTE For a detailed view of the flow connections refer to the section “Flow
connections” in chapter 1 of the reference manuals of the individual modules.
NOTE If a single stack of the 1100 modules becomes too high, e.g. if additional
modules like a G1327A ALS Thermostat or a second detector are added or if
your bench is too high, a two or more stack configuration may be a better
setup. Separate the stacks where appropriate and place the stacks beside each
other. This may require an additional (longer) capillary between the two
modules, where the stack has been separated. You should avoid any
configuration that adds unwanted delay volume between Sampler and
Column or dispersion volume between column and the detector(s).
21
Installing the Pump
Optimizing the Stack Configuration
Figure 1 Front View of the Recommended Stack Configuration for a one Pump setup
(isocratic system) with an Agilent 220 micro plate sampler and one of the 1100
Detectors
Solvent cabinet
Preparative pump
Micro plate
sampler Detector
NOTE For details on flow connections between the individual modules refer to
Chapter 1: “Flow connections” in the reference manuals of the individual
modules.
22
Installing the Pump
Optimizing the Stack Configuration
Figure 2 Rear View of the Recommended Stack Configuration for a one Pump setup
(isocratic system) with an Agilent 220 micro plate sampler and one of the 1100
Detectors
Analog signal
to recorder
GPIB or LAN
to LC ChemStation
AC power
NOTE For details about control, remote and data cables and for flow connections to
and from the Agilent 220 MPS refer to Chapter 4: “Connecting the Agilent 220
micro plate sampler to 1100 Series LC” in the Agilent 220 micro plate sampler
reference manual (PN G2250-90002).
23
Installing the Pump
Optimizing the Stack Configuration
Figure 3 Front View of the Recommended Stack Configuration for a dual Pump setup (binary
gradient system) with an Agilent 220 micro plate sampler and one of the 1100
Detectors
Solvent cabinet
Preparative pumps
Micro plate
sampler Detector
NOTE For details on flow connections between the individual modules refer to
Chapter 1: “Flow connections” in the reference manuals of the individual
modules.
24
Installing the Pump
Optimizing the Stack Configuration
Figure 4 Rear View of the Recommended Stack Configuration for a dual Pump setup (binary
gradient system) with an Agilent 220 micro plate sampler and one of the 1100
Detectors
Analog signal
to recorder
GPIB or LAN
to LC ChemStation AC power
NOTE For details about control, remote and data cables and for flow connections to
and from the Agilent 220 micro plate sampler refer to Chapter 4: “Connecting
the Agilent 220 micro plate sampler to 1100 Series LC” in the Agilent 220 micro
plate sampler reference manual (PN G2250-90002).
25
Installing the Pump
Optimizing the Stack Configuration
Figure 5 Front View of the Recommended Stack Configuration for a dual Pump setup (binary
gradient system) with the Agilent 220 micro plate sampler and the Agilent
Quadrupol LC/MS detector
Solvent cabinet
Preparative pump A
Preparative pump B
NOTE For details on flow connections between the individual modules refer to
Chapter 1: “Flow connections” in the reference manuals of the individual
modules.
26
Installing the Pump
Optimizing the Stack Configuration
Figure 6 Rear View of the Recommended Stack Configuration for a dual Pump setup (binary
gradient system) with the Agilent 220 micro plate sampler and the Agilent
Quadrupol LC/MS detector
Remote start
cable
Network cable
LAN to LC
ChemStation
AC power
NOTE For details about control, remote and data cables and for flow connections to
and from the Agilent Quadrupol LC-MS Detector refer to Chapter 2:
“Installation - Connecting the LC/MSD” in the Agilent 1100 Series LC/MSD
System Installation Manual (PN G1946-90002). For details on flow
connections between the individual modules refer to Chapter 1: “Flow
connections” in the reference manuals of the individual modules.
27
Installing the Pump
Installing the preparative pump
Status Lamp
Power Switch
Serial number
28
Installing the Pump
Installing the preparative pump
Figure 8 Power Connector and Safety Lever at Rear of the Preparative Pump
Safety lever
Power connector
4 Connect the power cable to the power connector at the rear of the module
(see Figure 8).
5 If required, connect additional interface and control cables to the preparative
pump (Figure 9). Refer to the documentation of the Agilent 1100 Series
control module or ChemStation for LC for more information.
NOTE In an Agilent 1100 Series system, the individual modules are connected by a
CAN cable. The Agilent 1100 control module can be connected to the CAN bus
on any of the modules in the system. The Agilent Chemstation can be
connected to the system by one GPIB cable at any of the modules that have a
GPIB port, however, it is recommended to connect the GPIB cable to the
detector. A second possibility to connect the 1100 system to the ChemStation
is via LAN, preferably at the detector. A LAN board can be delivered as an
option with one of the 1100 modules and installed in the BCD board slot. For
more information about connecting the control module or ChemStation refer
to the respective user manual.
WAR NI N G Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
29
Installing the Pump
Installing the preparative pump
8-bit configu-
ration switch
Remote
CAN-DC-out
RS232C CAN cable to next module
CAN-bus
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal Agilent 1100 Series
module data transfer and synchronization.
• The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
common shut down, prepare, and so on.
• The RS-232 connector may be used to control the preparative pump from
a computer through an RS-232 connection, using appropriate software.
This connector needs to be activated by the 8-bit configuration switch
module in the upper right corner of the rear of the pump. The software
needs the appropriate drivers to support this communication. See your
software documentation for further information.
• The power input socket accepts a line voltage of 100 – 240 Volts AC ± 10 %
with a line frequency of 50 - 60 Hz. max. power consumption is 250 VA
(Volt-Amps). There is no voltage selector on your preparative pump
because the power supply has wide-ranging capability. There are no
externally accessible fuses. The security lever at the power input socket
prevents that the preparative pump cover is taken off when line power is
still connected.
• The interface (BCD) board slot is used for external contacts, BCD output
and for LAN communication.
• The CAN-DC-out provides 24 Volts DC power for external CAN devices
like a switch valve. Max. permanent power consumption is
100 mA/channel or 2 A for 2 sec.
30
Installing the Pump
Flow Connections for a single (isocratic) Preparative Pump
WAR NI N G When opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
1 Remove the front cover by pressing the snap fasteners on both sides.
31
Installing the Pump
Flow Connections for a single (isocratic) Preparative Pump
4 Connect the solvent tubes from the bottle head assemblies to the inlet
adapters of the pump. Fix the tubes in the clips of solvent cabinet and
preparative pump.
5 By holding a piece of sanding paper around the waste tubing connect it to the
electromagnetic purge valve (EMPV) and place it into your waste system.
6 If the preparative pump is not part of an Agilent 1100 System stack or placed
on the bottom of a stack, connect the corrugated waste tube to the waste
outlet of the pump leak handling system.
NOTE In order to guarantee an error-free handling of any leakage that may occur in
the system, all modules of a stack must be thoroughly aligned on top of each
other. Outstanding knobs at the rear of the top covers of the modules and
holes as counterparts at the rear of the bottom covers of the modules will help
to align the instruments nicely on top of each other by making them slide into
the correct position. This will then allow any possible spills to be routed safely
from one instrument to the next one with the help of the built in leak drainage
system. (Each module has a funnel to catch drops coming from the above
instrument, from there spills a lead through a waste drain tube into the leak
pan and out from there through an outlet on to the next module).
32
Installing the Pump
Flow Connections for a single (isocratic) Preparative Pump
Inlet ball
valves EMPV out to
next module
From seal wash solvent bottle to seal wash pump EMPV to waste
Seal wash pump to pump head Waste outlet - corrugated waste tubing
33
Installing the Pump
Flow Connections for a dual (binary gradient) Preparative Pump
WAR NI N G When opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
1 Remove the front covers of both pumps by pressing the snap fasteners on both
sides.
34
Installing the Pump
Flow Connections for a dual (binary gradient) Preparative Pump
4 Connect the solvent tubes from the bottle head assemblies to the inlet
adapters of the pumps. Fix the tubes in the clips of solvent cabinet and
preparative pumps.
5 By holding a piece of sanding paper around the waste tubing connect it to the
electromagnetic purge valve (EMPV) of each pump and place it into your
waste system.
6 If one of the preparative pumps is placed on the bottom of a stack, connect the
corrugated waste tube to the waste outlet of the pump leak handling system.
NOTE In order to guarantee an error-free handling of any leakage that may occur in
the system, all modules of a stack must be thoroughly aligned on top of each
other. Outstanding knobs at the rear of the top covers of the modules and
holes as counterparts at the rear of the bottom covers of the modules will help
to align the instruments nicely on top of each other by making them slide into
the correct position. This will then allow any possible spills to be routed safely
from one instrument to the next one with the help of the built in leak drainage
system. (Each module has a funnel to catch drops coming from the above
instrument, from there spills a lead through a waste drain tube into the leak
pan and out from there through an outlet on to the next module).
35
Installing the Pump
Flow Connections for a dual (binary gradient) Preparative Pump
11 Install the outlet capillary (G1361-67302) to the mixer and make the necessary
connections the other side of the capillary (normally a capillary to an injector
or injection valve).
12 Install the seal wash tubings, as shown in Figure 13. You need - for each of the
pumps - a tubing from your seal wash solvent bottle to the seal wash pump,
one from there to the pump head and one from the pump head to waste.
13 Install the front covers.
14 Press in the power switch to turn on the preparative pumps.
36
Installing the Pump
Flow Connections for a dual (binary gradient) Preparative Pump
Leak funnel
EMPV out to
Inlet ball bottom pump
valves G1361-67306
From seal wash
solvent bottle
Female T-piece
Tubing: Seal wash to
waste G1361-67305
Mixer
G1361-67304
From seal wash solvent bottle to seal wash pump EMPV to waste 0890-0615
Waste outlet - corrugated waste tubing Mixer out to next module G1361-67302
37
Installing the Pump
Get the System Ready for the First Analysis
When you are using the system for the first time it is recommended to prime
it to remove all the air and the possible contamination introduced in the flow
path during the installation.
WAR NI N G When opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
NOTE The pump should never be used for priming empty tubings (never let the pump
run dry). Use a syringe to draw enough solvent for completely filling the
tubings to the pump inlet before continuing to prime with the pump.
38
Installing the Pump
Get the System Ready for the First Analysis
39
Installing the Pump
Get the System Ready for the First Analysis
40
2
2 Optimizing Performance
42
Optimizing Performance
Solvent Information
Solvent Information
Always filter solvents through 0.4 µm filters, small particles can permanently
block the capillaries and valves. Avoid the use of the following
steel-corrosive solvents:
• Solutions of alkali halides and their respective acids (for example, lithium
iodide, potassium chloride, and so on).
• High concentrations of inorganic acids like sulfuric and nitric acid,
especially at higher temperatures (replace, if your chromatography
method allows, by phosphoric acid or phosphate buffer which are less
corrosive against stainless steel).
• Halogenated solvents or mixtures which form radicals and/or acids, for
example:
2CHCl3 + O2 → 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the
stabilizing alcohol.
• Chromatographic grade ethers, which can contain peroxides (for example,
THF, dioxane, di-isopropylether) such ethers should be filtered through
dry aluminium oxide which adsorbs the peroxides.
• Solvents containing strong complexing agents (e.g. EDTA).
• Mixtures of carbon tetrachloride with 2-propanol or THF dissolve stainless
steel.
43
Optimizing Performance
Prevent Blocking of Solvent Inlet Filters
Contaminated solvents or algae growth in the solvent bottle will reduce the
lifetime of the solvent filter and will influence the performance of the
preparative pump. This is especially true for aqueous solvents or phosphate
buffers (pH 4 to 7). The following suggestions will prolong lifetime of the
solvent filter and will maintain the performance of the preparative pump.
• Use sterile, if possible amber, solvent bottles to slow down algae growth.
• Filter solvents through filters or membranes that remove algae.
• Exchange solvents every two days or refilter.
