Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.

57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 1 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Permanent Facility

1 ELECTRO- AC Current (45Hz) 1 mA to 100 mA 0.093% to 0.092% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

2 ELECTRO- AC Current (45Hz) 100 mA to 1 A 0.092% to 0.12% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

3 ELECTRO- AC Current (45Hz) 100 uA to 1 mA 0.11% to 0.093% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

4 ELECTRO- AC Voltage (1MHz) 10 mV to 100 0.59% to 0.12% Agilent 3458A


TECHNICAL- mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

5 ELECTRO- AC Voltage (20 kHz) 1 V to 10 V 0.037% to 0.095% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 2 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
6 ELECTRO- AC Current (10 kHz) 200 uA to 2 0.2% Fluke 5720A / Direct
TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

7 ELECTRO- AC Current (10Hz) 2 mA to 20 mA 0.039% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

8 ELECTRO- AC Current (10Hz) 20 mA to 200 0.039% to 0.041% Fluke 5720A / Direct


TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

9 ELECTRO- AC Current (10Hz) 200 µA 0.047% to 0.039% Fluke 5720A / Direct


TECHNICAL- to 2 mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

10 ELECTRO- AC Current (50Hz) 10 A to 18 A 0.59% to 1.78% Fluke 5500A & 5220A


TECHNICAL- Calibrator with 50 turn
ALTERNATING Current Coil
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 3 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
11 ELECTRO- AC Current (50Hz) 18 A to 900 A 2.2% to 0.86% Fluke 5500A & 5220A
TECHNICAL- Calibrator with 50 turn
ALTERNATING current coil
CURRENT (< 1 GHZ)
(Source)

12 ELECTRO- AC Voltage (20Hz to 40 Hz) 2 mV 0.35% to 0.068% Fluke 5720A Direct


TECHNICAL- to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

13 ELECTRO- AC Voltage (20Hz to 40Hz) 2 V to 0.039% Fluke 5720A / Direct


TECHNICAL- 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

14 ELECTRO- AC Voltage (20Hz to 40Hz) 20 mV 0.068% to 0.046% Fluke 5720A / Direct


TECHNICAL- to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

15 ELECTRO- AC Voltage (20Hz to 40Hz) 20 V to 0.039% to 0.036% Fluke 5720A / Direct


TECHNICAL- 200 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 4 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
16 ELECTRO- AC Voltage (20Hz to 40Hz) 200 mV 0.046% to 0.039% Fluke 5720A / Direct
TECHNICAL- to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

17 ELECTRO- Capacitance (100Hz) 1 uF to 30 uF 5.8mF to 4.5mF Fluke 5500A Calibrator


TECHNICAL- / Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

18 ELECTRO- Capacitance (1kHz) 0.35 nF to 300 28mF to 3.3mF Fluke 5500A Calibrator
TECHNICAL- nF / Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

19 ELECTRO- AC Current (1 kHz) 1 mA to 100 0.059% to 0.058% Agilent 3458A


TECHNICAL- mA Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

20 ELECTRO- AC Current (1 kHz) 10 µA to 2 mA 0.063% to 0.020% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 5 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
21 ELECTRO- AC Current (1 kHz) 10 uA to 100 0.46% to 0.11% Agilent 3458A
TECHNICAL- uA Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

22 ELECTRO- AC Current (1 kHz) 100 mA to 1 A 0.058% to 0.14% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

23 ELECTRO- AC Current (1 kHz) 100 uA to 1 mA 0.11% to 0.059% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

24 ELECTRO- AC Current (1 kHz) 2 A to 10 A 0.88% to 0.41% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

25 ELECTRO- AC Current (1 kHz) 2 mA to 20 mA 0.020% to 0.021% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 6 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
26 ELECTRO- AC Current (1 kHz) 200 mA to 2 A 0.024% to 0.071% Fluke 5720A / Direct
TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

27 ELECTRO- AC Current (10 kHz) 2 mA to 20 0.2% to 0.19% Fluke 5720A / Direct


TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

28 ELECTRO- AC Current (10 kHz) 20 mA to 200 0.19% to 0.16% Fluke 5720A / Direct
TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

29 ELECTRO- AC Current (10 kHz) 200 mA to 2 A 0.16% to 0.94% Fluke 5720A / Direct
TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

30 ELECTRO- AC Current (1kHz) 20 mA to 200 0.021% to 0.024% Fluke 5720A / Direct


TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 7 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
31 ELECTRO- AC High Voltage (50Hz) 1 kV to 5 kV 3.2% to 3.8% Fluke 80k-40 HV probe
TECHNICAL- with Fluke 115
ALTERNATING Multimeter
CURRENT (< 1 GHZ)
(Source,Measure)

32 ELECTRO- AC Voltage (100 kHz to 1 MHz) 2 V 0.33% to 0.35% Fluke 5720A / Direct
TECHNICAL- to 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

33 ELECTRO- AC Voltage (100 kHz to 1 MHz) 20 0.72% to 0.47% Fluke 5720A / Direct
TECHNICAL- mV to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

34 ELECTRO- AC Voltage (100 kHz to 1 MHz) 0.47% to 0.33% Fluke 5720A / Direct
TECHNICAL- 200 mV to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

