82-P0801/E

Single channel meter for the


calibration of compression
machines

INSTRUCTION MANUAL
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 2
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

INDEX

1. GENERAL DESCRIPTION .................................................................................................................. page 3

2. MAIN TECHNICAL CHARACTERISTICS OF HARDWARE .............................................................. 3

3. MAIN FUNCTIONS ............................................................................................................................... 3

4. INSTALLATION AND INITIAL CONNECTIONS ................................................................................. 4

5. USE OF THE INSTRUMENT ............................................................................................................... 4

6. FORCE MEASUREMENT .................................................................................................................... 5

7. USE OF PRINTER DURING ACQUISITION OF READINGS ............................................................. 9

8. FILE MANAGEMENT ........................................................................................................................... 10

9. RS232 SERIAL PORT AND PRINTER PORT ..................................................................................... 12

10.CONNECTIONS ................................................................................................................................... 12

11.SOLUTION TO MOST COMMON PROBLEMS .................................................................................. 13


MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 3
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

1. GENERAL DESCRIPTION

Digimax Plus is a microprocessor measuring instrument with one channels, used to read and elaborate signals from one
or more load cells (not simultaneously) as reference loads for the calibration of compression machines (all the load cells
described in our general catalogue, section 82, can be used). Each load cell supplied by Controls with the 82-P0801/E is
complete with a calibration certificate which shows the relationship between the number of divisions shown on the display
and the applied load. The instrument can be interfaced with the dedicated software program 82-P0804/E2.
The instrument is housed in an ergonomically designed light alloy using with an anti-scratch polycarbonate front panel with
ten key touch keyboard and large mono-chromatic graphical display. The rear panel houses the parts for the connection
of load cell, remote sampling button, serial ports and main feed with anti noise filter, fuse and on/off switch.

2. MAIN TECHNICAL CHARACTERISTICS OF HARDWARE

- Multi-layer printed circuit board with state of the art technology and latest SMD components.
- Hitachi H8S 16 bit processor
- Battery operated RAM with internal clock/calendar which functions even when the instrument is switched off.
- Analogical channel for connection to load cell with extensometric bridge (2mV/V) with 262000 point resolution.
- Serial output 9600 baud for connection to serial printer
- Serial output 38 400 baud for connection to PC
- Mono-chromatic graphical display 240 x 128 pixels
- 10 key touch keyboard
- Remote button reading control samples

3. MAIN FUNCTIONS

The P0801/E can be used to acquire readings from any load extensometric load cells (strain gauges) or four bridge cells.
It has one input with very high resolution and allows the display of applied load in terms of divisions.
The function of the P0801/E is completely governed by its internal software organised in various menus. Interface with the
operator is made via a graphical display (240 x 128 pixels) and 10 key touch keyboard, of which four one interactive with
the specific functions of each screen. The main characteristics of the software are:
- language selection between Italian, English, French, German and Spanish.
- Clock/calendar with possibility to re-set time and date
- Input of operator name, type of test, description of load cell
- Mono-channel mode (force indicator): performs test with display of applied load in divisions.
- Hold function via remote control sampling button.
- Real time or deferred print out of readings, or real time serial communication with PC using program 82-P0804/E2 in remote
mode.
- Permanent saving of readings. Up to 50 force indication tests, each of which can contain up to 4 groups of 12 readings
for a total of 2400 readings. Each saved test also includes all its descriptive data.
- File management which display of saved tests, transmission to serial printer through dedicated port, transmission to PC
through RS232 port, delete of single test or complete memory re-set.
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 4
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

4. INSTALLATION AND INITIAL CONNECTIONS

4.1 Conveniently place the P0801/E near the compression machine to be calibrated so that the load cell can be connect
and the display easily read.

4.2 Connect the mains cable ensuring before hand that the mains supply corresponds to that indicated on the table
attached to the instrument.

4.3 Connect the load cell to the “Ch1” port.

4.4 Connect the remote control sampling button to the “Hold” port

4.5 Connect the printer cable (if printer used) to the “printer” port. Connect the PC serial cable (if PC used) to the RS232
port (see chapter 9 for use of these two ports).

4.6 When all the connection have been made correctly, switch on the unit using the on/off switch on the rear panel.

5. USE OF THE INSTRUMENT

5.1 When the unit is switched on the display will light up and show the “Controls” logo for a few seconds and then the main
menu.

