You are on page 1of 91

Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO.

301/A, 9TH MAIN ROAD, 3RD


CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 1 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Permanent Facility

1 ELECTRO- Capacitance (1 kHz) 1 nF to 100 nF 0.056% to 0.061% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

2 ELECTRO- Capacitance (1 kHz) 100 nF to 1 µF 0.05% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

3 ELECTRO- Capacitance (1kHz) 1 µF to 100 µF 0.05% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

4 ELECTRO- Current (300 Hz to 10 1 A to 20 A 0.079% to 0.102% Direct Method with


TECHNICAL- kHz) using master 8½ DMM
ALTERNATING Fluke 8508 A
CURRENT (< 1 GHZ)
(Measure)

5 ELECTRO- Current (300 Hz to 10 100 µA to 100 mA 0.03% Direct Method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 2 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
6 ELECTRO- Current (300 Hz to 10 100 mA to 1 A 0.041% to 0.079% Direct Method using
TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

7 ELECTRO- Current (50 Hz) 20 A to 1000 A 0.5% to 1.0% Direct method master
TECHNICAL- Using Shunt With DMM
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

8 ELECTRO- Inductance (1kHz) 100 µH to 100 mH 0.236% to 0.065% Direct Method using
TECHNICAL- master LCR meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

9 ELECTRO- Inductance (1kHz) 100 mH to 10 H 0.065% to 0.073% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

10 ELECTRO- Voltage (20 Hz to 50 1 mV to 100 mV 0.820% to 0.014% Direct method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 3 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
11 ELECTRO- Voltage (20 Hz to 50 1 V to 100 V 0.017% to 0.011% Direct Method using
TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508A
CURRENT (< 1 GHZ)
(Measure)

12 ELECTRO- Voltage (20 Hz to 50 100 mV to 1 V 0.014% to 0.017% Direct method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508A
CURRENT (< 1 GHZ)
(Measure)

13 ELECTRO- Voltage (45 Hz to 1 100 V to 1000 V 0.01% to 0.007% Direct Method using
TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

14 ELECTRO- Voltage (50 kHz to 500 1 V to 20 V 0.07% to 0.230% Direct Method using
TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

15 ELECTRO- Capacitance (1 kHz) 220 pF to 300 nF 5.79% to 0.27% Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & Decade Capacitance
(Source) Box

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 4 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
16 ELECTRO- Capacitance (100 Hz) 1 µF to 30 µF 1.55% to 0.56% Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & Decade Capacitance
(Source) Box

17 ELECTRO- Capacitance(1 kHz) 300 nF to 100 µF 0.27% to 1.13% Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & Decade Capacitance
(Source) Box

18 ELECTRO- Current (1 kHz) 3.3 mA to 1 A 0.15% to 0.073% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

19 ELECTRO- Current (1 kHz) 30 µA to 330 µA 0.54% to 0.17% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

20 ELECTRO- Current (1 kHz) 330 µA to 3.3 mA 0.17% to 0.15% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 5 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
21 ELECTRO- Current (10Hz to 1kHz) 330 µA to 3.3 mA 0.25% to 0.26% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

22 ELECTRO- Current (10Hz to 1kHz) 3.3 mA to 3 A 0.26% to 0.23% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

23 ELECTRO- Current (1kHz) 1 A to 10 A 0.073% to 0.12% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

24 ELECTRO- Current (45 Hz to 5 3 A to 20 A 0.23% to 0.13% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

25 ELECTRO- Current (50 Hz) 20 A to 1000 A 0.68% to 0.37% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & current coil
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 6 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
26 ELECTRO- Inductance (1 kHz) 100 µH to 10 H 2.35 % to 1.14 % Direct Method using
TECHNICAL- master Decade
ALTERNATING Inductance Box
CURRENT (< 1 GHZ)
(Source)

27 ELECTRO- Power 50 Hz @ UPF 1.2 W to 0.16% Direct Method using


TECHNICAL- 4.8 kW,0.01 A to 20 master Standard
ALTERNATING A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

28 ELECTRO- Power 50Hz @ 0.5 Lag, 6 W 0.46% Direct Method using


TECHNICAL- to 2.4 kW,0.01 A to 20 master Standard
ALTERNATING A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

29 ELECTRO- Power 50Hz @ 0.8 Lead 9.6 0.20% Direct Method using
TECHNICAL- W to 3.8 kW,0.01 A to master Standard
ALTERNATING 20 A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

30 ELECTRO- Power 50Hz @0.2 Lag, 2.4 W 0.56% Direct Method using
TECHNICAL- to 960 W,0.01 A to 20 master Standard
ALTERNATING A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 7 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
31 ELECTRO- Power Factor 0.2 Lag to UPF 0.003 Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

32 ELECTRO- Power Factor 0.2 Lead to UPF 0.0003 Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

33 ELECTRO- Voltage (10 Hz to 10 1 V to 30 V 0.080% to 0.042% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

34 ELECTRO- Voltage (10 Hz to 10 100 mV to 1 V 0.086% to 0.080% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

35 ELECTRO- Voltage (10 Hz to 10 30 mV to 100 mV 0.12% to 0.086% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 8 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
36 ELECTRO- Voltage (100kHz) 30 mV to 30 V 0.452% to 0.054% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

37 ELECTRO- Voltage (100kHz) 30 V to 100 V 0.054% to 0.024% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

38 ELECTRO- Voltage (45 Hz to 1 30 V to 300 V 0.20% to 0.022% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

39 ELECTRO- Voltage (45 Hz to 1 300 V to 1000 V 0.022% to 0.035% Direct Method by using
TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

40 ELECTRO- Voltage (450kHz) 30 mV to 3 V 1.116% to 0.121% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 9 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
41 ELECTRO- Voltage (45Hz to 1 mV to 3 mV 0.712% to 0.24% Direct Method by using
TECHNICAL- 10kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

42 ELECTRO- Voltage (45Hz to 3 mV to 30 mV 0.24% to 0.042% Direct Method by using


TECHNICAL- 10kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

43 ELECTRO- Current 1 A to 20 A 0.080% to 0.052% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

44 ELECTRO- Current 1 mA to 100 mA 0.002% to 0.008% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

45 ELECTRO- Current 10 µA to 100 µA 0.0010% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

46 ELECTRO- Current 100 µA to 1 mA 0.001% to 0.002% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

47 ELECTRO- Current 100 mA to 1 A 0.062% to 0.080% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

48 ELECTRO- Current 20 A to 1000 A 1.19% Direct Method using


TECHNICAL- DIRECT 8½ DMM Fluke 8508
CURRENT (Measure) with Shunt

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 10 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
49 ELECTRO- Resistance 1 Gohm to 10 Gohm 0.116% to 0.133% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

50 ELECTRO- Resistance 1 Mohm to 10 Mohm 0.001% to 0.002% Direct Method by Using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

51 ELECTRO- Resistance 1 mohm to 100 mohm 0.061% to 0.207% Direct Method by Using
TECHNICAL- DIRECT master LCR Meter &
CURRENT (Measure) DRB

52 ELECTRO- Resistance 1 ohm to 100 ohm 0.0015% to 0.0010% Using LCR Meter &
TECHNICAL- DIRECT DRB
CURRENT (Measure)

53 ELECTRO- Resistance 10 kohm to 100 kohm 0.0007% to 0.0008% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

54 ELECTRO- Resistance 10 Mohm to 100 Mohm 0.002% to 0.008% Direct method by using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

55 ELECTRO- Resistance 100 Mohm to 1 Gohm 0.008% to 0.116% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

56 ELECTRO- Resistance 100 mohm to 150 0.206% Direct Method by Using


TECHNICAL- DIRECT mohm master LCR Meter &
CURRENT (Measure) DRB

57 ELECTRO- Resistance 100 ohm to 10 kohm 0.0010% to 0.0007% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 11 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
58 ELECTRO- Resistance 100 kohm to 1 Mohm 0.0008% to 0.001% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

59 ELECTRO- Resistance 150 mohm to 1 ohm 0.206% to 0.0015% Direct Method by using
TECHNICAL- DIRECT maste LCR Meter &
CURRENT (Measure) DRB

60 ELECTRO- Voltage 1 mV to 100 mV 0.44% to 0.0003% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Measure) Calibrator Fluke 5522A

61 ELECTRO- Voltage 1 V to 100 V 0.0002% to 0.0003% Direct Method by using


TECHNICAL- DIRECT master DMM Fluke
CURRENT (Measure) 8508A

62 ELECTRO- Voltage 100 mV to 1 V 0.0003% to 0.0003% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Measure) Calibrator Fluke 5522A

63 ELECTRO- Voltage 100 V to 1000 V 0.0004% to 0.0004% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Measure) Calibrator Fluke 5522A

64 ELECTRO- Current 1 A to 10 A 0.027% to 0.063% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

65 ELECTRO- Current 1 mA to 100 mA 0.016% to 0.011% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

66 ELECTRO- Current 10 µA to 100 µA 0.25% to 0.023% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 12 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
67 ELECTRO- Current 10 A to 20 A 0.063% to 0.11% Direct Method by using
TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

68 ELECTRO- Current 100 µA to 1 mA 0.023% to 0.016% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

69 ELECTRO- Current 100 mA to 1 A 0.011% to 0.027% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

70 ELECTRO- Current 20 A to 200 A 0.5% Direct Method using


TECHNICAL- DIRECT Current Source
CURRENT (Source)

71 ELECTRO- Current 200 A to 1000 A 0.66% to 0.32% Direct Method by using


TECHNICAL- DIRECT master Fluke 5522A
CURRENT (Source) with current coil 5500A

72 ELECTRO- DC Power 10 W to 1 kW,1V to 0.08% to 0.45% Using Calibrator Fluke


TECHNICAL- DIRECT 1000V, 1mA to 1000A. 5522A with Current
CURRENT (Source) Coil Direct Method.

73 ELECTRO- DC Power (1V to 1 KW to 19.9 KW 0.45% to 0.65% Using Calibrator Fluke


TECHNICAL- DIRECT 1000V, 1mA to 1000A) 5522A with current coil
CURRENT (Source) Direct Method.