• If the application permits add 0.0001 to 0.001 Molar sodium azide to the
solvent.
• Place a layer of argon on top of your solvent.
• Avoid exposure of the solvent bottles to direct sunlight.
• Filter HPLC grade (dry) Acetonitrile before use. The dryer the Acetonitrile,
the stronger the tendency to form polymers and therefore block the
system. Refilter at least every two days.
WAR NI N G When opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
44
Optimizing Performance
Prevent Blocking of Solvent Inlet Filters
NOTE Never use the system without solvent filters installed to avoid possible
blockages of the system.
45
Optimizing Performance
How to Configure the Seal Wash Function
Always use the seal wash function which is delivered with your pump.
Organic Solvents
For all organic non buffered solutions, we recommend to use the same
solvent for the seal wash as the one that is currently pumped by your
preparative pump. In this case, the seal wash pump has to be operated in
large intervals, only. Basically it is enough to keep the back of seals wet with
the help of the seal wash option.
NOTE Beware that the seal wash tubing is made of silicon. Never use any solvents
for the seal wash, that are not compatible with silicone.
46
Optimizing Performance
How to Optimize the Compressibility Compensation Setting
The compressibility compensation default settings are 75 × 10-6 /bar (best for
most aqueous solutions). Under normal conditions the default settings
reduce the pressure pulsation to values (below 2 % of system pressure for a
complete system with a preparative scale sized column installed) that will be
sufficient for most applications. If the compressibility values for the solvents
used differ from the default settings, it is recommended to change the
compressibility values accordingly. Compressibility settings can be
optimized by using the values for various solvents described in Table 7 on
page 48. If the solvent in use is not listed in the compressibility table, when
using premixed solvents and if the default settings are not sufficient for your
application the following procedure can be used to optimize the
compressibility settings:
NOTE Use the preparative pump with a flow rate of at least 20 ml/min.
1 Start the preparative pump with the adequate flow rate. The system pressure
should be between 50 and 150 bar
2 Before starting the optimization procedure, the flow must be stable. Before
starting the procedure, check the tightness of the system with the help of the
built in pressure test (see “Pressure Test” on page 92).
3 Verify the tightness of the pump itself (piston/seal combinations, valves, filters
fittings, etc.) with the help of the built in Leak Test (see “Leak Test” on page
99).
4 Your pump must be connected to a ChemStation or handheld controller, the
pressure and %-ripple can be monitored with one of these instruments.
5 Starting with a compressibility setting of 50 × 10-6 /bar increase the value in
steps of 10. The compressibility compensation setting that generates the
smallest pressure ripple is the optimum value for your solvent composition.
6 Repeat step 1 through step 5 for each preparative pump in a system with a
multiple pump setup.
47
Optimizing Performance
How to Optimize the Compressibility Compensation Setting
Acetone 126
Acetonitrile 115
Benzene 95
Carbon tetrachloride 110
Chloroform 100
Cyclohexane 118
Ethanol 114
Ethyl acetate 104
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
Toluene 87
THF 95
Water 46
48
3
3 Troubleshooting and
Test Functions
Status Indicators
The preparative pump is provided with two status indicators which indicate
the operational state (prerun, run, and error states) of the preparative pump.
The lower left indicates the power supply status, the upper right indicates the
preparative pump status.The status indicators provide a quick visual check of
the operation of the preparative pump (see “Status Indicators” on page 52).
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the preparative
pump generates an error message in the user interface. The following pages
describe the meaning of these error messages. For each message, a short
description of the failure, a list of probable causes of the problem, and a list
of suggested actions to fix the problem are provided (see “Error Messages”
on page 55).
Pressure Test
The pressure test is a quick test designed to determine the pressure tightness
of the system. After exchanging flow path components (e.g., pump seals or
injection seal), use this test to verify the system is pressure tight up to 400 bar
(see “Pressure Test” on page 92).
Leak Test
The leak test is a diagnostic test designed to determine the pressure tightness
of the preparative pump. When a problem with the preparative pump is
suspected, use this test to help troubleshoot the preparative pump and its
pumping performance (see “Leak Test” on page 99).
50
Troubleshooting and Test Functions
EMPV Cleaning
Depending on the application, particles can sometimes be collected in the
EMPV valve. This cleaning procedure is designed to remove the particle
deposits. This procedure should always be performed when the EMPV is
suspected of being leaky, or contaminated with particles (see “EMPV
Cleaning” on page 106).
51
Status Indicators
Two status indicators are located on the front of the preparative pump. The
lower left one indicates the power supply status, the upper right one
indicates the instrument status.
Status indicator
52
Troubleshooting and Test Functions
Power Supply Indicator
The power supply indicator is integrated into the main power switch. When
the indicator is illuminated (green) the power is ON.
NOTE The power switch stays pressed in and a green indicator lamp in the power
switch is on when the preparative pump is turned on. When the line power
switch stands out and the green light is off, the preparative pump is turned off.
The instrument status indicator on the top right corner of the preparative
pump’s front indicates one of four possible instrument conditions:
• When the status indicator is OFF (and power switch light is on), the
preparative pump is in a prerun condition, and is ready to begin an
analysis.
• A green status indicator, indicates the preparative pump is performing an
analysis (run mode).
• A yellow indicator indicates a not-ready condition. The preparative pump
is in a not-ready state when it is waiting for a specific condition to be
reached or completed (for example, immediately after changing a
setpoint), or while the purge valve is open or while a self-test procedure is
running.
• An error condition is indicated when the status indicator is red. An error
condition indicates the preparative pump has detected an internal problem
which affects correct operation of the instrument. Usually, an error
condition requires attention (for example, leak, defective internal
components). An error condition always interrupts the analysis.
• A flashing yellow status indicator indicates that the module is in its
resident mode. Call your local service provider for assistance upon
observing this error condition.
53
Troubleshooting and Test Functions
Instrument Status Indicator
• A flashing red status indicator indicates a severe error during the startup
procedure of the module. Call your local service provider for assistance
upon observing this error condition.
54
Troubleshooting and Test Functions
Error Messages
Error Messages
55
Troubleshooting and Test Functions
Time-out
Time-out
56
Troubleshooting and Test Functions
Shutdown
Shutdown
57
Troubleshooting and Test Functions
Remote Time-out
Remote Time-out
58
Troubleshooting and Test Functions
Synchronization Lost
Synchronization Lost
59
Troubleshooting and Test Functions
Leak
Leak
60
Troubleshooting and Test Functions
Leak Sensor Open
The leak sensor in the preparative pump has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current
to change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes • Leak sensor not connected to the PPM board.
• Defective leak sensor.
• Leak sensor incorrectly routed, being pinched by a metal component.
Suggested Actions ❏ Ensure the leak sensor is connected correctly.
❏ Exchange the leak sensor.
61
Troubleshooting and Test Functions
Leak Sensor Short
The leak sensor in the preparative pump has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current
to change within defined limits. If the current increases above the upper
limit, the error message is generated.
Probable Causes • Defective leak sensor.
• Leak sensor incorrectly routed, being pinched by a metal component.
Suggested Actions ❏ Exchange the leak sensor.
62
Troubleshooting and Test Functions
Compensation Sensor Open
63
Troubleshooting and Test Functions
Compensation Sensor Short
64
Troubleshooting and Test Functions
Fan Failed
Fan Failed
65
Troubleshooting and Test Functions
Open Cover
Open Cover
WAR NI N G Any adjustment, maintenance or repair of the instrument with the top
cover removed and with the power cord plugged in is strictly
forbidden.
WAR NI N G Disconnect the instrument from the line and unplug the power cord
before maintenance.
66
Troubleshooting and Test Functions
Pressure Above Upper Limit
The system pressure has exceeded the upper pressure limit. This error
condition automatically results in the opening of the purge valve.
Probable Causes • Upper pressure limit set too low.
• Blockage in the flowpath (after the pressure sensor).
• Defective pressure sensor.
• Defective PPM board.
Suggested Actions ❏ Ensure the upper pressure limit is set to a value suitable for the analysis.
❏ Check for blockage in the flowpath.
❏ Exchange the Multi Assembly.
❏ Exchange the PPM board.
67
Troubleshooting and Test Functions
Pressure Below Lower Limit
The system pressure has fallen below the lower pressure limit.
Probable Causes • Lower pressure limit set too high.
• Air bubbles in the mobile phase.
• Leak.
• Defective pressure sensor.
• Defective PPM board.
Suggested Actions ❏ Ensure the lower pressure limit is set to a value suitable for the analysis.
❏ Ensure bubble free flow. Purge the preparative pump.
❏ Ensure solvent inlet filters are not blocked.
❏ Inspect the pump head, capillaries and fittings for signs of a leak.
❏ Purge the preparative pump. Run a pressure test to determine whether the
seals or other pump components are defective.
❏ Exchange the Multi Assembly.
❏ Exchange the PPM board.
68
Troubleshooting and Test Functions
Pressure Sensor not Connected
The pressure readings, read by the pump ADC (analog-digital converter), are
missing.
The ADC reads the pressure readings from the pressure sensor every 1ms. If
the readings are missing for longer than 10 seconds, the error message is
generated.
Probable Causes • Pressure sensor not connected (should be connected to P29 on the PPM
board) or connected to the seal wash cable connector (P23).
• Defective pressure sensor.
• Defective PPM board.
Suggested Actions ❏ Ensure the pressure sensor connector is clean, connected and seated
correctly.
❏ Exchange multi assembly.
❏ Exchange the PPM board.
69
Troubleshooting and Test Functions
Wrong Data from Pressure Sensor
The pressure sensor delivers readings within the range of -5 to -25 bar.
Probable Causes • Defective pressure sensor.
• Defective PPM board.
Suggested Actions ❏ Exchange the pressure sensor.
❏ Exchange the PPM board.
70
Troubleshooting and Test Functions
Motor Drive Temperature too High
71
Troubleshooting and Test Functions
Motor Drive Fault
The motor drive chip has issued a fault signal. The current drawn by the
pump motor exceeded the maximum limit or under voltage lockout.
Blockages in the flowpath are usually detected by the pressure sensor, which
result in the preparative pump switching off when the upper pressure limit is
exceeded. If a blockage occurs before the pressure sensor (i.e. the pressure
increase cannot be detected by the pressure sensor), the preparative pump
will continue to pump. As pressure increases, the pump drive draws more
current. When the current reaches the maximum limit, the preparative pump
is switched off, and the error message is generated.
Probable Causes • Flow path blockage in front of the pressure sensor.
• Blocked outlet ball valve.
• High friction (partial mechanical blockage) in the drive assembly.
• Defective drive assembly.
• Defective PPM board.
Suggested Actions ❏ Ensure the capillaries and frits between the pump head and pressure
sensor (at the rear of the multi assembly) are free from blockage. Clean the
filters in the multi assembly.
❏ Exchange the outlet ball valve.
❏ Remove the pump head assembly. Ensure there is no mechanical blockage
of the pump head assembly or drive assembly.
❏ Exchange the drive assembly.
❏ Exchange the PPM board.
72
Troubleshooting and Test Functions
Motor Maximum Power Fault
The motor control has determined that the set maximum for the motor power
was exceeded for longer than 1 second. To protect the circuitry from damage
this error event is issued in this case.
Blockages in the flowpath are usually detected by the pressure sensor in the
multi assembly, which results in the preparative pump switching off when the
upper pressure limit is exceeded. If a blockage occurs before the pressure
sensor (i.e. the pressure increase cannot be detected by the pressure sensor),
the preparative pump will continue to pump. As pressure increases, the pump
drive draws more current. When the current reaches the maximum limit, the
preparative pump is switched off, and the error message is generated.
Probable Causes • Flow path blockage in front of the pressure sensor.
• Blocked outlet ball valve.
• High friction (partial mechanical blockage) in the drive assembly.
• Defective drive assembly.
• Defective PPM board.