Note the presence of four message on the right hand side of the display, near the keys F1, F2, F3 and F4 which describe
the functions of the four keys. These are:

F1 = ESCAPE this function is used to refuse the last input command or return to the previous menu. The same
function is achieved by pressing the ESC key.

F2 = DOWN moves the cursor down by one position

F3 = UP moves the cursor up by one position

F4 = ENTER this function is used to accept the last input command or confirm a selection. The same function
is achieved by pressing the “ENT” key.
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 5
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

The other elements that form this screen are:

1) State of serial communication. This normally shows “----------“. During remote function mode (see below) it gives
information on the connection of the PC with specific messages.

2) State of back-up battery. The P0801/E is fitted with a re-chargeable battery which is used to supply the memory
of the instrument when the unit is switched off, thus avoiding the loss of saved data and settings. The fully
charged battery has a voltage of approx. 4.2 to 4.3 V and in this state can maintain stored data for about 4 –
6 months.
Each time the instrument is switched on the battery is re-charged. It is quite normal that after a long period of
non use the indicator is not at its maximum value; after a few minutes an increase in the charge level will be
noted. The time required to fully re-charge the battery (from completely flat) is approximately 24 hours.
We recommend that the unit is not left unused for a period over 4 – 6 months. If testing is not to be made,
periodically switch on the unit and leave it on for a few hours. The total discharge of the battery can cause the
loss of stored data and the set up configuration of the instrument.

3) Version of installed firmware

4) Date and time

5.2 Next select the work mode, selecting “remote” if the instrument is to be used via a PC with the 82-P0804/E2 program,
or selecting "Local" if the instrument is used by its keyboard.

USE OF THE INSTRUMENT IN REMOTE MODE

In this mode, the P0801/E acts as the interface between the load cell and a PC with program 82-P0804/E2. The connection
between the P0801/E and the PC is through a 9 pin serial cable supplied with the unit.
Connect the RS232 port of the P0804/E to the serial port (COM) of the PC. After selecting the type of test (measure force
or apply force) select “REMOTE”. This display will show the divisions read and the keyboard is inactive accept the “ESC”
key which allows return to the previous menu. An the other functions of the unit or disabled and this cannot be used, the
operative sequence, data acquisition, saving and elaboration of data are made via the PC.

Start up the software on the PC referring to the specific instruction manual. When the serial communication is made the
display of the P0801/E shows a message indicating the COM port being used. Eventual errors in serial communications will
be shown in the same field.

USE OF THE INSTRUMENT IN LOCAL MODE

In this case the P0801/E is autonomous and operated via its keyboard and the read data are shown directly on the display.

6. FORCE MEASUREMENT

When this menu is accessed the following screen appears:

The following sub-menus can be accessed:


- Options
- Test
- File
Make desired selection and confirm to pass to next menu
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 6
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

6.1 OPTIONS

This menu allows the input of data and time, activation of print and automatic saving of readings, language selection.

Setting of CLOCK: use the vertical arrows keys or to make selection and ENTER key to move from one field
to another and ESC to confirm the setting and return to the previous menu.
Activation of Auto-print and Auto-store function: operate the selection via the ENTER key and press the vertical keys
and to pass from one function to the other. Activated function is indicated by an “X” symbol next to the option.

If the Auto-print function is active, during the data acquisition, each time the remote sample button is pressed a print
out is made of the reading.
If Auto-store function is active, during the data acquisition, each time the remote sample button is pressed the read data
is saved in the memory.

Language selection: use the vertical arrows keys and to select between English, French, German, Spanish
or Italian.
Press ENT to confirm the selection.

6.2 TEST

6.2.1 The descriptive parameters are used if the test is to be saved and are saved together with the readings, they are:
- Operator name
- Type of the test
- Type of cell
- Notes

The operator name is a 10 character alphanumerical code. Use the vertical arrows keys and to set the

characters and the horizontal arrows keys and to pass from one character to another. All the numerical and
alphabet (upper and lower key) characters are available together with a series of symbols. Press “ENT” at the end of
the input and to press to the next selection.

Test type is a 10 character alphanumerical code. Use the vertical arrows keys and to set the characters and

the horizontal arrows keys and to pass from one character to another. All the numerical and alphabet (upper
and lower key) characters are available together with a series of symbols. Press “ENT” at the end of the input and to
press to the next selection.