74 ELECTRO- Oscilloscope 1 mV to 130 V 0.29% to 0.096% Direct Method


TECHNICAL- DIRECT
CURRENT (Source)

75 ELECTRO- Resistance 0.1 Mohm to 1 Gohm 5.89% to 4.91% Direct Method using
TECHNICAL- DIRECT master of high stability
CURRENT (Source) Decade Megohm Box

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 13 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
76 ELECTRO- Resistance 1 Gohm to 100 Gohm 4.91% to 8.61% Direct Method using
TECHNICAL- DIRECT master of high stability
CURRENT (Source) Decade Megohm Box

77 ELECTRO- Resistance 1 kohm to 10 kohm 0.0036% to 0.0035% Direct Method using


TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

78 ELECTRO- Resistance 1 mohm 0.609% Direct Method using


TECHNICAL- DIRECT fixed resistor box
CURRENT (Source)

79 ELECTRO- Resistance 1 Mohm to 10 Mohm 0.0041% to 0.014% Direct Method using


TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

80 ELECTRO- Resistance 1 ohm to 10 ohm 0.10% to 0.015% Direct Method using


TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

81 ELECTRO- Resistance 10 kohm to 100 kohm 0.0035% to 0.0034% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

82 ELECTRO- Resistance 10 mohm 0.610% Direct Method using


TECHNICAL- DIRECT fixed resistor box
CURRENT (Source)

83 ELECTRO- Resistance 10 Mohm to 300 Mohm 0.014% to 0.38% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

84 ELECTRO- Resistance 10 ohm to 100 ohm 0.015% to 0.004% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 14 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
85 ELECTRO- Resistance 100 Gohm to 1 Tohm 5.81% Direct Method using
TECHNICAL- DIRECT master of high stability
CURRENT (Source) Decade Megohm Box

86 ELECTRO- Resistance 100 ohm to 1 kohm 0.004% to 0.0036% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

87 ELECTRO- Resistance 100 kohm to 1 Mohm 0.0034% to 0.0041% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

88 ELECTRO- Resistance 150 mohm 0.609% Direct Method using


TECHNICAL- DIRECT fixed resistor box
CURRENT (Source)

89 ELECTRO- Resistance 150 mohm to 1 ohm 0.611% to 0.10% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A:

90 ELECTRO- Resistance 300 Mohm to 1 Gohm 0.38% to 1.78% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

91 ELECTRO- Resistance 50 mohm 0.610% Direct Method using


TECHNICAL- DIRECT sing fixed resistor box
CURRENT (Source)

92 ELECTRO- Voltage 1 mV to 10 mV 0.240% to 0.023% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

93 ELECTRO- Voltage 1 V to 10 V 0.0015% to 0.0016% Direct Method using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A:

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 15 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
94 ELECTRO- Voltage 10 mV to 100 mV 0.023% to 0.0039% Direct Method by using
TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

95 ELECTRO- Voltage 10 V to 1000 V 0.0016% to 0.0023% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

96 ELECTRO- Voltage 100 mV to 1 V 0.0039% to 0.0015% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

97 ELECTRO- Amplitude-Sq wave / 1 mV to 55 Vp-p 0.6% to 0.3% Direct Method by using


TECHNICAL- Oscilloscope (1 kHz) master Standard
ELECTRICAL Calibrator Fluke 5522A
EQUIPMENT (Source)

98 ELECTRO- Bandwidth - 50 kHz to 1 GHz 0.23% to 0.87% Direct Method using


TECHNICAL- Oscilloscope master Standard
ELECTRICAL Calibrator Fluke 5522A
EQUIPMENT (Source)

99 ELECTRO- Temperature Indicator / 600 °C 0.53°C to 0.39°C Direct Method by using


TECHNICAL- Recorder / Controller to 1750 °C Master Standard
TEMPERATURE (B type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

100 ELECTRO- Temperature Indicator / -200 °C to 900 °C 0.48°C to 0.27°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (E type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 16 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
101 ELECTRO- Temperature Indicator / -200 °C to 1200 °C 0.28°C to 0.31°C Direct Method by using
TECHNICAL- Recorder / Controller (J Master used- Stnd
TEMPERATURE type) Calibrator Fluke 5522A
SIMULATION & 6½ DMM
(Measure)

102 ELECTRO- Temperature Indicator / -200 °C to 1350 °C 0.26°C to 0.48°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (K type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

103 ELECTRO- Temperature Indicator / -200 °C to 1300 °C 0.48°C to 0.35°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (N type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

104 ELECTRO- Temperature Indicator / 10 °C to 1750 °C 0.67°C to 0.48°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (R type) Calibrator fluke 5522A
SIMULATION
(Measure)

105 ELECTRO- Temperature Indicator / -200 °C to 800 °C 0.05 °C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (RTD type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 17 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
106 ELECTRO- Temperature Indicator / 10 °C to 1450 °C 0.56°C to 0.54°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (S type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

107 ELECTRO- Temperature Indicator / -200 °C to 400 °C 0.60°C to 0.17°C Direct Method by using
TECHNICAL- Recorder / Controller (T Master Standard
TEMPERATURE type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

108 ELECTRO- Temperature Indicator / 600 °C to 1750 °C 0.47°C to 0.11°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (B type) Calibrator fluke 5522A
SIMULATION (Source)

109 ELECTRO- Temperature Indicator / -200 °C to 900 °C 0.20°C to 0.45°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (E type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

110 ELECTRO- Temperature Indicator / -200 °C to 1200 °C 0.15°C to 0.25°C Direct Method by using
TECHNICAL- Recorder / Controller (J Master Standard
TEMPERATURE type) Calibrator fluke 5522A
SIMULATION (Source)

111 ELECTRO- Temperature Indicator / -200 °C to 1350 °C 0.37°C to 0.48°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (K type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 18 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
112 ELECTRO- Temperature Indicator / -200 °C to 1300 °C 0.45°C to 0.30°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (N Type ) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

113 ELECTRO- Temperature Indicator / 10 °C to 1750 °C 0.62°C to 0.44°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (R type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

114 ELECTRO- Temperature Indicator / -200 °C to 800 °C 0.071°C to 0.083°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (RTD type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

115 ELECTRO- Temperature Indicator / 10 °C to 1750 °C 0.61°C to 0.51°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (S type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

116 ELECTRO- Temperature Indicator / -200 °C to 400 °C 0.55°C to 0.16°C Direct Method by using
TECHNICAL- Recorder / Controller (T Master Standard
TEMPERATURE type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

117 ELECTRO- Frequency 10 Hz to 1 GHz 0.0011% to 0.0024% Direct Method using


TECHNICAL- TIME & master Frequency
FREQUENCY Counter Philips- PM
(Measure) 6669

118 ELECTRO- Timer 1 sec to 9000 sec 0.06% to 0.05% Comparison Method by
TECHNICAL- TIME & using master Digital
FREQUENCY Timer Beltronics - 501
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 19 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
119 ELECTRO- Frequency 1 Hz to 1 GHz 0.004% to 0.006% Direct Method using
TECHNICAL- TIME & master Standard
FREQUENCY (Source) Calibrator Fluke 5522A

120 ELECTRO- Timer Marker / 1 ns to 5 s 0.0003 sec Direct Method using


TECHNICAL- TIME & Oscilloscope master Standard
FREQUENCY (Source) Calibrator Fluke 5522A

121 MECHANICAL- Acceleration (Peak) 6.6 m/s2 2.96% Using Digital Vibration
ACCELERATION AND @80Hz Meter by Comparison
SPEED Method

122 MECHANICAL- Acceleration 10.0 m/s2 2.83% Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

123 MECHANICAL- Acceleration 20.1 m/s2 2.48 % Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

124 MECHANICAL- Acceleration 40.2 m/s2 2.44 % Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

125 MECHANICAL- Acceleration 50.1 m/s2 2.41 % Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

126 MECHANICAL- Acceleration 12.3 m/s2 2.56 % Using Digital Vibration


ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

127 MECHANICAL- Acceleration 20.5 m/s2 2.49 % Using Digital Vibration


ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 20 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
128 MECHANICAL- Acceleration 28.8 m/s2 2.49 % Using Digital Vibration
ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

129 MECHANICAL- Acceleration 4.0 m/s2 3.43 % Using Digital vibration


ACCELERATION AND (Peak)@40Hz Meter By Comparison
SPEED Method

130 MECHANICAL- Acceleration 11.5 m/s2 2.63 % Using Digital Vibration


ACCELERATION AND (Peak)@80Hz Meter by Comparison
SPEED Method

131 MECHANICAL- Acceleration 16.7 m/s2 2.53 % Using Digital Vibration


ACCELERATION AND (Peak)@80Hz Meter by Comparison
SPEED Method

132 MECHANICAL- Acceleration 2.8 m/s2 5.04 % Using Digital Vibration


ACCELERATION AND (Peak)@80Hz Meter by Comparison
SPEED Method

133 MECHANICAL- Displacement 0.009 mm 11.87% Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

134 MECHANICAL- Displacement 0.030 mm 4.36% Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

135 MECHANICAL- Displacement 0.050 mm 3.24% Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

136 MECHANICAL- Displacement 0.126 mm 2.56% Using Digital Vibration


ACCELERATION AND (Peak)@160Hz Meter by Comparison
SPEED Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 21 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
137 MECHANICAL- Displacement 0.065 mm 2.89% Using Digital Vibration
ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

138 MECHANICAL- Displacement 0.196 mm 2.47% Using Digital Vibration


ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

139 MECHANICAL- Displacement 0.263 mm 2.45% Using Digital Vibration


ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

140 MECHANICAL- Displacement 0.459 mm 2.44% Using Digital Vibration


ACCELERATION AND (Peak)@40Hz Meter by Comparison
SPEED Method

141 MECHANICAL- Displacement 0.026 mm 4.79% Using Digital Vibration


ACCELERATION AND (Peak)@80 Hz Meter by Comparison
SPEED Method

142 MECHANICAL- Displacement 0.011 mm 10.27% Using Digital Vibration


ACCELERATION AND (Peak)@80Hz Meter by Comparison
SPEED Method

143 MECHANICAL- Displacement 0.046 mm 3.08% Using Digital Vibration


ACCELERATION AND (Peak)@80Hz Meter by Comparison
SPEED Method

144 MECHANICAL- Displacement 0.067 mm 2.76% Using Digital Vibration


ACCELERATION AND (Peak)@80Hz Meter by Comparison
SPEED Method

145 MECHANICAL- Stroboscope, 25000 rpm to 45000 0.008% By Comparison method


ACCELERATION AND tachometer (Non rpm by using master
SPEED Contact Type) Tachometer Calibrator

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 22 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
146 MECHANICAL- Stroboscope, 60 rpm to 25000 rpm 0.7% to 0.008% By Comparison method
ACCELERATION AND tachometer (Non by using master
SPEED Contact Type) Tachometer Calibrator

147 MECHANICAL- Tachometer ( Contact 100 rpm to 5000 rpm 1.4% to 0.12% Comparison Method by
ACCELERATION AND Type) using master
SPEED Tachometer Calibration

148 MECHANICAL- Velocity (Peak) @40Hz 82.0 mm/s 2.42% Using Digital Vibration
ACCELERATION AND Meter by Comparison
SPEED Method

149 MECHANICAL- Velocity (Peak) 16.1 mm/s 2.52% Using Digital Vibration
ACCELERATION AND mm/s@40Hz Meter by Comparison
SPEED Method

