You are on page 1of 8

DENTAL-3078; No.

of Pages 8
ARTICLE IN PRESS
d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx

Available online at www.sciencedirect.com

ScienceDirect

journal homepage: www.intl.elsevierhealth.com/journals/dema

Wear of polyetherketoneketones — Influence of


titanium dioxide content and antagonistic material

Thomas Kewekordes, Sebastian Wille ∗ , Matthias Kern


Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry,
Christian-Albrechts-University, Kiel, Germany

a r t i c l e i n f o a b s t r a c t

Article history: Objective. The aim of this laboratory study was to analyze the influence of titanium dioxide
Received 17 July 2017 (TiO2 ) content and antagonistic material on the wear of polyetherketoneketones (PEKKs).
Received in revised form Methods. Twenty-four disk-shaped specimens of two PEKK materials containing either
18 December 2017 10 wt% or 20 wt% TiO2 particles (P10 and P20) were dynamically loaded in a chewing simula-
Accepted 21 December 2017 tor with 49 N and additional thermal cycling (5–55 ◦ C). Subgroups of 8 specimens each were
Available online xxx loaded with spherical antagonists made from either steatite ceramic (St), zirconia (Zr), or
the same PEKK material (P10 or P20). After 120,000, 240,000, 480,000, 840,000, and 1,200,000
Keywords: loading cycles the vertical substance loss and the volume loss of the loaded specimens
PEKK were evaluated using a laser scanner. Data were checked considering the normal distribu-
Chewing simulator tion (Shapiro–Wilk test) and were inspected for significant differences by means of single
Wear factor variance analyses and post hoc pair comparison (Games-Howell test).
TiO2 Results. After 1,200,000 chewing cycles, statistical analyses revealed a significant influence of
Laser scanning microscope the antagonistic material. A significant difference was also found between the tested PEKKs
Thermal cycling if Zr was used as the antagonist. The volume loss ranged from between 0.073 mm3 (P20-P20)
and 0.228 mm3 (P10-St), and the vertical substance loss ranged between 73.71 ␮m (P20-P20)
and 115.268 ␮m (P10-Zr).
Significance. The inclusion of TiO2 particles influences the wear behavior of PEKK materials.
© 2018 The Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.

corrosion can result in a metallic taste and pain [4]. More-


1. Introduction over corrosion products can cause diseases such as gingivitis,
lichenoid reactions or mucosal necrosis [5].
Because of the constantly increasing demands of patients for
Ceramics in dentistry have – in addition to positive
aesthetic and high-quality dental restorations, metals have
characteristics such as good biocompatibility [6], natural
been replaced as dental restoratives by ceramics and polymers
appearance [7], low plaque accumulation [8] and high com-
[1]. The corrosion of metals can increase surface roughness
pressive strength [2] – the disadvantages of lower bending
that could serve as retention sites for plaque [2,3], and galvanic
and tensile strength than metals [9]. Their brittleness is


Corresponding author at: Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry, Christian-Albrechts
University, Arnold-Heller-Straße 16, D-24105 Kiel, Germany.
E-mail addresses: thomas.kewekordes@gmx.de (T. Kewekordes), swille@proth.uni-kiel.de (S. Wille), mkern@proth.uni-kiel.de (M.
Kern).
https://doi.org/10.1016/j.dental.2017.12.009
0109-5641/© 2018 The Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
2 d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx

Fig. 1 – Chemical structures of PEKK and PEEK.

