Professional Documents
Culture Documents
Agilent Enviro Analysis Ebook v2
Agilent Enviro Analysis Ebook v2
Advances in
Environmental Analysis
Table of Contents
Volatile Organic Compounds Analysis in Soils and Sediments Using the Agilent 8697
Headspace Sampler
This application note describes volatile organic compounds analysis in soil and sediments using the Agilent 8697
headspace sampler, 8860 GC, and 5977B GC/MSD system.
An Alternate Testing Protocol for EPA 1613B using Agilent Triple Quadrupole GC/MS
A study for the determination of 2,3,7,8-substituted tetra- through octa-chlorinated dibenzo-p-dioxins and
dibenzofurans.
a collaboration between
Application Note
Environmental
Author Abstract
Jie Zhang
This application note describes volatile organic compounds analysis in soil and
Agilent Technologies
sediments using the Agilent 8697 headspace sampler, 8860 GC, and 5977B
(Shanghai) Co., Ltd, China
GC/MSD system. The system performance in terms of repeatability, linearity, limit
of detection, limit of quantitation, and method recovery rate were evaluated with
good results. The area repeatability was in the range of 1.0 to 4.3%; the LOD and
LOQ in the quartz sand blank was from 0.51 to 1.21 µg/kg and from 1.7 to 4.1 µg/kg,
respectively. The recovery rate for the soil samples at spiked concentrations of
50 and 125 µg/kg was 78.2 to 125.9% and 71.7 to 108.7%. The linearity across
the tested concentration range is excellent, with the R2 of all components better
than 0.996. The test results met or exceeded the requirements of Chinese standard
HJ 642-2013.
Introduction connects to Agilent smart GCs, including Matrix modifier: 500 mL of organic‑free
the 8860, 8890, and Intuvo 9000 GC water was adjusted to pH ≤2 with
Volatile organic compounds (VOCs) with integrated communication. Users phosphoric acid and saturated with
are widely used in many industries as can access the browser interface of the analytical-grade sodium chloride.
solvents or chemical intermediates. smart GC or workstation to configure Working solution: The VOCs and
VOCs leaked or emitted into industry or set the headspace parameters. surrogate stock solutions were mixed
waste pollute the soil and sediments Compared to previous headspace and diluted by methanol to 10 mg/L
that the wastewater flows through. products, the 8697 headspace sampler as the working solution. The IS stock
Considering that many VOCs have has more automated diagnostic solution was diluted to 50 mg/L by
adverse environmental effects and soil functions, which can be executed methanol for later use.
remediation is costly, decisions regarding from the browser interface of the GC.
the significance of contamination and With the help of this user-initiated Calibration standards and soil
cleanup must be based on accurate diagnosis process, it is easy to know sample preparation
VOC measurement. In China, the soil whether the 8697 headspace sampler The quartz sand was weighed at 2 g.
pollution prevention law was passed by is in good health and ready for sample The sand and 5 mL matrix modifier
the National People’s Congress on 31 analysis. Clear guidance on the browser were added to a 20 mL headspace
August 2018 as the first comprehensive interface, in text or in image formats, vial. Aliquots of VOCs/surrogates and
framework law for addressing soil makes the headspace maintenance or IS working solutions were spiked into
pollution, creating new obligations troubleshooting more straightforward. the modifier solution quickly, and then
and potential liabilities for land- Besides the smart maintenance and the vials were sealed immediately and
use‑rights holders. The law requires the diagnosis features, the 8697 headspace shaken to well mix the standard. The
land‑use-rights holder to investigate sampler provides 48-vial capacity and final calibration standards are prepared
soil conditions upon the government’s 12-position air bath vial oven for precise approximately at 4, 10, 20, 50, and
identification of soil pollution risk. temperature control of every sample 100 µg/L, and the internal standards
Headspace and purge-and-trap methods throughout its equilibration time, which were spiked with the concentration of
are used for VOCs contamination meet the routine sample throughput 50 µg/L.
measurement in soil and sediments, requirement in most commercial
The soil sample was weighed at 2 g in
with the method used dependent on testing labs.
a 20 mL sample vial. Matrix modifier of
the sample concentration. The static In this application note, VOCs in soil were 5 mL was added then spiked with IS to
headspace method features easy analyzed on the 8697 headspace/8860 50 µg/L. The vial was then quickly sealed
operation and good repeatability. It GC/5977B GC/MSD platform by and shaken to mix well.
allows use of an autosampler and following Chinese standard HJ 642‑2013.
minimizes carryover. The linearity, repeatability, LOD, and Instrumentation and analytical
LOQ for the targeted 36 VOCs were conditions
Chinese standard HJ 642-2013 is an
environmental protection standard evaluated to show the system’s excellent An 8860 GC was equipped with a
for determination of VOCs in soil and performance for VOCs analysis. split/splitless inlet. The 8697 headspace
sediments by headspace GC/MS. It sampler was used to extract and transfer
gives guidelines for handling of VOCs Experiment the VOCs in the sample to the GC. The
in soil and sediments when using a gas in the headspace of sample vials
headspace method. Chemicals and standards went through the 8697 transfer line and
Stock solution: A mixture of entered the split/splitless inlet before
The 8697 headspace sampler is a separation on an analytical column.
36 volatile organic compounds
platform that can introduce VOCs in the The sample was then analyzed using a
(VOCs) in methanol at 1,000 mg/L;
soil and sediments to a GC or GC/MSD 5977B GC/MSD configured with an inert
internal calibration standard of
platform for analysis as required by extraction ion source.
fluorobenzene, chlorobenzene-d5, and
HJ 642-2013 standard. The 8697
1,2-dichlorobenzene-d4 in methanol
headspace sampler has smart features
at 2,000 mg/L; and 2,000 mg/L
that are developed to improve user
surrogate standard of toluene-d8 and
experience and expand its diagnostic
4-bromofluorobenzene in methanol
capability. The 8697 headspace sampler
were used.
2
Agilent MassHunter Acquisition software Table 1. Analytical conditions of the Agilent 8697 headspace sampler/8860 GC/5977B GC/MSD system.
version 10.0 was used for data collection.
Parameters Setpoints
MassHunter Qualitative Analysis version
Inlet Temperature 250 °C
B.08.00 and MassHunter Quantitative
Liner 4 mm id Ultra Inert inlet liner, split (p/n 5190-2295), glass wool removed
Analysis version B.08.00 were used for
Column Flow Constant flow, 1.2 mL/min
peak identification and quantitation.
Split Ratio 10:1
The analytical conditions are listed in 40 °C (2 min),
Table 1. Oven Program 8 °C/min to 90 °C (4 min),
then 6 °C/min to 200 °C (10 min)
Results and discussion Column Agilent J&W DB-624 GC column, 60 m × 0.25 mm, 1.40 µm (p/n 122-1364)
Target Mass Rel to Mass Lower Limit % Upper Limit % Rel. Abn % Raw Abn Pass/Fail
95 95 100 100 100 96,889 Pass
3
vials of 20 µg/L calibrants were run for linearity, with correlation coefficients
The data of the calibration standards
repeatability assessment. The response (R2) of regression formula better than
were acquired in selected ion monitoring
RSD% of the 38 VOCs (i.e., 36 analytes 0.996. The calibration curves of four
(SIM) mode. The total ion chromatogram
and 2 surrogates) are in the range of 1.0 representative compounds eluting at
(TIC) in Figure 1 showed the separation
to 4.3% (Figure 2), which demonstrated the early, middle, and late part of the
and detection for the 20 µg/L standard.
excellent sampling and detection chromatogram are shown in Figure 3.
The instrument repeatability was verified precision. Instrument linearity were The linearity results were achieved on
based on the analyte absolute response. evaluated from 4 to 100 µg/L in 5 mL of the 6 mm MSD drawout plate. When
System linearity performance was matrix modifier with 2 g of quartz sand using a 3 mm drawout plate, there were
verified based on analytes’ quantitated as control matrix, corresponding to 10 to several compounds with R2 less than
concentrations instead of absolute 250 µg/kg analytes in real samples. All 0.99, which did not meet the HJ standard
area response, because HJ 642-2013 targeted components showed good requirement.
uses ISTD method for quantitation. Six
×105 1. Vinyl chloride 16. 1,1,2-Trichloroethane 27. Bromoform 30. 1,3,5-Trimethylbenzene
3.8 2. 1,1-Dichloroethene 17. Tetrachloroethylene 28. 1,2,3-Trichloropropane 31. 1,2,4-Trimethylbenzene
3.6 3. Methylene chloride 18. Dibromochloromethane 29. 1,1,2,2,-Tetrachloroethane 32. 1,3-Dichlorobenzene
3.4 4. trans-1,2-Dichloroethene 19. 1,2-Dibromoethane 33. 1,4-Dichlorobenzene
3.2 5. cis-1,2-Dichloroethene 20. Chlorobenzene 34. 1,2-Dichlorobenzene
6. 1,2-Dichloroethane 21. Ethylbenzene 23,24 35. 1,2,4-Trichlorobenzene
3.0
2.8 7. Chloroform 22. 1,1,1,2-Tetrachloroethane 36. Hexachlorobutadiene
20 22
2.6 8. 1,1,1-Trichloroethane 23,24. m-xylene, p-xylene
9. Carbon tetrachloride 25,26. Styrene, o-xylene 21
2.4
10. 1,2-Dichloroethane
2.2
11. Benzene
Counts
Response precision
5.0
4.5
4.0
3.5
Area RSD%
3.0
2.5
2.0
1.5
1.0
0.5
0
Compound name
Figure 2. Area precision of six vials of 20 µg/L calibration standards in 5 mL matrix modifier.
