You are on page 1of 5

ANALYTICAL SCIENCES FEBRUARY 2005, VOL.

21 143
2005 © The Japan Society for Analytical Chemistry

Sample Preparation for Evaluation of Detection Limits


in X-ray Fluorescence Spectrometry
M. K. TIWARI,† A. K. SINGH, and K. J. S. SAWHNEY

Synchrotron Utilization Division, Centre for Advanced Technology, Indore-452 013, India

The influence of analyte mass concentration on determination of detection limits in X-ray fluorescence spectrometry has
been investigated experimentally. Both the total reflection X-ray fluorescence (TXRF) and the conventional energy-
dispersive X-ray fluorescence techniques have been used to derive the dependence of analyte mass concentration on the
values of detection limits. Results obtained indicate that values of detection limits are optimum, or in other words, they
are closer to the true detection limit of the technique, when analyte concentrations are in the range of 10 times of the
detection limit.

(Received June 30, 2004; Accepted September 10, 2004)

described in detail by Pajek et al. They employed the process


Introduction of “Censoring” for determining the detection sensitivities in
TXRF. The parameters that improve the detection limits of
X-ray fluorescence spectrometry is one of the widely used various X-ray spectrometric techniques, such as PIXE,
techniques for elemental analysis of materials. Using the WDXRF, synchrotron excited XRF and γ-ray spectroscopy,
energy-dispersive X-ray fluorescence (EDXRF) analysis have been discussed thoroughly by several workers.16–19
technique, non-destructive multielement analysis at ppm levels Furthermore, it has also shown that in X-ray tube excited XRF,
can be performed for elements ranging from sodium to uranium the detection limit can be improved by properly selecting the
in the periodic table.1,2 The technique finds several applications primary filters.20 Moreover, a more elaborate theoretical
to a variety of fields such as archeological, geological and description for detection limit calculation in TXRF with various
environmental.3–6 Total reflection X-ray fluorescence (TXRF) modes of analysis, viz. grazing incidence (TXRF) and grazing
technique is a variant of the conventional EDXRF technique,7–9 emission (GE-XRF) conditions, has been given by Sanchez.21
where the principle of total external reflection is used to restrict With developments/innovations in the fields of X-ray sources,
the penetration of the primary beam into the specimen substrate. sample preparation, nuclear electronics, detection systems and
This yields a substantial reduction of scattered background, data processing procedures in the last decades, the XRF
which improves the detection limit of the system to sub- techniques have also been refined to give better and better DLs.
ppm/ppb levels. Synchrotron radiation based X-ray fluorescence (SR-XRF) and
The term “detection limit” (DL) is an important aspect of any total reflection X-ray fluorescence (SR-TXRF) techniques are
technique because it is an indication of the sensitivity of the the examples of such trends, where parts per trillion (ppt) level
technique. As the name itself indicates, the detection limit is a or, in absolute mass, femto gram detection sensitivities have
measure of the minimum quantity that can be detected by an been achieved.22–24 Normally, XRF techniques are comparative
instrument or technique under the given experimental in nature and type-standards are generally used to calibrate the
conditions (sample geometry, sample matrix, analysis time, spectrometers. Similarly, synthetic standards are employed to
etc.). A through description about detection limits can be determine the DLs. In the literature, no guidelines are available
obtained from publications of Currie et al.10–14 The detection for the selection of analyte concentration to be used for detection
values for an element are largely influenced by these limit determination. Our experience, over the years, has been
parameters. For example, in a multielement sample, the that the DLs tend to depend on the analyte concentration in the
presence of several peaks in the fluorescence spectrum can synthetic standard employed to determine the DLs.
modify the spectral background and hence change the DL value We have undertaken a systematic study to resolve this issue;
of the element of interest. Such a limitation sometimes results our results give guidelines for what the analyte mass
in “non-detects” i.e. some concentrations that cannot be concentrations should be for achieving the best possible DLs
measured directly.15 The problem of non-detects often occurs under the given experimental conditions. This study indicates
for biological and geological samples where the sample matrix that, to derive the optimum detection limit values, one should
strongly influences the spectral background. The presence of locate the analyte mass concentration in the synthetic standards
non-detects in XRF and TXRF analysis techniques has been in the range of 10 times the detection limit.

