P R E P A R I N GY O U RL A B

APPLICATION NOTE

PinAAcle Atomic
500/900 AA Absorption
Spectrometers Preparation Checklist
Suitable working area
Exhaust vent
Atomic absorption gases
Handling of gas cylinders and safety practices
Drain vessels
Graphite furnace and flame requirements
Facilities requirements
Important accessories and consumables

This document provides detailed instructions regarding the space, accessories and utilities
required to operate PerkinElmer’s PinAAcle™ 500 and 900 family of atomic absorption (AA)
spectrometers, THGA and HGA graphite furnace systems, and other major AA accessories.
PerkinElmer AA instruments are complete systems with the exception of the following items
which must be provided by the analyst: suitable working area; exhaust vent(s); gases and
regulators; and light sources. When a THGA or HGA graphite furnace system is being used,
additional items may be required, which are described in Section 8. General information on
each of the required items is given in the appropriate section of this document.

PinAAcle 500 Flame AA Spectrometer PinAAcle 900 Series AA Spectrometers

1 Suitable Working Area The environment in which any instrument is housed is


an important consideration. The room temperature
should be between 15 and 35 °C (59-95 °F) with a
WARNING: The use of atomic absorption instruments
maximum rate of change of 3 °C (5 °F) per hour. For
without adequate ventilation to outside air may
optimum instrument performance, the room
constitute a health hazard. For example, the combustion
temperature should be controlled at 20 +/-2 °C. The
of halogenated hydrocarbons produces toxic vapors.
instrument should be located away from
Extreme care should
direct sources of heat or cold. The relative humidity be taken that exhaust gases are vented properly.
should be between 20 and 80%, non-condensing.
FIAS (flow injection for atomic spectroscopy) systems and
A relatively dust-free environment is necessary. This is
flame autosamplers should be placed on a cart or table close
especially important when working with ultra-trace
to the AA spectrometer sample compartment to keep FIAS and/or
techniques, such as graphite furnace sampling. Other
autosampler tubing to a minimum length. The FIAS system
important considerations are to locate the instrument in
and/or autosampler can also be placed to the side of the
an area free of corrosive fumes and
instrument – however, maximum performance may be
vibration and in an area that does not receive direct
compromised due to the increased length of tubing required.
sunlight. This precision instrumentation is not designed to
With self aspiration of the nebulizer, the uptake rate is greatly
withstand vibrations from earthquakes. The instrument
affected by the length of the tubing.
should be located away from radiators and at least 6 feet
away from any A/C or heating ducts. 2 Exhaust Vent
AA spectrometers are normally installed on laboratory work A venting system is required to remove the combustion fumes
benches or tables. The benchtop or area in which the AA and vapors from the flame or graphite furnace for AA instruments.
instrument is to be installed should be large enough to Exhaust venting is important for a number of reasons:
accommodate the instrument and all accessories (dimensions of
• It will protect laboratory personnel from toxic vapors
those items are given in Section 10). A custom PinAAcle heavy-
which may be produced by some samples.
duty rolling bench, with locking casters and a lower shelf for
storage, can be ordered (Part No. • It will tend to remove the effects of room drafts
N0777783). Make sure that there is space at the rear and sides and the laboratory atmosphere on flame stability.
of the system for air to circulate freely. Do not block the fan • It will help to protect the instrument from corrosive
operating on both sides of the instrument. The back of the vapors which may originate from the samples.
instrument should not be placed permanently against a wall,
• It will remove dissipated heat which is produced by the
as the instrument must be accessible from the back for
flame or furnace.
servicing purposes. An accessible space of at least 60 cm (24
in.) should be available behind the instrument. If this is not A flow rate of approximately 5700-8500 L/min (200-300 cubic
possible, the table or bench on which the instrument is feet/ min) is required for air/acetylene flame and 1000 L (35
mounted should be on wheels so that it can be easily moved. cubic feet/ min) is recommended for furnace operation. A
The vent drop should be flexible so that it can be moved out flow rate of approximately 7000-8500 L/min (250-300 cubic
of the way when the instrument is serviced. It is much easier feet/min) is required for nitrous oxide/acetylene flame
to install and service an instrument that is on a rolling bench. operation. Though we do not specify a maximum flow rate,
more than 20% above the recommended flow rate of 300
A means of electrically grounding the instrument and all
cubic feet/min will also affect stability of the flame, thus
accessories must be available.
the precision of measurement. It is strongly recommended
Atomic Absorption System Layout that the instrument not be placed in a chemical hood! If a
In addition to the basic requirements for atomic chemical hood must be used, arrangements should be made
spectroscopy systems, preparation of your laboratory for to keep out corrosive vapors and backdrafts from other
atomic absorption (AA) spectrometers equipped with hoods.
graphite furnaces requires that consideration be given to
Sample preparation should not be carried out in the same
the installation needs of the furnace power supply.
hood where the instrument is located.

Figure 1. Blower and vent kit dimensions.


2
PerkinElmer offers an accessory, Blower and Vent Kit (Part When installing such a venting system, all connections should be made
No. 03030448 for 230 VAC; Part No. 03030447 for 115 VAC), with metal screws or rivets. Solder must not be used. The blower
which will fulfill the exhaust requirements for all AA should be located at least 4 meters (12 feet) and not more than 6.5
instruments (see Figure 1). The blower capacity is for one meters (20 feet) from the flame or the graphite furnace and should
instrument. Included in the kit are a rotary blower with exhaust to the atmosphere or into a considerably wider exhaust duct.
capacitor and hardware, a vent Under these conditions, the following temperatures
to be located above the instrument, and an adapter to
permit connection of the blower and vent with suitable
metal tubing. The adapter and vent are made of stainless
steel sheets.
NOTICE: Local electrical codes do not permit PerkinElmer
Service Engineers to install the blower and vent assembly.
The metal tubing required to connect the vent to the blower
and to provide suitable exhaust from the blower is not
included in the Blower and Vent Kit. Flexible stainless steel
tubing is recommended for this purpose and can be obtained
from the companies listed
in Table 1 and from various other firms. In some instances,
this type of flexible metal tubing is sold only in minimum
lengths of 3 meters (10 feet).
Table 1. U.S. Suppliers of Flexible Metal Duct Tubing.
Flexaust Co. Potomac Rubber
11 Chestnut St. Co. 9011 Hampton
Amesbury, MA 01913 Overlook Capital
(508) 388-9700 Heights, MD 20743
(301) 336-7400
Triplex Inc. Fox Manufacturing Inc.
1142 Kress St. P.O. Box 1047
Houston, TX 77020 Clarkdale, AZ 86324
(713) 672-7521 (602) 634-5897

For the names of suppliers in other areas, contact


your PerkinElmer representative.
The capacity of the blower depends on the duct length and
the number of elbows or bends used to install the system. If
an
excessively long duct system or a system with many bends is
used, a stronger blower may be necessary to provide
sufficient exhaust volume. Alternatively, smooth stainless
steel tubing may be used instead of flexible stainless steel
tubing where flexibility is not required to reduce system
friction loss or “drag.” A length of smooth stainless steel
ducting has 20-30% less friction loss than
a comparable length of flexible ducting. When smooth
stainless steel tubing is used, elbows must be used to turn
corners. These elbows should turn at a center line radius of 45
degrees to reduce friction losses, and the number of elbows
should be minimized.
If a rigid tubing system is used, it is strongly recommended
that flexible tubing be used from the vent hood to the ceiling
to facilitate hood alignment and service access to the
instrument. The exhaust vent cannot prevent removal of the
instrument covers.
The dimensions for the various parts of the Blower and Vent
Kit are shown in Figure 1. The vent i.d. is slightly larger
than the tubing o.d. to allow for tubing tolerances. A slight
gap between the two units is normal.