4
×10–1 ×10–1
A B
2.2 2.2
2.0 Vinyl chloride 1,1,2-Trichloroethane
2.0
1.8 y = 0.213223 * x + 7.451340E-004 1.8 y = 0.218556 * x + 6.038936E-004
1.6 R2 = 0.99952361 R2 = 0.99955801
Relative responses
Relative responses
1.6
1.4 1.4
1.2 1.2
1.0 1.0
0.8 0.8
0.6 0.6
0.4 0.4
0.2 0.2
0 0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
Relative concentration Relative concentration
×10–1
C D
2.4 1,3,5-Trimethylbenzene 6.0 Hexachlorobutadiene
2.2 y = 2.308429 * x – 0.030655 5.5 y = 0.610358 * x + 0.008732
2 R2 = 0.99657925 5.0 R2 = 0.99702957
Relative responses
Relative responses
1.8 4.5
1.6 4.0
1.4 3.5
1.2 3.0
1 2.5
0.8 2.0
0.6 1.5
0.4 1.0
0.2 0.5
0 0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
Relative concentration Relative concentration
Figure 3. Calibration curves for representative compounds: (A) Vinyl chloride with R2 = 0.9995; (B) 1,1,2-trichloroethane with R2 = 0.9995; (C) 1,3,5-trimethylbenzene
with R2 = 0.9965; (D) hexachlorobutadiene with R2 = 0.9970.
The method recovery performance was This met the standard requirements LOD and LOQ in real soil or sediments
tested on 2 g of real soil samples spiked on surrogate recovery rate in real sample. The LOD and LOQ for VOCs in
with 20 and 50 µL 10 mg/L calibration sample. The recovery rates of 36 target quartz sand was between 0.51 to 1.21
standards (corresponding to 50 and VOCs were between 78.2 to 125.9% at µg/kg and 1.7 to 4.1 µg/kg, good enough
125 µg/kg VOCs in soil sample). The 50 µg/kg and between 71.7 to 108.7% at to detect the targeted VOC compounds
soil sample without spiking was tested 125 µg/kg (Figure 5). The recovery rates at the single‑digit µg/kg level as specified
as blank, and the difference between demonstrated equivalent performance by the HJ 642-2013 method.
the spiked sample and the blank was to the reference results in HJ 642-2013
used for recovery rate calculation. The standard.
chromatograms of one soil blank and The instrument detection limits (IDL) for
two spiked soil samples are shown 36 targeted VOCs and two surrogates
in Figure 4. The surrogate’s recovery were calculated based on quantitation
rate was: precision of 4 µg/L standards in
– 103.5% for toluene-d8 and 109.7% for eight vials. The translated method LOD
4-bromofluorobenzene at 50 µg/kg and LOQ (in the unit of µg/kg) in blank
– 93.6% for toluene-d8, and 95.9% for quartz sand based on IDL were shown in
4-bromofluorobenzene at 125 µg/kg Table 3 (Appendix) as a reference of the
5
×104
1.5
Counts
Soil blank
1.0
0.5
0
×105
2.0
1.5
Counts
1.5
1.0
0.5
0
5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34
Acquisition time (min)
110%
85%
60%
Compound name
6
Conclusion Reference
This application note demonstrated 1. Chinese Environmental Protection
that the 8697 headspace sampler is an Industry Standard HJ 642-2013:
excellent means of introducing VOCs Soil and Sediment–Determination
trapped in soils and sediments to GC of Volatile Organic Compounds–
for identification and quantitation. The Headspace-Gas Chromatography/
combination of the 8697 headspace Mass Method. Chinese Ministry of
sampler with the 8860 GC and 5977B Environmental Protection (published
GC/MSD system delivered good on 21 January 2013).
repeatability, which was demonstrated
in response precision from 1.0 to 4.3%
for 38 VOCs. The calibration curves
of targeted compounds in the tested
concentration range (4 to 100 µg/L)
showed satisfactory linearity, with all
correlation coefficient better than 0.996.
The LOD and LOQ achieved on the
described system showed equivalent
performance as demonstrated in the
HJ 642-2013 standard. The recovery
rate of spiked soil samples at 50 and
125 μg/kg was 78.2 to 125.9% and 71.7
to 108.7%, demonstrating that the 8697
headspace sampler coupled with the
8860 smart GC can provide reliable and
accurate VOCs analysis in real-world soil
and sediments samples.
7
Appendix
Table 3. Instrument linearity, LOD, area precision, and recovery rate at applied operation conditions.
Recovery Rate
Response LOD LOQ
Name RT/min CF Formula CF R2 RSD% (μg/kg) (μg/kg) 50 μg/kg 125 μg/kg
Vinyl chloride 4.997 y = 0.213223 * x + 7.451340E-004 0.9995 1.6 1.22 4.05 112.4% 100.1%
1,1-Dichloroethene 7.271 y = 0.280056 * x + 0.002079 0.9994 1.9 0.62 2.06 108.5% 97.7%
Methylene chloride 8.004 y = 0.251349 * x + 0.002364 0.9992 1.1 0.73 2.43 107.4% 98.9%
trans-1,2-Dichloroethene 8.435 y = 0.287348 * x + 0.002651 0.9991 1.8 0.83 2.77 105.6% 94.8%
1,2-Dichloroethane 10.040 y = 0.297696 * x + 0.002091 0.9994 1.8 0.75 2.48 103.1% 94.6%
Chloroform 10.566 y = 0.369739 * x + 0.005035 0.9992 1.2 0.60 1.99 107.5% 97.5%
1,1,1-Trichloroethane 10.999 y = 0.329690 * x + 0.002463 0.9995 1.5 0.62 2.06 107.6% 96.0%
Carbon tetrachloride 11.336 y = 0.318289 * x + 0.002059 0.9996 1.7 0.57 1.89 105.1% 93.3%
1,2-Dichloroethane 11.746 y = 0.193792 * x + 0.001507 0.9993 1.2 0.74 2.45 103.4% 95.4%
Benzene 11.750 y = 1.051625 * x + 0.006656 0.9996 1.8 0.72 2.41 104.7% 95.2%
1,2-Dichloropropane 13.689 y = 0.291754 * x + 8.914385E-004 0.9998 1.4 0.65 2.16 114.3% 103.7%
Bromodichloromethane 14.281 y = 0.338244 * x + 0.001356 0.9997 1.8 0.65 2.15 110.6% 100.4%
Toluene 16.069 y = 1.164353 * x + 0.001734 0.9997 3.7 0.85 2.85 104.2% 93.9%
1,1,2-Trichloroethane 17.217 y = 0.218556 * x + 6.038936E-004 0.9996 1.6 0.67 2.24 104.6% 95.5%
Tetrachloroethylene 17.652 y = 0.384694 * x + 0.002042 0.9997 2.5 0.68 2.26 105.0% 91.7%
Dibromochloromethane 18.272 y = 0.287628 * x + 6.739824E-004 0.9992 2.4 0.71 2.38 100.8% 92.1%
1,2-Dibromoethane 18.622 y = 0.236576 * x + 2.992586E-004 0.9991 2.3 0.82 2.74 99.2% 92.1%
Chlorobenzene 19.884 y = 1.062315 * x + 0.003659 0.9997 2.4 0.68 2.28 99.7% 89.8%
m,p-Xylene 20.442 y = 2.580795 * x – 0.023563 0.9983 3.5 0.70 2.34 106.4% 97.0%
Styrene 21.510 y = 1.204945 * x – 0.009297 0.9989 3.6 0.69 2.29 108.2% 98.1%
o-Xylene 21.537 y = 1.637392 * x – 0.019908 0.9974 3.5 0.79 2.64 85.9% 86.2%
Bromoform 22.055 y = 0.350684 * x + 1.159518E-005 0.9990 2.4 0.78 2.61 111.5% 96.9%
4-Bromofluorobenzene 22.902 y = 0.892720 * x + 0.002945 0.9998 3.2 0.75 2.49 109.7% 95.9%
1,2,3-Trichloropropane 23.249 y = 0.801680 * x + 0.002143 0.9994 1.8 0.75 2.48 113.2% 96.4%
1,1,2,2,-Tetrachloroethane 23.397 y = 0.617940 * x + 0.001429 0.9995 1.7 0.74 2.48 116.5% 99.1%
1,2,4-Trimethylbenzene 25.046 y = 2.284967 * x – 0.033058 0.9963 4.3 0.82 2.74 94.7% 84.4%
1,3-Dichlorobenzene 25.823 y = 1.737283 * x + 0.007524 0.9997 3.1 0.68 2.26 100.4% 88.9%
1,4-Dichlorobenzene 26.054 y = 1.710684 * x + 0.009649 0.9997 3.1 0.73 2.45 98.2% 86.7%
1,2-Dichlorobenzene 27.052 y = 1.626089 * x + 0.010158 0.9995 2.7 0.64 2.12 97.4% 87.6%
1,2,4-Trichlorobenzene 31.265 y = 1.080196 * x + 0.006445 0.9996 4.1 0.95 3.17 78.2% 71.7%
Hexachlorobutadiene 31.715 y = 0.610358 * x + 0.008732 0.9995 2.5 0.51 1.69 96.3% 78.9%
www.agilent.com/chem
DE4429.2920023148
Authors Abstract
Coreen Hamilton and
This study provides data used to create an alternate testing protocol for the U.S.
Xinhui Xie,
Environmental Protection Agency (EPA) to use for Agilent 7010B Triple Quadrupole
SGS AXYS Analytical Services
GC/MS analysis of tetra- through octa-dioxins and furans that is equivalent to
Ltd.
EPA Method 1613B. EPA Method 1613B is used for the determination of the
Tarun Anumol, 17 toxic tetra- through octa-chlorinated Dibenzo-p-Dioxins and Dibenzofurans
Anastasia Andrianova, and (CDDs/CDFs) in aqueous, solid, and tissue matrices by isotope dilution gas
Dale Walker, chromatography/high-resolution mass spectrometry (GC/HRMS) using magnetic
Agilent Technologies, Inc. sector instruments. Traditionally used for dioxins analysis because of their high
sensitivity, GC/HRMS instruments are expensive to maintain, require a highly
specialized skill set to operate, and are being phased out by manufacturers.