Suggested Actions ❏ Ensure the capillaries and frits between the pump head and pressure
sensor (at the rear of the multi assembly) are free from blockage. Clean the
filters in the multi assembly.
❏ Exchange the outlet ball valve.
❏ Remove the pump head assembly. Ensure there is no mechanical blockage
of the pump head assembly or drive assembly.
❏ Exchange the drive assembly.
❏ Exchange the PPM board.
73
Troubleshooting and Test Functions
Motor Direction
Motor Direction
The motor control has determined that the motor turns in opposite direction.
Probable Causes • Cables connected in wrong orientation after a repair.
Suggested Actions ❏ Ensure the cables are correctly connected (on the main board and at the
pump drive) and reconnect with new orientation, if necessary.
74
Troubleshooting and Test Functions
Pump Off While Analysis
This error message is generated, if the pump was turned off, during an
analysis (Power Off or pump in a Not-Ready condition).
Probable Causes • Power Off or pump in a Not-Ready condition.
Suggested Actions ❏ Turn On the Pump.
❏ Eliminate the cause of the Not-Ready condition.
75
Troubleshooting and Test Functions
Index Missing
Index Missing
The pump motion control expects one index signal for every turn of the drive
wheel. The error message is generated, if this signal is not received within
one complete turn of the drive wheel.
Probable Causes • Defective or disconnected index cable.
• Defective Pump drive defective index board on pump drive.
• Defective main board.
Suggested Actions ❏ Ensure the connector connected and seated correctly.
❏ Exchange the pump drive assembly.
❏ Exchange the main board.
76
Troubleshooting and Test Functions
Position Error
Position Error
The motion control of the pump drive could not reach the desired condition
within a certain time frame. The deviation from the desired position
exceeded the error limit.
Probable Causes • Blocked flow path.
• Mechanical blockage of the pump drive assembly.
Suggested Actions ❏ Ensure that there is no blockage in the flow path before the pressure
sensor.
❏ Exchange the pump drive assembly.
77
Troubleshooting and Test Functions
Spurious Index
Spurious Index
The pump motion control expects one (and only one) index signal for every
turn of the drive wheel. This error message is generated, if this signal is
received more often than within one complete turn of the drive wheel.
Probable Causes • Missing reflections to light switch because of contamination on drive
wheel.
• Defective Pump drive.
• Defective main board.
Suggested Actions ❏ Ensure the drive wheel is clean.
❏ Exchange the pump drive assembly.
❏ Exchange the main board.
78
Troubleshooting and Test Functions
Encoder not Connected
There is an encoder wheel mounted to the shaft of the pump motor, that gives
pulses to the motion control according to the rotation of the shaft. The error
message is generated if the short bridge on the connector is not sensed.
Probable Causes • Defective or disconnected pump encoder connector or cable.
• Defective main board.
Suggested Actions ❏ Ensure the connector is clean, and seated correctly.
❏ Exchange main board.
79
Troubleshooting and Test Functions
Purge Valve (EMPV) Fuse Blown
The electronic fuse has detected that the current through the purge valve was
too high.
Probable Causes • Defective purge valve.
• Defective PPM board.
Suggested Actions ❏ Exchange the drive of the EMPV.
❏ Exchange the PPM board.
80
Troubleshooting and Test Functions
Electro-Magnetic-Proportional-Valve (EMPV) Missing
Electro-Magnetic-Proportional-Valve
(EMPV) Missing
81
Troubleshooting and Test Functions
No Pump Head
No Pump Head
82
Troubleshooting and Test Functions
Wait for Volume Timed Out
The wait time limit until a given volume has been pumped, has been
exceeded.
Probable Causes • Wrong limit set for the time limit.
• Flow setting too low.
Suggested Actions ❏ Set limit accordingly.
❏ Set flow accordingly.
83
Troubleshooting and Test Functions
Wait for Pressure Timed Out
The limit, for waiting until a given pressure is reached, has been exceeded.
Probable Causes • Wrong limit set for the time limit.
• Leak.
Suggested Actions ❏ Set limit accordingly.
❏ Check for leaks with the help of the built in pressure or leak test and
correct them, where necessary.
84
Troubleshooting and Test Functions
Composition Cluster Partner Lost
85
Troubleshooting and Test Functions
Solvent Counter Zero
86
Troubleshooting and Test Functions
No Motor Motion
No Motor Motion
87
Troubleshooting and Test Functions
Seal Wash Current Low
The current sensing of the seal wash motor was lower than expected.
Probable Causes • Defective or disconnected seal wash pump motor.
• Defective main board.
Suggested Actions ❏ Ensure the cable is correctly connected.
❏ Exchange the seal wash pump motor.
❏ Exchange main board.
88
Troubleshooting and Test Functions
Seal Wash
Seal Wash
WAR NI N G If the seal wash pump is turned on, while the pressure sensor is
connected to the main board port that should have the seal wash pump
installed, this can result in a defective pressure sensor (requires
exchange of complete multi assembly).
Probable Causes • Pressure sensor and seal wash pump connector were connected to the
wrong position on the main board (identical connectors allow to plug into
the wrong positions by mistake)
Suggested Actions ❏ Switch the connectors of the seal wash pump cable and the pressure
sensor to meet the correct positions (P23 for seal wash cable, P29 for the
pressure sensor cable from the multi assembly).
89
Troubleshooting and Test Functions
Wait for Encoder Position Timed Out
90
Troubleshooting and Test Functions
24V Supply for External Devices Low
Voltage sensing of the 24 Volts CAN outlet shows low voltage, probably fuse
has blown.
Probable Causes • Excessive current has been drawn by an external device connected to the
external 24 CAN-DC out.
• Defective main board.
Suggested Actions ❏ Ensure, that any device, connected to the external 24 CAN-DC out, does
not draw more than 100 mA as a permanent current or a maximum of 2 A
for no longer than 2 seconds. The error condition will be reset
automatically, when the excessive current is not drawn from the external
24 CAN-DC out port any more.
❏ Exchange main board.
91
Troubleshooting and Test Functions
Pressure Test
Pressure Test
Description
The pressure test is a quick, built-in test designed to demonstrate the
pressure-tightness of the system. The test should be used when problems
with leaks are suspected, or after maintenance of flow-path components
(e.g., pump seals, injection seal) to prove pressure tightness up to 350 bar.
The test involves monitoring the flow rate, required to keep a constant
pressure of 350 bar, as the preparative pump runs through a predefined
pumping sequence. The resulting flow rate provides information about the
pressure tightness of the system.
The last connection on the high pressure side of the system (usually the
fitting that connects to the inlet of the column) is blocked by a blank nut, and
then the test is run using isopropyl alcohol (IPA), while monitoring the flow
rate. The screen for the pressure test, performed with the ChemStation, is
shown in Figure 15.
92
Troubleshooting and Test Functions
Pressure Test
Step 1 After the preparation of the system by thoroughly flushing the entire system
with Isopropanol and removing all contaminations and potential air bubbles,
releasing the pressure in the system and positioning the blank nut, the test
begins with an automated flushing of the pump at a flow rate of 25 ml/min
with an open EMPV.
Step 2 The EMPV is closed and next, piston A begins pumping solvent with a flow
rate of 2 ml/min until reaching a pressure of 150 bar, then the pump delivers
at a flow rate of 0.2 ml/min until reaching a pressure of 250 bar, then the
pump delivers at a flow rate of 0.1 ml/min until reaching a pressure of
345 bar, then the pump switches into its “Keep pressure mode”.
NOTE To test the pressure tightness of all individual components in the flow path of
the pump itself use the leak test, see “Leak Test” on page 99.
Step 3 When the system pressure reaches 350 bar, the preparative pump switches to
its “Keep-pressure-mode”. The flow rate to compensate for leaks should from
this point onwards be no more than 10 µl/minute. The results of the test are
automatically evaluated and updated. So it is possible to tighten some fittings
in order to remove leaks, while monitoring the flow rate, required to keep the
constant pressure. Figure 16 shows the result of the pressure test.
93
Troubleshooting and Test Functions
Pressure Test
94
Troubleshooting and Test Functions
Running the Pressure Test
NOTE Make absolutely sure that all parts of the flow path that are part of the
test are very thoroughly flushed with IPA before starting to pressurize
the system! Any trace of other solvents or the smallest air bubble
inside the flow path definitely will cause the test to fail!
95
Troubleshooting and Test Functions
Running the Pressure Test
96
Troubleshooting and Test Functions
Evaluating the Results
The sum of all leaks between the pump and the blank nut will be indicated by
flow rate of >10 µl/minute to keep the pressure constant. Note that small
leaks may cause the test to fail, but solvent may not be seen leaking from a
module.
NOTE Please notice the difference between an error in the test and a failure of the
test! An error means that during the operation of the test there was an
abnormal termination. If a test failed, this means that the results of the test
where not within the specified limits.
NOTE Do not overtighten fittings! After the first major increase of resistance when
tightening a previously loose fitting, the fitting should not be tightened any
further than an additional 1/8 to 1/4 wrench turn.
NOTE Often it is only a damaged blank nut itself (poorly shaped from overtightening)
that causes a failure of the test. Before investigating on any other possible
sources of failure make sure that the blank nut you are using is in good
condition and properly tightened!
• If the test fails again, insert the blank nut at the outlet of the previous
module in the stack and repeat the pressure test. Exclude each module one
by one to determine which module is leaking.
• If the pump is determined to be the source of the leak, run the EMPV
cleaning procedure, repeat the pressure test and if the test fails again, run
the leak test.
• Large internal leaks at the EMPV can easily be identified by observing a
flow out through the waste tubing of the EMPV, while the EMPV should be
closed.
97
Troubleshooting and Test Functions
Evaluating the Results
98
Troubleshooting and Test Functions
Leak Test
Leak Test
Description
The leak test is a quick, built-in test designed to demonstrate the
pressure-tightness of the pump. The test should be used when problems with
leaks are suspected, or after maintenance of flow-path components (e.g.,
pump seals, filters, EMPV or capillaries) to prove pressure tightness up to
350 bar. The test involves monitoring the pump drive steps needed to build up
a pressure of 350 bar and the flow rate, required to keep a constant pressure
of 350 bar, as the preparative pump runs through a predefined pumping
sequence. The resulting flow rate provides information about the pressure
tightness of the pump. Each step is separately performed and evaluated
separately for piston 1 and piston 2.
The outlet of the EMPV is blocked by a blank nut, and then the test is run
using isopropyl alcohol (IPA), while monitoring the flow rate. The screen for
the leak test, performed with the ChemStation, is shown in Figure 15.
99
Troubleshooting and Test Functions
Leak Test
Step 1 After the preparation of the pump by thoroughly flushing the pump with
Isopropanol and removing all contaminations and potential air bubbles,
releasing the pressure in the system and positioning the blank nut, the test
begins with an automated flushing of the pump at a flow rate of 25 ml/min
with an open EMPV.
Step 2 The EMPV is closed and next, piston 1 begins pumping solvent with a flow
rate of 2 ml/min until reaching a pressure of 150 bar, then the pump delivers
at a flow rate of 0.2 ml/min until reaching a pressure of 250 bar, then the
pump delivers at a flow rate of 0.1 ml/min until reaching a pressure of
345 bar, then the pump switches into its “Keep pressure mode”.
The result of the encoder value position shows how many metering steps
were needed to build up the pressure of 350 bar for each individual piston.
For piston 1 an encoder value position from 3500 - 16000 is allowed, for
piston 2 a value from 19500 - 32000 is allowed. Any failure to pressurize the
system within the specified ranges normally indicates an air bubble in the
pump.
Step 3 The flow rate to compensate for leaks should from this point onwards be no
more than 10 µl/minute. The results of the test are automatically evaluated.