3
have been measured during operation of a nitrous For flame operation, the air supply should provide a
oxide-acetylene flame: 310 °C at the vent intake; minimum of 28 liters/min (1 cubic foot/min) at a minimum
160 °C at 2.4 meters (8 feet) from the vent intake; pressure of 300 kPa (43.5 psig). The pneumatics of the
105 °C at the blower intake; and 50 °C at the PinAAcle 500, 900F, 900H and 900T systems require an air
blower motor housing near the front bearing. pressure working range of 300-400 kPa (43.5-58 psig). It is
recommended to have a line pressure gauge installed in the
Instructions for installation (Part No. 09936775) lab in close proximity to the instrument, if the tank is not in
are provided with the Blower and Vent assembly. close proximity to the instrument.
The blower provided in the
PerkinElmer accessory kit requires a line voltage of It is desirable to have a water and oil trap or filter between
the compressor and the instrument gas control system. The
115 V or 230 V. Additional recommendations on the use of an Air Filter Accessory (Part No. N0775325) or an
venting system include: Air/Acetylene Filter Accessory (Part No. N9301398) is
• Make sure the duct casing is installed using fireproof strongly recommended to remove entrained water, oil, water
construction. Route ducts away from sprinkler heads. aerosols and solid particles

• Locate the discharge outlet as close to the from compressed air lines. ventilation.
blower as possible. All joints on the discharge Filter elements need to be Cylinders of
side should be airtight, especially if toxic vapors replaced every 6-12 compressed air can
are being carried. months, provided the also be used but are
compressed air supply is recommended only as
• Equip the outlet end of the system with a
properly maintained. an emergency or
backdraft damper and take the necessary
precautions to keep the exhaust outlet away from WARNING: When using short-term solution for
open windows or inlet vents and to extend it an air or air/acetylene the following reasons:
above the roof of the building for proper dispersal accessory, never exceed • A standard #1 size air
of the exhaust. 1050 kPa (150 psig) for cylinder contains about
the air input pressure. 6200 liters (220 cubic
• Equip the exhaust end of the system with an
exhaust stack to improve the overall efficiency If there is any doubt feet) of air at standard
of the system. regarding the usability of temperature and
a central air supply pressure
• Make sure the length of the duct that enters into
(insufficient volume or (STP). A premix burner-
the blower is a straight length at least ten times
pressure or excessive oil nebulizer system uses about
the duct diameter. An elbow entrance into the
or water 20 liters/min (0.7 cubic
blower inlet causes a loss in efficiency.
contamination), the quality feet/min), and, therefore,
• Design local exhaust ventilation systems of the supply should be a cylinder will last only
individually for each specific AA instrument. checked prior to the arrival about five hours. Unless an
Also, the opening of the exhaust vent should be of the instrument. A small, instrument is used only a
large enough to cover the graphite furnace or oil-less air compressor few
flame area completely. meeting the stated hours per day, changing
• Provide make-up air in the same quantity as is requirements is available cylinders becomes a
exhausted by the system. An “airtight” lab from PerkinElmer (AA Ultra nuisance as well as being
will cause an efficiency loss in the exhaust Quiet Oil-Less Air expensive.
system. Compressor 115V/60HZ - • Occasionally, cylinder air
• Ensure that the system is drawing properly Part No. may be obtained from a
by using an anemometer (Part No. N3134065) N9306345; AA Ultra liquefaction process
to accurately measure the exhaust flow rate. Quiet Oil-Less Air during which the oxygen-
Compressor 220V/50- 60 to-nitrogen ratio can
• Equip the blower with a pilot light located near
Hz - Part No. N9306346). change. Therefore, it is
the instrument to indicate to the operator when
the blower is on. Air compressors are not uncommon to find
generally uncomfortably other than 20% oxygen in
3 Atomic Absorption Gases noisy to have in the air cylinders. This can
NOTICE: Standards for cylinder dimensions, regulator immediate vicinity of cause erratic burner
connections, gas fittings, etc. vary from country to the instrument. operation and non-
country. The information provided here is for the U.S. Whenever possible, it is reproducible analytical
Contact your PerkinElmer representative for advisable to locate them results and, in extreme
information on the specific requirements of your at some distance from cases, may provide a
area. laboratory workers in an potential safety hazard.
area providing suitable In general, if cylinder air
Compressed Air
4
is to be used, it is air/acetylene flames, Using a heat-combustion requiring either CGA 300 or
important to specify the acetylene flow used value of 1,450 BTU per CGA 510 fittings as well
compressed air is about 4 liters/min or cubic foot, the heat given as a connector for attaching
rather than 0.14 cubic feet/min. off would be the fuel hose assembly
breathing air (i.e. approximately 12,300 BTU supplied with the
medical grade) or an per hour (3,600 W). An instrument. The fuel hose
unspecified form. air/acetylene flame can assembly is constructed of
WARNING: For be used with all red neoprene, reinforced
safe operation, PerkinElmer burner heads with high tensile-strength
oxygen must but is most commonly rayon cord, and provides a
NEVER be used used with the supplied 10 rated working pressure of
with PerkinElmer cm (4-inch) burner head. about 1700 kPa (250 psig).
premix burner Suitable acetylene typically The connectors are
systems. has a minimum purity permanently mounted at
specification of 99.6% with each end of the hose
The use of air cylinders
the actual assay being about assembly for connection to
requires the use of a
99.8%. In general, ordinary the pressure regulator and
suitable dual-stage
welding-grade acetylene is instrument gas controls, and
regulator. A regulator
adequate for most AA use left-hand threads as per
for cylinders with a CGA
analyses, though sometimes a accepted practice for fuel
590 connection is
particular tank may be gas connections. (See
available from
contaminated. Special higher Section 5 for more details.)
PerkinElmer (Part No.
purity “atomic absorption” The rigorous gas hose testing
03030264).
grade acetylene is also is negated if a fitting is
NOTICE: In order to removed and replaced with
available from some
minimize flame ignition another fitting.
vendors, and its use is
problems, the following
recommended when the It may be desirable to have
size piping is
available welding-grade an acetylene filter between
recommended for long
acetylene is not sufficiently the acetylene tank and the
distances of the air
pure. instrument gas control
source to the laboratory
system to remove particulates
where it is used in our A size 1A acetylene
and acetone droplets from
instrumentation: cylinder contains about
acetylene, protecting the gas
• > 100 feet < 300 8,500 liters (300 cubic feet)
controls and AA burner
feet - 3/8" I.D. of acetylene and usually
copper or stainless system from contamination
lasts about 30 hours of
steel; and corrosion. An Acetylene
burning time with an
• > 100 meters - 1/2" Filter (Part No. N9301399)
air/acetylene flame. The
I.D. copper or and an Air/Acetylene Filter
cylinder requires an
stainless steel. Accessory (Part No.
acetylene pressure
Acetylene N9301398) are available from
regulator, which can be
For the overwhelming PerkinElmer.
obtained from the supplier
majority of analyses, Some countries also require
of the gas or from
acetylene is the the use of a flashback
PerkinElmer (Part No.
preferred fuel gas with arrestor, such as
03030106).
AA spectrometers. PerkinElmer Part No.
CAUTION: Acetylene may N9300068, in the acetylene
Air/acetylene is the
react with copper to fuel line.
preferred flame for
form a potentially
the determination of Acetylene is normally
explosive compound.
about 35 elements by supplied dissolved in
Copper tubing or fittings
AA. All PerkinElmer AA acetone, and a small amount
for acetylene gas must be
flame instruments of acetone carryover with the
strictly avoided.
require acetylene for acetylene is normal.
the fuel. The The PerkinElmer Acetylene However, as tank pressure
temperature of the Regulator Assembly includes falls, the relative amount of
air/acetylene flame is an adapter so that the acetone entering the gas
approximately 2300 pressure regulator can be stream increases and can give
°C. For most connected to cylinders erratic results, particularly for

