Professional Documents
Culture Documents
Operator’s Manual
Note: The product complies with the Council Directive 2011/65/EU.
© 2013-2018 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issued Date is 2018-12.
Shenzhen Mindray Bio-medical Electronics Co., Ltd. (hereinafter called Mindray) owns the
intellectual property rights to this Mindray product and this manual. This manual may refer to
information protected by copyright or patents and does not convey any license under the patent
rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
Mindray is strictly forbidden.
derivative work of this manual in any manner whatsoever without the written permission of
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only
if:
• all installation operations, expansions, changes, modifications and repairs of this product are
conducted by Mindray authorized personnel;
• the electrical installation of the relevant room complies with the applicable national and local
requirements;
• the product is used in accordance with the instructions for use.
I
This equipment must be operated by skilled/trained clinical professionals.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the
improper use or application of the product or the use of parts or accessories not approved by
Mindray or repairs by people other than Mindray authorized personnel.
II
Customer Service Department
Tel: 0049-40-2513175
Fax: 0049-40-255726
III
X-ray Protection
X-ray system may cause injury if not properly used. Accordingly, read thoroughly and
understand the instructions herein before attempting to place the system in operation.
Although this system is built to the highest safety standards and incorporates a high degree of
protection against X-radiation other than the useful beam, no practical design of system can
provide complete protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly, unwisely, or unknowingly
exposing themselves or others to X-radiation.
It is important that everyone working with X-radiation should be properly trained and take
adequate steps to insure protection against injury. The manufacturer assumes that all operator
and service personnel authorized to use, install, calibrate and maintain this system is cognizant
of the danger of excessive exposure to X-radiation, is sufficiently trained and has the required
knowledge for it. The system herein described is sold with the understanding that the
manufacturer, its agents, and representatives are not liable for injury or damage which may
result from exposure to X-radiation.
Various protective materials and devices are available. It is recommended to use such materials
and devices.
IV
Safety Precautions
1. Advisory Symbols
NOTICE are used regarding safety and other important instructions. Their application and
meaning are described below. Please understand their meanings clearly before reading this
manual.
Signal word Meaning
INDICATES AN IMMINENTLY HAZARDOUS SITUATION
WHICH, IF NOT HEEDED OR AVOIDED, WILL CAUSE
SERIOUS PERSONAL INJURY, DEATH OR
SUBSTANTIAL PROPERTY DAMAGE.
INDICATES A POTENTIALLY HAZARDOUS SITUATION
WHICH, IF NOT HEEDED OR AVOIDED, COULD CAUSE
SERIOUS PERSONAL INJURY, DEATH, OR
SUBSTANTIAL PROPERTY DAMAGE.
Indicates a potentially hazardous situation which, if not
heeded or avoided, will or could cause minor personal
injury or substantial property damage.
Alerts readers to pertinent facts and conditions. It
represents information that is important to know but
does not necessarily relate to possible injury or
damage to equipment.
2. Safety Symbols
The table below explains the safety symbols used in the system.
Symbol Description
V
Do not touch the hot lamp and the cover. Risk of burn.
Warning message
Error message
No sitting
Alternating current
Earth (ground)
Emergency stop
Loudspeaker
Date of Manufacture
VI
Manufacturer
Serial Number
3. Safety Precautions
Please observe the following precautions to ensure patient and operator safety when using this
system.
VII
• DO NOT OPEN THE COVERS AND FRONT PANEL OF THE SYSTEM. SHORT
CIRCUIT OR ELECTRIC SHOCK MAY RESULT WHEN THE SYSTEM
HARDWARE IS EXPOSED AND POWERED ON. ANY SERVICE OR UPGRADE
OF THE SYSTEM SHOULD BE CONDUCTED BY THE SERVICE PERSONNEL
TRAINED AND AUTHORIZED BY MINDARY.
• DO NOT allow the patient to contact the control parts of the system.
• DO NOT demount or replace any component of the system. Only
professional authorized maintenance personnel are allowed for
maintenance.
• DO NOT modify the system software; otherwise, it may lead to system failure
and image data loss. DO NOT use any software that is not supplied by
Mindray.
• DO NOT replace any hardware component of the system; otherwise, it may
lead to system failure.
VIII
Version Change Record
Version Date
V1.0 2013.08.28
V2.0 2013.10.21
V3.0 2013.12.31
V4.0 2014.05.14
V5.0 2014-08-10
V6.0 2015-12-10
Table of Contents
Section Page
Chapter 1 Safety ..................................................................................................................1
Chapter 2 System Components .........................................................................................7
2.1 Image Acquisition Workstation ....................................................................................... 8
2.1.1 Exposure Handswitch ....................................................................................... 10
2.2 X-ray Tube Support Assembly ......................................................................................11
2.2.1 X-ray Tube Assembly ........................................................................................ 12
2.2.2 Collimator.......................................................................................................... 18
2.3 Digital Wall Stand (Optional) ........................................................................................ 22
2.3.1 Vertical Movement of the Detector Box (Cassette Tray) .................................. 22
2.3.2 Lateral Positioning Bar (Optional)..................................................................... 23
2.3.3 Grid ................................................................................................................... 24
2.4 Integrated Floating Patient Table (Optional) ................................................................ 26
2.5 System Control Box ..................................................................................................... 29
2.6 Portable Detector ......................................................................................................... 30
2.6.1 Placing and Locking the Portable Detector ...................................................... 30
2.6.2 Maintaining the Portable Detector .................................................................... 31
2.7 Mobile Stand (Optional)................................................................................................ 33
2.7.1 Features ............................................................................................................ 33
2.7.2 Operation .......................................................................................................... 37
Chapter 3 System Startup and Shutdown ......................................................................39
3.1 System Startup ............................................................................................................. 39
3.2 System Shutdown ........................................................................................................ 39
3.3 DROC Software Overview ........................................................................................... 41
3.4 Login ............................................................................................................................ 42
3.5 Logout .......................................................................................................................... 43
Chapter 4 Patient Exam ....................................................................................................45
4.1 Patient Registration and Management......................................................................... 45
4.1.1 Patient Registration .......................................................................................... 46
4.1.2 Registered Information Management ............................................................... 53
4.1.3 Querying Patient Information ............................................................................ 56
4.1.4 Traditional Radiography.................................................................................... 58
4.2 EXAM & EXPOSURE................................................................................................... 59
4.2.1 Patient Information Display............................................................................... 60
4.2.2 View Operation ................................................................................................. 60
4.2.3 Setting Exposure Preferences .......................................................................... 65
4.2.4 Component Status Display ............................................................................... 72
4.2.5 Exposure ........................................................................................................... 74
4.2.6 Closing an Exam............................................................................................... 76
4.3 Image View and Process ............................................................................................. 78
4.3.1 Patient and Exposure Dose Information ........................................................... 79
4.3.2 Image Display ................................................................................................... 79
4.3.3 Image Operation ............................................................................................... 81
4.3.4 Image Post-processing ..................................................................................... 92
4.3.5 Image Confirmation......................................................................................... 102
Chapter 5 History Image ................................................................................................. 103
5.1 History Image Query .................................................................................................. 103
5.2 Editing Information ..................................................................................................... 105
5.3 Deleting Data .............................................................................................................. 105
5.4 Data Protection ........................................................................................................... 106
5.5 Data Backup ............................................................................................................... 106
5.6 Data Recovery ............................................................................................................ 107
5.6.1 Image Preview ................................................................................................ 108
5.6.2 Related Operation ........................................................................................... 109
5.7 Query/Retrieve (Q/R) Function (Optional).................................................................. 110
Chapter 6 Image Printing Management ........................................................................ 113
6.1 Image Adjustment....................................................................................................... 115
6.2 Zoom .......................................................................................................................... 116
6.3 Print ............................................................................................................................ 116
6.3.1 Setting Print Format ........................................................................................ 116
6.3.2 Selecting Print Orientation .............................................................................. 118
6.3.3 Setting Print Parameters ................................................................................. 119
6.3.4 Adding and Removing a Film .......................................................................... 120
6.3.5 Removing an Image ........................................................................................ 120
6.3.6 Turning Over Pages ........................................................................................ 121
6.3.7 Printing Image(s)............................................................................................. 121
6.4 Exit.............................................................................................................................. 121
Chapter 7 Manual Image Stitching (Optional) .............................................................. 123
7.1 Shifting the Images Translationally ............................................................................ 123
7.2 Selecting the Stitching Point ....................................................................................... 123
Chapter 8 System Configuration ................................................................................... 127
8.1 DICOM Service Configuration .................................................................................... 128
8.1.1 DICOM Local Configuration ............................................................................ 128
8.1.2 DICOM Worklist Configuration ........................................................................ 129
8.1.3 DICOM Storage Configuration ........................................................................ 132
8.1.4 DICOM Print Configuration ............................................................................. 134
8.1.5 DICOM MPPS Configuration (Optional) ......................................................... 137
8.1.6 DICOM Storage Commit (SC) Configuration (Optional) ................................. 138
8.1.7 DICOM Q/R Configuration (Optional) ............................................................. 139
8.2 DICOM Printer Configuration ..................................................................................... 140
8.3 Patient Information Display Configuration .................................................................. 142
8.4 Patient Information Print Configuration ...................................................................... 143
8.5 Log System Configuration .......................................................................................... 144
8.6 User Management Configuration ............................................................................... 145
8.7 Software Configuration............................................................................................... 147
8.8 Exam Configuration.................................................................................................... 156
8.9 Other Configuration .................................................................................................... 157
8.10 Backup Management ................................................................................................. 159
8.11 Network Configuration ............................................................................................... 161
8.12 License Information .................................................................................................... 162
Chapter 9 Parameter Manager .......................................................................................163
9.1 Configuring the View Definition .................................................................................. 164
9.1.1 Configuring the Exam Bodypart...................................................................... 164
9.1.2 Configuring the Sub-bodypart ......................................................................... 165
9.1.3 Configuring Views of the Sub-bodypart .......................................................... 168
9.1.4 Configuring a View.......................................................................................... 169
9.2 Configuring the View Parameter ................................................................................ 173
9.3 Configuring the Exposure Parameter......................................................................... 177
9.4 Configuring the Post-processing Parameter .............................................................. 180
9.5 Backup and Parameters Recovery ............................................................................ 184
Chapter 10 System Menu..................................................................................................187
10.1 DICOM Job Manager ................................................................................................. 187
10.2 Statistic ....................................................................................................................... 187
10.3 Update of PCB Software ............................................................................................ 188
10.4 Software Version and Copyright ................................................................................ 189
Chapter 11 Report Module (Optional) .............................................................................191
Chapter 12 Detector Calibration ......................................................................................193
Chapter 13 Periodic Maintenance....................................................................................197
13.1 Operator Tasks ........................................................................................................... 197
13.2 Service Tasks ............................................................................................................. 198
Safety
Chapter 1 Safety
This chapter explains the safety considerations, general equipment and patient related
precautions.
1.2 Safety
For continued safe use of this system, follow the instructions in this Manual.
Study this manual carefully before using the equipment and keep it at hand for
quick reference.
1
Safety
• MIND CLOSELY THE BODY PART (HANDS, FEET, FINGERS, ETC.) OF THE
PATIENT, AND PREVENT IT FROM BEING HURT BY THE MOVEMENTS OF
THE SYSTEM. PATIENT HANDS SHOULD BE KEPT AWAY FROM THE
MOBILE COMPONENTS OF THE SYSTEM.
2
Safety
1.3 Responsibilities
Always be alert to safety when you operate this equipment. You must be
familiar enough with the equipment to recognize any malfunctions that can be
a hazard. If a malfunction occurs or a safety problem is known to exist, do not
use this equipment until qualified personnel correct the problem.
It is the user’s responsibility to provide the means for audio and visual
communication between the operator and the patient.
3
Safety
Read the accompanying End User License Agreement (“EULA”) for operating
system carefully before using the system. If you do not agree to this EULA, do
not use the system. Any use of this system will constitute your reading and
agreement to this EULA.
Any object in the path of the primary beam produces secondary (scattered) radiation. The
intensity of the secondary radiation is dependent upon the energy and intensity of the primary
beam and the atomic number for the object material struck by the primary beam. Secondary
radiation may be of greater intensity than that of the radiation reaching the film. Take protective
measures to safeguard against it.
An effective protective measure is the use of lead shielding. To minimize dangerous exposure,
use such items as lead screens, lead impregnated gloves, aprons, thyroid collars, etc. The lead
screen should contain a minimum of 2.0 mm of lead or equivalent and personal protective
devices (aprons, gloves, etc.) must contain a minimum of 0.25 mm of lead or equivalent. For
confirmation of the local requirements at your site, refer to your “Local Radiation Protection
Rules” as provided by your Radiation Protection Advisor.
Before operation, persons qualified and authorized to operate this system should be familiar with
the Recommendations of the International Commission on Radiological Protection, contained in
Annals Number 60 of the ICRP, and should have been trained in use of the system.
Choose appropriate technical parameters for each program to reduce the exposure dose of
X-ray to the lowest level, yet reach the best effect of image for diagnose at the same time. This
system meets the requirements of IEC 60601-1-3 (for medical electrical equipment).
THE OPERATOR SHOULD MAKE THE DISTANCE FROM THE FOCAL SPOT TO
THE SKIN AS LARGE AS POSSIBLE, IN ORDER TO KEEP THE ABSORBED
DOSE TO THE PATIENT AS LOW AS REASONABLY ACHIEVABLE.
4
Safety
The most effective method of determining whether or not the existing protective measures are
adequate is the use of instruments to measure the exposure. These measurements should be
taken at all locations where the operator or any portion of the body may be exposed. Exposure
must never exceed the accepted tolerable dose.
A frequently used, but less accurate, method of determining the amount of exposure is the
placement of film at strategic locations. After a specified period of time, develop the film to
determine the amount of radiation.
A common method of determining whether personnel have been exposed to excessive radiation
is the use of personal radiation dosimeters. These consist of X-ray sensitive film or
thermoluminescent material enclosed within a holder that may be worn on the body. Even
though this device only measures the radiation which reaches the area of the body on which they
are worn, they do provide a reasonable indication of the amount of radiation received.
5
Safety
1.8 Contraindication
Pregnant women and infants should avoid the examination by this system as much as possible.
If there is any justified reason for them to conduct the examination, take protective actions as
much as possible.
6
System Components
There are two kinds of configuration for the DigiEye280 system: the DR system.equipped with a
portable detector and the system equipped with no detector (an IP panel or cassette needs to be
provided by the user).
Their only difference is the detector boxes on the wall stand and in the table assembly.
The standard room layout for the DR system equipped with a detector is reverse installation
mode (see the figure below). That is, the wall stand is on the left side of the table assembly.
For the system equipped with no detector, the standard room layout is normal installation mode
(see the figure below). That is, the wall stand is on the right side of the table assembly.
7
System Components
8
System Components
• Check the system clock of the computer periodically, to ensure the accurate
time of patient examination.
9
System Components
Position Description
OFF
PREPARE
EXPOSE
Operate the handswitch as below. Prepare and record the exposure process:
1. Verify that the patient and the operator console are ready for making exposures.
2. Press the Prep/Expose button to the PREPARE position. The corresponding icon on the
Exam & Exposure Screen is shown as . It shows the generator starts to produce X-ray.
At the same time, the buzzer will make a sound. The progress bar on the
Exam & Exposure Screen shows the exposure progress.
10
System Components
4. Release the Prep/Expose button after the exposure is completed, or when it reaches the
preset exposure time. When the exposure is completed, the buzzer stops buzzing.