Step 4 Steps 2 and three are automatically repeated for piston 2
100
Troubleshooting and Test Functions
Leak Test
101
Troubleshooting and Test Functions
Running the Leak Test
NOTE Make absolutely sure that all parts of the flow path that are part of the
test are very thoroughly flushed with IPA before starting to pressurize
the pump! Any trace of other solvents or the smallest air bubble inside
the flow path definitely will cause the test to fail!
NOTE It is recommended to run the EMPV cleaning procedure (see “EMPV Cleaning”
on page 106) prior to running the Leak test, to remove any potential deposits
from the EMPV valve.
The slopes and plateaus are evaluated automatically. “Evaluating the Results”
on page 103 describes the evaluation and interpretation of the leak test
results.
102
Troubleshooting and Test Functions
Evaluating the Results
Defective or leaky components in the pump’s flow path lead to changes in the
leak-test pressure plot. Typical failure modes are described below.
NOTE Please notice the difference between an error in the test and a failure of the
test! An error means that during the operation of the test there was an
abnormal termination. If a test failed, this means that the results of the test
where not within the specified limits.
NOTE Often it is only a damaged blank nut itself (poorly shaped from overtightening)
that causes a failure of the test. Before investigating on any other possible
sources of failure make sure that the blank nut you are using is in good
condition and properly tightened!
103
Troubleshooting and Test Functions
Evaluating the Results
Error Limits Exceeded for One or Both of the Encoder Values and One
or Both of the Maximum allowed Flow Rates in the Constant Pressure
Mode
104
Troubleshooting and Test Functions
Evaluating the Results
Externally leaking inlet or outlet valve in Ensure the valves are installed correctly.
channel 1. Tighten the valves.
Internally leaking inlet valve in channel 1 Clean or replace the inlet valve in channel 1.
(back flow).
Internally leaking outlet valve in channel 2 Clean or replace the outlet valve in channel 2.
(back flow).
Leaking seal or scratched plunger in channel Exchange the pump seals in channel 1. Check
1. the plunger for scratches. Exchange if
scratched.
Loose or leaky fitting or capillary or fitting in Tighten fitting or replace capillary if
channel 1. necessary in channel 1.
Externally leaking inlet or outlet valve in Ensure the valves are installed correctly.
channel 2. Tighten the valves.
Internally leaking inlet valve in channel 2 Clean or replace the inlet valve in channel 2.
(back flow).
Internally leaking outlet valve in channel 1 Clean or replace the outlet valve in channel 1.
(back flow).
Leaking seal or scratched plunger in channel Exchange the pump seals in channel 2. Check
2. the plunger for scratches. Exchange if
scratched.
Loose or leaky fitting or capillary or fitting in Tighten fitting or replace capillary if
channel 2. necessary in channel 2.
105
Troubleshooting and Test Functions
EMPV Cleaning
EMPV Cleaning
NOTE Large internal leaks at the EMPV can easily be identified by observing a flow
out through the waste tubing of the EMPV, while the EMPV should be closed.
The EMPV outlet is plugged with a blank nut (01080-83202). After a short
flushing routine the EMPV is closed and the pressure is increased to
approximately 350 bar. Then the EMPV is opened and the pressure is
released very fast. The procedure is repeated 5 times in a sequence.
Figure shows, the ChemStation’s screen for the procedure:
106
Troubleshooting and Test Functions
EMPV Cleaning
1 Plug the pump outlet with a blank nut (01080-83202) at EMPV outlet.
2 Execute procedure (In the ChemStation’s Diagnosis view select ‘Tests’ and
choose the pump - EMPV Cleaning. In the Control Module go to System
Screen (F5, choose ‘System’)- Tests (F3) - select the pump and the EMPV
Cleaning procedure.).
NOTE This procedure can also be performed with the Control module by manually
entering a flow rate that pressurizes the pump that is blocked with a blanking
nut (01080-83202) at the EMPV outlet to 350 bar. Lower the flow rate with
increasing pressure. As the pressure approaches 350 bar, the pressure should
be released suddenly by turning on the EMPV. This procedure should be
repeated 5 times. Run a leak test (see “Leak Test” on page 99 after having
performed the cleaning procedure, to ensure tightness of the EMPV.
3 For better removal of any left-over particles, it will help to clean the ball of
the valve with an additional procedure, if the previous step did not help:
Using a 7/16” wrench remove deposits by turning the holding screw of the
EMPV body back and forth about 1/4 wrench turn several times.
Figure 21 Removing Deposits from the EMPV with the Help of a Wrench:
107
Troubleshooting and Test Functions
EMPV Cleaning
108
4
Simple Repairs
The preparative pump is designed for easy repair. The most frequent repairs
such as plunger seal change and filter change can be done from the front of
the preparative pump with the preparative pump in place in the system stack.
These repairs are described in “Simple Repair Procedures” on page 114.
WAR NI N G When opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
WAR NI N G The following procedures require opening the main cover of the
preparative pump. Always ensure the preparative pump is
disconnected from the line power when the main cover is removed. The
security lever at the power input socket prevents that the pump cover
is taken off when line power is still connected. To prevent personal
injury, the power cable must be removed from the preparative pump
before opening the pump cover. Do not connect the power cable to the
preparative pump while the covers are removed.
110
Repairing the Pump
Simple Repairs
WAR NI N G To disconnect the preparative pump from line, unplug the power cord.
The power supply still uses some power, even if the switch on the front
panel is turned off.
WAR NI N G NEVER remove the cover of the power supply. Charged capacitor
devices may still contain voltages hazardous for the operator’s health,
even if the pump is NOT connected to line power. The power supply
does not contain any user serviceable parts.
NOTE The electronics of the preparative pump will not allow operation of the
preparative pump when the top cover and the top foam are removed. A safety
light switch on the main board will inhibit the operation of the preparative
pump. Always operate the preparative pump with the top foam and top covers
in place.
111
Repairing the Pump
Using the ESD Strap
WAR NI N G Do not let liquid drip into the preparative pump. It could cause shock
hazard and it could damage the preparative pump.
112
Repairing the Pump
Overview
Overview
Figure 23 shows the main assemblies of the preparative pump. The pump
heads and its parts do require normal maintenance (for example, seal
exchange) and can be accessed from the front (simple repairs). Replacing
internal parts will require to remove the module from its stack and to open
the top cover.
113
Repairing the Pump
Simple Repair Procedures
The procedures described in this section can be done with the preparative
pump in place in the system stack.
114
Repairing the Pump
Exchanging the Inlet/Outlet Ball Valves
NOTE The inlet valves do not necessarily have to be replaced in pairs. Use the built
in Leak test (see “Leak Test” on page 99) to identify the valve, that is not
working properly.
Valve assembly
Valve adapter
115
Repairing the Pump
Exchanging the Inlet/Outlet Ball Valves
NOTE The outlet valves do not necessarily have to be replaced in pairs. Use the built
in Leak test (see “Leak Test” on page 99) to identify the valve, that is not
working properly.
1 Disconnect the solvent outlet capillary at the valve adapter, with a 1/4 inch
wrench. Beware of leaking solvents.
2 Using a 17-mm wrench loosen the outlet valve adapter and remove the valve
adapter from pump head.
3 Lift the valve adapter out of the housing in the pumphead.
4 With a pair of tweezers carefully remove valve assembly.
NOTE The side of the valve assembly that has the two holes must always face in flow
direction!
3 Insert the valve adapter with the new valve assembly into the housing and
screw tight.
4 Reconnect all capillaries and tubings.
NOTE Don’t reverse inlet and outlet valve adapters. The outlet valve adapters have a
smaller hole for fitting of the SST outlet capillary. The inlet adapters are
designed for the finger tight fittings of the inlet tubing.
116
Repairing the Pump
Exchanging the Filter Cup
NOTE Whenever replacing the filters on the multi assembly is required because of a
pump seal change or because excessive back pressure indicates the need for
a replacement, make sure to replace the filter cup first and then the filter frit
(see“Exchanging the Filter Frit” on page 119), after having pumped another
200 ml of solvent.
NOTE Generally, all components which are in direct contact with solvents should be
handled wearing gloves.
117
Repairing the Pump
Exchanging the Filter Cup
Filter Screw
Filter Cup
4 Before reinserting the new filter screw into the multi assembly, wash the filter
housing thoroughly with a wash bottle, containing Isopropanol, to remove any
contamination.
5 Install a new filter into the filter screw.
6 Reinsert the filter and filter screw and turn until finger tight, then another 1/8
turn using a 17 mm wrench.
NOTE Do not overtighten the filter screw. Excessive force on the filter screw doesn’t
help in case of a leak.
7 Pump about 200 ml of solvent before replacing the filter frit installed at the
multi assembly: the filter frit between the outlet of the multi assembly and the
EMPV (see “Exchanging the Filter Frit” on page 119).
118
Repairing the Pump
Exchanging the Filter Frit
When required When the back pressure exceeds 10 bar pumping 50 ml/min of pure water with open purge
valve. After 200 ml solvent have been pumped through the pump after a filter cup exchange.
Tools required Wrench 7/16 inch
Wrench 1/4 inch
Hex-key 4 mm
Wash bottle containing Isopropanol
Parts required Filter frit 2µm, PN 5022-2192
NOTE Whenever replacing the filters on the multi assembly is required because of a
pump seal change or because excessive back pressure indicates the need for
a replacement, make sure to replace the filter cup first (see “Exchanging the
Filter Cup” on page 117) and then the filter frit, after having pumped another
200 ml of solvent.
119
Repairing the Pump
Exchanging the Filter Frit
Filter frit
Hole
Filter holder
Filter holder
Filter frit
6 Reconnect capillary between multi assembly and EMPV using a 1/4 inch
wrench.
120
Repairing the Pump
Exchanging the Complete EMPV or just the Valve
NOTE For this procedure it is especially important to follow the instructions given
below. A failure to align the EMPV valve correctly might result in a failure of
the EMPV or in a defective EMPV.
121
Repairing the Pump
Exchanging the Complete EMPV or just the Valve
Figure 28 Electro-Magnetical-Purge-Valve
Valve body
Capillary from filter
EMPV valve
Capillary to system
Screws
122
Repairing the Pump
Exchanging the Complete EMPV or just the Valve
123
Repairing the Pump
Exchanging the Seal Wash Pump Cassette
1 Disconnect seal wash tubing at the connectors, and place the ends of each
tubing at a level above seal wash bottle (to avoid leaking solvents).
2 By pressing the upper and lower clip remove the seal wash cassette.
3 Replace the defective cassette and slide the new cassette on to the shaft
pressing upper and lower clip. Push forward until the clips snap into place.
The preferred orientation of the cassette is as shown in picture below.
4 Reconnect seal wash tubing.
124
Repairing the Pump
Removing and Disassembling the Pump Head Assembly
WAR NI N G Never start the pump when the pump head is removed. This may
damage the pump drive.
NOTE Seals should always be replaced as pairs (2x front seals + 2x back seals)!
125
Repairing the Pump
Removing and Disassembling the Pump Head Assembly
2 Using a 4-mm hexagonal key step wise loosen 3 Place the pump head on a flat surface
and remove the four pump head screws and facedown. Pull out the plungers from the
remove the pump head from the pump drive. spring assembly.
Spring assembly
4 Loosen the two screws on the spring 5 Remove secondary seals between the
assembly and carefully pull the spring pumphead and the spring assembly, by using
assembly and pumphead apart. the plungers.
Secondary seals
Screws
Spring assembly
Hold here
Pump head
126
Repairing the Pump
Removing and Disassembling the Pump Head Assembly
6 Using a 10-mm wrench loosen and remove the 7 Carefully lift up the rear pump head.
two special screws.
Special screws
Pump head rear
8 Carefully remove the seals from the pump 9 Remove the O-rings from the rear pump head
head front using the pistons. using a toothpick, if they require
replacement.
O-rings
Seals
127
Repairing the Pump
Reassembling the Pump Head Assembly
128
Repairing the Pump
Reassembling the Pump Head Assembly
2 Take rear pump head and place on top of 3 Insert and tighten special screws using a
front pump head. Beware that the O-rings are 10 mm wrench.
in proper place and the seal wash fitting is
facing in same direction as outlet valves.