Type of cell is a 10 character alphanumerical code used to identify the load cell used for the test. Use the vertical arrows
keys and to set the characters and the horizontal arrows keys and to pass from one character to
another. All the numerical and alphabet (upper and lower key) characters are available together with a series of
symbols. Press “ENT” at the end of the input and to press to the next selection.

Notes is an 18 character field for additional description. Use the vertical arrows keys and to set the characters
and the horizontal arrows keys and to pass from one character to another. All the numerical and alphabet (upper
and lower key) characters are available together with a series of symbols. Press “ENT” at the end of the input and to
press to the next selection.

These fields are only descriptive and as such have no influence on the acquired data.

To exit this menu press ESCAPE


MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 7
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

6.3 PERFORM TEST

Once the necessary inputs have been made, to start a test press the PERFORM TEST function. The P0801/E will show
the perform test screen. This is the main function of the instrument which shows the applied load in DIVISIONS, it allows
readings to be saved and printed. The operational sequence relative to performing the test is made within this menu.

6.4 PERFORMING TEST WITH DATA ACQUISITION

6.4.1 The menu is shown below and consists of:

1) Date
2) Time
3) Operator name
4) Test number
5) Type of test
6) Descriptive data of load cell
7) Notes
8) Number of reading made
9) Reading in divisions
10) Activation of print and save function
11) Zero command
12) Save command
13) New command
14) Print command
8
11
12
9
13
14
4
3
5
7 10
1 6
2

6.4.2 The instrument is now ready for the test, however some preliminary operations should be made to ensure good stability
and repeatability of results:
- Position the load cell between the compression machine platens. Refer to the compression machine instruction
manual for use of distance pieces etc.
- Check that the load cell is of a suitable capacity; refer to the calibration certificate issued by Controls to obtain
the conversion between the number of divisions shown on the display and the applied load in Engineering units.
It is important to note that the readings shown on the display are divisions.
- Check that the load cell is connected to the “load cell” port of the P0801/E and wait at least 15 minutes after
turning on the unit before preceding the test.
- After this period zero the reading, which initially will be several thousand divisions, with the zero key (F1).
- Load the load cell up to its maximum capacity using a medium load rate (of course not exceeding the maximum
capacity of the compression machine) and then unload to zero. Wait for approximately one minute and then zero
the readings again which could be several tens of divisions.
- Repeat this load/unload procedure for 4 to 6 times until the residual reading after unloading is just a few
divisions.

The unit is now ready to be used to check the calibration of the compression machine.
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 8
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

6.4.3 Before proceeding with other instructions it is important to give a brief explanation of the calibration verification
procedure and the way in which the P0801/E reads, saves and prints data:

- Calibration verification of compression machines consists in comparing the readings given by the compression
machine (digital or gauge) with the reference load cell. In this case the reference load cell is the one connected
to the P0801/E.
- The verification is made at a few load values, normally 5 (corresponding to 20%, 40%, 60%, 80% and 100%
of full scale) or more, repeating the readings three or four times. It is recommended that the reference standards
being used is consulted in this regard.
- The P0801/E allows the readings to be saved in files of groups each of which consists of a maximum of four
series of 12 readings each. It is possible to save up to 50 groups of readings for a total of 2400 individual
readings; each group is identified by the descriptive parameters input in the “Test Descriptive” menu, that is the
test number, operator name, type of test, load cell description, notes, date and time.
- Using the file management menu it is possible to access filed data so as to display them on the display, transmit
them to a PC via the RS232 port or to a printer via the printer port. It is also possible to print the data whilst the
readings are made.
- To make a reading the remote sampling button connect to the unit by a 3 m cable. This operation “freezes” the
reading on the display of the P0801/E whilst the button is kept pressed, allowing the operator to intervene on
the controls of the compression machine at the same time. If the Auto-print function is active, upon pressing the
remote sampling button, the reading is transmitted to the printer connected to the printer port. If the “Auto-store”
function is active the readings will be saved automatically. If both functions are active the reading will be both
printed and saved. If neither of the functions is active then the reading will only be shown the display but not
saved.
- To evaluate the relationship between the number of divisions and the corresponding load values, the calibration
certificate issued by Controls with each load cell must be consulted.

