You are on page 1of 1
31.8 Input of a temperature program requires setting of the following items, Start Temp. Dry 50 to 1000 °C Ash.. 50 to 2800 °C Atomiize 50 to 2800 °C Clean. 50 to 3000 °C (60 to 2800°C for other than Pyro Tube C HR and D HR) NOTE: The maximum temperature varies with the type of cuvette: 3000°C for Pyro Tube C HR and Pyro Tube D HR, 2800°C for Pyro Tube HR and Tube HR, 2600 °C for platform HR, and 2100 °C for Cup HR. Although entry is allowed up to 2800 °C on the window, temperature will not rise beyond the above limits due to the relationship between cuvette shape and resistance. © End Temp. Set the end temperature of dry and ash stages. You cannot enter each end temperature of atomize and clean stages because the temperature will be automatically set at the same temperature as for start. Dry 50 to 1000 °C Ash 50 to 2800 °C AOMIZE ...erene. 50 0 2800 °C Clean 50 to 3000 °C (50 to 2800°C for other than Pyro. Tube C HR and D HR) + Ramp Time Temperature can be raised linearly from the start temperature to end temperature in the set time. Input range Dry 1 to 100s Ash... 1 to 100 s