Professional Documents
Culture Documents
Length Stabilization
Guide of operation
Version 1.1
August 23, 2021
2 System overview 5
3 Menu structure 9
3.1 Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 PID menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Generator menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4 Save/load menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.5 Info menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.6 Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4 Examples 13
4.1 Scanning the cavity length . . . . . . . . . . . . . . . . . . . . . . . . 13
4.2 Stabilization of the cavity length using PID11 . . . . . . . . . . . . . 14
4.3 Minimum intensity lock . . . . . . . . . . . . . . . . . . . . . . . . . 16
5 Fault diagnosis 17
5.1 Controller does not start . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.2 Screen is dark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1
1 Before you start
• These guidelines try to describe the 2-channel PID controller correctly and
thoroughly. However, technical developments and customer-specific modifica-
tions can result in discrepancies between the guidelines and the actual res-
onator.
• Only well-trained operators experienced in working with lasers should set up,
use, and maintain the PID controller and a connected resonator for second har-
monic generation. Improper use can result in serious danger for the operator
and others.
1
1 Before you start
We strongly recommend the operator to consult the responsible laser safety officer
during the design phase of the laser system and at before starting the operation.
• Turn of and disconnect all voltage sources from the cavity housing before
maintenance or repair.
• Avoid the usage of high voltages for the operation of the piezoelectric actuator,
whenever possible.
2
1.3 Additional general remarks
1. Select the value by a quick click and turn the rotary encoder. To change the
digit to be changed, briefly push the rotary encoder.
2. Select the value by pushing the corresponding button for more than 1 second.
A number field will appear.
3
2 System overview
Fig. 2.1 shows the front panel of the two-channel digital PID controller. The de-
vice provides two main inputs IN 1 and IN 2 to connect error signals generated
by differential photodetectors. The two main outputs OUT 1 and OUT 2 can
drive two independent piezoelectric actuators. Important: Never drive 50 Ω loads,
this will damage the internal amplifier. The auxiliary inputs AUX 1 and AUX 2
can be employed for optical power monitoring via two independent photodetectors.
AUX 3 is not required for the length stabilization of SHG cavities. The central
element of the front panel is a touch display as user interface. The pushable rotary
encoder in the lower right can be used to easily modify the PID parameters.
The technical specifications are listed in Tab. 2.1, the limit values of the PID pa-
rameters in Tab. 2.2.
Figure 2.1: Picture of the front panel of the two-channel digital PID controller.
5
2 System overview
6
Table 2.2: Limit values of the control parameters
7
3 Menu structure
• MinIntens
9
3 Menu structure
10
3.5 Info menu
behind Load, all current settings are overwritten by the settings stored in that slot.
3.6 Exit
After selecting Exit and accepting, the system will be shut down safely. Always
use this functionallity to protect the file system, as sudden removal of the supply
voltage can result in memory damage.
11
4 Examples
This chapter explains the use of the PID controller with a SHG cavity step by step
by in three examples. Fig. 4.1 schematically depicts a setup required for scanning
and stabilizing the length of the resonator making use of the minimum intensity lock
option.
Figure 4.1: Schematic wiring of the SHG setup. Note that either the circulating
pump light or SHG monitoring signal can be connected to the auxiliary
input channel. Before any signal is connected to the inputs of the PID
controller, please verify that the maximum of the absolute value must
not exceed 1 V.
13
4 Examples
14
4.2 Stabilization of the cavity length using PID11
Go to PID → PID 11 menu. We recommend the settings listed in Tab. 4.1 as start
parameters.
Figure 4.3: Schematic SHG signal (red) and error signal (green) during cavity length
scan before locking the cavity length. The depicted signal shapes are
ideal curves for a fully optimized resonator.
Then go back and switch PID 11 on. Now the settings can be modified to
achieve a stable and robust lock. In case the system does not lock, use a negative
gain. At the end of the optimization the gain is significantly smaller than 1, we
suggest to install an RF attenuator in between the differential photodetector and
the input of the PID controller and then increase the gain. Otherwise, the division
operation in the controller caused by the low gain can result in significant loss in
15
4 Examples
lock performance.
Figure 4.4: Schematic SHG signal (red) and error signal (green) during cavity length
scan before locking the cavity length. The threshold for the minimum
intensity lock is chosen such that the system does not lock on resonator
Eigenmoden with low SHG power.
16
5 Fault diagnosis
17