Introduction ................................................................ 5
1
over-voltages often associated with local AC mains connec-
tions. The user should assume all measurement, control, and
data I/O connections are for connection to Category I sources
unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present.
Lethal voltage may be present on cable connector jacks or
test fixtures. The American National Standards Institute
(ANSI) states that a shock hazard exists when voltage levels
greater than 30V RMS, 42.4V peak, or 60VDC are present.
A good safety practice is to expect that hazardous voltage
is present in any unknown circuit before measuring.
Users of this product must be protected from electric shock
at all times. The responsible body must ensure that users are
prevented access and/or insulated from every connection
point. In some cases, connections must be exposed to poten-
tial human contact. Product users in these circumstances
must be trained to protect themselves from the risk of electric
shock. If the circuit is capable of operating at or above 1000
volts, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power
circuits. They are intended to be used with impedance limit-
ed sources. NEVER connect switching cards directly to AC
mains. When connecting sources to switching cards, install
protective devices to limit fault current and voltage to the
card.
Before operating an instrument, make sure the line cord is
connected to a properly grounded power receptacle. Inspect
the connecting cables, test leads, and jumpers for possible
wear, cracks, or breaks before each use.
When installing equipment where access to the main power
cord is restricted, such as rack mounting, a separate main in-
put power disconnect device must be provided, in close prox-
imity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or
any other instruments while power is applied to the circuit
under test. ALWAYS remove power from the entire test sys-
tem and discharge any capacitors before: connecting or dis-
connecting cables or jumpers, installing or removing
2
switching cards, or making internal changes, such as install-
ing or removing jumpers.
Do not touch any object that could provide a current path to
the common side of the circuit under test or power line
(earth) ground. Always make measurements with dry hands
while standing on a dry, insulated surface capable of with-
standing the voltage being measured.
The instrument and accessories must be used in accordance
with its specifications and operating instructions or the safety
of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments
and accessories, as defined in the specifications and operat-
ing information, and as shown on the instrument or test fix-
ture panels, or switching card.
When fuses are used in a product, replace with same type and
rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections
for measuring circuits, NOT as safety earth ground connec-
tions.
If you are using a test fixture, keep the lid closed while power
is applied to the device under test. Safe operation requires the
use of a lid interlock.
3
sociated information very carefully before performing the
indicated procedure.
The CAUTION heading in a manual explains hazards that
could damage the instrument. Such damage may invalidate
the warranty.
Instrumentation and accessories shall not be connected to
humans.
Before performing any maintenance, disconnect the line
cord and all test cables.
To maintain protection from electric shock and fire, replace-
ment components in mains circuits, including the power
transformer, test leads, and input jacks, must be purchased
from Keithley Instruments. Standard fuses, with applicable
national safety approvals, may be used if the rating and type
are the same. Other components that are not safety related
may be purchased from other suppliers as long as they are
equivalent to the original component. (Note that selected
parts should be purchased only through Keithley Instruments
to maintain accuracy and functionality of the product.) If you
are unsure about the applicability of a replacement compo-
nent, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based
cleaner. Clean the exterior of the instrument only. Do not ap-
ply cleaner directly to the instrument or allow liquids to enter
or spill on the instrument. Products that consist of a circuit
board with no case or chassis (e.g., data acquisition board for
installation into a computer) should never require cleaning if
handled according to instructions. If the board becomes con-
taminated and operation is affected, the board should be re-
turned to the factory for proper cleaning/servicing.
4
Introduction
This quick reference guide includes summary informa-
tion on front panel and IEEE-488 operation for the
Model 7001 Switch System. For detailed information,
consult the Model 7001 Instruction Manual.
5
Operation Summary
Channel List and Scan List
A Channel List specifies the channels to be closed or
opened, and a Scan List specifies the channels to be
scanned. Each channel in the list is made up of either a
two integer CHANNEL number (for non-matrix type
cards) or a three integer CHANNEL number (for matrix
type cards). Examples:
6
with a hyphen (-). The M1 entry represents a channel
pattern that is stored in memory (see STORE and
RECALL).
7
Note that the OPEN ALL key will open all channels,
including those not specified in the Channel List.
:OPEN ALL
Scanning Channels
Channels are scanned in the order that they are pre-
sented in the Scan List.
8
The scan can be terminated at any time by pressing the
OPEN ALL key. All closed channels will open and the
Model 7001 will return to the idle state (ARM indicator
off).
IEEE-488 operation
Over the bus, a scan is configured using the :TRIGger
subsystem commands.
:INITiate
:ABORt
All closed channels will open and the Model 7001 will
return to the idle state.
9
To recall a channel pattern from memory:
:MEMory:SAVe M36
:MEMory:RECall M36
10
Menu Structures
The desired menu is displayed by pressing the appropri-
ate key; MENU, SCAN CONFIGURATION or CARD
CONFIGURATION. Once in a menu structure, place
the cursor ( and keys) on the desired selection and
press ENTER. Use the keypad to enter parameters. Use
the EXIT key to back out of the menu structure and dis-
able the menu.
11
Table 1: MAIN MENU structure (Cont.)
