Professional Documents
Culture Documents
21 143
2005 © The Japan Society for Analytical Chemistry
Synchrotron Utilization Division, Centre for Advanced Technology, Indore-452 013, India
The influence of analyte mass concentration on determination of detection limits in X-ray fluorescence spectrometry has
been investigated experimentally. Both the total reflection X-ray fluorescence (TXRF) and the conventional energy-
dispersive X-ray fluorescence techniques have been used to derive the dependence of analyte mass concentration on the
values of detection limits. Results obtained indicate that values of detection limits are optimum, or in other words, they
are closer to the true detection limit of the technique, when analyte concentrations are in the range of 10 times of the
detection limit.
Fig. 1 Variation of Zn detection limits with analyte concentration determined using EDXRF
technique for different acquisition times. Excitation source: Cd109. a, Solid square (s); b, solid circle
(s); c, solid triangle (g) show the measurements performed for acquisitions time 1000, 5000 and
10000 s, respectively whereas a solid line (m) shows the polynomial fit.
detection limit has been defined in terms of standard counting of analyte. From Eq. (1), it is clear that, for a specified counting
error of the background intensity.1 If σP and σB are the counting time, the minimum detectable concentration of analyte or
errors of the individual peak and the background, respectively, detection limit depends mainly on two parameters: the
then the standard counting error for a net count can be background intensity IB and the net area intensity per unit
expressed as, analyte mass (IA/CA).
—————
σ = √ σ P2 + σ B2
Experimental
In the limit of detection one can assume σP ≈ σB. Therefore
Both the EDXRF and TXRF techniques have been used to
——
σ ≈ √ 2σB perform X-ray fluorescence measurements. For EDXRF, an in-
house developed EDXRF spectrometer25 has been employed
For 95% confidence level, the counting error would be that comprises of a Si(Li) detector of energy resolution ∼155 eV
at 5.9 keV and an EG&G ORTEC 92X spectrum master that has
——
2σ ≈ 3ρB = 3 √ NB a built-in spectroscopy amplifier, a multi channel analyzer and a
HV bias supply. A 25 mCi Cd109 annular radioisotope is used as
By taking into account the counting time T and the slope m of an excitation source. Several synthetic standards, ranging in
fluorescence intensity vs. analyte concentration curve, one can three decades of analyte concentrations, were prepared. Zn and
write the expression for minimum detectable concentration for Cr elements were chosen as the analyte elements and their
an analyte element as synthetic standards were prepared in the form of pellets of mass
∼100 mg/cm2, by mixing their appropriate salts in cellulose
3 NB matrix. Moreover, to confirm the reproducibility of system
CDL = m
T geometry, we prepared synthetic standards of various elements
in various concentrations range and we analyzed them
or separately. The system geometry was found to reproduce
within ±0.5%. To perform the same studies using the TXRF
technique, researchers have used the indigenously developed
CDL = 3 IB (1)
IA TXRF spectrometer.26 Aqueous residues of U of different
CA concentrations were prepared from liquid solutions and polished
Si substrates were used as the sample carriers. Mo X-ray tube
where, IB = background intensity (NB/T); IA = net area intensity; (17.4 keV) was used as the excitation source.
CA = mass concentration of analyte; NB = background counts in
a given time T; m = IA/CA, i.e., slope of analyte counts-
concentration curve; CDL = minimum detectable concentration
ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21 145
Fig. 2 Variation of relative uncertainty for a signal at different Fig. 3 Influence of analyte concentration on the background
analyte concentrations (expressed in DL units) for two different risk intensity and slope m of analyte fluorescence intensity–concentration
levels, 5% and 0.05%, calculated using Eq. (2). The variation of self- calibration curve. Analysis time, 10000 s; analyte, Zn; slope m, (å);
absorption, A-factor, for Zn-Kα signal at different analyte background intensity, (©).
concentrations is also shown in this figure. The figure shows that for
analyte mass ranging in 10 – 30 times of DL value, both self-
absorption and relative uncertainties are optimum.
In the limit of detection where self-absorption effects27 for
analyte elements are significantly low, LQ simply represents the
smallest analyte mass that can be expected to generate a net
signal intensity with a relative error less than a selected limit r.