35 ELECTRO- AC Voltage (100 kHz to 1MHz) 2 2.4% to 0.72% Fluke 5720A / Direct
TECHNICAL- mV to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 8 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
36 ELECTRO- AC Voltage (100 kHz) 1 V to 10 V 0.099% to 0.36% Agilent 3458A
TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

37 ELECTRO- AC Voltage (100 kHz) 10 mV to 0.59% to 0.037% Agilent 3458A


TECHNICAL- 100 mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

38 ELECTRO- AC Voltage (100 kHz) 10 V to 100 0.36% to 0.043% Agilent 3458A


TECHNICAL- V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

39 ELECTRO- AC Voltage (100 kHz) 100 mV to 1 0.037% to 0.099% Agilent 3458A


TECHNICAL- V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

40 ELECTRO- AC Voltage (1MHz) 1 V to 10 V 0.12% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 9 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
41 ELECTRO- AC Voltage (1MHz) 100 mV to 1 V 0.12% Agilent 3458A
TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

42 ELECTRO- AC Voltage (20 kHz to 100 kHz) 2 0.47% to 0.11% Fluke 5720A / Direct
TECHNICAL- mV to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

43 ELECTRO- AC Voltage (20 kHz to 100 kHz) 2 0.021% Fluke 5720A / Direct
TECHNICAL- V to 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

44 ELECTRO- AC Voltage (20 kHz to 100 kHz) 20 0.11% to 0.083% Fluke 5720A / Direct
TECHNICAL- mV to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

45 ELECTRO- AC Voltage (20 kHz to 100 kHz) 20 0.021% to 0.023% Fluke 5720A / Direct
TECHNICAL- V to 200 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 10 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
46 ELECTRO- AC Voltage (20 kHz to 100 kHz) 0.083% to 0.021% Fluke 5720A / Direct
TECHNICAL- 200 mV to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

47 ELECTRO- AC Voltage (20 kHz) 10 mV to 100 0.056% to 0.019% Agilent 3458A


TECHNICAL- mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

48 ELECTRO- AC Voltage (20 kHz) 10 V to 100 V 0.095% to 0.026% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

49 ELECTRO- AC Voltage (20 kHz) 100 mV to 1 V 0.019% to 0.037% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

50 ELECTRO- AC Voltage (20 kHz) 100 V to 700 0.026% to 0.10% Agilent 3458A
TECHNICAL- V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 11 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
51 ELECTRO- AC Voltage (40 Hz to 1 kHz) 1 V to 0.019% to 0.037% Agilent 3458A
TECHNICAL- 10 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

52 ELECTRO- AC Voltage (40 Hz to 1 kHz) 10 mV 0.043% to 0.042% Agilent 3458A


TECHNICAL- to 100 mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

53 ELECTRO- AC Voltage (40 Hz to 1 kHz) 10 V 0.037% to 0.026% Agilent 3458A


TECHNICAL- to 100 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

54 ELECTRO- AC Voltage (40 Hz to 1 kHz) 100 0.042% to 0.019% Agilent 3458A


TECHNICAL- mV to 1 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

55 ELECTRO- AC Voltage (40 Hz to 1 kHz) 100 V 0.026% to 0.10% Agilent 3458A


TECHNICAL- to 700 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 12 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
56 ELECTRO- AC Voltage (40 Hz to 20 kHz) 2 mV 0.21% to 0.048% Fluke 5720A / Direct
TECHNICAL- to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

57 ELECTRO- AC Voltage (40 Hz to 20 kHz) 2 V 0.01% Fluke 5720A / Direct


TECHNICAL- to 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

58 ELECTRO- AC Voltage (40 Hz to 20 kHz) 20 0.048% to 0.021% Fluke 5720A / Direct


TECHNICAL- mV to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

59 ELECTRO- AC Voltage (40 Hz to 20 kHz) 20 V 0.01% to 0.012% Fluke 5720A / Direct


TECHNICAL- to 200 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

60 ELECTRO- AC Voltage (40 Hz to 20 kHz) 200 0.021% to 0.01% Fluke 5720A / Direct
TECHNICAL- mV to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 13 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
61 ELECTRO- AC Voltage (50 Hz to 1 kHz) 1000 0.13% to 0.11% Fluke 5720A / Direct
TECHNICAL- V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

62 ELECTRO- DC Current 1 µA to 100 µA 0.58 % to 0.58% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

63 ELECTRO- DC Current 10 mA to 1 A 0.0029% to 0.014% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

64 ELECTRO- DC Current 100 µA to 10 mA 0.058% to 0.0029% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

65 ELECTRO- DC High Voltage 1 kV to 20 kV 1.90% to 6.72% Fluke 80k-40 HV probe


TECHNICAL- DIRECT with Fluke 115
CURRENT (Measure) Multimeter

66 ELECTRO- DC Voltage 1 V to 10 V 0.00096% to 0.00093% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

67 ELECTRO- DC Voltage 10 V to 1000 V 0.00093% to 0.0012% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

68 ELECTRO- DC Voltage 100 mV to 1 V 0.0014% to 0.00096% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 14 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
69 ELECTRO- Resistance 1 ohm to 100 k ohm 0.058% to 0.0014% Agilent 3458A
TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

70 ELECTRO- Resistance 10 M ohm to 100 M 0.0073% to 0.0059% Agilent 3458A


TECHNICAL- DIRECT ohm Multimeter
CURRENT (Measure)

