Professional Documents
Culture Documents
Electrolyte Analyzer
About this Manual
Statement
This manual will help you understand the operation and maintenance
of the product better. It is reminded that the product shall be used
strictly complying with this manual. User’s operation failing to comply
with this manual may result in malfunction or accident for which the
manufacturer cannot be held liable.
The user shall understand that nothing in this manual grants him,
expressly or implicitly, any right or license to use any of the intellectual
properties of the manufacturer.
The manufacturer reserves the right to modify and update this manual
without prior notice.
I
The manufacturer reserves the final explanation right to this manual.
Warning
Make sure to use the instrument under the provisions of this Operation
Manual. Otherwise, the instrument may not work normally, the test
result is unreliable and it may damage the instrument parts and
endanger people’s safety.
Preface
The LCD touch screen and thermal printer are convenient for users to
operate. The advanced ion selectivity electrode (ISE) test method is
used to ensure the accuracy and reliability of the results. Storage of
more than 2000 results can meet the general need.
II
This operation manual includes instructions and precautions of
installation, operation, maintenance and service for the electrolyte
analyzer. Please read this manual carefully before operating the
instrument.
This operation manual mainly intends to help users learn about the
safety, installation, structure, analysis principle, operating procedures,
maintenance and service, problems and troubleshooting, etc. of the
Electrolyte Analyzer. Please follow this manual to operate the
instrument.
III
Table of Contents
2.1 Safety................................................................................... 11
IV
3.1 Working conditions .............................................................. 15
3.2 Installation............................................................................ 16
3.2.1 Installer......................................................................... 16
3.2.2 Damage check ............................................................. 16
3.2.3 Packing list ................................................................... 17
3.2.4 Installation .................................................................... 17
4.8 Setting.................................................................................. 45
4.8.1 Contrast ....................................................................... 46
4.8.2 Unit selection ............................................................... 47
4.8.3 Reference range .......................................................... 47
4.8.4 Date and time............................................................... 49
4.8.5 COM setting ................................................................. 49
V
4.8.6 Printer setting ............................................................... 51
4.8.7 Test item setting ........................................................... 51
4.8.8 Exit ............................................................................... 52
VI
Chapter 8 Transportation, Storage and the Outer Packing
Instructions .................................................................................... 71
VII
Chapter 1 Instrument introduction
ln(ai × fi)
2.303RT
E = E0 ±
nF
Among them:
It can be seen from the above formula that under certain condition,
there is a linear relationship between E and ln(ai×fi), so we can get “ai”
by testing E.
1
Chapter 1 Instrument introduction
2
Chapter 1 Instrument introduction
Note: The object may not be exactly the same with the picture, please
refer to the real object provided.
3
Chapter 1 Instrument introduction
Note: The object may not be exactly the same with the picture, please
refer to the real object provided.
Fig. 1.1
4
Chapter 1 Instrument introduction
After doing the Power On Self Test, the instrument enters into the main
interface, as shown in Fig. 1.1. There are six sub-menus, simply click
on the corresponding menu to make a selection.
All reagents are only used for in vitro diagnostic, usually stored at room
temperature (18~30℃), and shall be used within the period of validity.
Please use the reagent as soon as possible, and store the remaining
reagent in the refrigerator to maintain its effect, otherwise some
parameters will get changed after a long time placement. In order to
ensure the accuracy and reliability of the test results and trouble-free
operation of the instrument, please use the original reagents.
The symbols listed in the table below are applicable to this operation
manual.
5
Chapter 1 Instrument introduction
Symbol Meaning
6
Chapter 1 Instrument introduction
Symbol Meaning
the real operation interface.
Note: Corrosion
The screen printings and labels listed in the table below are applicable
to this operation manual.
Label Meaning
Production date
~ Alternating current
7
Chapter 1 Instrument introduction
Power on
Power off
A, B Standard A, Standard B
1.4.3 Pictures
Pictures in this manual are only for illustration or examples, not for
other purposes. Please refer to the real object while in operation,
modification may be made without prior notice.