Results and Discussion This can be understood more clearly from Fig. 2. Here, the
relative uncertainty for a signal at different levels of analyte
Figures 1a, b and c show the variation of detection limits of Zn mass (expressed in DL units) has been plotted for two different
with various concentrations of Zn for analysis times of 1000 s, risk level 5% (k = 1.645) and 0.05% (k = 3.291) along with the
5000 s and 10000 s, respectively. Since the DLs depend self absorption of Zn-Kα signal. For analyte masses ranging
inversely on the square root of analysis time, the DLs are better around 10 – 30 times of DL, the relative uncertainties are
for larger analysis times (Fig. 1). The figure shows that the significantly low, e.g. comparable to acceptable limits of
dependency of DL on analyte mass concentration generates statistical errors (∼0.5 – 2%) and at the same time self-
broad minima and that the DLs are minimum for analyte absorption effects do not affect severely the analyte signal
concentrations ranging from 10 – 30 times of DL. Moreover, intensity. Beyond this range, at higher analyte concentrations,
one can see that the selection of an analyte concentration about though the relative uncertainties improve, the sample self-
10 times that of DL is more appropriate for getting the absorption effects become dominant. Figure shows that
minimum DL value. For example, for 1000 s (5000 s) analysis selecting analyte concentration in the range of 10 – 30 times of
time, the minimum of DL for Zn is ∼59 (∼27) ppm which DL would yield a fluorescence signal under acceptable
corresponds to analyte mass of 600 (300) ppm. Similarly, for uncertainties. Self-absorption effects for analyte signal can be
analysis time of 10000 s, Zn DL is minimum (∼18 ppm) for the ignored in this range. Though one can use any analyte mass
analyte mass of 200 ppm. The appropriateness of analyte between 10 – 30 times of DL value for detection limit
concentration corresponding to 10 times that of DL can be measurement, it is a common practice to use an analyte mass
explained by the Currie hypothesis of “third detection limit”.10 close to the DL value i.e. ∼10 times of DL.
Beside detection limit, which is a characteristic of a The applicability of analyte mass corresponding to 10 times of
measurement, Currie given a concept of third detection limit, DL is shown in Fig. 3. Here, the net area intensity per unit
“determination limit”, LQ, above which the precision of a analyte concentration and the background fluorescence intensity
measurement can be made good enough. The relationship have been plotted as a function of Zn analyte concentration.
between determination limit LQ and detection limit CDL can be This figure shows that net area intensity of Zn-Kα per unit
approximated by16 analyte concentration, also referred to as the slope in Eq. (1),
show a peaking behavior for Zn analyte concentration of ∼200
CDL ppm. On the other hand, the background intensity increases
LQ = ——
2kr monotonously, though this increase is very small. This is
because, at higher analyte concentrations, apart from sample
where, k defines the risk factor for a signal in terms of a fractile absorption effects, the tailing of the fluorescence peak increases,
of a Gaussian distribution and r is the required relative leading to higher background below the fluorescence peak,
uncertainty. which in turn reduces the value of slope m. If one lowers the
analyte concentration, the background that appears due to
If LQ is set to n times of CDL i.e. LQ = nCDL then one can derive a increase of tailing width of the fluorescence peak can be
relation, reduced significantly. In this case it becomes almost constant
and equal to the value that one obtains in case of a blank
1 sample. By optimizing the width of the fluorescence peak
r = —— (2)
2nk within 3σ limit and keeping background line well-known in the
area of the peak, the optimum value of m can be obtained for
146 ANALYTICAL SCIENCES FEBRUARY 2005, VOL. 21
inference can also be drawn from the TRXF analysis techniques 15. M. Pajek and A. Kubala-Kukus, in “10th International
results, thereby indicating widespread possible applications of Conference on Total Reflection X-ray Fluorescence
this guideline especially in calibration of XRF spectrometers for Analysis and the 39th Annual Conference on X-ray
their DLs under given experimental conditions. Chemical Analysis (TXRF 2003)”, 2003, Sept. 14 – 19,
Hyogo, Japan, 33.
16. Lars-Erik De Geer, Appl. Radiat. Isot., 2004, 61, 151.
Acknowledgements 17. L. A. Currie, Appl. Radiat. Isot., 2004, 61, 145.