150 MECHANICAL- Velocity (Peak)@40Hz 48.7 mm/s 2.43% Using Digital Vibration
ACCELERATION AND Meter by Comparison
SPEED Method

151 MECHANICAL- Velocity (Peak)@40Hz 65.1 mm/s 2.42% Using Digital Vibration
ACCELERATION AND Meter by Comparison
SPEED Method

152 MECHANICAL- Velocity(Peak)@160Hz 39.9 mm/s 2.42% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

153 MECHANICAL- Velocity(Peak)@160Hz 69.7 mm/s 2.41% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

154 MECHANICAL- Velocity(Peak)@160Hz 9.9 mm/s 2.80% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 23 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
155 MECHANICAL- Velocity(Peak)@160Hz 99.3 mm/s 2.41% Using Digital Vibration
ACCELERATION AND Meter by Comparison
SPEED Method

156 MECHANICAL- Velocity(Peak)@80Hz 13.2 mm/s 2.50% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

157 MECHANICAL- Velocity(Peak)@80Hz 22.8 mm/s 2.47% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

158 MECHANICAL- Velocity(Peak)@80Hz 33.3 mm/s 2.44% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

159 MECHANICAL- Velocity(Peak)@80Hz 5.7 mm/s 3.00% Using Digital Vibration


ACCELERATION AND Meter by Comparison
SPEED Method

160 MECHANICAL- Sound Level Meter 114 dB @ 1kHz 0.82dB Using Sound level
ACOUSTICS Calibrator

161 MECHANICAL- Sound Level Meter 94 dB @ 1kHz 0.82dB Using Sound level
ACOUSTICS Calibrator

162 MECHANICAL- 3 Point Micrometer, 0 to 100 mm 3.6µm Using master Master


DIMENSION (BASIC Res : 0.001 mm Ring Gauges
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 24 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
163 MECHANICAL- Bevel Protractor, Res: 0° 5.66arc of min Using profile projector
DIMENSION (BASIC 1min to 360 °
MEASURING
INSTRUMENT,
GAUGE ETC.)

164 MECHANICAL- Bore Gauge (0.001mm) Upto 2mm 2.4µm Using master
DIMENSION (BASIC Transmisson only Electronic Dial gauge
MEASURING calibration Tester
INSTRUMENT,
GAUGE ETC.)

165 MECHANICAL- Caliper (Analog / Dial / 0 mm to 1000 mm 8.6µm Using master Caliper
DIMENSION (BASIC Digital) Res: 0.01mm Checker & Gauge
MEASURING Block
INSTRUMENT,
GAUGE ETC.)

166 MECHANICAL- Caliper Checker 0 mm to 1000 mm 8µm Using master Gauge


DIMENSION (BASIC Block Set & Height
MEASURING Measuring System
INSTRUMENT,
GAUGE ETC.)

167 MECHANICAL- Coating Thickness 0 to 2000 µm 3.4µm Using master Standard


DIMENSION (BASIC Gauge (Res: 0.1 µm) Thickness Foils
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 25 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
168 MECHANICAL- Combination Set, 0 ° to 180 ° 13.2 arc of min Using Profile Projector
DIMENSION (BASIC Res:1°
MEASURING
INSTRUMENT,
GAUGE ETC.)

169 MECHANICAL- Comparator Dial Stand Up to 300 mm 7µm Using master Lever dial
DIMENSION (BASIC ( Flatness) & Linear height master
MEASURING
INSTRUMENT,
GAUGE ETC.)

170 MECHANICAL- Depth Micrometer, 0 mm to 300 mm 4µm Using Master Gauge


DIMENSION (BASIC Res: 0.001mm Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

171 MECHANICAL- Depth Vernier ( Analog 0 mm to 600 mm 15.2µm Using master Gauge
DIMENSION (BASIC / Dial / Digital), Res: block Set
MEASURING 0.01mm
INSTRUMENT,
GAUGE ETC.)

172 MECHANICAL- Dial Calibration Tester / 0 to 25mm 1.3µm Using master ULM
DIMENSION (BASIC LVDT probe with
MEASURING Indicator, Res:
INSTRUMENT, 0.001µm
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 26 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
173 MECHANICAL- Dial Gauge- Plunger 0 mm to 1 mm 2.3µm Using master
DIMENSION (BASIC type ( Analog / Digital), Electronic Dial Gauge
MEASURING Res: 0.001mm Calibration Tester
INSTRUMENT,
GAUGE ETC.)

174 MECHANICAL- Dial Gauge- Plunger 0 mm to 10 mm 2.6µm Using master


DIMENSION (BASIC type ( Analog / Digital), Electronic Dial gauge
MEASURING Res: 0.001mm Calibration Tester
INSTRUMENT,
GAUGE ETC.)

175 MECHANICAL- Dial Gauge- Plunger 0 mm to 50 mm 6.3µm Using master


DIMENSION (BASIC type ( Analog / Digital), Electronic Dial
MEASURING Res: 0.001mm Calibration Gauge
INSTRUMENT, Tester
GAUGE ETC.)

176 MECHANICAL- Dial Thickness Gauge, Upto 10mm 7µm Using master Gauge
DIMENSION (BASIC Res: 0.01 mm Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

177 MECHANICAL- Elongation gauge 14.7 mm to 81 mm 24µm Using Digital Caliper by


DIMENSION (BASIC Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 27 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
178 MECHANICAL- Engineering Square / Up to 350 mm 7.0µm Using master Digital
DIMENSION (BASIC Angle Plates Dial Gauge & Linear
MEASURING (Squareness) Height Master
INSTRUMENT,
GAUGE ETC.)

179 MECHANICAL- External Micrometer 0 mm to 450 mm 3µm Using master Gauge


DIMENSION (BASIC (Analog / Digital / Block Set
MEASURING Flange / Point / V Anvil
INSTRUMENT, / Pitch ) , Res : 0.001
GAUGE ETC.) mm

180 MECHANICAL- External Micrometer 0 mm to 1000 mm 7µm Using master Gauge


DIMENSION (BASIC (Analog / Digital / Block Set
MEASURING Flange / Point / V Anvil
INSTRUMENT, / Pitch ) , Res : 0.01
GAUGE ETC.) mm

181 MECHANICAL- Feeler gauge 0.02 mm to 1 mm 2.1µm Using master Universal


DIMENSION (BASIC Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

182 MECHANICAL- Flankiness Gauge 4.89 mm to 33.9 mm 7µm Using Profile Projector
DIMENSION (BASIC By Comparison Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 28 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
183 MECHANICAL- Foils Upto 2000 µm 1.2µm Using master Universal
DIMENSION (BASIC Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

184 MECHANICAL- Gauge Block Set 0.5 mm to 25 mm 0.12 µm Using master Gauge
DIMENSION (BASIC Block Grade 'k'
MEASURING comparator system
INSTRUMENT,
GAUGE ETC.)

185 MECHANICAL- Gauge Block Set 25 mm to 50 mm 0.12µm Using master Gauge


DIMENSION (BASIC Block Grade 'k'
MEASURING
INSTRUMENT,
GAUGE ETC.)

186 MECHANICAL- Gauge Block Set 50 mm to 75 mm 0.14µm Using master Gauge


DIMENSION (BASIC Block Grade 'k'
MEASURING
INSTRUMENT,
GAUGE ETC.)

187 MECHANICAL- Gauge Block Set 75 mm to 100 mm 0.20µm Using master Gauge
DIMENSION (BASIC Block Grade 'k'
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 29 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
188 MECHANICAL- Granite Square Upto 600 mm 13.5µm Using Lever Dial
DIMENSION (BASIC (Squareness, Flatness) Gauge and Linear
MEASURING height Master
INSTRUMENT,
GAUGE ETC.)

189 MECHANICAL- Groove Dial / Leg Upto 150 mm 7.0µm Using master Gauge
DIMENSION (BASIC Caliper (OD & ID) Block Set
MEASURING Gauge, Res: 0.025mm
INSTRUMENT,
GAUGE ETC.)

190 MECHANICAL- Height Gauge ( Dial / 0 mm to 1000 mm 11.8µm Using master Caliper
DIMENSION (BASIC Digital & Analog ) Res: Checker
MEASURING 0.01mm
INSTRUMENT,
GAUGE ETC.)

191 MECHANICAL- Height Gauge ( Dial / 0 mm to 600 mm 11.2µm Using master Caliper
DIMENSION (BASIC Digital & Analog ) Res: Checker
MEASURING 0.01mm
INSTRUMENT,
GAUGE ETC.)

192 MECHANICAL- Height Measuring 0 mm to 1000 mm 4.6µm By using Master Gauge


DIMENSION (BASIC System, Res: 0.1 µm BLock long Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 30 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
193 MECHANICAL- Internal Micrometer / Upto 300 mm 8µm Using Gauge Block Set
DIMENSION (BASIC Stick Micrometer & gauge Block
MEASURING (Including Extension Accessories
INSTRUMENT, Rod ) Res: 0.01mm
GAUGE ETC.)

194 MECHANICAL- Lever Dial Gauge 0 to 0.14 mm 2.4µm Using master


DIMENSION (BASIC (0.001mm) Electronic Dial gauge
MEASURING Calibration Tester
INSTRUMENT,
GAUGE ETC.)

195 MECHANICAL- Lever Dial Gauge 0 to 0.2 mm 3.1µm Using master


DIMENSION (BASIC (0.002mm) Electronic Dial gauge
MEASURING Calibration Tester
INSTRUMENT,
GAUGE ETC.)

196 MECHANICAL- Lever Dial Gauge, 0.01 0 to 1 mm 5.3µm Using master


DIMENSION (BASIC mm Electronic Dial gauge
MEASURING calibration Tester
INSTRUMENT,
GAUGE ETC.)

197 MECHANICAL- Measuring Pins ( Upto 25 mm 2µm Using Universal length


DIMENSION (BASIC Grade 1& 2) Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 31 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
198 MECHANICAL- Measuring Tape / Pi Upto 50000 mm (541 + L/100)µm(L in Using master Scale &
DIMENSION (BASIC Tape, Res: 0.5 mm mm) Tape Calibration
MEASURING 1.00 mm lc Machine
INSTRUMENT,
GAUGE ETC.)

199 MECHANICAL- Micrometer Head, Res Up to 25 mm 1.3µm Using Master Lever


DIMENSION (BASIC : 0.001mm Dial Gauge & Linear
MEASURING Height Master
INSTRUMENT,
GAUGE ETC.)

200 MECHANICAL- Parallel Block Up to 1000 mm 10µm Using master Lever


DIMENSION (BASIC Dial Gauge
MEASURING
INSTRUMENT,
GAUGE ETC.)

