You are on page 1of 26

CALSYTECH, NO.

38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL


Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 1 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Permanent Facility
ELECTRO-
TECHNICAL-
Precision Shunt with
Alternating
1 AC Current@50Hz Multimeter by Direct 20 A to 100 A 1.63%
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Current@50Hz to
2 Transmille 8081 by 100 mA to 20 A 0.077 % to 0.164 %
Current (< 1 1kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Current@50Hz to
3 Transmille 8081 by 20 µA to 100 mA 0.15 % to 0.077 %
Current (< 1 1kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL- AC Energy- Active
Energy Calibration
Alternating (1Ø and 3Ø) (COSØ 63.5 V to 300 V@ 1A
4 Source Educetech by 0.24%
Current (< 1 = 0.5 TO UPF) @ to 15A
Direct Method
GHz) 50Hz
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 2 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- High Voltage Probe
Alternating AC High Voltage @ Fluke 80K40 with
5 1 kV to 28 kV 7 % to 8.38 %
Current (< 1 50Hz DMM by comparison
GHz) Method
(Measure)
ELECTRO-
TECHNICAL-
AC Digital Power Meter 110 W to 6000
Alternating
6 POWER@50Hz(UPF Yokogawa by Direct W,110V to 400V@1A 0.38 % to 0.36 %
Current (< 1
1Ø) Method to 15A
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Voltage @1kHz
7 Transmille 8081 by 100 mV to 700 V 0.081 % to 0.05 %
Current (< 1 to 10kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Voltage @50Hz
8 Transmille 8081 by 100 V to 1000 V 0.035 % to 0.032 %
Current (< 1 to 1kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Voltage @50Hz
9 Transmille 8081 by 50 mV to 100 V 0.054 % to 0.035 %
Current (< 1 to 1kHz
Direct Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 3 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Digital Scientific
Alternating
10 Capacitance@1kHz make LCR Meter by 1 nF to 1 µF 0.68%
Current (< 1
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Digital Scientific LCR
Alternating
11 Inductance@1kHz Meter by Direct 5 mH to 10 H 1.4%
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
Multi Product
TECHNICAL-
AC Current @ 50Hz Calibrator Transmille
12 Alternating 1 A to 20 A 0.14 % to 0.20 %
to 1kHz 3050 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Current @ 50Hz Calibrator Transmille
13 Alternating 100 µA to 1 A 0.89 % to 0.14 %
to 1kHz 2041 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Current @ 50Hz Calibrator Transmille
14 Alternating 20 µA to 100 µA 4.01 % to 0.89 %
to 1kHz 2041 by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 4 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Multi Product
TECHNICAL-
AC Current @ 50Hz Calibrator with
15 Alternating 20 A to 1000 A 0.77 % to 0.58 %
to 60Hz current coils by
Current (< 1
Direct Method
GHz) (Source)
ELECTRO-
TECHNICAL- AC POWER(1Ø and using Energy Cal 63.5 W to 6000 W
16 Alternating 3Ø) (COSØ = 0.5 TO Source by Direct ,63.5 V to 300V@1A 0.24%
Current (< 1 UPF) @ 50Hz Method to 20A
GHz) (Source)
ELECTRO-
TECHNICAL- Discrete Standard
AC 100 ohm, 1k & 100k
17 Alternating Resistors by Direct 0.164%
Resistance@1kHz Ohm
Current (< 1 Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Voltage @ 1kHz Calibrator Transmille
18 Alternating 10 V to 700 V 0.2 % to 0.34 %
to 10kHz 2041 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Voltage @ 1kHz Calibrator Transmille
19 Alternating 100 mV to 10 V 0.22 % to 0.2 %
to 10kHz 2041 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Voltage @ 50Hz Calibrator Transmille
20 Alternating 1 V to 1000 V 0.081 % to 0.09 %
to 1kHz 2041 by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 5 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Multi Product
TECHNICAL-
AC Voltage @ 50Hz Calibrator Transmille
21 Alternating 20 mV to 1 V 0.36 % to 0.081 %
to 1kHz 2041 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance @ 1