responsible for spontaneous fractures or chipping when the


Table 1 – Group codes of the tested material
strength of the ceramic is exceeded [10]. Also, ceramics are combinations.
sensitive to processing and application errors [11]. As an alter-
Group code Tested material Antagonistic material
native to ceramics, the indications for polymers in dentistry
have expanded in recent years. The thermoplastic polymers, P10-St PEKK + 10 wt% TiO2 Steatite
P20-St PEKK + 20 wt% TiO2 Steatite
polyetheretherketones (PEEKs) have long been used in ortho-
P10-P10 PEKK + 10 wt% TiO2 PEKK + 10 wt% TiO2
pedics and traumatology as a substitute for metal components
P20-P20 PEKK + 20 wt% TiO2 PEKK + 20 wt% TiO2
because of their positive material characteristics [12]. In den- P10-Zr PEKK + 10 wt% TiO2 Zirconia
tistry, PEEKs have been used as materials for removable P20-Zr PEKK + 20 wt% TiO2 Zirconia
dentures and implant superstructures [13]. In an in vitro study,
the accumulation of biofilm was lower on implant abutments
2. Materials and methods
made of PEEKs than on those made of zirconium dioxide or
titanium [14]. Furthermore, a maxillary obturator prosthesis
2.1. Sample preparation
made of PEEK was reported to have good mucosal reactions in
areas where the mucosa was in contact with the denture [15].
Two PEKKs with a different portion of TiO2 particles were
PEKKs are also thermoplastic polymers with good biocompati-
investigated. The percentages of TiO2 were 10 wt% (Pekkton
bility [16,17] and improved chemical and mechanical qualities
ivory, Cendres + Métaux, Biel, Switzerland) and 20 wt% (Pekk-
because of the additional ketone group [18]. In addition, Moore
ton pearl, Cendres + Métaux, Biel, Switzerland). The size of the
et al. found in a rat study that PEKKs had a lower inflammatory
TiO2 particles was in the micrometer range. The specimens
response than polymethylmethacrylates [19].
were manufactured by milling technology (VhF, Ammerbuch,
Attrition and abrasion are important phenomena in the
Germany). The shape of the examined PEKK-specimens was a
wear of restorative materials [20]. The wear of natural teeth is
flat, oval plate with a width of 8 mm and a length of 11 mm.
to a certain degree a physiological process and should not be
The specimens were polymerized into metal sleeves with den-
disrupted by dental restorations. Ideally, the wear of enamel
ture acrylic resin (ProBase cold, Ivoclar Vivadent, Ellwangen,
in contact with restorative materials should not exceed the
Germany). These were clamped in special holders for the
physiological wear [21] of about 20 ␮m–40 ␮m per year [22]. If
chewing simulator. Afterwards, the specimens were ground
restorative materials have a wear behavior that differs from
under water cooling with rotating silicon carbide paper (600
natural tooth structure, the wear of the antagonistic natu-
grit) and polished (2500 grit silicon carbide paper).
ral teeth might be affected [23]. Therefore, an ideal dental
restorative material should replicate the wear behavior of nat-
2.2. Wear measurements
ural tooth structure. Reduced wear of the occlusal surface of
artificial dentures can lead to extended strain on antagonist
Twenty-four specimens of each investigated material were
teeth and their periodontium that may result in periodontal
loaded into a chewing simulator (Chewing Simulator CS-
or functional problems due to occlusal interferences [24–26].
4, SD Mechatronic, Feldkirchen-Westerham, Germany). The
Both PEKK and PEEK belong to the polyaryletherketone
specimens were randomly allocated to three subgroups with
(PAEK) family and are thermoplastic resins. The chemical
different antagonists. Six-mm diameter balls of St (Hoechst
structure of all PAEKs is based on the same aromatic rings
Ceram Tec, Wunsiedel, Germany), Zr (BCE Special Ceramics
and differs according to their ratio of keto- and ether-groups
GmbH, Mannheim, Germany) and PEKK (Cendres + Métaux,
(Fig. 1). According to the manufacturerı́s specification, the
Biel, Switzerland) (n = 8 each group) were used as the antago-
physical characteristics of PEKKs can be similar to those of
nists to simulate physiologic occlusal point contacts (Table 1).
dentin, and oxide particles can influence the mechanical char-
Table 2 shows the materials used.
acteristics and the color of PEKK materials (Cendres + Métaux,
The antagonistic balls were attached in special holders
Biel, Switzerland). PEKK materials have been successfully used
with modeling resin (Pattern Resin, GC Europe N.V., Leu-
for implanted prostheses [27–29].
ven, Belgium). Before each test series, impressions of the
The aim of this study was to evaluate the influence on wear
surfaces of the unloaded specimens were made with low-
against different antagonistic materials of two TiO2 concentra-
viscosity silicone (Express 2 Ultra-Light Body Quick, 3 M, St.
tions in PEKK.
Paul/Minnesota, USA). Afterwards, the specimens were each
The null hypothesis of this study was that no different
loaded dynamically with the parameters presented in Table 3
wear behavior would be found for P10 and P20 loaded with
in the chewing simulator with thermal cycling (5–55 ◦ C) and a
antagonists made from St, Zr, or the same PEKK material.
loading force of 49 N. After 120,000, 240,000, 480,000, 840,000

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx 3

Table 2 – Lot numbers of the used materials.


Pekkton pearl specimens Cendres + Métaux, Biel, Switzerland LOT X0223-2118m-AL-04
Pekkton ivory specimens Cendres + Métaux, Biel, Switzerland LOT 000230666
Pekkton pearl balls Cendres + Métaux, Biel, Switzerland LOT X0223-2118m-AL-04
Pekkton ivory balls Cendres + Métaux, Biel, Switzerland LOT 181265
Zirconia balls BCE Special Ceramics GmbH, Mannheim, Germany LOT 10.07.09/07
Express 2Ultra-Light Body Quick 3 M, St. Paul/Minnesota, USA LOT 523987
Steatite ceramic balls Hoechst Ceram Tec, Wunsiedel, Germany