201 MECHANICAL- Pistol Caliper (Res: 0 to 100 mm 10µm Using master Gauge
DIMENSION (BASIC 0.01 mm) Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

202 MECHANICAL- Pitch Micrometer, Res : 0 to 200 mm 7µm Using master Gauge
DIMENSION (BASIC 0.01mm Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 32 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
203 MECHANICAL- Plain Plug Gauges 1 mm to 300 mm 1.5µm Using master Universal
DIMENSION (BASIC Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

204 MECHANICAL- Plain Ring Gauges 4 mm to 300 mm 2.3µm Using master Universal
DIMENSION (BASIC Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

205 MECHANICAL- Portable Roughness Up to 100 µm 6.41 µm Using Surface


DIMENSION (BASIC tester Ra/ Rz Roughness Master
MEASURING Block
INSTRUMENT,
GAUGE ETC.)

206 MECHANICAL- Radius Gauges 0.5 mm to 25.0 mm 6µm Using master Profile
DIMENSION (BASIC projector
MEASURING
INSTRUMENT,
GAUGE ETC.)

207 MECHANICAL- Setting Rods & 25 mm to 975 mm 10µm Using master Gauge
DIMENSION (BASIC Extensions Rods Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 33 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
208 MECHANICAL- Snap Gauges / Dial 10 mm to 300 mm 2.5 µm Using Universal Length
DIMENSION (BASIC Snap Gauges / Width Machine
MEASURING Gauges
INSTRUMENT,
GAUGE ETC.)

209 MECHANICAL- Spirit Level / Frame Upto 0.12 mm/m 15µm/m Using master
DIMENSION (BASIC Level, Res : 10µm / m Electronic Level
MEASURING
INSTRUMENT,
GAUGE ETC.)

210 MECHANICAL- Steel scale / Taper 0 mm to 1000 mm 271µm Using master scale &
DIMENSION (BASIC scale Res: 0.5 / 1 mm Tape Calibration
MEASURING Machine
INSTRUMENT,
GAUGE ETC.)

211 MECHANICAL- Straight Edge 0 mm to 1000 mm 13.5µm Using Linear Height


DIMENSION (BASIC Master and lever Dial
MEASURING Gauge
INSTRUMENT,
GAUGE ETC.)

212 MECHANICAL- Tape Scale Calibrator 0 mm to 1000 mm 10.2µm Using Slip Gauge Set
DIMENSION (BASIC and Long Slip Gauges
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 34 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
213 MECHANICAL- Taper Thread Plug 0 mm to 100 mm 3µm Using master Universal
DIMENSION (BASIC Gauge ( Effective Dia) Length machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

214 MECHANICAL- Taper Thread Ring 0 mm to 110 mm 4.8µm Using master Universal
DIMENSION (BASIC Gauge ( Effective Dia) Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

215 MECHANICAL- Test Sieves 0.16 mm to 1 mm 5µm Using master profile


DIMENSION (BASIC projector
MEASURING
INSTRUMENT,
GAUGE ETC.)

216 MECHANICAL- Test Sieves 1 mm to 100 mm 18µm Using master profile


DIMENSION (BASIC projector
MEASURING
INSTRUMENT,
GAUGE ETC.)

217 MECHANICAL- Thread Pitch Gauge, 0.3 mm to 6.0 mm 3 µm & 5.1 arc min Using master Profile
DIMENSION (BASIC Set Pitch Flank Angle Projector
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 35 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
218 MECHANICAL- Thread Pitch Gauge, 5.1 min of arc Using master Profile
55 ° to 60 °
DIMENSION (BASIC Set Pitch Flank Angle Projector
MEASURING
INSTRUMENT,
GAUGE ETC.)

219 MECHANICAL- Thread plug Gauge ( Up to 300 mm 2.1µm Using master Universal
DIMENSION (BASIC Effective Dia) Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

220 MECHANICAL- Thread Ring Gauge ( 3 mm to 300 mm 2µm Using master Universal
DIMENSION (BASIC Effective Dia) Length Machine
MEASURING
INSTRUMENT,
GAUGE ETC.)

221 MECHANICAL- Ultra Sonic Thickness Upto 200mm 7.3µm Using master Gauge
DIMENSION (BASIC Gauge Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

222 MECHANICAL- Universal Length Upto 100 mm 0.7µm Using master Gauge
DIMENSION (BASIC Measuring Machine Block Set
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 36 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
223 MECHANICAL- 'V' - Block (Parallelism) Up to 200 mm 6.6µm Using master Mandrel
DIMENSION (BASIC symmetry, squareness & Lever Dial Gauge &
MEASURING Linear Height Master
INSTRUMENT,
GAUGE ETC.)

224 MECHANICAL- Weld Fillet Gauge 0 mm to 25 mm 10.1µm for linear & 3 Using Profile projector
DIMENSION (BASIC min arc By Comparsion Method
MEASURING
INSTRUMENT,
GAUGE ETC.)

225 MECHANICAL- wet / Dry Film 0 mm to 200 mm 8.68µm Using Profile Projector
DIMENSION (BASIC Applicator By Comparision
MEASURING Method
INSTRUMENT,
GAUGE ETC.)

226 MECHANICAL- Wire gauge 0.3 mm to 10 mm 8.8 µm Using Profile Projector


DIMENSION (BASIC By Comparision
MEASURING Method.
INSTRUMENT,
GAUGE ETC.)

227 MECHANICAL- Push Pull Gauge , 10 N to 1000 N 1.4N Using master Stain
FORCE PROVING Force gauge, Spring Less Weights and
INSTRUMENTS Balance hanger of weights 10,
50, 100 N as per VDI /
VDE 2624 Part 2.1

228 MECHANICAL- Shore Hardness Tester 0 to 100 Shore A&D 2.2 Shore A&D Using Rubber
HARDNESS TESTING Hardness Tester
MACHINES ASTM-D-2240-15

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 37 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
229 MECHANICAL- Digital / Dial Vaccum 0 bar to (-) 0.75 bar 0.014 bar Digital Pressure Gauge
PRESSURE Gauges / Transmitters / with uncertainty of
INDICATING DEVICES Transducers. 0.0008 bar. Procedure
based on ISO 3567 &
ISO 27893 standards.

230 MECHANICAL- PRESSURE 0 bar to 1000 bar 0.15 bar Digital Pressure gauge
PRESSURE (HYDRAULIC)Digital / with uncertainty of
INDICATING DEVICES Analog Pressure 0.021 bar. Procedure
Gauges, Pressure based on DKD-R 6-1
Indicator With guidelines.
Transducer /
Transmitter, Pressure
Switch

231 MECHANICAL- PRESSURE 0 bar to 700 bar 0.29 bar Digital Pressure gauge
PRESSURE (HYDRAULIC)Digital / with uncertainty of
INDICATING DEVICES Analog Pressure 0.182 bar. Procedure
Gauges, Pressure based on DKD-R 6-1
Indicator With guidelines.
Transducer /
Transmitter & Pressure
Switches

232 MECHANICAL- PRESSURE 0 bar to 20 bar 0.014 bar Digital Pressure gauge
PRESSURE (PNEUMATIC) Digital / with uncertainty of
INDICATING DEVICES Dial Gauges / Pressure 0,01bar. Procedure
Indicator With based on DKD-R 6-1
Transmitter / guidelines.
Transducer.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 38 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
233 MECHANICAL- PRESSURE 0 mbar to 500 mbar 0.31 mbar Digital Manometer with
PRESSURE (PNEUMATIC) uncertainty of 0.06
INDICATING DEVICES Pressure gauges, mbar. Procedure based
Magnehelic gauges, on DKD-R 6-1
Differential Pressure guidelines.
Transmitters, Low
Pressure Indicators/
Transducers,
Barometer

234 MECHANICAL- PRESSURE 0 bar to 70 bar 0.04 bar Digital Pressure gauge
PRESSURE (PNEUMATIC)Digital / with uncertainty of
INDICATING DEVICES Dial Gauges / Pressure 0.022 bar. Procedure
Indicator With based on DKD-R 6-1
Transmitter / guidelines.
Transducer.

235 MECHANICAL- PRESSURE 0 mbar to 2000 mbar 2.14 mbar Digital Manometer with
PRESSURE (PNEUMATIC)Pressur uncertainty of 1.03
INDICATING DEVICES e gauges, Magnehelic mbar. Procedure based
gauges, Differential on DKD-R 6-1
Pressure Transmitters, guidelines.
Low Pressure
Indicators/
Transducers,
Barometer.

236 MECHANICAL- Torque Wrench (Type- 200 Nm to 1000 Nm 1.2 % Using Torque wrench
TORQUE I)Class B,C,D,E and Calibration System as
MEASURING Type II class A B C D per IS/ISO 6789
DEVICES

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 39 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
237 MECHANICAL- Torque Wrench (Type- 0 Nm to 20 Nm 0.36 % Using Torque wrench
TORQUE I)Class B,C,D,E, Type Calibration System as
MEASURING II, A B C D per IS/ISO 6789
DEVICES

238 MECHANICAL- Torque Wrench (Type- 20 Nm to 200 Nm 1.5% Using Torque wrench
TORQUE I)Class B,C,D,E, Type Calibration System as
MEASURING II, Class A, B, C, D per IS/ISO 6789,
DEVICES

239 MECHANICAL- UTM, Universal Testing 100 N to 1000 KN 0.80% Using master Proving
TENSION CREEP Machine (Compression Ring and load cells as
AND TORSION Mode) per IS 1828 Part 1
TESTING MACHINE 2015

240 MECHANICAL- UTM, Universal Testing 200 kN to 2000 KN 0.80 % Using master Proving
TENSION CREEP Machine (Compression Ring and load cells as
AND TORSION Mode) per IS 1828 Part 1
TESTING MACHINE 2015

241 MECHANICAL- UTM, Universal Testing 0.5 KN to 50 KN 0.99% Using master Load
TENSION CREEP Machine (Tensile Cells as per IS 1828
AND TORSION Mode) Part 1 2015
TESTING MACHINE

242 MECHANICAL- DENSITYHydrometer , 0.600 g / ml to 1.8 g/ ml 0.0009 g/ml using reference


VOLUME L.C= 0.0005 g/ ml hydrometer and
calibration liquid of
appropriate density.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 40 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
243 MECHANICAL- Measuring Cylinder 10 ml 0.8 ml Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

244 MECHANICAL- Measuring Cylinder 100 ml 1.9 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.001 g)

245 MECHANICAL- Measuring Cylinder 1000 ml 11.5 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.001g)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 41 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
246 MECHANICAL- Measuring Cylinder 10000 ml 75.0 ml Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 g)

