ADVANCE AND
PREVENTIVE
MAINTENANCE OF
ESU
TECHNICAL
SERVICE
IKATEMI EVENT ON
BANDUNG
SATURDAY,10 APRIL 2021
PRESENTED BY
MUHAMMAD HIDAYAT FAISAL (08118587360)
NATIONAL SERVICE COORDINATOR ENERGY BASED DEVICE PRODUCT
PT. MEDTRONIC INDONESIA
Te r m i n o l o g y
Basic
1. CAUTER/
ELECTROCAUTER
2. ESU
SS-501SX
(ELECTROSURGICAL
UNIT )
3. DIATHERMI
4. VALEY
5. VALLEYLAB
VLLS10GEN
SS-200E
VLFT10GEN
VLFX8GEN
FOUR BASIC ELECTROSURGICAL PROCEDURES
MONOPOLAR
Cutting (vaporization)
Coagulate (Latin for “to curdle”)
Fulguration (Latin for lightning) – arching (“spraying”) between
electrode and tissue to stop heavy bleeding
Desiccation (Latin to dry up) – touching the tissue to achieve
hemostasis in individual vessels
Ablation (Latin to carry away) – heats and coagulates large
volumes of tissue (e.g., liver or tumor)
Bipolar (between two poles) often used to achieve
hemostasis during brain surgery
Blend (cut and coagulate)
Vessel sealing (uses bipolar technology)
2-29
Electrical
Basic Theory
Electricity always . . .
• Seeks ground (its source)
• Seeks path of least resistance
FREQUENCY SPECTRUM
9-9 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-11 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-12 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
GENERATOR
COMPARISON
-C
9-13 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
KEY DIFFERENCES
SIZE & WEIGHT
Dimensions -C
9-14 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-15 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
Tampak Depan
6. Pemberitahuan Untuk:
1. Defibrilator
2. FT10 energimya ter isolasi
3. Unit tegangan tinggi
1 2 34 5 4. REM™ patient return
6 7 electrode
7. Soket untuk REM connector
1.On/Off: Tombol On / off
2.Restore Settings: Tombol Reset untuk membalikan settingan
terakhir
3.Audio Volume: Tombol untuk menaik atau turunkan volume
4.Service/Settings: Tombol untuk ke service Menu
5.REM indicator : merah jika rem terconnect tapi tidak terpasang ke
pasien, hijau jikau rem terpasang ke pasien, dan abu-abu ketida
instrument REM belum terconnect ke unit.
9-16 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
TAMPAK BELAKANG
6. Port monopolar 1
7. Port Monopolar 2, (membutuhkan
7 6 adapter untuk menggunakan standart
footswitch)
9 8 8. Port Bipolar
5 4 3 2 1 9. Port Footswitch Ligasure / Bipolar
Reseksi
1. Connector Kabel Power
2. Grounding soket
3. Port WI-fi
4. Port ethenet kabel
5. Port penghubung antara unit dengan ECG
9-17 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
Tampilan Layar Sentuh
Berwarna menandakan
unit sudah menyala atau
Aktif , dan warna akan
menyala di tiap-tiap
quadrant yang terpasang
atau port terkoneksi oleh
instrument
9-18 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
SETTING MONOPOLAR
Permintaan untuk memasangkan
REM pad kepada pasien, dan jika
setelah dipasangkan kepasien
tetap menyala maka, periksa
penempelan PAD terhadap tubuh
dan periksa koneksi Port dan
connector sudah terpasang
sempurna atau tidak.
9-19 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
Monopolar Tersedia dua Mode
CUT & COAG,
Dan untuk mengeset settingan
daya silahkan tekan tanda atas
dan bawah
9-20 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
BIPOLAR SETUP
9-21 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
AUTO BIPOLAR
9-23 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
AUTO BIPOLAR TIME DELAY
9-24 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
LIGASURE
No Feature Benefit
1 Safety REM – Return Electrode Monitoring. Memonitor kualitas kontak, sehingga apabila tidak menempel dengan utuh
maka alarm generator akan berbunyi dan generator otomatis tidak berfungsi.
2 Tissue Fect Sensing Technology. Kemampuan generator untuk bereaksi seketika (434,000/detik) terhadap
perubahan kerapatan jaringan.
3 High PER – Power Efficiency Rating 99% Ukuran kemampuan generator dalam mengeluarkan power sesuai dengan
power yang ditentukan untuk berbagai macam tipe jaringan.
