FEATURES
• 0.10 % accuracy
• 5 lb to 500000 lb
• mV/V output (standard); 0 Vdc to 5 Vdc or 4 mA to 20 mA
(optional)
• Stainless steel
• Double diaphragm design
Model 43
PERFORMANCE SPECIFICATIONS MECHANICAL SPECIFICATIONS
Characteristic Measure Characteristic Measure
Load ranges9 5 lb to 500000 lb Maximum allowable load 150 % FS2
Non-linearity, 5 lb to 25 lb ±0.2 % full scale Weight See table
Non-linearity, Material 17-4PH stainless steel
±0.1 % full scale
50 lb to 500000 lb Less than 200,000 lb
Hysteresis, 5 lb to 25 lb ±0.1 % full scale Material greater than or Carbon steel
Hysteresis, equal to 300000 lb
±0.08 % full scale
50 lb to 500000 lb Deflection See table
Non-repeatability, 5 to 25 lb ±0.1 % full scale Natural frequency See table
Non-repeatability,
±0.03 % full scale
50 lb to 500000 lb RANGE CODES
Output (tolerance), Range Available ranges Range Available ranges
2 mV/V ±0.5 % full scale
5 lb to 25 lb Code Code
Output (tolerance), AT 5 lb DV 10000 lb
3 mV/V ±0.5 % full scale
50 lb to 500000 lb
AV 10 lb EJ 15000 lb
Operation Compression
BL 25 lb EL 20000 lb
Resolution Infinite
BN 50 lb EN 30000 lb
Standard calibration 5-point calibration, compression only
BR 100 lb EP 50000 lb
CN 250 lb ER 75000 lb
ENVIRONMENTAL SPECIFICATIONS
CR 500 lb ET 100000 lb
Characteristic Measure
CV 1000 lb FJ 150000 lb
Temperature, operating -54 °C to 121 °C [-65 °F to 250 °F]
DL 2000 lb FL 200000 lb
Temperature, compensated 15 °C to 71 °C [60 °F to 160 °F]
DN 3000 lb FN 300000 lb
Temperature effect, zero 0.002 % full scale/°F
DP 4000 lb FP 400000 lb
Temperature effect, span 0.002 % full scale/°F
DR 5000 lb FR 500000 lb
DT 7500 lb
ELECTRICAL SPECIFICATIONS
Characteristic Measure
Strain gage type Bonded foil
DEFLECTIONS AND RINGING FREQUENCIES
Excitation (calibration) 10 Vdc Capacity (lb) Deflection @ Natural ringing Weight
full scale (in) frequency (Hz) kg (lb)
Excitation (acceptable) Up to 15 Vdc or Vac
5 to 25 0.001 2000 0.36 [0.8]
Insulation resistance 5000 mOhm @ 50 Vdc
50 to 1000 0.002 4500 0.68 [1.5]
Bridge resistance 350 ohm (nominal)
(tolerance) 2000 to 5000 0.002 11000 0.73 [1.6]
Zero balance (tolerance) ±1 % full scale 7500 to 30000 0.003 9000 2.90 [6.4]
Shunt calibration data Included 50000 to 100000 0.004 13000 2.99 [6.6]
Electrical termination (std) PTIH-10-6P 150000 to 200000 0.004 11000 5.31 [11.7]
5 lb to 5000 lb 300000 0.005 10000 10.8 [23.8]
Electrical termination (std) MS3102E-14S-6P 400000 0.007 8400 12.02
7500 lb to 500000 lb [26.5]
Mating connector PT06A-10-6S or equiv. (AA111) 500000 0.007 4800 41.00
5 lb to 5000 lb (not incl.) [90.4]
Mating connector MS3106A-14S-6S (AA121)
7500 to 500000 lb (not incl.)
Wiring code: cable4,5,6 R (+) Supply R (+) Supply R (+) Supply R (+) Supply R (+) Supply R (+) Supply
Bl Output common Bl Output common* Bl Output common* Bl Output common* Bl (+) Output Bl (+) Output
G Supply return G Supply return* G Supply return* G Supply return* W Case ground W Case ground
W (+) Output W (+) Output W (+) Output W (+) Output
B Shunt cal 1 B Shunt cal 1 B Shunt cal 1 B Shunt cal 1
Br Shunt cal 2 Br Shunt cal 2 Br Shunt cal 2 Br Shunt cal 2
* Black and green wires are internally connected.
** Pins B and C are internally connected.
*** See our Web site (http://measurementsensors.honeywell.com) for the most up-to-date information regarding intrinsically safe approvals, ref. #008-0547-00.
Model 43
Customer supplied
Chart recorder
Alarm panel
Data acquisition
Computer
Mating connectors & cables PLC
PT06A-10-6S, std.
PT06E-10-6S
MS3106A-14S-6S Display units
SC500
SC2000
SC2001
SC3004
GM (unamp. only)
GM-A (amp. only)
NK (unamp. only)
HH (unamp. only)
WARNING
PERSONAL INJURY
• DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
Find out more
Honeywell serves its customers WARNING
through a worldwide network of MISUSE OF DOCUMENTATION
sales offices, representatives • The information presented in this datasheet is for
and distributors. For application reference only. DO NOT USE this document as product
assistance, current specifica- installation information.
tions, pricing or name of the • Complete installation, operation and maintenance
information is provided in the instructions supplied with
nearest Authorized Distributor,
each product.
contact your local sales office.
Failure to comply with these instructions could result in
To learn more about Honeywell’s
death or serious injury.
test and measurement products,
call +1-614-850-5000, visit
http://measurementsensors.
honeywell.com, or e-mail inqui-
ries to info.tm@honeywell.com