Optik
journal homepage: www.elsevier.de/ijleo
a r t i c l e i n f o a b s t r a c t
Article history: The first-order astigmatism-free conditions in crossed Czerny-Turner (C-T) spectrometer are derived. To
Received 8 March 2012 our best knowledge, there is no report about this kind of research. For the requirements of the optical
Accepted 18 July 2012 parameters of the relative aperture of 8 and the working wavelength ranging from 780 nm to 1020 nm,
according to the derived formula, the astigmatism-free crossed C-T spectrometer is built up with ZEMAX
software. The correspondent coma-free crossed C-T spectrometer is also built up. The two initial struc-
Keywords:
tures of crossed C-T optical systems are then optimized, and the performance merits are compared. It
Czerny-Turner
is shown that, the astigmatism-free crossed C-T spectrometer has a superior optical performance, with
Crossed-structure
Astigmatism-free
the RMS of spot diagram only 12–52% that in the correspondent coma-free crossed C-T spectrometer.
Coma-free This indicates that the astigmatism-free crossed C-T spectrometer possesses not only a better energy
concentration along the slit direction, which is beneficial to spectrometers requiring larger condenser
capacity, but also a smaller spot diameter along the direction perpendicular to the slit, which implies a
higher spectrum resolution achieved.
© 2012 Elsevier GmbH. All rights reserved.
0030-4026/$ – see front matter © 2012 Elsevier GmbH. All rights reserved.
http://dx.doi.org/10.1016/j.ijleo.2012.07.009
2540 Y. An et al. / Optik 124 (2013) 2539–2543
Fig. 2. Variations of the normalized linear dispersion as incident angle of grating for Fig. 3. The sagittal and tangential image distances of off-axis mirror.
two types of C-T structures.
Firstly, with the common variable factor d 2 /dˇ1 , we solve the 4. The comparison of structures and results of optimization
simultaneous equations of (7) and (18) to acquire the distance
from grating to focusing mirror Lgf . The partial derivatives can be 4.1. The comparison of two structures
obtained from Eqs. (1) and (2):
Considering the theoretical formulas in Section 2, under the
∂SS 2SS Lsc R1 sin 1
= (24) conditions of the grating of groove spacing d = 1/450 lp/mm, wave-
∂2 2Lsc (R1 sec 2 + R2 cos 1 ) − R1 R2 length range from 780 nm to 1020 nm, radius of 100 mm for
collimating and focusing mirrors, and the relative aperture of 8,
∂ST −2ST R2 (2Lsc sec 1 − R1 ) cos2 ˛ tan ˇ sec2 ˇ as listed in Table 1, we simulate the astigmatism-free crossed C-
= (25)
∂ˇ 2Lsc (R1 sec 2 + R2 cos 1 cos2 ˛/ cos2 ˇ) − R1 R2 cos2 ˛/ cos2 ˇ T structures as the incident angle of grating varying. It is shown
that in a range of the incident angles of grating from −10◦ to 0◦ ,
∂ST −2ST Lsc R1 sec 2 tan 2
a rational astigmatism-free crossed C-T structure can be formed.
= (26)
∂2 2Lsc (R1 sec 2 + R2 cos 1 cos2 ˛/ cos2 ˇ) − R1 R2 cos2 ˛/ cos2 ˇ
Table 1
Secondly, we use the zeroth-order divergent illumination con- The design parameters of the optical system.
dition of Eqs. (1)–(3) to determine the distance from entrance silt
Wavelength range (nm) 780–1020
to collimating mirror Lsc and the distance from focusing mirror to Central wavelength (nm) 898
image plane Lfi . Finally, we evaluate Eqs. (4) and (22) to determine Grating of groove spacing (lp/mm) 1/450
the tilted angle of image plane D . Since there is no constraint on Lcg , The radius of collimating mirror R1 (mm) 100
we set Lcg = Lsc × cos 1 for convenience, and then we can determine The radius of focusing mirror R2 (mm) 100
F-number 8
all the anastigmatic parameters of crossed C-T spectrometer.
2542 Y. An et al. / Optik 124 (2013) 2539–2543
Table 2
The initial parameters of astigmatism-free crossed structure and coma-free crossed
structure (unit mm).
i (◦ ) −3.86 −3.86
1 (◦ ) 8 8
2 (◦ ) 8 8
Lsc 43.5 50.0
Lcg 41.8 42.3
Lgf 39.6 42.3
Lfi 60.2 50.0
Fig. 6. The astigmatism blur height as a function of wavelength for two structures.
Fig. 5. The coma-free crossed C-T spectrometer. Fig. 7. Spot diagrams of astigmatism-free C-T structure at typical wavelengths.
Y. An et al. / Optik 124 (2013) 2539–2543 2543
Fig. 10. The ratio of RMS of spot diagram between two structures.
Acknowledgments
References
[1] N. Hagen, D.J. Brady, Coded aperture DUV spectrometer for standoff Raman
spectroscopy, Proc. SPIE 7319 (2009) 73190D–73191D.
[2] Q.S. Xue, S.R. Wang, F.T. Li, L.G. Yu, L.Q. Wang, Limb imaging spectrometer for
atmospheric remote sensing, Opt. Prec. Eng. 18 (4) (2010) 824–831.
Fig. 9. The RMS of spot diagram of two optical systems after optimization. [3] A.B. Shafer, L.R. Megill, L. Droppleman, Optimization of the Czerny-Turner spec-
trometer, J. Opt. Soc. Am. 54 (7) (1964) 879–887.
[4] A.B. Shafer, Correcting for astigmatism in the Czerny-Turner spectrometer and
in a comparison between Figs. 6 and 9, that the astigmatism blur spectrograph, Appl. Opt. 6 (1) (1967) 159–160.
height decreases dramatically by the optimization procedure. The [5] G.R. Rosendahl, Contributions to the optics of mirror systems and gratings
with oblique incidence. III. Some applications, J. Opt. Soc. Am. 52 (4) (1962)
maximum RMS value decreases from 300 m to 36 m in the wave- 412–415.
length range of 780–1020 nm for the astigmatism-free crossed C-T [6] Q.S. Xue, S.R. Wang, F.T. Li, Aberration-corrected Czerny-Turner imaging spec-
structure, and it is from 600 m to 60 m for the coma-free crossed trometer with a wide spectral region, Appl. Opt. 48 (1) (2009) 11–16.
[7] K.S. Lee, K.P. Thompson, J.P. Rolland, Broadband astigmatism-corrected Czerny-
C-T structure. The ratio of the RMS value of the spot diagram of Turner spectrometer, Opt. Express 18 (22) (2010) 23378–23384.
the astigmatism-free crossed C-T structure to that of the coma-free [8] M.W. McDowell, Design of Czerny-Turner spectrographs using divergent grat-
crossed C-T structure is shown in Fig. 10. It can be seen that the ing illumination, Opt. Acta 22 (1975) 473–475.
[9] D.R. Austin, T. Witting, I.A. Walmsley, Broadband astigmatism-free Czerny-
RMS of the astigmatism-free crossed C-T structure is much smaller
Turner imaging spectrometer using spherical mirrors, Appl. Opt. 48 (19) (2009)
than that of the coma-free crossed C-T structure, with the small- 3846–3853.
est percentage of 12% and the largest percentage of 52% in the [10] Z. Lin, S.F. Fan, Spectroscopic Instrumentology, Machine Industry Press, 1989.