5
elements such as cylinder pressure drops source to the laboratory where (acetylene). Unless
calcium, tin, chromium, below 600 kPa (85 psig) it is used in our separated by a minimum of
molybdenum and others may cause damage to instrumentation: 20 feet (6.1 meters), there
whose sensitivity is valves or tubing within • > 100 feet < 300 feet should be a fire-resistive
highly dependent on the burner gas control - 3/8" I.D. stainless partition between the
the fuel/oxidant ratio. system due to high steel - do not use nitrous oxide cylinders and
For this reason, acetone carryover. copper; the flammable gas
acetylene tanks should Such damage from • > 100 meters - 1/2" I.D. cylinders. The use of
be replaced when the acetone is not covered stainless steel - do not use nitrous oxide requires a
cylinder pressure drops by instrument copper. number of accessories and
to about 600 kPa (85 warranties. Improper precautions. A size 1A
Nitrous Oxide
psig). storage of an acetylene cylinder of nitrous oxide
The nitrous oxide/acetylene
IMPORTANT: Failure to tank could also cause contains about 14,800
flame has a maximum
change the acetylene acetone to enter the liters (520 cubic feet) and
temperature of
cylinder before the burner gas control will typically last for 10 to
approximately 2800 °C and
system. 12 hours of burning time.
is used for the
Since the acetylene is cylinder valve, it is a potentially determination of elements Cylinders of nitrous oxide
dissolved in acetone, the hazardous practice and one that form refractory oxides. (99.0% minimum purity) are
pressure drop is not linear which should never be It is also used to overcome available from local
with gas removal, and a attempted in the presence of chemical interferences suppliers. A dual-stage
pressure of 600 kPa (85 an open flame, sparks or other that may occur in flames regulator is recommended
psig) indicates that the possible sources of ignition. of lower (and is mandatory in some
cylinder is nearly empty, CAUTION: Acetylene-line temperatures. For the nitrous countries).
assuming the cylinder is pressure from the cylinder to oxide/acetylene flame, the Nitrous oxide is supplied
at room temperature. the instrument should never acetylene flow is about 14 in the liquid state,
Acetylene tanks should be allowed to exceed 103 kPa liters/min or 0.5 cubic feet initially at a pressure of
always be stored and (15 psig). At higher pressures, per min. Using a heat of about 5000 kPa (750 psig).
operated in a vertical acetylene can spontaneously combustion value of 1,450 Since nitrous oxide is in
position, rather than decompose or explode. BTU per cubic foot, the heat liquid form, the pressure
horizontally, to prevent PerkinElmer recommends that a given off would be gauge does not give a
liquid acetone from maximum acetylene line approximately 43,000 BTU per true indication of how
reaching the cylinder pressure of 80-95 kPa (12-14 hour (12,500 W). For much nitrous oxide
valve. Acetylene gas psig) be used to provide a N2O flame operation, the remains in the cylinder
reasonable margin of safety. It nitrous oxide supply should until the pressure starts
cylinders should not be
provide a minimum pressure
stored with oxygen or is recommended to have a line of 300 kPa (43.5 psig). The to
nitrous oxide cylinders or pressure gauge installed in the pneumatics of the fall rapidly as the residual
adjacent to oxygen- lab in close proximity to the PinAAcle 500, 900F, 900H and gas is drawn off. There is
charging facilities. New instrument, if the tank is not 900T systems require a nitrous no drop in pressure until all
tanks should be in close proximity to the oxide pressure working range of the liquid nitrous oxide
positioned vertically for instrument. of 300-400 kPa (43.5-58 psig). is consumed. The tank is
at least 8 hours prior to Cylinder valves should be It is recommended to have a basically empty when all of
use. Storage of acetylene closed to avoid the possibility line pressure gauge installed the liquid nitrous oxide is
tanks should be at a of pressure regulators failing in the lab in close proximity consumed.
location above 7 °C (45 and gas lines being subjected to the instrument, if the tank When nitrous oxide is
°F) and away from direct to the full cylinder pressure. is not in close proximity to the rapidly removed from the
sunlight. The practice of Both fuel and oxidant gas instrument. cylinder, the expanding gas
“cracking the valve” of an lines should be relieved of causes cooling of the
Cylinders of nitrous oxide
acetylene tank (that is, pressure at the end of the cylinder pressure regulator
stored inside buildings
opening the valve slightly working day or if the and the regulator
should not be located in
for a very short period instrument is diaphragm sometimes
close proximity to cylinders
prior to attaching the to remain unused for an freezes. This can create
of flammable gases
regulator) is not extended period. erratic flame conditions or,
recommended. While such in the most extreme case,
NOTICE: In order to minimize
an action will clear the a flashback. It is therefore
flame ignition problems, the
valve opening of dust or advisable to heat the
following size piping is
dirt particles and may regulator using either a
recommended for long
remove acetone from the built-in heater or an
distances of the acetylene