Table 3 – Test parameters of the chewing simulation. 3. Results


Cold/hot bath temperature 5 ◦ C/55 ◦ C
Vertical movement 6 mm The mean volume and the vertical substance loss of the spec-
Rising speed 55 mm/s imens, the standard deviation after different chewing cycles
Descending speed 30 mm/s
and the results of pair-wise comparisons of the means are
Weight per specimen 49 N
summarized in Tables 4 and 5. The progress of the volume
Kinetic energy 2250 × 10 − 6 J
Dwell time 60 s and vertical substance loss of the two tested PEKKs are shown
Horizontal movement 0.3 mm in Figs. 2 and 3. After 1,200,000 cycles, the mean volume
Forward speed 30 mm/s loss ranged from 0.073 mm3 (P20-P20) to 0.228 mm3 (P10-St)
Backward speed 50 mm/s and the vertical substance loss ranged from 73.7 ␮m (P20-
Cycle frequency 1.3 Hz P20) to 115.2 ␮m (P10-Zr). The Shapiro–Wilk-Test revealed only
occasionally significant deviations of the normal distribu-
and 1,200,000 chewing cycles, the vertical substance loss and
tion so that an analysis of variance was used for statistical
the volume loss of the loaded specimens were evaluated.
analysis, which is robust against slight deviations of nor-
The specimens were cleaned with oil-free compressed air,
mal distribution [30]. The 1-way ANOVA evaluated to what
and impressions (Express 2 Ultra-Light Body Quick, 3 M, St.
extent significant effects appeared between the different com-
Paul/Minnesota, USA) were made after corresponding chewing
binations of material after 120,000, 240,000, 480,000, 840,000
cycles. For this purpose the cartridge filled with low-viscosity
and 1,200,000 chewing cycles. Significant effects among all
silicone was inserted into the dispenser. After discharging a
six groups for both loss of volume and vertical substance
low amount of silicone, it was slowly put on the edge of the
were found after all chewing cycles (ANOVA, p ≤ 0.05). Testing
specimens under permanent material contact until the entire
the homogeneity of variance of the different combinations of
specimen was covered with silicone. At this point the spec-
material showed that these were not homogeneous (Levene
imens were left behind in the chewing simulator to avoid
Test, p ≤ 0.001), so subsequently post hoc-pair comparisons
an incorrect position afterwards. A timer was set to 5 min to
via the Games-Howell-Test were made to test single material
prevent underrunning the setting time of the silicone. After
combinations for significant differences in each case. This test
5 min the silicon impressions of the specimens were carefully
revealed a significant difference between the material com-
removed and sorted into labeled boxes in the same order as
binations P10-Zr and P20-Zr (p = 0.036) in volume loss after
the measuring chambers. In the following the specimens were
1,200,000 chewing cycles. Furthermore, after 1,200,000 chew-
fixed with high viscosity silicone to be in a defined position
ing cycles, significant differences were indicated between the
for the process with the laser scanning microscope VK-x 100
material combination P10-Zr and P10-P10 (p = 0.042), between
(Keyence, Neu-Isenburg, Germany). All impressions were mea-
P20-St and P20-Zr (p = 0.025) and between P20-St and P20-P20
sured three-dimensionally with a laser scanning microscope
(p = 0.016).
at ×10 magnification using a semiconductor laser with a wave-
The vertical substance loss after 1,200,000 chewing cycles
length of 658 nm with an output power of 0.95 mW and pulse
with the material combination P10-Zr and P20-Zr (p = 0.026)
durations of 1 ns. The medium during the chewing simula-
and with the combination P20-St and P20-P20 (p = 0.032) were
tion was deionized water. The volume loss and the vertical
also significantly different. Vertical substance loss with the
substance loss were evaluated with the software VK-Analyse
material combination P10-P10 showed a large increase until
(Keyence). To investigate the influence of increasing roughness
480,000 chewing cycles when it remained almost constant.
of the specimens on the vertical substance loss, the Ra value
The roughness measurements of this group revealed an aver-
at the abraded surface and the initial Ra of the antagonists was
age roughness of Ra = 5.26 ␮m after 120,000 chewing cycles.
evaluated with the laser scanning microscope. Furthermore,
This value increased to Ra = 11.71 ␮m after 480,000 chewing
the Vickers hardness of P10 and P20 was measured with the
cycles and fell to Ra = 7.32 ␮m after 1,200,000 chewing cycles.
hardness tester 3212 Zwick with 2 kg load. For this purpose, 3
Roughness measurements of the other groups did not show
measurements were performed on 5 specimens each.
this behavior and are given in Table 7. Roughness measure-
ments of the antagonists revealed an average roughness of
2.3. Statistical analysis
Ra = 2.04 ␮m for Zr, Ra = 4.21 ␮m for St, Ra = 4.13 ␮m for P20
and Ra = 3.91 ␮m for P10. The Vickers hardness measurements
Data were evaluated for normal distribution with the
revealed an average hardness of 33.2 ± 1.5 HV 2 for P10 and
Shapiro–Wilk test followed by 1-way analysis of variance
32.1 ± 0.5 HV 2 for P20.
(ANOVA) and post hoc-pair comparison with the Games-
Howell test (SPSS 20.0).