22 Alternating Capacitance box by 1 nF to 10 µF 0.4 % to 1 %
kHz
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
TECHNICAL-
Inductance Box by
23 Alternating Inductance@1kHz 1 mH to 10 H 5.43%
Direct Method
Current (< 1
GHz) (Source)
ELECTRO-
TECHNICAL- Energy Cal Source
24 Alternating Power Factor Educetech by Direct 0.5 PF to 1 PF 0.002PF
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL-
Precision Multimeter
25 DIRECT DC Current 1 µA to 10 µA 0.026 % to 0.046 %
by Direct Method
CURRENT
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
26 DIRECT DC Current Transmille 8081 by 1 A to 20 A 0.02 % to 0.09 %
CURRENT Direct Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 6 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Precision Multimeter
27 DIRECT DC Current Transmille 8081 by 10 µA to 100 µA 0.046 % to 0.003 %
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
28 DIRECT DC Current Transmille 8081 by 100 µA to 100 mA 0.002 % to 0.005 %
CURRENT Direct Method
(Measure)
ELECTRO-
Precision Shunt with
TECHNICAL-
Yokogawa Power
29 DIRECT DC Current 20 A to 100 A 0.63%
meter by Direct
CURRENT
Method
(Measure)
ELECTRO-
High Voltage Probe
TECHNICAL-
Fluke 80k40 with
30 DIRECT DC High Voltage 1 kV to 40 kV 3.8%
DMM by Direct
CURRENT
Method
(Measure)
ELECTRO-
Precision Multi
TECHNICAL-
meter Transmille
31 DIRECT DC Voltage 1 mV to 100 mV 0.041 % to 0.001 %
8081 by Direct
CURRENT
Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
32 DIRECT DC Voltage Transmille 8081 by 100 mV to 1000 V 0.001%
CURRENT Direct Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 7 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Precision Multimeter
Resistance (2Wire
33 DIRECT Transmille 8081 by 1 Ohm to 100 Ohm 0.003 % to 0.0015 %
/4Wire)
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Resistance(2Wire 1 MOhm to 10
34 DIRECT Transmille 8081 by 0.0021 % to 0.004 %
/4Wire) MOhm
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Resistance(2Wire
35 DIRECT Transmille 8081 by 10 mOhm to 1 Ohm 0.09 % to 0.003 %
/4Wire)
CURRENT Direct Method
(Measure)
ELECTRO-
Precision Multimeter
TECHNICAL-
Resistance(2Wire Transmille 8081 in 10 MOhm to 100
36 DIRECT 0.05 % to 0.55 %
/4Wire) electrometer mode GOhm
CURRENT
by Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Resistance(2Wire 0.0015 % to 0.0022
37 DIRECT Transmille 8081 by 100 Ohm to 1 MOhm
/4Wire) %
CURRENT Direct Method
(Measure)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator Transmille
38 DIRECT DC Current 1 µA to 10 µA 2.3 % to 0.24 %
2041 by Direct
CURRENT
method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 8 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Multi Product
39 DIRECT DC Current Calibrator by Direct 1 A to 20 A 0.03 % to 0.1 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
40 DIRECT DC Current Calibrator by Direct 10 µA to 100 µA 0.24 % to 0.034 %
CURRENT Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator Transmille
41 DIRECT DC Current 100 µA to 1 A 0.034 % to 0.031 %
2041 by Direct
CURRENT
Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator & Current
42 DIRECT DC Current 20 A to 1000 A 0.8 % to 0.35 %
Coil by Direct
CURRENT
Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator Transmille
43 DIRECT DC Voltage 1 mV to 10 mV 0.42 % to 0.05 %
2041 by Direct
CURRENT
Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator Transmille
44 DIRECT DC Voltage 10 mV to 100 mV 0.05 % to 0.011 %
2041 by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 9 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Multiproduct
TECHNICAL-
Calibrator Transmille
45 DIRECT DC Voltage 100 mV to 1000 V 0.01 % to 0.005 %
2041 by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
Power/Digital Power 10 W to 12 kW 10V
46 DIRECT Calibrator by Direct 0.03 % to 0.1 %
Meter to 600V@ 1A to 20A