However, current GC/MS/MS (GC/TQ) technology provides many of the specificity
and sensitivity advantages of HRMS for the analysis of regulated dioxins and furans,
without the cost and complexity, and with added versatility and robustness. This
application note describes a method developed in collaboration with SGS AXYS
Analytical Services Ltd., SGS AXYS Method 16130, that uses the Agilent 7890B
gas chromatograph coupled with an Agilent 7010B Triple Quadrupole GC/MS.
Performance factors investigated included sensitivity, linearity, method detection
limits (MDLs), recovery, and results compared to reference material. The GC/TQ
results met the QA/QC and performance specifications described in Method
1613B for the analysis of polychlorinated dioxins and furans (PCDDs/PCDFs)
in environmental matrices. Overall, the GC/TQ method produced accurate data
for real-world sample matrices, offering a lower cost, more efficient alternative
to GC/HRMS.
Introduction specifications in Method 1613B for the are spiked prior to extraction. After
analysis of polychlorinated dioxins and extraction, 37Cl4-labeled 2,3,7,8-TCDD is
Dioxins are pollutants of concern due to furans (PCDDs/PCDFs) in environmental added to each extract to measure the
the adverse effects of trace-level chronic matrices. The method--SGS AXYS efficiency of the cleanup process. After
exposure, persistence in the environment, Method 16130--is approved by the US cleanup, the extract is concentrated
and bio-accumulation in the food chain.1 EPA as an alternate testing protocol to near dryness. Immediately prior to
For this reason, they are monitored by for analyzing the Dioxins in EPA 1613B. injection into the GC for GC/TQ analysis,
environmental agencies worldwide. Performance factors investigated internal standards were added to
The U.S. Environmental Protection in this application note included each extract.
Agency (EPA) has promulgated sensitivity, linearity, method detection
Method 1613B for the determination limits (MDLs), recovery, and results for GC/TQ analysis and instrumentation
of the 17 toxic 2,3,7,8-substituted reference materials. GC/TQ analysis was carried out with a
tetra- through octa-chlorinated 7890B gas chromatograph coupled with
The EPA has reviewed the SGS AXYS
CDDs/CDFs in aqueous, solid, and a 7010B Triple Quadrupole GC/MS. The
Method 16130 using the 7010B Triple
tissue matrices by isotope dilution 7890B gas chromatograph was equipped
Quadrupole GC/MS and supporting
gas chromatography/high-resolution with a 60 meter Agilent DB-5 column
validation data submitted by SGS
mass spectrometry (GC/HRMS) (part number 122-5061). All GC/HRMS
AXYS, and has determined that it
using magnetic sector instruments. data used for comparison were also
meets requirements as an alternate
As originally written, Method 1613B collected using a DB-5 column operated
testing protocol for measurement
requires a high mass resolution of under similar conditions. The 7010B
of 2,3,7,8-substituted tetra- through
≥10,000, which can only be achieved Triple quadrupole GC/MS was operated
octa-chlorinated dibenzo-p-dioxins
using GC/HRMS. Traditionally, magnetic in the MRM mode and equipped with
and dibenzofurans (PCDDs/PCDFs) in
sector MS instruments have been a high-efficiency EI source (HES).
wastewater with performance similar to
used for this analysis due to lack of The GC/TQ parameters are provided
the methods listed in 40 CFR Part 136.
better alternatives. However, magnetic in Table 1.
Though the EPA has not yet promulgated
sector MS instruments are expensive to The GC/TQ system was tuned
the method or published it in the CFR
maintain and require a highly specialized to Agilent specifications using
at the time of this publication, on a
skill set to operate. In addition, with perfluorotributylamine (PFTBA) and
facility-by-facility basis laboratories
suppliers discontinuing or phasing the default HES tune. The method
may seek approval from their regional
out manufacture of magnetic sector stipulates that the system is ready to
authority to use the method in measuring
GC/HRMS instruments, an alternate operate as long as the vendor-specified
PCDDs/PCDFs in wastewater in per the
technique that provides data of the same tune criteria are met. Method 1613B
Clean Water Act (CWA) program.
quality, with easier and more robust requires a mass resolution check every
operation, is required.
Experimental 12 hours. The analogous parameter
MS/MS technology offers many of the when using MS/MS is a mass calibration
specificity and sensitivity advantages Sample preparation and extraction and tuning check. Every 12 hours the
of HRMS methods without the need Sample cleanup is required to maintain mass calibration was monitored by
for high mass resolution, or the cost the MS instrument in good condition, measuring the amount of peak drift
and complexity of HRMS instruments. and to avoid mass fluctuations and from the expected masses for PFTBA.
Approval of a method that uses changes in ionization efficiency due to If the peak apex had shifted more than
GC/MS/MS (GC/TQ) for determination background matrix. For this application 0.3 amu from the expected value, then
of dioxins and furans has the potential note, analyses were performed using the instrument was recalibrated.
to lower laboratory costs. Developed real-world sample extracts from four The need for lock mass monitoring
in collaboration with SGS AXYS matrices (aqueous, solids, biosolids, of the GC/HRMS system for Method
ANALYTICAL SERVICES LTD, this and tissues) that had been archived 1613B was replaced by use of a
application note describes a GC/TQ at SGS AXYS Analytical Services after stability reference compound in the
method using an Agilent 7890B gas preparation and extraction per EPA GC/TQ method. A small but constant
chromatograph coupled with an Agilent Method 1613B.1 In this procedure, amount of PFTBA, the reference
7010B Triple Quadrupole GC/MS that stable isotope-labeled analogs of 15 compound used for tuning and mass
meets the QA/QC and performance of the 2,3,7,8-substituted CDDs/CDFs calibration, was introduced and the
2
MRM transition 414.0 & 264.0 was Table 1. GC/TQ parameters.
monitored throughout the run. Any
Parameter Value
changes in the ionization efficiency and
Gas Chromatograph
ion transmission can be observed as
Model Agilent 7890B gas chromatograph
a change in the reference compound
Column Agilent DB-5, 60 m × 0.25 mm, 0.1 µm (p/n 122-5061)
signal intensity.
Column Pneumatics Constant flow, He carrier gas
Two transitions were monitored for
Injector Mode Splitless
each of the native PCDD/PCDF analytes
Injector Liner Inlet liner, splitless, double taper, deactivated (p/n 5181-3315)
and their corresponding 13C-labeled
Injection Volume 1.0 µL
analogues. Two masses from the
Injector Temperature 290 °C
molecular ion cluster were used as the
Flow Rate 0.93 mL/min
transition precursors, each with its own
90 °C for 2 min,
product ion (loss of neutral CO35Cl). 22 °C/min to 200 °C,
Temperature Program
MRM delivers a unique product ion 1 °C/min to 215 °C, hold 10 min,
5.2 °C/min to 300 °C, hold 2.7 min
that can be monitored and quantified
Total Run Time 51.05 min
in a complicated matrix, providing the
Equilibration Time 0.1 min
selectivity needed for PCDD/PCDF
Mass Spectrometer
analysis. The triple-stage selection
Model Agilent 7010B Triple Quadrupole GC/MS
process for ions reaching the detector
Ionization Mode EI, 70 eV
results in low noise and thus a high
Acquisition Mode MRM
signal-to-noise ratio (S/N) and good
sensitivity and selectivity for analytes. Filament Current 100 μA
for each analyte and labeled compound Quench Gas He at 2.25 mL/min
software was used for data acquisition, Ion Source Temperature 290 °C
3
Table 2. MRM transitions (continued).
Cleanup Standard
37
Cl4-2,3,7,8-TCDD 327.9 & 262.9 33 - C12-1,2,3,4-TCDD
13
13
C12-1,2,3,7,8-PeCDD 367.9 & 303.9 25 365.9 & 301.9 25 C12-1,2,3,4-TCDD
13
13
C12-1,2,3,4,7,8-HxCDD 401.9 & 337.9 25 403.9 & 339.9 25 C12-1,2,3,7,8,9-HxCDD
13
13
C12-1,2,3,6,7,8-HxCDD 401.9 & 337.9 25 403.9 & 339.9 25 C12-1,2,3,7,8,9-HxCDD
13
13
C12-1,2,3,4,6,7,8-HpCDD 435.8 & 371.9 25 437.8 & 373.9 25 C12-1,2,3,7,8,9-HxCDD
13
13
C12-OCDD 469.8 & 405.8 26 471.8 & 407.8 26 C12-1,2,3,7,8,9-HxCDD
13
13
C12-2,3,7,8 -TCDF 315.9 & 252.0 33 317.9 & 254.0 33 C12-1,2,3,4-TCDD
13
13
C12-1,2,3,7,8-PeCDF 351.9 & 287.9 35 349.9 & 285.9 35 C12-1,2,3,4-TCDD
13
13
C12-2,3,4,7,8-PeCDF 351.9 & 287.9 35 349.9 & 285.9 35 C12-1,2,3,4-TCDD
13
13
C12-1,2,3,4,7,8-HxCDF 385.9 & 321.9 35 387.9 & 323.9 35 C12-1,2,3,7,8,9-HxCDD
13
13
C12-1,2,3,6,7,8-HxCDF 385.9 & 321.9 35 387.9 & 323.9 35 C12-1,2,3,7,8,9-HxCDD
13
13
C12-1,2,3,7,8,9-HxCDF 385.9 & 321.9 35 387.9> 323.9 35 C12-1,2,3,7,8,9-HxCDD
13
13
C12-2,3,4,6,7,8-HxCDF 385.9 & 321.9 35 387.9 & 323.9 35 C12-1,2,3,7,8,9-HxCDD
13
13
C12-1,2,3,4,6,7,8-HpCDF 419.8 & 355.9 36 421.8 & 357.9 36 C12-1,2,3,7,8,9-HxCDD
13
13
C12-1,2,3,4,7,8,9-HpCDF 419.8 & 355.9 36 421.8 & 357.9 36 C12-1,2,3,7,8,9-HxCDD
13
Recovery Standards
13
C12-1,2,3,4-TCDD 331.9 & 268.0 24 333.9 & 270.0 24
13
C12-1,2,3,7,8,9-HxCDD 401.9 & 337.9 25 403.9 & 339.9 25
Cl-DPE Transitions
Descriptor Type Substance
1 375.8 & 305.9 30 M+2 HxCDPE
As with the GC/HRMS Method 1613B, could be used for this application. RTL were applied to the MS/MS data.