6
externally connection to the regulator which will keep the liquid argon cool
supplied heat instrument gas vent the tank as enough to remain in the
source, such as controls (see Section necessary in order to liquid state. Gaseous
an electrical 5). argon tanks do not require be stored with oxygen,
resistance heating NOTICE: In order to venting and consequently or nitrous oxide
tape. minimize flame ignition can be stored for extended cylinders, or adjacent to
CAUTION: All problems, the following periods without loss. oxygen-charging
lines carrying size piping is A dual-stage cylinder facilities.
nitrous oxide recommended for long regulator that can be used • Cylinders of nitrous
should be free distances of the with either gaseous argon or oxide stored inside
of grease, oil nitrous oxide source to nitrogen is available from buildings should not be
or other the laboratory where PerkinElmer (Part No. located in close
organic it is used in our 03030284). The regulator proximity to cylinders of
material, as it instrumentation: has a CGA 580 fitting, and flammable gases
is possible for • > 100 feet < 300 feet - 3/8" includes a color-coded hose (acetylene). Unless
spontaneous I.D. copper or stainless with 1/4-inch Swagelok® separated by a minimum
combustion to steel;
fittings to permit direct of 20 feet, there shall be
occur. Cylinders • > 100 meters - 1/2" I.D. connection to the regulator a fire-resistive partition
of nitrous oxide copper or stainless steel.
and to the instrument gas between the nitrous
should be Argon controls (see Section 5). oxide cylinders and the
considered high- Argon is required for flammable gas
pressure cylinders Table 2. Quality Criteria for
external and internal cylinders.
Argon.
and should be gas streams through
handled with care Criterion • Fasten all gas cylinders
the THGA or HGA
at all times. Purity securely to an immovable
graphite furnace or
Oxygen bulkhead or a permanent
A dual-stage as a carrier gas with
Nitrogen wall.
heated nitrous mercury/ hydride
Water
oxide pressure systems such as the • When gas cylinders are
regulator for use FIAS or FIMS flow stored in confined
injections systems or areas, such as a room,
with gas 4 Gas Line Connections
cylinders with a the MHS-15. The ventilation should be
PerkinElmer AA
CGA 326 quality criteria listed adequate to prevent
instruments include the
connection is in Table 2 for argon toxic or explosive
hoses necessary for
available from are suitable for these accumulations. Move or
connection to gas lines
PerkinElmer applications. store gas cylinders only
(see Table 3). It is the
(Part No. Normally, for in a vertical position
responsibility of the user
03030204 [115 graphite furnace with the valve cap in
to provide the appropriate
VAC] or operation, gaseous place.
gas lines, regulators,
03030349 [230 argon is used, • Locate gas cylinders away
connectors and valves to
VAC]). These although either liquid from heat or ignition
which the hoses are
regulators provide or gaseous argon can sources, including heat
connected.
pressure control be used. The choice lamps. Cylinders have a
from 350-520 kPa of liquid argon or 5 Handling of pressure-relief device
(50-75 psig) and gaseous argon tanks Gas Cylinders that will release the
contain an integral is determined and Other contents of the cylinder if
thermostatted primarily by the Safety the temperature exceeds
heater to prevent availability of each Practices 52 °C (125 °F).
freezing of the and the usage rate.
NOTICE: The • When storing cylinders
regulator Liquid argon is
permanent installation external to a building,
diaphragm. A usually less
of gas supplies is the the cylinders should
color-coded hose expensive per unit
responsibility of the be stored so that they
with suitable volume to purchase,
user and should are protected against
connectors at but cannot be stored
conform to local safety temperature extremes
each end is for extended periods.
and building codes. (including the direct
supplied with the If liquid argon is
• Flammable gas rays of the sun) and
regulators to used, the tank should
cylinders (such as should be stored
provide be fitted with an
acetylene) should not above ground on a
over-pressure
7
suitable floor. • Use only approved gas leak tests by cyanide gas may be
• Mark gas cylinders regulators and hose applying a soap released. Never allow an
clearly to identify connectors. Left- solution to all joints acidic solution to enter
the contents and hand thread fittings and seals. a drain bottle that
status (full, empty, marked with hatch WARNING: Contact contains a cyanide
etc.). marks are used for between acetylene gas solution.
fuel gas tank and copper or silver (or • Take suitable precautions
• Full cylinders should
not be stored with connections, high concentrations of when using volatile organic
empty cylinders. whereas right-hand silver salts), liquid solvents. A potentially
fittings are used for mercury or gaseous flammable organic vapor
• Do not attempt to
refill gas cylinders. oxidant and support chlorine can produce “cloud” can form around
gas connections. potentially unstable the opening of the sample
• Use galvanized iron acetylides. Always clean vessel. Feeding the
tubing, steel, the burner thoroughly capillary tubing through a
wrought iron or other after analyzing solutions small hole in a covered
tubing that will not with high silver, sample container is one
react chemically with copper, or mercury way of reducing the
acetylene. Never use concentrations, and possibility for ignition.
copper tubing with aspirate solution • Never view the flame,
acetylene. Joints continuously during the hollow cathode lamps
may be welded or analysis to prevent any (HCLs), electrodeless
made up of threaded residues from drying. discharge lamps (EDLs) or
or flanged fittings, • Periodically check deuterium background
typically stainless for the presence corrector lamps directly
steel, aluminum or of acetylene in without protective
brass composed of the laboratory eyewear. Potentially
less than 65% atmosphere, hazardous ultraviolet
copper. Rolled, especially near radiation may be emitted.
forged or cast steel the ceiling. Ordinary safety glasses will
or malleable iron in general provide sufficient
• When the equipment is
fittings may also be protection, but additional
turned off (for
used. Cast iron side shields will ensure a
example, at the end of
fittings cannot be further margin of safety.
the working day), close
used safely for Safety glasses will also
all gas cylinder valves
acetylene lines. provide mechanical
tightly at the tank.
• Arrange gas hoses protection for the eyes.
Bleed the remainder of
where they will the line to the • Never leave the flame
not be damaged atmosphere before the unattended.
or stepped on and exhaust fan (vent) is • Zeeman background-
where things will turned off. corrected AA
not be dropped on instruments generate
• When using premix
them. a strong magnetic
burners with cyanide
• Never run solutions, check the pH of field. People with
acetylene at the liquid trap and drain cardiac pacemakers
a pressure vessel. The pH of the are advised not to
higher than liquid should be greater operate or frequent
100 kPa (15 than 10. If the liquid is the vicinity of
psig). At even slightly acidic, Zeeman corrected
pressures highly toxic hydrogen instruments while
above this they are in operation.
level, 6 Drain Vessels the effluent from the AA burner
acetylene A specially-configured drain drain.
may vessel is supplied with all The drain vessel should NOT be
spontaneousl PerkinElmer AA instruments stored in an enclosed storage area.
y explode. with burner systems. That Rather, the drain vessel should be
• Perform periodic vessel must be used to gather stored in plain sight of the