71 ELECTRO- Resistance 100 K ohm to 10 M 0.0014% to 0.0073% Agilent 3458A


TECHNICAL- DIRECT ohm Multimeter
CURRENT (Measure)

72 ELECTRO- DC Current 10 µA 0.011% to 0.006% Fluke 5720A / Direct


TECHNICAL- DIRECT to 2 mA Method
CURRENT (Source)

73 ELECTRO- DC Current 10 A to 18 A 0.1% to 0.33% Fluke 5500A & Fluke


TECHNICAL- DIRECT 5220A / Direct Method
CURRENT (Source)

74 ELECTRO- DC Current 18 A to 900 A 2.77% to 0.86% Fluke 5500 & Fluke


TECHNICAL- DIRECT 5220A with 50 turn
CURRENT (Source) current coil

75 ELECTRO- DC Current 2 A to 10 A 0.14% to 0.1% Fluke 5500A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

76 ELECTRO- DC Current 2 mA to 20 mA 0.006% to 0.005% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

77 ELECTRO- DC Current 20 mA to 200 mA 0.005% to 0.007% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 15 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
78 ELECTRO- DC Current 200 mA to 2 A 0.007% to 0.011% Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

79 ELECTRO- DC Voltage 0.5 mV to 1 V 0.13% to 0.0012% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

80 ELECTRO- DC Voltage 1 V to 10 V 0.0012% to 0.00072% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

81 ELECTRO- DC Voltage 10 V to 100 V 0.00072% to 0.00089% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

82 ELECTRO- DC Voltage 100 V to 1000 V 0.00089% to 0.0011% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

83 ELECTRO- Resistance 1 k ohm 19m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

84 ELECTRO- Resistance 1 M ohm 31ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

85 ELECTRO- Resistance 1 ohm 240 µ ohm Fluke 5720A Calibrator


TECHNICAL- DIRECT / Direct Method
CURRENT (Source)

86 ELECTRO- Resistance 1.9 k ohm 22m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 16 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
87 ELECTRO- Resistance 1.9 M ohm 45 ohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

88 ELECTRO- Resistance 1.9 ohm 190 µ ohm Fluke 5720A Calibrator


TECHNICAL- DIRECT / Direct Method
CURRENT (Source)

89 ELECTRO- Resistance 10 M ohm 240 ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

90 ELECTRO- Resistance 10 ohm 87 µohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

91 ELECTRO- Resistance 100 k ohm 734 m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

92 ELECTRO- Resistance 100 ohm 658 µ ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

93 ELECTRO- Resistance 19 k ohm 181 m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

94 ELECTRO- Resistance 19 ohm 210 µohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

95 ELECTRO- Resistance 190 k ohm 1397m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 17 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
96 ELECTRO- Resistance 190 ohm 1255 µohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

97 ELECTRO- Resistance 2 ohm to 330 ohm 0.47% to 0.017% Fluke 5500A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

98 ELECTRO- Resistance 330 k ohm to 117 M 0.018% to 0.63% Fluke 5500A / Direct
TECHNICAL- DIRECT ohm Method
CURRENT (Source)

99 ELECTRO- Resistance 330 ohm to 330 k ohm 0.017% to 0.018% Fluke 5500A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

100 ELECTRO- DC Current 1 A to 30 A 0.28% Using shunt with


TECHNICAL- DIRECT Multimeter
CURRENT
(Source,Measure)

101 ELECTRO- DC Current 10 nA to 1 µA 1.1% to 0.58% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT
(Source,Measure)

102 ELECTRO- Resistance 10 k ohm 85 m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT
(Source,Measure)

103 ELECTRO- Resistance 100 M ohm 13270 ohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 18 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
104 ELECTRO- Resistance 19 M ohm 387ohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT
(Source,Measure)

105 ELECTRO- Temperature -190 °C to 790 °C 0.059°C to 0.14°C Fluke 5500A Calibrator
TECHNICAL- Simulation, RTD / Direct Method
TEMPERATURE PT385, 100 Ohm
SIMULATION (Source)

106 ELECTRO- Temperature -190 °C to 620 °C 0.037°C to 0.11°C Fluke 5500A Calibrator
TECHNICAL- Simulation, RTD / Direct Method
TEMPERATURE PT385, 1k Ohm
SIMULATION (Source)

107 ELECTRO- Temperature -200 °C to 800 °C 0.59°C to 0.26°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type E
SIMULATION (Source)

108 ELECTRO- Temperature 600 °C to 1800 °C 0.5°C Fluke 5500A Calibrator


TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type B
SIMULATION (Source)

109 ELECTRO- Temperature -200 °C to 1100 °C 0.36°C to 0.31°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type J
SIMULATION (Source)

110 ELECTRO- Temperature -190 °C to 1300 °C 0.42°C to 0.50°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type K
SIMULATION (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 19 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
111 ELECTRO- Temperature -200 °C to 1300 °C 0.07°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type N
SIMULATION (Source)

112 ELECTRO- Temperature 100 °C to 1700 °C 0.62°C to 0.76°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type R
SIMULATION (Source)

113 ELECTRO- Temperature 100 °C to 1700 °C 0.62°C to 0.76°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type S
SIMULATION (Source)

114 ELECTRO- Temperature -200 °C to 300 °C 0.79°C to 0.18°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type T
SIMULATION (Source)