individual PCDD/PCDFs were identified provides the same retention times on The non-2,3,7,8 substituted isomers
by comparing the GC retention time one Agilent GC/MS system to those and congeners were identified when
and MRM transition product ion ratio on another like system with the same retention times and ion-abundance
(primary/secondary transition, Table 3), nominal column. It also enables a single ratios were within predefined limits.
with the corresponding retention time GC to have the same retention time after Isomer specificity for 2,3,7,8-TCDD and
of the authentic standard and the the column is trimmed for maintenance. 2,3,7,8-TCDF was achieved using GC
theoretical transition product ion ratio. Shown in Table 3, the QC limits (±15% columns that resolve these isomers from
Though not used here, Agilent’s-patented of theoretical) of Method 1613B the other tetra-isomers.
retention time locking (RTL) technology
4
Table 3. Theoretical product ion ratios and ratio QC limits. Results and discussion
MRM Transition MRM Transition QC Limit **
5
A
×10 4
6 2378-TCDF
21.366 Cpd 15: TCDFs 21.366
2 27.976
1 9
2 8
0
×10 5
4 3 7
O
25.104 Cpd 16: 13C12-TCDFs 25.104 4 6
Clx Cly
3
x+y=4
Counts
×10 4 2378-TCDD
2.5 First eluted 22.885 25.111 Cpd17: TCDDs 25.111 Last eluted
1368-TCDD 1289-TCDD
2.0 26.351
Counts
1.5 26.605
28.138
1.0
0.5 1 9
O
0 2 8
×10 5
3 7
26.329 Cpd 18: 13C12-TCDDs 26.329 O
1.5 4 6
26.007 Clx Cly
x+y=4
Counts
1.0
0.5
0
20.5 21.0 21.5 22.0 22.5 23.0 23.5 24.0 24.5 25.0 25.5 26.0 26.5 27.0 27.5 28.0
Acquisition time (min)
Figure 1A. MRM chromatograms for tetrachlorinated dibenzofurans (TCDFs), labeled TCDF ISTD, tetrachlorinated dibenzodioxins (TCDDs), and labeled TCDD ISTD.
6
B
×10 4
4.00
3.75 41.854 Cpd 25: HxCDDs 41.856
3.50 123678-HxCDD
2.00
1.75
1.50
1.25
1.00
0.75
1 9
0.50 O
2 8
0.25
0
3 7
×10 5 O
1.2 4 6
41.843 Cpd 3: 13C12-HxCDDs 41.843 Clx Cly
1.1 x+y=6
1.0
0.9 42.243
0.8
0.7
Counts
0.6
0.5
0.4
0.3
0.2
0.1
0
38.5 39.0 39.5 40.0 40.5 41.0 41.5 42.0 42.5 43.0 43.5 44.0
Acquisition time (min)
7
×10 4
2.0
1.9
26.351
1.8
1.7
1.6
1.5
26.176 26.605
1.4
26.029
1.3 26.966
1.2
1.1
Counts
1.0
0.9 y
0.8
0.7
0.6
0.5
0.4
0.3
0.2
x
0.1
0
25.4 25.6 25.8 26.0 26.2 26.4 26.6 26.8 27.0 27.2 27.4 27.6
Acquisition time (min)
Figure 3. (A) Method setup for resolution check in MassHunter Quantitative Analysis; (B) front and rear valley height/peak height resolution calculated for
2,3,7,8-TCDD and its closest eluting isomers.
8
The 7010B triple quadrupole GC/MS is Table 4. %RSDs of the PCDD/PCDF response factors for the five days of initial calibrations.
equipped with a high-efficiency EI source
Date Acquired 19-AUG-19 21-AUG-19 06-JAN-20 07-JAN-20 08-JAN-20
that produces up to 20 times more ions
Data File ID DX9Z0415-A1 DX9Z0444-A1 DX9Z0830-A1 DX9Z0837-A1 DX9Z0853-A1
and maximizes ion transfer into the
Name RRF %RSD RRF %RSD RRF %RSD RRF %RSD RRF %RSD
quadrupole mass analyzer, allowing
2,3,7,8-TCDF 4.0 3.0 4.4 2.7 2.4
significantly more sensitivity while still
maintaining robustness. 1,2,3,7,8-PeCDF 3.7 2.8 3.4 2.9 2.7
9
Table 5. GC/TQ MDL results with comparison to Method 1613B MRLs. Recoveries
Aqueous Solid Tissue
Three sets of spiked clean matrix one
MDL and (MRL) MDL and (MRL) MDL and (MRL)
Compound in pg/L in pg/g in pg/g
each of aqueous (1 L), solids (10 g) and
2,3,7,8-TCDD 1.1 (10) 0.029 (1) 0.057 (0.5)
tissues (10 g) were run and the mean
percent recovery (n = 4) and percent
1,2,3,7,8-PeCDD 1.39 (50) 0.037 (5) 0.051 (2.5)
RSD calculated (Figure 6). Results were
1,2,3,4,7,8-HxCDD 1.05 (50) 0.042 (5) 0.061 (2.5)
compared and determined to conform to
1,2,3,6,7,8-HxCDD 1.08 (50) 0.045 (5) 0.033 (2.5)
Method 1613B IPR specifications.
1,2,3,7,8,9-HxCDD 1.78 (50) 0.064 (5) 0.067 (2.5)
1,2,3,4,6,7,8-HpCDD 1.19 (50) 0.070 (5) 0.032 (2.5) Proficiency, SRM, and CRM results
OCDD 9.4 (100) 0.311 (10) 0.085 (5) The evaluation report from Sigma‑Aldrich
2,3,7,8-TCDF 0.56 (10) 0.60 (1) 0.056 (0.5) RTC, Inc. concluded that both GC/HRMS
1,2,3,7,8-PeCDF 1.0 (50) 0.037 (5) 0.046 (2.5) and GC/TQ results obtained from the
2,3,4,7,8-PeCDF 1.25 (50) 0.039 (5) 0.033 (2.5) proficiency tests were acceptable and
1,2,3,4,7,8-HxCDF 0.89 (50) 0.032 (5) 0.029 (2.5) met study criteria and with an overall
1,2,3,6,7,8-HxCDF 1.11 (50) 0.031 (5) 0.046 (2.5) score of 100%. These results indicate
1,2,3,7,8,9-HxCDF 1.22 (50) 0.048 (5) 0.084 (2.5) the accuracy of PCDD/PCDF data from
2,3,4,6,7,8-HxCDF 1.26 (50) 0.026 (5) 0.034 (2.5) the 7010B Triple Quadrupole GC/MS
1,2,3,4,6,7,8-HpCDF 0.92 (50) 0.255 (5) 0.064 (2.5) analysis of the environmental matrices.
1,2,3,4,7,8,9-HpCDF 1.35 (50) 0.028 (5) 0.043 (2.5) The results of the GC/TQ analysis of
OCDF 2.81 (100) 0.365 (10) 0.113 (5) the solids SRM (NIST 1944) and tissue
CRM (EDF 2525) also demonstrated the
accuracy of the GC/TQ method.
100.000
Concentration (pg/g)
10.000
1.000
0.100
0.010
2,3,7,8-TCDD
1,2,3,7,8-PECDD
1,2,3,4,7,8-HXCDD
1,2,3,6,7,8-HXCDD
1,2,3,7,8,9-HXCDD
1,2,3,4,6,7,8-HPCDD
OCDD
TOTAL HEXA-DIOXINS
2,3,7,8-TCDF
1,2,3,7,8-PECDF
2,3,4,7,8-PECDF
1,2,3,4,7,8-HXCDF
1,2,3,6,7,8-HXCDF
1,2,3,7,8,9-HXCDF
TOTAL HEPTA-DIOXINS
TOTAL TETRA-FURANS
TOTAL PENTA-FURANS
TOTAL HEXA-FURANS
TOTAL HEPTA-FURANS
2,3,4,6,7,8-HXCDF
1,2,3,4,6,7,8-HPCDF
1,2,3,4,7,8,9-HPCDF
OCDF
TOTAL TETRA-DIOXINS
TOTAL PENTA-DIOXINS
Figure 4. Comparison of total PCDD/PCDF for a real-world biosolids sample determined by GC/TQ (blue bars) and
GC/HRMS (red bars).
10
Table 6. Fortified concentration, mean percent recovery (n = 4), and percent RSD for spiked clean matrix.
