Professional Documents
Culture Documents
Note 1: This Appendix C SNCS section is not used in the American markets.
Note 2: This Appendix C SNCS section applies to the XE-5000 only.
(8) When below screen appears, select Base-64 encoded X.509 and click Next.
(11) Click “Next” in the Certificate Export Wizard screen and click Finish.
(12) When the below screen appears, click OK.
(2) If the PC is connected to the internet via Proxy server, check the following values in the table.
Item Value Status
Hostname (Depend on your
environment)
Port (Depend on your
environment)
User name (Depend on your
(if required) environment)
Password (Depend on your
(if required) environment)
(1) Logoff the Windows and logon to Windows as “sysmex” (password “c” “9” “.” “0”). Insert the
Questra Smart Service (QSS) TotalAccess Ver.5.1 CD onto the CD-ROM drive. (QSS TOTAL
ACCESS SOFTWARE COMP. P/N: 063-3221-5)
(2) Run <CD-ROM drive letter>:\windows\setup.exe.
(4) When the Installation Key window appears, type “6D2A-08CB” and click Next.
(8) In the Enterprise Server setting window, follow the following steps.
Protocol: Select https.
Enterprise Server hostname: Type rs-ws.sncs-web.jp.
Enterprise Server port: Type 443.
Proxy server: Click the checkbox if it is necessary to connect to the internet via a proxy server.
Click Next..
(10) When the certification setting for Enterprise Server screen appears, type C:\cert\ca.cer. (or
use Browse… button.) Click Next.
(12) When the Enterprise Tunnel Server setting window appears, follow the following steps.
Protocol: Select https.
Enterprise Tunnel hostname: Type rs-rd.sncs-web.jp.
Enterprise Tunnel port: Type 443.
Proxy server: Click the checkbox if it is necessary to connect to the internet via a proxy server
on your environment.
Click Next.
-->
(3) When “Press any key to continue …” appears, press any key.
(4) If Questra is not installed properly, “Questa Service Agent does not work properly” will be
displayed and err.log will be generated in C:\Program Files\Questra\5.1\qsa\etc.
(5) After generating err.log, “Press any key to continue …” will be displayed. Press any key to
quit the test.bat.
(6) You can check where the problem is by observing err.log. You can correct the problem by
editing qsaconfig.xml or hacds.xml.
(16) Log off Windows, then log on with user name “sysmex”.
(17) Set back the account type of “XE-5000” user to “Limited” by selecting “Start“ Æ “Control
Panel” Æ “User Accounts.”
Note:
When executing Remote Access, ask to the customer to
start up the Remote_Start menu. (Start menu Æ Programs
Æ XE Æ Remote_Start.) This is to secure the customer’s
compliance for remote access.
NOTE
* If the screen loading does not stop and logon screen does not appear, check the
requirements for PC.
(4) In the logon screen, enter User Name and Password, and then click Login.
(6) Click Questra Service Agent, locate the target Device in Devices list and click it.
NOTE
* The icons of the Devices list show their connection status. Accessing is possible when the icon
is green.
- Green arrow: Online with Fast connect (LAN, broad-band)
- Green circle: Online
- Yellow circle: Online suspected Æ Refresh the screen to check the current status.
- Red circle: Offline
(12) When VNC window appears, the desktop of the device can be shared.
(13) When the remote access is finished, right-click WinVNC icon (as shown below) on the task
tray and select “Close VNC.” If “Close VNC” is not selected, the computer will be remote
accessed until the computer is restarted.
[OnLineQC]
ServerAddress = sncs.sysmex.co.jp
ClientAddress = IPU
From =E-mail address
To = E-mail address
Cc =
Bcc =
Subject = QC DATA
Folder = C:\Program Files\Sysmex\RmtMnt
Files = OnlineQC*.txt
AttachMents = ON
SendWindow=ON
[SystemSetting]
SendWindowTime=5
WaitTime=10
RetryNum=10