115 ELECTRO- Temperature -190 °C to 800 °C 0.07°C Agilent 3458A


TECHNICAL- Simulation, RTD Multimeter / Resistance
TEMPERATURE PT385, 100 ohm Method
SIMULATION
(Source,Measure)

116 ELECTRO- Temperature 800 °C to 1800 °C 0.5°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type B
SIMULATION
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 20 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
117 ELECTRO- Temperature -200 °C to 800 °C 0.08°C 7.5 Digit Multimeter
TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type E
SIMULATION
(Source,Measure)

118 ELECTRO- Temperature -200 °C to 1100 °C 0.08°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type J
SIMULATION
(Source,Measure)

119 ELECTRO- Temperature -190 °C to 1300 °C 0.05°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type K
SIMULATION
(Source,Measure)

120 ELECTRO- Temperature -200 °C to 1300 °C 0.09°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type N
SIMULATION
(Source,Measure)

121 ELECTRO- Temperature 100 °C to 1700 °C 0.07°C Agilent 3458A


TECHNICAL- Simulation, Multimeter
TEMPERATURE Thermocouple Type R
SIMULATION
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 21 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
122 ELECTRO- Temperature 150 °C to 1700 °C 0.14°C 7.5 digit Multimeter
TECHNICAL- Simulation, PXI- 4071
TEMPERATURE Thermocouple Type S
SIMULATION
(Source,Measure)

123 ELECTRO- Temperature -200 °C to 300 °C 0.1°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type T
SIMULATION
(Source,Measure)

124 ELECTRO- Frequency 1 Hz to 100 Hz 662ppm to 120ppm Agilent 3458A


TECHNICAL- TIME & Multimeter & HP
FREQUENCY 53131A Frequency
(Measure) Counter

125 ELECTRO- Frequency 10 MHz 0.76 ppm HP 58503A GPS Time


TECHNICAL- TIME & & Frequency Standard
FREQUENCY (Source)

126 ELECTRO- Frequency 10Hz to 1MHz 131ppm to 22ppm Fluke 5720A Calibrator
TECHNICAL- TIME & / Direct Method
FREQUENCY (Source)

127 MECHANICAL- Caliper 0 mm to 300 mm 9.30 µm Caliper checker and


DIMENSION (BASIC Gauge blocks as per IS
MEASURING 3651
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 22 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
128 MECHANICAL- Caliper 300 mm to 600 mm 9.40µm Caliper checker and
DIMENSION (BASIC Gauge blocks as per IS
MEASURING 3651
INSTRUMENT,
GAUGE ETC.)

129 MECHANICAL- Cylindrical Pins 0.5 mm to 20 mm 1.66 µm Using Laser Scan


DIMENSION (BASIC Micrometer as per IS
MEASURING 11103
INSTRUMENT,
GAUGE ETC.)

130 MECHANICAL- External Micrometer 0.5 mm to 25 mm 1.40 µm Gauge blocks as per


DIMENSION (BASIC 2967
MEASURING
INSTRUMENT,
GAUGE ETC.)

131 MECHANICAL- External Micrometer 25 mm to 150 mm 4.50µm Gauge Blocks as per


DIMENSION (BASIC 2967
MEASURING
INSTRUMENT,
GAUGE ETC.)

132 MECHANICAL- Height Gauge 0.5 mm to 600 mm 8.8 µm Caliper checker &
DIMENSION (BASIC Gauge blocks as per IS
MEASURING 2921
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 23 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
133 MECHANICAL- Plain Plug Gauge 0.5 mm to 40 mm 1.00 µm Gauge blocks and laser
DIMENSION (BASIC scan micrometer as per
MEASURING IS 3455, IS 6174
INSTRUMENT,
GAUGE ETC.)

134 MECHANICAL- Steel Ruler's / 0 m to 10 m 580 vL (where L is in Measuring Scale &


DIMENSION (BASIC Measuring Tape's mtr.)µm Tape calibration
MEASURING machine
INSTRUMENT,
GAUGE ETC.)

135 MECHANICAL- Pressure gauge/ 0 bar to 20 bar 0.066bar Druck DPI 603 Digital
PRESSURE calibrator/ transducer/ Pressure Guage as per
INDICATING DEVICES switch/ compound DKD R-6-1
gauge

136 MECHANICAL- Pressure gauge/ 0 bar to 700 bar 2.1bar Using Digital Pressure
PRESSURE calibrator/ transducer/ Gauge DPG700 as per
INDICATING DEVICES switch/ compound DKD R-6-1
gauge

137 MECHANICAL- Vacuum -0.95 bar to 0 bar 0.008bar Digital Pressure Gauge
PRESSURE gauge/transmitter/ DPI 603 as per IS 3567
INDICATING DEVICES switch /transmitter

138 THERMAL- SPECIFIC Analog / Digital / (25ºC) 15 % RH to 90 4.9% Standards temperature


HEAT & HUMIDITY Thermo hygrometer's / % RH humidity sensor with
Thermo Hygro graph's / Indicator & Humidity
Humidity Sensor's Chamber

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 24 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
139 THERMAL- SPECIFIC Analog / Digital / (50%) 10 ºC to 50 ºC 0.87 ºC Standards temperature
HEAT & HUMIDITY Thermo hygrometer's / humidity sensor with
Thermo Hygro graph's / Indicator & Humidity
Humidity Sensor's Chamber