Special screws
Seal wash fitting
4 Insert new secondary seals with the spring 5 Place Spring assembly on special screws and
facing the pump body. Press the seals firmly tighten the two flathead screws using a
into place. pozidrive#1. The RF-tag must be on the upper
side of the pumphead.
Flathead screws
Seals
Spring assembly
RF-tag
129
Repairing the Pump
Reassembling the Pump Head Assembly
Plungers
130
Repairing the Pump
Exchanging the Optional Interface Board
CA UT IO N The interface board is sensitive to electrostatic discharge. Always use the ESD
kit when handling electronic boards.
1 Switch off the preparative pump at the main power switch. Unplug the pump
from main power.
2 Disconnect cables from the interface board connectors.
3 Loosen the screws. Slide out the interface board from the preparative pump.
4 Install the new interface board. Secure screws.
5 Reconnect the cables to the board connector.
131
Repairing the Pump
Exchanging Internal Parts
WAR NI N G The following procedures require opening the main cover of the
preparative pump. Always ensure the preparative pump is
disconnected from the line power when the main cover is removed. The
security lever at the power input socket prevents that the pump cover
is taken off when line power is still connected. To prevent personal
injury, the power cable must be removed from the preparative pump
before opening the pump cover. Do not connect the power cable to the
preparative pump while the covers are removed.
WAR NI N G To disconnect the preparative pump from line, unplug the power cord.
The power supply still uses some power, even if the switch on the front
panel is turned off.
WAR NI N G NEVER remove the cover of the power supply. Charged capacitor
devices may still contain voltages hazardous for the operator’s health,
even if the pump is NOT connected to line power. The power supply
does not contain any user serviceable parts.
NOTE The electronics of the preparative pump will not allow operation of the
preparative pump when the top cover and the top foam are removed. A safety
light switch on the main board will inhibit the operation of the preparative
pump. Always operate the preparative pump with the top foam and top covers
in place.
132
Repairing the Pump
Exchanging Internal Parts
WAR NI N G When opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
The procedures in this section describe how to exchange internal parts. You
must remove the preparative pump from the stack in order to open the main
cover.
133
Removing the Top Cover and Foam
Preparations for this Switch off preparative pump at the main power switch and unplug the power cable.
procedure Disconnect the solvent inlet tubes from the inlet ball valve.
Beware of leaking solvents due to hydrostatic flow.
Remove leak funnel with the waste tube.
Remove the solvent cabinet from the preparative pump.
1 Remove the front cover by pressing the clip 2 Unplug the power cord and move the lever
fastener on both sides of the cover. towards the power socket.
(1)
Cover clip
Clip fastener (1)
134
Repairing the Pump
Removing the Top Cover and Foam
3 Lift the clips on both sides of the top cover 4 Unscrew the screws on the top plate and
(1). Remove the top cover (2). remove the plate by lifting its back first and
then sliding to the front.
(2)
(2) Screws
(1)
Cover (1)
135
Repairing the Pump
Exchanging the Multi Assembly
1 Remove top covers and foam (see “Removing 2 Unscrew the z-plane and lift it out of the
the Top Cover and Foam” on page 134). housing of the preparative pump.
136
Repairing the Pump
Exchanging the Multi Assembly
3 Remove all capillaries connected to the multi 4 Loosen the two holding screws of the multi
assembly. assembly.
(2) (3)
Screws
(1)
5 Disconnect the pressure sensor cable (P29) 6 Perform the opposite steps in reverse order to
from the PPM board and lift out the multi install a new multi assembly into the pump.
assembly from the pump.
137
Repairing the Pump
Exchanging the Preparative Pump Main Board (PPM Board)
1 Turn off the pump, disconnect all cables and remove the pump from the stack.
2 Remove the top covers and foam, see “Removing the Top Cover and Foam” on
page 134.
3 Disconnect all connectors from the main board.
NOTE When removing connectors, counterhold on connector P25 with one hand.
138
Repairing the Pump
Exchanging the Preparative Pump Main Board (PPM Board)
P4 P5 P6 P7
S1
P2 P3
P10
P8
P13
P17
P18 P19
P22
P23 P20
P25
P28 P29 P30 P27
4 Remove the connector screws from the remote connector and the additional
holding screw between the CAN interface ports and the CAN-DC OUT.
5 Remove the board. Place it on an ESD kit.
WAR NI N G The RFI spring plate sitting on the board connectors is very sharp! Be
careful, not to cut yourself, when removing it from the old board and
sliding it onto the new board.
6 In most cases the RFI spring plate (Radio Frequency Interference spring plate,
prevents radio emissions from the instrument to environment) remains on the
interface connectors of the board. Carefully remove the spring plate from the
old board and slide onto the new board before installing the new board in the
pump (the RFI spring plate is NOT part of an exchange board).
7 On the new board check the switch setting of address switch S1, see “Setting
the 8-bit Configuration Switch” on page 215.
139
Repairing the Pump
Exchanging the Preparative Pump Main Board (PPM Board)
NOTE An incorrect switch setting (e.g., TEST/BOOT) may cause the pump to turn
into a basic mode (yellow or red flashing status light). In such a case turn off
the pump, re-set the address switches, and turn on the pump again.
8 Install the new board and reconnect the connectors. Make sure that the board
is fitted correctly in the board recess holes at the rear panel.
9 Replace the connector and holding screws.
Screws
10 Replace foam and top covers, see “Replacing the Top Cover and Foam” on
page 151.
11 Reinstall the pump in the stack. Reconnect all cables. Turn on the preparative
pump. Continue with the section “Entering the Serial Number” on page 140.
140
Repairing the Pump
Exchanging the Preparative Pump Main Board (PPM Board)
3 Press FW Update (F5). Now, press the m key. This will display a box which
says Update Enter Serial#.
4 Press Enter. This will display the box labeled Serial#.
5 Letters and numbers are created using the up and down arrows. Into the box
labeled Serial#, enter the 10-character serial number for the preparative
pump. When the 10-character serial number is entered, press Enter to
highlight the complete serial number. Then, press Done (F6).
NOTE For firmware revisions below A02.00 it is very important never to press Done
if the Serial# box is blank. In this case, the module can no longer be recognized
by either the control module or the ChemStation. The main board must then
be replaced.
6 Turn the micro pump off, then on again. The Records screen should display the
correct serial number.
7 If a ChemStation is also connected, re-boot the ChemStation now as well.
NOTE The first two characters are letters, which should be capitalized.
The reply line will respond with RA 0000 SER followed by the module serial
number you just entered.
2 Turn off the preparative pump, then on again. Then, re-boot the ChemStation.
If the serial number you have just entered is different than the original module
serial number, you will be given the opportunity to edit the configure 1100
access screen during the re-boot of the ChemStation.
141
Repairing the Pump
Exchanging the Preparative Pump Main Board (PPM Board)
3 After boot-up, the serial number you have just entered can be seen under the
Instrument menu of the main user interface screen. The serial number of the
preparative pump can also be seen by typing the following command into the
command line:
print sendmodule$(lpmp, "ser?") The reply line will give the serial number of
the preparative pump.
http://www.agilent.com/chem
142
Repairing the Pump
Exchanging the Fan
1 Remove the top covers (see “Removing the Top Cover and Foam” on page
134).
2 Disconnect the fan cable (P95) at the main board and move the wires from the
multi assembly and the EMPV out of the way. Slide the fan upwards and move
it out of the unit.
3 Place the new fan into the recess. Make sure that the air flow is directed as
indicated (arrow on the fan points in the same direction as on the foam).
Connect the cable to the main board (P95).
4 Replace the top foam section, optional interface board, front cover, and top
cover (see “Replacing the Top Cover and Foam” on page 151)
143
Repairing the Pump
Exchanging or Maintaining the Pump Drive
WAR NI N G This is not a user-performable task. This part does not contain any
user serviceable parts. For safety reasons it is absolutely forbidden
that this task is performed by a customer! To avoid personal injury call
an Agilent Technologies service engineer or any service engineer
authorized by Agilent Technologies to perform this task.
144
Repairing the Pump
Exchanging Power Supply
WAR NI N G This is not a user-performable task. This part does not contain any
user serviceable parts. For safety reasons it is absolutely forbidden
that this task is performed by a customer! To avoid personal injury call
an Agilent Technologies service engineer or any service engineer
authorized by Agilent Technologies to perform this task.
WAR NI N G NEVER remove the cover of the power supply. Charged capacitor
devices may still contain voltages hazardous for the operator’s health,
even if the pump is NOT connected to line power. The power supply
does not contain any user serviceable parts.
145
Repairing the Pump
Exchanging the Leak Sensor
WAR NI N G This is not a user-performable task. This part does not contain any
user serviceable parts. For safety reasons it is absolutely forbidden
that this task is performed by a customer! To avoid personal injury call
an Agilent Technologies service engineer or any service engineer
authorized by Agilent Technologies to perform this task.
146
Repairing the Pump
Exchanging the Seal Wash Pump Drive
WAR NI N G This is not a user-performable task. This part does not contain any
user serviceable parts. For safety reasons it is absolutely forbidden
that this task is performed by a customer! To avoid personal injury call
an Agilent Technologies service engineer or any service engineer
authorized by Agilent Technologies to perform this task.
147
Repairing the Pump
Exchanging Power Switch or Light Pipe
WAR NI N G This is not a user-performable task. This part does not contain any
user serviceable parts. For safety reasons it is absolutely forbidden
that this task is performed by a customer! To avoid personal injury call
an Agilent Technologies service engineer or any service engineer
authorized by Agilent Technologies to perform this task.
148
Repairing the Pump
Exchanging Status Light Pipe
1 The status light pipe is clipped into the top 2 Replace the top cover, see “Replacing the Top
cover. Cover and Foam” on page 151.
3 Replace the preparative pump into the stack
and reconnect the cables and capillaries.
4 Turn on the preparative pump.
149
Repairing the Pump
Assembling the Main Cover
NOTE The cover kit contains all parts, but it is not assembled.
WAR NI N G In case you insert the left or right left / right side in the wrong
orientation or into the wrong side of the top cover, you may not be able
to remove the side from the top part.
1 Place the top part on the bench and insert the 2 Replace the cover.
left and right side into the top part.
Front
3 Replace the preparative pump into the stack and reconnect the cables and capillaries.
4 Turn on the preparative pump.
150
Repairing the Pump
Replacing the Top Cover and Foam
1 Replace the top foam section. 2 Make sure that the foam is installed correctly.
151
Repairing the Pump
Replacing the Top Cover and Foam
3 Location of foam in the light switch. 4 Replace the optional interface board or the
board cover plate.
Light switch
5 Replace the metal cover (slide the metal tabs 6 Replace the top cover.
into place (1) underneath the Z-Panel in the
front, then lower the back of the metal plate
(2)) and fix the two holding screws.
(2)
(1)
Holding
screws
152
Repairing the Pump
Replacing the Top Cover and Foam
7 Ensure clips are seated correctly, move lever 8 Replace the leak funnel with the waste tube.
back. Locate the lower end of the waste tube in the
holder of the leak pan.
Waste funnel
Lever
9 Place the preparative pump back to its Replace the front covers and reconnect the
position in the stack, replace solvent cabinet power cable to the preparative pump.
and reconnect all tubes and capillaries. Switch on the preparative pump.