13
C-2,3,7,8-TCDD 2,000 70 8 200 58 12 200 73 4
13
C-1,2,3,7,8-PECDD 2,000 74 9 200 62 15 200 78 5
13
C-1,2,3,4,7,8-HXCDD 2,000 81 4 200 64 10 200 71 9
13
C-1,2,3,6,7,8-HXCDD 2,000 79 5 200 61 9 200 70 9
13
C-1,2,3,4,6,7,8- HPCDD 2,000 87 5 200 69 12 200 74 9
13
C-OCDD 4,000 76 5 400 60 14 400 63 9
13
C-2,3,7,8-TCDF 2,000 67 7 200 53 11 200 65 3
13
C-1,2,3,7,8-PECDF 2,000 68 9 200 57 14 200 71 5
13
C-2,3,4,7,8-PECDF 2,000 69 9 200 57 15 200 74 4
13
C-1,2,3,4,7,8-HXCDF 2,000 77 5 200 63 9 200 66 10
13
C-1,2,3,6,7,8-HXCDF 2,000 78 6 200 61 9 200 68 8
13
C-1,2,3,7,8,9-HXCDF 2,000 75 4 200 60 12 200 73 8
13
C-2,3,4,6,7,8-HXCDF 2,000 79 5 200 62 10 200 70 9
13
C-1,2,3,4,6,7,8- HPCDF 2,000 77 6 200 62 9 200 66 9
13
C-1,2,3,4,7,8,9- HPCDF 2,000 83 5 200 67 12 200 71 12
37
Cl-2,3,7,8-TCDD 200 73 6 20 69 7 20 79 3
11
Note about potential interferences Conclusion References
In this study, the analysis of 36
real-world samples of four sample GC/TQ technology provides many of the 1. US EPA. Method 1613: Tetra-Through
matrices showed no interferences, and specificity and sensitivity advantages Octa-Chlorinated Dioxins and Furans
chromatography and quantified results of HRMS for the analysis of regulated by Isotope Dilution HRGC/HRMS,
for GC/TQ were equivalent to GC/HRMS. dioxins and furans without the cost September 1994. https://nepis.epa.
In addition, concentrated standards and complexity of HRMS instruments, gov/ (accessed December 1, 2020)
of PAH, alkylated PAH and chlorinated with added versatility and robustness.
pesticides showed no response when Approval of GC/TQ technology for
analyzed by the GC/TQ method. determination of dioxins and furans as
However, because there is incomplete an alternative testing protocol to Method
chromatographic separation of the 1316B has the potential to significantly
chlorinated diphenyl ethers (CDPEs) lower laboratory costs and increase
from PCDFs, a characteristic m/z for operational efficiency.
each chlorinated diphenyl ether must be This application note described and
monitored. If detected at the retention evaluated the GC/TQ SGS AXYS Method
time of any PCDFs, additional cleanup 16130 using the Agilent 7890B gas
must be performed per Method 1613B. chromatograph coupled with an Agilent
The GC/HRMS requirement to monitor 7010B Triple Quadrupole GC/MS. The
CDPEs and perform additional cleanup method will eventually be added into the
when detected remains when using the Federal Register. The results obtained
GC/TQ method. were determined to meet the QA/QC and
In addition, although there are small performance specifications in Method
mass differences (about 6 amu) between 1613B for the analysis of PCDDs/PCDFs
some PCBs and some PCDD/PCDFs in environmental matrices. Performance
at the same level of chlorination, factors investigated included sensitivity,
the DB-5 column provides complete linearity, MDLs, recovery, and results
chromatographic separation of these compared to reference material. The
compounds. However as with GC/HRMS, results of the performance tests
interferences from fragments of demonstrated that GC/TQ using 7010B
higher homolog PCBs are possible. It Triple Quadrupole GC/MS provides
is recommended that extract cleanup data of the same quality for real world
procedures include a step to remove samples representing complex matrices.
PCBs from sample extracts.
www.agilent.com/chem
DE44221.5046296296
Accelerated Determination of
Microplastics in Environmental
Samples Using Thermal
Extraction Desorption-Gas
Chromatography/Mass Spectrometry
(TED-GC/MS)
Authors Abstract
Ulrike Braun, Paul Eisentraut,
There is growing interest in quantifying microplastics in environmental
Korinna Altmann,
samples. This application note presents a thermal extraction desorption-gas
Maria Kittner, and
chromatography/mass spectrometry (TED-GC/MS) method that is well suited to
Erik Dümichen
automation and increased sample throughput. The method is also able to detect
Bundesanstalt für
all particle sizes in the sample as long as the limit of detection (LOD) is reached
Materialforschung und
and allows analysis of larger samples of 15 to 25 mg or more. Samples were
-prüfung (BAM),
decomposed by thermogravimetric analysis (TGA), and the gaseous decomposition
Berlin, Germany
products were trapped on a solid-phase sorbent, followed by thermal desorption‑gas
Kurt Thaxton and chromatography/mass spectrometry (TD-GC/MS) using an Agilent 5977B GC/MSD
Eike Kleine-Benne coupled to an Agilent 7890B GC. Target microplastic particle (MP) polymers were
GERSTEL Inc. identified in environmental samples including surface water, finished compost,
Tarun Anumol house dust, and drinking water. Quantification of MP polymers in environmental
Agilent Technologies, Inc. samples provided LODs of 0.06 to 2.2 µg, allowing the detection of MPs in trace
amounts with sample weights of up to 1 g. Method repeatability was adequate for
reliable quantification with RSDs of approximately 6 to 12%.
Introduction Thermo-analytical methods offer an allows analysis of larger samples of 15
alternative. Complementary to imaging to 25 mg or more, resulting in a more
Microplastic particles (MPs) are defined techniques, they provide information on representative sample and eliminating
as any solid plastic particles insoluble mass content that can be essential for the need to take multiple samples.
in water with any dimension between regulatory and monitoring purposes. This is in contrast to Py-GC/MS, where
1 µm and 1,000 µm (ISO/TC 21960). As temperature is increased, pyrolytic sample sizes are typically ten times less
Consisting mainly of synthetic polymers, decomposition of polymers occurs, and thus require running ten times the
MPs are observed nearly everywhere in forming smaller molecules that can be samples to achieve the same degree of
the environment. Though, the systematic analyzed using spectroscopic techniques representation, while also potentially not
occurrence of MPs in environmental such as IR spectrometry and mass covering a large enough sample volume
samples has not yet been sufficiently spectrometry. In addition, the thermal to achieve homogeneity.
investigated, it is necessary to fully decomposition products formed can An instrument system developed by the
understand their relevant sources generate structural information about Bundesanstalt für Materialforschung
and entry pathways into the food and the polymer analyzed. However using und -prüfung (BAM) in cooperation with
environmental ecosystem and to initiate such evolved gas analysis tools for GERSTEL GmbH & Co. KG adds a heated
sustainable prevention strategies. MPs environmental samples produces coupling device that enables a sample
are mainly formed in the environment unspecific signals, which cannot be robot to transfer thermal desorption
by fragmentation of macroplastics, unambiguously related to specific tubes from the pyrolysis gas flow of
and remain for an uncertain length of MP type. Pressurized liquid extraction the TGA furnace to the TD-GC/MS
time. This persistence increases the in combination with pyrolysis-gas system for automated analyses.4 This
likelihood that MPs will be ingested and chromatography/mass spectrometry application note describes use of a
accumulated in organisms, the risks (Py-GC/MS) can provide reliable 5977B GC/MSD coupled to a 7890B GC
of which are unclear and the subject determination of MPs in environmental as part a complete automated
of controversy. As a result, there is samples, however large quantities TED-GC/MS system, which is exclusively
increased interest in determining MPs of solvents, some of which are not available through GERSTEL and Agilent.
in environmental samples such as soils, environmentally friendly, must be Further details about the experimental
water, air, and organic matter. used and the additional extraction setup and methodology are provided
MP analysis commonly relies on step carries the risk of particle loss in the publications by Dümichen et al.,4
spectroscopic imaging techniques, such or contamination.2,3 Altmann et al.,5 and Eisentraut et al.6
as infrared (IR) or Raman microscopy Another approach, TED-GC/MS, is better
that provide particle numbers, and suited to automation and increased
polymer chemical identity, geometry, and sample throughput without an additional
size distributions.1 The disadvantage of sample preparation step, including
these techniques is that they are often reproducible automated analysis of
time-consuming and require extensive polymer decomposition processes.4-6 A
sample preparation. Small particles further advantage of TED-GC/MS is its
(<10 µm) cannot directly be analyzed ability to detect all particle sizes in the
by spectroscopic imaging, but are sample as long as the limit of detection
expected to be the relevant fraction for is reached. In this approach, the sample
toxicology. Quantification of MPs in is decomposed by TGA and the gaseous
environmental samples using traditional decomposition products are trapped
GC/MS or LC/MS methods has been on a solid-phase sorbent, followed by
challenging, because of their very high TD-GC/MS. MP samples can be very
molecular weights and poor solubility in heterogeneous. By using a TGA furnace
most solvents. with a relatively large oven, TED-GC/MS
2
Experimental GERSTEL’s PDMS Twister devices may The GC/MS system was a 7890B GC
be used for similar results. coupled to a 5977B GC/MSD. The 7890B
TED-GC/MS instrumentation A Thermal Desorption Unit (TDU 2, GC was equipped with cryo‑injection
TGA heats the sample in an GERSTEL) performed the thermal (CIS 4, GERSTEL). The analytes were
inert atmosphere, during which extraction step. It was mounted on top cryo-focused in the cooling trap (CIS)
measurements of mass versus of the coupling device, allowing the TD at –100 °C using liquid nitrogen.
temperature can be made. The technique tube to be automatically transferred When the thermal desorption process
is convenient, reliable, and robust for by the sample robot based on the was completed, the trapped analytes
analysis of sample weights up to 1 g and time program. Another TDU attached were quickly reheated, separated, and
temperatures up to 1,000 °C. TGA was to the GC/MS performed the thermal analyzed by GC/MS. The TD-GC/MS
performed using a thermogravimetric desorption step. parameters are presented in Table 2.
furnace equipped with an autosampler
(TGA2, METTLER TOLEDO) and 150
Table 1. TGA parameters for TED-GC/MS.