140 THERMAL- RTD ,Thermocouple, 50 ºC to 400 ºC 1.9ºC Using Standard RTD,


TEMPERATURE Thermistor Sensors Dry block calibrator &
with / without Indicator, Digital Multimeter
Temperature gauge/
Temperature bath,
controllers and
Transmitter with sensor

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 25 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
Site Facility

1 ELECTRO- AC Current (45Hz) 1 mA to 100 mA 0.093% to 0.092% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

2 ELECTRO- AC Current (45Hz) 100 mA to 1 A 0.092% to 0.12% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

3 ELECTRO- AC Current (45Hz) 100 uA to 1 mA 0.11% to 0.093% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

4 ELECTRO- AC Voltage (1MHz) 10 mV to 100 0.59% to 0.12% Agilent 3458A


TECHNICAL- mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

5 ELECTRO- AC Voltage (20 kHz) 1 V to 10 V 0.037% to 0.095% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 26 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
6 ELECTRO- AC Current (10 kHz) 200 uA to 2 0.2% Fluke 5720A / Direct
TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

7 ELECTRO- AC Current (10Hz) 2 mA to 20 mA 0.039% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

8 ELECTRO- AC Current (10Hz) 20 mA to 200 0.039% to 0.041% Fluke 5720A / Direct


TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

9 ELECTRO- AC Current (10Hz) 200 µA 0.047% to 0.039% Fluke 5720A / Direct


TECHNICAL- to 2 mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

10 ELECTRO- AC Current (50Hz) 10 A to 18 A 0.59% to 1.78% Fluke 5500A & 5220A


TECHNICAL- Calibrator with 50 turn
ALTERNATING Current Coil
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 27 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
11 ELECTRO- AC Current (50Hz) 18 A to 900 A 2.2% to 0.86% Fluke 5500A & 5220A
TECHNICAL- Calibrator with 50 turn
ALTERNATING current coil
CURRENT (< 1 GHZ)
(Source)

12 ELECTRO- AC Voltage (20Hz to 40 Hz) 2 mV 0.35% to 0.068% Fluke 5720A Direct


TECHNICAL- to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

13 ELECTRO- AC Voltage (20Hz to 40Hz) 2 V to 0.039% Fluke 5720A / Direct


TECHNICAL- 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

14 ELECTRO- AC Voltage (20Hz to 40Hz) 20 mV 0.068% to 0.046% Fluke 5720A / Direct


TECHNICAL- to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

15 ELECTRO- AC Voltage (20Hz to 40Hz) 20 V to 0.039% to 0.036% Fluke 5720A / Direct


TECHNICAL- 200 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 28 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
16 ELECTRO- AC Voltage (20Hz to 40Hz) 200 mV 0.046% to 0.039% Fluke 5720A / Direct
TECHNICAL- to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

17 ELECTRO- Capacitance (100Hz) 1 uF to 30 uF 5.8mF to 4.5mF Fluke 5500A Calibrator


TECHNICAL- / Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

18 ELECTRO- Capacitance (1kHz) 0.35 nF to 300 28mF to 3.3mF Fluke 5500A Calibrator
TECHNICAL- nF / Direct Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source)

19 ELECTRO- AC Current (1 kHz) 1 mA to 100 0.059% to 0.058% Agilent 3458A


TECHNICAL- mA Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

20 ELECTRO- AC Current (1 kHz) 10 µA to 2 mA 0.063% to 0.020% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 29 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
21 ELECTRO- AC Current (1 kHz) 10 uA to 100 0.46% to 0.11% Agilent 3458A
TECHNICAL- uA Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

22 ELECTRO- AC Current (1 kHz) 100 mA to 1 A 0.058% to 0.14% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

23 ELECTRO- AC Current (1 kHz) 100 uA to 1 mA 0.11% to 0.059% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

24 ELECTRO- AC Current (1 kHz) 2 A to 10 A 0.88% to 0.41% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

25 ELECTRO- AC Current (1 kHz) 2 mA to 20 mA 0.020% to 0.021% Fluke 5720A / Direct


TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 30 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
26 ELECTRO- AC Current (1 kHz) 200 mA to 2 A 0.024% to 0.071% Fluke 5720A / Direct
TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

27 ELECTRO- AC Current (10 kHz) 2 mA to 20 0.2% to 0.19% Fluke 5720A / Direct


TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

28 ELECTRO- AC Current (10 kHz) 20 mA to 200 0.19% to 0.16% Fluke 5720A / Direct
TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

29 ELECTRO- AC Current (10 kHz) 200 mA to 2 A 0.16% to 0.94% Fluke 5720A / Direct
TECHNICAL- Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

30 ELECTRO- AC Current (1kHz) 20 mA to 200 0.021% to 0.024% Fluke 5720A / Direct


TECHNICAL- mA Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 31 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
31 ELECTRO- AC High Voltage (50Hz) 1 kV to 5 kV 3.2% to 3.8% Fluke 80k-40 HV probe
TECHNICAL- with Fluke 115
ALTERNATING Multimeter
CURRENT (< 1 GHZ)
(Source,Measure)

32 ELECTRO- AC Voltage (100 kHz to 1 MHz) 2 V 0.33% to 0.35% Fluke 5720A / Direct
TECHNICAL- to 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

33 ELECTRO- AC Voltage (100 kHz to 1 MHz) 20 0.72% to 0.47% Fluke 5720A / Direct
TECHNICAL- mV to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