153
Repairing the Pump
Replacing the Top Cover and Foam
154
5
Table 10 Repair Parts — Pump Housing and Main Assemblies (Front View)
156
Parts and Materials
Pump Housing and Main Assemblies
1 6
5
8
3
2
4
12
11
10
13
157
Parts and Materials
Pump Housing and Main Assemblies
Table 11 Repair Parts — Pump Housing and Main Assemblies (Rear View)
3
4
158
Parts and Materials
Solvent Cabinet and Bottle-Head Assembly
8
1
7
2
3 5
9, 10
4
159
Parts and Materials
Hydraulic Path
Hydraulic Path
1 Tubing from solvent bottle, see “Solvent Cabinet and Bottle-Head G1361-60002
Assembly” on page 159)
2, 3 Tube clip, Leak funnel 5041-8387, 5041-8388
4 Leak drainage, same as item 7 (re-order 5 m) 5062-2463
5 Prep Capillary 3 (basis kit), EMPV out to next module, G1361-67302
SST 1/16” o.d., 0.6 mm i.d., 400 mm long
5a Prep Capillary 4, multi assembly to EMPV, G1361-67303
SST 1/16” o.d., 0.5 mm i.d., 160 mm long
6 Tubing PTFE 0.057” i.d., EMPV to waste (re-order 5 m) 5062-2461
7 Corrugated waste tube, 120 cm (re-order 5 m) 5062-2463
8 Prep Capillary 1, Outlet valve 1 to multi assembly top, G1361-67300
SST 1/16” o.d., 0.6 mm i.d., 173 mm long
9 Prep Capillary 2, Outlet valve 2 to multi assembly bottom, G1361-67301
SST 1/16” o.d., 0.6 mm i.d., 178 mm long
10 Outlet ball valve, see “Pump-Head Assembly” on page 162
11 Inlet ball valve, see “Pump-Head Assembly” on page 184
12 Tubing, Seal wash to waste 0890-1764
13 Tubing, Seal wash solvent bottle to seal wash pump 0890-1764
14 Tubing, Seal wash pump to pump head 0890-1764
Tube adapter (for item 13 and 14) 5042-6422
15 Prep Capillary 5 (gradient kit), Union to EMPV2 G1361-67304
SST 1/16” o.d., 0.6 mm i.d., 111 mm long
16 Prep Capillary 6 (gradient kit), Union to mixer G1361-67305
SST 1/16” o.d., 0.6 mm i.d., 40 mm long
17 Prep Capillary 7 (gradient kit), EMPV1 to union G1361-67306
SST 1/16” o.d., 0.6 mm i.d., 367 mm long
160
Parts and Materials
Hydraulic Path
12 4
10
5a
11
5
13 6
14
7
15
Gradient interconnection
Pump 1
EMPV 1
17
Pump 2
EMPV 2 Union 18
15
16
Mixer Column
5
161
Parts and Materials
Pump-Head Assembly
Pump-Head Assembly
NOTE Item 6 (plunger housing) and 14 (screw) are not available for replacement at
introduction. A complete assembly will be made available at a later stage.
162
Parts and Materials
Pump-Head Assembly
14
7
1
13
6
10 8
9
3
2 8
12
11
163
Parts and Materials
Electro Mechanical Proportional Valve (EMPV)
3
All fittings 1/16”
2
Fitting FRONT, Fitting
5 BACK, Fitting 1/16”,
re-order 10/pk
5062-2418
164
Parts and Materials
Multi Assembly Parts
Multi assembly with filters and pressure sensor (at the rear) G1361-60005
1 Filter screw assembly, includes item 2 G1361-60006
2 Holder filter G1361-22300
3 Filter cup 3150-0942
4 Filter screw G1361-44601
5, 6 SST Filter assembly 2 µm 5022-2192
Note:
5 Item 6 is part of item 5
165
Parts and Materials
Cover Parts
Cover Parts
1 Plastic cover kit (includes top, base and both sides) G1361-68703
2 Front plate G1361-68704
3 Logo plate, Agilent 1100 5042-1381
166
Parts and Materials
Sheet Metal Kit
* Included in item 1
167
Parts and Materials
Foam Parts
Foam Parts
upper foam
lower foam
transport foam
168
Parts and Materials
Preparative Pump Basis Kit
Table 20 Basis Kit Contents G1361-68708 (shipped with the first pump of a system)
169
Parts and Materials
Preparative Pump Gradient Kit
Table 21 Gradient Kit Contents G1361-68707 (shipped with every additional pump in a
gradient system)
170
Parts and Materials
Control Module (G1323B)
171
Parts and Materials
Power and Status Light Pipes
3
4
172
Parts and Materials
Leak Parts
Leak Parts
2 1
5
4
173
Cable Overview
WAR NI N G Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
174
Parts and Materials
Cable Overview
175
Parts and Materials
Analog Cables
Analog Cables
1 Shield Ground
2 Not connected
3 Center Signal +
4 Connected to pin 6
5 Shield Analog -
6 Connected to pin 4
7 Key
8 Not connected
1 Not connected
2 Shield Analog -
3 Center Analog +
176
Parts and Materials
Analog Cables
1 Not connected
2 Black Analog -
3 Red Analog +
177
Parts and Materials
Remote Cables
Remote Cables
178
Parts and Materials
Remote Cables
NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector.
179
Parts and Materials
Remote Cables
180
Parts and Materials
Remote Cables
181
Parts and Materials
Remote Cables
182
Parts and Materials
BCD Cables
8 7 6 5 4 3 2 1
BCD Cables 15 14 13 12 11 10 9
10 1 BCD 5 20
11 2 BCD 7 80
3 3 BCD 6 40
9 4 BCD 4 10
7 5 BCD 0\ 1
5 6 BCD 3 8
12 7 BCD 2 4
4 8 BCD 1 2
6 - Key
1 9 Digital ground
2 15 +5V Low
183
Parts and Materials
BCD Cables
1 1 BCD 5 20
2 2 BCD 7 80
3 3 BCD 6 40
4 4 BCD 4 10
5 5 BCD 0\ 1
6 6 BCD 3 8
7 7 BCD 2 4
8 8 BCD 1 2
9 9 Digital ground
NC 15 +5V Low
Connector Pin
18594-60520 Wire Color Agilent 1100 Signal Name BCD Digit
Green 1 BCD 5 20
Violet 2 BCD 7 80
Blue 3 BCD 6 40
Yellow 4 BCD 4 10
Black 5 BCD 0\ 1
Orange 6 BCD 3 8
Red 7 BCD 2 4
Brown 8 BCD 1 2
Gray 9 Digital ground
White 15 +5 Vt Low
184
Parts and Materials
Auxiliary Cable
Auxiliary Cable
Connector Pin
G1322-81600 Color Agilent 1100 Signal Name
White 1 Ground
Brown 2 Pressure signal
Green 3
Yellow 4
Grey 5 DC + 5 V IN
Pink 6 Vent
CAN Cable
185
Parts and Materials
External Contact Cable
5 1
External Contact Cable 10
15 11
6
One end of this cable provides a 15-pin plug to be connected to Agilent 1100
Series module’s interface board. The other end is for general purpose.
Connector Pin
G1103-61611 Color Agilent 1100 Signal Name
White 1 EXT 1
Brown 2 EXT 1
Green 3 EXT 2
Yellow 4 EXT 2
Grey 5 EXT 3
Pink 6 EXT 3
Blue 7 EXT 4
Red 8 EXT 4
Black 9 Not connected
Violet 10 Not connected
Grey/pink 11 Not connected
Red/blue 12 Not connected
White/green 13 Not connected
Brown/green 14 Not connected
White/yellow 156 Not connected
186
Parts and Materials
RS-232 Cable Kit
This kit contains a 9-pin female to 9-pin female Null Modem (printer) cable
and one adapter. Use the cable and adapter to connect Agilent Technologies
instruments with 9-pin male RS-232 connectors to most PCs or printers.
Instrument PC
DCD 1 1 DCD
RX 2 2 RX
TX 3 3 TX
DTR 4 4 DTR
GND 5 5 GND
DSR 6 6 DSR
RTS 7 7 RTS
CTS 8 8 CTS
RI 9 9 RI
187
Parts and Materials
LAN Cables
LAN Cables
Recommended Cables
For point to point connection (not using a network hub) use a twisted pair
cross over LAN cable (P/N 5183-4649, 10 feet long).
For standard network connections using a hub use category 5 UTP cables,
(P/N G1530-61480, 8 m long).
188
6
6 Introduction to the
Preparative Pump
The preparative pump is a cam driven pump, with fixed stroke and two
parallel pistons, each with (identical ball and seat) inlet and outlet valve.
Each of the two pump pistons is capable of delivering a maximum of
50 ml/min flow at up to 400 bar pressure.
The two cams have overlapping delivery cycles. Compression losses are
calculated and smooth delivery is ensured through variation of the motor
speed over the pump cycle.
The pistons are guided through two seals each. The gap between those seals
is connected to capillary fittings, this way seal wash is not only an option, but
integral part of every pump. A peristaltic pump, to automate seal wash, is
also part of the system.
190
Introduction to the Preparative Pump
Introduction to the Preparative Pump
PPM board
Pump head
Inlet valves
Electromagnetic
Leak plane purge valve
Leak sensor
Seal wash pump
Multi
Solvent
Assy
selection
with filters
valveand
pressure sensor (at the rear)
191
Introduction to the Preparative Pump
Introduction to the Preparative Pump
192
Introduction to the Preparative Pump
Introduction to the Preparative Pump
Multi-Assembly
with 2 filters and
Piston 2 pressure sensor
To Waste
193
Introduction to the Preparative Pump
Introduction to the Preparative Pump
194
Introduction to the Preparative Pump
Electrical Connections
Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal Agilent 1100 Series
module data transfer and synchronization.
• The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
common shut down, prepare, and so on.
• The RS-232 connector may be used to control the preparative pump from
a computer through an RS-232 connection, using appropriate software.
This connector needs to be activated by the 8-bit configuration switch
module in the upper right corner of the rear of the pump. The software
needs the appropriate drivers to support this communication. See your
software documentation for further information.
• The power input socket accepts a line voltage of 100 – 240 Volts AC ± 10 %
with a line frequency of 50 - 60 Hz. max. power consumption is 250 VA
(Volt-Amps). There is no voltage selector on your preparative pump
because the power supply has wide-ranging capability. There are no
externally accessible fuses. The security lever at the power input socket
prevents that the preparative pump cover is taken off when line power is
still connected.
• The interface (BCD) board slot is used for external contacts, BCD output
and for LAN communication.
• The CAN-DC-out provides 24 Volts DC power for external CAN devices
like a switch valve. Max. permanent power consumption is
100 mA/channel or 2 A for sec peak consumption.
WAR NI N G Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
WAR NI N G To disconnect the preparative pump from line, unplug the power cord.
The power supply still uses some power, even if the power switch on
the front panel is turned off.
195
Introduction to the Preparative Pump
Electrical Connections
8-bit configu-
ration switch
Remote
CAN-DC-out
RS232C
CAN cable to next module
CAN-bus
196
Introduction to the Preparative Pump
Instrument Layout
Instrument Layout
197
Introduction to the Preparative Pump
Early Maintenance Feedback (EMF)
EMF Counters
The preparative pump provides a series of EMF counters for the pump head.
Each counter increments with pump use, and can be assigned a maximum
limit which provides visual feedback in the user interface when the limit is
exceeded. Each counter can be reset to zero after maintenance has been
done. The preparative pump provides the following EMF counters:
• liquimeter pump head,
• seal wear pump head.
Liquimeter
The liquimeter displays the total volume of solvent pumped by the pump
head since the last reset of the counters. The liquimeter can be assigned an
EMF (maximum) limit. When the limit is exceeded, the EMF flag in the user
interface is displayed.
198
Introduction to the Preparative Pump
Early Maintenance Feedback (EMF)
199
Introduction to the Preparative Pump
The Electronics
The Electronics
200
Introduction to the Preparative Pump
The Electronics
Motor Drive
The drive comprises motor control and motor amplifier.
Pressure Converter
This block comprises a filter and amplifier for the pressure-sensor-signal, an
A/D converter with integrated multiplexer.
Leak Converter
This block comprises a PTC for the leak identification and a NTC for the
ambient temperature measurement. This assures that temperature changes
are not identified as leak. A leak would cool down the PTC and its change in
resistance results in a leak signal.