8
Chapter 1 Instrument introduction
The manufacturer warrants that this product will be free from defects in
materials and workmanship that occur within the warranty period. This
warranty shall not apply to damages caused by the following situations:
3) Man-made damage;
9
Chapter 1 Instrument introduction
1.6 Training
In order to help users properly use the Electrolyte Analyzer and give
full play to its performance, the manufacturer holds periodical free
training for users throughout the world, please visit the manufacturer’s
website or contact the local offices to get the training schedule.
Please carefully read this manual and the related reagents instructions,
if not operating in accordance with the provisions of this instrument
manual, it may damage the provided protection or greatly affect the
test results.
10
Chapter 2 Safety and precautions
The following are the warning symbols related to this analyzer, if ignore
these symbols, the users may run the risk of death or serious injury.
The following warning symbols are not sorted by their importance, all
the symbols have the equivalent importance.
2.1 Safety
Electric shock
1) When the power supply is turned on, unauthorized
service technicians are prohibited to open the
analyzer’s panel.
2) Avoid spilling any liquid onto the workbench, if liquids
penetrate into the analyzer, please turn off the power
immediately and contact the manufacturer promptly.
Danger
1) Remove the sample when hearing a "di" alarm
sound, and pay attention to the sharp parts of the
analyzer, which may cause body injury.
2) When the analyzer is working, do not touch the
moving parts, such as the sample probe, peristaltic
pump, fan, etc.
3) While making the “exit” operation, the sample probe
automatically descends to the suction tank, pay
attention to the sharp parts of the analyzer.
11
Chapter 2 Safety and precautions
Biohazard
1) All the test samples, calibrators, quality controls, etc.
should be regarded as infectious, gloves should be
worn while in touch;
2) All the waste liquids should be regarded as
infectious, gloves should be worn while in touch. All the
items contacted with the test samples, such as pipette
tips, measuring cups should be regarded as infectious,
gloves should be worn while in touch;
3) All the wastes are potentially infectious and should
be regarded as medical wastes and disposed according
to the current regulatory requirements;
4) When the instrument reaches its lifetime, it should be
disposed according to the requirements of the local
environmental protection department, cannot be
disposed and discarded as common wastes.
2.2 Precautions
Application Range
+
1) Applicable to the determination of potassium (K ),
+ - 2+
sodium (Na ), chloride (Cl ), calcium (iCa ), total
+
calcium (TCa), lithium (Li ) concentration and pH value
(pH) in human serum, urine, and body fluids.
2) When making a clinical judgment based on the test
results, please consider the clinical examination results,
other test results or clinical symptoms as well.
Operator
This analyzer can only be used by personnel who have
been trained and authorized by the manufacturer or its
12
Chapter 2 Safety and precautions
Application Environment
Please follow the installation environment and use
conditions specified in this manual to install the
instrument correctly. Otherwise, the test results may be
unreliable and the analyzer may be damaged.
Electromagnetic Interference
1) The analyzer is susceptible to electromagnetic
interference during running, which may affect the test
results and lead to incorrect operation. During its
running, please do not use devices that generate
electromagnetic waves, such as electric drill, mobile
phone, intercoms, etc.
2) The analyzer can emit electromagnetic waves during
running. Please do not install and use
electromagnetic-sensitive equipment near the analyzer.
Imperfect Earth
1) The power supply must be correctly grounded;
otherwise, there is a risk of electric shock.
2) The grounding impedance must be lower than 0.1Ω.
Imperfect earth may cause unstable test results and
electricity leakage of the shell, and thus lead to the risk
of electric shock.
Device Connection
1) For a non-permanent connection, please do not place
the device at a location where it is hard to disconnect.
2) For all the external switches or breakers and external
13
Chapter 2 Safety and precautions
Combustion
Do not use inflammable hazardous substances near the
analyzer, such as alcohol or ether, etc.
14
Chapter 3 Installation
Chapter 3 Installation
Content Requirements
Flat workbench with tilt < 1/200 and sufficient
intensity, which can bear the weight of 50kg;
Dust-free, without corrosive and inflammable gases,
Place
without heat source and wind source, without
mechanical vibration;
Avoid direct sunlight and with good ventilation.