µL alumina crucibles. Dümichen et
al determined that a TGA purge gas Parameter Value
flow rate of 30 mL/min N2 is the best Model METTLER TOLEDO large furnace TGA 2
compromise between analyte response Sample Mass 20 mg (standard conditions; but up to 1,000 mg possible)
and contamination, and that a heating Purge Gas 30 mL/min N2
rate of 10 °C/min provides the best Heating Rate 10 °C/min
average response.4 The TGA parameters Total Exhaust Flow 56 mL/min
are provided in Table 1. Heating Range 25 to 600 °C
PDMS adsorption bar was 2 mm by Mode Scan, 35 to 350 m/z (under specific circumstances 25 to 450 m/z for PE)
3
The complete TED-GC/MS instrument
setup is shown in Figure 1. Water filtrate,
air filter samples, neat polymers, soil,
compost, and other materials filled into
heat-resistant conventional or smart
filter crucibles can be measured directly
without additional sample preparation.
In this approach, the total TED-GC/MS
measurement time is approximately
two hours and 30 minutes: one hour
for TGA, one hour for GC/MS analysis,
and about 30 minutes for GC cool
down and TGA device bake out. After
the thermal extraction step and during
the GC/MS run, the TGA furnace can
be cleaned by heating to 1,000 °C
under oxidizing atmosphere. A new
sample measurement can then be
started after completion of the GC/MS
measurement. Complete automation
Figure 1. Complete TED-GC/MS instrument setup with the TGA furnace (right) and coupling device (top).
allows measurement of up to ten
The TD-GC/MS system based on the Agilent 5977B GC/MSD (left), and sample robot. The patented smart
samples per day, including blank runs filter crucible is shown bottom right. Photos courtesy of BAM.
between measurements. In general, a
sample mass of 20 mg is sufficient, • Fractions of filtered surface water characteristic fragment ions associated
but if needed, masses up to 1 g can • Finished compost with their decomposition products and
be investigated. retention times (tR) (Table 3). Polystyrene
• House dust deuterated on the benzene ring (d5PS)
Agilent MassHunter GC/MS Acquisition
• Drinking water in PET bottles was used as an internal standard to
Software revision B.07.04.2260 was
The technique is versatile and allows provide a relative reference value for
used for data acquisition. Agilent
other experiments, such as the analysis measurement quality control. One
Enhanced ChemStation software revision
of a PE/PP blend with unknown relative decomposition product of the internal
F.01.03.2357 was used for data analysis.
weight mass5 and the determination of standard is the deuterated dimer of
Samples adsorbed pollutants on activated carbon styrene d10-(2,4-diphenyl-1‑buten)
TED-GC/MS allows routine from wastewater treatment filters that with the characteristic fragment
determination of the polymers were analyzed per Dittmann et al.7 ions m/z 96 and 218, which can be
commonly used in manufacturing, distinguished from the decomposition
and thus of environmental concern, Polymer identification product of PS (2,4-diphenyl-1-butene) at
including those in tire materials. The polymers in the MP samples m/z 91 and 208 using MS, but also by
Polymers determined in this study were identified via detection of the chromatographic separation.
were polyethylene (PE), polypropylene
(PP), polystyrene (PS), poly(ethylene Table 3. Specific polymer markers and characteristic fragment ions.
4
Polymer quantification In the fractions of filtered surface water, identified in the fractions <100 μm. A
Quantification of polymers is typically PE and SBR were dominant, but small total of 73 μg/L of target polymers was
performed using standard addition. By amounts of PP and PS were also found. found in the surface water.
rerunning samples after adding polymer A component of tires, SBR is mainly
standards in known amounts (available
from BAM), the change in peak response
was used to determine polymer
mass contents.
A faster alternative is the use of
response factors obtained from
one‑point calibration experiments or
external calibration where polymers
in various levels are added to suitable
plastic-free matrices. It was determined
that for samples from the same or
similar origin, a standard addition for a
single, carefully selected sample was
suitable for quantification of the entire
set of samples.
Figure 2. Total ion chromatogram (TIC) from the TED-GC/MS analysis of an environmental sample.
Results and discussion The insets show extracted ion chromatograms that were used to identify the polymers based on their
characteristic ions.
14 0.15 PS
21 μg/mg and 13 μg/mg, respectively. 12
PET
PET
SBR
However, the polymer compositions of 10
SBR
PA 0.10
the samples differed. In house dust, the 8 Acrylate
PA
Acrylate
fiber polymers PET, PA, and PP were 6
found in the highest amounts, while the 4
0.05
compost mainly included PE and PP, 2
which are polymers commonly used 0 0
500 100 50 10 PE PP PS PET SBR PA Acrylate
for packaging. Surface water Bottled water
Figure 3. MP polymer content found in house dust, finshed compost, surface water, and bottled water by
TED-GC/MS. In the surface water analysis chart, the bars show the MP content of the different particle
fractions in μm resulting from the sampling.
5
Because bottled water typically contains Repeatability and the integration values were
lower levels of MP, a filter crucible with The repeatability of the TED-GC/MS normalized to the sample masses. The
5 μm pore size was used. Up to 10 L of method using the 5977B GC/MSD results showed good repeatability for
bottled or tap water were poured directly was determined by measuring specific reliable quantitation (Figure 5). Three
onto the filter and filtered using vacuum thermal decomposition products of of the degradation products had a
filtration with subsequent drying and PP (Table 4) in seven separate runs. Relative Standard Deviation (RSD) of
analysis of the filter contents. A total The mass fragments with the highest approximately 6% (Table 4). Repeatability
of 0.193 μg/L of PET was found in the intensities were chosen for integration, results could be improved using
bottled water. internal standards.
2
without environmental matrix and
for the respective polymer in fluvial
suspended particulate matter (SPM). Six
conventional standard polymers (PE, PP,
PS, PET, SBR, and PMMA) were analyzed. 1
and terephthalic acid dibut-3-enyl-ester Figure 4. LOD values for the neat polymers compared to the LOD of the polymer in suspended
(tR = 34.7 minutes, m/z = 203). particulate matter.
6
5E+08
4E+08
Peak area (normalized for PP mass)
3E+08
2E+08
1E+08
1 2 3 4 5 6 7
Measurement
Figure 5. Integration results for the repeatability determination normalized to the sample mass
for six selected decomposition product peaks. The average values are shown as a straight line.
7
References 5. Altmann, K. et al.
TED-GC/MS: Schnelle Bestimmung
1. Robey, D. Microplastics Analysis von Mikroplastik-Massegehalten
Doesn’t Need to Be So Hard. Agilent in verschiedenen Proben. Mitt
Technologies, publication number Umweltchem Ökotox. 2020, 2,
5994-1584EN, 2019. 55–57,
2. Dierkes, G.; Anumol, T. Quantification 6. Eisentraut, P. et al. Two Birds with
of Microplastics in Environmental One Stone—Fast and Simultaneous
Samples Using Pyrolysis and Analysis of Microplastics:
GC/MSD. Agilent Technologies, Microparticles Derived from
publication number 5994-2199EN, Thermoplastics and Tire Wear.
2020. Environmental Science & Technology
3. Dierkes, G. et al. Quantification of Letters 2018, 5(10), 608-613.
Microplastics in Environmental 7. Dittmann, D. et al. Specific
Samples Via Pressurized Liquid Adsorption Sites and Conditions
Extraction and Pyrolysis-Gas Derived by Thermal Decomposition
Chromatography. Analytical and of Activated Carbons and Adsorbed
Bioanalytical Chemistry 2019, Carbamazepine. Scientific Reports
411(26), 6959–6968. 2020, 10(1), 6695.
4. Duemichen, E. et al. Automated
Thermal Extraction-Desorption
Gas Chromatography Mass
Spectrometry: a Multifunctional
Tool For Comprehensive
Characterization of Polymers and
Their Degradation Products. Journal
of Chromatography A 2019, 1592,
133–142.
www.agilent.com/chem
DE.6298726852
Authors Abstract
Sujan Fernando and
1,4-Dioxane is an industrial chemical contaminate that may be found at trace levels
Alesia Haddad-Carroll
in consumer products.1–6 Government jurisdictions are beginning to regulate the
Clarkson University
amount of 1,4-dioxane allowed in consumer products. The allowable concentrations
Ron Honnold, Simon Jones, are expected to vary from state to state and country to country. Meanwhile, it has
Shannon Coleman, already been banned and deemed unsafe in cosmetics in Canada.1–6 There have
Joe Hendrick, and been several methods developed to test for 1,4-dioxane, but none of these methods
Tim Conjelko are adequate to detect 1,4-dioxane in consumer products with complex mixtures
Agilent Technologies, Inc. and solutions.7,8
The current study shows methodology for low level detection of 1,4-dioxane
in consumer products including cosmetics, liquid soaps, shampoos, and
cleaning products. The extraction and analysis were initially performed using
an Agilent 7694E headspace sampler attached to an Agilent 7890 GC with an
Agilent 5977A MSD using Agilent MassHunter software. Then, the analysis was
reproduced (with minor method modifications) on a newer system consisting of
an Agilent 7697A headspace sampler attached to an Agilent Intuvo 9000 GC with a
5977B MSD using MassHunter software.
The quantitation of the target analyte 1,4-dioxane was performed using isotope
dilution by adding deuterated 1,4-dioxane-d8 as an internal standard to all samples,
controls, and calibrators. This method has a linear quantitation range from 10 ng/g
to 20,000 ng/g (ppb). Sample size for liquid detergents was typically 2 mL and
for solids was in the range of 0.1 to 2.0 g, depending on the matrix. The method
detection limit was determined to be 7.1 ppb on the older system and 2.3 ppb on
the newer system. Alternate qualifiers were needed for confirmation in matrix due to
interferences at low levels.