Release the Prep/Expose button only after the buzzer stops buzzing, to avoid
inadequate dosage.
Adjust the field size of the collimator before pressing the Prep/ Expose button,
to protect the patient from absorbing excessive X-ray radiation.
11
System Components
Tube stand
X-ray Tube
Collimator
SID scale
Angle
indicator
Linear
movement
control button
Angulation
control button Operation
handle
To rotate the tube-collimator assembly on the vertical plane, press the angulation control button
on the left operation handle. To move the tube-collimator assembly vertically along the tube
stand or move the tube stand laterally along the ground rail, press the linear movement control
There is an ionizing radiation warning label on the left end cover of the X-ray tube and a high
12
System Components
Ionizing
radiation
warning label
High
temperature
warning label
To change the angulation of the X-ray tube assembly, press the angulation control button
on the left operation handle, grasp the handles and rotate the X-ray tube assembly on
the vertical plane. When the desired angulation is reached, release the angulation control
button to lock the angulation.
Rotation of the tube support
The tube support can be rotated ±180° manually on the horizontal plane along the tube
stand.
13
System Components
To rotate the tube support, step and hold down the foot pedal on the bottom of the tube
stand, then grasp the handles on the X-ray tube control panel and rotate the tube support
manually on the horizontal plane along the tube stand. When the desired position is reached,
release the foot pedal to lock the movement.
Note: The rotation is automatically notched every 90°.
Tube support
Tube stand
Foot pedal
14
System Components
There are two lock bolts on the rear top of the X-ray tube assembly (see the figure below).
To tilt the X-ray tube assembly, loosen the two lock bolts. When the desired angle is
reached, fasten the two lock bolts to lock the rotation.
Note: Lock the X-ray tube assembly tightly after tilting.
Lock bolts
Tilt angle
indicator
15
System Components
To move the X-ray tube assembly up and down along the tube stand, press the linear movement
control button on the right operation handle. Hold the handles on the panel and move the
X-ray tube assembly to the desired position. Then, release the button to lock the movement.
Release the control button only after the movement is stopped, to avoid severe
impact.
KEEP THE PATIENT OR ANY OTHER OBJECT IN THE ROOM AWAY FROM THE
MOVEMENT RANGE DURING THE MOVEMENT, TO AVOID EQUIPMENT
DAMAGES OR BODILY INJURIES.
There are SID scale and a label on the lateral side of the tube stand.
SID label
16
System Components
To move the tube stand transversely, press the linear movement control button on the right
operation handle. Hold the handles on the panel and move the tube stand to the desired position.
Then, release the button to lock the movement.
NOTE THE POTENTIAL PINCH POINT WHEN MOVING THE TUBE STAND
TRANSVERSELY.
17
System Components
2.2.2 Collimator
The operator may adjust the radiation field size to the anatomy through the collimator. This
section describes the controls and basic features of the collimator.
Do not touch the collimator rear side since the light bulb inside may be too hot.
5
4
2 6
1. Collimator transverse control knob: adjusting knob for formatting the X-direction radiation
field
2. Collimator longitudinal control knob: adjusting knob for formatting the Y-direction radiation
18
System Components
field
Do not turn on the collimator light repeatedly or keep it on for long time to avoid
overheating of the lamp inside.
4. Additional filter fix lever: Turn the fix lever and select the filter with desired thickness to
insert. The filter is 0.5 mm Al or 1.0 mm Al thick. If two filters are used together, the
thickness is 1.5 mm Al.
5. Rotation lock knob: Dial the lock knob to the right to unlock the collimator. Turn the
collimator to the desired position and then dial the lock knob back. The collimator can be
rotated along the beam axis. There are detents for rotation at every 15°.
Left 0º Right
7. DAP holder
After the SID is set, press the collimator light button③. Align the light field to the exposure field of
the patient. Adjust the field size according to the selected film. When the light field is centered
with the radiation field center, you can make exposures now.
Note: When the SID is 70 cm, the minimum radiation field is smaller than 10 mm×10 mm and the
maximum radiation field is greater than 280 mm×280 mm. When the SID is 200 cm, the minimum
19
System Components
radiation field is smaller than 10 mm×10 mm and the maximum radiation field is greater than 430
mm×430 mm.
Do not keep the collimator light on for a long time; otherwise, the temperature
of the cover around the light bulb is high. It may also shorten the service life of
the bulb.
Before replacing the light bulb, turn off the collimator light first, wait for five
minutes (or ten minutes in the event of a fan failure), and verify that the lamp
cover and the light have been fully cooled down.
1. Unscrew the two screws on the collimator rear cover by a cross screwdriver and then
detach the collimator rear cover.
2. Verify that the lamp cover has been fully cooled down. Detach the lamp cover by the cross
screwdriver.
20
System Components
3. Verify that the light bulb has been fully cooled down. Hold the bottom of the light bulb and
unplug it. If the light bulb is too tight to unplug, slightly unscrew the screws on the bottom of
the lamp holder.
4. Wear gloves before handling the new bulb to protect it from getting stained.
5. Insert the new bulb into the lamp holder. Note: It must be inserted vertically to the end.
6. Mount the lamp cover. Check whether the light works normally and whether there is any
offset between the light field and the X-ray field.
7. In the event of any offset, wait for cooling down of the lamp cover and the light bulb and
then check whether the light bulb is inserted vertically to the end.
21
System Components
Detector box
(cassette tray)
The detector box can move up and down along the wall stand. The wall stand can be placed on
the left or right side of the integrated patient table.
The digital wall stand can be configured with a fixed detector or portable detector. Traditional
cassette tray can be adopted here to make exposures.
Note: This system supports the cassette tray in compliance with the ISO4090 standard.
Make sure that the detector box is fastened before making exposures.
22
System Components
Detent lever
The detent lever is used to lock the movement of the detector box. To move the detector box,
turn the detent lever on the side of the detector box to the horizontal position. Pull the detector
box up or down along the wall stand. When the detector box reaches the desired position, turn
the detent lever to the vertical position clockwise to lock the movement.
Note: The movement can be locked or unlocked no matter the detector is inside the detector
box.
There is a slot (see the figure above) on each side of the detector box for inserting the bar. To
mount the lateral positioning bar, grasp and insert it to the slot. To remove the bar, just grasp
and pull it out. To rotate the positioning bar for 90°, grasp and raise the bar up. Rotate it to the
90° direction and then insert it to the slot.
23
System Components
Lateral
positioning bar
When positioning the patient in front of the wall stand, instruct the patient to
use the handgrip or lateral positioning bar.
• IF THE DETECTOR BOX IS LOWERED WITH THE LATERAL BAR, THE BAR
MAY HIT THE HEAD OF THE PATIENT.
• The patient cannot pull the lateral bar downward with great force. Otherwise,
the lock may be released and the detector box may drop slowly.
2.3.3 Grid
There is a slot in the detector box of the digital wall stand or the integrated patient table for
placing the grid. The system can automatically detect whether the grid is inserted and its
corresponding specifications.
24
System Components
Grid
The front side (with a central line and specification mark) of the grid should face
the X-ray tube during installation.
25
System Components
Place a protective cover over the patient table first and then lay down the
patient to avoid direct contacts.
Tabletop
Lock button of
the detector box
Detector box (table cassette tray) Foot pedal for tabletop floating
1. When the power button on the lateral side of the integrated table is not pressed down, the
power LED light on the control box is off; the tabletop and the table detector box can be
moved freely in the corresponding direction.
2. When the power button on the lateral side of the integrated table is pressed down, the power
LED light on the control box is on.
a) To float the tabletop of the integrated floating patient table longitudinally or transversely,
step and hold down the foot pedal on the front end of the integrated table. To lock the
floating of the tabletop, release the foot pedal.
b) When the lock button on the lateral side of the detector box is not pressed down, grasp
the handle of the detector box and move the detector box transversely. When the
detent mechanism on the tube stand is snapped to the slot of the detector box rear side,
the table detector box can be moved along with the tube stand.
c) Press down the lock button of the detector box, the detector box is locked and cannot
be moved freely. When the tube stand is moved transversely, the detector box is
locked and the detent mechanism is ejected from the slot automatically.
The detector box is under the tabletop of the integrated floating patient table and can be inserted
with a grid. The grid can be loaded or unloaded manually.
The body of the stationary lifting patient table can be equipped with a generator.
To prevent patient injuries when getting on or off the floating tabletop, move the X-ray tube
assembly up and the tube stand to one side.
26
System Components
• BEFORE THE PATIENT GETS ON/OFF THE PATIENT TABLE, MAKE SURE
THAT THE PATIENT TABLE HAS BEEN LOCKED. DO NOT STEP ON THE
FOOT PEDAL BY ACCIDENT.
There is a Finger Pinch warning label on the middle of the front beam of the tabletop.
27
System Components
WHEN INSERTING THE DETECTOR INTO THE DETECTOR BOX, NOTE THE
POTENTIAL FINGER PINCH HAZARD.
28
System Components
Power
PREP LED light indicator
EXPOSURE
LED light
Emergency
OFF switch
Power
switch
Press the switch to power on the system and the switch to power off the system.
To power off the system forcibly, press and hold down the switch for several seconds.
If the power indicator is blue, the system is in working state; if the power indicator is off, the
system is not in working state.
Press down the PREP handswitch button, the green exposure ready indicator is on. Fully press
down the handswitch button, the yellow exposure indicator is on. After the exposure, the
indicator is off.
To stop exposure in case of an emergency, press the red emergency OFF switch on the control
box.
29
System Components
• When loading or unloading the detector, hold the detector with hands to
avoid accidental drop of the detector.
• When inserting the portable detector to the detector box on the wall stand or
the integrated floating patient table, put the detector cable inside the slot to
protect it from being pressed.
• When not in use, put the detector in the detector stand (optional), detector
box, or other container supplied by Mindray. Do not put the detector on the
border position or lean it against the wall or table. Otherwise, the detector
may fall down unexpectedly.
• When the patient is in contact with the detector, do not use the defibrillator.
30
System Components
radiography.
Place the portable detector in the detector box drawer according to the detector placement label.
Then lock the portable detector as below.
When the portable detector is placed and locked in the detector box on the wall stand.
DO NOT TIDY THE DETECTOR CABLE WHEN THE SYSTEM IS POWERED UP.
TO PULL OUT THE DETECTOR CABLE, POWER OFF THE SYSTEM FIRST.
RECONNECT THE DETECTOR CABLE AFTER TIDYING THE CABLE. BEFORE
POWER-ON OF THE SYSTEM, VERIFY THAT THE CONNECTION OF DETECTOR
CABLE IS CORRECT.
31
System Components
• Do not overbend or fold the detector cable, especially the end connecting
the detector and the system.
To prolong the service life of the detector, clean the detector surface, detector cable, and grid
after each radiography examination. For details, refer to Chapter 13 of this manual.
Before each cleaning, the detector needs to be dried up for at least 60 seconds.
• Keep the detector and the grid dry before each use.
• Do not put the cleaning cloth on the detector or the grid for above 60
seconds.
32
System Components
Backplate
Handrail
Step stool
Base
2.7.1 Features
1. The caster brake of the mobile stand base is featured with four-caster interlocking system.
When the red brake pedal is depressed, all the four casters of the stand are locked and the
stand is irremovable. When the green pedal is depressed, the casters are all unlocked and
33
System Components
Green pedal
Red pedal
2. For easier patient positioning, this mobile stand is equipped with fixing strap connector,
infusion and arms hanger connector.
Fixing strap
connector
3. Handrails can be moved up or down vertically. The following figure shows the limit positions.
The backplate is equipped with anti-collision rubber strips to avoid collision between the
mobile stand backplate and the detector assembly. As a result, it is convenient for the stand
to align with the detector assembly.
34
System Components
Arms hanger
Anti-collision
Handrail upper rubber strip
limit position
Handrail lower
limit position
4. Handrails can be locked with the pin as shown in the figure below. After pulling out the pin,
you can move the handrail vertically. When it is floated to a proper pin hole position, loosen
the pin and insert it into the hole to lock the handrail.
Pin
Handrail
5. There is a stitching scale on the mobile stand backplate to help doctor determine the image
stitching position. On the backplate, there are four mounting holes. The doctor can adjust the
scale mounting position depending on the patient’s bodypart to be examined. For details of
the mounting holes, refer to the figure below (indicated by circles).
35
System Components
Stitching scale
6. There are antislip foot pads on both the base and the step stool.
36
System Components
2.7.2 Operation
1. Move the mobile stand to the desired position:
a) Depress the green pedal to unlock the casters;
b) Move the stand in front of the detector and make sure that the anti-collision rubber strips
are located outside the imaging region;
c) After adjusting the position, press down the red pedal with foot to lock the casters.
Green pedal
Red pedal
37
System Components
Bolt
Handrails
While installing the stitching scale, prevent it from overlapping with the
patient’s bodypart to be examined in the projection direction.
38
System Startup and Shutdown
This chapter explains the startup and shutdown procedures for the system.
• Before system startup, make sure that the working environment of the room
meets the requirements of Appendix A.2 and has been kept for over 10
minutes.
• After initial startup of the system, warm up the detector for at least 30
minutes.
After power indicators on the workstation control box are on, press the power switch on the
control box. The following components will be powered on or started:
• The image acquisition workstation starts up. When the DROC login screen is displayed, the
image acquisition workstation is ready.
• The generator starts up. If the generator beeps for alarm, find the corresponding error code
from the DROC screen and refer to Installation and Maintenance Manual for troubleshooting.
No alarm indicates normal startup.
• The image acquisition workstation is powered on. The monitor shows the startup process.
When the DROC login screen appears, the workstation is ready.
Do not shut down the system during image saving, sending or recording or
online upgrade of board programs.
• In power-on status, press the power switch on the control box. The system is closed.
• In power-on status, press the power button on the image acquisition workstation. Exit from
the DROC software. Then, the image acquisition workstation is turned off.
39
System Startup and Shutdown
detector.
40
System Startup and Shutdown
Function
Icons
The columns on the bottom of the screen respectively display: login name, prompt message
(including subsystem operational status and error messages), system status, and system time.
41
System Startup and Shutdown
• indicates the hard disk space. If the free space is less than 2G, the icon would blink.
Move the cursor to the icon, it would prompt: Insufficient disk space. If a new image is to be
saved, the message would prompt: “Insufficient disk space, delete history data automatically
or not?” If yes, part images will be deleted according to the automatic delete strategy.
3.4 Login
Click and enter the Patient Management Screen of the DROC software.
To ensure image quality, wait for at least two minutes on the Patient
Management Screen after the startup of the DROC software.
If the system is not authorized or the authorization has expired, the
corresponding prompts will be displayed on the login screen. The user
cannot make exposure or post-process images.
42
System Startup and Shutdown
3.5 Logout
Click on the right of the DROC interface. The logout screen is displayed.
Click to log out as the current user. The logout screen is displayed.
Enter the correct user name and password, and then click .
Click to close the DROC software and return to Windows OS (Note: Only the
43
System Startup and Shutdown
44
Patient Exam
The Registration & Exam Screen is the starting point for patient set up and selecting procedures
for acquisition.
The Registration & Exam Screen covers the whole process of patient exams, including Patient
Registration & Management Screen, Exam & Exposure Screen, Image View Screen, and Image
Printing Screen.
If the detector has not been calibrated for over one month, the bottom message column of the
screen will show: “Latest calibration has expired for XX days. It is suggested to re-calibrate the
detector.” For specific operation of detector calibration, refer to Chapter 12.