Enough space to place the analyzer, minimum area
Space
of 0.48m (length ≥80cm, width ≥60cm).
2
Environment
15℃~30℃
temperature
Relative
20%~80%
humidity
Atmospheric
86kPa~106kPa
pressure
Able to exchange air with outside, smooth air flow,
Ventilation
no direct wind blow to the analyzer.
~100-240V, 50/60Hz, power supply correctly
Power supply grounded and the grounding impedance is less than
0.1Ω.
15
Chapter 3 Installation
Once you find any of the above issues, please do not open the
package and do not power on the instrument. Otherwise, it may lead to
electric shock. Please notify the local distributor immediately.
16
Chapter 3 Installation
If the outer package is intact, please open the packaging box and
check the items in the box at the presence of the distributor or the
manufacturer:
3.2.4 Installation
Open the dedicated packaging box of this analyzer and carefully verify
if the instrument and accessories in the box are consistent with the
packing list. If there is any loss or damage, please contact the supplier
at once;
If all the items are there, carefully take out the instrument and put it on
a stable workbench with good ventilation, do not place it near the wall
or other places where are difficult for disconnection;
Take out the power cable, plug one end into the supply hub at the rear
panel and insert the other end into the grounded three-wire socket. For
a non permanent connection, please do not place the device at a
location where is hard to disconnect;
17
Chapter 3 Installation
with the steel pipe which is buried under the ground more than one
meter.
Take out the waste bottle from the packing box and put it at the right
side of the instrument, then put the waste tube into the bottle.
Take out the reagent bottle from the packing box, open the reagent
room, then put the reagent tube into the corresponding reagent bottle
and make sure the tubes reach the bottom of the bottle; please refer to
App. 2.
Switch on the power button at the rear panel, after the system self-test
and auto calibration, the instrument is ready to be used.
18
Chapter 4 Interface Structure and Operation
Interface menu structure and function are shown in the following table.
Sampli Aspirate
ng sample
Num Input the
Input
eric 13-digit ID
ID Click the
key NO.
Samp correspo
Directly
le nding
Input input the
type button on
NO. sample ID
(Seru the
Meas NO.
m, screen to
ure + Query the
CSF, realize
previous or
Urine, - the
Result next NO.
Other correspo
Print
s) Print nding
result
results function
s
Return to
Exit the
measurem
19
Chapter 4 Interface Structure and Operation
ent
interface
External
Bar
bar code
code
reader to
input
input
ID
13-digit ID
Return to
Exit the main
menu
Calibr Auto
ate calibration
Set the
Num Click the
correspond
Samp eric correspo
ing target
le Select key nding
value
type items button on
Save the
(Seru (K, the
Corre changes
m, Na…) screen to
ct and return
CSF, to be Enter realize
to the
Urine, correct the
previous
Other ed correspo
interface
s) nding
Return to
Exit function
the
20
Chapter 4 Interface Structure and Operation
previous
interface
without
saving the
changes
Delete the
Del input
number
Aspirate
sample/sta
Sam
ndard
pling
solution/Q
Autom
C solution
atic
Return to
the
Exit
correction
interface
Modify
Num factors of
eric the
Manual key correspond
ing ions
Rese Restore
t the
21
Chapter 4 Interface Structure and Operation
modified
factor to
the initial
value
Return to
the
Exit
correction
interface
Return to
Exit the main
interface
Aspirate
Sampl
QC Click the
ing
Samp solution correspo
le + Query the nding
type Quality previous or button on
-
(Seru control next NO. the
Meas
QC m, (QC1, Delete the screen to
ure
CSF, QC2, result with realize
Result Del
Urine, QC3) the shown the
Other No. correspo
s) Print nding
Print
results function
Exit Return to
22
Chapter 4 Interface Structure and Operation
the
previous
menu
Return to
the
Exit
previous
menu
Set the
Nume correspond
ric ing lot
key number
code
Save the
changes
and return
Lot Enter
to the
No.