6.4.4 Once all the preliminary operations have been made it is possible to start the compression machine. Zero any readings
on the display before the load cell is subjected to any load.
IMPORTANT: never zero the readings once the load cell has been subject to loading.
Proceed at a moderate load rate, and when one approaches the first load to be verified further reduce the load rate
as much as possible. The slower the load rate at making readings, the more accurate will be the readings. To freeze
the reading on the display press the remote sampling button (or F2 or SAVE on the P0804/E panel). It will be noted
that the reading in divisions remains blocked for a few seconds after saving and, in any case, for all the time the remote
sampling button remains pressed.

6.4.5 Now the P0801/E has “Captured” the division value of the load to be verified. Remember that it is possible to record
up to 12 readings in each series, above this the message “SERIES COMPLETE” and excess readings are not saved.
If the series of readings is complete, press ENTER to pass to the next series. The display will up date the series number
by one.

6.4.6 Repeat The same procedure for all the successive load values to be checked. Remember that in each series it is
possible to make up to 12 readings, above this the message SERIES COMPLETE appears and excess readings are
not saved. If the series is complete and correct, press ENTER to confirm and pass to the next series. In this case the
display will update the series number by one.
If the series of readings is not correct or is to be ignored press ESCAPE and re-start from point 6.4.4.
In any case it is not necessary to make all 12 readings to pass to the next series; by pressing “NEW” confirmation is
requested to pass to the next series. Confirm with ENTER.

6.4.7 Unload the load cell to zero, wait at least 30 seconds so that the zero reading is stable. Now the unit is ready for the
next series of readings which will form part of the same group as the series of readings just finished.

6.4.8 Repeat the operations described in 6.4.4, 6.6.5 and 6.4.6 to pass to the third and, if necessary, fourth series of readings
as described in point 6.4.7
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 9
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

6.4.9 At any moment it is possible to display of the current group by pressing the right hand arrows key. An example is given
below:

Press the same key again to return to the previous display.

6.4.10 For printing instructions see chapter 7

6.4.11 At the end of the test, when all the readings have been made, press ESC to return to the previous menu. To confirm
press ENTER or ESC to remain in the same menu. To make a new test enter the “Test Description” menu, the test
number will increase by one. Follows the instructions from point 6.2 onwards.

7. USE OF PRINTER DURING ACQUISITION OF READINGS

7.1 The P0801/E has a printer port for the connection of a Serial printer such as Controls printer model 82-P0171. For
further details of this port see chapter 9.

7.2 During the test the acquired readings can be printed by pressing, when F4 PRINT key. A report is printed containing
all the readings of the current test along with a heating as shown below:
***********************
DIGIMAX P801/E
***********************
05/06/02 11:18:01
***********************
Serial No : 98765432
Operator : OPERATOR_1
Test no. : 00002
Test Type : COMPRESS.
Load Cell : E0100/FS
Note:__________________
***********************
Batch no. Sample
01 01 -00001
01 02 008384
01 03 016876
01 04 025353
01 05 032663
01 06 042355
01 07 050834
01 08 059315
02 01 -00003
02 02 008482
02 03 016963
02 04 025439
02 05 033982
02 06 042460
02 07 050939

We recommend that the print is made when the current series of readings is complete.

7.3 It may be more convenient to activate the automatic print function “Auto-print” which will automatically print each
reading as the remote sampling button is pressed. See chapter 6.1. “Options”.
IMPORTANT: when the Auto-print function is active the F4 print key is deactivated.
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 10
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

8. FILE MANAGEMENT

8.1 The P0801/E has a permanent memory, that is it can maintain data even when it is switched off. This data can be
accessed through the “File” menu presented in the main menu, allowing the display of filed data, and then to print it
through a Serial printer or download to a PC via the RS232 port. The memory contain up to 50 tests made in the “force
measurement” mode.

8.2 The screen of the File Management menu is described below:

1) Operator name
2) Type of test
3) Descriptive data of load cell
4) Notes
5) Test number
6) Test date
7) Time of saving file
8) Print command
9) Download to PC command
10) CANCEL command
11) Single test or complete memory selection command.

8
1 9
2 10
3 11

4 5

6 7

8.3 The logic of the keys in this menu is as follows:

- With the vertical arrows keys and it is possible to scroll through the filed tests. Each test is identified
by its test number followed by “C”.
- Using the right hand arrows the summary table of readings of the selected test is displayed. Use the left
hand arrows key to return to the previous screen.
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 11
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

8.4 Following the logic above it is possible:

- To display the filed tests


- Print the selected tests (that displayed) by pressing the PRINT F1 key. Check before hand that the printer is
connected to the printer port and that it is on line. See also chapter 7 for further instructions on the use of the
printer. Before is an example of a print out.