Click the icon on the right side of the screen. The Registration & Exam Screen is
displayed.
Note: Image stitching is an optional function (an advanced option) for patient exam.
45
Patient Exam
Note: The system forbids radiographic exam without the patient registration procedure; however,
the traditional radiography mode is an exception.
Patient registration can be classified into three categories: non-local registration, local
46
Patient Exam
Click or update the worklist to acquire the patient information from the Worklist
for registration.
To enable Auto Refresh function of the Worklist, refer to Section 8.7.
Click to create the registration of patient on the local workstation. The screen is
displayed as below.
D E F
The Patient Local Registration Screen mainly includes the following areas:
A. patient query area
B. patient list area
C. operation area
D. patient information registration area
E. exam bodyparts list area
F. Registration button area
1. Firstly, enter the patient information at the patient information registration area (zone D in
47
Patient Exam
It is subject to the constraints: The characters can only be numbers, letters or two-byte
characters used by the Oriental countries. The length should be no longer than 64
bytes.
Before entering the PID, check whether it has been used before.
• Accession #: the accession number registered for the patient’s current check
It is subject to the constraints: The characters can only be numbers, letters or two-byte
characters used by the Oriental countries. The length should be no longer than 64
bytes.
• Patient Name: the full name of the patient, including Last Name, First Name and Middle
Name
It is subject to the constraints: The characters can only be numbers, letters or two-byte
characters used by the Oriental countries. The length should be no longer than 64
bytes.
• Operator: the name of the operator or technician who performs the exam, which can be
selected from the drop-down list box
It is subject to the constraints: The characters can only be numbers, letters or two-byte
characters used by the Oriental countries. The length should be no longer than 64
bytes.
• Gender: the sex of patient, select from one of the following:
• Age: enter the number directly, or the age can be calculated automatically according to
the patient’s birth date.
48
Patient Exam
If the patient is up to or over 1 year (1 Y), enter the number directly; if the patient is less
than 1 year, click and select “M” (month) or “D” (day) from the drop-down list box,
as shown below:
Once the button is clicked for registration, the patient’s age will be
updated automatically according to the entered birth date. The final age is
determined by the birth date, even though the number is entered manually.
• Birth Date: enter the patient’s birth date in the format yyyy/mm/dd. The birth year can be
calculated automatically according to the patient’s age.
Rest the cursor on the field of month, date, or year, and click or to increase or
decrease the number.
If only the Age is entered, the birth year will be generated automatically
according to the Age upon the click of . The birth month and date
• Physician: the name of the physician who applied for the exam, which can be selected
from the drop-down list box
It is subject to the constraints: The characters can only be numbers, letters or two-byte
characters used by the Oriental countries. The length should be no longer than 64
bytes.
2. Secondly, select the exam bodypart (bodypart to be examed) from the exam bodyparts list
(zone E in the red frame).
49
Patient Exam
A. Select the exam bodypart: Select the anatomical category of bodyparts from the list in
the left column, and then click to open the category. The category expands to show
the available exam bodyparts. Click to close the expanded items.
B. Add the exam bodypart: Click the box at the left of the exam bodypart name, and a
checkmark √ appears in the box, which indicates that the exam bodypart is selected.
The exam bodypart name then appears in the Selected Bodypart list in the right column.
To de-select the exam bodypart, click √ inside (left column) or double-click the
selected exam bodypart in the right column, and the bodypart name is removed from
the Selected Exam Bodypart list.
50
Patient Exam
register a new patient and start an exam. To start an exam for a registered patient, click
or double-click the patient record in the patient list (zone B in the red frame)
An exam bodypart must be selected after the input of the patient information;
otherwise, a message box will prompt: “Please select an exam bodypart!” if you
click .
X-ray exam immediately. Click to access the Exposure & Exam Screen directly.
The system will automatically generate patient ID, accession number, patient name and preset
exam bodypart for the patient, and show the information in the patient list (as shown below).
• Patient ID: consisting of the prefix (such as “PID”) and numbers (a date and time stamp
For example, “PID090115160728” means that the exam was initiated in year 09, month
01, day 15, hour 16, minute 07, and second 28. The following numbers are the same
format.
• Accession #: consisting of the prefix (such as “AN”) and numbers
• Patient Name: consisting of the prefix (such as “Emg”) and numbers
• Birth Date: the registration date, according to the local time
• Age: 0 Y
• Registration Time: the registration time displayed in the computer
• Physician: the name of the physician who applied for the exam
• Number of exam bodyparts: 1 (default setting)
Note: The prefix of Patient ID or Accession Number (such as “PID” and “AN”) can be configured.
Refer to Section 8.8 for details.
51
Patient Exam
52
Patient Exam
You can only edit or delete the registered information of unchecked patient on
the Patient Registration and Management Screen.
53
Patient Exam
The original patient information and record cannot be recovered after the edit!
• For X-rayed patient, the patient information should be edited at the History
Image Screen.
Select the patient record to be edited from the patient list; then click at the left
After editing the information manually, click to save the change and quit. Click
54
Patient Exam
Select the patient record to be deleted from the patient list, and then click to delete
the patient record and information. Before the delete, the following message box appears for
confirmation.
To delete the X-rayed patient’s information, switch to the History Image Screen.
55
Patient Exam
• Patient ID: Enter the patient ID as query criterion into the text box. The asterisk (*) or
blank is also applicable for blind search.
• Accession #: Enter the exam’s accession number as query criterion into the text box.
The asterisk (*) or blank is also applicable for blind search.
• Patient Name: Enter the patient’s name as query criterion into the text box. The asterisk
(*) or blank is also applicable for blind search.
• Period: Click to select the registration date for query from the drop-down list box.
If the “User defined” option is selected for the date, the specified range of time can be
set for query.
• Exam Status: Click to select the exam status for query from the drop-down list box.
1. Select a completed record from the patient list and then click . The following
dialog box is displayed.
56
Patient Exam
• To save the new exposure image and the previous exposure image in the same
exam, click . Then, only one record will be displayed during query.
• To save the new exposure image as a new exam, click . Then, two
records will be displayed during query: record of the new exposure and record of
the previous exposure.
2. Click and all matches will be displayed in the patient list, as shown below.
57
Patient Exam
radiography mode needs no installed FPD (flat panel detector) or CCD, and allows traditional
cassette or IP board to receive X-ray exposure.
In traditional radiography mode, the screen does not show patient information, examed bodypart
name, and AEC exposure mode. Operations, such as image acquisition, viewing and printing,
are unavailable in traditional radiography mode.
The traditional radiography mode can be set in “System Preferences -> Software Configuration
-> Exam process settings", as shown below.
Only when “Traditional radiography” is selected, the button will appear on the
58
Patient Exam
• Select the patient record from the list, double click the record or click .
radiography mode.)
After accessing the Exam & Exposure Screen, the system automatically loads the patient’s
information.
B
D F
59
Patient Exam
1. Click in front of the exam bodypart to expand all the available views (such as PA and
AP) of the bodypart. Click to close the expanded items.
2. Click the desired view of the bodypart to be X-rayed from the list, which will be shown in
blue once selected. The view map area above the view list displays the view name and
view map.
60
Patient Exam
Note: When the view is selected, the preset exposure parameters will be shown in the parameter
setting area (Zone C in red frame) and the generator will be set up.
If the view of the body has been X-rayed, the view name in the view list will be shown in bold, as
shown below.
Note: The X-rayed view can be exposed repeatedly and the image can be saved.
Note: If the system is not equipped with a detector, the view list loads all views automatically.
61
Patient Exam
Note: If the system is not equipped with a detector, this function is disabled (The
1. Click above the right of the view list, and a window appears for adding a new
bodypart.
62
Patient Exam
2. Select the anatomical category of bodyparts from the list in the left column, and then click
to open the category. The category expands to show the available exam bodyparts.
3. Click the box at the left of the exam bodypart name, and a checkmark √ appears in the
box, which indicates that the exam bodypart is selected. The exam bodypart name then
appears in the Selected Exam Bodypart list in the right column. To de-select the exam
bodypart, click √ inside (left column) or double-click the selected exam bodypart in the
right column, and the bodypart name is removed from the Selected Exam Bodypart list.
4. Click . The exam bodypart is added to the exam procedure of the patient. Click
Note: If the system is not equipped with a detector, this function is disabled (The
1. Select an exam bodypart from the list, and click to delete the exam bodypart.
delete.
63
Patient Exam
64
Patient Exam
The system is set to digital radiography mode by default. If you click the
There are default exposure parameters for each exam bodypart. Make manual adjustments if
necessary.
Note: If the image acquisition workstation is equipped with a Dose Area Product (DAP, optional).
2
The DAP information is displayed on the DICOM header tag (0018, 115e) (unit: dGy-cm ).
• If the system is not equipped with an ion chamber or AEC exposure is not
selected, modify the exposure preferences according to the body type of the
patient.
• Do not start exposure before setting proper exposure preferences for the
patient.
After changing the exposure preferences, you (with the administrator’s authority) may save the
If the preferences are changed by the administrator, the saved preferences will
be the default exposure preferences automatically next time and may be
unrecoverable.
65
Patient Exam
In AEC mode, the following exposure preferences are adjustable: kV, mA, patient size, focal spot,
ion chamber, and image density.
66
Patient Exam
The higher the kV and mA values are, the greater the radiation dosage received
by the patient will be.
Click or on the right of the field to increase or decrease the kV value by one unit (1
kVp). The kV cannot be adjusted once it reaches the max or min value. It is the same with the
mA.
In AEC mode, the mAs and ms are not adjustable, and the buttons are in gray, such as .
Click the corresponding icon (left side: small focal spot, right side: large focal spot) to select the
desired focal spot. The selected icon will be shown in deep blue, such as .
Selecting correct patient size is quite important. Incorrect selection may lead to
excessive radiation dosage or repeated exposures.
67
Patient Exam
Click the corresponding icon to select the desired patient size. Available options are: Pediatric,
Thin Adult, Medium Adult, and Fat Adult. The system default is Medium Adult. The selected icon
L R
Click the ion chamber icon (L: left/ C: center/ R: right) and the selected ion chamber will be
The bodypart to be examed should cover the selected ion chamber to achieve
effective exposure. Otherwise, it might prolong or shorten the exposure time.
Note: If no ion chamber is selected, the system will alarm and disable the exposure.
The higher the AEC density is, the greater the radiation dosage received by the
patient will be.
Density Setting:
Density Setting Percentage of mAs
+4 244%
+3 195%
+2 156%
+1 125%
68
Patient Exam
0 100%
-1 80%
-2 64%
-3 51%
-4 41%
Note: The default density (0) is set according to normal contrast and density demand. The
density should be enhanced for big and fat patients, whereas the density should be reduced for
small and thin patients. If the image shows obvious mottle noise, you may adjust the density to
suppress the noise. If the image is too dark or bright, you need to adjust the image processing
parameters instead of the AEC density.
The smaller the film sensitivity is, the greater the radiation dosage received by
the patient will be.
The range of film sensitivity is 100 to 1200. The setting of film sensitivity is related to the selected
Sensitivity Setting:
Sensitivity Setting Percentage of mAs
100 400%
200 200%
300 133%
400 100%
500 80%
600 67%
700 57%
800 50%
900 44%
1000 40%
1100 36%
1200 33%
69
Patient Exam
The higher the mA and ms value are, the greater the radiation dosage received
by the patient will be.
Click to select the manual exposure mode. The button is then shown as .
In TIME mode, the following exposure preferences are adjustable: kV, mA, ms, patient size and
focal spot. The adjustment of kV, focal spot and patient size is the same with that in AEC mode.
Refer to Section 4.2.3.1 for details.
The following text mainly explains how to adjust exposure dose by altering mA and ms manually.
Click or on the right of the mA and ms display field to adjust the mA and ms. The mA
or ms will change to the next value supported by the generator upon each click.
In TIME mode, the mAs (=mA × mS) is not adjustable. The adjustment to the mA and ms
respectively will lead to the change of the mAs value.
When the exposure time reaches the set ms, the system stops emitting X-ray.
The higher the kV and mAs values are, the greater the radiation dosage
received by the patient will be.
Click the button to select the manual exposure mode. The clicked button is shown
as .
In mAs mode, the following exposure preferences are adjustable: kV, mAs, patient size and focal
spot. The adjustment of kV, focal spot and patient size is the same with that in AEC mode and
TIME mode. Refer to Section 4.2.3.1 for details.
The following text mainly explains how to adjust exposure dose by altering mA and ms manually.
Click or on the right of the mAs display field to adjust the mAs. In mAs mode, mA
70
Patient Exam
71
Patient Exam
shown as .
If the grid status does not match the SID, prompt message will appear on the grid status
icon.
The heat capacity icon shows the heat capacity of the tube before cooling down. A
low percentage indicates that the tube is relatively cool and is ready for successive exposures.
If the heat capacity of the tube exceeds the maximum level, a warning message with (such
as “Heat capacity is abnormal”) will be displayed on the icon. The X-ray tube and the generator
will be disabled, and the system will forbid making exposures.
If so, release the handswitch and wait until the X-ray tube cools down. Proceed with X-ray
72
Patient Exam
: The detector/film is on the wall stand. Select the ion chamber of the wall stand for AEC
exposures.
: The detector/film is in the integrated patient table. Select the ion chamber of the table
for AEC exposures.
73
Patient Exam
4.2.5 Exposure
Note: If the system is not equipped with a detector, this function is disabled.
1. In digital radiography, press the expose button on the handswitch to make exposure and
image acquisition.
: It is not allowed to make exposure. (Refer to the message field for reasons, such as
appear in the message field, and the status icon will be or/ and .
If the self-recovery error occurs on the generator, it will disappear automatically in several
seconds. If the self-recovery fails, click the button to remove the error manually. If
the error still exists, restart the generator.
If the patient is positioned in front of the wall stand, have the patient suspend
respiration, if required.
74
Patient Exam
2. Hold down the expose button of the handswitch. After several seconds, a "beep" prompt
tone is heard and the exposure starts. Then, release the button.
Note: If the detector has not been used for a preset period of time, it will be in a dormant state to
save energy. Open a patient’s exam can “awake” the detector, and it takes several seconds to
recover and prepare the detector.
If the heat capacity of the X-ray tube exceeds the maximum limit, the X-ray tube and the
generator will be unusable until the capacity returns to normal range. Related message will pop
out.
If the patient is positioned in front of the wall stand, have the patient resume
normal respiration.
3. After the exposure, the preview of the acquired image appears in the image palette at the
bottom of the screen. The palette displays not more than 6 image previews at a time. If
there are over 6 images in the series, a scrollbar appears at the bottom for the view of the
rest images.
75
Patient Exam
After all images of the current exam are acquired, click to close the current exam.
When the exam is closed, the images of the current exam will be sent to PACS, and the screen
switches to the Patient Registration & Management Screen. The exam status of the patient will
be changed from SCHEDULED to COMPLETED. Another exam of the patient or the exam of
another patient may then be opened.
You may also click to close the current exam without finishing the acquisition of all
images (or even without acquiring any image). A check box will appear for confirmation. You may
reopen the exam to continue the image acquisition.
Note: If the current exam is closed and no new patient or exam is opened, the detector may
activate calibration preparation. If no exam is opened for a relatively long period of time, the
detector enables dormancy function to save the energy. Therefore, it is suggested to close the
current exam if the image acquisition has not been conducted for a long period of time, and
reopen the exam when the exposure is ready.