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
4 d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx

Table 4 – Mean volume loss and standard deviations of tested materials after various chewing cycles (in mm3 ; n = 8).
Same uppercase letters within a column show no statistically significant influence of the antagonistic material on the
wear of the tested PEKKs after the respective number of chewing cycles (p ≥ 0.05). Same lowercase letters within a row
show no statistically significant influence of the tested PEKKs on its volume loss after the respective number of chewing
cycles (p ≥ 0.05).
Cycles Antagonist Means of volume loss in mm3 Means of volume loss in mm3
and (standard deviation) of P10 and (standard deviation) of P20
120,000 St 0.018 (0.006)a A, a 0.049 (0.045) AB, a
Zr 0.011 (0.005) A, a 0.012 (0.003) A, a
PEKK 0.011 (0.004) A, a 0.006 (0.003) B, a

240,000 St 0.030 (0.010)a A, a 0.069 (0.061) A, a


Zr 0.029 (0.010) A, a 0.019 (0.003) A, a
PEKK 0.024 (0.005) A, a 0.012 (0.004) A, b

480,000 St 0.047 (0.017)a A, a 0.104 (0.070) AB, a


Zr 0.059 (0.021) A, a 0.039 (0.007) A, a
PEKK 0.041 (0.011) A, a 0.024 (0.006) B, a

840,000 St 0.106 (0.069)a AB, a 0.151 (0.072) A, a


Zr 0.131 (0.040) A, a 0.061 (0.010) A, b
PEKK 0.061 (0.018) B, a 0.049 (0.009) A, a

1,200,000 St 0.228 (0.156)a AB, a 0.212 (0.077) A, a


Zr 0.181 (0.063) A, a 0.085 (0.010) B, b
PEKK 0.087 (0.030) B, a 0.073 (0.012) B, a
a
Because of incorrect measurements, one specimen was excluded (n = 7).

Table 5 – Mean vertical substance loss and standard deviations of tested materials after various chewing cycles (in ␮m;
n = 8). Same uppercase letters within a column show no statistically significant influence of the antagonistic material on
the wear of the tested PEKKs after the respective number of chewing cycles (p ≥ 0.05). Same lowercase letters within a
row show no statistically significant influence of the tested PEKKs on its vertical substance loss after the respective
number of chewing cycles (p ≥ 0.05).
Cycles Antagonist Means of vertical substance loss in Means of vertical substance loss in
␮m and (standard deviation) of P10 ␮m and (standard deviation) of P20
120,000 St 25.740 (5.2)a A, a 36.838 (26.3) A, a
Zr 20.920 (4.5) A, a 18.772 (4.0) A, a
PEKK 19.274 (12.7) A, a 15.877 (4.1) A, a

240,000 St 32.584 (6.4)a A, a 46.580 (27.8) AB, a


Zr 39.030 (9.2) A, a 30.156 (4.4) A, a
PEKK 67.731 (30.9) A, a 21.605 (4.3) B, b

480,000 St 46.169 (8.1)a A, a 63.311 (25.8) A, a


Zr 57.903 (11.5) A, a 44.981 (5.5) A, a
PEKK 88.061 (42.4) A, a 34.933 (7.1) A, a

840,000 St 74.284 (32.8)a A, a 84.637 (25.3) A, a


Zr 94.191 (18.2) A, a 57.128 (4.5) A, b
PEKK 78.434 (28.7) A, a 51.326 (10.2) A, a

1,200,000 St 113.173 (46.3)a A, a 107.164 (21.4) A, a


Zr 115.269 (21.6) A, a 82.983 (7.9) AB, b
PEKK 85.306 (22.1) A, a 73.710 (7.6) B, a
a
Because of incorrect measurements, one specimen was excluded (n = 7).

Heintze et al. noted no major difference in the quantifi-


4. Discussion cation of substance loss by chewing simulations whether
measured by 3D laser scanning, with a mechanic sensor or
In this laboratory study, the two-body wear of two PEKKs
with an optic sensor [37]. However, in this study, the assump-
was tested in a chewing simulator with different antagonistic
tion was that volume loss of the tested specimens was more
materials. Different in vitro wear tests have been reported, par-
informative than vertical substance loss, because in the mea-
ticularly two-body wear [31,32] and three-body wear [33,34].
surement of height loss, singular peaks have a stronger impact
Three-body simulators use abrasive slurry, added between the
on the results of measurement by laser. About 240,000–250,000
antagonistic material [35,36]. For this study, two-body wear, in
chewing cycles simulate approximately one year in vivo;
which only specimen and antagonist were used, was chosen
therefore, 5 years were simulated in the present study [38–40].
to allow comparisons with other studies.

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx 5

Fig. 2 – Volume loss of P10 und P20 after 120,000, 240,000, 480,000, 840,000 and 1,200,000 chewing cycles in mm3 .