Fan Drive
The revolution of the fan is controlled by the main processor depending on
the internal heat distribution in the preparative pump. The fan provides a
PWM signal which is proportional to the revolution. This fan status signal is
used for diagnostics.
201
Introduction to the Preparative Pump
The Electronics
Index Board
The Index board is used to sense the movement and positioning of the pump
drive. It also comprises calibration information of the drive.
Electronic Fuses
The valve circuit is electronically fused on the board. Any error on the board
or shortages of the valve will activate the electronic fuse that will switch off
the supply voltage. This prevents the damage of components.
Onboard Battery
An onboard lithium battery buffers the electronic memories when the
preparative pump is turned off.
For safety information on lithium batteries, see “Lithium Batteries
Information” on page 254.
Interfaces
The preparative pump provides the following interfaces:
• two CAN connectors as interface to other Agilent 1100 Series modules,
• one RS-232C as interface to a computer,
• one REMOTE connector as interface to other Agilent products,
• one optional interface board.
• two CAN-DC-out for external CAN devices like a switch valve,
• LAN.
202
Introduction to the Preparative Pump
The Electronics
to A/D Converter
1 (see next page)
Encoder
Optocoupler
Transceiver (TAG)
2 to A/D Converter
Processor
(see next page)
FPGA
Seal wash pump Drive Control
ASIC
CAN-DC OUT
ON/OFF
CAN-DC OUT
203
Introduction to the Preparative Pump
The Electronics
ASIC
Amplifier &
ambient meas. Leak sensor
Drive Fan
Sensor
Control Signal
Data/Address
Diagnose signals
ϑ
204
Introduction to the Preparative Pump
The Electronics
P2
CAN P4, P5 CAN-DC OUT
RS232 P6 Power
+24VCAN
Remote P7 switch
P3
MIO P25 CAN-DC OUT
205
Introduction to the Preparative Pump
Firmware Description
Firmware Description
Resident System
This resident section of the firmware is identical for all Agilent 1100 series
modules. Its properties are:
• the complete communication capabilities (CAN, LAN and RS-232C),
• memory management,
• ability to update the firmware of the ‘main system’.
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN and RS-232C),
• memory management,
• ability to update the firmware of the ‘resident system’.
In addition the main system comprises the instrument functions that are
divided into common functions like
• run synchronization via APG remote
• error handling,
• diagnostic functions and so on,
or module specific functions like
• internal events such as motor control, flow rates and so on,
• calculation of compensation values for compressibility.
206
Introduction to the Preparative Pump
Firmware Description
Firmware Updates
Firmware updates can be done using your user interface:
• handheld control module with files from a PC-card or
• Agilent ChemStation with files from floppy disk
The file naming conventions are:
xxxxvvvv.DLB, where
xxxx is the product number, e.g. 1361 for the G1361A preparative pump), and
vvvv is the revision number, for example A402 is revision A.04.02.
For instructions refer to your user interface.
NOTE Update of main system can be done in the resident system only.
Update of the resident system can be done in the main system only.
main FW update
Resident System Main System
resident FW update
207
Introduction to the Preparative Pump
Optional Interface Boards
The Agilent 1100 Series modules have one optional board slot that allows to
add an interface board to the modules.
BCD Board
The BCD board provides a BCD output for the bottle number of the
Agilent 1100 Series autosampler and four external contacts. The external
contact closure contacts are relay contacts. The maximum settings are: 30 V
(AC/DC); 250 mA (fused). There are general purpose cables available to
connect the BCD output (see “BCD Cables” on page 183) and the external
outputs (see “Remote Cables” on page 178) to external devices.
Board
identification
RFI filter
12 BCD
Processor
BCD register
interface connector
Line driver
RFI filter
+ 250 mA
External External
contacts 4× contact
connector
208
Introduction to the Preparative Pump
Optional Interface Boards
LAN Board
The HP LAN board is actually an HP JetDirect card, which is a network
interface card used in HP printers.
NOTE One board is required per Agilent 1100 stack. If the Agilent 1100 stack
has a DAD, then the DAD MUST be the module used for LAN board
installation. If no DAD is present a pump should be used for the LAN
board installation.
A DAD, VWD, standard Pumps and ALS with main board version G13XX-66520
and greater and ChemStation software revision A.06.01 or above. The
following modules can be used with main board version G13XX-66500 and
greater and ChemStation revisions as indicated:
FLD A.06.03 or above,
RID, MWD A.06.04 or above,
WPS A.08.03 or above,
Cap. LC System A.08.03 or above,
Prep. Pump A.08.04 or above.
The following boards can be used with the Agilent 1100 modules.
209
Introduction to the Preparative Pump
Optional Interface Boards
Recommended Cables
For point to point connection (not using a network hub) use a twisted pair
cross over LAN cable (P/N 5183-4649, 10 feet long).
For standard network connections using a hub use category 5 UTP cables,
(P/N G1530-61480, 8 m long).
210
Introduction to the Preparative Pump
Agilent 1100 Series Interfaces
DA Detector Thermostatted
Well Plate MW Detector VW Detector Column Vacuum
Interface Type Pumps Autosampler Sampler FL Detector RI Detector Compartment Degasser
*
No for the G1361A Preparative Pump
**
The well plate sampler can be controlled through CAN via a detector that is connected via GPIB to a ChemStation.
*** No for the G1361A Preparative Pump
****
The vacuum degasser will have a special connector for specific use. For details see description of main board.
• The CAN connectors serve as the interface between the Agilent 1100
Series modules themselves,
• the GPIB connector serves as the interface between an Agilent1100
module and the Agilent ChemStation,
• RS-232C provides an interface to a computer,
• the APG remote connector serves as the interface between the
Agilent 1100 and other Agilent (APG remote compatible) products,
• the analog output connector(s) provide an analog signal output,
• the interface board slot (not common to all modules) provides specific
interfacing needs (external contacts, BCD, LAN and so on).
For identification and location of the connectors see Figure 31 on page 139.
211
Introduction to the Preparative Pump
Agilent 1100 Series Interfaces
WAR NI N G Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations, see “Cable Overview” on page 174.
CAN Interface
The CAN is an inter module communication interface. It is a 2 wire serial bus
system supporting high speed data communication and real-time
requirements.
212
Introduction to the Preparative Pump
Agilent 1100 Series Interfaces
213
Introduction to the Preparative Pump
Agilent 1100 Series Interfaces
RS-232C
The RS-232C connector is used to control the instrument from a computer
through RS-232C connection, using the appropriate software. This connector
can be configured with the configuration switch module next to the CAN-DC
OUT connector. The RS-232C is designed as DCE (data communication
equipment) with a 9-pin male SUB-D type connector. The pins are defined as:
1 In DCD
2 In RxD
3 Out TxD
4 Out DTR
5 Ground
6 In DSR
7 Out RTS
8 In CTS
9 In RI
Instrument PC
DCD 1 1 DCD
RX 2 2 RX
TX 3 3 TX
DTR 4 4 DTR
GND 5 5 GND
DSR 6 6 DSR
RTS 7 7 RTS
CTS 8 8 CTS
RI 9 9 RI
214
Introduction to the Preparative Pump
Setting the 8-bit Configuration Switch
The 8-bit configuration switch is located in the upper right corner of the rear
of the instrument, next to the CAN-DC OUT connectors. Switch settings
provide configuration parameters for serial communication protocol and
instrument specific initialization procedures.
The factory default setting for the preparative pump is:
all switches set to 0 (down).
Mode Select 1 2 3 4 5 6 7 8
Switches 1 and 2 define which set of parameters (for example, for RS-232C)
will be changed. Once the change has been completed, the instrument must
be powered up again in order to store the values in the non-volatile memory.
215
Introduction to the Preparative Pump
Setting the 8-bit Configuration Switch
Mode Select 1 2 3 4 5 6 7 8
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up
3 4 5 3 4 5
216
Introduction to the Preparative Pump
Setting the 8-bit Configuration Switch
0 7 Bit Communication
1 8 Bit Communication
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
Switches Parity
7 8
0 0 No Parity
0 1 Odd Parity
1 0 Even Parity
CA UT IO N A forced cold start erases all methods and data stored in non-volatile memory.
Exceptions are diagnose and repair logbooks which will not be erased.
217
Introduction to the Preparative Pump
Setting the 8-bit Configuration Switch
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Mode Select 1 2 3 4 5 6 7 8
TEST/BOOT 1 1 0 0 0 0 0 1
Mode Select 1 2 3 4 5 6 7 8
TEST/BOOT 1 1 0 0 1 0 0 0
218
Introduction to the Preparative Pump
The Main Power Supply Assembly
WAR NI N G To disconnect the instrument from line, unplug the power cord. The
power supply still uses some power, even if the power switch on the
front panel is turned off.
219
Introduction to the Preparative Pump
The Main Power Supply Assembly
220
7
NOTE The screens shown on the next pages are based on the following firmware
revisions:
Control Module firmware revision B.02.0x (G1323B).
HPLC Module firmware revision 4.1x.
NOTE In case the control module’s display seems to be frozen (hang-up due to a
communication problem on the CAN bus) unplug the control module from the
HPLC module and reconnect.
222
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Analysis view This is the wake-up screen, if the Agilent 1100 preparative pump is the only
configured Agilent 1100 module. It is used to enter the most common pump
method parameters.
The m-key allows access to the context sensitive menus. Date&Time allows
you to change time settings. Print Screen gives access to the print screen.
About tells you the current firmware revision and the serial# of your control
module. Setup view leads you to analysis view configuration for additional
Agilent 1100 modules. Restart re-boots the control module.
223
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Setup View In the Setup view, modules can be added or removed to the analysis view.
Select a module, you want to move, with the help of the selection keys. Use
function keys F7/F8 (Remove/Add) to move the highlighted module.
Changes must be activated with Done (F6).
224
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
On/Off From the Analysis view use the F7-key (On/Off) to proceed to the Turn On
screen. Press F8 (On) once to turn on the pump. If more than one module is
available, select F8 (Pump) or turn on the entire system by selecting the
System On button and pressing Enter.
Settings With the F1-key (Settings) you open a pull-down menu where you can select
the preparative pump module with the help of the selection or
alphanumeric keys.
225
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Settings In the Settings screen you can change the pump parameters. Here, you can
enter values for the flow or composition (for gradient prep. pump, only). In
addition, you have access to a different set of parameters available through
the F1-5 navigation-keys. The F7-key (Default) resets the pump to default
values. F8 (On/Off) opens a window to turn on the pump. The following
figures show the screens for an isocratic preparative pump or a gradient
preparative pump. Changes must be acknowledged with F6 (Done).
Use the m-key for the context sensitive menu. Reset will load the pump
default parameters. The Status command pulls up the module specific setup
screen.
226
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Settings - More... Select Flow Ramp / Compr. to enter appropriate for the flow ramp up speed
and the solvent compressibility (see “How to Optimize the Compressibility
Compensation Setting” on page 47). Changes must be activated with F6
(Done).
227
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Settings - Timetable With the F2-key (Timetable) you can list and edit the timetable for the
pump. Press F7-key (Insert) to add entries, F6-key (Delete) to remove or
Enter to edit selected/highlighted entries .
Edit the different sections of each timetable line. Use the pull-down menus, if
available. Use the F6-key (Done) to view the entered lines of the timetable or
F7 (Enter) to acknowledge current settings.
Use the m-key for the context sensitive menu. It gives you additional tools for
the timetable.
228
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Settings - Pressure With the F3-key (Pressure) you can change the settings for the pressure
limits.
Settings - Bottle With the F4-key (Bottle Fillings) you can adjust the settings for the bottle
Fillings fillings to their current state. Select F8 (Refill) to refill All bottles or a
Selected bottle to total volume. By selecting the respective tick-boxes, you
can also choose whether and when to have Not-Ready or Error conditions.