After you click the buttons on the right side of the Exam & Exposure Screen,
including “System”, “History” and “Exit”, they become invalid.
Select the image and click to access the Image Print Screen.
Since the image acquisition and thumbnail are unavailable in the traditional
radiography mode, you cannot access the Image Print Screen.
76
Patient Exam
Screen.
Since the image acquisition and thumbnail are unavailable in the traditional
radiography mode, you cannot access the Image View Screen.
77
Patient Exam
B C
You may also view and reprocess the acquired history images on the Image View Screen.
For the history images, the image view screen is shown as below.
78
Patient Exam
Note: The EI reflects the amount of exposure dosage received by the human tissue. The smaller
the value is, the lower the dosage will be; and vice versa.
79
Patient Exam
Besides displaying the image, the image display area also supports related operation on images,
such as:
• Left-click and drag the orientation mark
• Use the mouse wheel to zoom in/out the image
• Right-click the image to adjust the window/level
When cropping the image, you may cover the image outside the crop frame by a grey mask.
When the mouse pointer moves over the brim of cutting frame, it turns into a cursor to change
the rectangle area size.
80
Patient Exam
Click an orientation marker (e.g. ), and a mark (e.g. ) is placed on the image, as
shown below.
Click the mark on the image (e.g. ), and drag it to the desired position by pressing and
holding the left mouse button down. After you click , the mark is then integrated into
the image.
81
Patient Exam
After the click, the orientation marker is selected and shown in deep blue (such as ).
Re-click the marker to cancel the selection.
You may also self-define a mark by clicking the marker . A text box pops out for you to
enter mark or annotation. Click for confirmation, and the user-defined mark is
82
Patient Exam
ROI (Region
of Interested) Shows the window / level of ROI
window
Pan image
When the image is magnified for viewing, the screen can only show part of the image. Click
and the cursor changes to . You may move the image within the viewing area by
Crop image
Click and the cursor changes to . Left-click and drag the mouse to crop the image.
Besides, the crop frame can be adjusted at any status. Put the mouse over the edge or head of
83
Patient Exam
Note: The maximum boundary of the crop frame should not exceed the image area.
Zoom In/Out
In image viewing area, you may zoom in /out the image by moving the mouse’s roller ahead or
backward.
Note: When the true size icon is selected, the image will be shown in exact size and not
respond to the zoom operation. Once the icon is de-selected, the image can respond to the
zoom operation.
Click and drag the mouse on a region image to select an area to magnify.
Magnify image
Click and the cursor changes to . Left-click and drag the mouse on a region image
to select an area to magnify. After you release the left button, the magnifying glass is hidden.
84
Patient Exam
Click to view the true size of the image. No zoom operation is available in true size view.
Re-click to cancel the selection, and the image can respond to zoom operation.
Invert
Click to reverse light and dark areas of the selected image. The image is shown in default
mode to meet the physician’s practice of viewing the image.
Click to show ROI window. The window / level of ROI window are different from those of
the whole window, used for optimizing the regional image. The following is an instance of ROI
window application.
85
Patient Exam
Click to display the preset patient information. The figure below shows the content and
position (left top, left bottom, right top, right bottom) of the information.
86
Patient Exam
The display content and position of patient information can be configured. Refer to Section 8.7
for details.
87
Patient Exam
Right-click the mouse to adjust the window / level, the cursor will be changed to within the
image viewing area. The image brightness can be adjusted by moving the direction of the cursor.
88
Patient Exam
The measurement data is based on the image formed at the detector, and does
not represent the actual data of the X-rayed object.
Line measurement
This function allows the measurement of the actual distance between two points (on the
detector). Click and the cursor changes to . Left-click and drag the mouse to place a
line measurement annotation on the image.
After the annotation is successfully added, the cursor automatically switches to and can
select and activate the annotation. When the annotation is activated, you can move the
annotation by clicking and dragging the middle of the line; or adjust the length and direction of
the line by clicking and dragging the yellow square handles at the ends of the line; or remove it
by pressing the DELETE key on the keyboard. The following is an example of an activated line
measurement annotation.
89
Patient Exam
Note: The line measurement is in units of millimeter. If the line end is outside of the image area,
the measurement is invalid.
Angle measurement
This function allows the measurement of the angle on the image. Click and the cursor
changes to . Left-click and drag the mouse to place an angle measurement annotation on
the image.
After the annotation is successfully added, the cursor automatically switches to and can
select and activate the annotation. When the annotation is activated, you can move the
annotation by clicking and dragging any line of the angle; or adjust the point position and angle
by clicking and dragging the yellow square handles at the angle’s line; or remove it by pressing
the DELETE key on the keyboard. The following is an example of an activated angle
measurement annotation.
Note: The angle measurement is in units of degree. If any line end is outside of the image area,
the measurement is invalid.
90
Patient Exam
Profile measurement
This function allows the measurement of the profile on a line path. Click and the cursor
changes to . Press the left button of the mouse and drag it to place a profile measurement
annotation on the image. It will display the gray scale of the line segment between the starting
point and end point of the mouse. The vertical axis of the coordinate marks the max and min grey
value of the image pixel, as shown below.
After the annotation is successfully added, the cursor automatically switches to and can
select and activate the annotation. When the annotation is activated, you can view the grey scale
of a specific point by clicking and dragging the mouse, and the corresponding grey value is
shown on the horizontal axis of the coordinate. Moreover, you may remove the annotation by
pressing the DELETE key on the keyboard when it is selected and activated. The following is an
example of an activated profile measurement annotation.
Click and all the measurement annotations will be deleted from the image, excluding the
orientation marks.
2. Delete one
Click and a measurement annotation can be selected and activated. Press the DELETE
key on the keyboard to remove it. However, the orientation mark cannot be deleted in this way.
91
Patient Exam
Click on the Image View Screen and select the enhance mode from the down-pull
Click on the Image View Screen and a parameter regulating panel appears (as
shown below), which can make further adjustment on image post-processing parameters.
92
Patient Exam
After the adjustment, click on the top right side of the panel, and the regulated
effect is shown on the left image area. Click to save the adjusted settings, which
would be the default settings thereafter. This function is only open to the administrator.
Click and the parameters regulating panel disappears. The image would keep
The default settings of different views can be selected from the list to reprocess
the image.
The parameters regulating panel is different in the Enhance mode and the ESA mode. See the
following for detailed introduction.
93
Patient Exam
Simple mode:
It is a simplified enhance mode and contains four parameters: brightness, contrast, sharpness,
and de-noise. Normally, the values are “0.00” by default.
• Brightness: The adjustment range is -10 to 10. The smaller the value is, the darker the image
will be. The bigger the value is, the brighter the image will be.
• Contrast: the contrast range is -10 to 10. The smaller the value is, the weaker the image
contrast will be. The bigger the value is, the greater the image contrast will be.
• Sharpness: The adjustment range is -10 to 10. The smaller the value is, the weaker the
sharpness will be and there will be fewer details. The bigger the value is, the greater the image
sharpness will be and there will be more details.
• Noise reduction: The adjustment range is -10 to 10. The smaller the value is, the greater the
image noise will be; and vice versa.
Professional mode:
94
Patient Exam
95
Patient Exam
Control point
Input pixel
96
Patient Exam
Parameters adjustment:
• Abscissa of the control point (brightness): change the image brightness by adjusting the
brightness value. Increase the brightness value and the whole image (including bone tissue
and parenchyma) would be brightened; decrease the brightness value and the whole image
would be darkened.
• Ordinate of the control point (Y): change the image display effect of the parenchyma by
adjusting the Y value. Decrease the Y value and the image parenchyma would be brightened;
increase the Y value and the image parenchyma would be darkened.
Decrease the
Y value, and Increase the
the image Y value and
parenchyma the image
would be parenchyma
whitened would be
darkened.
• Slope of the control point (contrast grade): change the contrast grade of the whole image by
adjusting the contrast grade
97
Patient Exam
Increased
contrast grade
• Curve tail optimization parameter (tail): change the display effect of the parenchyma whose
grey value approximates the background and the adjustability of the parenchyma whose grey
value approximates the background by changing the shape of the curve tail. Compared with
the curve “tail parameter 1” and curve “tail parameter 2”, it can display more parenchyma
whose grey value approximates the background.
Dynamic compression
Compress the dynamic range of the image, so that the information displayed simultaneously
would be more abundant.
Parameters adjustment:
• Dynamic compression: Reducing dynamic compression can greatly reduce the whole image
brightness variation range, and shrink the apparent discrepancy between the black and white.
• Parameter Beta: mainly changing the displayable area of the high-density structures, thus
improving the displayed details of these areas.
• Compensation factor: used to control the correction degree of the grey value. The figure
below shows the trend of correction value varying with the grey value.
98
Patient Exam
Correction value
Grey value
Detail enhancement
Detail enhancement only changes the contrast degree of local details in the image, and does not
affect the display effect of the whole image.
The parameters range from 0.5 to 1.0.
99
Patient Exam
Noise restrain
100
Patient Exam
threshold value, the noise restrain impact of the pixel in surrounding area would be great on
the current pixel. Or else, the impact would be small. Thus the detail existing between the two
pixels is reserved.
• Sigma: Control the impact degree of pixels around the current pixel on noise restrain. At the
same distance to the current pixel, when the value is higher, the impact of surrounding pixel
on noise restrain is lower; and vice versa.
• Fine grain: Restrain fine grain noise. If the parameter is set “0”, noise restraining is maximized;
if the parameter is set “1.0”, noise restraining is minimized (actually no noise retraining); as
the parameter changes from 0 to 1 gradually, the noise restraining effect reduces accordingly.
• Middle-sized grain: Restrain middle-sized grain noise. If the parameter is set “0”, noise
restraining is maximized; if the parameter is set “1.0”, noise restraining is minimized (actually
no noise retraining); as the parameter changes from 0 to 1 gradually, the noise restraining
effect reduces accordingly.
• Large-sized grain: Restrain large-sized grain. If the parameter is set “0”, noise restraining is
maximized; if the parameter is set “1.0”, noise restraining is minimized (actually no noise
retraining); as the parameter changes from 0 to 1 gradually, the noise restraining effect
reduces accordingly.
Click . The image is reprocessed according to the system’s default settings, including
recovering the default image post-process parameters, initial image window/ level and
self-adapting window size, as well as inverted image display status when loading the image.
101
Patient Exam
Click , and you may conduct the following processes: cropping image, rasterizing
added mark on the image, and saving image.
Whether to accept or reject the image, it will then return to the Exam & Exposure screen.
Click and you may conduct the following processes: cropping image, rasterizing
added mark on the image, and saving image. Click the box in front of “Save image as…”, and
a checkmark √ appears inside, then the image is saved additionally.
Click to unsave the modification of the history image, and quit the Image View
Screen.
If the system is not authorized or the authorization has expired, you cannot
post-process images.
102
History Image
Press to access the History Image Screen, as shown below. The following functions can
be realized here:
• Searching and displaying patient information
• Editing patient information
• Deleting patient data
• Sending and printing patient image
• Patient data backup and recovery
103
History Image
Click to select the registration date for query from the drop-down list box.
If the “User-defined” option is selected for the Date, the specified range of time can be
set for query.
• Data source: the image data storage source, click to select the source from the
drop-down list box.
Local: only shows patient data stored locally
CD: only shows patient data stored at the CD
You may select a period and data source for query directly without entering any information.
2. Click and all matches will be displayed in the patient list, as shown below.
Click before the Patient ID item, and a sub list appears to show the patient’s exam bodypart,
exam date and exam status. Click the area outside of the sub list and the sub list disappears.
104
History Image
Press on the History Image Screen. The following edit box pops up.
The patient information can be checked and modified here. Click to save the edited
The original patient information and record cannot be recovered after the edit!
Select one or several patient records from the patient list, and then click . The
105
History Image
Click to delete the selected exam record and related images; click to
Only the administrator or the authorized user is entitled to delete the data.
1. First, select the patient record for protection from the list. Right-click the record. The
following box appears (as shown in the red circle).
2. Left-click the mouse and select “protection setting”, The record is then locked and
protected. To cancel the protection, just select “protection cancel”.
Click before the Patient ID item. A window appears for checking the data protection status.
In the figure below, the patient data is shown as “protected”. The data will not be deleted if the
button is pressed.
To cancel the protection over the patient data and images, select the record from the list, and
then right-click it. Select “unlock” and the protection status is shown as “unlocked”.
Press and the following window pops up. The selected image can be backed up or
106
History Image
The window displays the export directory, needed disc space and available disc space, etc.
• Backup type: DICOM, BMP, and JPEG are available formats. BMP and JPEG format images
need no backup and can be recorded directly. DICOM format images should be backed up
first and then recorded.
• Backup: only copy the images to designated directory. The backup result will show in the
screen.
• Burn(optional): startup the CD-writing program, and record the selected images in the list
(already backed up).
• Close: return to the History Image Screen.
If the burn function is not authorized or the authorization has expired, you
cannot post-process images.
Click and the data recovery window pops up. You may either select “Restore all
107
History Image
Press to directly import DICOM images to the local computer. Then, the following
window appears.
The data can be recovered only via DICOM images, and only the administrator
or the authorized user is entitled to data recovery.
108
History Image
Double-click an image in the preview area, and the screen is switched to the Image View Screen
for you to view and operate the image. Refer to Section 4.3 for detailed operation of Image View
Screen.
PACS transmission, you may click to send the selected patient data (several
patients may be selected from the list) to PACS, or click to send the selected
Click to send the selected patient data in the patient information list to the image
printing screen for printing, or click to send the selected previewed image(s) to the
109
History Image
The Query/Retrieve function screen is as shown below. Area A is the query condition area, area
B is the query result area and area C is the retrieve result area.
110
History Image
Query: The default examination period is the query date. Select the corresponding query
condition and then click . The matching patient records are displayed in area B. Click
to clear the query conditions and return to the default settings.
Retrieve: In the query result list, select one or several patient records and click . The
retrieved patient record(s) will be displayed in area C.
111
History Image
Click to exit and return to the history image screen. The patient and relevant images
retrieved from the server can be queried there.
112
Image Printing Management
To add new images for printing, you may switch between History Image Screen
and Image Printing Screen manually.
2. On the Exam & Exposure Screen, you may select the image at the preview area and then click
113
Image Printing Management
On the Image Printing Screen, you may lay out and print the required images, including selecting
the print node, print orientation, film size, image layout format, removing image /film, etc.
114
Image Printing Management
C
A
D
115
Image Printing Management
6.2 Zoom
6.3 Print
Click and the following print format appears for selection. The original default format
is single row and single column (as shown in the deep blue icon).
Different icons represent different print formats. Once selected, the icon will be shown in deep
blue. The image display area on the left screen will make corresponding regulation when the
116
Image Printing Management
117
Image Printing Management
Click to select the print orientation of the film: landscape or portrait (default orientation,
shown as ). Different icons represent different print orientations. Once selected, the icon
will be shown in deep blue, and the image display area on the left screen will be rearranged
according to the selected orientation. When the current film orientation is changed, the zoom
status of the images added to the film will be changed either, as shown below.
118
Image Printing Management
119
Image Printing Management
Click and a new blank film will be added to the left image display area as the end page.
The initial format of the new film is the current selected format.
At the print operation area, the current film can also be removed. Click and the
following prompt box pops up. Click and the current film (including the image on the
current film) is removed. The current screen will switch to the following film (if there is)
automatically.
Click at the print operation area to remove the selected image(s). The following
prompt box appears for confirmation. Press and the image(s) in the current window
will be deleted.