34 ELECTRO- AC Voltage (100 kHz to 1 MHz) 0.47% to 0.33% Fluke 5720A / Direct
TECHNICAL- 200 mV to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

35 ELECTRO- AC Voltage (100 kHz to 1MHz) 2 2.4% to 0.72% Fluke 5720A / Direct
TECHNICAL- mV to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 32 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
36 ELECTRO- AC Voltage (100 kHz) 1 V to 10 V 0.099% to 0.36% Agilent 3458A
TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

37 ELECTRO- AC Voltage (100 kHz) 10 mV to 0.59% to 0.037% Agilent 3458A


TECHNICAL- 100 mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

38 ELECTRO- AC Voltage (100 kHz) 10 V to 100 0.36% to 0.043% Agilent 3458A


TECHNICAL- V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

39 ELECTRO- AC Voltage (100 kHz) 100 mV to 1 0.037% to 0.099% Agilent 3458A


TECHNICAL- V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

40 ELECTRO- AC Voltage (1MHz) 1 V to 10 V 0.12% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 33 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
41 ELECTRO- AC Voltage (1MHz) 100 mV to 1 V 0.12% Agilent 3458A
TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

42 ELECTRO- AC Voltage (20 kHz to 100 kHz) 2 0.47% to 0.11% Fluke 5720A / Direct
TECHNICAL- mV to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

43 ELECTRO- AC Voltage (20 kHz to 100 kHz) 2 0.021% Fluke 5720A / Direct
TECHNICAL- V to 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

44 ELECTRO- AC Voltage (20 kHz to 100 kHz) 20 0.11% to 0.083% Fluke 5720A / Direct
TECHNICAL- mV to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

45 ELECTRO- AC Voltage (20 kHz to 100 kHz) 20 0.021% to 0.023% Fluke 5720A / Direct
TECHNICAL- V to 200 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 34 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
46 ELECTRO- AC Voltage (20 kHz to 100 kHz) 0.083% to 0.021% Fluke 5720A / Direct
TECHNICAL- 200 mV to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

47 ELECTRO- AC Voltage (20 kHz) 10 mV to 100 0.056% to 0.019% Agilent 3458A


TECHNICAL- mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

48 ELECTRO- AC Voltage (20 kHz) 10 V to 100 V 0.095% to 0.026% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

49 ELECTRO- AC Voltage (20 kHz) 100 mV to 1 V 0.019% to 0.037% Agilent 3458A


TECHNICAL- Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

50 ELECTRO- AC Voltage (20 kHz) 100 V to 700 0.026% to 0.10% Agilent 3458A
TECHNICAL- V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 35 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
51 ELECTRO- AC Voltage (40 Hz to 1 kHz) 1 V to 0.019% to 0.037% Agilent 3458A
TECHNICAL- 10 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

52 ELECTRO- AC Voltage (40 Hz to 1 kHz) 10 mV 0.043% to 0.042% Agilent 3458A


TECHNICAL- to 100 mV Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

53 ELECTRO- AC Voltage (40 Hz to 1 kHz) 10 V 0.037% to 0.026% Agilent 3458A


TECHNICAL- to 100 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

54 ELECTRO- AC Voltage (40 Hz to 1 kHz) 100 0.042% to 0.019% Agilent 3458A


TECHNICAL- mV to 1 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

55 ELECTRO- AC Voltage (40 Hz to 1 kHz) 100 V 0.026% to 0.10% Agilent 3458A


TECHNICAL- to 700 V Multimeter
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 36 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
56 ELECTRO- AC Voltage (40 Hz to 20 kHz) 2 mV 0.21% to 0.048% Fluke 5720A / Direct
TECHNICAL- to 20 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

57 ELECTRO- AC Voltage (40 Hz to 20 kHz) 2 V 0.01% Fluke 5720A / Direct


TECHNICAL- to 20 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

58 ELECTRO- AC Voltage (40 Hz to 20 kHz) 20 0.048% to 0.021% Fluke 5720A / Direct


TECHNICAL- mV to 200 mV Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

59 ELECTRO- AC Voltage (40 Hz to 20 kHz) 20 V 0.01% to 0.012% Fluke 5720A / Direct


TECHNICAL- to 200 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

60 ELECTRO- AC Voltage (40 Hz to 20 kHz) 200 0.021% to 0.01% Fluke 5720A / Direct
TECHNICAL- mV to 2 V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 37 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
61 ELECTRO- AC Voltage (50 Hz to 1 kHz) 1000 0.13% to 0.11% Fluke 5720A / Direct
TECHNICAL- V Method
ALTERNATING
CURRENT (< 1 GHZ)
(Source,Measure)

62 ELECTRO- DC Current 1 µA to 100 µA 0.58 % to 0.58% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

63 ELECTRO- DC Current 10 mA to 1 A 0.0029% to 0.014% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

64 ELECTRO- DC Current 100 µA to 10 mA 0.058% to 0.0029% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

65 ELECTRO- DC High Voltage 1 kV to 20 kV 1.90% to 6.72% Fluke 80k-40 HV probe


TECHNICAL- DIRECT with Fluke 115
CURRENT (Measure) Multimeter

66 ELECTRO- DC Voltage 1 V to 10 V 0.00096% to 0.00093% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

67 ELECTRO- DC Voltage 10 V to 1000 V 0.00093% to 0.0012% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

68 ELECTRO- DC Voltage 100 mV to 1 V 0.0014% to 0.00096% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 38 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
69 ELECTRO- Resistance 1 ohm to 100 k ohm 0.058% to 0.0014% Agilent 3458A
TECHNICAL- DIRECT Multimeter
CURRENT (Measure)