18. G. L. Long and J. D. Winefordner, Anal. Chem., 1983, 55,
The authors thank R. V. Nandedkar for fruitful discussions and 712A.
suggestions. 19. C. J. Sparks Jr., “Synchrotron Radiation Research”, ed. H.
Winick and S. Z. Doniach, 1980, Plenum Press, New York,
459.
References 20. M. N. Ingham and B. A. R. Vrebos, Adv. X-ray Anal.,
1994, 37, 717.
1. E. P. Bertin, “Principles and Practice of X-ray 21. H. J. Sanchez, Spectrochim. Acta Part B, 2001, 56, 2027.
Spectrometric Analysis”, 1975, Plenum Press, New York. 22. K. Baur, S. Brennan, B. Burrow, D. Werho, and P. Pianetta,
2. R. E. Van Grieken, “Hand Book of X-ray Spectrometry”, Spectrochim. Acta, Part B, 2001, 56, 2049.
2nd ed., 2002, Marcel Dekker, Inc., New York, 433 – 498. 23. N. Awaji et al., Jpn. J. Appl. Phys., 2000, 39, Pt. 2, No.
3. K. J. S. Sawhney and G. S. Lodha, Computers and 12A, L1252.
Geosciences, 1989, 15(7), 1115. 24. C. Streli, P. Wobrauschek, H. Aiginger, W. Ladisich, and
4. G. Misra, K. J. S. Sawhney, G. S. Lodha, V. K. Mittal, and R. Rieder, Adv. X-ray Anal., 1994, 37, 577.
H. S. Sahota, Appl. Radiat. Isot., 1992, 43(5), 609. 25. K. J. S. Sawhney, M. K. Tiwari, A. K Singh, and R. V.
5. S. Piorek, Nucl. Instr. Meth., 1994, A353, 528. Nandedkar, “Proc. 6th National Seminar on X-ray
6. V. Zaichick, N. Ovchjarenko, and S. Zaichick, Appl. Spectroscopy and Allied Areas”, ed. S. K. Joshi, B. D.
Radiat. Isot., 1999, 50(2), 283. Shrivastava, and A. P. Deshpande, 1998, Narosa, New
7. Y. Yoneda and T. Horiuchi, Rev. Sci. Instrum., 1971, 42, Delhi, 130.
1069. 26. M. K. Tiwari, B. Gowri Sankar, V. K. Raghuvanshi, R. V.
8. P. Wobrauschek and H. Aiginger, Anal. Chem., 1975, 47, Nandedkar, and K. J. S. Sawhney, Bull. Mater. Sci., 2002,
852. 25(5), 435.
9. R. Klockenkamper, “Total-Reflection X-ray Fluorescence 27. Sample absorption correction factor, A, in X-ray fluorescence
Analysis”, 1997, John Wiley and Sons, Inc., New York.
10. L. A. Currie, Anal. Chem., 1968, 40, 586. 1 – exp – µ(E0) µ(Ei)
sinφ + sinϕ M
can be evaluated from A = ,
11. L. A. Currie, “Lower limit of detection: definition and µ(E0) + µ(Ei) M
elaboration of a proposed position for radiological effluent sinφ sinϕ
and environmental measurements”, 1984, Report issued by where µ(E0) and µ(Ei) represent sample mass absorption
the US Nuclear Regulatory Commission, NUREG/CR- coefficients at incident energy E0 and characteristic
4007, 1 – 139. fluorescence line energy of the element of interest. φ and ϕ
12. L. A. Currie, Anal. Chim. Acta, 1999, 391, 103. are incident and exit angles; M represents the mass of the
13. L. A. Currie, J. Radioanal. Nucl. Chem., 2000, 245, 145. pellet in g/cm2. To determine the self-absorption correction
14. L. A. Currie, “Detection: international update, and some for analyte signal, normalization of A factor to that in case
emerging dilemmas involving calibration, the blank, and of blank sample is needed (A = 1 represents no absorption
multiple detection decisions”, 1997, Vol. 37, Chemometrics and A = 1.10 means 10% absorption of analyte intensity).
Intell. Lab Systems, 151 – 181.