8
operator, usually on the PerkinElmer Small Diameter Removing a lamp without
floor in front of the manufactures all of Lamp Adapter Kit, switching it off first may
instrument or on an open its hollow cathode (Part No. 03030870). damage the lamp
shelf underneath the and electrodeless permanently.
instrument table. discharge lamps. The
Table 3. Gas Line Connections (Note: Regulator, connector and fitting needs
The drain system should Lumina™ series of vary by country. For information on what is required in your area, consult
be checked regularly and hollow cathode lamps your local PerkinElmer Service Representative.)
replaced when are especially noted
Regulat Regulator Regulator
necessary. Follow the for spectral purity, Gas
or Part to Cylinder Connection to Gas Part
directions in the brightness, stability Number CGA No. Hose Assembly Numbe
instrument manuals and long life. Air 03030 590 1/4" Swagelok® 005705
regarding the proper Hollow cathode 264
placement of the drain Argon 03030 580 1/4" Swagelok® 005705
lamps (HCLs) are
284
tube in the drain vessel excellent for most
Nitrogen 03030 580 1/4" Swagelok® 005705
and the proper liquid elements; 284
level in the drain vessel. however, there are Nitrous 03030 326 1/4" N.P.T.* 004702
a number of Oxide 204
7 Atomic Absorption Acetylene 03030 510 or 9/16" L.H.T.** 005705
“difficult”
Source Lamps 106 300
elements for which
AA spectrometers require an improved light N.P.T. = Normal Pipe Thread, L.H.T. = Left-Hand Thread
different source lamps, source is * Supplied with 5/16" x 1/4" N.P.T. Swagelok ® Male connector Body (Part No. 09
depending on the desirable.
elements to be ** Supplied with Outlet Bushing (Part No. 09903031), 1/4" N.P.T. to 9/16" L.H.T
PerkinElmer System
determined and the PerkinElmer Air Compressor (Part No. N9306345 for 115 V/60 Hz; Part No. N930
2 electrodeless
instrument to be used. discharge lamps PerkinElmer Air Filter Assembly (Part No. N0580531) provides 1/4" Swagelok® inlet and outlet
Multielement lamps are (EDLs) provide to a furnace. PerkinElmer Air Dryer Filter Assembly (Part No. N0775325) provides 1/4" Sw
available for some improved PerkinElmer Air/Acetylene Filter Assembly (Part No. N9301398) provides 1/4" Swagelok® inlet a
elements, but most lamps performance in 9/16" LH (B size) outlet fittings for acetylene.
are constructed using a most instances. PerkinElmer Acetylene Filter (Part No. N9301399) provides 3/8" LH (A size) inlet and
single element to avoid EDLs are more
potential spectral intense than their
Also Available:
interferences and reduced 1. Part No. 09903032 Connector for joining two Part No. 00570559 fuel hose assemblie
corresponding
performance, especially HCLs. Most also 2. Part No. 09903898 Connector for joining two Part No. 00570567 air/argon hose assem
when using a graphite provide better 3. Part No. 09903196 Adapter, female 1/4" N.P.T. to male 1/4" Swagelok®
furnace. lamp life and 4. Part No. 09920223 Connector for joining two 00470258 nitrous oxide hose assemb
stability as well as 5.Part No. N3180520 Gas Hose Adapters Kit
better sensitivity.
EDLs do not require consumption capability of
8 Graphite Furnace and
a separate power Flame Requirements 800 VA. The operating range
supply as it is built of the electrical line should
Location
into the PinAAcle meet the required
The flame and furnace
systems. specifications of 85-253 VAC,
power supply is built into
A lamp mount or the PinAAcle 500 and 4A-8A, and 47-63 Hz for the
turret is supplied 900 spectrometers. PinAAcle 500 and 900F.
with all PerkinElmer Graphite furnaces require
Electrical Services
AA instruments and electrical power, cooling
The PinAAcle 500 and 900F
will accommodate all water and a supply of inert
flame-only systems require a
PerkinElmer HCLs or gas, normally argon. A
dedicated circuit and wiring
EDLs. Users who may minimum input voltage of
size selected according to
have lamps with 1.5 207 VAC is recommended
local regulations
inch diameters rather under load to enable the
which provide power
than the standard furnace to reach maximum
specifications that meet the
PerkinElmer 2 inch potential operating
following criteria: 100-230
diameter can adapt temperatures and heating
VAC +10/-15 VAC, 50 or 60
those lamps for use rate and, for some systems
Hz +/- 3 Hz, and single phase
in PerkinElmer lamp with Zeeman effect
alternating current with a
mounts with the background correction,
maximum power
9
proper magnetic field types of 3-wire circuits on the instrument line cord. will not fit into the IEC 60309
strength. For used in North America, The PinAAcle 900T and 900Z 250 VAC 32A receptacle, and
installations where the which provide power also ship standard with an the wires should not be cut
line voltage may drop adequate to run THGA IEC 60309 250 VAC 32A to fit the receptacle, as this
below this level, the use or HGA systems, are Receptacle Kit (Part No. will compromise power to
of a “buck boost”-type illustrated in Figure 3 W1036805), that the IEC plug the instrument and affect
transformer is (page 10). The PinAAcle inserts into. The PinAAcle heating of the furnace and
recommended to maintain 900H requires a 230 VAC 900H system is provided with instrument performance.
proper analytical (±10%), 50/60 Hz (+/- a 20-amp IEC 60309 16/20 A NOTICE: Do not replace the
operating conditions. 0.3 Hz), 20-amp line power connector plug on the IEC instrument power plugs
Please refer to the Buck- capable of delivering instrument line cord. The installed on PinAAcle
Boost Transformers table 10.1 KVA of peak PinAAcle 900H also ships instruments.
on page 12 for details. power. The operating standard with an IEC 309 250
Running a ZL furnace range of the electrical For all furnace systems, the
VAC 16/20A Receptacle Kit
below the minimum line should meet the electrical supply should
(Part No. N0770425), that
voltage can cause required specifications contain a “slow blow” circuit
the IEC plug inserts into. For
permanent damage to of 207-253 VAC, 20A, breaker capable of handling
all PinAAcle 900 systems, the
the power supply. and 47-63 Hz for the 300% of the rated current for
receptacle must be installed
An appropriately- PinAAcle 900H under periods of 3 seconds. Also,
by a licensed electrician.
rated female full load of furnace the AA spectrometer,
The 20-amp plug supplied
electrical connector atomization checked graphite furnace, Zeeman
with the PinAAcle 900H can
is required to during installation. magnet, computer and other
be adequately used for the
provide power for There should be a operation of the PinAAcle accessories should all be
the graphite dedicated circuit 900H, though the use of a connected to the same
furnace and flame allocated for each 30-amp plug provides electrical ground, and the
systems. installed instrument additional safety margin for power supply should be free