The following figures show the screens for an a binary gradient preparative
pump. Changes must be activated with F6 (Done).
229
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Settings - Runtimes With the F5-key (Runtimes) you can change the stop time and the post-run
time for this module individually. Changes must be activated with F6 (Done).
Status View This is an example if an Agilent 1100 binary gradient preparative pump is
configured standalone. Information on the actual flow rate, mobile phase
composition, pressure and %-Ripple, elapsed run time and the pressure plot
are shown. Press key F8 (Start) to start a run, key F7 (Rescale) to
maximize the signal.
230
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
231
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Signal plot Press F6-key (Plot) to enter the plot screen (available also from all other
views). Here you can observe online signal(s). If more than one signal is
configured, use the 1-2-3 numeric keys to switch between the signals.
The m-key allows you to Clear All signals, Maximize signals and Print the
plot.
232
Control Module Screens for the Preparative Pump
Screens available from the Analysis view
Method screens On the Analysis view use the F3-key (Method) to view the parameters in a
method. Use the F8-key (Save As) to save the method in the module(s). The
PC-Card-key (F2) is only active when a PCMCIA card is inserted in the
control module.
NOTE The PCMCIA card is only recognized, if inserted before last start-up.
Use F2-key (PC-Card) to transfer a method from/to a PCMCIA card. Use the
Right/Left selection keys to switch between PC-Card and Instrument
window. Use the Up/Down selection keys to select the method. Use the
F7/F8-keys (Copy) or Enter to move available methods between PCMCIA
card and Modules. F6 Deletes a selected method.
233
Control Module Screens for the Preparative Pump
Screens available from the System view
System view Use the Esc-key to until you receive Views on the F5-key. Choose System
from the pull-down menu. This screen shows the last activities in the system.
System - Control Use the F1-key (Control) and select the preparative pump.
234
Control Module Screens for the Preparative Pump
Screens available from the System view
In the preparative pump’s Control screen, you can turn On or Off (F8) the
pump, or Reset (F7) the pump. Here you receive information about the
not-ready conditions if needed.
Use F1 (Purge Task) to access the pump’s Purge Task. After selecting a
channel (gradient system, only), you can edit its Purge Time. Changes must
be activated with F6 (Done).
235
Control Module Screens for the Preparative Pump
Screens available from the System view
System - On the System view use the F2-key (Configure) and select the pump.
Configuration
This screen gives you information about installed options and allows you to
configure your pump to the filter and mixer type that is actually installed in
the pump. Changes must be activated with F6 (Done).
236
Control Module Screens for the Preparative Pump
Screens available from the System view
Press F5 (Gradient Pump) in the Config screen. Here you can configure,
which of the two isocratic preparative pumps that form a gradient
preparative pump represents channel A, which represents channel B. In case
you are not sure, you can press F8 (Identify) to identity, which pump
represents which channel. The status indicator of the respective instrument
will then start to blink. By pressing F7 (Break) you can separate the two
combined pumps and use each of them as an individual module. Changes
must be activated with Done (F6).
237
Control Module Screens for the Preparative Pump
Screens available from the Records screen
Records screen Use the Esc-key until you receive Views on the F5-key. Choose System from
the pull-down menu. Use the F4-key (Records) and select the pump. Errors
are reported either into the F2 (System Log) or F3 (Error Log).
EMF (Early Use the F1-key (EMF) to set EMF Limits. Choose menu item Setup Limits
Maintenance to set the amount of pumped solvent or seal wear, when you want to receive
Feedback) a warning for exceeded limits.
238
Control Module Screens for the Preparative Pump
Screens available from the Records screen
Press F7 (Reset) to reset the counters after exchanging parts that are
subject to wear. Changes must be activated with F6 (Done).
If a set limit has been exceeded, a message box will pop up.
NOTE An exceeded limit will not stop a sequence or run (information only, to plan
maintenance activities).
If you press Reset, the limits will be removed. Ignore will continue to keep
the EMF flag set.
System / Error Log Use the F2-key (System Log) or F3-key (Error Log) to look for errors on
the selected/ highlighted module.
239
Control Module Screens for the Preparative Pump
Screens available from the Records screen
Maintenance Log Use the F4-key (Maint Log) to review and enter maintenance activities in
the maintenance logbook.
Use the F7-key (Add) to add new maintenance activities. You can add a
comment to the activity into the line ‘Add’ using the control modules
alphanumeric keys. Changes must be activated with F6 (Done).
240
Control Module Screens for the Preparative Pump
Screens available from the Records screen
Firmware Update Use the Esc-key to receive Views on the F5-key. Choose System from the
pull-down menu. Use the F3-key (Records) to select the pump. Use the
F5-key (FW Update) to enter the Update section. If you want to update the
resident firmware (together with specific main firmware revisions), select
the file from the PCMCIA card (RESnnnn.DLB) and press execute. If you
want to update the main firmware, press F7-key (Transfer) to turn the
module into the resident mode (LED on module should blink yellow).
Use the Esc-key to receive Views on the F5-key. Choose System from the
pull-down menu. Use the F3-key (Records) to select the Generic module. In
this screen the resident firmware revision is shown.
241
Control Module Screens for the Preparative Pump
Screens available from the Records screen
Use the F5-key (FW Update) to enter the Update section. Select the a file
from the PCMCIA card (1361nnnn.DLB) and press execute.
When the update has finished, press F7-key (Transfer) to return the module
into the normal mode (LED on module should stay yellow).
242
Control Module Screens for the Preparative Pump
Screens available from the Records screen
Changing the serial In case the serial number of the module has to be added, use the m-key to
number open the menu Enter Serial#. The serial number becomes active after
restart of the module.
Type in the module’s Serial# as indicated on the front or rear label of the
module using the alphanumeric keys. Changes must be activated with F6
(Done).
243
Control Module Screens for the Preparative Pump
Diagnostics and Tests
Tests screen Use the Esc-key until you receive Views on the F5-key. Choose System from
the pull-down menu. Use the F3-key (Tests) to select the pump.
Leak Test Press F2 (Leak Test) on the Test screen to perform a leak test. Several steps
like purging the system, setting up Isopropanol as solvent and blocking the
purge valve outlet with a blank nut have to be performed before operating the
leak test. For details use the i-key to achieve context sensitive help, follow
the instructions on the screen and refer to “Running the Leak Test” on page
102. For evaluating the leak test refer to “Evaluating the Results” on page 103.
Pressure Test Use the F3-key (Pressure Test) to perform a pressure test of the system.
Several steps like purging the system, setting up Isopropanol as solvent and
blocking the column outlet with a blank nut have to be performed before
operating the pressure test. For details use the i-key to achieve context
sensitive help, follow the instructions on the screen and refer to “Running the
Pressure Test” on page 95. For evaluating the pressure test refer to
“Evaluating the Results” on page 97.
244
8
8 Specifications
Performance Specifications
Type Specification
246
Legal, Safety and Warranty
Information
Warranty Statement
Warranty Statement
Limitation of Warranty
Onsite warranty services are provided at the initial installation point. Installation and
onsite warranty services are available only in Agilent service travel areas, and only
in the country of initial purchase unless buyer pays Agilent international prices for
the product and services. Warranties requiring return to Agilent are not limited to the
country of purchase.
For installation and warranty services outside of Agilent’s service travel area,
Agilent will provide a quotation for the applicable additional services.
If products eligible for installation and onsite warranty services are moved from the
initial installation point, the warranty will remain in effect only if the customer
purchases additional inspection or installation services, at the new site.
The foregoing warranty shall not apply to defects resulting from:
1 improper or inadequate maintenance, adjustment, calibration, usage of corrosive
solvent as described in the solvent information note later in this chapter or
operation by buyer,
2 buyer-supplied software, hardware, interfacing or consumables,
3 unauthorized modification or misuse,
248
Warranty Statement
249
Warranty Statement
*
This warranty may be modified in accordance with the law of your country. Please consult your
local Agilent office for the period of the warranty, for shipping instructions and for the applicable
wording of the local warranty.
**
Warranty services are included as specified for Analytical products and options purchased
concurrently provided customer is located within a Agilent defined travel area. Agilent warranty
service provides for 8 a.m. to 5 p.m. on-site coverage Monday through Friday, exclusive of Agilent
holidays.
*** Columns and Consumables are warranted to be free from defects for a period of 90 days after
250
Safety Information
Safety Information
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
General
This is a Safety Class I instrument (provided with terminal for protective earthing)
and has been manufactured and tested according to international safety standards.
Operation
Before applying power, comply with the installation section. Additionally the
following must be observed.
Do not remove instrument covers when operating. Before the instrument is switched
on, all protective earth terminals, extension cords, auto-transformers, and devices
connected to it must be connected to a protective earth via a ground socket. Any
interruption of the protective earth grounding will cause a potential shock hazard
that could result in serious personal injury. Whenever it is likely that the protection
has been impaired, the instrument must be made inoperative and be secured against
any intended operation.
Make sure that only fuses with the required rated current and of the specified type
(normal blow, time delay, and so on) are used for replacement. The use of repaired
fuses and the short-circuiting of fuseholders must be avoided.
WAR NI N G Disconnect the instrument from the line and unplug the power cord
before maintenance.
251
Safety Information
Safety Symbols
Table 42 shows safety symbols used on the instrument and in the manuals.
Symbol Description
The apparatus is marked with this symbol when the user should refer
! to the instruction manual in order to prevent risk of harm to the
operator and to protect the apparatus against damage.
Eye damage may result from directly viewing the light produced by
the Xenon flash lamp used in this product. Always turn the xenon flash
lamp off before removing it.
WAR NI N G A warning alerts you to situations that could cause physical injury or
damage to the equipment. Do not proceed beyond a warning until you
have fully understood and met the indicated conditions.
252
Safety Information
CA UT IO N A caution alerts you to situations that could cause a possible loss of data. Do
not proceed beyond a caution until you have fully understood and met the
indicated conditions.
253
Lithium Batteries Information
NOTE Bij dit apparaat zijn batterijen geleverd. Wanneer deze leeg zijn, moet u ze niet
weggooien maar inleveren als KCA.
254
Radio Interference
Radio Interference
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Sound Emission
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the German Sound
Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB.
• Sound Pressure Lp < 70 dB (A)
• At Operator Position
• Normal Operation
• According to ISO 7779:1988/EN 27779/1991 (Type Test)
255
Solvent Information
Solvent Information
Solvents
Brown glass ware can avoid growth of algae.
Always filter solvents, small particles can permanently block the capillaries. Avoid
the use of the following steel-corrosive solvents:
• Solutions of alkali halides and their respective acids (for example, lithium iodide,
potassium chloride, and so on).
• High concentrations of inorganic acids like nitric acid, sulfuric acid especially at
higher temperatures (replace, if your chromatography method allows, by
phosphoric acid or phosphate buffer which are less corrosive against stainless
steel).
• Halogenated solvents or mixtures which form radicals and/or acids, for example:
2CHCl3 + O2 → 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly
with dried chloroform if the drying process removes the stabilizing alcohol.
• Chromatographic grade ethers, which can contain peroxides (for example, THF,
dioxane, di-isopropylether) such ethers should be filtered through dry aluminium
oxide which adsorbs the peroxides.
• Solutions of organic acids (acetic acid, formic acid, and so on) in organic
solvents. For example, a 1-% solution of acetic acid in methanol will attack steel.
• Solutions containing strong complexing agents (for example, EDTA, ethylene
diamine tetra-acetic acid).
• Mixtures of carbon tetrachloride with 2-propanol or THF.
256
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web site on
the Internet at:
http://www.agilent.com
Select “Products” - “Chemical Analysis”
It will provide also the latest firmware of the Agilent 1100 series modules for
download.
257
Agilent Technologies on Internet
258
Index
259
Index
260
Index
261
Index
262
s1
In This Book
• Installation,
• Optimizing performance,
• Repairing,
• Specifications and