247 MECHANICAL- Measuring Cylinder 20 ml 1.5 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

248 MECHANICAL- Measuring Cylinder 200 ml 2.3 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.001 g)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 42 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
249 MECHANICAL- Measuring Cylinder 2000 ml 12.0 ml Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01g)

250 MECHANICAL- Measuring Cylinder 5 ml 0.1 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

251 MECHANICAL- Measuring Cylinder 50 ml 1.5 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 43 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
252 MECHANICAL- Measuring Cylinder 500 ml 10.5 ml Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric method
based on ISO
4787:2010 & ISO
20461 with Electronic
weighing Balance (
Readability : 1 mg)

253 MECHANICAL- Measuring Cylinder 5000 ml 25.0 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 g)

254 MECHANICAL- Micropipette 1 µl 0.01 µl Using Weights of


VOLUME Accuracy Class E1 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 44 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
255 MECHANICAL- Micropipette 10 µl 0.03 µl Using Weights of
VOLUME Accuracy Class E1 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg)

256 MECHANICAL- Micropipette 100 µl 0.62 µl Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg
)

257 MECHANICAL- Micropipette 1000 µl 6.20 µl Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.01 mg )

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 45 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
258 MECHANICAL- Micropipette 2 µl 0.01µl Using Weights of
VOLUME Accuracy Class E1 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg)

259 MECHANICAL- Micropipette 20 µl 0.04 µl Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.01 mg /
0.1 mg)

260 MECHANICAL- Micropipette 200 µl 1.30 µl Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg
)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 46 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
261 MECHANICAL- Micropipette 2000 µl 12.5 µl Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.01 mg /
0.1 mg)

262 MECHANICAL- Micropipette 5 µl 0.02 µl Using Weights of


VOLUME Accuracy Class E1 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg)

263 MECHANICAL- Micropipette 50 µl 0.20 µl Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.001 mg
)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 47 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
264 MECHANICAL- Micropipette 500 µl 3.10 µl Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.01 mg )

265 MECHANICAL- Micropipette 5000 µl 32.0 µl Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 8655( part 6):
2002 with using master
Micro Balance (
Readability : 0.1 mg)

266 MECHANICAL- Pipette & burette 10 ml 0.03 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 mg)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 48 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
267 MECHANICAL- Pipette & burette 100 ml 0.21 ml Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

268 MECHANICAL- Pipette & burette 25 ml 0.06ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 mg )

269 MECHANICAL- Pipette & burette 5 ml 0.02 ml Using Weights of


VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 mg)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 49 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
270 MECHANICAL- Pipette & burette 50 ml 0.11 ml Using Weights of
VOLUME Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

271 MECHANICAL- Volumetric ( Flask / 10 ml 0.03ml Using Weights of


VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 mg)

272 MECHANICAL- Volumetric ( Flask / 100 ml 0.21 ml Using Weights of


VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 50 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
273 MECHANICAL- Volumetric ( Flask / 1000 ml 2.2 ml Using Weights of
VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 g)

274 MECHANICAL- Volumetric ( Flask / 20 ml 0.06 ml Using Weights of


VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 mg)

275 MECHANICAL- Volumetric ( Flask / 200 ml 0.41 ml Using Weights of


VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.001 g)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 51 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
276 MECHANICAL- Volumetric ( Flask / 2000 ml 4.60 ml Using Weights of
VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 g)

277 MECHANICAL- Volumetric ( Flask / 5 ml 0.02 ml Using Weights of


VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.01 mg )

278 MECHANICAL- Volumetric ( Flask / 50 ml 0.11 ml Using Weights of


VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.1 mg )

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 52 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
279 MECHANICAL- Volumetric ( Flask / 500 ml 1.03 ml Using Weights of
VOLUME Beaker) Accuracy Class E2 &
Distilled water by
Gravimetric methods
on ISO 4787:2010 &
ISO 20461 with master
using Balance (
Readability : 0.001 g)

280 MECHANICAL- ELECTRONIC 1 mg to 2.1 g 0.003mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of acccuracy
AND BALANCE ( Class I) Readability : Class E1 and
0.0001 mg procedure based OIML
R-76: 2006.

281 MECHANICAL- ELECTRONIC 1 mg to 1 kg 0.002 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class Ii) Readability : Class E1 & E2, and
1 mg procedure as per OIML
R-76: 2006.

282 MECHANICAL- ELECTRONIC 0.1 g to 20 kg 0.10 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of Accuracy
AND BALANCE (Class III) Readability : Class E2 & F1.
100 mg Procedure as per OIML
R-76: 2006.

283 MECHANICAL- ELECTRONIC 10 g to 100 kg 10.0 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class III)Readability : Class M1 and
10 g procedure as per OIML
R76:2006

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 53 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
284 MECHANICAL- ELECTRONIC 1 g to 100 kg 1.0 g Using Standard
WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE Readability : 1 g Class M1, and
procedure as per OIML
R76:2006.

285 MECHANICAL- ELECTRONIC 1 mg to 5.1 g 0.005 mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE ( Class I) Readability : Class E1 and
0.001 mg procedure as per OIML
R-76: 2006.

286 MECHANICAL- ELECTRONIC 1 mg to 200 g 0.06 mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE ( Class I) Readability : Class E1 and
0.1 mg procedure as per OIML
R-76: 2006.

287 MECHANICAL- ELECTRONIC 1 mg to 50 g 0.011 mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE ( Class I)Readability : Class E1 and
0.01 mg procedure as per OIML
R-76: 2006.

288 MECHANICAL- ELECTRONIC 0.1 mg to 5 kg 0.011 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of Accuracy
AND BALANCE (Class II)Rwadability: Class E1& E2 .
10 mg Procedure as per OIML
R 76:2006.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 54 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
289 MECHANICAL- ELECTRONIC 0.1 g to 35 kg 0.97 g Using Standard
WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class III) Readability : Class F1& M1.
100 mg Peocedure as per
OIML R-76: 2006.

290 MECHANICAL- ELECTRONIC 0.1 mg to 10 kg 0.037 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of Accuracy
AND BALANCE (Class III) Readability : Class E1& E2.
100 mg Procedure as per OIML
R-76: 2006.

291 MECHANICAL- ELECTRONIC 2 kg to 500 kg 20.0 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class III) Readability : Class F2. Peocedure
50 g as per OIML R76:2006

292 MECHANICAL- Electronic Weighing 0.1 mg to 2 kg 0.011 mg Using Weights of


WEIGHING SCALE Balance (Class Accuracy Class E1&
AND BALANCE III)Readability: 100 mg E2 . Procedure as per
OIML R 76:2006.

293 MECHANICAL- ELECTRONIC 2 kg to 1000 kg 50.0 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class IV)Readability : Class M1, procedure
100 g as per OIML R76:2006

294 MECHANICAL- MASS - Weights (E2 1g 0.004 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 55 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
295 MECHANICAL- MASS - Weights (E2 1 mg 0.002 mg Using Standard
WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance (d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

296 MECHANICAL- MASS - Weights (E2 10 g 0.012mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.01mg).
Procedure as per OIML
R 111-1 guidelines.

297 MECHANICAL- MASS - Weights (E2 10 mg 0.002 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

298 MECHANICAL- MASS - Weights (E2 2g 0.004 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 56 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
299 MECHANICAL- MASS - Weights (E2 2 mg 0.002 mg Using Standard
WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

300 MECHANICAL- MASS - Weights (E2 20 g 0.016 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.01mg).
Procedure as per OIML
R 111-1 guidelines.

301 MECHANICAL- MASS - Weights (E2 20 mg 0.002 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

302 MECHANICAL- MASS - Weights (E2 200 g 0.099 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.01mg).
Procedure as per OIML
R 111-1 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 57 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
303 MECHANICAL- MASS - Weights (E2 200 mg 0.003 mg Using Standard
WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111- 1, 2004.

304 MECHANICAL- MASS - Weights (E2 5g 0.006 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

305 MECHANICAL- MASS - Weights (E2 5 mg 0.002 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

306 MECHANICAL- MASS - Weights (E2 50 g 0.014 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.01mg).
Procedure as per OIML
R 111-1 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 58 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
307 MECHANICAL- MASS - Weights (E2 500 mg 0.003 mg Using Standard
WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

308 MECHANICAL- MASS - Weights (F1 1 kg 0.001 g Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 1mg).
Procedure as per OIML
R 111-1 guidelines.

309 MECHANICAL- MASS - Weights (F1 500 g 0.001 g Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 1mg).
Procedure as per OIML
R 111-1 guidelines.

310 MECHANICAL- MASS - Weights (F2 2 kg 0.009 g Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 10mg).
Procedure as per OIML
R 111-1 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 59 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
311 MECHANICAL- MASS - Weights (F2 20 kg 0.090 g Using Standard
WEIGHTS Class & Coarser) Weights of accuracy
Class E2 and
Electronic weighing
Balance(d: 100mg).
Procedure as per OIML
R 111-1 guidelines.

312 MECHANICAL- MASS - Weights (M1 10 kg 0.083 g Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E2 and
Electronic weighing
Balance(d: 100mg).
Procedure as per OIML
R 111-1 guidelines.

313 MECHANICAL- MASS - 5 kg 10.0mg Using E2 Class


WEIGHTS WEIGHTS(Accuracy Standard Weights and
Class F2 and coarser) Electronic weighing
balance(d: 10mg).
Procedure as per OIML
R111-1.

314 MECHANICAL- MASS - Weights(E2 100 g 0.089 mg Using Standard


WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.01mg).
Procedure as per OIML
R 111-1 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 60 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
315 MECHANICAL- MASS Weights (E2 50 mg 0.002 mg Using Standard
WEIGHTS Class & Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

316 MECHANICAL- Weights (E2 Class & 100 mg 0.003 mg Using Standard
WEIGHTS Coarser) Weights of accuracy
Class E1 and
Electronic weighing
Balance(d: 0.001mg).
Procedure as per OIML
R 111-1 guidelines.

