Calibration On KJ-8000 BILIRUBIN METER
Calibration On KJ-8000 BILIRUBIN METER
Calibration On KJ-8000 BILIRUBIN METER
In a sample case, the KJ-8000 test result is 05.6, while a reference test result is 04.1, customer will
choose that reference test result as a standard one. Here below is the procedure to decrease a test
result.
Procedure
1. Turn on meter, test forehead to get a test data 05.6
2. Use a pen tip to long press SETUP hole until test data to flash, now it is ready for set up
Long press SETUP hole Data flash mean ready for set up
3. Press RESET key to adjust data from 1-0
4. Use pen tip to long press SETUP hole until flash stops, adjusted data 04.1 saved
5. Turn off meter, turn it on again and test forehead, new result is 04.0, calibration succeed