Introduction 1,4-dioxane is not used as an ingredient There have been several methods
in consumer products but may be developed to test for 1,4-dioxane,
1,4-Dioxane is a heterocyclic organic present as a trace contaminant in some. primarily for soil, water, air, cleaning
compound, classified as an ether. products, and cosmetics. Without
The New York State Environmental
1,4-Dioxane (C4H8O2) is a colorless modification, none of the standard
Conservation Law has been amended
liquid or solid at cool temperatures methods can quantify 1,4-dioxane at the
to include limits on the maximum
(below 53 °F) with a sweet odor like single-digit parts per million (ppm) level.
allowable concentration of 1,4-dioxane
that of diethyl ether. 1,4-Dioxane is Consumer products are challenging to
in consumer products. The limit for
highly water-soluble, making it mobile analyze due to the complex matrices
household cleaning and personal care
in the environment, and it does not associated with viscous and foaming
products is set to 2 ppm effective
readily biodegrade. products like gels and detergents. Due to
December 31, 2022 and 1 ppm effective
Due to the myriad of problems attributed December 31, 2023. The maximum these method demands, there is a need
to 1,4-dioxane in the environment, it has allowable concentration of 1,4-dioxane for developing a testing method using
been identified as a chemical warranting in cosmetic products will be 10 ppm headspace GC and single ion monitoring
further research toward its removal effective December 31, 2022.10 (SIM) technology.7–8,12,13
from consumer products (cosmetic,
personal care, and cleaning products)
Ethylene oxide
by California DTSC (Department of Toxic O
Lauryl alcohol Laureth-X
Substances Control).9 In New York state, H3C OH H3C O
there is a law in place that restricts x OH
the amount of 1,4-dioxane allowed in
personal care, cleaning, and cosmetic Ethylene oxide rings Ethylene glycol
O open to form ethylene molecules dimerize O
products. The state of New York has glycol molecules to form 1,4-dioxane
HO OH
also adopted a first-in-the-nation drinking Side reaction:
O HO OH
water standard for 1,4-dioxane and set O
the maximum contaminant level of 1 part
per billion (ppb).11
1,4-Dioxane is a trace by-product formed
Laureth-X
during the synthesis of ethoxylated
H3C O
ingredients used in finished consumer x OH
products (cosmetic, personal care, SO2 Na+
and cleaning products). 1,4-Dioxane Followed by Sodium laureth-X sulfate O O–
neutralization
can form inadvertently by undesirable H3C O S
O x
side reactions during ethoxylation. O
It can form when two consecutive
Side reaction:
ethylene oxide units are cleaved from
Sodium laureth-3 sulfate
a chain of ethylene oxides and form
CH3 CH3
a ring of 1,4-dioxane. Or, it can form
SO2 can encourage an
when the ethylene oxide ring opens to O O internal SN2 reaction that O
form ethylene glycol, and two ethylene cleaves the ethoxylated
O O O chain, forming 1,4-dioxane, a OH
glycols dimerize to form 1,4-dioxane. S shortened ethoxylated
Both reaction pathways are shown in S OH alcohol, and regenerated O
O O O O O
acid catalyst S
Figure 1. These ingredients include O
certain detergents, foaming agents, O O
emulsifiers, and solvents identifiable Figure 1. Two reaction pathways for synthesizing 1,4-dioxane (Environment Canada and Health Canada
by the prefix, suffix, word, or syllables: 2010).
PEG, polyethylene, polyethylene glycol,
polyoxyethylene, -eth- (e.g. laureth
sulfate), or -oxynol-. The compound
2
Headspace has been used in several Experimental Supplies
applications for analyzing volatile – Volumetric flasks, Class A, 1 mL and
organic compounds with the aid of Acquisition method 10 mL with ground glass stoppers
gas chromatography. In headspace All analyses were performed on – Glass Pasteur pipets
analysis, the ambient volume above a two systems. The first system was
sample matrix is sampled where the – Volumetric air-displacement pipettes
an Agilent 7890 GC equipped with an
volatile compounds exist in gaseous 50 to 1,000 µL with disposable tips
Agilent 7694E headspace sampler
form at predictable levels. In this attached to an Agilent 5977A MSD. The – 20 mL glass headspace vials with
case, the sample is typically heated to second system was an Agilent Intuvo magnetic caps/PTFE-lined septa
volatilize the compounds of interest into 9000 GC system equipped with an – 12 mL screw cap glass vials
the headspace and allowed to reach Agilent 7697A headspace sampler
equlibrium, which leads to accurate and – Column dimensions:
attached to an Agilent 5977B MSD.
reproducible analytical results when 30 m × 0.25 mm, 0.25 µm
SIM was used to enhance sensitivity
sampled. A major advantage of this and selectivity. – Ultrahigh purity helium
approach is that it enables the analysis
After initial full autotune, a passing – Ultrahigh purity nitrogen
of only the volatile subset of compounds
present in a sample. The complex matrix check tune must be performed before – Methanol: pesticide residue grade,
that may otherwise interfere in the the start of a batch or every 24 hours. If purge and trap grade
case of direct injection of the sample the check tune does not pass, corrective
action must be performed, and then a full Standard preparation
is avoided.
autotune must be run. Make sure that clean glassware is used
Extraction techniques such as in the preparation of standards and
liquid‑liquid extraction or solid- GC method parameters are shown in
samples. Rinse all glassware with Milli-Q
phase extraction may yield cleaner Table 1. Headspace method details are
water prior to use. Use certified vials to
extracts for direct injection. However, in Table 2. MSD parameters are listed in
transfer neat standards if available.
these techniques are time and Tables 3 and 4.
solvent‑consuming and thus are
unattractive compared to headspace. Table 1. GC method parameters used for the analysis of 1,4-dioxane.
The consumer products analyzed
Agilent 7890 GC System Agilent Intuvo 9000 GC System
using the headspace method are highly
Inlet 60 °C 160 °C
complex and contain a myriad of volatile
Hold 0.1 min
and semivolatile compounds that are
Ramp 1 500 °C/min
co-extracted with the 1,4-dioxane. To
Final Temperature 250 °C
compensate for these matrix effects, the
Septa Purge 3 mL/min 3 mL/min
isotope dilution quantitation technique
Mode Split 1:1 Split 10:1
is used.12 Isotope dilution involves
Agilent inlet liner, Ultra Inert, splitless, single Agilent inlet liner, Ultra Inert, splitless, straight
the addition of a known and constant Liner
taper, glass wool (p/n 5190-3171) 2 mm (p/n 5190-6068)
amount of deuterated 1,4-dioxane Oven 40 °C 40 °C
(1,4-dioxane-d8) as an internal standard Hold 0 min 0.5 min
to each sample, quality control, Ramp 1 10 °C /min 10 °C/min
calibrator, and calibration verification.
Final Temperature 140 °C 140 °C
The deuterated analog of 1,4-dioxane
Hold 0 min 0 min
behaves the same as 1,4-dioxane,
Column Flow 2 mL/min 2 mL/min
both physically and chemically,
Agilent J&W DB-5ms Ultra Inert Intuvo GC
allowing for the reproducible and Column
GC column, like the DB-5,
column module, 30 m × 0.25 mm, 0.25 µm
30 m × 0.25 mm, 0.25 µm
accurate quantitation of 1,4-dioxane in (p/n 122-5532UI-INT)
complex matrices.
3
Internal standard (1,4-dioxane-d8) Table 2. Headspace method parameters used for the analysis of 1,4-dioxane.
1. Transfer an aliquot (~1 mL) of the Agilent 7694E Headspace Sampler Agilent 7697A Headspace Sampler
dioxane-d8 neat standard into a clean Incubation Temperature 100 °C 80 °C
vial. Add 25 µL of this solution to a
Incubation Time 10 min 10 min
25 mL volumetric flask, fill to mark
Loop Temperature 120 °C 80 °C
with Milli-Q water, and shake well.
Transfer Line 120 °C 140 °C
Label this solution as 1,000 ppm
Loop Fill and Pressure 0.2 min
dioxane-d8 internal standard.
Fill Pressure 15
2. Next, transfer 25 µL of the 1,000 ppm Loop Ramp 20 psi/min
dioxane-d8 standard to a 25 mL Loop Final Pressure 10
volumetric flask, fill to mark with Loop Equilibrium 0.05
Milli-Q water, and shake well. Label Vial Size 20 20
this solution as 1 ppm dioxane-d8 Vial Shake 5
internal standard.
Table 3. MSD method parameters used for the analysis of 1,4-dioxane.
Native standard (1,4-dioxane)
1. Transfer an aliquot (~1 mL) of Agilent 5977A MSD Agilent 5977B MSD
1,4-dioxane neat standard into a clean Transfer Line Temperature 280 °C 280 °C
4
Initial calibration verification (ICV) Laboratory control spike (LCS) Sample preparation and analysis
A second source standard should be An LCS should be analyzed per For liquid samples, measure 2 mL into
analyzed to verify the accuracy of the preparatory batch. The concentration a headspace vial along with 2 mL of
calibration. A certified reference material of the LCS should be greater than LOQ 1 ppm dioxane-d8 internal standard and
(CRM) containing 2,000 μg/mL of and less than the midlevel calibration. mix well.
1,4-dioxane (Sigma-Aldrich part number A spiked water sample at 0.5 ppm For semisolid or solid samples, measure
CRM48367) is used. The analyst may concentration is prepared and analyzed between 0.1 to 2 g depending on the
prepare any one (or all) of the following as follows: matrix into a headspace vial. It may be
ICV standards for analysis. Transfer 100 µL of 1,4-dioxane standard necessary to add water to the sample
– ICV at 10 ppm: Transfer 10 µL of at 10 ppm to 1,900 µL of Milli-Q water. to form a slurry if the sample is thick.