317 OPTICAL- OPTICAL Lux / Light Meter 100 lx to 10000 lx 3.5% Using Light Chamber
and Lux Meter By
Comparison Method

318 THERMAL- SPECIFIC Humidity (Digital & 20 %RH to 95 %RH@ 1.48%RH Using Standard
HEAT & HUMIDITY Analog Thermo 25 ºC Humidity Meter with
hygrometer) Humidity Generator
Temperature &
Humidity Meter,
Humidity Indicator,
Humidity &
Temperature
Transmitter with
Indicator

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 61 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
319 THERMAL- SPECIFIC Humidity (Digital & 5 °C to 50 °C@60%RH 0.36°C Using Standard
HEAT & HUMIDITY Analog Thermo Humidity Meter with
hygrometer) Humidity Generator
Temperature &
Humidity Meter,
Humidity Indicator,
Humidity &
Temperature
Transmitter with
Indicator

320 THERMAL- Glass Thermometer -80 °C to 150 °C 0.72°C Using SPRT with
TEMPERATURE Indicator & oil

321 THERMAL- Indicators of Oil Baths / 140 °C to 600 °C 0.35°C Using SPRT & S- Type
TEMPERATURE Low & High Temp Bath Thermocouple with
/ Dry bath / furnace / digital Indicator
oven/ incubators /
autoclave for industrial
applications - single
position

322 THERMAL- Indicators of Oil Baths / -95 °C to 140 °C 0.17°C Using SPRT & S-Type
TEMPERATURE Low & High Temp Bath Thermocouple with
/ Dry bath /furnace/ digital indicator
oven/ incubators /
autoclave for industrial
applications - single
position

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 62 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
323 THERMAL- Indicators of Oil Baths / 600 °C to 1200 °C 2.47°C Using SPRT & S-Type
TEMPERATURE Low & High Temp Bath Thermocouple with
/ Dry bath/ furnace / digital Indicator
oven / autoclave /
incubators for industrial
applications - single
position

324 THERMAL- Non-Contact 50 °C to 650 °C 6.19°C Using Standard Non-


TEMPERATURE Thermometer Contact IR Pyrometer,
Pyrometer Black Body (Emissivity
0.950)

325 THERMAL- Non-Contact 650 °C to 1200 °C 7.88°C Using Standardr Non-


TEMPERATURE Thermometer Contact IR Pyrometer,
Pyrometer Black Body (Emissivity
0.950)

326 THERMAL- Temperature (Digital & 5 °C to 50 °C 0.18 °C Using SPRT with


TEMPERATURE Analog Thermo Indicator & temperature
hygrometer) Generator
Temperature &
Humidity Meter,
Humidity Indicator,
Humidity &
Temperature
Transmitter with
Indicator

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 63 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
327 THERMAL- Thermocouples & RTD 140 °C to 600 °C 0.64°C Using SPRT with
TEMPERATURE Sensors/ Temperature Indicator & Dry Block
Sensors with indicators Bath
/ Temperature gauges /
Transmitters /
Temperature Switches
& Transmitters with
Indicatators

328 THERMAL- Thermocouples & RTD 600 °C to 1200 °C 2.53°C Using Fluke S Type
TEMPERATURE Sensors/ Temperature Thermocouple with
Sensors with indicators Indicator & Dry Block
/ Temperature gauges / Bath
Transmitters /
Temperature Switches
& Transmitters with
Indicatators

329 THERMAL- Thermocouples & RTD -95 °C to 140 °C 0.18°C Using SPRT with
TEMPERATURE Sensors/ Temperature Indicator & Dry block
Sensors with indicators Bath
/ Temperature gauges /
Transmitters /
Temperature Switches
& Transmitters with
Indicatators

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 64 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

1 ELECTRO- Capacitance (1 kHz) 1 nF to 100 nF 0.056% to 0.061% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

2 ELECTRO- Capacitance (1 kHz) 100 nF to 1 µF 0.05% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

3 ELECTRO- Capacitance (1kHz) 1 µF to 100 µF 0.05% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

4 ELECTRO- Current (300 Hz to 10 1 A to 20 A 0.079% to 0.102% Direct Method with


TECHNICAL- kHz) using master 8½ DMM
ALTERNATING Fluke 8508 A
CURRENT (< 1 GHZ)
(Measure)

5 ELECTRO- Current (300 Hz to 10 100 µA to 100 mA 0.03% Direct Method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 65 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

6 ELECTRO- Current (300 Hz to 10 100 mA to 1 A 0.041% to 0.079% Direct Method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

7 ELECTRO- Current (50 Hz) 20 A to 1000 A 0.5% to 1.0% Direct method master
TECHNICAL- Using Shunt With DMM
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

8 ELECTRO- Inductance (1kHz) 100 µH to 100 mH 0.236% to 0.065% Direct Method using
TECHNICAL- master LCR meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

9 ELECTRO- Inductance (1kHz) 100 mH to 10 H 0.065% to 0.073% Direct Method using


TECHNICAL- LCR Meter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

10 ELECTRO- Voltage (20 Hz to 50 1 mV to 100 mV 0.820% to 0.014% Direct method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 66 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

11 ELECTRO- Voltage (20 Hz to 50 1 V to 100 V 0.017% to 0.011% Direct Method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508A
CURRENT (< 1 GHZ)
(Measure)

12 ELECTRO- Voltage (20 Hz to 50 100 mV to 1 V 0.014% to 0.017% Direct method using


TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508A
CURRENT (< 1 GHZ)
(Measure)

13 ELECTRO- Voltage (45 Hz to 1 100 V to 1000 V 0.01% to 0.007% Direct Method using
TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

14 ELECTRO- Voltage (50 Hz) 1 kV to 25 kV 6.33% to 7.72% Direct Method by using


TECHNICAL- master H.V probe
ALTERNATING Fluke 80K-40 with
CURRENT (< 1 GHZ) DMM Fluke 17B
(Measure)

15 ELECTRO- Voltage (50 kHz to 500 1 V to 20 V 0.07% to 0.230% Direct Method using
TECHNICAL- kHz) master 8½ DMM Fluke
ALTERNATING 8508 A
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 67 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

16 ELECTRO- Capacitance (1 kHz) 220 pF to 300 nF 5.79% to 0.27% Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & Decade Capacitance
(Source) Box

17 ELECTRO- Capacitance (100 Hz) 1 µF to 30 µF 1.55% to 0.56% Direct Method using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & Decade Capacitance
(Source) Box

18 ELECTRO- Capacitance(1 kHz) 300 nF to 100 µF 0.27% to 1.13% Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & Decade Capacitance
(Source) Box

19 ELECTRO- Current (1 kHz) 3.3 mA to 1 A 0.15% to 0.073% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

20 ELECTRO- Current (1 kHz) 30 µA to 330 µA 0.54% to 0.17% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 68 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

21 ELECTRO- Current (1 kHz) 330 µA to 3.3 mA 0.17% to 0.15% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

22 ELECTRO- Current (10Hz to 1kHz) 330 µA to 3.3 mA 0.25% to 0.26% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

23 ELECTRO- Current (10Hz to 1kHz) 3.3 mA to 3 A 0.26% to 0.23% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

24 ELECTRO- Current (1kHz) 1 A to 10 A 0.073% to 0.12% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

25 ELECTRO- Current (45 Hz to 5 3 A to 20 A 0.23% to 0.13% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 69 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

26 ELECTRO- Current (50 Hz) 20 A to 1000 A 0.68% to 0.37% Direct Method by using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ) & current coil
(Source)

27 ELECTRO- Inductance (1 kHz) 100 µH to 10 H 2.35 % to 1.14 % Direct Method using


TECHNICAL- master Decade
ALTERNATING Inductance Box
CURRENT (< 1 GHZ)
(Source)

28 ELECTRO- Power 50 Hz @ UPF 1.2 W to 0.16% Direct Method using


TECHNICAL- 4.8 kW,0.01 A to 20 master Standard
ALTERNATING A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

29 ELECTRO- Power 50Hz @ 0.5 Lag, 6 W 0.46% Direct Method using


TECHNICAL- to 2.4 kW,0.01 A to 20 master Standard
ALTERNATING A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

30 ELECTRO- Power 50Hz @ 0.8 Lead 9.6 0.20% Direct Method using
TECHNICAL- W to 3.8 kW,0.01 A to master Standard
ALTERNATING 20 A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 70 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

31 ELECTRO- Power 50Hz @0.2 Lag, 2.4 W 0.56% Direct Method using
TECHNICAL- to 960 W,0.01 A to 20 master Standard
ALTERNATING A,120 V to 240 V Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

32 ELECTRO- Power Factor 0.2 Lag to UPF 0.003 Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

33 ELECTRO- Power Factor 0.2 Lead to UPF 0.0003 Direct Method using
TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

34 ELECTRO- Voltage (10 Hz to 10 1 V to 30 V 0.080% to 0.042% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

35 ELECTRO- Voltage (10 Hz to 10 100 mV to 1 V 0.086% to 0.080% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 71 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

36 ELECTRO- Voltage (10 Hz to 10 30 mV to 100 mV 0.12% to 0.086% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

37 ELECTRO- Voltage (100kHz) 30 mV to 30 V 0.452% to 0.054% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

38 ELECTRO- Voltage (100kHz) 30 V to 100 V 0.054% to 0.024% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

39 ELECTRO- Voltage (45 Hz to 1 30 V to 300 V 0.20% to 0.022% Direct Method by using


TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

40 ELECTRO- Voltage (45 Hz to 1 300 V to 1000 V 0.022% to 0.035% Direct Method by using
TECHNICAL- kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 72 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

41 ELECTRO- Voltage (450kHz) 30 mV to 3 V 1.116% to 0.121% Direct Method by using


TECHNICAL- master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

42 ELECTRO- Voltage (45Hz to 1 mV to 3 mV 0.712% to 0.24% Direct Method by using


TECHNICAL- 10kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

43 ELECTRO- Voltage (45Hz to 3 mV to 30 mV 0.24% to 0.042% Direct Method by using


TECHNICAL- 10kHz) master Standard
ALTERNATING Calibrator Fluke 5522A
CURRENT (< 1 GHZ)
(Source)

44 ELECTRO- Current 1 A to 20 A 0.080% to 0.052% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

45 ELECTRO- Current 1 mA to 100 mA 0.002% to 0.008% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

46 ELECTRO- Current 10 µA to 100 µA 0.0010% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 73 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

47 ELECTRO- Current 100 µA to 1 mA 0.001% to 0.002% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

48 ELECTRO- Current 100 mA to 1 A 0.062% to 0.080% Direct method by using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

49 ELECTRO- Current 20 A to 1000 A 1.19% Direct Method using


TECHNICAL- DIRECT 8½ DMM Fluke 8508
CURRENT (Measure) with Shunt

50 ELECTRO- Resistance 1 Gohm to 10 Gohm 0.116% to 0.133% Direct Method by Using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

51 ELECTRO- Resistance 1 Mohm to 10 Mohm 0.001% to 0.002% Direct Method by Using


TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

52 ELECTRO- Resistance 1 mohm to 100 mohm 0.061% to 0.207% Direct Method by Using
TECHNICAL- DIRECT master LCR Meter &
CURRENT (Measure) DRB

53 ELECTRO- Resistance 1 ohm to 100 ohm 0.0015% to 0.0010% Using LCR Meter &
TECHNICAL- DIRECT DRB
CURRENT (Measure)

54 ELECTRO- Resistance 10 kohm to 100 kohm 0.0007% to 0.0008% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 74 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

55 ELECTRO- Resistance 10 Mohm to 100 Mohm 0.002% to 0.008% Direct method by using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