70 ELECTRO- Resistance 10 M ohm to 100 M 0.0073% to 0.0059% Agilent 3458A


TECHNICAL- DIRECT ohm Multimeter
CURRENT (Measure)

71 ELECTRO- Resistance 100 K ohm to 10 M 0.0014% to 0.0073% Agilent 3458A


TECHNICAL- DIRECT ohm Multimeter
CURRENT (Measure)

72 ELECTRO- DC Current 10 µA 0.011% to 0.006% Fluke 5720A / Direct


TECHNICAL- DIRECT to 2 mA Method
CURRENT (Source)

73 ELECTRO- DC Current 10 A to 18 A 0.1% to 0.33% Fluke 5500A & Fluke


TECHNICAL- DIRECT 5220A / Direct Method
CURRENT (Source)

74 ELECTRO- DC Current 18 A to 900 A 2.77% to 0.86% Fluke 5500 & Fluke


TECHNICAL- DIRECT 5220A with 50 turn
CURRENT (Source) current coil

75 ELECTRO- DC Current 2 A to 10 A 0.14% to 0.1% Fluke 5500A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

76 ELECTRO- DC Current 2 mA to 20 mA 0.006% to 0.005% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

77 ELECTRO- DC Current 20 mA to 200 mA 0.005% to 0.007% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 39 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
78 ELECTRO- DC Current 200 mA to 2 A 0.007% to 0.011% Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

79 ELECTRO- DC Voltage 0.5 mV to 1 V 0.13% to 0.0012% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

80 ELECTRO- DC Voltage 1 V to 10 V 0.0012% to 0.00072% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

81 ELECTRO- DC Voltage 10 V to 100 V 0.00072% to 0.00089% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

82 ELECTRO- DC Voltage 100 V to 1000 V 0.00089% to 0.0011% Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

83 ELECTRO- Resistance 1 k ohm 19m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

84 ELECTRO- Resistance 1 M ohm 31ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

85 ELECTRO- Resistance 1 ohm 240 µ ohm Fluke 5720A Calibrator


TECHNICAL- DIRECT / Direct Method
CURRENT (Source)

86 ELECTRO- Resistance 1.9 k ohm 22m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 40 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
87 ELECTRO- Resistance 1.9 M ohm 45 ohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

88 ELECTRO- Resistance 1.9 ohm 190 µ ohm Fluke 5720A Calibrator


TECHNICAL- DIRECT / Direct Method
CURRENT (Source)

89 ELECTRO- Resistance 10 M ohm 240 ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

90 ELECTRO- Resistance 10 ohm 87 µohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

91 ELECTRO- Resistance 100 k ohm 734 m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

92 ELECTRO- Resistance 100 ohm 658 µ ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

93 ELECTRO- Resistance 19 k ohm 181 m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

94 ELECTRO- Resistance 19 ohm 210 µohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

95 ELECTRO- Resistance 190 k ohm 1397m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 41 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
96 ELECTRO- Resistance 190 ohm 1255 µohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

97 ELECTRO- Resistance 2 ohm to 330 ohm 0.47% to 0.017% Fluke 5500A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

98 ELECTRO- Resistance 330 k ohm to 117 M 0.018% to 0.63% Fluke 5500A / Direct
TECHNICAL- DIRECT ohm Method
CURRENT (Source)

99 ELECTRO- Resistance 330 ohm to 330 k ohm 0.017% to 0.018% Fluke 5500A / Direct
TECHNICAL- DIRECT Method
CURRENT (Source)

100 ELECTRO- DC Current 1 A to 30 A 0.28% Using shunt with


TECHNICAL- DIRECT Multimeter
CURRENT
(Source,Measure)

101 ELECTRO- DC Current 10 nA to 1 µA 1.1% to 0.58% Agilent 3458A


TECHNICAL- DIRECT Multimeter
CURRENT
(Source,Measure)

102 ELECTRO- Resistance 10 k ohm 85 m ohm Fluke 5720A / Direct


TECHNICAL- DIRECT Method
CURRENT
(Source,Measure)

103 ELECTRO- Resistance 100 M ohm 13270 ohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 42 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
104 ELECTRO- Resistance 19 M ohm 387ohm Fluke 5720A / Direct
TECHNICAL- DIRECT Method
CURRENT
(Source,Measure)

105 ELECTRO- Temperature -190 °C to 790 °C 0.059°C to 0.14°C Fluke 5500A Calibrator
TECHNICAL- Simulation, RTD / Direct Method
TEMPERATURE PT385, 100 Ohm
SIMULATION (Source)

106 ELECTRO- Temperature -190 °C to 620 °C 0.037°C to 0.11°C Fluke 5500A Calibrator
TECHNICAL- Simulation, RTD / Direct Method
TEMPERATURE PT385, 1k Ohm
SIMULATION (Source)

107 ELECTRO- Temperature -200 °C to 800 °C 0.59°C to 0.26°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type E
SIMULATION (Source)

108 ELECTRO- Temperature 600 °C to 1800 °C 0.5°C Fluke 5500A Calibrator


TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type B
SIMULATION (Source)