previous
interface
Return to
the
previous
Exit
interface
without
saving the
23
Chapter 4 Interface Structure and Operation
changes
Delete the
Del input
number
Calculate
SD value
Stat
and CV
value
Return to
the
Exit measurem
ent
interface
Increase or Click the
+
decrease correspo
the screen nding
Contr -
contrast button on
ast
Settin Return to the
gs Exit the setting screen to
interface realize
mmol/L After the
Units selected, correspo
mg/dL
the nding
24
Chapter 4 Interface Structure and Operation
instrument function
automatica
lly
performs
the unit
conversion
Return to
Exit the setting
interface
Nume Set the
ric correspond
key ing value
Save the
Sampl changes
Item
e type and return
(K, Enter
(Serum to the
REF Na…
, CSF, previous
STD )
Urine, interface
rang
Others) Return to
e key
the
previous
Exit
interface
without
saving the
25
Chapter 4 Interface Structure and Operation
changes
Delete the
Del input
number
Set the
Numeri
correspond
c key
ing value
Save the
changes
and return
Enter
to the
previous
interface
Set
Return to
Time
the
previous
Exit interface
without
saving the
changes
The cursor
Del moves
backward
26
Chapter 4 Interface Structure and Operation
27
Chapter 4 Interface Structure and Operation
print after
calibration
or test
Return to
Exit the setting
interface
ASP+ Increase or
decrease
the
ASP-
aspiration
Set Click the
volume
Vol correspo
Return to
nding
the
Exit button on
maintain
the
Maint interface
screen to
ain Deprot De-protein
realize
ein cleaning
the
Auto clean
correspo
the tubes
Clean Tube nding
with
ing function
reagent A
Return to
Exit the
maintenan
28
Chapter 4 Interface Structure and Operation
ce
interface
Sampli Aspirate
ng sample
Auto
STD A aspirate
STD A
Auto
ADC STD B aspirate
STD B
Return to
the
Exit maintenan
ce
interface
Feed
Feed paper to
the printer
Return to
Exit the main
interface
29
Chapter 4 Interface Structure and Operation
Note:
All the “Exit” operation or hearing the alarm sound, the
sample tubes and human body should stay away from
the probe and other moving parts, otherwise, it may
cause body injury or instrument damage.
Fig. 4.1
30
Chapter 4 Interface Structure and Operation
Fig. 4.2
31
Chapter 4 Interface Structure and Operation
Fig. 4.3
The interface of Fig.4.3 will stay for 5s on the screen and automatically
print out, then the system enters into the main interface. If there is any
abnormal, recalibrate or take other measures according to the actual
situation.
Note:
32
Chapter 4 Interface Structure and Operation
4.4 Measure
4.4.1 Measuring flow chart
Turn on
Selt-test
Auto-
Calibrate
Main Menu
Measure
Prepare
sample
ASP
Take away
sample
Result and
print
Cancel Next
Main Menu
Other
operations
In the main interface (see Fig. 4.4), click "Measure" to enter into the
measurement interface.
33
Chapter 4 Interface Structure and Operation
Fig. 4.4
Select sample type as shown in Fig. 4.5. After selecting the sample,
the sample probe automatically lifts from the suction tank and the liquid
system drains. Then the system enters into the measurement interface,
as shown in Fig. 4.6, in this interface, the barcode reader can scan to
input the 13-digit ID number (Refer to App. 7 for the detailed setting
methods).
Fig. 4.5
34
Chapter 4 Interface Structure and Operation
Click “Sampling” as shown in Fig. 4.6, when hearing a “di” alarm sound,
remove the sample immediately.
Fig. 4.6
The instrument will auto-print the results. If “Cancel” is set for print
setting or need re-print, it needs to be printed manually, enter into the
“Result” interface, search the results to be printed and click “Print” to
print again. Click “Exit” to return to the measurement interface, see Fig.
4.6.
The probe lifts from the suction tank, click “Sampling” to enter step 3)
to complete all the sample tests of the same type.
35
Chapter 4 Interface Structure and Operation
The test value is high or slope is low: use the cleaning solution to clean
the electrodes, then the electrode activating solution to activate the
electrodes, please refer to 4.9.3.
Long time not use the electrodes, use the distilled water to wash away
the residual liquid of the electrode membrane tube, then seal and store
it wetly.