Please contact your local with no grounding the power surge experienced of transients in excess of 50
PerkinElmer Service loops. during the furnace V over the nominal voltage.
Engineer to determine The THGA furnace and atomization step when peak Additional Furnace
the appropriate power is used. This supplied Requirements
Zeeman magnet of the
connector for your 20-amp plug can be replaced A water supply is required
PinAAcle 900Z and
laboratory. with a 30-amp plug, if to cool the furnace quickly
PinAAcle 900T systems
desired. It is recommended to ambient temperature
NOTICE: For 120 VAC operate from a single,
that 8-gauge (6 mm2) wire after reaching high
power flame-only dedicated electrical
be used for the electrical atomization temperatures.
instrument supply of 230 VAC
supply for PinAAcle 900Z, The water supply should be
installations, the use (±10%), 30 amp, 50 or
and PinAAcle 900T systems, free of sediment, have a
of an isolation 60 Hz (±0.3 Hz),
and that the length of the pH between 6.5 and 7.5,
ground receptacle single phase, capable of
wiring (circuit breaker to and a hardness not greater
(Part No. N3151390) delivering 10.1 KVA of
instrument connection) not than 6 °dH or 100 ppm. A
is strongly peak power. The
exceed 20 meters (65 feet). maximum flow of 2.5
recommend. It operating range of the
Although 10-gauge (6 mm2) L/min (0.6 gal/min) is used
should be a electrical line should
wire can be used for the for the THGA,
dedicated circuit meet the required
electrical supply of the 1.5 liters/min (0.4 gal/min)
with a specifications of 207-253
PinAAcle 900H system, 8- for the HGA at a
direct run of all three VAC, 30A, and 47-63 Hz
gauge (10 mm2) wire is temperature between 20 °C
wires to the power for the PinAAcle 900T
recommended. If the length and 40 °C. As both the flow
panel. There should and 900Z under full load
of the wiring of the rate and water temperature
not be any grounding of furnace atomization
dedicated circuit from the affect the cooling rate, it is
loops on the checked during
power panel exceeds desirable to be able to vary
dedicated circuit. installation. The
20 meters/65 feet, the use of 6- the flow rate to compensate
Electrical supply PinAAcle 900Z and 900T
gauge (16 mm2) is required. If for variations in cooling
circuitry, circuit systems are provided
6-gauge (16 mm2) wire is water temperature.
breakers and wiring size with a 30-amp IEC
used, install a junction box A suitable recirculating
for the graphite 60309 32A power
as close as possible to the cooling system is included
furnaces should be connector plug
instrument, and then run 8- with a number of THGA
selected according to gauge (6 mm2) wire to the systems. With an HGA
local regulations. Three receptacle. 6-gauge (16 mm2) system, the use of an
1
optional Recirculating turns the cooling
10 approximate power
Cooling System is system on during
Fac consumption for AA
strongly recommended. furnace operation. The
iliti spectrometers and major
All PinAAcle 900 cooling system will turn
es accessories are given in
furnace systems (T, H, off after the furnace
Re Table 5.
or Z), have a plug on has been idle for 5
qui PerkinElmer instruments
the back of the minutes. Since the
red will normally operate
instrument that relay controls the
for within the specified
supplies power to the cooling system, the power
switch on the back of the Per operating ranges of the
cooling system. There
cooling kin electrical line defined in
is also a relay that
El the Electrical Services
system should be left in psig). Purity specifications for me section on page 8,
the "on" position. Do not the gases to be used with r unless otherwise noted.
allow the cooling water HGA systems are given in AA If the power line is
temperature to drop Section 3. Ventilation is Ins unstable, fluctuates in
below 20 °C, or have too required to remove tru frequency or is subject
fast of a flow rate. This potentially toxic or corrosive me to surges or sags,
will cause condensation gases which can be nts additional control of the
on the furnace windows generated by the samples. incoming power may be
The following
and/or furnace. required. A means of
figure and tables
Condensation on the 9 Training Courses electrically grounding
provide the
furnace will cause poor Training courses for users of the instruments and
dimensions and
tube and contact cylinder PerkinElmer AA instruments accessories must be
power
life. are given at various available. Power to the
requirements for
CAUTION: To prevent locations. The courses cover instrument should be
PerkinElmer AA
combustion of the basic principles and clean from excessive
instruments and
graphite tube at high applications of atomic high frequency noise.
major accessories.
temperatures, the absorption and detailed
Dimensions are The ANSI-IEEE C62.41*
furnace is purged with instruction in the use of the
defined in Figure recommends <10 VAC
argon. instruments and major
2. There must be normal mode (signal to
accessories.
It is not recommended to a ground) and <1/2 VAC
use nitrogen as the To gain the most benefit loading dock common mode**
furnace purge gas. Its use from the training course, it available for (neutral to ground). Can
may lead to reduced is strongly recommended instrument delivery. be verified with the use
sensitivity for some that the attendee should There must also be a of an oscilloscope or
elements, and it is also have operated the 36-inch door logic analyzer and a
possible for nitrogen to instrument for at least available for a Power Probe power
react with the graphite several weeks prior to the crated entry of the meter tool.
tube at temperatures course. instrument into the * American National Standards
laboratory, or a 32- Institute (ANSI) is a private,
above 2300 °C to form For additional non-profit organization that
cyanogen, a toxic gas. information on atomic inch door to administers and coordinates
spectroscopy training accommodate entry the U.S. voluntary standards.
When operating the HGA
courses, please contact of an uncrated * Institute of Electrical and
Graphite Furnace systems Electronics Engineers (IEEE) is a
your local sales instrument. Internet professional association with its
at high temperatures, do
representative. access in the corporate office in New York
not look directly at the City.
laboratory where the
tube without suitable eye ** Excessive common mode
instrument will be
protection. (neutral to ground) noise can
installed is be caused by a poor building
Maximum gas preferred. Required ground. The NEC (National
Electrical Code) requires that
consumption is 0.7 services are shown in the building ground resistance
liters/min (0.19 cubic Table 4, and product does not exceed 25 ohms. This
feet/min) for THGA can be verified with an earth
dimensions and
ground test.
furnaces and 1.2
liters/min (0.3 cubic
feet/min) for
the HGA furnaces, both
at 300-450 kPa (44-66