120

100

80 P20-St
P10-St
60 P20-Zr
P10-Zr
40

20

0
0 240000 480000 720000 960000 1200000

Fig. 3 – Vertical substance loss of P10 und P20 after 120,000, 240,000, 480,000, 840,000 and 1,200,000 chewing cycles in ␮m.

Table 6 – Comparison of enamel wear studies after various chewing cycles in the chewing simulator.
Study Cycles Loading force Specimen/antagonist Wear
Burgess et al. [48] 100,000 10 N Enamel/enamel 0.17 mm3
200,000 10 N 0.27 mm3

Mehl et al. [43] 120,000 49 N Enamel/steatite 0.015 mm3


240,000 49 N 0.017 mm3
480,000 49 N 0.022 mm3
840,000 49 N 0.029 mm3
1,200,000 49 N 0.037 mm3

Ghazal et al. [49] 120,000 49 N Enamel/steatite 38 ␮m


240,000 49 N 41 ␮m
480,000 49 N 46 ␮m
840,000 49 N 53 ␮m
1,200,000 49 N 56 ␮m

Kern et al. [40] 120,000 49 N Enamel/steatite 0.022 mm3


240,000 49 N 0.029 mm3
480,000 49 N 0.045 mm3
840,000 49 N 0.067 mm3
1,200,000 49 N 0.083 mm3

Jung et al. [50] 240,000 49 N Feldspathic 0.119 mm3


240,000 49 N Porcelaina /enamel zirconiab /enamel 0.031 mm3

Kim et al. [51] 300,000 49 N Zirconiab /enamel feldspathic 0.04 mm3


300,000 49 N Porcelaina /enamel heat-pressed 0.11 mm3
300,000 49 N Ceramicc /enamel 0.06 mm3
a ®
Vita Omega 900 (Vita Zahnfabrik, Bad Säckingen, Germany).
b ®
Zirkonzahn prettau (Zirkonzahn GmbH, Bruneck, Italy).
c
IPS e.max Press (Ivoclar Vivadent, Liechtenstein).

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
6 d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx

Table 7 – Mean roughness measurements (Ra ) in ␮m of the material combinations.


Cycles P10-St P20-St P10-Zr P20-Zr P10-P10 P20-P20
120,000 3.398 3.721a 4.559 3.764 5.260 4.311
240,000 3.654 3.439a 4.450 3.999 10.240 4.234
480,000 4.291 3.584a 4.605 3.704 11.714 5.878
840,000 3.890 3.895a 4.552 3.907 8.535 6.287
1,200,000 3.655 4.077a 4.107 4.901 7.321 6.042
a
Because of incorrect measurements, one specimen was excluded (n = 7).