36
Chapter 4 Interface Structure and Operation
Long time not use or move, the reagents should be taken out and the
tubes drained.
When test urine or other non-serum body fluids, if need dilution, please
follow the related standards.
4.5 Calibration
The analyzer will do the calibration before the test to get the slope and
intercept.
The analyzer will do the auto-calibration, as shown in Fig 4.2 and 4.3,
and will auto print the calibration result.
37
Chapter 4 Interface Structure and Operation
4.6 Correction
After the auto calibration, the instrument will generate a factor. While
doing the clinical tests, users are suggested to buy the human
substrate calibrators with traceability for results correction.
Enter into the sample type selection interface, directly click to select
the sample type.
Fig. 4.7
Carefully check whether the ions’ target value is consistent with the
calibrators’ nominal value, if not, revise the target value to be the same
with the nominal value before re-correcting.
Change the target value: if the target value is not the same as the
38
Chapter 4 Interface Structure and Operation
calibrator’s nominal value, directly click the item, for example, if the
displayed target value for “K” is different from the nominal value,
directly click “K” in the interface of Fig. 4.7 to enter into the modification
interface, as shown in Fig. 4.8, directly click the corresponding
numbers to edit the result target value, then click “Enter” to save and
return to the interface of Fig. 4.7.
Fig. 4.8
Note:
The displayed target value on the interface must be
exactly the same with the calibrator’s nominal value.
Otherwise it will not be able to obtain reliable test
results after correction.
Auto correction
39
Chapter 4 Interface Structure and Operation
Fig. 4.9
Prepare the calibrator and put it under the probe, click “Sampling”, the
instrument automatically aspirates sample and displays the test value
and correction ratio, as shown in Fig. 4.10.
Fig. 4.10
40
Chapter 4 Interface Structure and Operation
Manual correction
If there is big discrepancy between the test results and the calibrator
result, click “Manual” in the interface of Fig. 4.7 to modify the
corresponding factor, the manual correction interface is as shown in
Fig. 4.11.
Fig. 4.11
If need to modify the slope and intercept of item “K”, click the
correspondingly a and b box to make the modification, as shown in Fig.
4.12, directly input the corresponding number, then click “Enter” to
save and return to the manual correction interface.
41
Chapter 4 Interface Structure and Operation
Fig. 4.12
After completing the factor modification for all ions, click “Exit” to return
to the correction interface of Fig. 4.7.
Click “Reset”, the modified factors will restore to the original value:
a=1.000, b=0.
In the main menu, click “QC” to select the sample type, as shown in
Fig.4.5.
After selecting the “Type”; it pops out a new interface, as shown in Fig.
4.13.
42
Chapter 4 Interface Structure and Operation
Fig. 4.13
Select one QC mode in Fig. 4.13 to enter the new interface, as shown
in Fig. 4.14.
Fig. 4.14
Measure
Click “Measure”, the probe will lift from the suction tank, and the
system enters into the interface of Fig.4.15.
43
Chapter 4 Interface Structure and Operation
Fig. 4.15
Put the QC solution under the probe, click “Sampling” to aspirate the
QC sample. After hearing a “di” alarm sound, remove the QC sample.
The instrument will auto-print the results. If “Cancel” is set for print
setting or need re-print, it needs to be printed manually, enter into the
“Result” interface, search the results to be printed, and click “Print” to
print again. Click “Exit” to return to the interface of Fig. 4.15.
Click “Exit”, the probe descends to the suction tank, and the system
returns to the interface of Fig. 4.14.
Lot No.
44
Chapter 4 Interface Structure and Operation
Fig.4.16.
Fig. 4.16
Directly click the corresponding numeric key to input the lot No..
4.8 Setting
In the main menu, click “Settings” to enter into the setting interface of
Fig. 4.17, where you can set the Contrast, Units, REF STD, Time,
COM and Printer.
45
Chapter 4 Interface Structure and Operation
Fig. 4.17
4.8.1 Contrast
Fig. 4.18
46
Chapter 4 Interface Structure and Operation
Click “Units” to enter the unit selection interface, as shown in Fig. 4.19.