12
Table 2: CONFIGURE SCAN menu structure (Cont.)
13
Table 2: CONFIGURE SCAN menu structure (Cont.)
14
Table 3: CARD CONFIG MENU structure
15
Default Conditions
Note: “No change” indicates that RESET does not affect the currently set condi-
tion.
16
Table 5: *RST and :SYSTem:PRESet
:SYSTem:PRESet
Command name *RST value value
:INITiate
:CONTinuous OFF OFF
:ARM
:SEQuence[1]
:LAYer[1]
:COUNt 1 1
:SOURce IMMediate IMMediate
:TCONfigure
:DIRection ACCeptor ACCeptor
:ASYNchronous
:ILINe 1 1
:OLINe 2 2
:LAYer2
:COUNt 1 INFinite
:DELay 0 0
:SOURce IMMediate IMMediate
:TIMer 0.001 0.001
:TCONfigure
:DIRection ACCeptor ACCeptor
:ASYNchronous
:ILINe 1 1
:OLINe 2 2
:TRIGger
:SEQuence1
:COUNt 1 Scan list length
:AUTO OFF ON
:DELay 0 0
:SOURce IMMediate MANual
:TIMer 0.001 0.001
:TCONfigure
:PROTocol ASYNchronous ASYNchronous
:DIRection ACCeptor SOURce
:ASYNchronous
:ILINe 1 1
:OLINe 2 2
:SSYNchronous
:LINE 1 1
17
Table 5: *RST and :SYSTem:PRESet (Cont.)
:SYSTem:PRESet
Command name *RST value value
[:ROUTe]
:CONFigure
:SLOT[1]
:STIMe 0 0
:SLOT2
:STIMe 0 0
:CPAir OFF OFF
:SCHannel OFF OFF
:DISPlay
:SMESsage OFF OFF
18
Error and Status Messages
19
Table 6: Error and status messages (Cont.)
20
Table 6: Error and status messages (Cont.)
EE = Error event
SE = Status event
21
PON = Power On
URQ = User Request * ESR ? PON URQ CME EXE DDE QYE OPC Standard Event
CME = Command Error (B15-B8) (B7) (B6) (B5) (B4) (B3) (B2) (B1) (B0) Status Register
EXE = Execution Error
DDE = Device - Dependent Error
QYE = Query Error
OPC = Operation Complete
&
&
&
OR &
&
22
To Event Summary &
Operation Event
Idle Arm Trig Set Condition Register
(B15 - B11) (B10) (B9) (B8) (B7) (B6) (B5) (B4) (B3) (B2) (B1) (B0)
Operation Event
Idle Arm Trig Set Status Register
(B15 - B11) (B10) (B9) (B8) (B7) (B6) (B5) (B4) (B3) (B2) (B1) (B0)
23
&
&
&
&
OR &
24
&
&
25
Table 7: IEEE-488.2 common commands and queries
(Cont.)
26
SCPI Command Subsystems
Command Description
:DISPlay
[ :WINDow[1]] Path to locate message to top portion of
display
:TEXT Path to control user text messages:
:DATA <a> Define ASCII message “a” (up to 20
characters).
:STATe <b> Enable (1 or ON) or disable (0 or
OFF) message.
:STATe? Read state.
:WINDow2 Path to locate message to bottom portion
of display:
:TEXT Path to control user text messages:
:DATA <a> Define ASCII message “a” (up to 32
characters).
:STATe <b> Enable (1 or ON) or disable (0 or
OFF) message.
:STATe? Read state.
:SMESsage <b> Enable (1 or ON) or disable (0 or OFF)
status messages.
:SMESsage? Read state of status message display.
Command Description
:OUTPut
:TTLn Path and command to specify output
line “n” (1-4):
:LSENse AHIGh| Set polarity; active high (AHIG) or
ALOW active low (ALOW).
:LSENse? Read polarity.
27
Table 10: ROUTe command summary
Command Description
[:ROUTe]
:CLOSe <list> Close specified channels:
:STATe? Read closed channels.
:CLOSe? <list> Read state of specified channels
(1=closed, 0=open).
:OPEN <list>|ALL Open specified (or all) channels.
:OPEN? <list> Read state of specified channels
(1=open, 0=closed).
:SCAN <list> Specify scan list:
:POINts? Read number of channels in
scan list.
:SCAN? Read scan list.
:FCHannels <list> Specifies channels that cannot
be closed (forbidden).
:FCHannels? Read forbidden channels.
:CONFigure: Configuration command path:
:SCHannel <b> Enable (1 or ON) or disable (0
or OFF) Single Channel.
:SCHannel? Read Single Channel mode.
:CPAir <b> Enable (1 or ON) or disable (0
or OFF) Channel Pairing.
:CPAir? Read Channel Pairing.
:SLOT1 Path to configure CARD 1:
:CTYPE <type> Specify card type.
:CTYPE? Read card type.
:POLE <n> Select pole mode (1, 2 or 4).
:POLE? Read pole mode.
:STIMe <num> Set relay settling time (0 to
99999.999 sec).
:STIMe? Read relay settling time.
28
Table 10: ROUTe command summary (Cont.)