In a clinical study, Lambrechts et al. reported the wear rate ever, the roughness of Zr antagonists was lower. This might
of enamel after 12 months as 51 ␮m for molars and 32 ␮m for be why the number of particles in the micrometer range sig-
premolars. After 48 months, the molar wear rate was 153 ␮m nificantly influences the wear resistance of the PEKK. The
and the premolar wear rate was 88 ␮m [41]. The vertical sub- lowest wear occurred when the PEKKs were tested against
stance loss after 840,000 chewing cycles (corresponding to 48 themselves. In this case, the wear compared well with the
months) in the present study (51 to 94 ␮m) was similar to that wear of natural enamel and St as antagonist. This effect of a
of the clinical study. However, the comparison of in vivo and lower hardness antagonistic material is well known [44]. Ver-
laboratory data is meaningful only to a limited extent because tical substance loss with the material combination P10-P10
the wear of natural teeth depends on more factors than just showed a large increase up to 480,000 chewing cycles, after
antagonistic contacts [42]. In the interest of comparability of which it remained almost constant. This could be attributed
the obtained data, Table 6 presents various wear tests with to an initial increase in roughness and subsequent leveling
natural enamel in different chewing simulations. By compar- (Fig. 3) as confirmed by roughness measurements. It might be
ing laboratory studies of the average dental enamel wear with that due to the lower portion of TiO2 particles in P10 compared
the test results, the wear of dental enamel was significantly to P20, during the chewing simulation some polymer matrix
lower than that of the tested PEKKs (volume loss and vertical is initially retained. Therefore, in contrast to P20 the mate-
substance loss) after 1,200,000 chewing cycles. However, when rial is not homogeneously worn off, as the polymer matrix is
the wear of the tested PEKK materials was compared with not completely removed with the TiO2 particles. In addition,
that of six dental composites with steatite as the antagonis- statistical analysis revealed significant differences regarding
tic material [43], the wear seemed to be of similar dimension the volume loss between P20-St and P20-Zr and also between
(Tables 4 and 5 compared with Table 6). P20-St and P20-P20 but no significant difference between P20-
It was assumed that the different amounts of TiO2 par- P20 and P20-Zr. The significant variations between P20-St and
ticles of the tested PEKKs resulted in a different hardness, P20-Zr were due to the varying roughness of the antagonists St
and therefore the wear resistance increased. However, these and Zr. In the case of P20-St and P20-P20, the significant con-
characteristics had no significant effect on wear except when trast was caused by the different hardness of the antagonists.
Zr was the antagonist. The increased wt% of TiO2 particles The tests showed no significant differences between P20-P20
increased wear resistance only for antagonist materials with and P20-Zr. The higher roughness of the antagonist P20 may
low roughness. The total loss of substance depends on the have been compensated for by the higher hardness of Zr. How-
hardness of the materials involved, the geometry of the par- ever, regarding P10, significant differences were only found
ticles involved, and the load and length of the contact area between the antagonists Zr and P10 visible. In this case, the
[44]. The load and the contact area were constant parame- higher hardness of the antagonist Zr does not seem to corre-
ters in the chewing simulation. As reported, the roughness of late with the higher roughness of the tested PEKK. Wear is not
the Zr antagonists was significantly less than the roughness a material property but a system property, so material strength
of all other antagonistic materials. The Vickers hardness of may play a role. However, relation of the values for the mate-
the antagonists used was about 680 MPa for St [45], 1300 MPa rial strength of the tested materials is similar to the relation of
for Zr (BCE Special Ceramics GmbH, Mannheim, Germany), the values for the hardness of these materials. This may lead
and 252 MPa for P10 (Cendres + Métaux, Biel, Switzerland). to an increase of the influence of the material. The use of St
According to the corresponding manufactures the compres- as antagonist caused nonsignificant differences between P10
sive strength is 830 MPa for St, 2200 MPa for Zr and 246 MPa and both the other antagonists because of its high standard
for P10 and the elastic modulus is 80 GPa for St, 200 GPa for Zr deviation.
and 5.1 GPa for P10. According to the manufacturer the values A limitation of this study is that it is a laboratory study.
for the compressive strength and elastic modulus for P20 are Its conditions only simulate the clinical situation to a lim-
similar to those of P10. ited extend, so the clinical behavior might differ from the test
The wear of the two tested PEKK materials with differ- results. Another limitation is that the only tested parame-
ent amounts of TiO2 particles against the St and the PEKK ter was wear. Furthermore, it has to be considered that flat
antagonists was not significantly different (p > 0.05). It appears specimens, probably due to the higher stress distribution in
that the TiO2 particles were removed from the bulk mate- the specimens [46], show higher material loss than anatomi-
rial with the polymer matrix material by the rougher St and cally shaped crowns [47]. In summary, the wear of the tested
PEKK antagonists. Therefore, the hard ceramic pigments did PEKKs was higher than that of comparable studies with natu-
not affect wear resistance against these antagonists. How- ral enamel but comparable with current restorative materials.