There are two ways of image removal: removing the images and window, re-arranging the page;
or removing the images only, without re-arranging the page. The image removal strategies can
be set at the “System Configuration”, as shown below.
120
Image Printing Management
If there are many images to be print, there would be several film pages. You may click
Click and the image of the current page is sent to the selected print node for
printing. Click and the images of all pages will be print out. The print number is the
currently set number.
Before printing, a prompt box would appear for confirmation.
6.4 Exit
Click to exit the Image Printing Screen, and return to the Exam & Exposure
121
Manual Image Stitching (Optional)
The stitching position can be adjusted manually in two ways: image translation and stitching point
selection.
1. Find out the region to be adjusted by viewing the image stitching result in Area B, and click the
left mouse button over the region until a blue box appears. Or, view the thumbnail image in Area
C to judge the approximate position of the stitching seam. Area A1 and A2 then immediately
display the upper and lower images corresponding to the selected stitching seam and their
stitching relationship.
2. Click the left mouse button and drag the cursor to translate the image in Area A2 to adjust the
stitching positions of images A1 and A2. The adjustment effect can be viewed in Area B.
Release the mouse after the adjustment.
3. To adjust the stitching relationship of the other two adjacent images, repeat steps 1 to 3.
4. After the stitching adjustment of all images, click to save the image.
1. Find out the region to be adjusted by viewing the image stitching result in Area B, and left-click
the mouse over the region until a blue box appears. Or, view the thumbnail image in Area C to
judge the approximate position of the stitching seam. Area A1 and A2 then immediately display
the upper and lower images corresponding to the selected stitching seam and their stitching
relationship.
3. Click the left mouse button over the images of Area A1 and A2 to select the stitching points. The
selected stitching point will be shown in a blue cross.
123
Manual Image Stitching (Optional)
4. Click and the system combines images in Area A1 and A2 by overlapping the
stitching points. The adjustment effect can be viewed in Area B. If the stitching effect is
124
Manual Image Stitching (Optional)
5. To adjust the stitching relationship of the other two adjacent images, repeat steps 1 to 4.
125
System Configuration
Click the icon on the right side of the screen and then select the item .A
prompt box appears for confirmation. Press to access the System Configuration
User identification is required for accessing the System Configuration Screen. Different levels of
users have different configuration authorities.
The administrator is entitled to configure the following on the System Configuration Screen:
DICOM service, printer, exam bodyparts, views, exposure preference, image process schemes,
patient information display, patient information printing, log system, user authority, software, and
backup management, network configuration and license information, etc.
General users have no above authorities and can only change the password on the User
Management Configuration screen.
127
System Configuration
On this page, you can configure AETitle, port and specific character set of DICOM local service.
128
System Configuration
On this page, you can configure DICOM Worklist node parameters and test connection.
• Enable worklist: once selected, the Worklist function will be enabled.
• AE Title: AE title for PACS node
• Port: port No. of the service provided by the DICOM node, value range: 0 to 65535
• IP address: configure the IP address of the DICOM node; four segments of numbers (0 to
255)
• Verify: test whether each DICOM node is connected.
129
System Configuration
Set Mapping
Click to enter the worklist mapping screen.
• Worklist Code mapping with local setting: Tick this item to enable mapping, or otherwise,
disable it.
• DICOM element of Worklist Code: Optional items include “Requested Procedure
ID”, ”Requested Procedure Description”, ”Requested Sequence Code Value”, “Requested
Sequence Code Meaning”, “Scheduled Procedure Step ID”, “Scheduled Procedure Step
Description”, “Scheduled Sequence Code Value” and “Scheduled Sequence Code Meaning”.
• Apply: Click it to enable the configuration.
• Add Worklist Code: Enter the information from the worklist end (e.g. “Chest PA”)
130
System Configuration
Select the corresponding sub-bodypart or view and click OK to make configurations. If incorrect
item is selected, click Cancel to reselect.
Finally, click OK and the configured item appears.
131
System Configuration
On this page, you can configure DICOM storage node parameters and test connection.
• Enable storage: once selected, the storage function is enabled.
• Nodes List: add and configure several storage nodes.
• Default node: use one or more nodes in the list as default nodes. Image will be transmitted
based on the default node(s) in DROC.
• AE Title: AE title for DICOM node
• Port: port No. of the service provided by the DICOM node
• IP address: configure the IP address of the DICOM node; four segments of numbers (0 to
255)
• Compensating curve: select the curve compensation file to compensate the curve of the
image sent to PACS
box and then click . The node will appear in the default node column.
Click and the following input box pops up. Enter the AE Title, port, IP address
and compensating curve of the node, and then click . The added node will appear in
132
System Configuration
the storage node list box, and the node attribute frame will display the attribute of the added
node.
Click to delete the node selected at the storage node list box.
133
System Configuration
On this page, you can configure DICOM print node parameters and test connection.
• Enable print: once selected, the print function will be enabled.
• Print Status: add and configure several print nodes.
• Default node: use one node as a default print node.
• AE Title: AE title for DICOM node
• Port: port No. of the service provided by the DICOM node
• IP address: configure the IP address of the DICOM node; four segments of numbers (0 to
255)
• Compensating curve: select the curve compensation file to compensate the curve of the
image to be printed
• Print Brightness: select the print brightness to adjust the print effect. Optional range is -100
to 100.
click , and the node will appear in the default node column.
To add or delete the node, refer to the operations below:
Click and the following input box pops up. Enter the AE Title, port, IP address
134
System Configuration
and compensating curve of the node, and then click to enter the printer attribute
setting page (The related printer setting may refer to Section 8.2). On the printer attribute setting
page, click and the added node will appear in the print node list box. The node
135
System Configuration
Click to delete the node selected at the storage node list box.
136
System Configuration
On this page, you can configure DICOM MPPS node parameters and test connection.
• Enable MPPS: Select it to enable the MPPS service
• AE Title: AE title for DICOM node
• Port: port No. of the service provided by the DICOM node
• IP address: the IP address of the DICOM node; four segments of numbers (0 to 255)
If the DICOM Mpps function is not authorized or the authorization expires, the
DICOM Mpps function is disabled.
137
System Configuration
On this page, you can configure DICOM SC node parameters and test connection.
• Enable Storage Commit: Select it to enable the Storage Commit service.
• AE Title: AE title for DICOM node
• Port: port No. of the service provided by the DICOM node
• IP address: the IP address of the DICOM node; four segments of numbers (0 to 255)
138
System Configuration
On this page, you can configure DICOM Q/R node parameters and test connection.
• Enable Query/Retrieve: Select it to enable the Query/Retrieve service.
• AE Title: AE title for DICOM node
• Port: port No. of the service provided by the DICOM node
• IP address: IP address of the DICOM node; four segments of numbers (0 to 255)
• Equipment type: Select the type of acquired image: DX or CR
Node configuration and test:
Click to test whether each node is connected successfully. If yes, the following
If the DICOM Q/R function is not authorized or the authorization expires, the
DICOM Q/R function is disabled.
139
System Configuration
Print parameters:
• Select printer: selecting the print node from the drop-down list box
• Copy number: setting the copy number of each printing; value range: 0-65535, 1~99 are
suggested
• Border density: setting the density between film images: WHITE, BLACK, or entering the
number (1~ 65535)
• Empty image density: setting the density of the surrounding area: WHITE, BLACK, or
entering the number (1~ 65535)
• Min. density: selecting the minimum density
• Max. density: selecting the maximum density
• Config. information: entering letters or numbers according to the conformance setting of the
printer
• Owner ID: the owner ID appears on the DROC screen, usually set as Mindray by default; you
may also change the owner ID according to the customer’s requirement
Film parameters:
• Film session label: the default setting is Mindray DX; it can also be modified according to the
customer’s requirement; enter letters or numbers.
• Film destination: option: PROCESSOR, MAGAZINE, BIN_1, BIN_2, …BIN_8
140
System Configuration
• Film orientation: setting the default film orientation, option: PORTRAIT, LANDSCAPE
• Medium type: option: BLUE FILM, PAPER, CLEAR FILM, MAMMO CLEAR FILM, MAMMO
BLUE FILM; the default setting is BLUE FILM
• Magnification Type: option: REPLICATE, BILINEAR, CUBIC, NONE
• Trim: whether to set the print rim, option: YES, NO
• Image display format: setting the default image display format, option: STANDRAD1\1,
STANDRAD1\2, STANDRAD1\3, STANDRAD2\1, STANDRAD2\2, STANDRAD2\3,
STANDRAD3\1, STANDRAD3\2, STANDRAD3\3
• Image polarity: setting the default image polarity, option: NORMAL and REVERSE
• Priority: option: HIGH, MED, LOW; the default setting is MED
After configuration, click to save and quit the current screen. Click to
apply the configuration and continue the other configuration items.
141
System Configuration
Patient information will be displayed respectively in the right four columns according to the set
display orientation. The information in hidden status is displayed in gray for easy distinction.
142
System Configuration
Patient information will be displayed respectively in the right four columns according to the set
display orientation. The information in hidden status is displayed in gray for easy distinction.
143
System Configuration
drop-down list.
Export the log: Click . Select the saving path of the log to be exported, and then click
Note: The export log is a .zip file. The log contains the database, configuration file, and DROC
version information. In this system, the log cannot be exported to the disc at present.
144
System Configuration
Add a user:
Click under the user list and the following window pops up. The user name,
password, and authority grade can be set here.
Note: The user name should be numbers, letters or Chinese characters. The password is the
same.
145
System Configuration
Change password:
Select the user item from the user list, and then click under the list. The
Delete a user:
Select the user item from the user list, and then click under the list. The
following confirmation window is displayed.
Authorization Allocation
Through the “authorization table”, specific operation authorization can be allocated to the normal
user.
146
System Configuration
• Send to PACS when close: select it to enable the function, the image will be sent to PACS
after the exam is closed.
• Simulated exposure: select it to enable the function at the Exam & Exposure Screen
147
System Configuration
• Adjust the display sequence of items in the list: Select an item in the list and then click or
148
System Configuration
149
System Configuration
• Set icon zoom ratio: click or to increase or decrease the icon zoom ratio.
• W/L adjust modes: distance mode, and speed mode
In the two modes, the W/ L step can be adjusted:
W step: input range: 1~ 10 (integer)
L step: input range: 1~ 10 (integer)
• Save image as …: select the item
• Enable automatic labeling: select the item
• Show patient information: select the item
• Preview image after exposure: select the item
• Save pure image after exposure: select the item
150
System Configuration
• Three Image delete methods: reserve always; deleting by time; and deleting by free disk
space
-> Reserve always: disable the automatic image delete function; the image should be
deleted manually
-> Deleting by time: enable the automatic image delete function, the image can be
deleted according to the set time (image-saving days); days input range: 1~32767
->Deleting by free disk space: enable the automatic image delete function, the image
can be deleted according to the disk space; range: 1 ~ 20 (unit: GB)
151
System Configuration
Note: If the images are deleted according to disk space, the images of a period long ago will be
deleted first, until the disk space is enough for image saving. Besides, the images of the recent
two days will be saved all the way, even if the disk space is not enough for saving.
• Warning before deleting image: select it to enable automatic warning, as shown below.
152
System Configuration
• Refreshing synchronize with worklist: select it to make the refreshing of the register interface
list synchronous with that of worklist.
• Auto refresh: select it to refresh the patient list automatically
• Interval (s): setting the interval of auto refresh, range: 30 ~ 32767
• Auto delete patients not examed: select it to automatically delete the non-examination
records before the selected days.
• Age-Calculating Method: “By Actual Date and Time” or “By Difference of Years” is optional.
• Operator: select or add an operator to the list.
• Physician: select or add an operator to the list.
• List column items: select the items which are then displayed in the patient registration list.
• Adjust the display sequence of items in the list: Select an item in the list and then click or
153
System Configuration
• Frame-delete mode: option: delete image and frame, delete only image
• Automatically load image that is sent to print: select it to enable the function
• Display scale: select it to enable the function
• Set print font zoom ratio: set the printed font size.
154
System Configuration
155
System Configuration
Enable exposure compensation: When the option is selected, if the current SID and grid are
inconsistent with the SID and grid configured for the view, the DROC will compensate mA and
ms automatically through algorithm to reduce the impact of incorrect SID and grid on image
quality.
Restart detector when detector startup fails: When the option is selected, the DROC will
power off and restart the detector automatically if it cannot connect the detector during startup,
and then try connection again after detector restarting.
156
System Configuration
Institution information:
Enter the institution name to be displayed or printed out, only English letters, Chinese characters,
numbers and blank are allowed (all blanks are not allowed; the blank at the head or the end will
be omitted automatically).
PID:
• Generate method: For the logging into the local host, there are two methods to create a PID:
Manual: not to adopt the automatic-generate method; the Patient ID is entered
manually
Automatic: The system will generate the Patient ID automatically, and you may set the
prefix for the automatically generated PID (PID for instance).
• Prefix: Enter the prefix for the PID.
Accession number:
• Generate method: For the logging in the local host, there are two methods to create an
accession number:
Manual: not to adopt the automatic-generate method; the accession number is entered
manually
Auto: The system will generate the accession number automatically, and you may set
157
System Configuration
the prefix for the automatically generated accession number (AN for instance).
• Prefix: Enter the prefix for the accession number.
Note: The prefixes of PID and accession number should be numbers, letters or Chinese
characters.
Equipment information:
Configure the equipment model support by the DROC software
Date format:
Set the date format in local habits.
Modify the system time.
158
System Configuration
Click and enter the backup name in the dialog box as below.
After confirmation, the added backup name will be displayed in the backup list.
Note: Only English letters, Chinese characters, numbers and blank can be entered (all blanks
are not allowed; the blank at the head or the end will be omitted automatically)
Select a backup name from the backup list, and then click . The
backup corresponding to the name will be deleted, and the backup name will also be deleted
from the backup list.
Select a backup name from the backup list, and then click . The
159
System Configuration
system settings backup corresponding to the name will replace the current configuration.
Click and the default factory settings will replace the current
configuration. The default settings will not appear in the backup list, and cannot be deleted.
160
System Configuration
Connection name: Select the network connection name of the IP address to be modified from
the drop-down list.
Adapter Type: The adapter type is displayed after the connection name is selected.
IP address setup: Obtain an IP address automatically or set the IP address manually.
If is selected, the IP address, subnet mask and default gateway can be
modified.
161
System Configuration
This page displays the authorization state and authorized functions (including optional functions)
of the system and the expiry date of the authorization.
162
Parameter Manager
On the right side of the DROC screen, click and then select
from the menu. On the pop-up dialog box, click . The ParaManager screen is displayed.
On the ParaManager screen, you (as an administrator) can configure view definition, view
parameters, exposure parameters, post-processing parameters, as well as parameter backup and
recovery.
Note: The default homepage of the ParaManager is the View Definition page. After you click the
, , and tabs on the default homepage
separately, the ParaManager will load and display the corresponding page.
163
Parameter Manager
Expand all/ Hide all: To display all exam bodyparts, sub-bodyparts, and views, select Expand
all; to hide all bodyparts, sub-bodyparts, and views, select Hide all.
Add a bodypart: To add a new bodypart, select Add a bodypart. The following dialog box is
displayed. Enter the name of the new bodypart and then click OK. The newly added bodypart is
in the list of the Root node.
164
Parameter Manager
Delete a bodypart: To delete a bodypart, select Delete… from the menu and then click OK on
the confirmation dialog box as below.