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Decade Resistance
Resistance 1 Gohm to 100
47 DIRECT Box Vaiseshika by 1.22 % to 6 %
2Wire/4wire Gohm
CURRENT Direct Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Resistance Calibrator Transmille
48 DIRECT 1 kOhm to 1 MOhm 0.011 % to 0.012 %
2Wire/4wire 2041 by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
49 DIRECT 1 mOhm 0.46%
2Wire/4wire Direct Method
CURRENT
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Resistance Calibrator Transmille
50 DIRECT 1 Mohm to 1 Gohm 0.012 % to 1.23 %
2Wire/4wire 2041 by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 10 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
51 DIRECT 1 Ohm 0.13%
2Wire/4wire Direct Method
CURRENT
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Resistance Calibrator Transmille
52 DIRECT 1 ohm to 1 kOhm 0.125 % to 0.012 %
2Wire/4wire 2041by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
53 DIRECT 10 mOhm 0.46%
2Wire/4wire Direct Method
CURRENT
(Source)
ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
54 DIRECT 100 µohm 0.46%
2Wire/4wire Direct method
CURRENT
(Source)
ELECTRO-
TECHNICAL- Standard
Resistance
55 DIRECT Resistor/Direct 100 mOhm 0.46%
2Wire/4wire
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
Bandwidth/Oscillosc
56 ELECTRICAL Calibrator by Direct 10 MHz to 600 MHz 1.3dB
ope
EQUIPMENT Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 11 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
DC
TECHNICAL- Multi Product
Amplitude/Oscillosco
57 ELECTRICAL Calibrator by Direct 12 mV to 300 V DC 0.37 % to 0.045 %
pe (2mV/Div to
EQUIPMENT Method
50V/Div)
(Source)
ELECTRO-
TECHNICAL- Multi Product
Time
58 ELECTRICAL Calibrator by Direct 2 ns to 5 sec 0.088 % to 0.081 %
Base/Oscilloscope
EQUIPMENT Method
(Source)
ELECTRO-
TECHNICAL- Temperature( Precision Multimeter
59 TEMPERATURE Type"J" & Direct/ -200 ° C to 1200 °C 0.16°C
SIMULATION Thermocouple) Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
60 TEMPERATURE by Direct 0 °C to 800 °C 0.16°C
"E" Thermocouple)
SIMULATION /Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
61 TEMPERATURE by Direct -140 °C to 1340 °C 0.16°C
"K" Thermocouple )
SIMULATION /Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
62 TEMPERATURE by Direct / -180 °C to 1300 °C 0.18°C
"N" Thermocouple )
SIMULATION Comparison Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 12 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
63 TEMPERATURE by Direct -100 °C to 800 °C 0.10°C
"RTD" )
SIMULATION /Comparison Method
(Measure)
ELECTRO-
Precision Multimeter
TECHNICAL-
Temperature( Type by
64 TEMPERATURE 0 °C to 1700 °C 0.27°C
"S" Thermocouple) Direct/Comparison
SIMULATION
Method
(Measure)
ELECTRO-
TECHNICAL- Temperature(Therm Precision Multimeter
65 TEMPERATURE ocouple Type "B" by Direct 600 °C to 1800 °C 0.68°C
SIMULATION Thermocouple) /Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature(Type
66 TEMPERATURE by Direct / 0 °C to 1700 °C 0.27°C
"R" Thermocouple )
SIMULATION Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature(Type
67 TEMPERATURE by Direct 0 °C to 400 °C 0.16°C
"T" Thermocouple )
SIMULATION /Comparison Method
(Measure)
ELECTRO-
Precision Multimeter
TECHNICAL-
/ Multi function
68 TIME & Frequency 100Hz to 600MHz 0.5% to 0.024%
counter by Direct
FREQUENCY
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 13 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Universal Counter
69 TIME & Time timer by comparison 100 ms to 1000 s 0.025 s to 3.4 s
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Multi Product
70 TIME & Frequency Calibrator by Direct 10 MHz to 600 MHz 0.004%
FREQUENCY Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
71 TIME & Frequency Calibrator by Direct 45 Hz to 10 MHz 0.0055 % to 0.004 %
FREQUENCY Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 14 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Site Facility