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx 7

An appropriate comparison of the results of this study with [7] Gallucci GO, Guex P, Vinci D, Belser UC. Achieving
the values determined in the vivo study by Lambrechts et al. natural-looking morphology and surface textures in anterior
is not possible. Using the approach of several test parame- ceramic fixed rehabilitations. Int J Periodontics Restorative
Dent 2007;27:117–25.
ters, for example, different chewing forces and a three-body
[8] Chan C, Weber H. Plaque retention on teeth restored with
wear simulation with various abrasive slurries, could improve full-ceramic crowns: a comparative study. J Prosthet Dent
comparability with the natural masticatory apparatus. 1986;56:666–71.
The null hypothesis was accepted after 1,200,000 chewing [9] Gehre G, Keramische Werkstoffe. In: Kappert H, Eichner K.
cycles using St or the same PEKK as the antagonistic mate- Zahnärztliche Werkstoffe und ihre Verarbeitung. 1996;6:348.
rial. Neither volume loss nor vertical substance loss showed [10] Rudolph H, Johannes M, Luthardt R.
significantly different values. However, a statistically signifi- Niederdruckspritzgießen von Hochleistungskeramik. Dtsch
Zahnärztl Z 2005;60:172–5.
cant difference was found between PEKK and Zr, so the null
[11] Kern M, Keramik A. Was hat sich langfristig bewährt, was
hypothesis was rejected for Zr as the antagonistic material. noch nicht? DDZ 2015;124:292–8.
[12] Kurtz SM, Devine JN. PEEK biomaterials in trauma,
orthopedic, and spinal implants. Biomaterials
5. Conclusion 2007;28:4845–69.
[13] Najeeb S, Zafar MS, Khurshid Z, Siddiqui F. Applications of
Within the limitations of this in vitro study, the following con- polyetheretherketone (PEEK) in oral implantology and
clusions were drawn: prosthodontics. J Prosthodont Res 2016;60:12–9.
[14] Hahnel S, Wieser A, Lang R. Biofilm formation on the surface
1. Depending on the antagonist roughness. the amount of of modern implant abutment materials. Clin Oral Implants
Res 2014;26:1297–301.
TiO2 particles had a significant influence on the wear
[15] Costa-Palau S, Torrents-Nicolas J, Brufau-de Barbara M. Use
behavior of PEKKs.
of polyetheretherketone in fabrication of a maxillary
2. Wear of the tested PEKKs with St as the antagonist was obturator prosthesis: a clinical report. J Prosthet Dent
comparable with wear of dental composites but not with 2014;112:680–2.
wear of dental enamel. [16] Converse GL, Conrad TL, Merrill CH, Roeder RK.
3. Against Zr or the same PEKK as the antagonist, the amount Hydroxyapatite whisker-reinforced polyetherketoneketone
of wear was comparable to the amount of wear of dental bone ingrowth scaffolds. Acta Biomater 2010;6:856–63.
[17] Converse GL, Conrad TL, Roeder RK. Mechanical properties
enamel against St.
of hydroxyapatite whisker reinforced
4. Wear of the tested PEKKs against themselves showed the polyetherketoneketone composite scaffolds. J Mech Behav
lowest rates in this study. Biomed Mater 2009;2:627–35.
[18] Copponnex T, DeCarmine A. Reevaluating thermoplastics.
How good is PEEK, really, for aesthetic and long-term
Acknowledgements structural applications? Med Device Technol 2009:26–7.
Available from:
The authors wish to thank Frank Lehmann and Rüdiger Möller http://www.cmdental.fr/medias/files/publications-pekkton
(Department of Prosthodontics, Propaedeutics and Dental -2014.pdf.
Materials, School of Dentistry, Christian-Albrechts-University [19] Moore R, Beredjiklian P, Rhoad R, Theiss S, Cuckler J,
Ducheyne P, et al. A comparison of the inflammatory
at Kiel) for technical support and Ulrike von Hehn and Jür-
® potential of particulates derived from two composite
gen Hedderich (Medistat , Kiel) for statistical advice. This
materials. J Biomed Mater Res 1997;34:137–47.
study was supported financially and with materials by Cen- [20] Mair LH. Wear in dentistry—current terminology. J Dent
dres + Métaux SA (Biel, Switzerland). 1992;20:140–4.
[21] Lambrechts P, Braem M, Vanherle G. Buonocore memorial
references lecture. Evaluation of clinical performance for posterior
composite resins and dentin adhesives. Oper Dent
1987;12:53–78.
[22] Seghi RR, Rosenstiel SF, Bauer P. Abrasion of human enamel
[1] Wall JG, Cipra DL. Alternative crown systems. Is the by different dental ceramics in vitro. J Dent Res
metal-ceramic crown always the restoration of choice? Dent 1991;70:221–5.
Clin North Am 1992;36:765–82. [23] Sulong MZ, Aziz RA. Wear of materials used in dentistry: a
[2] Kappert H, Eichner K. Zahnärztliche Werkstoffe und ihre review of the literature. J Prosthet Dent 1990;63:342–9.
Verarbeitung. Volume 2 Werkstoffe unter klinischen [24] DeLong R, Sasik C, Pintado MR, Douglas WH. The wear of
Aspekten. 6th ed. Stuttgart: Georg Thieme Verlag; 2008. p. 37. enamel when opposed by ceramic systems. Dent Mater
[3] Geis-Gerstofer J, Rupp F, Scheideler L. Korrosionsprüfung - 1989;5:266–71.
eine wichtige Basis für die Materialauswahl. Quintessenz [25] Mahalick JA, Knap FJ, Weiter EJ. Occusal wear in
Zahntech 2005;31:1343. prosthodontics. J Am Dent Assoc 1971;82:154–9.
[4] Opydo W, Opydo-Szymaczek J. Metallic dental materials in [26] Gallegos LI, Nicholls JI. In vitro two-body wear of three
patient’s oral cavity acting as electrodes of electrochemical veneering resins. J Prosthet Dent 1988;60:172–8.
cells. Mater Corros 2004;55:520–3. [27] Koehler S, Raslan F, Stetter C, Rueckriegel SM, Ernestus RI,
[5] Manaranche C, Hornberger H. A proposal for the Westermaier T. Autologous bone graft versus PEKK cage for
classification of dental alloys according to their resistance to vertebral replacement after 1- or 2-level anterior median
corrosion. Dent Mater 2007;23:1428–37. corpectomy. J Neurosurg Spine 2015:1–6.
[6] Pae A, Lee H, Kim HS, Kwon YD, Woo YH. Attachment and [28] Morrison RJ, Sengupta S, Flanangan CL, Ohye RG, Hollister SJ,
growth behaviour of human gingival fibroblasts on titanium Green GE. Treatment of severe acquired tracheomalacia with
and zirconia ceramic surfaces. Biomed Mater 2009;4:025005.

