Electrode

Technical Information
ELEC50 / ELEC60
Introduction Intended use and key features
The electrode (ELEC60/ELEC50) is The electrode is to be used as the control input(s) for a myoelectric
designed and manufactured by Steeper. It prosthesis. It is suitable for direct skin contact:
is the same size and shape as electrodes –– This electrode will control prosthetic devices from most major manufacturers.
from other manufacturers and can be –– Proportional output signals.
directly fitted into the cavities prepared –– Suitable for child or adult systems.
for these electrodes. For all related –– High sensitivity (2000-100,000 fold) and range (90-450Hz).
products please refer to Steeper Upper –– Able to function with threshold control using signals as low as 10V.
Limb Prosthetic Components Catalogue. –– Interference protection from common power sources, depending on electrode
type, and high frequency emitting devices.
i NOTE –– An Electrode Gain Control (EGC) is included to allow control of the myoelectric
Ensure the patient receives the information signal providing optimal control of the terminal device.
card included in the product packaging. –– Two fabrication options included in the kit.
–– The contacts are coated with pure titanium, a biocompatible material and an
WARNING excellent conductor of signal.
Please read the information contained in this –– No known contraindications.
manual carefully before using the product.
Safety information Standard compliance
–– Before use, check for any visible electrode damage, particularly the titanium The CE-marking has been carried out in accordance with the requirements
contacts. of the 93/42/EEC guidelines.
–– The electrode should be placed against undamaged skin, avoiding scarred or Steeper operate a UKAS approved quality management system and fully
unhealed areas. complies with the requirements of BS EN ISO 9001:2015, ISO27001:2005,
–– Do not disassemble the electrode. ISO50001 & TickIT Issue 5. This certifies that Steeper meets the
–– The electrode must only be used within the power range recommended in this appropriate international quality standards for design, manufacture and
technical literature. 8747
supply of prosthetic products and user software.
–– Do not immerse the electrode in water.
–– Do not attempt to modify the product, seek specialist advice if required. Steeper is registered with both the Medicines and Healthcare
–– Please ensure the patient receives the information card included in the product Regulatory Authority in the UK and the Food and Drugs Administration
packaging. of the United States Government for the manufacture and supply of
–– Sensitive to EMC radiation. prosthetics and orthotics products.
–– In case of an emergency remove forearm immediately. For internal batteries turn MHRA Registration N°: 0000006617 FDA Registration N°: 9612243
charge point switch to OFF position. For external batteries disconnect battery
from housing. IEC 60417-5333 BF Approved part.
Operating instruction manual.
Product details Mains power system map

ELEC50 ELEC60
Light grey Dark grey

–– Provides interference protection –– Provides interference protection


in countries operating 50Hz mains in countries operating 60Hz mains
power systems (see map). power systems (see map).
–– Appropriate control device for –– Appropriate control device for
prosthetic devices from all major prosthetic devices from all major
manufacturers. manufacturers.
50Hz electrode (ELEC50) should
–– Titanium contacts for optimum –– Titanium contacts for optimum be used in these areas.
conductivity of signal and conductivity of signal and
60Hz electrode (ELEC60) should
biocompatibility. biocompatibility. be used in these areas.
In the box Principal dimensions

B Imperial Metric

E25591
1. A - Electrode width 11
⁄16” 18mm
5.
B - Electrode length 1
1 ⁄16” 27mm
A
C - Electrode height 3
⁄8” 10mm

Vacuum kit Weight 1


⁄6 oz 4.4g
3. o o o o
Temperature range 5 F to 140 F -15 C to 60 C
4.
2. Max humidity

E25590
95%

Maximum atmospheric pressure 104Pa

C Power supply 5V to 16V*

Lamination kit 1.
2.
3.
Vacuum kit
Lamination kit
Electrode (ELEC50 or
4.
5.
6.
Adjustment tool
Spare suspension legs
User guide card
Bandwidth 90-450Hz

ELEC60) 7. Technical information *Steeper batteries recommended


Further information Gain level adjustment
To adjust and balance the electrode, turn the dial on the back of the product using the
A guide for using the lamination kit and Spares of the following parts can be
tool provided. Number 1 is the lowest setting, number 7 is the highest setting.
vacuum forming kit is available at: ordered using the appropriate part
number. 1. 2. 3.
www.steepergroup.com or
www.steeperusa.com
Cable Retaining Plug B30819
A printed copy is also available upon
request. Suspension Legs B31032

Lamination Kit
Vacuum Kit
Cable Retaining Plug ELECSP WARNING
Gain Level Do not force the dial
Adjustment Tool beyond the end stops.
Connecting electrode
1. 2. 3. 4. WARNING 5. 6.
Use NLU0010 silicone
Ensure blue side grease to seal the
of electrode cable cable retaining block.
(ECA=300, 600 or Remove any excess
1000) is facing towards after connection.
the electrode.
Electrode suspension leg removal Electrode suspension leg reattachment
If the electrode suspension legs become worn over time, they may need replacing. There 1. 2. 3.
are two replacements included. Firstly remove them as shown and then replace as shown. Apply a small amount Insert the suspension
Remove any excess
of NAD0107 glue to leg into the cavity.
1. 2. 3. adhesive after insertion.
the underside of the
electrode suspension
leg in the area shown.
Returns Warranty
Returns - If you need to return your product for any reason please contact Warranty - This product is covered by a 12 month warranty period against
your local distributor or customerservices@steepergroup.com or inquiries@ defective design and manufacture.
steeperusa.com providing full details. A returns number will be issued allowing
your items to be dealt with promptly. The electrode is a solid state maintenance free device created using ABS
plastic and pure titanium. Store in a cool dry place.
Disposal - If you no longer require your product please return it to your local
supplier or Steeper/SteeperUSA with a request to dispose of it safely. Customer Services / Technical Support - For technical assistance/support
on any Steeper product please contact us:
Legal - The design and manufacture of Steeper equipment and components
are subject to a policy of continuous reappraisal. The company therefore USA: Rest of World:
reserves the right to introduce changes and withdraw products without prior inquiries@steeperusa.com customerservices@steepergroup.com
notice.
210 481 4126 +44 (0) 870 240 4133
Notes
SteeperUSA Steeper Group
3619 Paesanos Parkway Unit 3, Stourton Link
Suite 200 Intermezzo Drive
San Antonio Leeds
TX 78231, USA LS10 1DF

210 481 4126 +44 (0) 870 240 4133


inquiries@steeperusa.com customerservices@steepergroup.com
www.steeperusa.com www.steepergroup.com

RSLLIT327 Issue 4
© 2017 Steeper

You might also like