Fig. 4.19
Directly select one unit, the results will do the auto-conversion. After
switching the units, the reference range needs to be re-input, set the
range according to the setting method in 4.8.3.
Click “REF STD” to enter into the interface of Fig. 4.5, and select the
sample type.
After selecting one of the sample types, enter into the interface of Fig.
4.20.
47
Chapter 4 Interface Structure and Operation
Fig. 4.20
Select the item to be modified, the left side value is the lower limit and
the right side value is the upper limit. Directly click the upper or lower
limits of the corresponding item to enter into the interface of Fig. 4.16.
Click the numeric keys to input the upper or lower limit value, and click
“Exit” to return to the interface of Fig. 4.20.
Click “Exit” to return to the interface of Fig. 4.17, and click “REF STD”
to select different sample type, then follow step 4.8.3.2 to 4.8.3.4 to
input the “REF STD” for testing items of all the sample types.
K: 1mmol/L=3.90983mg/dL
Na: 1mmol/L=2.29897mg/dL
Cl: 1mmol/L=3.54532mg/dL
48
Chapter 4 Interface Structure and Operation
Ca: 1mmol/L=4.00784mg/dL.
In the setting interface, click “Set Time” button to enter into the
interface of Fig. 4.21.
Fig. 4.21
Click the left or right arrow key to select the date or time, if the cursor
selects one number, it is highlighted, directly input the corresponding
number to do the modification, after inputting the number, the cursor
automatically moves to the next number. Complete all the date and
time modifications successively.
After the modifications, click “Enter” to save and exit, then return to the
setting interface.
In the setting interface, click “Set COM” to enter into the interface of Fig.
49
Chapter 4 Interface Structure and Operation
Fig. 4.22
Baud rate: click “Baud” to switch between the baud rates. The factory
default setting is 9600. Customers are not suggested to modify it by
themselves.
50
Chapter 4 Interface Structure and Operation
In the setting interface, click “Printer” to enter into the interface of Fig.
4.23 to set the print mode.
Fig. 4.23
Directly click “Print” or “Cancel” to select the print mode. If select “Print”,
the printer will automatically print after the sample test or calibration, if
select “Cancel”, the printer will not print the results.
In the setting interface, click “Set Test” to enter into the interface of Fig.
4.24 to select the items to be tested.
51
Chapter 4 Interface Structure and Operation
Fig. 4.24
4.8.8 Exit
4.9 Maintenance
52
Chapter 4 Interface Structure and Operation
Fig. 4.25
In the maintenance interface, click "Set Vol" to enter into the interface
of Fig. 4.26.
Click the "5000" button after the “Sample” to change the sample
volume.
Click the "5000" button after the "Sam step" to adjust the steps
between the sample and the electrodes.
Click the "5000" button after the "Cleaning" to change the volume of
the cleaning solution.
Click the "5000" button after the "A/B" to change the STD A/B volume.
53
Chapter 4 Interface Structure and Operation
Click the "5000" button after the "A/B step" to adjust the steps between
the STD A/B and the electrodes.
Exit: click “Exit” to save the adjustment and return to the “Maintain”
interface
Fig. 4.26
54
Chapter 4 Interface Structure and Operation
Fig. 4.27
Deprotein
When used for a period of time, the slope value and stability of the
electrode will decrease. Perform this function to activate the electrodes
with activating solution to increase the electrode slope value and
enhance stability.
4.9.2.1.2 When the probe descends to the suction tank, the instrument
follows the specified program to wash the whole liquid system
automatically. The working status will be displayed on the left side.
55
Chapter 4 Interface Structure and Operation
Tube
4.9.2.2.2 The probe descends to the suction tank, and the tubes are
drained.
Fig. 4.28
56
Chapter 4 Interface Structure and Operation
Click “STD A” or “STD B”, the probe descends to the suction tank and
aspirates the corresponding liquid to measure the mV value according
to the specified program.
Exit: click “Exit”, the probe descends to the suction tank, and the
system returns to the “Maintain” interface.
4.9.4 Feed
When replacing the print paper, use the “Feed” function to easily install
the print paper, the specific operation is as follows:
Take out the print paper from the printer and put in new paper.
