You are on page 1of 2

8/30/2018 Accessory: Absorbance Test Plate

We use cookies to provide visitors of our website the best possible Accept and close
experience. To learn more how we use cookies or how to block
cookies, please visit our cookie policy.

Absorbance Test Plate


Part Number: 7260522
For use with: Cytation 5, PowerWave HT, Synergy Neo2, Epoch 2, Cytation 1, Cytation Multi-
Mode, Epoch, Synergy H1, ELx808, Synergy HTX, 800 TS, Synergy LX

BioTek recognizes the importance of verifying the performance of analytical instrumentation.


Many regulatory agencies and laws such as CLIA 88 and Good Laboratory Practices (GLP) require
microplate readers to be checked periodically to ensure they continue to meet factory
specifications. BioTek's Test Plates make such testing easy, fast and accurate. By providing a
convenient method of monitoring the performance of your reader over time, the Absorbance Test
Plate can be an invaluable aid in diagnosing system performance of BioTek’s Absorbance
Microplate Readers, as well as a documentation tool in your research or clinical laboratory.
All BioTek automated absorbance readers can be quickly and simply tested for critical
performance specifications using two complementary methods uniquely developed by BioTek
Instruments:
BioTek's Absorbance Test Plate uses neutral density filters and precision machined
alignment holes to confirm performance specs in the visible range (400 – 800 nm). It also
includes a glass filter to test the wavelength accuracy of monochromator-based
absorbance systems.
BioTek's on-board Reader Diagnostic Self-Test verifies the reader's optical and electronic
performance.

https://www.biotek.com/products/accessories/absorbance-test-plate/ 1/2
8/30/2018 Accessory: Absorbance Test Plate

BioTek's Absorbance Test Plate helps meet regulatory instrumentation validation requirements by
We use
testing cookies
four criticaltospecifications:
provide visitors of our website the best possible Accept and close
experience. To learn more how we use cookies or how to block
Alignment
cookies, please visit our cookie policy.
Accuracy
Reproducibility
Linearity
Each Test Plate consists of the following:
NIST (NBS)-traceable neutral density glass filters for checking accuracy, reproducibility, and
linearity.
Glass filter for monochromator accuracy.
Blanking verification holes.
Alignment verification holes
Calibration data at 8 wavelengths is supplied for full range testing.
The Absorbance Test Plate includes a data sheet for these commonly used wavelengths: 405,
450, 490, 550, 620, 630, 690, and 750 nm. Custom wavelength data is available for an additional
charge (Contact Customer Care for more information).
BioTek's Absorbance Test Plate complies with the IVD (In Vitro Diagnostics) Directive.

RETURN TO THE ACCESSORY SEARCH

BioTek Instruments, Inc., headquartered in Winooski, VT, USA, is a worldwide leader in the design, manufacture,
and distribution of innovative life science instrumentation. These products enable life science research by
providing high performance, cost-effective analysis and quantification of biomolecules, biomolecular
interactions and cellular structure and function across diverse applications. BioTek espouses a "Think Possible"
approach that sets the tone for fresh ideas, unsurpassed customer service and original innovations.
Copyright BioTek Instruments, Inc. ©2018 Phone: (888) 451-5171 Contact Us

https://www.biotek.com/products/accessories/absorbance-test-plate/ 2/2