1
A = width
B = height to top of cover
C = depth
D = distance from center of atomizer compartment to right hand edge of instrument
E = protuberance of sample tray in front of instrument
F = height of flame shield above top of instrument cover

G = 10 cm (4 inches) in all cases H = 30 cm (12 inches) in all cases J = 15 cm (6 inches) in all cases

Figure 2. Spectrometer and accessories dimensions (Refer to Table 5).

1
Table 4. Required Services.
Ga Cool Compute Req. Electrica
Instrument/ ses ing r No. l
Accessory
Air N 2O C 2H 2 Ar Wa Controlle of Power
ter d Vents
PinAAcle 500 • • • • 1 a
PinAAcle 900F • • • • 1 a
PinAAcle 900Z • e • 1 b

PinAAcle 900H • • • • c • 1 b
PinAAcle 900T • • • • e • 1 b
FIAS 100/400 • f d a

FIMS 100/400 • • d a
Amalgamation • • f a
Attachment
MHS-15 • d

a = 85-253 VAC, 4-8 A, 47-63 Hz, single phase The FIMS 100/400 does not require a vent if only analyzing
mercury and the b = 207-253 VAC, 30 A, 47-63 Hz (under full instrument load), single phase charcoal filter is used.

c = 207-253 VAC, 20 A, 47-63 Hz (under full instrument load), single phase The FIAS, FIMS (when used to analyze for hydrides), and MHS-15
units will use the vent that is over the instrument they are connected to.
d = uses the same vent as the spectrometer
The FIMS
does not require a vent when used to analyze for mercury, since the
units
e = optional (included with some versions of the PinAAcle 900H/T/Z) are supplied with a mercury absorbing filter. If the
amalgamation unit f = controlled from the spectrometer computer/controller is attached to the FIAS, then it will use the instrument
vent. If the
amalgamation unit is attached to the FIMS, then it will use
the mercury absorbing filter supplied with the FIMS.

Figure 3. Furnace power circuit connection configurations.


Table 5. Instrument and Accessory Specifications (Refer to Figure 2).

Width Height Depth (c ( ( Po Weight


Instrument/Accessor
(cm (cm (cm m c c w (Kg)
y
)A )B )C ) m m er
D ) ) V
E F A
Spectrometers:
PinAAcle 500 67 64 63 26 19.5 6.6 800 57
PinAAcle 900F 95 64 68 66 19.5 8.5 800 94
PinAAcle 900Z 95 64 68 66 19.5 8.5 10100 127
PinAAcle 900H 95 64 68 66 19.5 8.5 10100 123
PinAAcle 900T 95 64 68 66 19.5 8.5 10100 141

Graphite Furnace Components:


AS 900 Autosampler 25 20 34 NA 6
HGA/THGA Cooling 20 37.5 50 140 18
System
Flame Autosamplers:
S23 57 45 53 50.4 9.5
S25 79 45 53 50.4 13.6

Mercury/Hydride System:
MHS-15 17 38 16 NA 3

Flow Injection Systems:


FIAS 100/400 42 18 41 600 11
FIMS 100/400 42 26 41 450 12

FAST Flame Sample Automation Systems:


Switching Valve 20 10 9 NA 0.1
Peripumps 1: 9; 2: 13 10 NA 1: 1.6; 2:
17 2.8
Autosamplers (a) SC2 2- 54 15 without 34 144 18
tray 77 probe 34 144 22
(b) SC4 4-tray 15 without
probe
Lenovo® ThinkCentre® Computers:
CPU 18 44 40 690 11
24" LCD Monitor 55 50 20 350 25
Keyboard 46 5 18 NA 2

Accessories:
Air Compressor 65 55 31 700 36
Heated N2O Regulator 150 2
HP® LaserJet® M601 42 40 41 350 20
Brother® HL-5440 Laser 36 24 38 600 10.5
Printer
Brother® Compact all-in-
35 30 35 300 8
one Inkjet Printer MFC-
J650DW 120V
Brother® Compact all-in-
35 36 35 350 9
one Inkjet Printer MFC-
J870DW 220V

* With the PinAAcle 900Z, 900H or 900T, the furnace autosampler projects NA: Not applicable (powered by another device or included
35 cm from the right front of the instrument in its storage position. with another component)
** Height at maximum sample arm upward travel.
11 Important Accessories and Consumables

Line Conditioners
Instrument
Instrument Models Description Frequen Part
Input/
Output cy No.
Voltage
PinAAcle 500/900F 840 VA Line Conditioner 120 60 N930751
PinAAcle 500/900F 750 VA Line Conditioner 220 H 7
z N930752
50 1
H
z

UPS Systems
Instrument Models Descripti Frequen Part
on cy No.
PinAAcle 500/900F 800 VA UPS 60 N077768
PinAAcle 500/900F 800 VA UPS H 1
z N077768
50 9
H

Buck-Boost Secondary Transformers


Instrument
Instrument Models Description Frequen Part
Input/
Output cy No.
Voltage
Transformer- 2.0 KVA
PinAAcle 900H/T/Z 16/3 50/60 N930752
16/32 VAC Buck-Boost
2 Hz 5
Transformer- 2.0 KVA
PinAAcle 900H/T/Z 12/2 50/60 N930752
12/24 VAC Buck-Boost
4 Hz 6