ELECTRO-
TECHNICAL-
Precision Shunt with
Alternating
1 AC Current@50Hz Multimeter by Direct 20 A to 100 A 1.63%
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Current@50Hz to
2 Transmille 8081 by 100 mA to 20 A 0.077 % to 0.164 %
Current (< 1 1kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Current@50Hz to
3 Transmille 8081 by 20 µA to 100 mA 0.15 % to 0.077 %
Current (< 1 1kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL- AC Energy- Active
Energy Calibration
Alternating (1Ø and 3Ø) (COSØ 63.5 V to 300 V@ 1A
4 Source Educetech by 0.24%
Current (< 1 = 0.5 TO UPF) @ to 15A
Direct Method
GHz) 50Hz
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 15 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- High Voltage Probe
Alternating AC High Voltage @ Fluke 80K40 with
5 1 kV to 28 kV 7 % to 8.38 %
Current (< 1 50Hz DMM by comparison
GHz) Method
(Measure)
ELECTRO-
TECHNICAL-
AC Digital Power Meter 110 W to 6000
Alternating
6 POWER@50Hz(UPF Yokogawa by Direct W,110V to 400V@1A 0.38 % to 0.36 %
Current (< 1
1Ø) Method to 15A
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Voltage @1kHz
7 Transmille 8081 by 100 mV to 700 V 0.081 % to 0.05 %
Current (< 1 to 10kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Voltage @50Hz
8 Transmille 8081 by 100 V to 1000 V 0.035 % to 0.032 %
Current (< 1 to 1kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Precision Multimeter
Alternating AC Voltage @50Hz
9 Transmille 8081 by 50 mV to 100 V 0.054 % to 0.035 %
Current (< 1 to 1kHz
Direct Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 16 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Digital Scientific
Alternating
10 Capacitance@1kHz make LCR Meter by 1 nF to 1 µF 0.68%
Current (< 1
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Digital Scientific LCR
Alternating
11 Inductance@1kHz Meter by Direct 5 mH to 10 H 1.4%
Current (< 1
Method
GHz)
(Measure)
ELECTRO-
Multi Product
TECHNICAL-
AC Current @ 50Hz Calibrator with
12 Alternating 20 A to 1000 A 0.77 % to 0.58 %
to 60Hz current coils by
Current (< 1
Direct Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Current@ 50Hz Calibrator Transmille
13 Alternating 2 A to 20 A 0.28 % to 0.17 %
to 1kHz 3050 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Current@ 50Hz Calibrator Transmille
14 Alternating 20 µA to 20 mA 1.76 % to 0.48 %
to 1kHz 3050 by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 17 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Multi Product
TECHNICAL-
AC Current@ 50Hz Calibrator Transmille
15 Alternating 20 mA to 2 A 0.48 % to 0.28 %
to 1kHz 3050 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- AC POWER(1Ø and using Energy Cal 63.5 W to 6000 W
16 Alternating 3Ø) (COSØ = 0.5 TO Source by Direct ,63.5 V to 300V@1A 0.24%
Current (< 1 UPF) @ 50Hz Method to 20A
GHz) (Source)
ELECTRO-
TECHNICAL- Discrete Standard
AC 100 ohm, 1k & 100k
17 Alternating Resistors by Direct 0.164%
Resistance@1kHz Ohm
Current (< 1 Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Voltage@ 50Hz Calibrator Transmille
18 Alternating 200 V to 1000 V 0.09%
to 1kHz 3050 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Multi Product
TECHNICAL-
AC Voltage@ 50Hz Calibrator Transmille
19 Alternating 50 mV to 200 V 0.22 % to 0.09 %
to 1kHz 3050 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance @ 1
20 Alternating Capacitance box by 1 nF to 10 µF 0.4 % to 1 %
kHz
Current (< 1 Direct Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 18 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Inductance Box by
21 Alternating Inductance@1kHz 1 mH to 10 H 5.43%
Direct Method
Current (< 1
GHz) (Source)
ELECTRO-
TECHNICAL- Energy Cal Source
22 Alternating Power Factor Educetech by Direct 0.5 PF to 1 PF 0.002PF
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL-
Precision Multimeter
