Professional Documents
Culture Documents
01 - AD&Vent - Introduction To Ventilation - JAF
01 - AD&Vent - Introduction To Ventilation - JAF
Introduction to Ventilation
VOLUME
A measure of the Vt delivered by the ventilator
to the patient
Reflects the volume of gas the patient breathes
Is usually expressed in ml for VT and in L
for minute volume
Ventilation
Parameters
PRESSURE
A measure of the impedance to gas flow rate
encountered in the ventilator breathing circuit and the
patient’s airways and lungs
Refers to the amount of pressure generated as a result
of airway resistance and lung-thorax compliance
Is expressed in cm H2O, mmHg, or Kilopascals (kPa)
(1mmHg = 1.36 cmH2O, 7.6mmHg = 1 kPa)
120
INSP
Flow SEC
L/min 1 2 3 4 5 6
120 EXP
Ventilation
Phases - Cycling
How the ventilator switches from inspiration to
expiration: the flow has been delivered to the volume
or pressure target
- Time cycled – the ventilator cycles to expiration
once a set time is reached
such as in pressure controlled ventilation
- Flow cycled– the ventilator cycles to expiration once
a set flow is reached
such as in pressure support
- Volume cycled - the ventilator cycles to expiration once
a set tidal volume has been delivered: this occurs in
volume controlled ventilation. If an inspiratory
pause is added, then the breath is both volume
and time cycled
Ventilation
Phases - Triggering
What causes the ventilator to cycle to inspiration.
Ventilators may be time triggered, pressure
triggered or flow triggered.
Time: The ventilator cycles at a set
frequency as determined by the
controlled rate.
Pressure: The ventilator senses the patient's
inspiratory effort by a decrease in
the baseline pressure.
Flow: Modern ventilators deliver a constant
flow around the circuit throughout the
respiratory cycle (flow-by). A
deflection in this flow by patient
inspiration, is monitored by the
ventilator and it delivers a breath.
This mechanism requires less work by
the patient than pressure triggering.
Ventilation
Flow Pattern
Constant V Constant Flow
Decelerating V
Decelerating Flow
T
Ventilation
Flow Pattern
Constant V Constant Flow
Flow continues at a
constant rate until the
set tidal volume is
T
delivered. Can cause
sustained high Paw
Ventilation
Flow Pattern
Decelerating V
Decelerating Flow
The flow pattern seen in
pressure targeted ventilation:
inspiration slows down as
alveolar pressure increases T
(there is a high initial flow).
Most intensivists and
respiratory therapists use this
pattern in pressure targeted
ventilation also, as it results in
a lower peak airway pressure
than constant and
accelerating flow, and better
distribution characteristics
Ventilation
Control
So, how does the ventilator know how much
flow to deliver?
Volume Controlled
(volume limited, volume targeted)
and Pressure Variable
Pressure Controlled
(pressure limited, pressure targeted)
and Volume Variable
Modes of Ventilation
Ventilation
Modes
Modes of ventilation describes the primary
method, how machine generates and
regulates the flow of gas into the lungs
Ventilation
Modes
Spontaneous Breathing
Controlled Modes: Volume Controlled Ventilation:
- Ventilator is active and patient is passive
Ventilator provides all breaths to the patient & does all work
necessary to maintain effective alveolar ventilation
1) VCV
2) PCV
Assisted Modes:
- Patient initiates and may or may not participate
the breath
Ventilator provides partial support while the patient must
do some or most of the work of breathing
1) SIMV - Synchronized Intermittent Mandatory Ventilation
2) Pressure Support Ventilation (PSV)
Ventilation
Types of Breath
What Causes the Ventilator to Cycle
Mandatory (controlled) - which is
determined by the
respiratory rate
Assisted (as in assist control,
synchronized intermittent