***********************
DIGIMAX P801/E
***********************
05/06/02 11:18:01
***********************
Serial No : 98765432
Operator : OPERATOR_1
Test no. : 00002
Test Type : COMPRESS.
Load Cell : E0100/FS
Note:__________________
***********************
Batch no. Sample
01 01 -00001
01 02 008384
01 03 016876
01 04 025353
01 05 032663
01 06 042355
01 07 050834
02 01 -00001
02 02 008485
02 03 016968
02 04 025440
02 05 033985
02 06 042462
02 07 050945
03 01 000001
03 02 008482
03 03 016963
03 04 025439
03 05 033982
03 06 042460
03 07 050939
04 01 000000
04 02 008480
04 03 016962
04 04 025438
04 05 033984
04 06 042461
04 07 050934

It is possible to print all the filed tests together. To do this press F4 so that the ALL message is displayed in reverse
characters and then press F1. Press F4 to return to single test print out.

It is possible to print all the filed tests together. To do this press F4 so that the ALL message is displayed in reverse
characters and then press F1. Press F4 to return to single test print out.

- Download to a PC via the RS232 port the displayed test or all the filed tests. Ensure that the PC is connected
(see chapter 9) and press F2 download to a PC. The transmitted data are identical to these obtained by printing.

- Delete the displayed test or all the filed tests. To delete the displayed test press F3 CANCEL and then confirm
with ENT or not with ESC. This operation will become necessary when the memory is full and free space must
be created. It is also possible to delete all the filed tests.
IMPORTANT: deleted tests cannot be recovered. To delete all the filed tests press F4 so that the ALL message
is displayed in reverse and then press F3. To return to the single test mode press F4.

Upon completion press ESC to exit the filed tests.


MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 12
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

9. RS232 SERIAL PORT AND PRINTER PORT

The P0801/E has two serial ports: the printer port for connection to a serial printer and an RS232 port for connection to a
PC. The printer port transmits with the following parameters: 9600 baud, 8 data bit, 1 stop bit, no parity. All transmitted fields
are ended with CR and LF. We recommend the use of Controls printer 82-P0171. The RS232 port transmits with the following
parameters: 38400 baud, 8 data bit, 1 stop bit, no parity. All the transmitted fields end with CR and LF. We recommend the
use of Controls software package “D-Terminal” (code 82-P0800/TRM) for data receival on the PC.

10.CONNECTIONS

The connection found on the rear of the instrument are:

Connector J1 - Printer output


Pin 2 = TX
Pin 3 = RX
Pin 5 = GND

Connector J2 - RS232 output for PC


Pin 2 = TX
Pin 3 = RX
Pin 5 = GND

Connector M1 - load sensor connection Channel1


Pin 1 = excitation + V (10 Vdc)
Pin 2 = Signal Output (3 mV/V max)
Pin 3 = GND – V excitation
Pin 4 = Single input
Pin 5 = Shield

Connector M3 - load sensor connection Channel 2

Pin 1 = excitation + V (10 Vdc)


Pin 2 = Signal Output (3 mV/V max)
Pin 3 = GND – V excitation
Pin 4 = Single input
Pin 5 = Shield

Connector M6 - Connections for Channel 3 and 4


NOT USED
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 13
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

11.SOLUTION TO MOST COMMON PROBLEMS

This chapter offers some advice for the identification and solution of problems that may be encountered.

Problem: The unit does not switch on


Possible solutions:
- Check that the mains cable is correctly connected.
- Check that the mains supply corresponds to that of the unit (see identification plate).
- Check the fuse next to the mains socket. Replace if necessary with 200mA 5x20 mm fuse.
- Check that the connector cables between the card and the front panel/display are in place. To do this remove the mains
cable, open the rear panel by removing the fixing screws so as to gain access to the card. Then check the connectors
marked A and B.
- Contact Controls Service Department.

Problem: The unit switches on but the display does not show anything
Possible solutions:
- The unit switches on but the display does not show anything.
- Check that the backlighting cable of the display is connected to the card. This cable is pink and must be connected to C
- Contact Controls Service Department.

Problem: The display works correctly but the keyboard does not
Possible solutions:
- Check that the connector cables between the card and the front panel/display are in place. To do this remove the mains
cable, open the rear panel by removing the fixing screws so as to gain access to the card. Then check the connectors
marked A and B.
- Contact Controls Service Department.

