Professional Documents
Culture Documents
We use cookies to provide visitors of our website the best possible Accept and close
experience. To learn more how we use cookies or how to block
cookies, please visit our cookie policy.
https://www.biotek.com/products/accessories/absorbance-test-plate/ 1/2
8/30/2018 Accessory: Absorbance Test Plate
BioTek's Absorbance Test Plate helps meet regulatory instrumentation validation requirements by
We use
testing cookies
four criticaltospecifications:
provide visitors of our website the best possible Accept and close
experience. To learn more how we use cookies or how to block
Alignment
cookies, please visit our cookie policy.
Accuracy
Reproducibility
Linearity
Each Test Plate consists of the following:
NIST (NBS)-traceable neutral density glass filters for checking accuracy, reproducibility, and
linearity.
Glass filter for monochromator accuracy.
Blanking verification holes.
Alignment verification holes
Calibration data at 8 wavelengths is supplied for full range testing.
The Absorbance Test Plate includes a data sheet for these commonly used wavelengths: 405,
450, 490, 550, 620, 630, 690, and 750 nm. Custom wavelength data is available for an additional
charge (Contact Customer Care for more information).
BioTek's Absorbance Test Plate complies with the IVD (In Vitro Diagnostics) Directive.
BioTek Instruments, Inc., headquartered in Winooski, VT, USA, is a worldwide leader in the design, manufacture,
and distribution of innovative life science instrumentation. These products enable life science research by
providing high performance, cost-effective analysis and quantification of biomolecules, biomolecular
interactions and cellular structure and function across diverse applications. BioTek espouses a "Think Possible"
approach that sets the tone for fresh ideas, unsurpassed customer service and original innovations.
Copyright BioTek Instruments, Inc. ©2018 Phone: (888) 451-5171 Contact Us
https://www.biotek.com/products/accessories/absorbance-test-plate/ 2/2