Rename a bodypart: To rename a bodypart, select Rename… from the menu, enter a new
name for the bodypart on the dialog box as below and then click OK.
165
Parameter Manager
Expand all/ Hide all: To display all sub-bodyparts and views of the sub-bodypart, select Expand
all; to hide all sub-bodyparts and views, select Hide all.
Add a new sub-bodypart: To add a new sub-bodypart, select Add a sub-bodypart. The
following dialog box is displayed. Enter the name of the new sub-bodypart and then click OK.
The newly added sub-bodypart is in the list of the sub-bodypart.
Note: Please enter the name of the newly-added sub-bodypart in the textbox and do not leave it
blank. Do not use the name of an existing sub-bodypart for the newly-added sub-bodypart under the
same exam bodypart (bodypart category) or a different bodypart.
Select a sub-bodypart and right-click on it. An operation menu is displayed as below.
Expand all/ Hide all: To display all views of the selected sub-bodypart, select Expand all; to
hide all views of the sub-bodypart, select Hide all.
Delete a sub-bodypart: To delete a sub-bodypart, select Delete a sub-bodypart. A dialog box is
166
Parameter Manager
displayed for confirmation. Click OK. The selected sub-bodypart is deleted from the list of
sub-bodyparts.
Rename a sub-bodypart: To rename a sub-bodypart, enter the new name of the sub-bodypart in
the pop-up dialog box and then click OK.
Remove a sub-bodypart: To move away the current sub-bodypart, select Remove from the
current bodypart. A dialog box is displayed for confirmation. Click OK. The selected
sub-bodypart is deleted from the list of sub-bodyparts.
167
Parameter Manager
. All views under the sub-bodypart are displayed in the list. Click on the
desired view and drag it to the sub-bodypart under the Root tree on the left side.
Remove a view from a sub-bodypart: To remove a view from a sub-bodypart, right-click on the
view and then select Remove from the current bodypart from the operation menu.
168
Parameter Manager
Edit a view: Select Edit. A dialog box is displayed. You may edit the name and picture of the
view, DICOM bodypart, and view type.
Note: Removing a view is different from deleting a view. If a view is removed, the definition of the
view is still in the view list. If you display all views, you can find the removed view. The view can be
169
Parameter Manager
drawn to the sub-bodypart again. If a view is deleted, the definition of the view is also deleted. You
cannot find the deleted view in the view list.
170
Parameter Manager
Enter the Exposure Parameter page. Edit the exposure parameter as needed and then
Enter the Post-processing Parameter page. Edit the post-processing parameter as needed
171
Parameter Manager
Note: If you do not want to modify the parameters, just click Next each time. The system will
generate the view definition according to the default parameters.
After you add a similar view, a message box is displayed as below:
Note: The name and picture of the newly-added view cannot be null. The name of the view in the
same bodypart or a different bodypart cannot be used before.
172
Parameter Manager
Rename a view parameter: To rename a view parameter, select Rename from the menu. A
dialog box is displayed. Enter the new parameter name and then click OK.
173
Parameter Manager
Save as a new view parameter: To save the current view parameter as a new parameter, select
Save as…. A dialog box is displayed. Enter the new parameter name and then click OK.
Delete a view parameter: To delete a view parameter, select Delete. However, you cannot
delete the factory preset view parameter. A warning message is displayed if a factory preset
view parameter is to be deleted.
Set the default parameter: To set a selected view parameter as a default parameter, select Set
as default from the menu.
Note: If a view parameter is set as default, the view parameter is highlighted in blue.
Modify the view parameter value: After you select a view parameter, details of the parameter are
displayed on the right side. Modify the values of the parameter and then click
to save the modified value.
Note: You cannot modify the factory preset parameter.
174
Parameter Manager
2 PA posterior anterior
3 LL left lateral
4 RL right lateral
175
Parameter Manager
• FatClipSizeX/FatClipSizeY: set a default clip size for the fat patient of the view during image
preview; it is the same for other body sizes; the size will be sent to the IPC for adjustment of the
collimator light field
• HMirror: select or deselect horizontal flip
• Rotate: select the clockwise rotation degree, 0, 90, -90, or 180
• StitchLength: set the length (range) of image stitching
Note: You may use the mirror tool to flip the image horizontally or the rotate tool to rotate the image
clockwise for better image study and diagnosis. For instance, you may use the mirror tool to flip the
default image of a thorax examination. Normally, use the mirror tool first and then the rotate tool.
If you click each node on the left parameter tree root, the parameter list on the right side displays all
parameter values of the node. For instance, click Head->Head. The parameter list displays all
parameter values, parameter origin, and parameter version of all views under the head bodypart
176
Parameter Manager
On the Exposure Parameter panel, click a Root node, panel, bodypart, sub-bodypart, and then
right-click a view. An operation menu is displayed.
• Set the default dosage: After you set the default dosage, all default dosages of the corresponding
node are changed.
• Rename the dosage: After you rename the dosage, the names of all dosages of the
corresponding node are changed.
177
Parameter Manager
• Rename an exposure parameter: To rename an exposure parameter, select Rename. Enter the
new name of the parameter in the dialog box.
• Save as a new exposure parameter: To save the exposure parameter as a new one, select Save
as…..
• Delete an exposure parameter: To delete an exposure parameter, select Delete. However, you
cannot delete the factory preset exposure parameter. A warning message is displayed if a factory
preset exposure parameter is to be deleted.
• Set the default exposure parameter: To set an exposure parameter of the current dosage type as
the default parameter of the selected view, select Set as default.
• Edit an exposure parameter: After you select an exposure parameter, details of the parameter
For the view and body size (fat, medium, thin and pediatric), the exposure parameter contains
values of the exposure style, ion chamber, AEC density, focal spot, grid, kVp, mA, ms, and SID.
178
Parameter Manager
• Exposure style: AEC (auto mode), Time (manual mode), or mAs (semi-auto mode)
• Ion chamber: You can combine the three ion chambers at will. Each ion chamber can be set as
ON (1) or OFF (0). There are eight combination modes. The value “000” indicates that three ion
chambers are OFF. You may set a default combination here.
• AEC density: the degree of density (only available in AEC mode), value range: -4 to +4
• Focal spot: large or small
• Grid: 0.0, 1.0, 1.3, or 1.8
• kVP: 40-150
• mA: cannot be edited in semi-auto mode
• ms: auto (cannot be edited in semi-auto mode)
• mAs: cannot be edited in auto or manual mode
Note: The mAs value is not set directly, but mAs =mA × mS
• SID (cm): 70-200
• Cu Filter: 0.0, 0.1, 0.2, 0.3
179
Parameter Manager
On the Post-processing Parameter panel, click a Root node, panel, bodypart, sub-bodypart, and
then right-click a view. An operation menu is displayed.
• Set the default style: To set the selected style as a default style, select Set as default. Then, the
default styles of all views under the Root node, panel, bodypart, and sub-bodypart are changed.
• Rename the style: To rename the selected style, select Rename. Then, the names of all styles
under the Root node, panel, bodypart, and sub-bodypart are changed.
After you select a post-processing parameter, right-click the parameter. An operation menu is
180
Parameter Manager
displayed.
181
Parameter Manager
• Brightness: the abscissa of the inflection point, input range: 1~16380 (integer only)
• Y: the Y coordinate of the inflection point, input range: 1~16380 (integer only)
• Contrast: the slop of the inflection point, input range: 0 ~10 (floating point numbers 0.10~10.00
only, precise to the two digits after the radix point)
• Tail: curve tail optimization, input range: 0 ~0.1 (floating point numbers 0.00~0.10 only, precise to
the two digits after the radix point)
B. Compress:
• Compression: floating point numbers 0.01~1.00 only, precise to the two digits after the radix
point
• Beta: floating point numbers 0.00~1.00 only, precise to the two digits after the radix point
• Correction factor: floating point numbers 0.00~1.00 only, precise to the two digits after the radix
point
C. Enhance:
• size parameter: each parameter matches a detail size grade, used to control the detail
enhancement degree of the size grade; floating point numbers 0.00~1.00 only, precise to the two
digits after the radix point
• Enhance factor: enter the floating number from 0.00 to 3.50 only, precise to the two digits after
the radix point
D. Noise:
• Noise restrain level: the degree of noise restrain effect, option: 0, 1, 2, and 3
• Noise threshold: floating point numbers 1.00~300.00 only, precise to the two digits after the
radix point
• Sigma: floating point numbers 0.00~1.00 only, precise to the two digits after the radix point
182
Parameter Manager
• Fine grain: ranges from 0.00 to 1.00, precise to the two digits after the radix point
• Middle-sized grain: ranges from 0.00 to 1.00, precise to the two digits after the radix point
• Large-sized grain: ranges from 0.00 to 1.00, precise to the two digits after the radix point
• Brightness: -10 to 10, the lower the value is, the darker the image will be; and vice versa
• Contrast: -10 to 10, the lower the value is, the weaker the image contrast and image layering the
will be; and vice versa
• Sharpness: -10 to 10, the lower the value is, the smaller the image sharpness and details will be;
and vice versa
• Denoise: -10 to 10, the lower the value is, the stronger the image noise will be; and vice versa
183
Parameter Manager
• System backup: Enter the backup information in the dialog box. Then, you can back up the view
parameter, exposure parameter, post-processing parameter of the system. The newly added
backup will be displayed in the backup list.
• Backup management:
• Recover the parameter: On the parameter management window, select the backup name to be
184
Parameter Manager
To exit from the Parameter Manager screen and enter the DROC start screen, click Logout on the
menu bar.
185
System Menu
Click on the right side of the screen. The following function buttons are displayed. For
detector calibration, refer to Chapter 12; for system configuration, refer to Chapter 8.
10.2 Statistic
The workload of each user ID can be checked here. The total exam patients and exposure times
187
System Menu
Click on the right side of the screen, and then select . The statistic
screen appears for the counting under different search terms.
Click on the right side of the screen, and then select . Common users have
no access to the update function. The following prompt box will be displayed.
188
System Menu
Click to start updating. The DROC program is closed and the following window is
displayed.
Click on the right side of the screen, and then select . The DROC’s
software version and copyright can be checked here. Click any area on the screen to close the
window.
189
System Menu
Click and the detailed version information will be displayed in the box below after the
190
Report Module (Optional)
To enter the Report Screen, click the icon on the right side of the DROC main screen.
To switch to the DROC main screen, click on the right side menu directly.
For operation details, refer to relevant manual on the report module.
191
Detector Calibration
Note: Calibrate and maintain the detector regularly to keep performance of the detector at a high
and stable level.
Tool: No (Calibrate with Mindary algorithm by default)
Calibration steps:
1. Align the light field center of the tube-collimator assembly to the detector center. Set the SID
to 130 cm. Verify that no grid is inside the detector and no object is between the detector and
the tube assembly.
Note: For the portable detector, calibrate when it is not in the detector box.
a) Put the portable detector on a flat table. Do not cover the portable detector.
b) Move the tube-collimator assembly above the detector. Turn on the collimator and center
the light field crosshair with the detector cross line.
c) Draw out the tape rule of the collimator to the front cover of the detector until the
distance is 130 cm.
2. Click the DROC application icon to run the DROC. The DROC login page is displayed. Enter
the user name and password, and then click OK. The DROC Registration & Management
page is displayed.
Note: After you click OK, the calibration file loading page is displayed. It takes about 30
seconds to load the calibration files.
193
Detector Calibration
3. Click on the right side of the DROC screen and then click .
Detector Prompt
position for
exposure
Cal.
stages
Realtime
cal. info
Cal.
parameters
setting
Note: indicates the detector of the wall stand; indicates the detector of the
integrated floating patient table; and indicates that the detector is not in the detector
box. If the detector is selected, the icon is in deep blue. Click on the icon to switch the
placement direction of the detector.
5. The calibration progress includes Dark calibration and X-ray calibration. First, click
for Dark calibration. A calibration confirmation window pops up. Click Yes to start Dark
calibration. During Dark calibration, and turn into grey and are in
unusable status. When the buttons are recovered to normal status, it indicates that dark
calibration is completed. The detailed info window displays the whole process of calibration.
This information is also kept in the log file.
Note: You should perform Dark calibration first and then X-ray calibration.
6. After Dark calibration, click for X-ray calibration. A calibration confirmation window
pops up. Click Yes to start X-ray calibration. The button turns to grey.
7. Manually adjust the collimator light field to the maximum size (fully cover the detector).
8. When “X-ray shot” turns red ( ), press the Prep/ Expose button on the top of the
194
Detector Calibration
9. Repeat the above step until the button is not in grey. The message “… successful”
is displayed in the Detail Info. column. It indicates that the calibration is completed. Click
.
10. Click on the right side of the DROC screen to return to the Registration & Exam
Screen.
Note: The following prompts may be displayed in the information column during calibration:
1. The dose is too high: It indicates that the dosage is too high. Check whether the SID is
correct. If yes, decrease the kV or ms value.
2. The dose is too low: It indicates that the dosage is too low. Check whether the SID is correct.
If yes, increase the kV or ms value.
3. Detecting object existence failed, an object might be on the detector!: It indicates that
the detector is covered by an object. Check whether there is any object on the detector or a
grid is inserted in the detector.
195
Periodic Maintenance
In order to ensure continued safe performance of the system, a periodic maintenance program
must be established. It is the user’s responsibility to supply or arrange for this service.
There are two levels of maintenance, the first consists of tasks which are performed by the
user/operator, and the second are those tasks to be performed by qualified X-ray service
engineer.
To avoid degrade or damage in safety performance of the equipment, it is suggested to clean the
outer cover of the equipment every quarter.
The tasks of this periodic maintenance shall include the following items:
1) Switch off the equipment and cut off the 3-phase power supply to the equipment.
2) Externally, check the proper cable connections between each major component in the X-ray
system.
3) Clean the equipment and stained components periodically, particularly the components with
direct contacts with patients, such as the detector box, remote controller and the integrated
floating patient table. Clean outer covers and surfaces of the equipment by using a cloth
moistened in warm water with mild soap. Wipe with a cloth moistened in clean water. Do not
use cleaners or solvents of any kind; otherwise, they may darkle the paint or stain the print of
197
Periodic Maintenance
the cover.
Note: Wipe the remote controller with a dry cloth after cleaning.
4) Disinfect the detector box and detector box handle on the mobile stand (optional) periodically.
It is suggested to use common disinfectants, such as acetaldehyde and amphoteric solution.
5) Plastic and rubber components can only be cleaned by specialized glass cleaner,
dishwashing detergent, soapy water, or laundry detergent.
198
Technical Specifications
A-1
Technical Specifications
Image processing system: DROC (V1.0), network communication meets the DICOM
3.0 standard, achieving management over patients and images.
A-2
Technical Specifications
consumption
A-3
Technical Specifications
0.1, 0.125, 0.16, 0.2, 0.25, 0.32, 0.4, 0.5, 0.64, 0.8, 1.0,
1.25, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.4, 8.0, 10, 12.5, 16,
20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320,
400, 500, 640
Allowable deviation: ≤ ± (10%+0.2 mAs)
A-4
Technical Specifications
(Optional)
• Tabletop size (L×W):2120 mm×825 mm
• Height: 650 mm (±20 mm)
• Tabletop movement range: longitudinal 900 mm, lateral 260 mm
• Longitudinal traverse of the detector box: 580 mm
• The detector box can move together with the tube stand.