4.9.5 Exit
57
Chapter 5 Maintenance and Service
Soft tissue
Clean gauze
84 disinfectant
If not clean the outside of the sample probe after testing serum, the
residual serum will flow down and finally gathered in the needle mouth
to slowly desiccate, it may block the probe after a long period of time,
which makes it unable to perform calibration and measurement. So
remember to clean the outside of the probe with soft tissues (cannot
58
Chapter 5 Maintenance and Service
→“Cleaning” to
In the main menu interface, click “Maintain”
1
lift the probe from the suction tank;
Gently wipe the needle tip with soft tissue until there are no
2
visible attachments.
Note: Corrosion
Cleaning solution is chemical corrosive, protective
gloves should be worn while in operation.
59
Chapter 5 Maintenance and Service
The waste tube and its interface are located at the right side of the
analyzer, check the waste tube and waste bottle in the following order:
Open the front door of the analyzer, the peristaltic pump tube is located
at the downside, check the pump tubes in the following order:
60
Chapter 5 Maintenance and Service
Check the filling solution volume of each electrode, if lack of the filling
solution, it will greatly affect the test results, please refer to App. 4 for
the specific maintenance method.
5.3.1 Check the electrodes at least once a week, clean the dried
solution and sample, and keep the measuring electrode clean.
5.3.2 Check the filling solution volume of the electrodes. Make sure the
Ag/AgCl reference electrode is always immersed in the filling solution.
If the reference electrode filling solution is used too much, add the
filling solution by the syringe or replace the filling solution. Refer to App.
4 for the specific methods.
5.3.3 Immerse the electrodes into the deprotein solution for 5 minutes
to guarantee the electrodes’ state (refer to the bottle label for the usage
of the deprotein).
5.4.2 In order to make the probe and tubes play its good function,
make sure their inner surface is clean and smooth. Perform the
maintenance and service as required. It is recommended to do
maintenance at least once a month. If there is protein residue at the
inner surface of the tubes, follow the steps in the table below to do the
cleaning:
61
Chapter 5 Maintenance and Service
62
Chapter 6 Problems and Troubleshooting
63
Chapter 6 Problems and Troubleshooting
If the problems are not solved or there are other problems, please
contact the after-sales service unit.
64
Chapter 6 Problems and Troubleshooting
Possible
Error Solutions
reasons
Pump tube Replace with a new pump
damaged tube
Check whether the pipeline
Pipeline clogged
clogged or leaking
Not aspirate sample
Peristaltic pump Pump tube too tight or the
not working motor damaged
Pump tube too Replace a pump tube with
long an appropriate length
Distribution Replace with a new
valve positioning distribution valve positioning
sensor failure sensor
Distribution Replace with a new
valve shaft distribution valve shaft,
failure or check whether the valve is
Not aspirate STD
clogged clogged or not
A/B
The aspiration
position Replace with a new suction
seriously tank
abraded
Pipeline clogged Check whether the pipeline
or air leaking is clogged or leaking
Switch to manually adjust
Inadequate sample Liquid sensor the sampling mode to check
filling the electrodes failure if need to replace the liquid
leads to abnormity sensor or liquid tube
Inadequate Increase the sample
65
Chapter 6 Problems and Troubleshooting
If the problems are not solved or there are other problems, please
contact the after-sales service unit.
66
Chapter 6 Problems and Troubleshooting
67
Chapter 6 Problems and Troubleshooting
Sampling probe
Motor
68
Chapter 7 Common Consumables and Ordering Methods
Ordering Methods
69
Chapter 7 Common Consumables and Ordering Methods
Ordering methods:
70
Chapter 8 Transportation, Storage and the Outer Packing Instructions
8.1.2 Storage
71
Chapter 8 Transportation, Storage and the Outer Packing Instructions
72
Chapter 8 Transportation, Storage and the Outer Packing Instructions
Pictures are for reference only, please refer to the actual pictures on
the real packing box.
73
Chapter 9 Appendix
Chapter 9 Appendix
Open the printer cover and take down the roller (take it out horizontally
with care), then take out the remaining paper.