To whom correspondence should be addressed.
E-mail: mktiwari@cat.ernet.in Theoretical Background
K. J. S. S. present address: Diamond Light Source Ltd., Rutherford
Appleton Laboratory, Chilton, Didcto-OX11 0QX, UK. Mathematically, as per the IUPAC recommendation, the
144 ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21

Fig. 1 Variation of Zn detection limits with analyte concentration determined using EDXRF
technique for different acquisition times. Excitation source: Cd109. a, Solid square (s); b, solid circle
(s); c, solid triangle (g) show the measurements performed for acquisitions time 1000, 5000 and
10000 s, respectively whereas a solid line (m) shows the polynomial fit.

detection limit has been defined in terms of standard counting of analyte. From Eq. (1), it is clear that, for a specified counting
error of the background intensity.1 If σP and σB are the counting time, the minimum detectable concentration of analyte or
errors of the individual peak and the background, respectively, detection limit depends mainly on two parameters: the
then the standard counting error for a net count can be background intensity IB and the net area intensity per unit
expressed as, analyte mass (IA/CA).
—————
σ = √ σ P2 + σ B2
Experimental
In the limit of detection one can assume σP ≈ σB. Therefore
Both the EDXRF and TXRF techniques have been used to
——
σ ≈ √ 2σB perform X-ray fluorescence measurements. For EDXRF, an in-
house developed EDXRF spectrometer25 has been employed
For 95% confidence level, the counting error would be that comprises of a Si(Li) detector of energy resolution ∼155 eV
at 5.9 keV and an EG&G ORTEC 92X spectrum master that has
——
2σ ≈ 3ρB = 3 √ NB a built-in spectroscopy amplifier, a multi channel analyzer and a
HV bias supply. A 25 mCi Cd109 annular radioisotope is used as
By taking into account the counting time T and the slope m of an excitation source. Several synthetic standards, ranging in
fluorescence intensity vs. analyte concentration curve, one can three decades of analyte concentrations, were prepared. Zn and
write the expression for minimum detectable concentration for Cr elements were chosen as the analyte elements and their
an analyte element as synthetic standards were prepared in the form of pellets of mass
∼100 mg/cm2, by mixing their appropriate salts in cellulose
3 NB matrix. Moreover, to confirm the reproducibility of system
CDL = m
T geometry, we prepared synthetic standards of various elements
in various concentrations range and we analyzed them
or separately. The system geometry was found to reproduce
within ±0.5%. To perform the same studies using the TXRF
technique, researchers have used the indigenously developed
CDL = 3 IB (1)
 IA  TXRF spectrometer.26 Aqueous residues of U of different
 CA  concentrations were prepared from liquid solutions and polished
Si substrates were used as the sample carriers. Mo X-ray tube
where, IB = background intensity (NB/T); IA = net area intensity; (17.4 keV) was used as the excitation source.
CA = mass concentration of analyte; NB = background counts in
a given time T; m = IA/CA, i.e., slope of analyte counts-
concentration curve; CDL = minimum detectable concentration
ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21 145