23 DIRECT DC Current 1 µA to 10 µA 0.026 % to 0.046 %
by Direct Method
CURRENT
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
24 DIRECT DC Current Transmille 8081 by 1 A to 20 A 0.02 % to 0.09 %
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
25 DIRECT DC Current Transmille 8081 by 10 µA to 100 µA 0.046 % to 0.003 %
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
26 DIRECT DC Current Transmille 8081 by 100 µA to 100 mA 0.002 % to 0.005 %
CURRENT Direct Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 19 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Precision Multimeter
27 DIRECT DC Current Transmille 8081 by 100 mA to 1 A 0.005 % to 0.02 %
CURRENT Direct Method
(Measure)
ELECTRO-
Precision Shunt with
TECHNICAL-
Yokogawa Power
28 DIRECT DC Current 20 A to 100 A 0.63%
meter by Direct
CURRENT
Method
(Measure)
ELECTRO-
High Voltage Probe
TECHNICAL-
Fluke 80k40 with
29 DIRECT DC High Voltage 1 kV to 40 kV 3.8%
DMM by Direct
CURRENT
Method
(Measure)
ELECTRO-
Precision Multi
TECHNICAL-
meter Transmille
30 DIRECT DC Voltage 1 mV to 100 mV 0.041 % to 0.001 %
8081 by Direct
CURRENT
Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
31 DIRECT DC Voltage Transmille 8081 by 100 mV to 1000 V 0.001%
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Resistance (2Wire
32 DIRECT Transmille 8081 by 1 Ohm to 100 Ohm 0.003 % to 0.0015 %
/4Wire)
CURRENT Direct Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 20 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Precision Multimeter
Resistance(2Wire 1 MOhm to 10
33 DIRECT Transmille 8081 by 0.0021 % to 0.004 %
/4Wire) MOhm
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Resistance(2Wire
34 DIRECT Transmille 8081 by 10 mOhm to 1 Ohm 0.09 % to 0.003 %
/4Wire)
CURRENT Direct Method
(Measure)
ELECTRO-
Precision Multimeter
TECHNICAL-
Resistance(2Wire Transmille 8081 in 10 MOhm to 100
35 DIRECT 0.05 % to 0.55 %
/4Wire) electrometer mode GOhm
CURRENT
by Direct Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Resistance(2Wire 0.0015 % to 0.0022
36 DIRECT Transmille 8081 by 100 Ohm to 1 MOhm
/4Wire) %
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Multi Product
37 DIRECT DC Current Calibrator by Direct 1 µA to 100 µA 2.3 % to 0.06 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
38 DIRECT DC Current Calibrator by Direct 100 µA to 100 mA 0.06 % to 0.02 %
CURRENT Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 21 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Multi Product
TECHNICAL-
Calibrator Transmille
39 DIRECT DC Current 100 mA to 20 A 0.02 % to 0.1 %
3050 by Direct
CURRENT
Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator & Current
40 DIRECT DC Current 20 A to 1000 A 0.8 % to 0.35 %
Coil by Direct
CURRENT
Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
DC Resistance 4W & Calibrator Transmille 10 Mohm to 100
41 DIRECT 0.07 % to 0.6 %
2W 3050 by Direct Mohm
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
DC Resistance 4W & 10 ohm to 100
42 DIRECT Calibrator by Direct 0.12 % to 0.06
2W kOhm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
DC Resistance 4W & 100 kOhm to 10
43 DIRECT Calibrator by Direct 0.06 % to 0.07 %
2W Mohm
CURRENT Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
DC Resistance 4W & Calibrator Transmille 100 MOhm to 1
44 DIRECT 0.6 % to 1.23 %
2W 2041 by Direct GOhm
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 22 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Multi Product
45 DIRECT DC Voltage Calibrator by Direct 1 mV to 10 mV 0.46 % to 0.052 %
CURRENT Method
(Source)
ELECTRO-
Multi Product
TECHNICAL-
Calibrator Transmille
46 DIRECT DC Voltage 10 mV to 100 mV 0.052 % to 0.014 %
3050 by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
47 DIRECT DC Voltage Calibrator by Direct 100 mV to 1000 V 0.014 % to 0.008 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