56 ELECTRO- Resistance 100 Mohm to 1 Gohm 0.008% to 0.116% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

57 ELECTRO- Resistance 100 mohm to 150 0.206% Direct Method by Using


TECHNICAL- DIRECT mohm master LCR Meter &
CURRENT (Measure) DRB

58 ELECTRO- Resistance 100 ohm to 10 kohm 0.0010% to 0.0007% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

59 ELECTRO- Resistance 100 kohm to 1 Mohm 0.0008% to 0.001% Direct Method by Using
TECHNICAL- DIRECT master 8½ DMM Fluke
CURRENT (Measure) 8508 A

60 ELECTRO- Resistance 150 mohm to 1 ohm 0.206% to 0.0015% Direct Method by using
TECHNICAL- DIRECT maste LCR Meter &
CURRENT (Measure) DRB

61 ELECTRO- Voltage 1 kV to 20 kV 1.70% to 1.67% Direct Method by using


TECHNICAL- DIRECT master H.V probe
CURRENT (Measure) Fluke 80K-40 with
DMM Fluke 17B

62 ELECTRO- Voltage 1 mV to 100 mV 0.44% to 0.0003% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Measure) Calibrator Fluke 5522A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 75 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

63 ELECTRO- Voltage 1 V to 100 V 0.0002% to 0.0003% Direct Method by using


TECHNICAL- DIRECT master DMM Fluke
CURRENT (Measure) 8508A

64 ELECTRO- Voltage 100 mV to 1 V 0.0003% to 0.0003% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Measure) Calibrator Fluke 5522A

65 ELECTRO- Voltage 100 V to 1000 V 0.0004% to 0.0004% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Measure) Calibrator Fluke 5522A

66 ELECTRO- Voltage 20 kV to 40 kV 1.67% to 1.65% Direct Method by using


TECHNICAL- DIRECT master H.V probe
CURRENT (Measure) Fluke 80K-40 with
DMM Fluke 17B

67 ELECTRO- Current 1 A to 10 A 0.027% to 0.063% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

68 ELECTRO- Current 1 mA to 100 mA 0.016% to 0.011% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

69 ELECTRO- Current 10 µA to 100 µA 0.25% to 0.023% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

70 ELECTRO- Current 10 A to 20 A 0.063% to 0.11% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 76 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

71 ELECTRO- Current 100 µA to 1 mA 0.023% to 0.016% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

72 ELECTRO- Current 100 mA to 1 A 0.011% to 0.027% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

73 ELECTRO- Current 20 A to 200 A 0.5% Direct Method using


TECHNICAL- DIRECT Current Source
CURRENT (Source)

74 ELECTRO- DC Power 10 W to 1 kW,1V to 0.08% to 0.45% Using Calibrator Fluke


TECHNICAL- DIRECT 1000V, 1mA to 1000A. 5522A with Current
CURRENT (Source) Coil Direct Method.

75 ELECTRO- DC Power (1V to 1 KW to 19.9 KW 0.45% to 0.65% Using Calibrator Fluke


TECHNICAL- DIRECT 1000V, 1mA to 1000A) 5522A with current coil
CURRENT (Source) Direct Method.

76 ELECTRO- Oscilloscope 1 mV to 130 V 0.29% to 0.096% Direct Method


TECHNICAL- DIRECT
CURRENT (Source)

77 ELECTRO- Resistance 0.1 Mohm to 1 Gohm 5.89% to 4.91% Direct Method using
TECHNICAL- DIRECT master of high stability
CURRENT (Source) Decade Megohm Box

78 ELECTRO- Resistance 1 Gohm to 100 Gohm 4.91% to 8.61% Direct Method using
TECHNICAL- DIRECT master of high stability
CURRENT (Source) Decade Megohm Box

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 77 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

79 ELECTRO- Resistance 1 kohm to 10 kohm 0.0036% to 0.0035% Direct Method using


TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

80 ELECTRO- Resistance 1 mohm 0.609% Direct Method using


TECHNICAL- DIRECT fixed resistor box
CURRENT (Source)

81 ELECTRO- Resistance 1 Mohm to 10 Mohm 0.0041% to 0.014% Direct Method using


TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

82 ELECTRO- Resistance 1 ohm to 10 ohm 0.10% to 0.015% Direct Method using


TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

83 ELECTRO- Resistance 10 kohm to 100 kohm 0.0035% to 0.0034% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

84 ELECTRO- Resistance 10 mohm 0.610% Direct Method using


TECHNICAL- DIRECT fixed resistor box
CURRENT (Source)

85 ELECTRO- Resistance 10 Mohm to 300 Mohm 0.014% to 0.38% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

86 ELECTRO- Resistance 10 ohm to 100 ohm 0.015% to 0.004% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 78 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

87 ELECTRO- Resistance 100 Gohm to 1 Tohm 5.81% Direct Method using


TECHNICAL- DIRECT master of high stability
CURRENT (Source) Decade Megohm Box

88 ELECTRO- Resistance 100 ohm to 1 kohm 0.004% to 0.0036% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

89 ELECTRO- Resistance 100 kohm to 1 Mohm 0.0034% to 0.0041% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

90 ELECTRO- Resistance 150 mohm 0.609% Direct Method using


TECHNICAL- DIRECT fixed resistor box
CURRENT (Source)

91 ELECTRO- Resistance 150 mohm to 1 ohm 0.611% to 0.10% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A:

92 ELECTRO- Resistance 300 Mohm to 1 Gohm 0.38% to 1.78% Direct Method using
TECHNICAL- DIRECT Standard Calibrator
CURRENT (Source) Fluke 5522A

93 ELECTRO- Resistance 50 mohm 0.610% Direct Method using


TECHNICAL- DIRECT sing fixed resistor box
CURRENT (Source)

94 ELECTRO- Voltage 1 mV to 10 mV 0.240% to 0.023% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 79 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

95 ELECTRO- Voltage 1 V to 10 V 0.0015% to 0.0016% Direct Method using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A:

96 ELECTRO- Voltage 10 mV to 100 mV 0.023% to 0.0039% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

97 ELECTRO- Voltage 10 V to 1000 V 0.0016% to 0.0023% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

98 ELECTRO- Voltage 100 mV to 1 V 0.0039% to 0.0015% Direct Method by using


TECHNICAL- DIRECT master Standard
CURRENT (Source) Calibrator Fluke 5522A

99 ELECTRO- Amplitude-Sq wave / 1 mV to 55 Vp-p 0.6% to 0.3% Direct Method by using


TECHNICAL- Oscilloscope (1 kHz) master Standard
ELECTRICAL Calibrator Fluke 5522A
EQUIPMENT (Source)

100 ELECTRO- Bandwidth - 50 kHz to 1 GHz 0.23% to 0.87% Direct Method using
TECHNICAL- Oscilloscope master Standard
ELECTRICAL Calibrator Fluke 5522A
EQUIPMENT (Source)

101 ELECTRO- Temperature Indicator / 600 °C 0.53°C to 0.39°C Direct Method by using
TECHNICAL- Recorder / Controller to 1750 °C Master Standard
TEMPERATURE (B type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 80 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

102 ELECTRO- Temperature Indicator / -200 °C to 1200 °C 0.28°C to 0.31°C Direct Method by using
TECHNICAL- Recorder / Controller (J Master used- Stnd
TEMPERATURE type) Calibrator Fluke 5522A
SIMULATION & 6½ DMM
(Measure)

103 ELECTRO- Temperature Indicator / -200 °C to 1350 °C 0.26°C to 0.48°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (K type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

104 ELECTRO- Temperature Indicator / -200 °C to 1300 °C 0.48°C to 0.35°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (N type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

105 ELECTRO- Temperature Indicator / 10 °C to 1750 °C 0.67°C to 0.48°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (R type) Calibrator fluke 5522A
SIMULATION
(Measure)

106 ELECTRO- Temperature Indicator / -200 °C to 800 °C 0.05 °C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (RTD type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 81 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

107 ELECTRO- Temperature Indicator / 10 °C to 1450 °C 0.56°C to 0.54°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (S type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

108 ELECTRO- Temperature Indicator / -200 °C to 400 °C 0.60°C to 0.17°C Direct Method by using
TECHNICAL- Recorder / Controller (T Master Standard
TEMPERATURE type) Calibrator fluke 5522A
SIMULATION & 6½ DMM
(Measure)

109 ELECTRO- Temperature Indicator / 600 °C to 1750 °C 0.47°C to 0.11°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (B type) Calibrator fluke 5522A
SIMULATION (Source)

110 ELECTRO- Temperature Indicator / -200 °C to 900 °C 0.20°C to 0.45°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (E type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

111 ELECTRO- Temperature Indicator / -200 °C to 1200 °C 0.15°C to 0.25°C Direct Method by using
TECHNICAL- Recorder / Controller (J Master Standard
TEMPERATURE type) Calibrator fluke 5522A
SIMULATION (Source)

112 ELECTRO- Temperature Indicator / -200 °C to 1350 °C 0.37°C to 0.48°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (K type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 82 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

113 ELECTRO- Temperature Indicator / -200 °C to 1300 °C 0.45°C to 0.30°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (N Type ) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

114 ELECTRO- Temperature Indicator / 10 °C to 1750 °C 0.62°C to 0.44°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (R type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

115 ELECTRO- Temperature Indicator / -200 °C to 800 °C 0.071°C to 0.083°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (RTD type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

116 ELECTRO- Temperature Indicator / 10 °C to 1750 °C 0.61°C to 0.51°C Direct Method by using
TECHNICAL- Recorder / Controller Master Standard
TEMPERATURE (S type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

117 ELECTRO- Temperature Indicator / -200 °C to 400 °C 0.55°C to 0.16°C Direct Method by using
TECHNICAL- Recorder / Controller (T Master Standard
TEMPERATURE type) Calibrator fluke 5522A
SIMULATION (Source) & 6½ DMM

118 ELECTRO- Frequency 10 Hz to 1 GHz 0.0011% to 0.0024% Direct Method using


TECHNICAL- TIME & master Frequency
FREQUENCY Counter Philips- PM
(Measure) 6669

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 83 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

119 ELECTRO- Timer 1 sec to 9000 sec 0.06% to 0.05% Comparison Method by
TECHNICAL- TIME & using master Digital
FREQUENCY Timer Beltronics - 501
(Measure)

120 ELECTRO- Frequency 1 Hz to 1 GHz 0.004% to 0.006% Direct Method using


TECHNICAL- TIME & master Standard
FREQUENCY (Source) Calibrator Fluke 5522A

121 ELECTRO- Timer Marker / 1 ns to 5 s 0.0003 sec Direct Method using


TECHNICAL- TIME & Oscilloscope master Standard
FREQUENCY (Source) Calibrator Fluke 5522A

122 MECHANICAL- Centrifuge / Centrifugal 10 RPM to 15000 RPM 1.1% to 1.2% By Comparison method
ACCELERATION AND Motor / Speed (Non- by using Tachometer
SPEED Contact)

123 MECHANICAL- Bench Centre Upto 500mm 7.0µm Using Taper Mandrel
DIMENSION (BASIC (Parallelity and
MEASURING coaxiality)
INSTRUMENT,
GAUGE ETC.)