Spares Kits Instrument Models Descripti Qty Part


on No.
PinAAcle 500/900F Spares Kit 1 (Stainless Steel Neb) N31605
20
Tubing-Tygon SE-200 per 1 FT/0.3 M 4 0250798
7
Burner Head Viton O-Ring 1 0990221
9
End Cap O-Ring - Large 1 0990214
7
End Cap O-Ring - Small 4 0920025
3
Apiezon Grease Assy 1 0303040
5
Neb Poly Tubing 10 FT AA 0.1 0990826
5
Stainless Steel Neb O-Ring (0.364 ID 0.070 WD) 2 0990201
5
Stainless Steel Neb O-Ring (0.176 ID 0.070 WD) 2 0990200
5
Stainless Steel Neb O-Ring (0.208 ID 0.070 WD) 1 0990223
9
Stainless Steel Neb O-Ring (0.070 ID 0.040 WD) 1 0990210
2
Nebulizer PTFE Washer 1 0303180
2
Burner Quick Connect O-Rings 6 0990204
7
Media-Fan Filter (5/pack) 2 N201115
6
Clamp- Hose 7/32-5/8 SST (for 02507987) 2 0990300
4
Silicone O-Ring Lube, 2 Grams 1 0920007
8
PinAAcle 500/900F Spares Kit 2 (High Sensitivity Neb) N31605
21
Tubing-Tygon SE-200 per 1 FT/0.3 M 4 0250798
7
Spares Kits, continued Instrument Models Descripti Qty Part No.
on
PinAAcle 900Z Spares Kit N3160522
THGA Contact Cylinders -1 Pair 1 B0504035
AS 900 Sampling Probe 1 B0129258
Quad-Ring 25 X 3.53 1 B0501236
PTFE Tubing (Clear) 1.75 ID X 2.45 OD 1 B0017998
Media-Fan Filter (5/pack) 2 N2011156
Ferrule Front 2.45 1 B3001333
Ferrule Back 2.45 1 B3001334
PTFE-Hose ID 1.0 X WD 0.3 1 B0029792
Back Ferrule 1/16" 1 B0080196
Front Ferrule 1/16" 1 B0080197
Cleaning Tips (10/pack) 1 B0505368
Outlet-Valve CPL 1 B3002006
Inlet-Valve CPL 1 B3002007
PinAAcle 900H Spares Kit 1 (Stainless Steel Neb) N3160525
Contact Cylinders – 1 Set 1 B0128495
AS 900 Sampling Probe 1 B0129258
Collar - Window Holder 1 B3120294
Tubing-Tygon SE-200 per 1 FT/0.3 M 4 02507987
Burner Head Viton O-Ring 1 09902219
End Cap O-Ring - Large 1 09902147
End Cap O-Ring - Small 4 09200253
Apiezon Grease Assy 1 03030405
Neb Poly Tubing 10 FT AA 0.1 09908265
Stainless Steel Neb O-Ring (0.364 ID 0.070 WD) 2 09902015
Stainless Steel Neb O-Ring (0.176 ID 0.070 WD) 2 09902005
Stainless Steel Neb O-Ring (0.208 ID 0.070 WD) 1 09902239
Stainless Steel Neb O-Ring (0.070 ID 0.040 WD) 1 09902102
Nebulizer PTFE Washer 1 03031802
Burner Quick Connect O-Rings 6 09902047
PTFE Tubing (Clear) 1.75 ID X 2.45 OD 1 B0017998
Back Ferrule 1/16" 1 B0080196
Front Ferrule 1/16" 1 B0080197
Cleaning Tips (10/pack) 1 B0505368
Ferrule Front 2.45 1 B3001333
Ferrule Back 2.45 1 B3001334
Outlet-Valve CPL 1 B3002006
Inlet-Valve CPL 1 B3002007
Media-Fan Filter (5/pack) 2 N2011156
Clamp- Hose 7/32-5/8 SST (for 02507987) 2 09903004
PinAAcle 900H Spares Kit 2 (High Sensitivity Neb) N3160526
Contact Cylinders – 1 Set 1 B0128495
AS 900 Sampling Probe 1 B0129258
Collar - Window Holder 1 B3120294
Tubing-Tygon SE-200 per 1 FT/0.3 M 4 02507987
Burner Head Viton O-Ring 1 09902219
Spares Kits, continued Instrument Models Descripti Qty Part No.
on
PinAAcle 900H Spares Kit 2 (High Sensitivity Neb) continued N3160526
End Cap O-Ring - Large 1 09902147
End Cap O-Ring - Small 4 09200253
Apiezon Grease Assy 1 03030405
High Sensitivity Neb O-Ring (0.364 ID 0.070 2 09902015
WD)
O-Ring-Metric 2 ID 1 WD 1 09926127
Burner Quick Connect O-Rings 6 09902047
Nebulizer Tubing Assembly 1 B3150458
Ceramic Ball (Neb) 1 B0505086
PTFE Tubing (Clear) 1.75 ID X 2.45 OD 1 B0017998
Back Ferrule 1/16" 1 B0080196
Front Ferrule 1/16" 1 B0080197
Cleaning Tips (10/pack) 1 B0505368
Ferrule Front 2.45 1 B3001333
Ferrule Back 2.45 1 B3001334
Outlet-Valve CPL 1 B3002006
Inlet-Valve CPL 1 B3002007
Media-Fan Filter (5/pack) 2 N2011156
Clamp- Hose 7/32-5/8 SST (for 02507987) 2 09903004
PinAAcle 900T Spares Kit 1 (Stainless Steel Neb) N3160523
THGA Contact Cylinders -1 Pair 1 B0504035
AS 900 Sampling Probe 1 B0129258
Quad-Ring 25 X 3.53 1 B0501236
Tubing-Tygon SE-200 per 1 FT/0.3 M 4 02507987
Burner Head Viton O-Ring 1 09902219
End Cap O-Ring - Large 1 09902147
End Cap O-Ring - Small 4 09200253
Apiezon Grease Assy 1 03030405
Neb Poly Tubing 10 FT AA 0.1 09908265
Stainless Steel Neb O-Ring (0.364 ID 0.070 WD) 2 09902015
Stainless Steel Neb O-Ring (0.176 ID 0.070 WD) 2 09902005
Stainless Steel Neb O-Ring (0.208 ID 0.070 WD) 1 09902239
Stainless Steel Neb O-Ring (0.070 ID 0.040 WD) 1 09902102
Nebulizer PTFE Washer 1 03031802
Burner Quick Connect O-Rings 6 09902047
PTFE Tubing (Clear) 1.75 ID X 2.45 OD 1 B0017998
Ferrule Front 2.45 1 B3001333
Ferrule Back 2.45 1 B3001334
PTFE-Hose 1.0 ID X 0.3 WD 1 B0029792
Back Ferrule 1/16" 1 B0080196
Front Ferrule 1/16" 1 B0080197
Outlet-Valve CPL 1 B3002006
Inlet-Valve CPL 1 B3002007
Cleaning Tips (10/pack) 1 B0505368
Media-Fan Filter (5/pack) 2 N2011156
Clamp- Hose 7/32-5/8 SST (for 02507987) 2 09903004
Spares Kits, continued Instrument Models Descripti Qty Part
on No.
PinAAcle 900T Spares Kit 2 (High Sensitivity Neb) N31605
24
THGA Contact Cylinders -1 Pair 1 B050403
5
AS 900 Sampling Probe 1 B012925
8
Quad-Ring 25 X 3.53 1 B050123
6
Tubing-Tygon SE-200 per 1 FT/0.3 M 4 0250798
7
Burner Head Viton O-Ring 1 0990221
9
End Cap O-Ring - Large 1 0990214
7
End Cap O-Ring - Small 4 0920025
3
Apiezon Grease Assy 1 0303040
5
Stainless Steel Neb O-Ring (0.364 ID 0.070 WD) 2 0990201
5
Nebulizer Tubing Assembly 1 B315045
8
Ceramic Ball (Neb) 1 B050508
6
O-Ring-Metric 2 ID 1 WD 1 0992612
7
Burner Quick Connect O-Rings 6 0990204
7
PTFE Tubing (Clear) 1.75 ID X 2.45 OD 1 B001799
8
PTFE-Hose 1.0 ID X 0.3 WD 1 B002979

Every day you count on PerkinElmer to provide you with solutions that deliver reliable performance, control operating costs
and maximize operational time. Our complete portfolio of consumables, parts, supplies, training and service helps you meet
both routine and demanding measurement challenges. We invest heavily in testing and validating our products to ensure
you receive guaranteed compatibility and performance – on-time, for every instrument in your laboratory.

For a complete listing of AA consumables, please visit www.perkinelmer.com/supplies

PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com

For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2020, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their

respective owners. 122446 (009362D_03) PKI

You might also like