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009
DENTAL-3078; No. of Pages 8
ARTICLE IN PRESS
8 d e n t a l m a t e r i a l s x x x ( 2 0 1 8 ) xxx–xxx

a patient-specific, 3D-printed, permanent tracheal splint. [40] Kern M, Strub JR, Lu XY. Wear of composite resin veneering
JAMA Otolaryngol Head Neck Surg 2017;143:523–5. materials in a dual-axis chewing simulator. J Oral Rehabil
[29] Adamzyk C, Kachel P, Hoss M, Gremse F, Modabber A, Holzle 1999;26:372–8.
F, et al. Bone tissue engineering using [41] Lambrechts P, Braem M, Vuylsteke-Wauters M, Vanherle G.
polyetherketoneketone scaffolds combined with autologous Quantitative in vivo wear of human enamel. J Dent Res
mesenchymal stem cells in a sheep calvarial defect model. J 1989;68:1752–4.
Craniomaxillofac Surg 2016;44:985–94. [42] Smith BG, Bartlett DW, Robb ND. The prevalence, etiology
[30] Toutenburg H, Knöfel P. Six Sigma Methoden und Statistik and management of tooth wear in the United Kingdom. J
für die Praxis. 2nd ed. Berlin: Springer-Verlag; 2008. p. 239. Prosthet Dent 1997;78:367–72.
[31] Hu X, Marquis PM, Shortall AC. Influence of filler loading on [43] Mehl C, Scheibner S, Ludwig K, Kern M. Wear of composite
the two-body wear of a dental composite. J Oral Rehabil resin veneering materials and enamel in a chewing
2003;30:729–37. simulator. Dent Mater 2007;23:1382–9.
[32] Knobloch LA, Kerby RE, Seghi R, van Putten M. Two-body [44] Abebe M, Appl F. Theoretical analysis of the basic mechanics
wear resistance and degree of conversion of of abrasive processes: Part II: Model of the ploughing
laboratory-processed composite materials. Int J Prosthodont process. Wear 1988;126:267–83.
1999;12:432–8. [45] Shortall AC, Hu XQ, Marquis PM. Potential countersample
[33] Frankenberger R, Garcia-Godoy F, Lohbauer U, Petschelt A, materials for in vitro simulation wear testing. Dent Mater
Kramer N. Evaluation of resin composite materials. Part I: 2002;18:246–54.
in vitro investigations. Am J Dent 2005;18:23–7. [46] Heintze SD, Cavalleri A, Forjanic M, Zellweger G, Rousson V.
[34] Reich SM, Petschelt A, Wichmann M, Frankenberger R. Wear of ceramic and antagonist — a systematic evaluation
Mechanical properties and three-body wear of veneering of influencing factors in vitro. Dent Mater 2008;24:433–49.
composites and their matrices. J Biomed Mater Res A [47] Wimmer T, Huffmann AM, Eichberger M, Schmidlin PR,
2004;69:65–9. Stawarczyk B. Two-body wear rate of PEEK, CAD/CAM resin
[35] Mair LH, Stolarski TA, Vowles RW, Lloyd CH. Wear: composite and PMMA: effect of specimen geometries,
mechanisms, manifestations and measurement. Report of a antagonist materials and test set-up configuration. Dent
workshop. J Dent 1996;24:141–8. Mater 2016;32, e127–36.
[36] Xu HH, Quinn JB, Giuseppetti AA, Eichmiller FC, Parry EE, [48] Burgess JO, Janyavula S, Lawson NC, Lucas TJ, Cakir D.
Schumacher GE. Three-body wear of dental resin Enamel wear opposing polished and aged zirconia. Oper
composites reinforced with silica-fused whiskers. Dent Dent 2014;39:189–94.
Mater 2004;20:220–7. [49] Ghazal M, Yang B, Ludwig K, Kern M. Two-body wear of resin
[37] Heintze SD, Cavalleri A, Forjanic M, Zellweger G, Rousson V. and ceramic denture teeth in comparison to human enamel.
A comparison of three different methods for the Dent Mater 2008;24:502–7.
quantification of the in vitro wear of dental materials. Dent [50] Jung YS, Lee JW, Choi YJ, Ahn JS, Shin SW, Huh JB. A study on
Mater 2006;22:1051–62. the in-vitro wear of the natural tooth structure by opposing
[38] DeLong R, Sakaguchi RL, Douglas WH, Pintado MR. The wear zirconia or dental porcelain. J Adv Prosthodont 2010;2:111–5.
of dental amalgam in an artificial mouth: a clinical [51] Kim MJ, Oh SH, Kim JH, Ju SW, Seo DG, Jun SH, et al. Wear
correlation. Dent Mater 1985;1:238–42. evaluation of the human enamel opposing different Y-TZP
[39] Sakaguchi RL, Douglas WH, DeLong R, Pintado MR. The wear dental ceramics and other porcelains. J Dent 2012;40:979–88.
of a posterior composite in an artificial mouth: a clinical
correlation. Dent Mater 1986;2:235–40.

Please cite this article in press as: Kewekordes T, et al. Wear of polyetherketoneketones — Influence of titanium dioxide content and antagonistic
material. Dent Mater (2018), https://doi.org/10.1016/j.dental.2017.12.009

You might also like