124 MECHANICAL- Height Measuring 0 mm to 1000 mm 5.4µm Using Gauge Block Set
DIMENSION (BASIC System, Res: 0.1µm
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 84 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

125 MECHANICAL- Profile Projector / Video 50 min of arc Using Glass Angle
0 ° to 360 °
DIMENSION (BASIC Measuring System / gauge, Glass Scale
MEASURING Micro Scope
INSTRUMENT,
GAUGE ETC.)

126 MECHANICAL- Profile Projector / Video Mag.: 10x to 100x 0.6% Using Glass Angle
DIMENSION (BASIC Measuring System / gauge, Glass Scale
MEASURING Micro Scope
INSTRUMENT,
GAUGE ETC.)

127 MECHANICAL- Profile Projector / Video 0 mm to 200 mm 2.4 µm Using Glass Angle
DIMENSION (BASIC Measuring System / gauge, Glass Scale
MEASURING Micro Scope (L.C.
INSTRUMENT, 0.001 mm)
GAUGE ETC.)

128 MECHANICAL- Surface Plate ( Granite 6000 mm to 2000 mm 2.0 sqrt(L+ W/150)µm, Using Electronic Level
DIMENSION (BASIC / Cast Iron Surface L W in mm
MEASURING Plate)
INSTRUMENT,
GAUGE ETC.)

129 MECHANICAL- Tape Scale Calibrator 0 mm to 1000 mm 10.2µm Using Slip Gauge Set
DIMENSION (BASIC and Long Slip Gauges
MEASURING
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 85 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

130 MECHANICAL- Brinell Hardness HBW 5/750 & HBW 7.1% & 3.65% Using Standard
HARDNESS TESTING Testing Machine 10/3000 Hardness Blocks
MACHINES IS:1500 (Part 2):2013

131 MECHANICAL- Rockwell Hardness HRA, HRBW, HRC 1.0 HRA, 1.2 HRBW, Using Standard
HARDNESS TESTING 0.85 HRC Hardness Blocks IS
MACHINES 1586(Part2):2018

132 MECHANICAL- Vickers Hardness HV1, HV5, HV10, 5.12%HV1, 3.72 HV5, Using Standard
HARDNESS TESTING Testing Machine HV30 2.5%HV10, 2.0% HV30 Hardness Blocks
MACHINES IS:1501(Part2)2013

133 MECHANICAL- Digital / Dial Vaccum 0 bar to (-) 0.75 bar 0.014 bar Digital Pressure Gauge
PRESSURE Gauges / Transmitters / with uncertainty of
INDICATING DEVICES Transducers. 0.0008 bar. Procedure
based on ISO 3567 &
ISO 27893 standards.

134 MECHANICAL- PRESSURE 0 bar to 1000 bar 0.15 bar Digital Pressure gauge
PRESSURE (HYDRAULIC)Digital / with uncertainty of
INDICATING DEVICES Analog Pressure 0.021 bar. Procedure
Gauges, Pressure based on DKD-R 6-1
Indicator With guidelines.
Transducer /
Transmitter, Pressure
Switch

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 86 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

135 MECHANICAL- PRESSURE 0 bar to 700 bar 0.29 bar Digital Pressure gauge
PRESSURE (HYDRAULIC)Digital / with uncertainty of
INDICATING DEVICES Analog Pressure 0.182 bar. Procedure
Gauges, Pressure based on DKD-R 6-1
Indicator With guidelines.
Transducer /
Transmitter & Pressure
Switches

136 MECHANICAL- PRESSURE 0 bar to 20 bar 0.014 bar Digital Pressure gauge
PRESSURE (PNEUMATIC) Digital / with uncertainty of
INDICATING DEVICES Dial Gauges / Pressure 0,01bar. Procedure
Indicator With based on DKD-R 6-1
Transmitter / guidelines.
Transducer.

137 MECHANICAL- PRESSURE 0 mbar to 500 mbar 0.31 mbar Digital Manometer with
PRESSURE (PNEUMATIC) uncertainty of 0.06
INDICATING DEVICES Pressure gauges, mbar. Procedure based
Magnehelic gauges, on DKD-R 6-1
Differential Pressure guidelines.
Transmitters, Low
Pressure Indicators/
Transducers,
Barometer

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 87 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

138 MECHANICAL- PRESSURE 0 bar to 70 bar 0.04 bar Digital Pressure gauge
PRESSURE (PNEUMATIC)Digital / with uncertainty of
INDICATING DEVICES Dial Gauges / Pressure 0.022 bar. Procedure
Indicator With based on DKD-R 6-1
Transmitter / guidelines.
Transducer.

139 MECHANICAL- PRESSURE 0 mbar to 2000 mbar 2.14 mbar Digital Manometer with
PRESSURE (PNEUMATIC)Pressur uncertainty of 1.03
INDICATING DEVICES e gauges, Magnehelic mbar. Procedure based
gauges, Differential on DKD-R 6-1
Pressure Transmitters, guidelines.
Low Pressure
Indicators/
Transducers,
Barometer.

140 MECHANICAL- ELECTRONIC 1 mg to 2.1 g 0.003mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of acccuracy
AND BALANCE ( Class I) Readability : Class E1 and
0.0001 mg procedure based OIML
R-76: 2006.

141 MECHANICAL- ELECTRONIC 1 mg to 1 kg 0.002 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class Ii) Readability : Class E1 & E2, and
1 mg procedure as per OIML
R-76: 2006.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 88 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

142 MECHANICAL- ELECTRONIC 0.1 g to 20 kg 0.10 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of Accuracy
AND BALANCE (Class III) Readability : Class E2 & F1.
100 mg Procedure as per OIML
R-76: 2006.

143 MECHANICAL- ELECTRONIC 10 g to 100 kg 10.0 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class III)Readability : Class M1 and
10 g procedure as per OIML
R76:2006

144 MECHANICAL- ELECTRONIC 1 mg to 5.1 g 0.005 mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE ( Class I) Readability : Class E1 and
0.001 mg procedure as per OIML
R-76: 2006.

145 MECHANICAL- ELECTRONIC 1 mg to 200 g 0.06 mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE ( Class I) Readability : Class E1 and
0.1 mg procedure as per OIML
R-76: 2006.

146 MECHANICAL- ELECTRONIC 1 mg to 50 g 0.011 mg Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE ( Class I)Readability : Class E1 and
0.01 mg procedure as per OIML
R-76: 2006.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 89 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

147 MECHANICAL- ELECTRONIC 0.1 mg to 5 kg 0.011 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of Accuracy
AND BALANCE (Class II)Rwadability: Class E1& E2 .
10 mg Procedure as per OIML
R 76:2006.

148 MECHANICAL- ELECTRONIC 0.1 g to 35 kg 0.97 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class III) Readability : Class F1& M1.
100 mg Peocedure as per
OIML R-76: 2006.

149 MECHANICAL- ELECTRONIC 0.1 mg to 10 kg 0.037 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of Accuracy
AND BALANCE (Class III) Readability : Class E1& E2.
100 mg Procedure as per OIML
R-76: 2006.

150 MECHANICAL- ELECTRONIC 2 kg to 500 kg 20.0 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class III) Readability : Class F2. Peocedure
50 g as per OIML R76:2006

151 MECHANICAL- Electronic Weighing 0.1 mg to 2 kg 0.011 mg Using Weights of


WEIGHING SCALE Balance (Class Accuracy Class E1&
AND BALANCE III)Readability: 100 mg E2 . Procedure as per
OIML R 76:2006.

152 MECHANICAL- ELECTRONIC 2 kg to 1000 kg 50.0 g Using Standard


WEIGHING SCALE WEIGHING BALANCE Weights of accuracy
AND BALANCE (Class IV)Readability : Class M1, procedure
100 g as per OIML R76:2006

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 90 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

153 THERMAL- SPECIFIC Humidity Chambers / 20 %RH to 95 %RH 2.6%RH, 0.7°C Using Data Loggers by
HEAT & HUMIDITY Climatic Chamber / @25°C& 10°C to 50°C Multi channel Method
Environmental @ 50%RH
Chamber/Humidity
Generator/Humidity
Source(Single or Multi
Point)

154 THERMAL- Thermal Freezers -80 °C to 200 °C 1.6°C Using RTD / TC and
TEMPERATURE Chambers / Ovens / data Logger (Mapping)
Incubators (Industrial
Purpose) &
Furnace,Cold room,
Freezer, Deep Freezer,
Temperature
Chamber,Hot Air Oven,
Furnace, Liquid Bath,
Muffle Furnace (Multi
Point)

155 THERMAL- Thermal Freezers/ 200 °C to 1200 °C 2.5°C Using RTD


TEMPERATURE Chambers / Ovens / Thermometer & N Type
Incubators (Industrial Thermocouples with
Purpose) & Furnance - Multichannel
multipoint Temperature Recorder

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name VAIDYANATHESHWARA INSTRUMENTS, NO. 301/A, 9TH MAIN ROAD, 3RD
CROSS, RAJIV GANDHI NAGAR, BENGALURU, KARNATAKA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2473 Page No. : 91 / 91

Validity 03/07/2020 to 26/10/2021 Last Amended on -

S.No Discipline / Group Measurand or Measurement range * Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

156 THERMAL- Thermocouples & RTD 140 °C to 600 °C 0.64°C Using SPRT with
TEMPERATURE Sensors/ Temperature Indicator & Dry Block
Sensors with indicators Bath
/ Temperature gauges /
Transmitters /
Temperature Switches
& Transmitters with
Indicatators

157 THERMAL- Thermocouples & RTD 600 °C to 1200 °C 2.53°C Using Fluke S Type
TEMPERATURE Sensors/ Temperature Thermocouple with
Sensors with indicators Indicator & Dry Block
/ Temperature gauges / Bath
Transmitters /
Temperature Switches
& Transmitters with
Indicatators

158 THERMAL- Thermocouples & RTD -95 °C to 140 °C 0.18°C Using SPRT with
TEMPERATURE Sensors/ Temperature Indicator & Dry block
Sensors with indicators Bath
/ Temperature gauges /
Transmitters /
Temperature Switches
& Transmitters with
Indicatators

* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.

This is annexure to 'Certificate of Accreditation' and does not require any signature.

You might also like