Power/Digital Power 10 W to 12 kW 10V
48 DIRECT Calibrator by Direct 0.03 % to 0.1 %
Meter to 600V@ 1A to 20A
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Decade Resistance
Resistance 1 Gohm to 100
49 DIRECT Box Vaiseshika by 1.22 % to 6 %
2Wire/4wire Gohm
CURRENT Direct Method
(Source)
ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
50 DIRECT 1 mOhm 0.46%
2Wire/4wire Direct Method
CURRENT
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 23 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
51 DIRECT 1 Ohm 0.13%
2Wire/4wire Direct Method
CURRENT
(Source)
ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
52 DIRECT 10 mOhm 0.46%
2Wire/4wire Direct Method
CURRENT
(Source)
ELECTRO-
TECHNICAL-
Resistance Standard Resistor by
53 DIRECT 100 µohm 0.46%
2Wire/4wire Direct method
CURRENT
(Source)
ELECTRO-
TECHNICAL- Standard
Resistance
54 DIRECT Resistor/Direct 100 mOhm 0.46%
2Wire/4wire
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
Bandwidth/Oscillosc
55 ELECTRICAL Calibrator by Direct 10 MHz to 600 MHz 1.3dB
ope
EQUIPMENT Method
(Source)
ELECTRO-
DC
TECHNICAL- Multi Product
Amplitude/Oscillosco
56 ELECTRICAL Calibrator by Direct 12 mV to 300 V DC 0.37 % to 0.045 %
pe (2mV/Div to
EQUIPMENT Method
50V/Div)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 24 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Multi Product
Time
57 ELECTRICAL Calibrator by Direct 2 ns to 5 sec 0.088 % to 0.081 %
Base/Oscilloscope
EQUIPMENT Method
(Source)
ELECTRO-
TECHNICAL- Temperature( Precision Multimeter
58 TEMPERATURE Type"J" & Direct/ -200 ° C to 1200 °C 0.16°C
SIMULATION Thermocouple) Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
59 TEMPERATURE by Direct 0 °C to 800 °C 0.16°C
"E" Thermocouple)
SIMULATION /Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
60 TEMPERATURE by Direct -140 °C to 1340 °C 0.16°C
"K" Thermocouple )
SIMULATION /Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
61 TEMPERATURE by Direct / -180 °C to 1300 °C 0.18°C
"N" Thermocouple )
SIMULATION Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature( Type
62 TEMPERATURE by Direct -100 °C to 800 °C 0.10°C
"RTD" )
SIMULATION /Comparison Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 25 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Precision Multimeter
TECHNICAL-
Temperature( Type by
63 TEMPERATURE 0 °C to 1700 °C 0.27°C
"S" Thermocouple) Direct/Comparison
SIMULATION
Method
(Measure)
ELECTRO-
TECHNICAL- Temperature(Therm Precision Multimeter
64 TEMPERATURE ocouple Type "B" by Direct 600 °C to 1800 °C 0.68°C
SIMULATION Thermocouple) /Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature(Type
65 TEMPERATURE by Direct / 0 °C to 1700 °C 0.27°C
"R" Thermocouple )
SIMULATION Comparison Method
(Measure)
ELECTRO-
TECHNICAL- Precision Multimeter
Temperature(Type
66 TEMPERATURE by Direct 0 °C to 400 °C 0.16°C
"T" Thermocouple )
SIMULATION /Comparison Method
(Measure)
ELECTRO-
Precision Multimeter
TECHNICAL-
/ Multi function
67 TIME & Frequency 100Hz to 600MHz 0.5% to 0.024%
counter by Direct
FREQUENCY
Method
(Measure)
ELECTRO-
TECHNICAL- Universal Counter
68 TIME & Time timer by comparison 100 ms to 1000 s 0.025 s to 3.4 s
FREQUENCY Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
CALSYTECH, NO. 38,NORTH MADA STREET,NANDAMBAKKAM, CHENNAI, TAMIL
Laboratory Name :
NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2193 Page No 26 of 26
Validity 16/03/2021 to 15/03/2023 Last Amended on -

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Multi Product
69 TIME & Frequency Calibrator by Direct 10 MHz to 600 MHz 0.004%
FREQUENCY Method
(Source)
ELECTRO-
TECHNICAL- Multi Product
70 TIME & Frequency Calibrator by Direct 45 Hz to 10 MHz 0.0055 % to 0.004 %
FREQUENCY Method
(Source)
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.

This is annexure to 'Certificate of Accreditation' and does not require any signature.