Fig. 2 Variation of relative uncertainty for a signal at different Fig. 3 Influence of analyte concentration on the background
analyte concentrations (expressed in DL units) for two different risk intensity and slope m of analyte fluorescence intensity–concentration
levels, 5% and 0.05%, calculated using Eq. (2). The variation of self- calibration curve. Analysis time, 10000 s; analyte, Zn; slope m, (å);
absorption, A-factor, for Zn-Kα signal at different analyte background intensity, (©).
concentrations is also shown in this figure. The figure shows that for
analyte mass ranging in 10 – 30 times of DL value, both self-
absorption and relative uncertainties are optimum.
In the limit of detection where self-absorption effects27 for
analyte elements are significantly low, LQ simply represents the
smallest analyte mass that can be expected to generate a net
signal intensity with a relative error less than a selected limit r.
Results and Discussion This can be understood more clearly from Fig. 2. Here, the
relative uncertainty for a signal at different levels of analyte
Figures 1a, b and c show the variation of detection limits of Zn mass (expressed in DL units) has been plotted for two different
with various concentrations of Zn for analysis times of 1000 s, risk level 5% (k = 1.645) and 0.05% (k = 3.291) along with the
5000 s and 10000 s, respectively. Since the DLs depend self absorption of Zn-Kα signal. For analyte masses ranging
inversely on the square root of analysis time, the DLs are better around 10 – 30 times of DL, the relative uncertainties are
for larger analysis times (Fig. 1). The figure shows that the significantly low, e.g. comparable to acceptable limits of
dependency of DL on analyte mass concentration generates statistical errors (∼0.5 – 2%) and at the same time self-
broad minima and that the DLs are minimum for analyte absorption effects do not affect severely the analyte signal
concentrations ranging from 10 – 30 times of DL. Moreover, intensity. Beyond this range, at higher analyte concentrations,
one can see that the selection of an analyte concentration about though the relative uncertainties improve, the sample self-
10 times that of DL is more appropriate for getting the absorption effects become dominant. Figure shows that
minimum DL value. For example, for 1000 s (5000 s) analysis selecting analyte concentration in the range of 10 – 30 times of
time, the minimum of DL for Zn is ∼59 (∼27) ppm which DL would yield a fluorescence signal under acceptable
corresponds to analyte mass of 600 (300) ppm. Similarly, for uncertainties. Self-absorption effects for analyte signal can be
analysis time of 10000 s, Zn DL is minimum (∼18 ppm) for the ignored in this range. Though one can use any analyte mass
analyte mass of 200 ppm. The appropriateness of analyte between 10 – 30 times of DL value for detection limit
concentration corresponding to 10 times that of DL can be measurement, it is a common practice to use an analyte mass
explained by the Currie hypothesis of “third detection limit”.10 close to the DL value i.e. ∼10 times of DL.
Beside detection limit, which is a characteristic of a The applicability of analyte mass corresponding to 10 times of
measurement, Currie given a concept of third detection limit, DL is shown in Fig. 3. Here, the net area intensity per unit
“determination limit”, LQ, above which the precision of a analyte concentration and the background fluorescence intensity
measurement can be made good enough. The relationship have been plotted as a function of Zn analyte concentration.
between determination limit LQ and detection limit CDL can be This figure shows that net area intensity of Zn-Kα per unit
approximated by16 analyte concentration, also referred to as the slope in Eq. (1),
show a peaking behavior for Zn analyte concentration of ∼200
CDL ppm. On the other hand, the background intensity increases
LQ = ——
2kr monotonously, though this increase is very small. This is
because, at higher analyte concentrations, apart from sample
where, k defines the risk factor for a signal in terms of a fractile absorption effects, the tailing of the fluorescence peak increases,
of a Gaussian distribution and r is the required relative leading to higher background below the fluorescence peak,
uncertainty. which in turn reduces the value of slope m. If one lowers the
analyte concentration, the background that appears due to
If LQ is set to n times of CDL i.e. LQ = nCDL then one can derive a increase of tailing width of the fluorescence peak can be
relation, reduced significantly. In this case it becomes almost constant
and equal to the value that one obtains in case of a blank
1 sample. By optimizing the width of the fluorescence peak
r = —— (2)
2nk within 3σ limit and keeping background line well-known in the
area of the peak, the optimum value of m can be obtained for
146 ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21

Fig. 5 Variation of U detection limits with analyte concentration


Fig. 4 X-ray fluorescence spectra recorded for various determined using TXRF technique. Excitation source: Mo X-ray
concentrations of Zn analyte. Analysis time, 10000 s; excitation tube, 40 kV, 25 mA. Open circles (A) shows the measurements
source, Cd109 radioisotope source. The figure shows that background performed for an acquisition time of 500 s whereas the solid line
region/intensity under a fluorescence peak increases non-linearly (m) shows the polynomial fit. The inset shows the variation of net
with increasing analyte concentrations. area intensity for uranium Lα (slope m) and the background intensity
at different analyte concentrations.