Problem: The load values shown on the display are always zero or full-scale values or are evidently unstable
Possible solutions:
- Check that the load cell cable is correctly connected
- Check that the connectors are not damaged or rusted
- Check that the load cell cable is not damaged, particularly near its connectors
- Check that the load cell cable is connected to its connectors (i.e. the individual wires inside the connectors).
- Check that the load cell has not been damaged
- Contact Controls Service Department

Problem: The load values displayed are not coherent


Possible solutions:
- Check the calibration of the unit
- Contact Controls Service Department

Problem: The printer connected to the printer port does not print
Possible solutions:
- Check the connection between the printer port of the P0801/E and the serial port of the printer
- Check that the correct cable (supplied by Controls) is being used
- Check that the cable is not damaged, particularly near the connectors
- Check that the printer is switched on
- Check that the printer is on line
- Contact Controls Service Department

Problem: The printer connected to the Printer Port does not print correctly
Possible solutions:
- Check that the cable is not damaged, particularly near the connectors
- Check that the printer is set correctly: 9600 baud. See the instruction manual of the printer
- Contact Controls Service Department
MANUALE D'ISTRUZIONI Single channel meter for the calibration of pag. 14
compression machines
Rev. 1
Mod. 82-P0801/E
INSTRUCTION MANUAL 6 feb 2006

Problem: The PC connected to the RS232 Port does not receive data
Possible solutions:
- Check the connection between the RS232 port of the P0801/E and the serial port of the PC.
- Check that the cable is not damaged, particularly near the connectors
- Check that the cable is a pin to pin type
- Check that the PC is set to receive data on the serial port to which the P0801/E is connected. See PC instruction manual
- Check the settings of the PC serial port being used is correct (see chapter 9)
- Try to use a different serial port of the PC
- Contact Controls Service Department

Problem: The PC connected to the RS232 Port receives data incorrectly


Possible solutions:
- Check that the cable is not damaged, particularly near the connectors
- Check that the cable is a pin to pin type
- Check that the PC is set to receive data on the serial port to which the P0801/E is connected. See PC instruction manual
- Check the settings of the PC serial port being used is correct (see chapter 9)
- Try to use a different serial port of the PC
- Contact Controls Service Department
MANUALE D'ISTRUZIONI
INSTRUCTION MANUAL
NOTE

Italy Poland United Kingdom


CONTROLS S.r.l. Controls Polska Sp z o.o. Controls Testing Equipment Ltd
6, Via Aosta ul. Minska 63 Controls House
I-20063 Cernusco s/N. (MI) Italy 03-808 Warszawa, POLAND Icknield Way
Tel. +39-0292184.1 Tel.: +48 22 33 08 100/101 Tring, Hertfordshire HP23 4JX - U.K.
Fax +39-0292103333 Fax: +48 22 33 08 102 Tel. +44-1442 828311
E-Mail: controls@controls.it E-Mail: controls#@controls.pl Fax +44-1442 828466
Web Site: www.controls.it Internet: www.controls.pl E-Mail: sales@controlstesting.co.uk
Web Site: www.controlstesting.co.uk
France
Controls S.à r.l.
Zone Artisanale
68130 Walheim, FRANCE
tel. +33-03 89 402655 United kingdom
fax +33-03 89 402645 Wykeham Farrance LTD
E-Mail: controls@controls.fr Spain Chiltern House
Equipos de Ensayo Controls S.a. Unit 4, knaves Beech Business Centre
Mexico Poligono industrial arboledas Loudwater - HIGH WYCOMBE
Equipos de Ensaye Controls S.a. De C.V. C/Sabina, Nave 5, Manzana D Bucks
Rio Churubusco N°. 27, Col. Portales 45200 - Illescas (Toledo), SPAIN HP10 9QY - United Kingdom
C.P. 03300, Mexico, D.F. Tel: +34-902 180843 (4 lineas) Tel. +44-1628 521000
Tel. +52-55 56726186 Fax: +34-902 180846 Fax +44-1628 530300
Fax +52-55 55393229 E-Mail: controls@eecontrols.es E-Mail: sales@wfi.co.uk
E-Mail: controlsmexico@mexis.com Web Site: www.eecontrols.es Web Site: www.wfi.co.uk

You might also like