mandatory ventilation,
pressure support)
Spontaneous (no additional assistance
in inspiration)
Ventilation
Modes Pattern
There Are Only A Few Different Modes of Ventilation:
Tplat
Vmax
Tinsp
Constant Flow
- Parameters
Target Pressure (Pset) in cmH2O
Inspiratory time (Ti) in secs. or I:E Ratio
Respiratory Rate in breaths/min
PEEP Level in cmH2O
Modes of Ventilation
PCV vs. VCV
PCV VCV
Pressure
Pset + PEEP
PEEP
Flow
Volume
Vt
Ti Te Time
Modes of Ventilation
PCV vs. VCV
- Parameters
Tidal Volume PEEP
Respiratory rate (set ≠ meas) Pressure Support level
Inspiratory time Trigger level
Pressure Limit Inspiratory pause
Modes of Ventilation
Synchronized Intermittent Mandatory Vent
Variables
Mandatory Breath Spontaneous breaths
Trigger: Trigger:
- Patient: Flow or pressure
- Time
Cycling
- Patient: Flow or pressure - Under pressure support: flow
Cycling: Control Variable
- volume - Pressure
Control Variable
- Flow
Modes of Ventilation
Synchronized Intermittent Mandatory Vent
Deliver a minimum minute volume
Patient is allowed to breathe spontaneously between mandatory
breaths
Pressure Support can be added
Timing for spontaneous and mandatory breaths is defined through a
time window
Ventilator always deliver a certain amount of mandatory breaths
Modes of Ventilation
SIMV and Pressure Support
SIMV w/o PSV
SIMV w/PSV
PS Breath
Mandatory breath
Modes of Ventilation
Synchronized Intermittent Mandatory Vent
Advantage:
- Patient is able to breath spontaneously
- Can be used as weaning tools
Disadvantage:
- Excessive work of breathing
Resistance of the endotracheal tube and demand valves:
Solved by Pressure Support ventilation
Modes of Ventilation
Synchronized Intermittent Mandatory Vent
SIMV in ICU
• Known as a Weaning Mode, to overcome fighting ventilation
• Ensures minimum level of ventilation (mandatory breaths)
SIMV in anesthesia
• Ensures Minimum Minute Ventilation in spontaneous breathing
• With or without PSV
• Can be used during maintenance and off-set of anesthesia
PS + PEEP
PEEP
Trigger
M F
C
N B
I A
Q
P K
R
Volume Compensation
Compliance Compensation
• The effect of the gas being compressed in the dead space of the
breathing system (Circuit compliance and compression loss) is to
reduce the volume (TV) that is delivered to the patient.
• To calculate this effect it is necessary to measure the capacity or
compliance of the system (Cs). This is done during the pre-use check.
• The system is first pressured to 10cmH20 to check for leaks. Then the
system is raised a further 30cmH20, the volume needed to produce
the pressure rise is recorded.
• Volume in (ml) / Pressure = Dead space compliance (Cs)
• The Total compliance of the system and patient (Ct) is measured
after 2 or 3 stable breaths
• The TV can then be adjusted to compensate for the lost volume due
to compression within the breathing system.
• Set TV x (1+ (Cs / Ct –Cs)) = New TV
Volume Compensation
Compliance Compensation
Example
• System test measurement using 240ml to give 30cmH20
pressure rise
• Volume in (ml) / Pressure = Dead space compliance Cs
• 240ml / 30cmH20 = 8 = Cs
• Ventilator running on patient with set 500ml TV gave 20cmH20
peak Pressure
• 500ml / 20cmH20 = 25 = Ct
• Set TV x (1+(Cs / Ct –Cs)) = New TV
• 500ml x (1+(8 / 25-8)) = 735ml
• 735ml is the actual ventilator output to give 500ml at the patient
Volume Compensation
Fresh Gas Compensation
• Fresh gas flow adds to the delivered Tidal Volume TV
during the inspired period
• To compensate for this a reduction in the delivered
volume needs to be made
• Set TV – (Fresh gas flow (ml/Min) x Inspired Time
(sec)/60)
• Example FG 5 LPM, TV 600ml, 10 BPM, I:E 1:2.0
• 600 – (5000 x 2 / 60) = 434ml
• The ventilators effective TV is 434ml to allow for FG of
5LPM to give the patient 600ml