analyte mass ∼10 times of DL. Figure 4 shows a more


elaborated picture of this non-linear behavior of the background ppb, then 100 ppb (0.1 ppm) of analyte concentration would be
intensity, wherein fluorescence spectra for various analyte appropriate for determining this detection limit. In TXRF, as
concentrations have been plotted. The figure shows that, at we have used aqueous samples for determination of DLs, apart
large analyte concentrations the background region/intensity from the reasons discussed above that influences background
under a fluorescence peak is higher compared to that at lower intensity, there is a certain probability of increased surface
analyte concentrations, at the same time the net area intensity roughness at large volumes of analyte. The inset of Fig. 5
behaves linearly at lower concentrations, which is not strictly shows the variation of net area intensity and background
true at higher analyte concentrations because of sample intensity as a function of analyte (uranium) concentration. It is
absorption effects. Since slope m as well as the background found that net area intensity varies like it does in cases of
intensity IB governs the detection limit value as defined in Eq. conventional XRF measurements, while the increase of
(1), the detection limits improve for analyte concentrations background is more rapid at large analyte concentrations. This
ranging in 100 – 500 ppm and a minimum is obtained at analyte is mainly due to relatively large surface roughnesses induced at
concentration of ∼200 ppm. Further reduction of analyte higher analyte mass/volumes.
concentration leads to poor statistics of analyte peak, resulting We have also taken into account the total errors contributed in
in less than optimum values for DL. It is often true at very low DL measurement as well as in preparation of analyte
analyte concentrations that the number of counts that appears concentration. Since the detection limit depends linearly on
under a fluorescence peak sometimes may not be sufficient to analyte concentration, the net error in these two parameters has
support the Gaussian approximation i.e. the shape of the been calculated partially. We have assumed that all
fluorescence peak no longer remains Gaussian; rather, it shifts measurement errors; viz., errors in net area intensity, sample
to a Poisson distribution. This not only results in large errors in preparation error, statistical error, or geometry reproducibility,
net area estimation but it also sometimes becomes difficult to influence the DL value and have been determined using
define properly a window/ROI for a fluorescence peak in a synthetic standards of known concentrations. The total error
multichannel pulse height analyzer (MCA) spectra. Although contributed to DL measurement has been found to vary from 2 –
the detection limit values determined for two analytes Cr and Zn 8%. The error is higher at low analyte concentration and in this
are found to be minimum for analyte concentrations ranging case the statistical errors are dominant. On the other hand, to
from 10 – 30 times of DLs, the dependency of DL on analyte estimate errors in preparation of analyte concentrations, we
concentrations still shows that the selection of analyte prepared three sets of synthetic standards and analyzed each
concentration corresponding to 10 times of DLs would be better standard for each concentration of analyte element. The net
for getting optimum values of DL. error in analyte mass concentration has been determined
To further investigate the influence of analyte concentrations indirectly by measuring the fluorescence intensity. The total
on detection limit values, an independent study using TXRF error in analyte concentration has been found to vary from 2 –
technique has also been performed. The detection limit values 10 %.
in TXRF technique are generally determined by employing
aqueous residue generated from synthetic or standard solutions.
We used various concentrations of synthetic-standard solution Conclusions
of uranium for determining the detection limit of uranium using
this technique. The dependency of detection limit on analyte The systematic study presented here gives guidelines for the
concentration studied using TXRF technique, Fig. 5, reconfirmed most suitable analyte mass concentrations for achieving the best
that the best DL is obtained for analyte mass concentration possible DLs in XRF techniques. The results obtained indicate
corresponding to ∼10 times of DL value. For instance, if the that the optimum value of detection limit will correspond to
true detection limit of the TXRF technique for uranium is 10 analyte mass ~10 times of the best detection limit. The same
ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21 147