109 ELECTRO- Temperature -200 °C to 1100 °C 0.36°C to 0.31°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type J
SIMULATION (Source)

110 ELECTRO- Temperature -190 °C to 1300 °C 0.42°C to 0.50°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type K
SIMULATION (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 43 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
111 ELECTRO- Temperature -200 °C to 1300 °C 0.07°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type N
SIMULATION (Source)

112 ELECTRO- Temperature 100 °C to 1700 °C 0.62°C to 0.76°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type R
SIMULATION (Source)

113 ELECTRO- Temperature 100 °C to 1700 °C 0.62°C to 0.76°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type S
SIMULATION (Source)

114 ELECTRO- Temperature -200 °C to 300 °C 0.79°C to 0.18°C Fluke 5500A Calibrator
TECHNICAL- Simulation, / Direct Method
TEMPERATURE Thermocouple Type T
SIMULATION (Source)

115 ELECTRO- Temperature -190 °C to 800 °C 0.07°C Agilent 3458A


TECHNICAL- Simulation, RTD Multimeter / Resistance
TEMPERATURE PT385, 100 ohm Method
SIMULATION
(Source,Measure)

116 ELECTRO- Temperature 800 °C to 1800 °C 0.5°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type B
SIMULATION
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 44 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
117 ELECTRO- Temperature -200 °C to 800 °C 0.08°C 7.5 Digit Multimeter
TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type E
SIMULATION
(Source,Measure)

118 ELECTRO- Temperature -200 °C to 1100 °C 0.08°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type J
SIMULATION
(Source,Measure)

119 ELECTRO- Temperature -190 °C to 1300 °C 0.05°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type K
SIMULATION
(Source,Measure)

120 ELECTRO- Temperature -200 °C to 1300 °C 0.09°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type N
SIMULATION
(Source,Measure)

121 ELECTRO- Temperature 100 °C to 1700 °C 0.07°C Agilent 3458A


TECHNICAL- Simulation, Multimeter
TEMPERATURE Thermocouple Type R
SIMULATION
(Source,Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 45 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
122 ELECTRO- Temperature 150 °C to 1700 °C 0.14°C 7.5 digit Multimeter
TECHNICAL- Simulation, PXI- 4071
TEMPERATURE Thermocouple Type S
SIMULATION
(Source,Measure)

123 ELECTRO- Temperature -200 °C to 300 °C 0.1°C 7.5 digit Multimeter


TECHNICAL- Simulation, PXI-4071
TEMPERATURE Thermocouple Type T
SIMULATION
(Source,Measure)

124 ELECTRO- Frequency 1 Hz to 100 Hz 662ppm to 120ppm Agilent 3458A


TECHNICAL- TIME & Multimeter & HP
FREQUENCY 53131A Frequency
(Measure) Counter

125 ELECTRO- Frequency 10Hz to 1MHz 131ppm to 22ppm Fluke 5720A Calibrator
TECHNICAL- TIME & / Direct Method
FREQUENCY (Source)

126 MECHANICAL- Pressure gauge/ 0 bar to 20 bar 0.066bar Druck DPI 603 Digital
PRESSURE calibrator/ transducer/ Pressure Guage as per
INDICATING DEVICES switch/ compound DKD R-6-1
gauge

127 MECHANICAL- Pressure gauge/ 0 bar to 700 bar 2.1bar Using Digital Pressure
PRESSURE calibrator/ transducer/ Gauge DPG700 as per
INDICATING DEVICES switch/ compound DKD R-6-1
gauge

128 MECHANICAL- Vacuum -0.95 bar to 0 bar 0.008bar Digital Pressure Gauge
PRESSURE gauge/transmitter/ DPI 603 as per IS 3567
INDICATING DEVICES switch /transmitter

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name MICRO PRECISION CALIBRATION INDIA PRIVATE LIMITED, NO.57/65, SIPPOY
LANE, AVADI ROAD, KARAYANCHAVADI, CHENNAI, TAMIL NADU , INDIA

Accreditation Standard ISO/IEC 17025:2005

Certificate Number CC-2490 Page No. : 46 / 46

Validity 19/11/2018 to 18/11/2020 Last Amended on -

S.No Discipline / Group Measurand or Measurement range Calibration and Calibration or


Reference and additional Measurement Measurement
Material/Type of parameters where Capability(CMC)(±) Method or
instrument or applicable(Range procedure)
material to be and Frequency)
calibrated or
measured / Quantity
Measured /
Instrument
129 MECHANICAL- UTM, Universal Testing 1 kN to 50 kN 0.70% Using precision loadcell
TENSION CREEP Machine Compression with Indicator as per IS
AND TORSION and Tension 1828:2005 (Part I)
TESTING MACHINE

130 THERMAL- RTD ,Thermocouple, 50 ºC to 400 ºC 1.9ºC Using Standard RTD,


TEMPERATURE Thermistor Sensors Dry block calibrator &
with / without Indicator, Digital Multimeter
Temperature gauge/
Temperature bath,
controllers and
Transmitter with sensor

131 THERMAL- Temperature Indicator -50 ºC to 400 ºC 1.9 ºC Standard RTD with
TEMPERATURE of cold chamber, oven, Indicator
incubator, furnace, bath
at single position

This is annexure to 'Certificate of Accreditation' and does not require any signature.