4 Force Triverse. Kontrol setting power dari area steril, Valleylab mode.
5 Simultaneous coag Two pencils activating at the same time Kemampuan generator bekerja secara otomatis untuk mengaktifkan bipolar
split power sesuai waktu yang akan di set oleh operator, maka operator akan fokus ke area
operasi tanpa harus mencari footswitch.
6 Auto bipolar Kemampuan generator bekerja secara otomatis untuk mengaktifkan bipolar
sesuai waktu yang akan di set oleh operator, maka operator akan fokus ke area
operasi tanpa harus mencari footswitch.
9-26 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
WORKSHOP AND
ACTION
FT10
MAINTENANCE &
CALIBRATION
INTRODUCTION TO THE CALIBRATION PROCESS
THREE CALIBRATION LEVELS
9-29 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
PREVENTIVE MAINTENANCE
FT10
9-30 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-31 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-32 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-33 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
MECHANICAL DESIGN
EASE OF REPAIR FOR BIOMEDS
PCBA Stacking
Layer 3 EPAC
RF PCBA
Layer 2 EPAC
Layer 1 EPAC
Controller PCBA
9-34 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
TOUCH SCREEN CALIBRATION
PART 1 - NINE TARGETS
9-37 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
TOUCH SCREEN CALIBRATION
PART 2 - FIVE TARGETS
9-38 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
TOUCH SCREEN VERIFICATION
19TARGETS
9-39 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
TROUBLESHOOTING
9-41 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
LOG FILES
DEEPER AND MORE DETAILED THAN IN THE PAST
If you drill down, you can discover a lot of information about an error
Event
• Provides in-depth information about any event
Details
And…there is also an Error Histogram, which will show how many times
this same error has occurred
Let’s walk through a typical error. In this case, it is a faulty fan, with an
error code of 333.
9-42 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 1
9-43 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 2
Select LOGS
9-44 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 3
Select Error
9-45 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 4
9-46 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 5
Before selecting the error, move the scroll bar to the right to get more
information
9-47 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 6
Select the error line and the Error Details screen will appear
9-48 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 7
Again, move the scroll bar right to get more information (if it exists).
9-49 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 8
Select the SHOW EVENTS tab and the Event Log screen will appear.
9-50 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 9
Scroll down and look for Error 333 with the time stamp of 2015-10-15 11:22:47
9-51 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 10
Move the scroll bar right to get more information about all the information
above and below the error.
9-52 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
REVIEWING AN ERROR IN A LOG FILE
FAN ERROR 333 – STEP 11
If you select the line in the log file, it will provide additional details about that
event.
9-53 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
ERROR HISTOGRAM
FAN ERROR 333 – STEP 12
If you want to see how many times that same error has occurred, go to the Error
Histogram screen.
9-54 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
FT10 OUTPUT TESTING
RF OUTPUT TESTS
GENERAL SETUP INSTRUCTIONS
9-56 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
VERIFYING THE REM FUNCTION
in the table. 60 Ω 53 Ω - 67 Ω
0Ω 0Ω-4Ω
NOTE: If the REM ZMag reading exceeds 2000 Ω, the
Open (no decade box) < 2000 Ω
Steering Relay PCBA will need to be replaced to ensure
the proper performance of REM.
9-57 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
TESTING COAG MODE WITH A 500 Ω LOAD
9-59 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
PVT MONOPOLAR
9-60 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
TESTING BIPOLAR WITH 100 Ω LOAD
9-61 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
TESTING LIGASURE & BPR MODES WITH 50, 100, & 500 Ω LOADS
9-65 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
VERIFYING CROSS COUPLING – MONO 1 CURRENT
9-66 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
VERIFYING CROSS COUPLING – MONO 2 CURRENT
9-68 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
VERIFYING CROSS COUPLING – BIPOLAR CURRENT
9-69 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
RF OUTPUT TESTS
CHECKING HIGH FREQUENCY LEAKAGE CURRENT
9-71 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
SERVICE DOCUMENTATION & HELP FILES
BIOMED CONNECT™ IS THE PRIMARY LOCATION
Training materials, technical documentation and help files will be updated on the
Biomed Connect™ website.
9-75 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.
9-76 This presentation is intended for company-internal audiences only. It is NOT Zinc approved.