inference can also be drawn from the TRXF analysis techniques 15. M. Pajek and A. Kubala-Kukus, in “10th International
results, thereby indicating widespread possible applications of Conference on Total Reflection X-ray Fluorescence
this guideline especially in calibration of XRF spectrometers for Analysis and the 39th Annual Conference on X-ray
their DLs under given experimental conditions. Chemical Analysis (TXRF 2003)”, 2003, Sept. 14 – 19,
Hyogo, Japan, 33.
16. Lars-Erik De Geer, Appl. Radiat. Isot., 2004, 61, 151.
Acknowledgements 17. L. A. Currie, Appl. Radiat. Isot., 2004, 61, 145.
18. G. L. Long and J. D. Winefordner, Anal. Chem., 1983, 55,
The authors thank R. V. Nandedkar for fruitful discussions and 712A.
suggestions. 19. C. J. Sparks Jr., “Synchrotron Radiation Research”, ed. H.
Winick and S. Z. Doniach, 1980, Plenum Press, New York,
459.
References 20. M. N. Ingham and B. A. R. Vrebos, Adv. X-ray Anal.,
1994, 37, 717.
1. E. P. Bertin, “Principles and Practice of X-ray 21. H. J. Sanchez, Spectrochim. Acta Part B, 2001, 56, 2027.
Spectrometric Analysis”, 1975, Plenum Press, New York. 22. K. Baur, S. Brennan, B. Burrow, D. Werho, and P. Pianetta,
2. R. E. Van Grieken, “Hand Book of X-ray Spectrometry”, Spectrochim. Acta, Part B, 2001, 56, 2049.
2nd ed., 2002, Marcel Dekker, Inc., New York, 433 – 498. 23. N. Awaji et al., Jpn. J. Appl. Phys., 2000, 39, Pt. 2, No.
3. K. J. S. Sawhney and G. S. Lodha, Computers and 12A, L1252.
Geosciences, 1989, 15(7), 1115. 24. C. Streli, P. Wobrauschek, H. Aiginger, W. Ladisich, and
4. G. Misra, K. J. S. Sawhney, G. S. Lodha, V. K. Mittal, and R. Rieder, Adv. X-ray Anal., 1994, 37, 577.
H. S. Sahota, Appl. Radiat. Isot., 1992, 43(5), 609. 25. K. J. S. Sawhney, M. K. Tiwari, A. K Singh, and R. V.
5. S. Piorek, Nucl. Instr. Meth., 1994, A353, 528. Nandedkar, “Proc. 6th National Seminar on X-ray
6. V. Zaichick, N. Ovchjarenko, and S. Zaichick, Appl. Spectroscopy and Allied Areas”, ed. S. K. Joshi, B. D.
Radiat. Isot., 1999, 50(2), 283. Shrivastava, and A. P. Deshpande, 1998, Narosa, New
7. Y. Yoneda and T. Horiuchi, Rev. Sci. Instrum., 1971, 42, Delhi, 130.
1069. 26. M. K. Tiwari, B. Gowri Sankar, V. K. Raghuvanshi, R. V.
8. P. Wobrauschek and H. Aiginger, Anal. Chem., 1975, 47, Nandedkar, and K. J. S. Sawhney, Bull. Mater. Sci., 2002,
852. 25(5), 435.
9. R. Klockenkamper, “Total-Reflection X-ray Fluorescence 27. Sample absorption correction factor, A, in X-ray fluorescence
Analysis”, 1997, John Wiley and Sons, Inc., New York.
10. L. A. Currie, Anal. Chem., 1968, 40, 586. 1 – exp –  µ(E0) µ(Ei) 
sinφ + sinϕ  M
can be evaluated from A = ,
11. L. A. Currie, “Lower limit of detection: definition and  µ(E0) + µ(Ei)  M
elaboration of a proposed position for radiological effluent  sinφ sinϕ 
and environmental measurements”, 1984, Report issued by where µ(E0) and µ(Ei) represent sample mass absorption
the US Nuclear Regulatory Commission, NUREG/CR- coefficients at incident energy E0 and characteristic
4007, 1 – 139. fluorescence line energy of the element of interest. φ and ϕ
12. L. A. Currie, Anal. Chim. Acta, 1999, 391, 103. are incident and exit angles; M represents the mass of the
13. L. A. Currie, J. Radioanal. Nucl. Chem., 2000, 245, 145. pellet in g/cm2. To determine the self-absorption correction
14. L. A. Currie, “Detection: international update, and some for analyte signal, normalization of A factor to that in case
emerging dilemmas involving calibration, the blank, and of blank sample is needed (A = 1 represents no absorption
multiple detection decisions”, 1997, Vol. 37, Chemometrics and A = 1.10 means 10% absorption of analyte intensity).
Intell. Lab Systems, 151 – 181.

You might also like