• Filter: ≤1.2 mmAl/ 100 kV
• Model: E7876X
Nominal tube voltage: 150 kV
Focal spot: 1.2 mm (large focus) / 0.6 mm (small focus)
Target angle: 12°
Tube assembly heat capacity: 975 kJ (1354 kHu)
• Model: H1086X
Nominal tube voltage: 150 kV
Focal spot: 1.2 mm (large focus) / 0.6 mm (small focus)
Target angle: 12°
Tube assembly heat capacity: 900 kJ (1250 kHu)
• Model: E7240X
Nominal tube voltage: 150 kV
Focal spot: 1.2 mm (large focus) / 0.6 mm (small focus)
Target angle: 12°
Tube assembly heat capacity: 900 kJ (1250 kHu)
Note: For specification details of these X-ray tube assemblies, see Appendix C.
A-5
Technical Specifications
Minimum radiation field: On the plane that is vertical to the reference axis and 100
cm away from the focal spot, the maximum radiation field can be selected. Its
length and width both cannot exceed 5 cm.
A-6
EMC Guidance & Manufacturer’s Declaration
B-1
EMC Guidance & Manufacturer’s Declaration
B-2
EMC Guidance & Manufacturer’s Declaration
TABLE 1
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
The DigiEye 280 Radiography System is intended for use in the electromagnetic
environment specified below. The customer of the user of the DigiEye 280 Radiography
System should assure that it is used in such an environment.
ELECTROMAGNETIC ENVIROMENT-
-
EMISSIONS TEST COMPLIANCE
GUIDANCE
The DigiEye 280 Radiography System uses RF
energy only for its internal function. Therefore, its
RF emissions
Group1 RF emissions are very low and are not likely to
CISPR 11
cause any interference in nearby electronic
equipment.
RF emissions
Class A
CISPR 11
Harmonic The DigiEye 280 Radiography System is suitable
Emissions Not applicable for use in all establishments other than domestic
IEC61000-3-2 and those directly connected to the public
Voltage low-voltage power supply network that supplies
Fluctuations / buildings used for domestic purposes.
Not applicable
Flicker Emissions
IEC61000-3-3
B-3
EMC Guidance & Manufacturer’s Declaration
TABLE 2
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The DigiEye 280 Radiography System is intended for use in the electromagnetic environment
specified below. The customer or the user of DigiEye 280 Radiography System should
assure that it is used in such an environment.
IMMUNITY IEC 60601 TEST COMPLIANCE ELECTROMAGNETIC
TEST LEVEL LEVEL ENVIRONMENT-GUIDANCE
Electrostatic Floors should be wood, concrete
Discharge(ESD) or ceramic tile. If floors are
±8 kV contact; ±8 kV contact;
IEC 61000-4-2 covered with synthetic material,
±15 kV air ±15kV air
the relative humidity should be at
least 30%.
Electrical fast ±2 kV for power ±2 kV for power The power supply quality should
Transient / burst supply lines; supply lines; be that of a typical commercial or
IEC 61000-4-4 ±1 kV for input / ±1 kV for input / hospital environment.
output lines output lines
Surge ±1 kV line (s) to ±1 kV line (s) to The power supply quality should
IEC 61000-4-5 line (s); line (s); be that of a typical commercial or
±2 kV line (s) to ±2 kV line (s) to hospital environment.
earth earth
Voltage dips, Mains power quality should be
Short that of a typical commercial or
interruptions hospital environment. If the user
and voltage of the MobiEye 700/700T
0 % UT; 1 cycle 0 % UT; 1 cycle
variation on MOBILE RADIOGRAPHY
70% UT for 25/30 70% UT for 25/30
power supply SYSTEM requires continued
cycle cycle
input voltage operation during power mains
at 0° at 0°
IEC 61000-4-11 interruptions, it is recommended
that the MobiEye 700/700T
0 % UT; 250/300 0 % UT; 250/300
MOBILE RADIOGRAPHY
cycle cycle
SYSTEM be powered from an
uninterruptible power supply or a
battery.
Power Power frequency magnetic fields
frequency should be at levels characteristic
(50/60 HZ) 30 A/m 30 A/m of a typical location in a typical
magnetic field commercial or hospital
IEC 61000-4-8 environment.
NOTE: UT is the A.C. power voltage prior to application of the test level.
B-4
EMC Guidance & Manufacturer’s Declaration
TABLE 3
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The DigiEye 280 Radiography System is intended for use in the electromagnetic environment
specified below. The customer or the user of DigiEye 280 Radiography System should assure
that it is used in such an environment.
IMMUNITY IEC 60601-1-2 COMPLIANCE ELECTROMAGNETIC
TEST TEST LEVEL LEVEL ENVIRONMENT-GUIDANCE
Portable and mobile RF
communications equipment should be
3 Vrms
3 Vrms used no closer to any part of the
0,15 MHz – 80
0,15 MHz – 80 system, including cables, than the
MHz
Conduced RF MHz recommended separation distance
6 Vrms in ISM
IEC 6 Vrms in ISM calculated from the equation applicable
bands
61000-4-6 bands between to the frequency of the transmitter.
between
0,15 MHz and Recommended separation distance
0,15 MHz and
80 MHz
80 MHz
d = 1.2 x P
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
B-5
EMC Guidance & Manufacturer’s Declaration
B-6
EMC Guidance & Manufacturer’s Declaration
TABLE 4
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATION DEVICE AND SYSTEM
The DigiEye 280 Radiography System is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the DigiEye 280
Radiography System can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters)
and the DigiEye 280 Radiography System as recommended below, according to the
maximum output power of the communications equipment.
Rated Maximum Separation Distance According to Frequency of Transmitter
Output power of m
Transmitter 150kHz -80MHz 80MHz-800MHz 800MHz-2.7GHz
W d=1.2√P d=1.2√P d=2.3√P
0.01 0.12 0.12 0.23
100 12 12 23
For transmitters at a maximum output power not listed above, the recommended separation
distanced in meters (m) can be determined using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
If system image distortion occurs, it may be necessary to position system further from sources
of conducted RF noise or to install external power source filter to minimize RF noise to an
acceptable level.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
B-7
EMC Guidance & Manufacturer’s Declaration
B-8
Specifications of X-ray Assembly
C-1
Technical Data TD
ROTANODE™
E7240X
E7240FX
General Data
Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage:
Radiographic ...................................................................................................................... 150 kV
Nominal Focal Spot Value (IEC60336:1993):
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s):
50 Hz 60 Hz
Large Focus ..................................................................................................... 29 kW 30 kW
Small Focus ..................................................................................................... 14 kW 15 kW
★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
Motor Ratings:
Stator: XS-AV
Starting Running
Driven Frequency [Hz] 50/60 50/60
Input Power [W] 1050 270 43
1) 3)
Voltage [V] 200 100 40
2)
Current [A] 6.0 3.0 1.2
4)
Min. Speed Up [s] 0.8 1.5 -
Capacitor [μF] 24 24 24
Note: 1) Applied voltage between common and main terminal.
2) Common current.
3) The every applied voltage must be never exceeded 110% of the above specification.
4) The speed-up time is allowed up to 110% of the above specification.
Anode Speed:
50 Hz ............................................................................................................ Minimum 2700 min-1
60 Hz ............................................................................................................ Minimum 3200 min-1
Stator Resistance:
Common-Main Winding ...................................................................................................... 27.5 Ω
Common-Auxiliary Winding ................................................................................................ 58.0 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 °C
Mode of Operation ........................................................................................................... Intermittent
Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 479 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Anode Angle ............................................................................................................... 12 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ................................................................. 1.3 mm Al / 75 kV IEC60522:1999
Radiation Protection (To meet the requirements of IEC60601-1-3):
Leakage Technique Factor .................................................................................. 150 kV, 3.4 mA
X-ray Coverage ................................................................................ 430 × 430 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 16 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number:
E7240X ............................................................................................................................. XH-121
E7240FX ........................................................................................................................... XH-126
-2-
E7240X / E7240FX
-3-
E7240X / E7240FX
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No Condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No Condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-4-
E7240X / E7240FX
-5-
E7240X / E7240FX
<5 % UT
(>95 % dip in UT )
for 5 sec
Power frequency 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of a
magnetic field typical location in a typical commercial
IEC61000-4-8 or hospital environment.
Note: UT is the A.C. mains voltage prior to application of the test level.
-6-
E7240X / E7240FX
10 Vrms [V2] V d = [
12
] P
150 kHz to 80 MHz V 2
in ISM bandsa
23
d = [ ] P 800 MHz to 2.5 GHz
E1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,c should be less than the
compliance level in each frequency ranged.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
-7-
E7240X / E7240FX
Note: 1 At 80 MHz and 800 MHz, the higher frequency range applies.
2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, object and people.
a.
The ISM (industrial, scientific and medical) bands between 150kHz and 80MHz are 6,765MHz
to 6,795MHz; 13,553MHz to 13,567MHz; 26,957MHz to 27,283MHz; and 40.66MHz to
40.70MHz.
b.
The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the
frequency range 80MHz to 2.5GHz are intended to decrease the likelihood that mobile/portable
communications equipment could cause interference if it is inadvertently brought into patient
areas. For this reason, an additional factor of 10/3 is used in calculating the recommended
separation distance for transmitters in these frequency ranges.
c.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the [EQUIPMENT or SYSTEM] is used
exceeds the applicable RF compliance level above, the [EQUIPMENT or SYSTEM] should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the [EQUIPMENT or SYSTEM].
d.
Over the frequency range 150kHz to 80MHz, field strengths should be less [V1] V/m.
-8-
E7240X / E7240FX
7
d = [ ] P 800 MHz to 2.5 GHz
E1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the
compliance level in each frequency rangeb.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
-9-
E7240X / E7240FX
Warning
1. Housing Rupture
Never input over rated power to X-ray tube assembly.
If the input power exceeds tube specification, it causes the over temperature of anode, insert tube
glass shatter and ultimately the following serious problems due to generating over-pressure by oil
vaporization inside housing assembly.
In such a critical condition causing housing rupture by over load, the safety thermal switch can not
protect X-ray tube even if it works.
* Housing sealing parts rupture.
* Human injury including burns due to hot oil escape.
* Fire accident due to flaming anode target.
The X-ray generator should have a protective function which manages input power to be within
tube specification.
2. Electric Shock
To avoid the risk of electric shock, this equipment must only be connected to a supply with
protective earth.
- 10 -
E7240X / E7240FX
Cautions
1. Over Rating
X-ray tube assembly can be broken with applying just one over rated shot.
Please read the technical data sheets carefully and follow the instructions.
2. Permanent Filtration
The total filtration and the distance between X-ray focal spot and human body are regulated legally.
They should be complied with the regulation.
4. Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or
failure. If the serious problems caused by the above risk is expected, you are requested to have a
contingency plan to avoid such a case.
5. New Application
If you use the product with new application not to be mentioned in this specification or with
different type of X-ray generator, please contact to us for confirming its availability.
- 11 -
E7240X / E7240FX
1. Qualified Persons
Only qualified persons who have technical training and professional knowledge can handle X-ray
tube assembly.
2. Fragile Glass
X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or
pulsed shock over 19.6m/s2 (2G).
Careful handling is required to treat or transport.
3. Ground Terminal
X-ray tube assembly has ground terminal. Ground cable should be connected.
4. High Voltage
All X-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the
high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled and high
voltage capacitors/cables discharged.
6. Rotating Anode
(1) Test rotation of the anode with the large focus filament at the preheat condition. At this time, do
not apply the tube voltage.
(2) When the hand switch "READY" button is pressed, the filament will become brighter and the
anode target will start rotating. The anode for normal speed will reach the predetermined
number of revolution in 0.4 to 1 sec., and it will continue rotation for some time after the stator
drive input is cut off.
The anode for high speed shall be automatically braked to reduced speed and continue
coasting after the drive input to the stator is cut off.
- 12 -
E7240X / E7240FX
(3) Ensure that the anode target rotates in the proper direction.
X-ray tube should always rotate counter-clockwise, viewing from the cathode side.
(4) If the inside of any tube unit cannot be observed (for example, the tube unit with a beryllium
window, etc.), check the tube unit to ensure again that the terminal numbers of the drive circuit
of the low-voltage terminals are properly connected.
7. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.
8. Protective Cover
X-ray tube assembly is not allowed to be used without the protective cover attached.
9. Handling
Appropriate jig or tools are required for tube installation to avoid physical damages.
- 13 -
E7240X / E7240FX
Function:
When housing inner pressure reaches to the working pressure of the Rupture Disc, the Rupture
Disc collapses and vaporized oil and hot oil flow through the collapsed Rupture Disc.
The external cover can protect the initial impact of oil flow, however, it can not protect oil drop or oil
leakage.
Caution:
1. Never touch the Rupture Disc or remove the label and the screws.
Otherwise, it could cause collapse of Rupture Disc.
2. Please note the Rupture Disc is not the device to protect X-ray tube from destruction when over
rated power input but just to reduce the impact of housing rupture.
- 14 -
E7240X / E7240FX
Caution in Operation
1. X-ray Radiation
X-ray tube assembly should have the beam limiting equipment mounted on the X-ray port to
protect unnecessary radiation.
2. Dielectric Oil
X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for
human health, if it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.
3. Operation Atmosphere
X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas.
4. Disposal
X-ray tube housing is lined with lead to protect unnecessary radiation.
The lead powder or vapor is harmful for human health.
Dispose the scrapped products according to the requirement of local regulation.
7. Operating Limits
Prior to usage, please confirm the environmental condition is within the operating limits.
8. Any Malfunction
Please contact to TETD immediately, if any malfunction of the X-ray tube assembly is noticed.
- 15 -
E7240X / E7240FX
Caution Labels
These labels are caution label to notify the user of the following point.
2. "Housing end caps must always be mounted to the tube housing assembly properly and correctly
for X-ray protection and safety."
- 16 -
E7240X / E7240FX
CE MARKING OF CONFORMITY
SERIAL NUMBER
MANUFACTURER
DATE OF MANUFACTURE
PROTECTIVE EARTH
EARTH
- 17 -
E7240X / E7240FX
In order to use "ROTANODE" for a long time, the seasoning before every day's start, adjust the tube
current and the cooling after every day's finish are recommendable.
Seasoning
1. When "ROTANODE" is used first after the installation or after the pause more than two weeks,
recommend the seasoning of "ROTANODE". Refer to the seasoning schedule. Execute the
seasoning when "ROTANODE" is unstable.
3. If the tube current is unstable on the way to raise a tube voltage, lower the tube voltage until
disappearing the instability, and raise the tube voltage slower than before.
4. The seasoning is carried out by service engineers who has sufficient technical and safe
knowledge.
110
TUBE VOLTAGE
C C
90
70
50
2
0
0 4 8 12 16 20 24 28 32 36 TIME [min]
Continuous Load Short time load
Tube Current 1mA Half of permissible tube current
0.1s,one exposure / min
Recess
- 18 -
E7240X / E7240FX
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
60kV 50kV 40kV 70kV 60kV 50kV 40kV
500 200
70kV 80kV
80kV 90kV
400
90kV 150 100kV
100kV 110kV
50
100
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
70kV 60kV 50kV 40kV 80kV 70kV 60kV 50kV 40kV
500 200
90kV
80kV
100kV
400 90kV
150 110kV
TUBE CURRENT [mA]
100kV
110kV 125kV
300 125kV
150kV
150kV 100
200
50
100
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
- 19 -
E7240X / E7240FX
80kV
400 100kV 18
125~150kV
300 16
50kV
200 40kV 14
Ef
100 12
0 10
4.5 4.7 4.9 5.1 5.3 5.5 5.7 5.8
200 14
125~150kV
FILAMENT VOLTAGE [V]
150 12
TUBE CURRENT [mA]
100kV
Ef
100 10
80kV
60kV
50 50kV 8
40kV
0 6
4.0 4.2 4.4 4.6 4.8 5.0 5.2
- 20 -
E7240X / E7240FX
Thermal Characteristics
900
180W
800
COOLING
700
HEAT STORAGE [kJ]
600
500
400
300
200
100
HEATING
0
0 20 40 60 80 100 120 140 160 180 200
TIME [min]
100
COOLING
80
HEAT STORAGE [kJ]
60
40
60W
20
HEATING
0
0 1 2 3 4 5 6 7 8
TIME [min]
The heating curves are showing example of average
input power to anode in operation.