CRM stock to a headspace vial Add 2 mL of 1 ppm dioxane-d8 internal Add 2 mL of 1 ppm dioxane-d8 internal
containing 1990 µL of Milli-Q water. standard and mix well. standard to the sample and mix well
Add 2 mL of 1 ppm dioxane-d8 prior to analysis.
internal standard and mix well. Matrix spike and matrix
spike duplicate
– ICV at 1 ppm: Dilute the CRM stock Results and discussion
A matrix spike should be performed one
to 20 µg/mL by combining 10 µL of
per preparatory batch. The following 7890 GC system with 7694E
stock with 990 µL of Milli-Q water
procedure describes the spiking of a headspace sampler
in a 2 mL autosampler vial Transfer
sample with 1 ppm of 1,4-dioxane. A seven-point calibration curve was used
100 µL of CRM at 20 µg/mL to a
headspace vial containing 1,900 µL Combine 100 µL of the 1,000 ppm for quantification in the range of 13 ppb
of Milli-Q water. Add 2 mL of 1 ppm 1,4-dioxane standard with 900 µL of to 10 ppm. Based on five calibration
dioxane-d8 internal standard and Milli-Q water in a 2 mL ASV to prepare curves (spanning three years), the R2
mix well. a 100 ppm 1,4-dioxane standard. Next, value was greater than 0.997 in each
spike 2 mL of a sample with 20 µL of the calibration. The average response factor
– ICV at 0.1 ppm: Dilute the CRM stock
100 ppm standard in a headspace vial. of the curves was 1.13 with a relative
to 20 µg/mL by combining 10 µL of
Add 2 mL of 1 ppm dioxane-d8 internal standard deviation (RSD) of 11.6%.
stock with 990 µL of Milli-Q water
standard and mix well. Calibration data are shown in Figure 2.
in a 2 mL ASV. Transfer 10 µL of
CRM at 20 µg/mL to a headspace Prepare a duplicate spiked sample
vial containing 1,990 µL of Milli-Q following the same procedure
water. Add 2 mL of 1 ppm dioxane-d8 for analysis.
internal standard and mix well.
12
Continuing calibration verification
y = 1.0115x
A midlevel calibration standard 10 R² = 0.9996
(standard 4) should be injected after
every 10 samples and at the end of 8
Response ratio
the sequence. 6
Method blank 4
A method blank must be analyzed with
each preparatory batch of samples. 2
and analyzed. Figure 2. Calibration curve for 1,4-dioxane analysis on the Agilent 7890 GC system
with Agilent 7694E headspace sampler.
5
The initial calibration was verified Table 5. Matrix spike and matrix spike duplicate results for 1,4-dioxane analysis on the
with the use of a certified reference Agilent 7890 GC system with Agilent 7694E headspace sampler.
material that was diluted and analyzed Type Concentration (ppm) Percent Recovery Percent Deviation
at 10, 1, and 0.1 ppm. In each case, the
Sample 1.69
percent difference was less than 10%.
Matrix Spike 2.63 94
For continuing calibration verification,
Matrix Spike Duplicate 2.58 90 1.6%
a midlevel calibration standard
(standard 4) was repeatedly injected
over a period of five days from the initial
calibration. The percent difference was
less than 13.5%. ×102
3
Quality control for this method was
Counts
chromatogram is shown in Figure 3. The 1,4-Dioxane 2.115 88, 58 50 ppb 49 ppb 0.0025 7.1 ppb 13 ppb
6
Intuvo 9000 GC system with 7697A
headspace sampler
Method parameters from the 7890 GC
system with 7694E headspace sampler
can be transferred to the Intuvo 9000 GC
system with 7697A headspace sampler.
The conditions for the 7697A headspace
sampler need minimal optimization.
The 7697A has onboard pneumatics
control to achieve active backpressure
control. This difference changes the loop
fill behavior and requires new method
parameters. The use of the method
conversion wizard (Figure 4) will give
recommended starting conditions, which
then can be optimized.14
A nine-point calibration curve was used
for quantification in the range of 6.5 ppb
to 20 ppm. Based on five calibration
curves (spanning nine days), the R2 value
was greater than 0.9979, as shown in Figure 4. Screen capture from the method conversion wizard for the Agilent 7697A headspace sampler.
Figure 5. Average response factor of the
curves was 1.1 with an RSD of 5.44%.
×101 1,4-Dioxane ×104
The initial calibration was verified 2.2 y = 1.119808x 6.8
with the use of a certified reference 2.0 R2 = 0.9980 6.6
1.8 Avg. RF RSD = 5.443341 6.4
material that was diluted and analyzed 1.6
6.2
Relative responses
6.0
at 10, 1, and 0.1 ppm. In each case, the
ISTD responses
1.4 5.8
5.6
percent difference was less than 10%. 1.2
5.4
1.0
For continuing calibration verification, 5.2
0.8 5.0
a midlevel calibration standard 0.6 4.8
4.6
(standard 4) was repeatedly injected 0.4 4.4
over a period of nine days from the initial 0.2 4.2
0 4.0
calibration, and the percent difference 3.8
was less than 10%. -1 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Relative concentration
Quality control for this method was also
monitored throughout data collection. Figure 5. Calibration curve for 1,4-dioxane analysis on the Agilent Intuvo 9000 GC system with an
Agilent 7697A headspace sampler.
Method blanks yielded nondetectable
levels, proving there was no system
carry over. Three laboratory control Table 7. Matrix spike and matrix spike duplicate results for 1,4-dioxane
spikes were analyzed, and the accuracy analysis on the Agilent Intuvo 9000 GC system with Agilent 7697A
ranged from 90 to 110%. A matrix spike headspace sampler.
and matrix spike duplicate were also
Type Concentration (ppm) Percent Recovery
analyzed. The matrix spike samples
Sample 0.0
were analyzed after spiking with a known
Matrix Spike 0.994 94.2
amount (1 ppm) of 1,4-dioxane. The
Matrix Spike Duplicate 0.995 95.4
results are shown in Table 7.
7
The MDL for 1,4-dioxane was
calculated based on EPA methodology
(EPA 821‑R‑16-006). The MDL was
determined by spiking a detergent
(predetermined to contain nondetectable
levels of 1,4-dioxane) at a concentration
of 50 ppb 1,4-dioxane. Thirty replicates
of the spiked sample were injected over
nine days, representative data are shown
in Figure 6. The MDL was determined to
be 2.3 ppb (Table 8).
Figure 6. Example chromatography from one of the 30 spiked sample injections used for method
detection limit (MDL) calculation of 1,4-dioxane on the Agilent Intuvo 9000 GC system with an
Agilent 7697A headspace sampler.
Table 8. Method detection limit (MDL) for the analysis of 1,4-dioxane on the Agilent Intuvo 9000 GC
system with an Agilent 7697A headspace sampler.
Best practices
Replace reference materials when responses do not pass criteria, are low compared to past calibrations, or reach their expiration date.
Recalibrate when the continuing calibration verification no longer passes within 20% of true value, or when maintenance has been performed.
8
Conclusion References 9. Work Plan Implementation:
1,4-Dioxane in Personal Care and
These methods present a sensitive, 1. 1,4-Dioxane. Environmental Working Cleaning Products. California
robust, and selective method to Group 2008. Department of Toxic Substances
determine 1,4-dioxane in consumer 2. Spath, D. 1,4-Dioxane Action Level. Control 2019.
products including: cosmetic, personal California Office of Environmental 10. Kaminsky, T. An Act to Amend the
care, and cleaning products. The Health Hazard Assessment 1998. Environmental Conservation Law, in
benefits of using the Agilent 5977 MSD
3. Integrated Risk Information System: Relation to Prohibiting Household
SIM capabilities with the headspace
Chemical Assessment Summary Cleansing Products, Cosmetic
sampler cannot be underestimated in
for 1,4 Dioxane (CASRN 123-91-1). Products and Personal Care
reducing sample matrix interference
United States Environmental Products that Contain 1,4-Dioxane.
and improving signal-to-noise. This
Protection Agency 2008. The New York State Senate, bill
instrument configuration provides
S4389B, 2020.
high selectivity and sensitivity with a 4. Report on Carcinogens, 11th Edition.
more confidence-driven solution for National Toxicology Program, U.S. 11. 1,4-Dioxane. Wikipedia.org 2021.
the analysis of 1,4-dioxane. Lower Department of Health and Human 12. Giri, A.; Zelinkova, Z.; Wenzl, T.
detection limits cannot be achieved due Services 2008. Experimental Design-Based Isotope-
to the myriad of interferences in the gel, 5. Chemicals known to the state to Dilution SPME-GC/MS Method
viscous, and foaming mixtures. One cause cancer or reproductive toxicity. Development for the Analysis of
approach to reduce matrix interferences California Office of Environmental Smoke Flavoring Products. Food
and achieve lower detection limits Health Hazard Assessment 2008. Additives & Contaminants: Part A
would be the use of a triple quadrupole 2017, (34)12, 2069–2084.
6. Natural Consumer Products Found
GC/MS system. 13. Munch, J.; Grimmett, P. Method
Contaminated with Cancer-Causing
1,4-Dioxane in Groundbreaking 522 - Determination of 1,4-Dioxane
Analysis Released by OCA. Organic in Drinking Water by Solid
Consumers Association 2008. Phase Extraction (SPE) and Gas
Chromatography Mass Spectrometry
7. Black, R.; Hurley, F.; Havery, D.
(GC/MS) with Selected Ion
Occurrence of 1,4-Dioxane in
Monitoring (SIM). United States
Cosmetic Raw Materials and
Environmental Protection Agency
Finished Cosmetic Products. Journal
2008.
of AOAC International 2001, 84(3),
666–670. 14. System Parameter and Performance
Comparison Between Agilent 7697A
8. Tahara, M.; Obama, T.; Ikarashi, Y.
and Agilent G1888A Headspace
Development of analytical method
Samplers for USP <467>. Agilent
for determination of 1,4-dioxane
Technologies technical overview,
in cleaning products. International
publication number 5991-5182EN,
Journal of Cosmetic Science 2013,
2014.
35(6), 575–580.
9
www.agilent.com/chem
DE44279.723407778