Professional Documents
Culture Documents
EXPERIMENT #1
LIQUID LEVEL CONTROL USING ON/OFF CONTROLLER
(inflow curve). Find a function in terms of system parameters to describe level change
of water in tank if height of water in tank is h0 and pump is off at the beginning of the
experiment (outflow curve).
4. If the tank is empty and a step reference input h = h0h(t) with height of h0 is applied to
the system, draw a rough graph showing how water level in the tank and pump
flowrate change.
5. Read the lab document and learn how the experiment setup is run.
2
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
3
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
1. Startup screen
It is the very first screen when the software is run (Figure 4). Check Table 2 for details.
4
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
1.Stop simulation
2.Start simulation
3.Experiment selection
4.Load picture to the graphics screen
5.Delete and restart graphics
Figure 6. Start stop simulation menu for 2 point closed loop control
5
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
Digital input/output
Save and print Save desired data (ASCII format) and print
6
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
Experiment 2:
i. Enter the reference value of liquid level in tank B101 (w) and deadband (range of
manip.).
ii. Start the simulation.
iii. As the experiment is running, change deadband after some time by increasing and
decreasing it such that you get at least 5 different value. Run the system for a while
within deadband after each increase or decrease. Observe the system output on
computer screen. Save the graphics of liquid height change and take a screean shot.
iv. Stop the simulation.
Experiment 3:
i. Enter the reference value of liquid level in tank B101 (w) and set deadband (range of
manip.) as zero.
ii. Start the simulation.
iii. Wait until tank is empty. Observe the system output on computer screen. Save the
graphics of liquid height change and take a screen shot.
iv. Stop the simulation.
Note: Do not ever change the opening of the valves during experiments.
5. REPORT
Unorganized reports in which experiment data are given in random order will not be
accepted. Report should hold qualities of a technical document with a proper proofreading.
Part of the grading of the report will depend on presentation, grammar, shape of graphs and
figures. You should consider the following in your report:
7
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
4
3
Select the connection port between computer and Easyport from COM menu at the
top. (COM1 usually) In case a different port is selected, software should be restarted.
There are small switches in the highlighted section (1) for observing and controlling
digital inputs and outputs. If mouse is placed on one of the switches for more than 2
seconds, a small note is shown displaying which output is controlled by that switch
(Figure 8).
Controls in the highlighted section (2) are used for adjusting signal levels of analog
outputs.
Highlighted section (3) shows 4 analog input channels. Channel numbers are displayed
on the left. Physical value can be calculated with the formula below using sensor data:
For example, if the sensed value is within 0-200 bar and the signal is between 0-10V,
factor should be 20. If 5 bar is assumed to be zero, offset should be -5. Thus, a new
offset should be set each time.
Table 4 contains further information about the channels.
1
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
The switch in the highlighted section (4) is used to select between software simulation
and experiment via Easyport. If any change is applied, you should save and restart the
software.
Devices connected to digital I/O signals can be seen in Table 5.
Devices connected to analog I/O signals can be seen in Table 6.
Input Output
Details Details
Signal Signal
2
MAK 305L SYSTEM DYNAMICS AND CONTROL LABORATORY
1. Spherical valve
2. Solenoid
3. 5/2 direction valve (V102)
4. Winding SYPAR type driver
5. Sensor box