Command Description
[:ROUTe]
:CONFigure:
:SLOT2 Path to configure CARD 2:
:CTYPE <type> Specify card type.
:CTYPE? Read card type.
:POLE <n> Select pole mode (1, 2 or 4).
:POLE? Read pole mode.
:STIMe <num> Set relay settling time (0 to
99999.999 sec).
:STIMe? Read relay settling time.
:MEMory Path to program memory:
:SAVE M<num> Saves channel pattern in mem-
ory (1-100).
:RECall M<num> Recalls channel pattern from
memory (1-100).
Command Description
:SENSe[1]
:TTL1 Command path for input port.
:DATA? Read digital input port.
29
Table 12: SOURce command summary
Command Description
Command Description
:STATus
:OPERation Path to control operation event
registers:
[:EVENt]? Read status register.
:CONDition? Read condition register.
:PTRansition <n> Program positive transition reg-
ister (0 to 1122).
:PTRansition? Read positive transition register.
:NTRansition <n> Program negative transition reg-
ister (0 to 1122).
:NTRansition? Read negative transition register.
:ENABle <n> Program enable register (0 to
1122).
:ENABle? Read enable register.
:ARM Path to control arm event regis-
ters:
30
Table 13: STATus command summary (Cont.)
Command Description
:STATus
:OPERation
:ARM
[:EVENt]? Read status register.
:CONDition? Read condition register.
:PTRansition <n> Program positive transition
register (0 or 2).
:PTRansition? Read positive transition regis-
ter.
:NTRansition <n> Program negative transition
register (0 or 2).
:NTRansition? Read negative transition regis-
ter.
:ENABle <n> Program enable register (0 or
2).
:ENABle? Read enable register.
:SEQuence Path to control the sequence
event registers:
[:EVENt]? Read status register.
:CONDition? Read condition register.
:PTRansition <n> Program positive transition
register (0 to 6).
:PTRansition? Read positive transition reg-
ister.
:NTRansition <n> Program negative transition
register (0 to 6).
:NTRansition? Read negative transition
register.
:ENABle <n> Program enable register (0
to 6).
:ENABle? Read enable register.
:TRIGger Path to control trigger event
registers:
[:EVENt]? Read status register.
:CONDition? Read condition register.
:PTRansition <n> Program positive transition
register (0 or 2).
:PTRansition? Read positive transition reg-
ister.
31
Table 13: STATus command summary (Cont.)
Command Description
:STATus
:OPERation
:ARM
:TRIGger
:NTRansition <n> Program negative transition
register (0 or 2).
:NTRansition? Read negative transition
register.
:ENABle <n> Program enable register (0
or 2).
:ENABle? Read enable register.
:PRESet Return status registers to default
states.
:QUEue Path to access error queue:
[:NEXT]? Read most recent error message.
:ENABle <list> Specify error and status mes-
sages for queue.
Command Description
:SYSTem
:ERRor? Read Error/Status Queue.
:POSetup RST| Select Power-on setup.
PRESet|SAV0|
SAV1|SAV2|SAV3|
SAV4|SAV5|SAV6|
SAV7|SAV8|SAV9
:POSetup? Read power-on setup.
:PRESet Return instrument to default condi-
tions.
32
Table 14: SYSTem command summary
Command Description
Command Description
33
Table 15: Trigger command summary (Cont.)
Command Description
34
Table 15: Trigger command summary (Cont.)
Command Description
:ARM[:SEQuence[1]]
:LAYer2 Path to program Arm Layer 2
(scan layer) of scan:
:IMMediate Loop around control source.
:COUNt <n>| Program number of scans (1
INFinite to 9999, or INFinite).
:COUNt? Read scan count.
:DELay <num> Program delay (0 to
99999.999 sec)
:DELay? Read delay.
:SOURce HOLD| Select Arm Layer 2 control
IMMediate| source.
MANual|BUS|
TLINk|EXTernal
|TIMer
:SOURce? Read source.
:SIGNal Bypass scan control source
one time only.
:TIMer <num> Set timer interval (0 to
99999.999 sec).
:TIMer? Read timer interval.
:TCONfigure Path to configure Triggers:
:DIRection Enable (SOUR) or disable
SOURce| (ACC) Bypass.
ACCeptor
:DIRection? Read direction.
:ASYNchronous Path to configure asyn-
chronous mode:
:ILINe <n> Select input line (1 to 6).
:ILINe? Read input line.
:OLINe <n> Select output line (1 to
6).
:OLINe? Read output line.
35
Table 15: Trigger command summary (Cont.)
Command Description
36
Table 15: Trigger command summary (Cont.)
Command Description
:TRIGger
:TCONFigure
:ASYNchronous
:OLINe <n> Select output line (1 to 6).
:OLINe? Read output line.
:SSYNchronous Path to configure semi-
synchronous Trigger Link
mode:
:LINE <n> Select trigger line (1 to 6).
:LINE? Read trigger line.
37
38
Specifications are subject to change without notice.
All Keithley trademarks and trade names are the property of Keithley
Instruments, Inc. All other trademarks and trade names are the
property of their respective companies.