1) Put the paper roll into the feed slot, pay attention to the different
side of the thermal print paper, and make sure the thermal side is
downward. Then put the paper under the printer roller, make sure
the print paper trend is vertical to the printer roller. (In order to
avoid paper jam while printing). Please refer to Fig. 9.1.1
Fig. 9.1.1
74
Chapter 9 Appendix
While installing the printer roller, align the gear side first, then lay flat
the roller and click it to the right position.
The reagent room is located at the right side of the instrument. Open
the reagent room door to take out the old reagent.
Put in the new reagent, and insert the tube into the new reagent bottle,
notice that the letters on the tube should be consistent with the letters
on the bottle.
Make sure the tube end reaches the bottom of the reagent bottle.
Open the front panel of the instrument, in the lower right corner, there
are peristaltic pump, pump tube and tube connectors. Please refer to
Fig. 9.3.1.
75
Chapter 9 Appendix
Fig. 9.3.1
Take out the pump tube connectors from the baffle plate to remove the
pump tube.
Open the front panel of the instrument, there are aligned electrodes,
such as K, Na, etc.. Please refer to Fig. 9.4.1.
Take out the electrode module: Uplift the electrode module from the
slot to take out the whole electrode module.
Take out the electrodes: Loose the left and right spinning nut of
76
Chapter 9 Appendix
Fig. 9.4.1
After installing all electrodes, tighten the left and right spinning nuts of
the electrode module.
Put the electrode module into the slot, and connect the connectors with
tubes. Notice that the electrode connectors should be consistent with
the interface symbols;
77
Chapter 9 Appendix
Note: Pictures are for reference only, please see the real product for
detailed information.
78
Chapter 9 Appendix
The electrodes must be filled with filling solution before use and the
dedicated syringe needle (the metal part is 8-10mm in length) is
needed to add in the filling solution to avoid poking the electrode
membrane. The electrode chamber should be filled with 2/3 to 3/4
filling solution (see Fig. 9.5.2), then tighten the conductive pole, make
sure there are no bubbles between the filling solution and the
membrane. Clean and dry the electrode surface and install them
according to the electrode order.
While the electrode is in use, there will be pollution, poisoning and the
filling solution reducing, etc., reflected by the abnormal electrode drift
and electrode response (including electrode mV value abnormal and
electrode mV difference value abnormal). For the electrode pollution,
use the deprotein solution to clean the contaminant attached on the
electrode membrane, if seriously polluted, use the cotton thread to
clean the attachments by pulling it back and forth inside the electrode
hole (all the electrodes are strictly prohibited using hard object to poke).
If the filling solution decreased, toss out the remaining solution, then
refill with new filling solution.
79
Chapter 9 Appendix
9.6 MSCOMM
80
Chapter 9 Appendix
1 Frame header 02
1 Version 1
1 Instrument type 3
1 Instrument No. 0
12 Time ASCII 0-9
1 Sample type ASCII 0-3
5 No. ASCII 0-99999
13 ID ASCII 13 digits
1 Item name ASCII: K, Na, Cl, iCa, TCa, pH
3 Integer data ASCII 0-999
1 Decimal point ASCII .
2 Decimal data ASCII 0-99
3 Unit ASCII: mmol/L, mg/dL
1 Frame tail 03
81
Chapter 9 Appendix
Unit code:
2) All the sending data is ASCII code, if using serial port tool, switch
to the character format.
82
Chapter 9 Appendix
Sample test:
QC sample:
Analysis is as follows:
Version: 1
Instrument type: 3
No.: 00007
Unit: mg/dL
83
Chapter 9 Appendix
Connect the barcode reader to the COM RS232 2, the reader will
prompt a “di” alarm sound.
Set the serial port mode to Baud rate: 9600, Parity: None, Data Bit: 8,
as follows:
Complete setting
After completing the settings, enter into the “Measure” interface of Fig.
4.6, use the barcode reader to scan a 13-digit barcode, after hearing a
“di” alarm sound, check whether the displayed ID No. is consistent with
the barcode to be input, if not, reset it again.
84
Item No.: P01.91.300253-00