- 21 -
E7240X / E7240FX
φ140
ANODE
4-M6 8 DEEP
170
°
12
41.27
125.4
479
92
90
φ
‹
45
11.1
4-M5 8 DEEP
90
156
11.1
L S
CATHODE
TERMINAL CONNECTIONS
11.1
121
9 8 7 6 5 1 2 3 4 E
53
NC NC NC NC
81
C1 M A
φ69
ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and No.5 or 6
FOCAL SPOT in series circuit.
C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS B
-1.5mm≦A≦1.5mm
TERMINAL CONNECTIONS
-1.5mm≦B≦1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE
- 22 -
E7240X / E7240FX
72
ANODE
4-M6 8 DEEP
CENTRAL RAY AND REFERENCE AXIS φ152.4
11.1
170
12°
41.27
125.4
479
92
90
φ
°
45
90
156
4-M5 8 DEEP
11.1
L S
CATHODE
81
C
2-M63.5, P=1.27
11.1
9 8 7 6 5 1 2 3 4 E
53
81
NCNC NC NC
C1 M A
φ69
ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS B
-1.5mm≦A≦1.5mm
TERMINAL CONNECTIONS
-1.5mm≦B≦1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE
- 23 -
E7240X / E7240FX
EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402
For Sales & Technical Services, please contact the following representative:
・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.
Technical Data TD
ROTANODE™
E7876X
General Data
Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage (IEC60613:2010):
Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Nominal Focal Spot Value (IEC60336:2005):
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s):
50 Hz 60 Hz
Large Focus ..................................................................................................... 50 kW 54 kW
Small Focus ..................................................................................................... 20 kW 22 kW
Nominal Radiographic Anode Input Power (IEC60613:2010):
50 Hz 60 Hz
Large Focus ..................................................................................................... 43 kW 48 kW
Small Focus ..................................................................................................... 18 kW 21 kW
★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
Motor Ratings:
Stator: XS-RA
Starting Running
Driven Frequency [Hz] 50/60 50/60
Input Power [W] 1450 80
1) 3)
Voltage [V] 240 58
2)
Current [A] 6.5 1.5
4)
Min. Speed Up [s] 0.6 -
Capacitor [µF] 24 24
Note 1) Applied voltage between common and main terminal.
2) Common current.
3) The every applied voltage must be never exceeded 110% of the above specification.
4) The speed-up time is allowed up to 110% of the above specification.
Anode Speed:
50 Hz ............................................................................................................ Minimum 2700 min-1
60 Hz ............................................................................................................ Minimum 3200 min-1
Stator Resistance:
Common-Main Winding ...................................................................................................... 27.5 Ω
Common-Auxiliary Winding ................................................................................................ 58.0 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 ℃
Mode of Operation ........................................................................................................... Intermittent
Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 479 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Anode Angle ............................................................................................................... 12 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Filtration:
Permanent Filtration ................................................................ 1.3 mm Al / 75 kV IEC60522:1999
Available Additional Filter combination (0.4 - 1.5 mm) ..................... Maximum 2.8 mm Al / 75 kV
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor .................................................................................. 150 kV, 3.4 mA
X-ray Coverage ................................................................................ 430 × 430 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 16 kg
High Voltage Receptacle ........................... To meet requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Housing:
Model Number .................................................................................................................. XH-121
-2-
E7876X
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-3-
E7876X
As the result of reviewing the test requirements in the following tables, an X-ray tube assembly
as a component of X-ray system or apparatus does not fall within the scope of the EMC directive of
IEC60601-1-2: 2007. Therefore, X-ray system manufacturer shall conduct the required test and
evaluate EMC.
CAUTION: The diagnostic X-ray system can only allow to use the designated cables and
components. If the different cables or components are used, they may deteriorate
the performance of electromagnetic emissions and immunity.
The following tables are the copies of statements in IEC60601-1-2: 2007.
-4-
E7876X
<5 % UT
(>95 % dip in UT )
for 5 sec
Power frequency 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of a
magnetic field typical location in a typical
IEC61000-4-8 commercial or hospital environment.
Note: UT is the A.C. mains voltage prior to application of the test level.
-5-
E7876X
10 Vrms [V2] V d = [
12
] P
150 kHz to 80 MHz V 2
in ISM bandsa
23
d = [ ] P 800 MHz to 2.5 GHz
E1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,c should be less than the
compliance level in each frequency ranged.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
-6-
E7876X
Note: 1 At 80 MHz and 800 MHz, the higher frequency range applies.
2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, object and people.
a.
The ISM (industrial, scientific and medical) bands between 150kHz and 80MHz are 6,765MHz
to 6,795MHz; 13,553MHz to 13,567MHz; 26,957MHz to 27,283MHz; and 40.66MHz to
40.70MHz.
b.
The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the
frequency range 80MHz to 2.5GHz are intended to decrease the likelihood that mobile/portable
communications equipment could
cause interference if it is inadvertently brought into patient areas. For this reason, an additional
factor of 10/3 is used in calculating the recommended separation distance for transmitters in
these frequency ranges.
c.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the [EQUIPMENT or SYSTEM] is used
exceeds the applicable RF compliance level above, the [EQUIPMENT or SYSTEM] should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the [EQUIPMENT or SYSTEM].
d.
Over the frequency range 150kHz to 80MHz, field strengths should be less [V1] V/m.
-7-
E7876X
-8-
E7876X
Warning
2. Electric Shock
To avoid the risk of electric shock, this equipment must only be connected to a supply with
protective earth.
-9-
E7876X
Cautions
2. Permanent Filtration
The total filtration and the distance between X-ray focal spot and human body are regulated legally.
They should be complied with the regulation.
4. Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or
failure. If the serious problems caused by the above risk is expected, you are requested to have a
contingency plan to avoid such a case.
5. New Application
If you use the product with new application not to be mentioned in this specification or with
different type of X-ray generator, please contact to us for confirming its availability.
- 10 -
E7876X
2. Fragile Glass
X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or
pulsed shock over 19.6m/s2 (2G).
Careful handling is required to treat or transport.
3. Ground Terminal
X-ray tube assembly has ground terminal. Ground cable should be connected.
4. High Voltage
All X-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the
high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled and high
voltage capacitors/cables discharged.
6. Rotating Anode
(1) Test rotation of the anode with the large focus filament at the preheat condition. At this time, do
not apply the tube voltage.
(2) When the hand switch "READY" button is pressed, the filament will become brighter and the
anode target will start rotating. The anode for normal speed will reach the predetermined
number of revolution in 0.4 to 1 sec., and it will continue rotation for some time after the stator
drive input is cut off.
The anode for high speed shall be automatically braked to reduced speed and continue
coasting after the drive input to the stator is cut off.
- 11 -
E7876X
(3) Ensure that the anode target rotates in the proper direction.
X-ray tube should always rotate counter-clockwise, viewing from the cathode side.
(4) If the inside of any tube unit cannot be observed (for example, the tube unit with a beryllium
window, etc.), check the tube unit to ensure again that the terminal numbers of the drive circuit
of the low-voltage terminals are properly connected.
7. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.
8. Protective Cover
X-ray tube assembly is not allowed to be used without the protective cover attached.
9. Handling
Appropriate jig or tools are required for tube installation to avoid physical damages.
- 12 -
E7876X
Function:
When housing inner pressure reaches to the working pressure of the Rupture Disc, the Rupture
Disc collapses and vaporized oil and hot oil flow through the collapsed Rupture Disc.
The external cover can protect the initial impact of oil flow, however, it can not protect oil drop or oil
leakage.
Caution:
1. Never touch the Rupture Disc or remove the label and the screws.
Otherwise, it could cause collapse of Rupture Disc.
2. Please note the Rupture Disc is not the device to protect X-ray tube from destruction when over
rated power input but just to reduce the impact of housing rupture.
- 13 -
E7876X
Caution in Operation
1. X-ray Radiation
X-ray tube assembly should have the beam limiting equipment mounted on the X-ray port to
protect unnecessary radiation.
2. Dielectric Oil
X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for
human health, if it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.
3. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.
4. Disposal
X-ray tube housing is lined with lead to protect unnecessary radiation.
The lead powder or vapor is harmful for human health.
Dispose the scrapped products according to the requirement of local regulation.
7. Operating Limits
Prior to usage, please confirm the environmental condition is within the operating limits.
8. Any Malfunction
Please contact to TETD immediately, if any malfunction of the X-ray tube assembly is noticed.
Optional Parts
Following are additional Aluminium Filters as optional parts.
KITX-104 : A set of 0.4mm Aluminium Filters.
KITX-105 : A set of 0.5mm Aluminium Filters.
KITX-106 : A set of 0.6mm Aluminium Filters.
- 14 -
E7876X
Caution Labels
These labels are caution label to notify the user of the following point.
1. "Always have service personnel install or remove the high voltage cable."
2. "Housing end caps must always be mounted to the tube housing assembly properly and correctly
for X-ray protection and safety."
- 15 -
E7876X
CE MARKING OF CONFORMITY
SERIAL NUMBER
MANUFACTURER
DATE OF MANUFACTURE
PROTECTIVE EARTH
EARTH
CAUTION
- 16 -
E7876X
In order to use "ROTANODE" for a long time, the seasoning before every day's start, adjust the tube
current and the cooling after every day's finish are recommendable.
Seasoning
1. When "ROTANODE" is used first after the installation or after the pause more than two weeks,
recommend the seasoning of "ROTANODE". Refer to the seasoning schedule. Execute the
seasoning when "ROTANODE" is unstable.
3. If the tube current is unstable on the way to raise a tube voltage, lower the tube voltage until
disappearing the instability, and raise the tube voltage slower than before.
4. The seasoning is carried out by service engineers who has sufficient technical and safe
knowledge.
A,B
110
C C
90
70
50
2
0
0 4 8 12 16 20 24 28 32 36 TIME [min]
Continuous Load Short time load
Tube Current 1mA Half of permissible tube current
0.1s,one exposure / min
Recess
- 17 -
E7876X
500
110kV 200 110kV
125kV 125kV
400
150kV 150 150kV
300
100
200
50
100
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
Refer to IEC60613:2010
500
125kV 200 125kV
100
200
50
100
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
Refer to IEC60613:2010
- 18 -
E7876X
700 40kV 14
50kV
60kV
80kV
500 100kV 12
400 125-150kV 11
Ef
300 10
200 9
100 8
0 7
3.9 4.1 4.3 4.5 4.7 4.9 5.1 5.3 5.4
50kV
8
60kV
FILAMENT VOLTAGE [V]
300
TUBE CURRENT [mA]
80kV
100kV 7
200 125-150kV
Ef 40kV 6
100
5
0 4
3.9 4.1 4.3 4.5 4.7 4.9 5.1 5.2
- 19 -
E7876X
Thermal Characteristics
600
500
400
300
200
HEATING
100
0
0 20 40 60 80 100 120 140 160 180 200
TIME [min]
180
160
COOLING
140
HEAT STORAGE [kJ]
120
100
80
60
142W
40
20
HEATING
0
0 1 2 3 4 5 6 7 8 9 10
TIME [min]
- 20 -
E7876X
Dimensional Outline
Unit: mm
φ140
ANODE
4-M6 8 DEEP
170
12°
41.27
125.4
479
92
90
φ
°
45
11.1
4-M5 8 DEEP
90
156
11.1
L S
CATHODE
TERMINAL CONNECTIONS
11.1
121
9 8 7 6 5 1 2 3 4 E
53
NC NC NC NC
81
C1 M A
φ69
ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and No.5 or 6
FOCAL SPOT in series circuit.
C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS B
-1.5mm≦A≦1.5mm
TERMINAL CONNECTIONS
-1.5mm≦A≦1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR
: CENTRAL X-RAY
NC : NON-CONNECTION
ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL
: IEC60526 TYPE
- 21 -
E7876X
EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402
For Sales & Technical Services, please contact the following representative:
・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.
X-ray Tube Housing Assembly
H1086X‐0.6‐1.2
H1086Y‐0.6‐1.2
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
Description
The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective tube
X-ray tube assembly with double-focus rotating anode X-ray tube H1086 is intended to be used
for all routine diagnostic examinations with conventional or digital radiographic and fluoroscopy
workstations of OEM (Original Equipment Manufacturer).
H1 086 X
Housing one KL86tube High voltage socket with 90 degree direction
H1 086 Y
Housing one KL86 tube High voltage socket with 270 degree direction
Barometric pressure From 700 hPa to 1060hPa From 500 hPa to 1060 hPa
1-phase stator
Test point C-P C–S
Winding resistance ≈18.0 ... 22,0 Ω ≈45.0 ... 55,0 Ω
Max.permissible operating voltage(run-up) 230V±10%
Recommend operating voltage(run-up) 160V±10%
Braking voltage 70VDC
Run-on voltage in exposure 80 V rms
Run-on voltage in fluoroscopy 20V -40V rms
Run-up time(depending on starter system) 1.2s
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
Emission curves of the cathode
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
X-ray tube assembly dimensional drawings
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
Notes!!!
In the interest of complying with legal requirements concerning the environmental compatibility of
products (protection of natural resources, avoidance of waste) we endeavor to reuse components and to
return them to the production cycle. We guarantee the reliability and functionality, quality and life of these
components by taking extensive quality assurance measures including all test procedures, just as for brand
new components. Reused parts are “qualified-as- good-as-new”.
The requirements of IEC 62309 are fulfilled.
Kailong Medical Solution has performed a conformity assessment in accordance with the MDD for this
component.
The tube assembly is not a sterile product. The housing surfaces are designed for easy cleaning
disinfection (damp cloth).
Applications with contacts to animal or human tissue or in body cavities are not intended.
Use the original packaging for returning the tube assembly to the manufacturer.
With increased tube assembly load one or two fans can be used in addition. In single fan operation
the fan must be provided basically on the anode side. The fans are operated at an alternating voltage of 24
V (50/60 Hz) and a power of approx. 20 W. The installation and the connection of the fans are described in
the installation instructions for the fan. Fans must be cleaned once annually. Your customer service can do
this correctly.
Unless otherwise indicated, all load data specified apply to six-pulse, twelve-pulse, multi-pulse or DC
voltage generators.
Caution! Handle with Care.
X-ray Tube Housing Assembly
H1086X,H1086Y‐0.6‐1.2
Caution
◆ At high housing temperature take care to avoid contact when the protective cover is missing.
◆ Prior to an examination it is your responsibility as the user to ensure that all safety features are
functional and that the product is ready for operation.
Caution, operation of the tube assembly only with housing covers fitted!
◆ To ensure the safety of both the user and the patient, the customer service must be notified
immediately when a defect has been detected, and the operation of the tube assembly concerned
must be terminated.
◆ Collimators with a weight up to 300N may be fastened to the collimator flange of the X-ray
tube assembly.
P/N: 046-004982-00 (7.0)